Emerson Model Total Chlorine Analyzer with Model 56 Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 51-TCL-56/rev.A
May 2011
Total Chlorine Analyzer
with Model 56
Model TCL-56
ESSENTIAL INSTRUCTIONS
WARNING
READ THIS PAGE BEFORE PROCEEDING!
ELECTRICAL SHOCK HAZARD
Rosemount Analytical designs, manufactures, and tests its
products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain
them to ensure they continue to operate within their normal
specifications. The following instructions must be adhered
to and integrated into your safety program when installing,
using, and maintaining Rosemount Analytical products.
Failure to follow the proper instructions may cause any one
of the following situations to occur: Loss of life; personal
injury; property damage; damage to this instrument; and
warranty invalidation.
• Read all instructions prior to installing, operating, and
servicing the product. If this Instruction Manual is not the
correct manual, telephone 1-800-654-7768 and the
requested manual will be provided. Save this Instruction
Manual for future reference.
• If you do not understand any of the instructions, contact
your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on
and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation
Instructions of the appropriate Instruction Manual and
per applicable local and national codes. Connect all
products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to
install, operate, update, program, and maintain the
product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can
affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may
result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is
being performed by qualified persons, to prevent electrical shock and personal injury.
Making cable connections to and servicing
this instrument require access to shock hazard level voltages which can cause death or
serious injury.
Be sure to disconnect all hazardous voltage
before opening the enclosure.
Relay contacts made to separate power
sources must be disconnected before serving.
Electrical installation must be in accordance
with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable
national or local codes.
Unused cable conduit entries must be
securely sealed by non-flammable closures
to provide enclosure integrity in compliance
with personal safety and environmental protection requirements.
The unused conduit openings need to be
sealed with NEMA 4X or IP65 conduit plugs
to maintain the ingress protection rating
(IP65).
For safety and proper performance this
instrument must be connected to a properly
grounded three-wire power source.
Proper relay use and configuration is the
responsibility of the user.
No external connection to the instrument of
more than 69VDC or 43V peak allowed with
the exception of power and relay terminals.
Any violation will impair the safety protection
provided
Do not operate this instrument without front
cover secured. Refer installation, operation
and servicing to qualified personnel.
WARNING
This product is not intended for use in the light industrial, residential or commercial environment, per the
instrument’s certification to EN50081-2.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
DANGER
CAUTION
HAZARDOUS AREA INSTALLATION
Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel.
This device is not Intrinsically Safe or
Explosion Proof.
SENSOR/PROCESS APPLICATION COMPATIBILITY
Wetted materials may not be compatible with
process composition and operating conditions. Application compatibility is entirely the
responsibility of the user.
To secure and maintain an intrinsically safe
installation, the certified safety barrier, transmitter, and sensor combination must be
used. The installation system must comply
with the governing approval agency (FM,
CSA or BASEEFA/CENELEC) hazardous
area classification requirements. Consult
your analyzer/transmitter instruction manual
for details.
Proper installation, operation and servicing
of this device in a Hazardous Area Installation is entirely the responsibility of the user.
About This Document
This manual contains instructions for installation and operation of the Model TCL56
Total Chlorine Analyzer.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date
A
5/11
Notes
This is the initial release of the product manual. The manual
has been reformatted to reflect the Emerson documentation
style and updated to reflect any changes in the product offering.
MODEL TCL-56
TABLE OF CONTENTS
MODEL TCL-56
TOTAL CHLORINE ANALYZER
TABLE OF CONTENTS
Section
1.0
1.1
1.2
1.3
1.4
1.5
1.6
Title
DESCRIPTION AND SPECIFICATIONS.................................................................
Features ..................................................................................................................
Specifications — Sample Conditioning System ......................................................
Specifications — Model 56 Analyzer .......................................................................
Specifications — Model 499ACL-02 Sensor............................................................
Performance Specifications — Complete System .................................................
Ordering Information and Accessories ...................................................................
Page
1
1
3
3
4
4
5
2.0
PRINCIPLES OF OPERATION ..............................................................................
7
3.0
3.1
3.2
INSTALLATION ......................................................................................................
Unpacking and Inspection ......................................................................................
Installation................................................................................................................
8
8
8
4.0
4.1
4.2
4.3
4.4
4.5
WIRING ..................................................................................................................
Prepare Analyzer Conduit Openings .......................................................................
Provide Power to the Sample Conditioning System ................................................
Make Power, Alarm, Output, and Sensor Connections in the Analyzer ...................
Sensor Wiring ..........................................................................................................
Apply Power to the Analyzer and Complete Quick Start..........................................
13
13
13
14
15
16
5.0
5.1
5.2
5.3
5.4
START-UP ..............................................................................................................
Prepare the Reagent ...............................................................................................
Zero the Sensor .......................................................................................................
Start Sample Flow....................................................................................................
Begin Operation and Calibrate the Sensor ..............................................................
17
17
17
17
17
6.0
6.1
6.2
6.3
6.4
6.5
6.6
DISPLAY AND OPERATION ..................................................................................
Main Display ............................................................................................................
Keypad.....................................................................................................................
Operation .................................................................................................................
Hold .........................................................................................................................
Main Display ............................................................................................................
Security....................................................................................................................
18
18
18
19
21
22
23
7.0
7.1
7.2
7.3
7.4
7.5
7.6
7.7
PROGRAMMING THE ANALYZER .......................................................................
Entering the Program Menus...................................................................................
Outputs ....................................................................................................................
Relays......................................................................................................................
Measurement...........................................................................................................
Temperature.............................................................................................................
Security....................................................................................................................
Restoring Default Settings .......................................................................................
24
24
24
25
27
27
28
28
i
MODEL TCL-56
TABLE OF CONTENTS
TABLE OF CONTENTS CONT'D.
8.0
8.1
8.2
8.3
8.4
8.5
8.6
CALIBRATION .......................................................................................................
Introduction ..............................................................................................................
Entering the Calibration Menus ...............................................................................
Calibrating Temperature ..........................................................................................
Calibrating the Free Chlorine Sensor ......................................................................
Calibrating the Analog Outputs ................................................................................
Reset .......................................................................................................................
29
29
29
29
30
30
31
9.0
9.1
9.2
9.3
9.4
9.5
DATA AND EVENT LOGGING AND RETRIEVAL .................................................
Overview..................................................................................................................
Configuration ...........................................................................................................
Downloading Data and Events ................................................................................
Viewing Events ........................................................................................................
Date and Time .........................................................................................................
33
33
33
34
34
34
10.0
10.1
10.2
GRAPHICAL DISPLAY ..........................................................................................
Overview..................................................................................................................
Configuration ...........................................................................................................
35
35
35
11.0
DIGITAL COMMUNICATIONS ................................................................................
36
12.0
12.1
12.2
12.3
MAINTENANCE .....................................................................................................
Analyzer...................................................................................................................
Total Chlorine Sensor ..............................................................................................
Sample Conditioning System...................................................................................
37
37
38
40
13.0
13.1
13.2
13.3
13.4
13.5
13.6
13.7
13.8
TROUBLESHOOTING ...........................................................................................
Overview..................................................................................................................
Reading and Troubleshooting Fault and Warning Messages ..................................
Sensor Diagnostics..................................................................................................
Troubleshooting Calibration Problems.....................................................................
Other Troubleshooting .............................................................................................
Other Troubleshooting - General .............................................................................
Simulating Inputs - Chlorine.....................................................................................
Simulating Inputs - Temperature..............................................................................
47
47
47
48
48
48
50
51
52
14.0
14.1
14.2
14.3
RETURN OF MATERIAL ........................................................................................
General ....................................................................................................................
Warranty Repair.......................................................................................................
Non-Warranty Repair ...............................................................................................
53
53
53
53
LIST OF TABLES
Number
Title
Page
12-1
Spare Parts .............................................................................................................
39
12-2
Replacement Parts and Reagent for Sample Conditioning System .......................
46
ii
MODEL TCL-56
TABLE OF CONTENTS
LIST OF FIGURES
Number
2-1
3-1
3-2
3-3
3-4
3-5
4-1
4-2
4-3
4-4
4-5
6-1
6-2
7-1
7-2
7-3
7-4
7-5
12-1
12-2
12-3
12-4
12-5
12-6
12-7
12-8
12-9
12-10
12-11
12-12
12-13
12-14
12-15
12-16
12-17
13-1
13-2
13-3
13-4
Title
Schematic of Sample Conditioning System and Analyzer ....................................
Panel Mounting Dimensions ..................................................................................
Pipe and Wall Mounting Dimensions ......................................................................
Installing the Sample Conditioning Enclosure.........................................................
TCL Case Dimensions ............................................................................................
Reagent Tubing Assembly ......................................................................................
Power Wiring ..........................................................................................................
Analog output connections......................................................................................
Alarm relay connections..........................................................................................
Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Model 56 Analyzer........
Wiring Sensor with Standard Cable to Model 56 Analyzer......................................
Main Display ..........................................................................................................
Analyzer keypad .....................................................................................................
Menu Tree for the Outputs Submenu .....................................................................
Menu Tree for the Relays Submenu .......................................................................
Menu Tree for the Measurements Submenu ..........................................................
Menu Tree for the Temperature Submenu ..............................................................
Menu Tree for the pH Diagnostic Setup Submenu .................................................
Sensor Board Connections ....................................................................................
Sensor Parts ...........................................................................................................
Replacing Reagent Tubing......................................................................................
Replacing Sample Tubing .......................................................................................
Peristaltic Pump Tubing ..........................................................................................
Pump Cover ............................................................................................................
Removing the Cover ...............................................................................................
Inserting New Tube .................................................................................................
Replace the Cover ..................................................................................................
Bottom of the Cover ................................................................................................
Tracks .....................................................................................................................
Disconnection .........................................................................................................
Removal of Screws .................................................................................................
Connect the Air Inlet and Outlet Tubing to the Air Pump.........................................
Schematic ...............................................................................................................
Remove the Air Inlet Fitting.....................................................................................
Slide Pump Assembly Out of the Air Pump Body....................................................
Disconnecting Sample and Reagent Tubing Prior to Checking Flow......................
Simulating Chlorine.................................................................................................
Three Wire RTD Configuration ...............................................................................
Simulating RTD Inputs ............................................................................................
iii
Page
7
9
10
11
12
12
13
14
15
16
16
18
18
24
25
27
27
28
37
39
40
41
42
42
42
42
43
43
43
44
44
44
44
45
45
49
51
52
52
MODEL TCL-56
SECTION 1
DESCRIPTION AND SPECIFICATIONS
SECTION 1.
DESCRIPTION AND SPECIFICATIONS
Model TCL Sample Conditioning System
•
•
•
•
NO METAL WETTED PARTS. Ideal for seawater.
LOW SAMPLE FLOW (about 15 mL/minute) means little waste.
REAGENT-BASED SYSTEM measures true total chlorine.
FIVE GALLONS OF REAGENT lasts two months.
Model 56 Chlorine Analyzer
•
•
•
•
LARGE, PROGRAMMABLE, FULL COLOR DISPLAY with easy to use interface.
FOUR INDEPENDENT OUTPUTS with HART digital communication standard.
FOUR FULLY PROGRAMMABLE RELAYS.
DATA LOGGER and graphical display standard.
Model 499A CL-02 Sensor
• MEMBRANE-COVERED AMPEROMETRIC SENSOR.
• NO TOOLS REQUIRED to change membrane.
• MAINTENANCE TAKES ONLY A FEW MINUTES a month.
• VARIOPOL CONNECTOR OPTION allows the sensor to be replaced without removing and
rewiring cable.
1.1 FEATURES
MODEL TCL SAMPLE CONDITIONING SYSTEM
The sample conditioning system permits a single sensor to measure total chlorine in water. The sample conditioning system continuously injects a solution of acetic
acid (vinegar) and potassium iodide into the sample.
The acid lowers the pH to between 3.5 and 4.5 and
allows total chlorine in the sample to quantitatively react
with the potassium iodide to produce iodine. The sensor
measures the iodine concentration, and the analyzer
displays the total oxidant concentration in ppm as Cl2.
MODEL 56 CHLORINE ANALYZER
The Model 56 chlorine analyzer is intended for the
continuous determination of chlorine in water. When
used with the TCL and 499ACL-02 sensor it measures
total chlorine.
The analyzer is housed in a corrosion resistant NEMA
4X enclosure. It is suitable for panel, pipe, or wall
mounting. The large, high resolution, full color display
shows total chorine concentration and temperature
in 0.5 inch (13 mm) high characters. Six other
user-selectable variables can be shown in smaller
characters. Operation of the analyzer is through a
membrane keypad.
Calibration and programming screens are simple and
intuitive. Plain language prompts in nine languages
guide the user. Information about program settings
and calibration, as well as troubleshooting guidance is
available on-screen at the press of a button.
The analyzer has four fully programmable analog outputs, with HART digital communication superimposed
on output 1. Profibus DP digital communications is
available as an option. PID control is standard.
Four fully programmable relays are available. Relays
can be configured as simple high/low setpoint alarms
or can be used to perform a number of timer functions
1
MODEL TCL-56
as well as time proportional control (TPC). For more
information about relay actions refer to Product Data
Sheet 71-56.
A data and event logger are standard. The data logger
stores up to 30 days of data, and the event logger
stores up to 300 events. Events can be viewed on
screen, and data can be shown in a full color dual
graphics display. Data and events can also be downloaded through a USB port on the front panel of the
analyzer.
MODEL 499A CL-02 SENSOR
The Model 499ACL-02 total chlorine sensor is used in
the TCL sample conditioning system. Although the sensor is called a chlorine sensor, it really measures iodine.
The iodine comes from the reaction between oxidants in
the sample and the acetic acid/potassium iodide
reagent added by the sample conditioning system.
The sensor consists of a gold cathode and a silver
anode in an electrolyte solution. A silicone membrane,
permeable to iodine, is stretched over the cathode.
The analyzer applies a voltage to the cathode sufficiently negative to reduce all the iodine reaching it.
Because the concentration of iodine in the sensor is
always zero, a concentration gradient continuously
forces iodine from the sample through the membrane
into the sensor.
2
SECTION 1
DESCRIPTION AND SPECIFICATIONS
The reduction of iodine in the sensor generates a current directly proportional to the diffusion rate of iodine
through the membrane, which is directly proportional to
the concentration of iodine in the sample. Because the
iodine concentration depends on the amount of total
chlorine in the sample, the sensor current is ultimately
proportional to the total chlorine concentration.
The permeability of the membrane to iodine is a function
of temperature. A Pt100 RTD in the sensor measures
the temperature, and the analyzer uses the temperature
to compensate the total chlorine reading for changes in
membrane permeability.
Sensor maintenance is fast and easy. Replacing the
membrane requires no special tools or fixtures. Simply
place the membrane assembly on the cathode and
screw the retainer in place. Installing a new membrane
and replenishing the electrolyte takes only a few minutes.
MODEL TCL-56
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1.2 SPECIFICATIONS — SAMPLE CONDITIONING SYSTEM
GENERAL
SAMPLE REQUIREMENTS
Enclosure: Fiberglass reinforced polyester, NEMA 3
(IP53) suitable for marine environments
Dimensions: 14.5 x 13.0 x 8.6 in. (369 x 329 x 218 mm)
Mounting: Wall
Ambient Temperature: 32° - 122°F (0 - 50°C)
Ambient Humidity: 0 - 90% (non-condensing)
Power: 115 Vac, 6.9 W, 50/60 Hz;
230 Vac, 7.0 W, 50/60 Hz
Hazardous Location: The TCL sample conditioning system has no hazardous location approvals.
Pumps:
EN 809:1998
Weight/Shipping Weight: 14 lb/16 lb (6.5 kg/7.5 kg)
Inlet Connection: compression fitting, accepts 1/4 in. OD
tubing
Drain Connection: 3/4 in. barbed fitting (must drain to
open atmosphere)
Inlet Pressure: <100 psig (791 kPa abs)
Flow: at least 0.25 gph (15 mL/min)
Temperature: 32 - 122°F (0 - 50°C)
Total Alkalinity: <300 mg/L as CaCO3. For samples containing <50 mg/L alkalinity, consult the factory.
SAMPLE CONDITIONING SYSTEM
Reagent: Potassium iodide in vinegar.
Reagent Usage: 5 gallons lasts approximately 60 days.
Reagent Pump: Fixed speed peristaltic pump, about
0.2 mL/min
Sample Pump: Fixed speed peristaltic pump, about 11 mL/min
1.3 SPECIFICATIONS — MODEL 56 ANALYZER
Case: Polycarbonate
Display: Full color LCD, 3.75 x 2.20 in. (95 x 56 mm); display can be customized by the user.
Languages: English, French, German, Italian, Spanish,
Portuguese, Chinese, Russian, and Polish.
Ambient Temperature and Humidity: 14 to 140°F
(-10 to 60°C); RH 5 to 95% (non-condensing). Between
23 and 131°F (-5 to 55°C) there is no visible degradation
in display response or performance.
Storage temperature: -4 to 140°F (-20 to 60°C)
Power: 85 to 265 VAC, 47.5 to 65.0 Hz, 20 W
RFI/EMI: EN-61326
LVD: EN-6101-01
Outputs: Four 4-20 or 0-20 mA isolated current outputs;
assignable to measurement or temperature; fully
scalable; maximum load 550 Ω. HART digital signal is
superimposed on output 1.
Alarms and Timers: Four relays, fully configurable as a
setpoint alarm, interval timer, TPC, bleed and feed
timer, delay timer, date and time timer, and fault alarm.
Relays: Form C, SPDT, epoxy sealed.
Relay Contact ratings:
5 A at 28 VDC or 300 VAC (resistive)
1/8 HP at 120/240 VAC
Control features: PID control (analog output) and time
proportional control or TPC (relays) are standard.
Data logger: Data automatically stored every 30 seconds for
30 days; older data removed to make room for new data.
The following data are automatically stored: date and
time, ppm, temperature, and raw sensor current.
Event logger: Stores up to 300 events with data and time
stamp: faults, warnings, calibration data, calibration
results (pass or fail), power on/off cycles, and hold on/off.
Alarm relay activation and deactivation can also be
stored. Older events are automatically removed to make
room for new events.
Data and event downloading: through USB port on front
panel.
Graphical display: Dual graphical display shows measurement data on the y-axis and time on the x-axis. The
y-axis is fully assignable and scalable. The x-axis can
be set to one hour, one day, seven days, or 30 days.
Digital communications: HART digital communications is
standard. Profibus DP is optional.
Weight/Shipping Weight: (rounded up to nearest lb or
nearest 0.5 kg): 1.5 kg (3 lb)/2.0 kg (4 lb)
3
MODEL TCL-56
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1.4 SPECIFICATIONS — MODEL 499ACL-02 SENSOR
Wetted Parts: Gold, Noryl®1 (PPO), Viton®2, EPDM, Silicone
Dimensions: 1.0 x 5.6 in. (25.4 x 143 mm)
Cable: 25 ft. (7.6m) standard
Pressure Rating: 0 to 65 psig (101 to 549 kPa)
Temperature Rating: 32 to 122°F (0 to 50°C)
Electrolyte Capacity: Approximately 25 mL
Electrolyte Life: Approximately 4 months
Weight/Shipping Weight: 1 lb/3 lb (0.5 kg/1.5 kg)
1 Noryl is a registered trademark of General Electric.
2 Viton is a registered trademark of DuPont Performance Elastomers.
1.5 PERFORMANCE SPECIFICATIONS — COMPLETE SYSTEM
Linear Range: 0 to 20 ppm (mg/L) as Cl2 (for higher ranges, consult factory)
Linearity (per ISO 15839): 0-10 ppm: 2%; 0-20 ppm: 3%
Response Time: Following a step change in concentration, the reading reaches 90% of final value within 7 minutes at
25°C.
Drift: At about 1.5 ppm in clean water and constant temperature, drift is typically less 0.05 ppm over two weeks.
Detection Limit (per ISO 15839): 0.02 ppm (mg/L) in clean water at room temperature
4
MODEL TCL-56
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1.6 ORDERING INFORMATION AND ACCESSORIES
Model TCL Reagent-Based Chlorine System. The TCL is used for the continuous determination of total chlorine in water.
The TCL consists of a sample conditioning system, a reagent carboy, a sensor, and an analyzer. Reagents must be
ordered separately. Regent kits for 0-5 ppm and 0-10 ppm chlorine are available. For higher ranges, consult the
factory. See ACCESSORIES - Sample Conditioning System.
MODEL
TCL
REAGENT-BASED CHLORINE SYSTEM
CODE
11
12
POWER (required selection)
115 V 50/60 Hz
230 V 50/60 Hz
CODE
280
281
ANALYZER (optional selection)
56-03-24-38-HT analyzer, with alarm relays, HART
56-03-24-38-DP analyzer, with alarm relays, Profibus DP
CODE
30
31
32
SENSOR (optional selection)
499ACL-02-54 sensor with standard cable
499ACL-02-54-60 sensor with optimum EMI/RFI cable
499ACL-02-54-VP sensor with Variopol 6 fitting (interconnecting cable must be ordered separately)
ACCESSORIES — SAMPLE CONDITIONING SYSTEM
PN
24134-00
24134-01
9160578
9322052
24153-00
9100204
9100132
9380094
9380095
9380091
24151-00
24135-00
9380090
9380093
9380092
24152-00
PN
Description
Air pump, 115 Vac, 50/60 Hz
Air pump, 230 Vac, 50/60 Hz
Air pump repair kit
Check valve for air injection line
Carboy for reagent, 5 gal/19 L, includes cap
Fuse, 0.25 A, 250 V, 3AG, slow blow for option -11 (115 Vac)
Fuse, 0.125 A, 250 V, 3AG, slow blow for option -12 (230 Vac)
Reagent pump, 115 Vac, 50/60 Hz
Reagent pump, 230 Vac, 50/60 Hz
Reagent pump replacement tubing
Reagent tubing replacement kit
Reagent uptake tubing, 6 ft (1.8 m), includes weight
Sample pump, 115 Vac, 50/60 Hz
Sample pump, 230 Vac, 50/60 Hz
Sample pump replacement tubing
Sample tubing replacement kit
Description
1
1
1
1
4
1
1
1
1
1
1
1
1
1
1
1
Weight*
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (1.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
lb (0.5 kg)
Ship Weight**
1 lb (0.5 kg)
1 lb (0.5 kg)
1 lb (0.5 kg)
1 lb (0.5 kg)
5 lb (2.0 kg)
1 lb (0.5 kg)
1 lb (0.5 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
2 lb (1 kg)
Weight*
Ship Weight**
24165-00
Acetic acid, 2 x 2.5 gal (9.5 L) bottles/case, with 25 g potassium iodide
(0-5 ppm total chlorine)
45 lb (20.5 kg)
48 lb (22.0 kg)
24165-01
Acetic acid, 2 x 2.5 gal (9.5 L) bottles/case, with 50 g potassium iodide
(0-10 ppm total chlorine)
45 lb (20.5 kg)
48 lb (22.0 kg)
24164-00
Potassium iodide, 25 g, sufficient for 5 gallons (19 L) of vinegar
(0-5 ppm total chlorine)
1 lb (0.5 kg)
1 lb (0.5 kg)
24164-01
Potassium iodide, 50 g, sufficient for 5 gallons (19 L) of vinegar
(0-10 ppm total chlorine)
1 lb (0.5 kg)
1 lb (0.5 kg)
*Weights are rounded up to the nearest whole pound or 0.5 kg.
5
ACCESSORIES — 1055-24 Analyzer
PN
9240048-00
23820-00
DESCRIPTION
WEIGHT*
SHIP WEIGHT*
Tag, stainless steel, specify marking
1 lb (0.5 kg)
1 lb (0.5 kg)
Pipe mounting kit
2 lb (1.0 kg)
3 lb (1.5 kg)
WEIGHT*
SHIP WEIGHT*
ACCESSORIES — 54eA Analyzer
PN
DESCRIPTION
2002577
Wall and two inch pipe mounting kit
2 lb (1.0 kg)
3 lb (1.5 kg)
23545-00
Panel mounting kit
23554-00
Cable glands, kit (Qty 5 of PG 13.5)
2 lb (1.0 kg)
1 lb (0.5 kg)
3 lb (1.5 kg)
1 lb (0.5 kg)
9240048-00
Stainless steel tag (specify marking)
1 lb (0.5 kg)
1 lb (0.5 kg)
WEIGHT*
SHIP WEIGHT*
ACCESSORIES — Sensor
PN
DESCRIPTION
23501-02
Total Chlorine Membrane, includes one membrane assembly and
one O-ring
1 lb (0.5 kg)
1 lb (0.5 kg)
23502-02
Total Chlorine Membrane Kit, includes 3 membrane assemblies and
three O-rings
1 lb (0.5 kg)
1 lb (0.5 kg)
9210438
Total Chlorine Sensor Fill Solution, 4 oz (120 mL)
1 lb (0.5 kg)
2 lb (1.0 kg)
FOR FIRST TIME VARIOPOL INSTALLATIONS
PART #
DESCRIPTION
23747-06
Interconnecting cable, VP 6, 2.5 ft (0.8 m)
23747-04
Interconnecting cable, VP 6, 4 ft (1.2m)
23747-02
Interconnecting cable, VP 6, 10 ft (3.0 m)
23747-07
Interconnecting cable, VP 6, 15 ft (4.6 m)
23747-08
Interconnecting cable, VP 6, 20 ft (6.1 m)
23747-09
Interconnecting cable, VP 6, 25 ft (7.6 m)
23747-10
Interconnecting cable, VP 6, 30 ft (9.1 m)
23747-03
Interconnecting cable, VP 6, 50 ft (15.2 m)
23747-11
Interconnecting cable, VP 6, 100 ft (30.5 m)
*Weights are rounded up to the nearest whole pound or 0.5 kg.
6
MODEL TCL-56
SECTION 2
PRINCIPLES OF OPERATION
SECTION 2.
PRINCIPLES OF OPERATION
Total chlorine by definition is the iodine produced in a sample when it is treated with
potassium iodide at a pH between 3.5 and 4.5.
Typically, acetic acid (or vinegar) is used to
adjust the pH.
The total chlorine analyzer consists of a sample conditioning system, which injects the
reagent into the sample, and a sensor and
analyzer, which measure the amount of iodine
produced. Figure 2-1 shows the sample conditioning system. The sample enters the sample
conditioning enclosure and flows to an overflow sampler from which the sample pump
takes suction. Excess sample drains to waste.
At the same time, the reagent pump draws
reagent, a solution of potassium iodide in
vinegar, from the reagent carboy and injects it
into the suction side of the sample pump. The
sample and reagent mix as they pass through
the pump, and total chlorine in the sample is
converted to the chemically equivalent amount
of iodine. The flow rates are 11 mL/min for the
sample and 0.2 mL/min for the reagent.
FIGURE 2-1. Schematic of Sample Conditioning
System and Analyzer.
The treated sample next enters the flow cell. Bubbles injected into the flow cell produce turbulence, which
improves the stability of the reading. A membrane-covered amperometric sensor in the flow cell measures the
concentration of iodine. The analyzer receives the raw signal from the sensor and displays the concentration of
total chlorine. Display units are ppm (mg/L) chlorine as Cl2. The treated sample leaves the flow cell and drains
to waste along with the excess sample.
7
MODEL TCL-56
SECTION 3
INSTALLATION
SECTION 3.
INSTALLATION
3.1 UNPACKING AND INSPECTION
Inspect the shipping containers. If there is damage, contact the shipper immediately for instructions. Save the
boxes. If there is no apparent damage, unpack the containers. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analytical immediately.
3.2 INSTALLATION.
3.2.1 General Information
1. Although the analyzer and sample conditioning system are suitable for outdoor use, do not install them in direct
sunlight or in areas of extreme temperature.
CAUTION
The TCL Total Chlorine sample conditioning system
is NOT suitable for use in hazardous areas.
2. Install the analyzer and sample conditioning system in an area where vibration and electromagnetic and radio
frequency interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer and sample conditioning system.
4. The analyzer is suitable for panel, pipe or wall mounting. The sample conditioning enclosure must be mounted on a wall. Provide adequate room beneath the enclosure for the 5-gallon reagent carboy.
5. Be sure that the distance between the analyzer and sample conditioning cabinet does not exceed the length
of the sensor cable.
3.2.2 Install the Analyzer
1. Refer to the appropriate figure for installation details.
Type of Mounting
Figure
Panel
3-1
Wall or Pipe
3-2
2. See section 4.1 for more information about the conduit openings.
3. See Section 4.2 for wiring instructions.
8
MODEL TCL-56
SECTION 3
INSTALLATION
INCH
MILLIMETER
FIGURE 3-1 PANEL MOUNTING DIMENSIONS
9
INCH
MILLIMETER
FIGURE 3-2 PIPE AND WALL MOUNTING DIMENSIONS
(Mounting bracket PN:23820-00)
10
MODEL TCL-56
SECTION 3
INSTALLATION
3.2.3 Install the Sample Conditioning Enclosure
1. Refer to Figures 3-3, 3-4, and 3-5 for installation details.
2. Connect the sample line to the sample conditioning system. Use ¼-inch OD hard plastic or stainless steel tubing.
If dechlorinated water is being measured, provide a way for occasionally substituting a chlorinated water sample
for the dechlorinated sample. Chlorinated water is needed to calibrate the sensor and to check its response.
3. If a grab sample tap is not already available, install one in the process piping. Choose a point as close as possible to the sample line supplying the TCL. Be sure that opening the sample valve does not appreciably alter
the flow of sample to the instrument.
4. Connect the drain to a length of ¾-inch ID flexible plastic tubing. The sample must drain to open atmosphere.
5. Find the reagent tubing and fitting in the plastic bag taped to the inside of the enclosure door. Screw the
reagent fitting onto the bulkhead fitting at the bottom left of the enclosure. Pass the reagent tubing through the
hole in the carboy cap. Be sure the plastic weight will be inside the carboy when the cap is in place. Attach the
reagent tubing to the barbed connector. See Figure 3-5.
6. Place the blue plastic carboy beneath the enclosure. Screw the cap and tubing assembly on to the carboy. To
prepare reagent, see Section 5.2.
3.2.4 Install the Sensor
1. From inside the sample conditioning enclosure, thread the sensor cable or VP cable through the gland on the
upper left side. Leave about one foot of cable inside the enclosure.
2. Wire the cable to the analyzer. Refer to Section 4.4.
3. Remove the nut and adapter from the flow cell. Slip the nut over the end of the sensor. Thread the adapter
onto the sensor. Hand tighten only. If you are using a VP cable, connect the cable to the sensor. The connector and receptacle are keyed to ensure proper mating. Once the key has slid into place, tighten the connection by turning the knurled ring clockwise. Remove the protective cap from the end of the sensor.
4. Insert the sensor in the flow cell. Hand tighten the nut.
analyzer
process piping
sensor cable
sample inlet 1/4
inch OD tubing
TCL
enclosure
drain
3/4 inch ID barbed fitting
sample tap
reagent
bottle
FIGURE 3-3. Installing the Sample Conditioning Enclosure
11
MODEL TCL-56
SECTION 3
INSTALLATION
INCH
MILLIMETER
FIGURE 3-4. TCL Case Dimensions
FIGURE 3-5. Reagent tubing assembly
12
MODEL TCL-56
SECTION 4
WIRING
SECTION 4.
WIRING
4.1 PREPARE ANALYZER CONDUIT OPENINGS
The analyzer enclosure has six conduit openings. Four conduit openings are fitted with conduit plugs.
Conduit openings accept 1/2-inch conduit fittings or PG 13.5 cable glands. To keep the case watertight, block
unused openings with NEMA 4X or IP65 conduit plugs.
NOTE
Use watertight fittings and hubs that comply with the requirements of UL514B. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer (UL508-26.16).
4.2 PROVIDE POWER TO THE SAMPLE CONDITIONING SYSTEM
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local code.
NOTE
Provide a switch or breaker to disconnect the sample conditioning cabinet from the
main power supply. Install the switch or breaker near the unit and identify if as the
disconnecting device for the sample conditioning system.
1. Be sure the pump switches on the wiring access panel are in the off position.
2. Remove the four screws securing the wiring access panel. Pull the panel out of the way to reveal the power
terminal strip.
3. Insert the power cable through the strain relief connection labeled power (see Figure 3-4). Wire the power
cable to the terminal strip as shown in Figure 4-1. Do not apply 230 Vac power to a 115 Vac TCL (Model option
-11). Doing so will damage the instrument.
4. Leave the pump power switches off until ready to start up the unit. See Section 5.
Model option -11 115 Vac only
Model option -12 230 Vac only
FIGURE 4-1. Power Wiring
13
MODEL TCL-56
SECTION 4
WIRING
4.3 MAKE POWER, ALARM, OUTPUT, AND SENSOR CONNECTIONS IN THE ANALYZER
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local code.
4.3.1 Power
Wire AC mains power to the power supply board, which is mounted on the left hand side of the enclosure beneath
the gray plastic cover. To remove the cover, grab it by the upper edges and pull straight out. The power connector
is at the bottom of the board. See Figure 4-3. Bring the power cable through the conduit opening just below the
connector. Unplug the connector from the board and wire the power cable to it. Lead connections are marked on
the connector. (L is live or hot; N is neutral; the ground connection has the standard symbol.)
Replace the cover. The two tabs on the back edge of the cover fit into slots at the rear of the enclosure, and the
three small slots in the front of the cover snap into the three tabs next to the relay terminal strip. See Figure 4-3.
Once the tabs are lined up, push the cover to snap it in place.
AC power wiring should be 14 gauge or greater. Run the power wiring through the conduit opening nearest the
power terminal. Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the
switch or breaker near the analyzer and label it as the disconnecting device for the analyzer.
4.3.2 Analog output wiring
Four analog current outputs are
located on the main circuit board,
which is attached to the inside of
the enclosure door. Figure 3-1
shows the location of the terminals, the outputs they are
assigned to, and the polarity.
For best EMI/RFI protection, use
shielded output signal cable
enclosed in earth-grounded metal
conduit.
Keep output signal wiring separate
from power wiring. Do not run signal and power or relay wiring in
the same conduit or close together
in a cable tray.
FIGURE 4-2. Analog output connections.
The analog outputs are on the main board near
the hinged end of the enclosure door.
14
MODEL TCL-56
4.3.3
SECTION 4
WIRING
Alarm wiring.
WARNING
Exposure to some chemicals may degrade the sealing
properties used in the following devices: Zettler
Relays (K1-K4) PN AZ8-1CH12DSEA
The alarm relay terminal strip is located on the power
supply board, which is mounted on the left hand side of
the enclosure beneath the gray plastic cover. To remove
the cover, grab it by the upper edges and pull straight
out. The relay terminal strip is at the top of the board.
See Figure 4-3. Bring the relay wires through the rear
conduit opening on the left hand side of the enclosure
and make connections to the terminal strip.
Replace the cover. The two tabs on the back edge of
the cover fit into slots at the rear of the enclosure, and
the three small slots in the front of the cover snap into
the three tabs next to the relay terminal strip. See
Figure 4-3. Once the tabs are lined up, push the cover
to snap it in place.
Keep alarm relay wiring separate from signal wiring.
Do not run signal and power or relay wiring in the
same conduit or close together in a cable tray.
FIGURE 4-3. Alarm relay
connections.
4.4 SENSOR WIRING
1.
Shut off power to the analyzer.
2.
Locate the chlorine signal board.
Slot 1 (left)
communication
Slot 2 (center)
input 1 (chlorine)
Slot 3 (right)
input 2 (optional)
3.
Insert the sensor cable through the conduit opening nearest the chlorine board.
4.
Slide the board forward to gain access to the wires and terminal screws.
5.
Connect the sensor cable to the chlorine board. Refer to Figure 4-4 or 4-5.
6.
Once the cable has been connected, slide the board fully into the enclosure while taking up the excess
cable through the conduit opening. If you are using a cable gland, tighten the gland nut to secure the cable
and ensure a sealed enclosure.
15
MODEL TCL-56
FIGURE 4.4. Wiring Sensor with Optimum EMI/RFI
or Variopol Cable to Model 56 Analyzer
SECTION 4
WIRING
FIGURE 4.5 Wiring Sensor with Standard Cable
to Model 56 Analyzer
4.5 APPLY POWER TO THE ANALYZER AND COMPLETE QUICK START
1. Once all wiring connections are secured and verified, apply power to the analyzer.
2. When the analyzer is powered up for the first time, Quick Start screens appear. The first quick start screen
has two control boxes, one for language and the other for temperature units.
a. The cursor, shown by dark blue backlighting, will be on the language control box. To change the
language, press the ENTER/MENU key. A list of available languages, shown two at a time, will
appear. Using the up and down keys, scroll (see section 4.2) to display the choices. Press
ENTER/MENU to select the desired language. Press the down key to move to the cursor to the
temperature control box. To change units, press ENTER/MENU and scroll to either °F or °C.
Press ENTER/MENU to store the selection.
b. To move to the next screen, use the navigation keys to move the cursor to NEXT and press
ENTER/MENU.
3. The next screen lists navigation rules. Press ENTER/MENU for the next screen.
4. The next step is to configure sensor 1. Sensor 1 is the total chlorine sensor. The screen has two control boxes.
a. For measurement choose total chlorine.
b. Choose the desired units, mg/L or ppm
5. Move the cursor to NEXT and press ENTER/MENU. The display will change to show some basic keypad operation guidelines. Press ENTER/MENU to show the main display.
6. The outputs, alarms, display configuration, and data logging are all assigned to default values. The default
value for data logging is disabled. To change the settings refer to Sections 6.5, 7.0, and 9.0.
16
MODEL TCL-56
SECTION 5
START-UP
SECTION 5.
START-UP
NOTE
Complete Section 4 before starting this section.
5.1 PREPARE THE REAGENT
WARNING
The reagent contains potassium iodide dissolved in distilled vinegar or 5%
acetic acid. Avoid contact with skin and eyes. Wash thoroughly after using.
1. DO NOT PREPARE THE SOLUTION UNTIL READY TO USE.
2. Position the blue plastic carboy under the sample conditioning cabinet. Unscrew the cap and reagent tube
assembly.
3. Add the potassium iodide reagent to the carboy. See the table.
Expected range,
ppm as Cl2
Amount of KI needed
per 5 gal (19 L) of vinegar
Part number
0 – 5 ppm
25 grams
24164-00
0 – 10 ppm
50 grams
24164-01
0 – 20 ppm
2 x 50 grams
24164-01
4. Add five gallons (19 L) of distilled white vinegar one gallon (4 L) at a time. Swirl the carboy after each addition
5. Screw the cap on the carboy. Be sure the reagent uptake tube extends to the bottom of the carboy.
6. If it hasn’t already been connected, connect the reagent tube to the small fitting on the bottom left hand side
of the enclosure.
NOTE
The shelf life of the potassium iodide vinegar solution is at least two months if stored in the blue
carboy. Do not store the reagent in a container other than the blue carboy. The reagent is sensitive to sunlight, which the blue carboy effectively blocks.
5.2 ZERO THE SENSOR
1. Place the sensor in a beaker of deionized water or simply place the sensor in air.
2. Let the sensor operate until the sensor current is stable, then zero the sensor. See Section 8.4.
5.3 START SAMPLE FLOW
Adjust the sample flow until a slow stream of liquid is running down the inside tube of the sampling cup.
5.4 BEGIN OPERATION AND CALIBRATE THE SENSOR
1. Turn on the reagent and sample pump switches. Observe that liquid begins to fill the flow cell. The sample flow
is about 11 mL/min, so the flow cell will fill rather slowly. Also observe that the air pump is operating. The pump
will produce very vigorous bubbling in the flow cell.
2. Once the flow of reagent starts, it takes about two minutes for the reagent to reach the flow cell. If the concentration of total chlorine in the sample is greater than about 0.5 ppm, the treated sample in the flow cell will
be pale yellow. Sample containing more chlorine will be dark yellow.
3.
Monitor the sensor current. Once the reading is stable, calibrate the unit. See Section 8.4. It may take thirty
minutes or longer for the reading to stabilize when the sensor is first put in service.
17
MODEL TCL-56
SECTION 6
OPERATION
SECTION 6.0
DISPLAY AND OPERATION
6.1. MAIN DISPLAY
The analyzer has a four line display. See Figure 6-1. The display can be customized to meet user requirements. See
Section 6.5. Fault or warning messages, if appropriate, appear at the bottom of the screen. See Section 13.1.
1.
7.
ppm
pH
T1: 25.0qC O1: 12.00 mA O3: 12.00 mA
T2: 25.0qC O2: 12.00 mA O4: 12.00 mA
Warning Press INFO key
FIGURE 6-1. Main Display
The following abbreviations are used in the lower two lines of the display. The number following the abbreviation
refers to the sensor, alarm relay, or output.
O
output
M
measurement
T
temperature (live)
AL
alarm relay
Tm
temperature (manual)
I
sensor current (chlorine)
6.2. KEYPAD
Local communication with the analyzer is through the membrane keypad. See Figure 6-2.
FIGURE 6-2. Analyzer keypad.
18
MODEL TCL-56
SECTION 6
OPERATION
6.3 OPERATION
The operation of the Model 56 can best be understood from the following example.
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
q
Calibrate
Data storage and retrieval
Program
HART
Hold
Time and date
Display setup
Reset
q
q
1. With the main display
showing (Figure
6-1), press the ENTER/MENU key.
q
q
The main menu, shown at left, will appear. Pressing the ENTER/MENU
key will bring up the main menu only if the main display is showing.
Note that the current reading and temperature for sensor 1 (S1) and
sensor (S2), if applicable, always appear at the top of the screen.
The cursor (dark blue backlit field) is on the Calibrate button. Press the
down key to move the cursor to the Program button.
q
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Output
1
Analog/PID/Simulate
To select a different program submenu use the right key to move the cursor
to the desired tab and press ENTER/MENU.
Analog
Assign S1 measurement
Range
4-20 mA
Scale
Linear
Dampening
sec
0
NEXT
q q
q
q
2. Press the ENTER/MENU
key.q The
cursor is on the outputs tab and the
q
first screen in the outputs sub-menu is showing.
S1: 1.00 ppm 25.0qCq
BACK
S2: 7.00 pH 25.0qCq
Outputs Relays Measure Temperature Security
Output
1
Analog/PID/Simulate
3. To enter the outputs submenu, press the down key. The cursor moves to the
first control box, Output. The Model 56 has four analog outputs, and this
control lets the user select which output to configure.
Analog
Assign S1 measurement
Range
4-20 mA
Scale
Linear
Dampening
sec
0
NEXT
q q
S1: 1.00 ppm 25.0qC
q
BACK
S2: 7.00 pH 25.0qC
q
Outputs Relays Measure Temperature Security
1
1
2
Assign S1 measurement
Output
Analog/PID/Simulate
Range
4. The default is output 1. To select a different output, press the
ENTER/MENU key. A list of the available outputs, shown two at a time,
appears. To view the list, press or press and hold the up or down key.
To select and store the highlighted selection, press ENTER/MENU.
4-20 mA
Scale
Linear
Dampening
sec
0
NEXT
BACK
S
5. To move from one control box to another, press the up or down key.
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Output
1
Analog/PID/Simulate
6. Some controls require
the user to qselect an item from a list. Others, like the
q
dampening control, require the user to enter a number. Move the cursor to
Dampening at the bottom of the screen.
Analog
Assign S1 measurement
Range
4-20 mA
Scale
Linear
Dampening
0
sec
NEXT
q
BACK
q
q
q
19
MODEL TCL-56
SECTION 6
OPERATION
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Output
1
Analog/PID/Simulate
Analog
Assign S1 measurement
Range
4-20 mA
Scale
Linear
Dampening
sec
NEXT
7. The default dampening
value is q0 seconds. To change the value, press
q
ENTER/MENU. The dark blue back-lighting will disappear indicating that
a number can be entered. Use the numeric keypad to enter the desired
number. If you make an error, press the left key to erase the digit last
entered. To store the number, press ENTER/MENU.
BACK
S
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Dampening applies to the output only, not the main
display. Increasing the dampening time reduces the
noise on the output, but increases the response time.
q
q
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Outputs Relays Measure
q Temperature Security
q
Output
1
Analog/PID/Simulate
Analog
8. Every control box has an information or help screen associated with it. To
view the information screen for the control box the cursor is on, press the
INFO key. The information screen for the Dampening control is shown at
left. To close the information screen, press any key.
9. A NEXT and BACK button are at the bottom of the screen. The NEXT button
means that additional control boxes are available on at least one more
screen. To view the next screen, use the navigation keys (either down or
right) to move the cursor to NEXT and press ENTER/MENU.
Assign S1 measurement
Range
4-20 mA
Scale
Linear
Dampening
sec
NEXT
S1: 1.00 ppm 25.0qC
BACK
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Range: 0/4 mA
0.000
ppm
Range: 20 mA
10.00
ppm
Fault
Fault current
10. The next screen in the Outputs sub-menu appears. The cursor is on the
outputs tab. To enter the screen, press the down navigation key.
Fixed
22.00
mA
BACK
11. To return to the previous screen, move the cursor to BACK and press
ENTER/MENU.
12. To return to the main menu, press EXIT.
20
MODEL TCL-56
SECTION 6
OPERATION
6.4 HOLD
6.4.1 Purpose
To prevent unwanted alarms and improper operation of control systems or dosing pumps, place the alarm
relays and outputs assigned to the sensor in hold before removing the sensor for maintenance. Hold is also
useful if calibration, for example, buffering a pH sensor, will cause an out of limits condition. During hold,
outputs assigned to the sensor remain at the last value, and alarms assigned to the sensor remain in their
present state.
6.4.2 Using the Hold Function.
The hold function uses certain programming features not discussed in Section 6.3.
q
S1: 1.00 ppm 25.0qC
Hold what?
q
S2: 7.00 pH 25.0qC
Sensor 1 output(s) and alarm relay(s)
Sensor 2 output(s) and alarm relay(s)
Analyzer will remain in hold until taken out of hold. To
take analyzer out of hold, move the cursor to the
checked item and press ENTER.
APPLY
BACK
S1: 1.00 ppm 25.0qC
q
Hold what?
S2: 7.00 pH 25.0qC
q
9 Sensor 1 output(s) and alarm relay(s)
9
Sensor 2 output(s) and alarm relay(s)
1.
7.
1. With the main display showing, press ENTER/MENU. The main menu will
appear. Choose Hold. The screen shown at left appears. The cursor is on
ppmoutputs and relays associated with sensor 1,
the first check box. To hold
press ENTER/MENU. A check will appear in the check box. To put
sensor 2 on hold also, pH
move the cursor to the sensor 2 line and press
qq
ENTER/MENU
to check the sensor 2 hold box.
qq
2. To activate Hold, move the cursor to the APPLY button at the bottom left
of the screen and press ENTER/MENU. The selected sensor outputs and
alarm relays will remain on hold until taken out of hold. However, if power
is lost then restored, hold will automatically be turned off.
Analyzer will remain in hold until taken out of hold. To
take analyzer out of hold, move the cursor to the
checked item and press ENTER.
APPLY
BACK
3. The screen describes how to take the analyzer out of hold. Be sure to
press APPLY once the box has been unchecked.
q
1.
7.
ppm
4. A message stating which sensors are in hold will appear in the fault/warning
banner at the bottom of the main display.
pH
T1: 25.0qC O1: 12.00 mA O3: 12.00 mA
T2: 25.0qC O2: 12.00 mA O4: 12.00 mA
S1 Hold
q
21
MODEL TCL-56
SECTION 6
OPERATION
6.5 MAIN DISPLAY
6.5.1 Configuring the main display
The main display can be configured to meet specific user requirements.
q
q
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Graphics Display setup Tag Language Warning
View graph
Upper graph
q
Lower graph
Variable (y-axis) Measure 1
Measure 2
Y-axis (maximum)
10.00
ppm
0.00
pH
Y-axis (minimum)
0.00
ppm
14.00
pH
X-axis (time)
q
1. With the main qdisplay showing,q press ENTER/MENU. The main menu will
appear. Choose Display Setup. The screen shown at left appears.
1 day
1 day
BACK
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Graphics Display setup Tag Language Warning
q
q
2. Move the cursor the Display setup tab and press ENTER/MENU. The
screen shown at left appears.
Configure main display
Set brightness
BACK
q
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Graphics Display setup Tag Language Warning
Measure 1
Measure 2
Temp 1
Output 1 mA
Output 3 mA
Temp 2
Output 2 mA
Output 4 mA
3. Choose Configure main display. The screen at left appears. The position
of each control box corresponds to the position of the variable in the main
display. Move the cursor to the control box and press ENTER/MENU. Use
the up and down keys to scroll through the list of variables and press
ENTER/MENU to select the desired variable for display.
BACK
q
6.5.2 Setting brightness
Move the cursor to the Set brightness button in the screen shown in step 2 in Section 6.5.1 and press
ENTER/MENU. Then, move the cursor to the Display brightness control and select the desired brightness. The
information screen gives recommendations about setting the brightness level especially in areas where the
ambient temperature exceeds 121°F (50°C).
22
MODEL TCL-56
SECTION 6
OPERATION
6.6 SECURITY
6.6.1 How the Security Code Works
Security codes prevent accidental or unwanted changes to program settings or calibrations. There are three levels
of security.
a. A user can view the main display and diagnostic screens only.
b. A user has access to the calibration and hold menus only.
c. A user has access to all menus.
S1: 1.00 ppm 25.0qC
S2: 1.00 pH 25.0qC
1.
If a security code has been programmed, pressing a sub-menu button
(See section 6.3) will cause the security screen shown at left to appear.
2.
Enter the three digit security code.
3.
If the entry is correct, the requested sub-menu will appear and the user
has access to all the sub-menus the code entitles him to.
4.
If the entry is wrong, the invalid code screen appears.
Enter security code
0
6.6.1 Assigning Security Codes
See Section 7.6.
6.6.2 Bypassing Security Codes
Call the factory.
23
MODEL TCL-56
SECTION 7
PROGRAMMING THE ANALYZER
SECTION 7.0
PROGRAMMING THE ANALYZER
7.1 ENTERING THE PROGRAM MENUS
q
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
Calibrate
1. With the main display showing, press ENTER/MENU to display the main
menu. Move the cursor to Program and press ENTER/MENU.
Data storage and retrieval
Program
HART
Hold
Time and date
Display setup
Reset
S1: 1.00 ppm 25.0qC
q
2. Move the cursor to the tab showing the desired sub menu and press
ENTER/MENU. A fifth tab, not shown, labeled pH diagnostics setup, will
be present if one of the sensors is a pH sensor.
S2: 7.00 pH 25.0qC
q
Outputs Relays Measure Temperature Security
Output
Analog/PID/Simulate
1
Analog
Assign S1 measurement
Range
4-20 mA
Scale
Dampening
Linear
0
sec
NEXT
BACK
7.2 OUTPUTS
7.2.1 Menu Tree
Figure 7-1 is the Outputs menu tree.
Outputs
Output 1, 2, 3, or 4
Analog
PID
Simulate
Assign
Assign
Range
Range
Scale
Setpoint
Dampening
LRV
Range 0/4 mA
URV
Range 20 mA
Proportional band
Fault
Integral
Fault current
Derivative
Fault
Fault current
FIGURE 7-1. Menu tree for the Outputs sub menu.
24
Hold at X mA
MODEL TCL-56
SECTION 7
PROGRAMMING THE ANALYZER
7.2.2. Settings
Move the cursor to the appropriate control box and make the desired setting. For more information about the control box the cursor is on press INFO. To close the information screen, press any key.
7.3 RELAYS
7.3.1 Menu Tree
Figure 7-2 is the Relays menu tree.
Relays
Delay timer
Date and time
Fault
Simulate
Assign relay
Assign relay
Assign relay
Assign relay
Type (delay timer)
Type (date and time)
Type (fault)
Energize/de-energize
Assign measurement
High/low
Setpoint
Deadband
Week 1
Normal state
Set time
Set duration
Week 2
On time
Delay time
Normal state
Set time
Set duration
FIGURE 7-2. Menu tree for the Relays sub menu.
25
MODEL TCL-56
SECTION 7
PROGRAMMING THE ANALYZER
7.3.2. Settings
q
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
Outputs Relays Measure Temperature Security
Explanation of relay actions
q
q
1. A large numberq of relay actionsq are available in the Model 56. For more
information about a relay action, move the cursor the Explanation of relay
actions button and press ENTER/MENU.
Configure relay
Simulate relay action
BACK
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
q
q Temperature Security
Outputs Relays Measure
Setpoint alarm info
Delay timer info
Interval timer info
Date and time timer info
TPC info
Totalizer based timer info
Bleed and feed info
Fault info
BACK
q
S1: 1.00 ppm 25.0qC
The totalizer-based relay timer is not available with the Model 56 offered as
standard options with the TCL. It is available only if one of the measurements
is flow.
To configure a relay, press EXIT to return to the screen in step 1.
S2: 7.00 pH 25.0qC
3. Move the cursor to the Configure relay button and press ENTER/MENU.
A screen similar to the one at left will appear.
Outputs Relays Measure Temperature Security
Relay
Type
2. The screen at left appears. Select the desired relay action and press
INFO to display the information screen. To close the information screen,
press any key.
1
Setpoint
Assign S1 measurement
Logic
High
Setpoint
10.00
ppm
Deadband
0.00
ppm
NEXT
BACK
4. Move the cursor to the appropriate control box and make the desired setting. For more information about the control the cursor is on press INFO.
To close the information screen, press any key.
q
26
MODEL TCL-56
SECTION 7
PROGRAMMING THE ANALYZER
7.4 MEASUREMENT
7.4.1 Menu Tree
Figure 7-3 is the Measurements menu tree.
Measurement
Sensor 1 or 2
Sensor 1 (Cl)
Sensor 2
Settings available depend on
the identity of sensor 2 (if
present)
Measurement
Units
Resolution
Filter
FIGURE 7-3. Menu tree for the Measurements sub menu.
7.4.2. Settings
Move the cursor to the appropriate control box and make the desired setting. For more information about the control
the cursor is on press INFO. To close the information screen, press any key.
7.5 TEMPERATURE
7.5.1 Menu Tree
Figure 7-4 is the Temperature menu tree.
Temperature
Sensor 1 or 2
Sensor 1 (Cl)
Sensor 2
Units
Units
Automatic/manual*
Automatic/manual*
* If manual, specify manual temperature.
* If manual, specify manual temperature.
FIGURE 7-4. Menu tree for the Temperature sub menu.
7.5.2. Settings
Move the cursor to the appropriate control box and make the desired setting. For more information about the control
the cursor is on press INFO. To close the information screen, press any key.
27
MODEL TCL-56
SECTION 7
PROGRAMMING THE ANALYZER
7.6 SECURITY
7.6.1 Menu Tree
Figure 7-5 is the security setup menu tree.
Security
Calibration/hold
All
FIGURE 7-5. Menu tree for the pH diagnostic setup sub menu.
7.6.2. Settings
Move the cursor to the appropriate control box and make the desired setting. For more information about the control the cursor is on press INFO. To close the information screen, press any key.
7.7 RESTORING DEFAULT SETTINGS
See section 8.6.
28
MODEL TCL-56
SECTION 8
CALIBRATION
SECTION 8.0
CALIBRATION
8.1 INTRODUCTION
The calibrate menu allows the user to do the following:
1. Calibrate the RTD (temperature sensing element) in the chlorine and pH sensors.
2. Calibrate the chlorine sensor.
3. Calibrate the analog outputs.
8.2 ENTERING THE CALIBRATION MENUS
q
q
S1: 1.00 ppm 25.0qC
Calibrate
S2: 7.00 pH 25.0qC
1.
With the main display showing, press ENTER/MENU to display the main
menu. The cursor will be on Calibrate. Press ENTER/MENU.
2.
Choose the sensor (measurement or temperature) or output to be calibrated.
Sensor 1 (S1) is the free chlorine sensor; sensor 2 (S2) is the pH sensor
(if present).
Data storage and retrieval
Program
HART
Hold
Time and date
Display setup
Reset
BACK
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
q
q
S1 Measurement
Output 1
S1 Temperature
Output 2
S2 Measurement
Output 3
S2 Temperature
Output 4
BACK
8.3 CALIBRATING TEMPERATURE
To calibrate the temperature device in the sensor, choose S1 temperature or S2 temperature and follow the prompts.
If you want more information about a calibration step, press the INFO key. Once the calibration is complete, the screen
will show the results of the calibration. The screen will also show some acceptance criteria to help you determine
whether to accept the calibration. Press the INFO key for an information screen to aid with troubleshooting if the
calibration results are not acceptable.
29
MODEL TCL-56
SECTION 8
CALIBRATION
8.4 CALIBRATING THE FREE CHLORINE SENSOR
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Why is calibration necessary? To find out press INFO.
Otherwise, choose the desired calibration method.
Zero
Grab
1. Choosing sensor 1 (total chlorine) in section 8.2 causes the screen
shown at left to appear. There are two steps to calibrating a free chlorine
sensor, measuring the zero current (zero) and determining the slope of the
calibration curve (grab). Because stable total chlorine standards in the
ppm range do not exist, the sensor must be calibrated against the results
of a laboratory test run on a grab sample.
BACK
2. To zero the sensor, select Zero and follow the prompts. For more
information about preparing the zero solution and measuring the zero
current, press the INFO key when prompted.
If the zero step is successful, the analyzer will display the zero complete
screen and the measured zero current. The screen will also show the
typical zero current for the sensor and the recommended acceptance
criterion. You will be asked to accept the zero current. Press the INFO
key for an information screen to aid with troubleshooting if the results
are not acceptable.
If the zero current is badly in error, the analyzer will display the zero
failed screen. Press the INFO key for troubleshooting.
3. To calibrate the sensor response in chlorinated water, select Grab and follow
the prompts. Be sure the sensor is installed in the flow cell in the TCL and
the sample is flowing down the inside tube of the overflow sampler. Also,
verify that reagent is being delivered to the sample and that the air pump
is working.
If the calibration is successful, the analyzer will display the calibration
complete screen and the sensitivity (nA/ppm). The screen will also show
the typical sensitivity range for the sensor and the recommended acceptance
criterion. You will be asked to accept the calibration. Press the INFO key
for an information screen to aid with troubleshooting if the calibration is
not acceptable.
If the sensitivity is badly in error, the analyzer will display the calibration
failed screen. Press the INFO key for troubleshooting.
8.5 CALIBRATING THE ANALOG OUTPUTS
Choose the appropriate output in section 8.2 and follow the prompts to trim the selected output. If the calibration is
successful the trim complete screen will appear. If the entered value is more than 1.0 mA different from the simulated
output current, the analyzer will display the possible error screen, and you will be asked to accept the calibration.
Press the INFO key for an information screen to aid with troubleshooting if the calibration is not acceptable.
30
MODEL TCL-56
SECTION 8
CALIBRATION
8.6 RESET
8.6.1 Purpose
There are three resets.
1. Reset all user settings, including calibration and program settings, to the factory default values. The analyzer
will return to Quick Start. The event logger and data logger (See section 8.0) will be unaffected.
2. Reset sensor calibration to the default value. The analyzer will clear all user-entered calibration data for the
selected sensor. It will leave all other user-entered data unaffected.
3. Reset the analog output calibration for the selected output to the default value. The analyzer will leave all other
user-entered settings unchanged.
8.6.2 Procedure
1. With the main display showing, press ENTER/MENU to display the main menu. Move the cursor to Reset and
press ENTER/MENU.
2. Check the desired boxes and press APPLY.
31
MODEL TCL-56
32
SECTION 8
CALIBRATION
MODEL TCL-56
SECTION 9
DATA EVENT LOGGING AND RETRIEVAL
SECTION 9.0
DATA AND EVENT LOGGING AND RETRIEVAL
9.1. OVERVIEW
Data and event logging is a standard feature in the Model 56 analyzer. However, the feature must be enabled.
When data/event logging is enabled, the Model 56 analyzer will automatically store the following events with date
and time stamp: faults, warnings, calibration data, calibration results (pass or fail), power on/off cycles, hold on/off,
and new sensor board detected. At the user’s discretion the analyzer will also store alarm activation and deactivation
as events. The event logger holds 300 events. When the capacity of the logger is reached, the oldest events are
removed to make room for new events.
When data/event logging is enabled, the analyzer will automatically store the following measurement data for total
chlorine: date and time, ppm chlorine, temperature, and sensor current.
The analyzer can store up to 30 days of data. When the capacity of the logger is reached, the oldest data are
removed to make room for new data. Data storage frequency is once every 30 seconds.
9 .2. CONFIGURATION
S1: 1.00 ppm 25.0qC
q
Calibrate
S2: 7.00 pH 25.0qC
q
Data storage and retrieval
Program
HART
Hold
Time and date
Display setup
Reset
q
S1: 1.00 ppm 25.0qC
Configure
Download
1. With the main display showing, press MENU/ENTER. Choose Data storage
and retrieval
q
S2: 7.00 pH 25.0qC
View events
Data/event logger is currently
Disabled
2. The screen shown at left appears. The data logger is currently disabled
(default). To enable the data logger, move the cursor to the Enable
data/event logger button and press ENTER/MENU.
Enable data/event logger
Disable data/event logger
BACK
3. Make the appropriate date and time settings and choose which alarm
relay activations and deactivations to record as events.
NOTE
Setting the date or time to an earlier value than the one currently showing
will cause data to be lost from the data/event logger. Download data
before resetting time or date. See section 9.3.
33
MODEL TCL-56
SECTION 9
DATA EVENT LOGGING AND RETRIEVAL
9.3. DOWNLOADING DATA AND EVENTS
To download data or events, move the cursor to the download tab and press ENTER/MENU. Unscrew the USB
port cover in the lower right hand corner of the front panel and insert a USB flash drive in the port. Press the
appropriate button to download data or events. Downloading may take as long as 20 minutes. During download,
the display and keypad are frozen, but all other analyzer functions continue.
Downloaded data and events are stored in a spreadsheet. There is a separate spreadsheet for every day of data.
The filename for downloaded data is dl mmddyy or dl ddmmyy, depending on the date and time format selected by the user. The filename for downloaded events is el mmddyy or el ddmmyy.
9.4. VIEWING EVENTS
The event log can be viewed on the Model 56 display. Move the cursor the View events tab and press ENTER/MENU.
Move the cursor to the View Events button and press ENTER/MENU.
To scroll through the list of events move the cursor to the DOWN or UP key at the bottom of the screen and press
and hold ENTER/MENU.
9.5. DATE AND TIME
The date and time can also be reset from the main menu by pressing the Time and Date button.
NOTE
Setting the date or time to an earlier value than the one showing will cause data to be lost from
the data/event logger. Download data before resetting time or date. See section 9.3.
34
MODEL TCL-56
SECTION 10
GRAPHICAL DISPLAY
SECTION 10.0
GRAPHICAL DISPLAY
10.1. OVERVIEW
The Model 56 has a dual graphical display. Each graph can be configured to meet user requirements, although
the time axis on both graphs must be the same. The time scale can be one hour, one day, seven days, or 30 days.
10.2. CONFIGURATION
q
q
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
Calibrate
Data storage and retrieval
Program
HART
Hold
Time and date
Display setup
Reset
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Graphics Display setup Tag Language Warning
View graph
q
Upper graph
Variable (y-axis) Measure 1
Lower graph
Measure 2
Y-axis (maximum)
10.00
ppm
0.00
pH
Y-axis (minimum)
0.00
ppm
14.00
pH
X-axis (time)
1 day
q
q
1. With the main display
showing, press
MENU/ENTER. Choose Display setup.
q
q
2. The screen shown at left appears. Configure the displayed variable,
the maximum and minimum values for the y-axis, and the time scale.
To view the graphs, move the cursor to the View graph button and
press ENTER/MENU.
1 day
BACK
q
S1: 1.00 ppm 25.0qC
2
S2: 7.00 pH 25.0qC
S1 ppm
+
0
8
3. The time axis can be expanded or shrunk. To expand the time scale,
use the left or right navigation keys to move the pair of dotted green
lines to the area of interest. Press the up navigation key to expand the
graph. To shrink the time axis, press the down navigation key.
+
S2 pH
7
12/09 03:13
12/09 09:13
12/0915:13
12/09 21:13
12/10 03:13
BACK
q
35
MODEL TCL-56
SECTION 11
DIGITAL COMMUNICATIONS
SECTION 11.
DIGITAL COMMUNICATIONS
The Model 56 analyzer supplied with the TCL has HART communications as a standard
feature and Profibus DP as an option. For more information refer to the Model 56 HART
or Profibus DP Addendum Manuals.
36
MODEL TCL-56
SECTION 12
MAINTENANCE
SECTION 12.
MAINTENANCE
12.1 ANALYZER
The Model 56 analyzer used with the TCL needs little routine maintenance.
Clean the analyzer case and front panel by wiping with a clean soft cloth dampened with water ONLY. Do not use
solvents, like alcohol, that might cause a buildup of static charge.
The chlorine sensor circuit board (PN 24203-01) is replaceable. If you have a dual input analyzer, consult the
Model 56 instruction manual for the part number of the other board.
To replace the board
1.
Turn off power to the analyzer.
2.
Loosen the four screws holding the front panel in place and let the front panel drop down.
3.
Loosen the gland fitting and carefully push the sensor cable up through the fitting as you pull out the circuit
board.
4.
Once you have access to the terminal strip, disconnect the sensor.
5.
Unplug the sensor board from the main board. See Figure 12.1.
6.
Slide the replacement board partially into the board slot. Plug the sensor board into the main board and reattach
the sensor wires.
7.
Carefully pull the sensor cable through the gland fitting as you push the sensor board back into the enclosure.
8.
Close the front panel.
9.
Turn on power.
FIGURE 12-1. Main board showing connections
to sensor board(s).
37
MODEL TCL-56
SECTION 12
MAINTENANCE
12.2 TOTAL CHLORINE SENSOR
12.2.1 General
When used in clean water, the total chlorine sensor requires little maintenance. Generally, the sensor needs maintenance when the response becomes sluggish or noisy or when readings drift follow calibration. Maintenance
frequency is best determined by experience. If the sensor is used in potable water, expect to clean the membrane every month and replace the membrane and electrolyte solution every three months. Sensors used in dirty
water require more frequent maintenance and calibration. However, if experience shows the sensor is holding
calibration and not drifting appreciably between calibration intervals, the maintenance interval can be extended.
12.2.2 Cleaning the membrane.
Keep the membrane clean. Clean the membrane with water sprayed from a wash bottle. Use a soft tissue to
gently wipe the membrane.
12.2.3 Replacing the membrane.
1. Hold the sensor with the membrane facing up.
2. Unscrew membrane retainer. Remove the membrane assembly and O-ring. See Figure 12-2.
3. Inspect the cathode. If it is tarnished, clean it by gently rubbing in the direction of the existing scratches (do
not use a circular motion) with 400-600 grit silicon carbide finishing paper. Rinse the cathode thoroughly with
water.
4. Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. Fill the cup with electrolyte solution. Set aside.
5. Put a new O-ring in the groove.
6
Place a drop of electrolyte solution on the cathode. Invert the membrane assembly and place it over the cathode stem.
7. Screw the membrane retainer back in place.
8. Hold the sensor with the membrane pointing down. Shake the sensor a few times, as though shaking down a
clinical thermometer.
12.2.4 Replacing the membrane and electrolyte solution.
CAUTION
Fill solution may cause irritation. Avoid contact
with skin and eyes. May be harmful if swallowed.
1. Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 12-2.
2. Hold the sensor over a container with the cathode pointing down.
3. Remove the fill plug and allow the electrolyte solution to drain out.
4. Wrap the plug with two turns of pipe tape and set aside. Remove old tape first.
5. Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. Fill the cup with electrolyte solution. Set aside.
6. Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution
(PN 9210438) through the fill hole until the liquid overflows. Tap the sensor near the threads to release trapped
air bubbles. Add more electrolyte solution if necessary.
7. Place the fill plug in the electrolyte port and begin screwing it in. After several threads have engaged, rotate
the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten.
8. Place a new O-ring in the groove around the cathode post. Cover the holes at the base of the cathode stem
with several drops of electrolyte solution.
9. Insert a small blunt probe, like a toothpick with the end cut off, through the pressure equalizing port. See Figure
12-2.
38
MODEL TCL-56
SECTION 12
MAINTENANCE
NOTE
Do not use a sharp probe. It will puncture the bladder and destroy the sensor.
Gently press the probe against the bladder several times to force liquid through the holes at the base of the
cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes
remain covered with electrolyte solution.
10. Place a drop of electrolyte solution on the cathode, then place the membrane assembly over the cathode.
Screw the membrane retainer in place.
11. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte
solution has been replenished.
FIGURE 12-2. Sensor Parts
TABLE 12-1. Spare Parts
33523-00
9550094
33521-00
23501-02
23502-02
9210438
Electrolyte Fill Plug
O-Ring, Viton 2-014
Membrane Retainer
Total Chlorine Membrane Assembly: includes one membrane assembly and one O-ring
Total Chlorine Membrane Kit: includes 3 membrane assemblies and 3 O-rings
Total Chlorine Sensor Fill Solution, 4 oz (120 mL)
39
MODEL TCL-56
SECTION 12
MAINTENANCE
12.3 SAMPLE CONDITIONING SYSTEM
12.3.1 Reagent
The sample conditioning reagent lasts about 2 months. Before putting fresh reagent in the carboy, discard any
small amount of remaining reagent. To prepare the reagent refer to the procedure in Section 5.1. See Table 12-2
for ordering information.
12.3.2 Sample and reagent tubing.
Periodically inspect sample and reagent tubing for cracks and leaks. Replace tubing if it is damaged.
After a period of time, the sample tubing may become plugged with suspended matter. The tubing is flexible and
difficult to clean mechanically. Plugged sample tubing is best replaced.
Replacement tubing kits are available. See Table 12-2 for part numbers.
To replace reagent tubing:
1. Reagent tubing is shown in Figure 12-2.
2. Turn off sample and reagent pumps.
3. Luer fittings connect the reagent tubing to the pump. Disconnect the tubing by turning the fitting in the direction of the arrows shown in Figure 12-5.
4. Disconnect the other end of the suction tubing from the barb on the reagent inlet fitting in the bottom of the
enclosure. Disconnect the other end of discharge tubing from the reagent injection tee.
5. Install the replacement tubing. Note that the discharge tubing is longer than the suction tubing.
FIGURE 12-3. Replacing Reagent Tubing
40
MODEL TCL-56
SECTION 12
MAINTENANCE
To replace sample tubing:
1. Sample tubing and tees are shown in Figure 12-4.
2. Turn off the sample and reagent pumps.
3. Luer fittings connect the sample tubing to the pump. Disconnect the tubing by turning the fitting in the
direction of the arrows shown in Figure 12-5.
4. Disconnect the other end of the sample pump suction tubing from the overflow sampler. Pull the reagent
injection tubing off the reagent injection tee.
5. Disconnect the other end of the sample pump discharge tubing from the flow cell. Pull the air injection tubing
off the air injection tee.
6. Disconnect the sample inlet and drain tubing.
7. Install the replacement sample pump suction and discharge tubing assemblies. The assemblies look similar.
To tell the difference, note the air injection tee in the discharge tubing assembly has a larger diameter barb
than the reagent injection tee in the suction tubing assembly.
8. Install replacement sample inlet and drain tubing. The sample inlet tubing is longer than the drain tubing.
FIGURE 12-4. Replacing Sample Tubing
41
MODEL TCL-56
SECTION 12
MAINTENANCE
12.3.3 Peristaltic pump tubing.
The expected life of the peristaltic pump tubing is one year.
To replace pump tubing:
1. Turn off the sample and reagent pumps.
2. The reagent and sample tubing is connected to the pump tubing
with luer fittings. See Figure 12-5. Disconnect the fittings from
the pump by turning the fitting in the direction of the arrow.
FIGURE 12-5.
3. Using your thumb and forefinger gently pinch the sides of the
pump cover. Slide the cover upwards to remove it. See Figure
12-6.
FIGURE 12-6.
4. Using your thumb as shown in Figure 12-7, push the tubing fitting straight outward until the fitting slides out of the socket.
Repeat the process for the other fitting.
5. Remove and discard the pump tubing.
FIGURE 12-7.
6. Insert the new tubing one end at a time. Tongues on the sides
of the gray fittings at the ends of the tube fit into receiving
grooves in the pump casing. See Figure 12-8. Push the fitting
into place until it clicks. Gently stretch the tubing over the rollers
and insert the other fitting into the receiving socket on the other
side of the pump.
42
FIGURE 12-8.
MODEL TCL-56
SECTION 12
MAINTENANCE
7. Replace the pump cover.
a. Place the cover on the pump casing. See Figure 12-9.
FIGURE 12-9.
b. Be sure the pins at the bottom of the cover (Figure 12-10)
ride on the tracks in the pump casing.
FIGURE 12-10.
c.
The position of the track is outlined in Figure 12-11. The pins
on the pump cover must ride in these tracks as the cover is
pushed into place. Gently squeeze the ends of the cover to
guide the pins.
FIGURE 12-11.
d. Push down until the cover snaps into place.
8. Reconnect the tubing.
43
MODEL TCL-56
SECTION 12
MAINTENANCE
12.3.4 Replacing the air pump
WARNING
1. Disconnect power to the analyzer.
2. Refer to Figure 12-12. Disconnect the reagent
and air injection tubes. Disconnect the suction
and discharge tubing by turning the Luer fitting
in the direction shown in the figure. Disconnect
the air pump inlet tubing from the barbed fitting
in the bottom of the enclosure.
HAZARDOUS
VOLTAGE
CAN CAUSE
SEVERE INJURY
OR DEATH.
DISCONNECT
POWER BEFORE
SERVICING.
9241136/B
3. Remove the four screws (circled in Figure 12-13)
holding the air pump access panel. Pull out the pump and panel.
4. Disconnect the air inlet and outlet tubing from the air pump. See
Figure 12-14.
5. Remove the five screws (surrounded by squares in Figure 12-13)
holding the air pump to the access panel.
6. Remove the four screws holding the wiring access panel.
FIGURE 12-12.
7. Disconnect the air pump power wires from the terminal strip.
See Figure 12-15. Discard the old air pump.
8. Remove the five screws holding the rubber base of the replacement air pump to the body.
9. Using the five screws removed in step 6, attach the replacement
air pump to the access panel.
10. Connect the air pump power wires to the terminal strip.
11. Replace the wiring access panel.
12. Connect the air inlet and outlet tubing to the air pump. See
Figure 12-14. The conical end of the check valve points in the
direction of the air flow.
13. Replace the air pump access panel.
14. Connect the sample pump tubing to the pump. Connect the
reagent and air injection tubing. Connect the air inlet tubing to
the barbed fitting at the bottom of the enclosure.
check
valve
FIGURE 12-13.
Model option -11 115 Vac only
Model option -12 230 Vac only
air outlet
air inlet
FIGURE 12-14.
FIGURE 12-15.
44
MODEL TCL-56
12.3.5 Replacing the air pump diaphragm
and check valves.
1. Disconnect power to the analyzer.
2. Refer to Figure 12-12. Disconnect the
reagent and air injection tubes. Disconnect
the suction and discharge tubing by turning
the Luer fitting in the direction shown in the
figure. Disconnect the air pump inlet tubing
from the barbed fitting in the bottom of the
enclosure.
SECTION 12
MAINTENANCE
WARNING
HAZARDOUS
VOLTAGE
CAN CAUSE
SEVERE INJURY
OR DEATH.
DISCONNECT
POWER BEFORE
SERVICING.
9241136/B
3. Remove the four screws (circled in Figure 12-13) holding the air
pump access panel. Pull out the pump and panel.
FIGURE 12-16.
4. Disconnect the air inlet and outlet tubing from the air pump. See
Figure 12-14.
5. Remove the five screws (surrounded by squares in Figure 12-13)
holding the air pump to the access panel.
6. Pull the rubber base off the pump.
7. Using needle nose pliers, remove the air inlet fitting from the
side of the air pump. See Figure 12-16.
8. Slide the pump assembly out of the air pump body. See Figure
12-17.
9. Following instructions on the package (PN 9160518), replace
the diaphragm and check valves.
FIGURE 12-17.
10. Slide the pump assembly back into the pump body and replace
the barbed inlet fitting.
11. Replace the rubber base and screw the pump access panel
back onto the air pump.
12. Connect the air inlet and outlet tubing to the air pump. See
Figure 12-14. The conical end of the check valve points in the
direction of the air flow.
13. Replace the air pump access panel.
14. Connect the sample pump tubing to the pump. Connect the
reagent and air injection tubing. Connect the air inlet tubing to
the barbed fitting at the bottom of the enclosure.
45
MODEL TCL-56
SECTION 12
MAINTENANCE
TABLE 12-2. Replacement Parts and Reagent for Sample Conditioning System
PN
24134-00
Air pump, 115 Vac, 60 Hz
24134-01
Air pump, 230 Vac, 50 Hz
9160578
Air pump repair kit
9322052
Check valve for air injection line
24153-00
Carboy for reagent, 5 gal/19 L, includes cap
9100204
Fuse, 0.25 A, 250 V, 3AG, slow blow for option -11 (115 Vac)
9100132
Fuse, 0.125 A, 250 V, 3AG, slow blow for option -12 (230 Vac)
9380094
Reagent pump, 115 Vac, 50/60 Hz
9380095
Reagent pump, 230 Vac, 50/60 Hz
9380091
Reagent pump replacement tubing
24151-00
Reagent tubing replacement kit (see Section 12.3.2)
24135-00
Reagent uptake tubing, 6 ft (1.8 m), includes weight
9380090
Sample pump, 115 Vac, 50/60 Hz
9380093
Sample pump, 230 Vac, 50/60 Hz
9380092
Sample pump replacement tubing
24152-00
Sample tubing replacement kit (includes tees, see Section 12.3.2)
PN
46
Description
Description
24165-00
Acetic acid, 2 x 2.5 gal (9.5 L) bottles/case, with 25 g potassium iodide
24165-01
Acetic acid, 2 x 2.5 gal (9.5 L) bottles/case, with 50 g potassium iodide
24164-00
Potassium iodide, 25 g, sufficient for 5 gallons (19 L) of vinegar (for 0-5 ppm total chlorine)
24164-01
Potassium iodide, 50 g, sufficient for 5 gallons (19 L) of vinegar (for 0-10 ppm total chlorine)
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
SECTION 13.
TROUBLESHOOTING
13.1 OVERVIEW
The analyzer continuously monitors itself and the sensor(s) for problems. When the analyzer identifies a problem,
the word warning or fault appears intermittently at the bottom of the display. To read the fault or warning message
and troubleshooting information, press INFO. See section 13.2.
A warning means the instrument or sensor is usable, but steps should be taken as soon as possible to correct the
condition causing the warning. Warning messages can be turned off. To turn off warning messages, go to the main
menu and choose Display setup. Scroll to the Warning tab and turn off warning messages.
A fault means the measurement is seriously in error and is not to be trusted. A fault condition might also mean that
the analyzer has failed. Fault conditions should be corrected immediately. When a fault occurs, the analog output
goes to 22.00 mA or to the value programmed in Section 7.2. Fault messages cannot be turned off.
13.2 READING AND TROUBLESHOOTING FAULT AND WARNING MESSAGES
q
S1: 1.00 ppm 25.0qC
q
S2: 7.00 pH 25.0qC
Faults – fix now
Outputs information
Warnings – fix soon
Analyzer information
Sensor 1 information
HART information
1. With the main display showing, press the INFO key. The screen at left
appears. Move the cursor to the appropriate button and press
ENTER/MENU.
Sensor 2 information
BACK
S1: 1.00 ppm 25.0qC
Warn: S2 Out of Range
q
Press INFO for troubleshooting.
S2: 7.00 pH 25.0qC
2. A screen like the one at left will appear showing all the warning or fault
messages. For troubleshooting information press the INFO key
q
BACK
47
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
13.3 SENSOR DIAGNOSTICS
Sensor diagnostic readings are often useful in troubleshooting measurement problems.
S1: 1.00 ppm 25.0qC
S2: 7.00 pH 25.0qC
Faults – fix now
Outputs information
Warnings – fix soon
Analyzer information
Sensor 1 information
HART information
1. With the main display showing, press the INFO key. Move the cursor to
the Sensor 1 or Sensor 2 information button and press ENTER/MENU.
Sensor 2 information
BACK
2. A list of sensor diagnostics will appear. For more information about a specific
diagnostic measurement, move the cursor to the diagnostic of interest and
press the INFO key.
13.4 TROUBLESHOOTING CALIBRATION PROBLEMS
If a calibration attempt results in an error or a likely error, the analyzer will display the appropriate warning screen.
For troubleshooting suggestions, press the INFO key.
13.5 OTHER TROUBLESHOOTING
Problem
See Section
Readings are too low
11.5.1
Process readings are erratic or wander
11.5.2
Readings drift
11.5.3
Readings are too high
11.5.4
13.5.1 Readings are low
1. Does the reagent carboy contain reagent? Is the reagent uptake tubing below the level of the reagent? Has
potassium iodide been added to the acetic acid (vinegar) reagent?
2. Is there adequate flow to the overflow sampler? Excess sample should be flowing down the inside tube of the
overflow sampler.
3. Does the reagent contain the correct amount of potassium iodide? See the table.
Expected range,
ppm as Cl2
Amount of KI needed
per 5 gallons of vinegar
Part number
0 – 5 ppm
25 grams
24164-00
0 – 10 ppm
50 grams
24164-01
0 – 20 ppm
2 x 50 grams
24164-01
4. Was the comparison or calibration sample tested as soon as it was taken? Chlorine solutions can be unstable.
Test the sample immediately after collecting it. Avoid exposing the sample to sunlight.
5. Is the membrane fouled or coated? A dirty membrane inhibits diffusion of iodine through the membrane, reducing
sensor current. Clean the membrane by rinsing it with a stream of water from a wash bottle. Wipe gently with a
soft tissue.
48
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
6. Are the reagent and sample pumps running? If a pump is not running, check the fuse and replace it if necessary. See Table 12-2 for part numbers. If the fuse is okay, replace the pump.
7. Are all tube fittings tight? Pay particular attention to the luer fittings that connect the tubing to the pumps.
8. Does the pump tubing element need replacing? Remove the tubing from the pump and inspect it. If the tubing
appears permanently pinched or deformed, replace the tubing. Refer to Section 12.3.3 for instructions on how
to remove and replace the tubing elements. The expected life of a tubing element is about one year.
9. Is the sample flow to the sensor about 11 mL/min? If the sample flow is too low, the total chlorine reading will
be low. If the flow is too high, the ratio between the sample flow and reagent flow will be too high, and there
might be insufficient reagent to properly react with the total chlorine in the sample. To check sample flow…
a. Turn off the reagent and sample pumps.
b. Disconnect the luer fitting on the discharge of the sample pump. See A in Figure 13-1.
c.
Hold a small beaker under the discharge port.
d. Start the sample pump and collect sample for two minutes.
e. Measure the volume of sample collected in the beaker. After two minutes, the volume should be about
22 mL.
10. Is the reagent flow about 0.2 mL/min? If the reagent flow is too low, there might be insufficient acetic acid to
lower the sample pH and insufficient potassium iodide to react with total chlorine in the sample. To check
reagent flow…
a. Turn off the reagent and sample pumps.
b. Disconnect the reagent tubing at the injection tee. See B in Figure 13-1.
c.
Place the end of the tubing in a 5 mL graduated cylinder.
d. Start the reagent pump and collect reagent for ten minutes.
e. Note the volume of reagent collected in the graduated cylinder. After ten minutes the volume should be
about 2 mL.
FIGURE 13-1. Disconnecting sample (A) and reagent (B) tubing prior to checking flow.
49
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
13.5.2 Process readings are erratic or wander
1. Is the sensor properly wired to the analyzer? See Section 4.3. Verify that all connections are tight.
2. Readings can be erratic when a new sensor is first placed in service. Readings usually stabilize after about an
hour.
3. Is the air pump working? There should be a vigorous stream of bubbles in the flow cell. The bubbles help mix
the sample and keep carbon dioxide bubbles off the membrane. Carbon dioxide forms when bicarbonate alkalinity in the sample reacts with acetic acid. The bubbles accumulate on the membrane and eventually break
away, causing the total chlorine reading to wander.
4. Is the membrane damaged or loose? Replace the membrane if necessary.
5. Is the space between the membrane and cathode filled with electrolyte solution and is the flow path between
the electrolyte reservoir and membrane clear? Often the flow of electrolyte can be started by simply holding
the sensor with the membrane end pointing down and sharply shaking the sensor a few times as though shaking down a clinical thermometer. If shaking does not work, try clearing the holes around the cathode stem. Hold
the sensor with the membraned end pointing up. Unscrew the membrane retainer and remove the membrane
assembly. Use the end of a straightened paper clip to clear the holes at the base of the cathode stem.
13.5.3 Readings drift
1. Is the sample temperature changing? Membrane permeability is a function of temperature. The time constant
for the 499ACL-01 sensor is about five minutes. Therefore the reading may drift for a while after a sudden temperature change.
2. Is the membrane clean? For the sensor to work properly, iodine must diffuse freely through the membrane. A
coating on the membrane will interfere with the passage of iodine, resulting in a gradual downward drift in readings. The coating will also slow the response on the sensor to step changes. Clean the membrane by rinsing
is with a stream of water from a wash bottle. Wipe the membrane with a soft tissue.
3. Is the sensor new or has it recently been serviced? New or rebuilt sensors may require several hours to stabilize.
4. Is the flow of sample past the sensor about 11 mL/min? See Section 13.5.1 step 9 for more information.
5. Is the reagent flow about 0.2 mL/min? See Section 13.5.1 step 10 for more information.
13.5.4 Readings are too high
1. Is the sample conditioning reagent clear and colorless? If the reagent is pale yellow, results will be high. The
pale yellow color is caused by iodine, which comes from the reaction between atmospheric oxygen and potassium iodide. The reaction is catalyzed by sunlight. The purpose of the blue carboy is to protect the reagent
from sunlight.
2. Is the sensor fill solution fresh? An old, discolored fill solution may produce a high reading.
13.6 OTHER TROUBLESHOOTING — GENERAL
Problem
See Section
Current output is too low
13.6.1
Alarm relays do not operate properly
13.6.2
13.6.1 Current Output Is Too Low.
Load resistance is too high. Maximum is 550Ω.
13.6.2 Alarm Relays Do Not Operate Properly
1. Verify the relays are properly wired.
2. Verify the relays are properly configured.
50
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
13.7 SIMULATING INPUTS — CHLORINE
To check the performance of the analyzer, use a decade box and
1.5V battery to simulate the current from the sensor. The battery, which
opposes the polarizing voltage, is necessary to ensure that the sensor
current has the correct sign.
1. Disconnect the anode and cathode leads from terminals 8 and 10
on TB1 and connect a decade box and 1.5V battery as shown in
Figure 13-1. It is not necessary to disconnect the RTD leads.
2. Set the decade box to 1.4MΩ.
R
7
ANOD SHLD
8
ANOD
9
CATH SHLD
10 CATH
3. Note the sensor current. It should be about 960 nA. The actual
value depends on the voltage of the battery. To view the sensor
FIGURE 13-2. Simulating Chlorine
current, go to the main display and press INFO. Choose sensor 1
information. The input current is the second line in the display.
Change the decade box resistance and verify that the correct current is shown. Calculate current from the
equation:
Vbattery — 250 (voltages in mV)
current (nA) =
resistance (MΩ)
The voltage of a fresh 1.5 volt battery is about 1.6 volt (1600 mV).
51
MODEL TCL-56
SECTION 13
TROUBLESHOOTING
13.8 SIMULATING INPUTS —
TEMPERATURE
13.8.1 General.
The Model 56 accepts a Pt100 RTD. The Pt100 RTD is
in a three-wire configuration. See Figure 13-3.
13.8.2 Simulating temperature
To simulate the temperature input, wire a decade box to
the analyzer as shown in Figure 13-4.
To check the accuracy of the temperature measurement, set the resistor simulating the RTD to the values
indicated in the table and note the temperature readings. The measured temperature might not agree with
the value in the table. During sensor calibration an offset
might have been applied to make the measured temperature agree with a standard thermometer. The offset
is also applied to the simulated resistance. The analyzer
is measuring temperature correctly if the difference
between measured temperatures equals the difference
between the values in the table to within ±0.1°C.
FIGURE 13-3. Three-Wire RTD Configuration.
Although only two wires are required to connect
the RTD to the analyzer, using a third (and sometimes fourth) wire allows the analyzer to correct
for the resistance of the lead wires and for
changes in the lead wire resistance caused by
temperature changes.
For example, start with a simulated resistance of 103.9 Ω,
which corresponds to 10.0°C. Assume the offset from the
sensor calibration was -0.3 Ω. Because of the offset, the
analyzer calculates temperature using 103.6 Ω. The
result is 9.2°C. Now change the resistance to 107.8 Ω,
which corresponds to 20.0°C. The analyzer uses 107.5 Ω
to calculate the temperature, so the display reads
19.2°C. Because the difference between the displayed
temperatures (10.0°C) is the same as the difference
between the simulated temperatures, the analyzer is
working correctly.
FIGURE 13-4. Simulating RTD Inputs.
Temp. (°C)
0
10
20
25
30
40
50
60
70
80
85
90
100
52
Pt 100 (Ω)
100.0
103.9
107.8
109.7
111.7
115.5
119.4
123.2
127.1
130.9
132.8
134.7
138.5
MODEL TCL-56
SECTION 14
RETURN OF MATERIAL
SECTION 14.
RETURN OF MATERIAL
14.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)
number.
14.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Emerson Process Management, Liquid Division
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
14.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 11.2.
NOTE
Consult the factory for additional information regarding service or repair.
53
MODEL TCL-56
54
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90
days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1.
2.
3.
4.
5.
Location type of service, and length of time of service of the device.
Description of the faulty operation of the device and the circumstances of the failure.
Name and telephone number of the person to contact if there are questions about the returned material.
Statement as to whether warranty or non-warranty service is requested.
Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
8
Credit Cards for U.S. Purchases Only.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011