HACH pH ELECTRODE CLEANING & MAINTENANCE GUIDE
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
pH ELECTRODE CLEANING & MAINTENANCE GUIDE
Throughout the life of your pH electrode, you want to make sure you are getting the best performance and most accurate
results possible. However minor items like air bubbles, crystallization, low electrolyte solution, KCl leakage and minor
contamination can cause issues. Issues which can result in a lower confidence in the results being provided and additional
purchases of replacement electrodes. This guide will provide tips to help you maintain and keep your pH electrodes clean,
extending your electrode’s life, while allowing you to maintain a high level of confidence in your readings.
1. First Use & Conditioning
3. Regular Maintenance
pH electrodes are
2
typically shipped with
a storage cap to
protect and hydrate
1
the glass bulb (1).
Refillable pH
electrodes are
additionally shipped
with the refill hole
sealed to prevent
electrolyte leakage during shipment. This is often a piece of tape which
must be removed before first use (2). During shipment, the electrolyte
can move or shift, sometimes allowing air bubbles into the glass bulb,
or drying of the reference element.
Because of this, it is recommended that the pH
electrode be conditioned before first use. After
unpacking, check for sufficient electrolyte inside
the electrode and refill if necessary (for refillable
electrodes). Then check that the glass bulb is
completely filled with electrolyte and that there
are no visible air bubbles. If there are, shake the
electrode in a downward motion to force air
bubbles out of the glass bulb.
Air in the
pH Glass Bulb
The pH electrode will then need to be conditioned per the electrode’s
user manual (typically this requires soaking the electrode for several
minutes in the sample or pH buffers). The response time for a new,
conditioned pH electrode in pH buffers at 25°C is typically less than
30 seconds.
2. Closing & Opening the KCl Refill Hole
Refillable pH electrodes have a refill hole where
electrolyte can be added. The level or height of
the inner liquid is important for a refillable pH
electrode junction.
If sufficient electrolyte solution is inside the
electrode, the hydrostatic pressure of the liquid
column will allow electrolyte to flow out of the
electrode at a faster flow rate, and additionally
will prevent sample solution from going inside
the electrode.
Electrode Refill
Hole Stopper
• Slow stabilization times
• Erroneous or erratic readings
• Difficulty in calibrating
Proper maintenance will ensure
faster measurements, improve pH
accuracy and extend the lifetime of
the electrode.
Regular pH electrode maintenance
requires that a pH electrode be
stored in the recommended pH
storage solution between
measurements, and that the
electrolyte filling solution be refilled
as necessary. For best electrode
performance, do not let the
reference junction dry out.
The pH electrode will need to be
cleaned regularly, depending on
the nature of the sample. The most
suitable cleaning solution is one
which interacts in a more selective
way with the contamination.
Typically fats, oils and grease
(FOG) contaminations should be
cleaned with a non-ionic surfactant
solution or methanol, proteins such
as food stuff should be cleaned
with an acidic pepsin cleaning
solution and mineral deposits
should be cleaned with an acidic
cleaning solution. Refer to your
user manual for your electrode’s
specific guidelines; however,
typical cleaning/maintenance
guidelines are as follows:
Electrode in Beaker
Rinse Electrode
Proteins: Cleaned with an acidic
pepsin cleaning solution.
Typically, the faster the electrolyte flow rate, the
faster the pH stabilization time. However this is
dependent on the junction type used. Infiltration
of the sample solution can contaminate the
reference element and cause unstable or
unreliable measurements.
Fats, oils: Cleaned with a non-ionic
surfactant solution or methanol.
Open the refill hole plug during use and close it
Refillable pH Electrode
when not in use.
with Stopper
hach.com/HQdGuide-pH
There are some key signs that will
help the electrode user understand
when to clean.
Afterward, rinse the electrode with
deionized water, then store in the
recommended pH storage
solution.
Hach Electrode
Cleaning Solution
4. Regular Cleaning of the Glass Bulb
and Reference Junction
For optimal response time, it may be necessary to clean the pH
electrode’s glass bulb and reference junction of contaminants and
precipitations. The pH electrode will need to be cleaned regularly
following the steps outlined in the pH electrode manual. Typically,
it is recommended to soak the electrode for a few minutes in warm
deionized water, or in a cleaning solution (Hach Electrode Cleaning
Solution) to maintain the reference junction.
Correctly Functioning Ceramic Junction
Outflow of Electrolyte (Red Dye)
6. Exchange of Electrolyte Filling Solution
If the electrolyte filling
solution is contaminated,
or if a different electrolyte is
needed, use a syringe with
needle to empty the entire
fluid contents of the
electrode. Withdraw the filling
solution slowly and carefully
to ensure that the pH
electrode is not damaged
during this process. Refer to
the pH electrode manual to
determine the correct
replacement filling solution to
use to refill the electrode.
Contaminated
Reference Junction
KCl
Electrode in Beaker
Rinse Electrode
Hach Electrode
Cleaning Solution
5. Refilling pH Electrode Electrolyte
Refer to the specific pH electrode manual
to determine the correct electrode filling
solution to use. Remove the filling hole
plug, and fill the pH electrode with
electrolyte to just below the refill hole.
Leave a small amount of space below
the refill hole to help minimize electrolyte
leakage and build up of KCl crystallization
at the filling hole. Replace the filling hole
plug (or leave open if next step is to take
a measurement).
KCl w/Ag
Sat. KCl
KCl w/Ag
Sat. KCl
7. Cleaning ISFET pH Sensors
Refill with KCl
Using Toothbrush to Clean an ISFET pH Sensor
ISFET pH sensors are robust and can be
cleaned with a toothbrush, even on the
ISFET surface. Do NOT use sharp tools,
abrasive materials, or cleaning powder.
A mild soap or detergent can be used to
help remove oil or fat. Pepsin cleaning
solution should routinely be used for
samples that contain proteins. After
cleaning, rinse completely with deionized
water.
Refill with KCl
KCl
Remove Fluid Contents
of Electrode
Don't forget to clean the reference
junction as well. Since ISFET electrodes
are stored dry and are commonly exposed
Clean ISFET pH Sensor
to food stuff, they commonly get KCl
crystallization (see Tip 13) and dirty reference junction (see Tip 4).
Make sure that the electrode cap is always placed on the electrode tip
when not in use. This will create a humid environment to prevent the
reference junction from drying out and clogging.
8. Air Bubble in Glass Bulb
10. Contamination Inside the Electrode
The electrolyte inside a pH electrode can shift during shipment,
during horizontal storage, and from general use. This movement can
allow air bubbles into the glass bulb causing erratic readings or
calibration errors. Before each use, it is recommended to inspect
that the glass bulb is completely filled with electrolyte and that there
are no visible air bubbles.
If the pH electrode has visible air bubbles, shake the electrode in a
downward motion like a thermometer. This will force air bubbles out
of the glass bulb.
Some samples can cause unintended
biological growth inside the pH
electrode. This contamination affects
the performance of the pH electrode.
Thiourea solution (example: Hach
KS410 Solution) can remove this type
of organic growth. Allow the electrode
to soak for a few hours in thiourea,
followed by a thorough rinsing with
deionized water.
Dirty Gel Electrolyte in
Hach IntelliCAL PHC101
pH Electrode
Air in the pH Glass Bulb
9. KCl Leakage & Salt Build-Up
Shaking the Electrode
Electrode in Beaker
Salt crystals can accumulate when water in a salt solution evaporates
during contact with air (1) (2). As a result, liquid junctions of the pH
electrode can occasionally become clogged (3) if not stored properly.
The salt crystallization is not harmful to the electrode, nor does it
inhibit the electrode’s performance if treated properly. These salt
crystals can be removed by rinsing the area with warm water (4). Salt
crystals can also be minimized by properly covering the reference
junction(s) with storage solution when the electrode is not in use,
along with maintaining the electrolyte level in refillable electrode (5).
(1) KCl salt
crystallization
(2) KCl salt
crystallization
(3) KCl salt
crystallization
11. Dirty Glass Bulb
Dirty samples or samples that remain on the glass surface can
result in incorrect pH readings. Refer to your user manual for
specific guidelines for the most suitable cleaning solution for the
specific contamination you have; however, a typical cleaning
procedure for a dirty glass bulb is as follows:
General contaminants: Soak the glass bulb for up to 16 hours in
Hach Electrode Cleaning Solution. Rinse in deionized water. Soak
the electrode in pH 4 buffer for up to 20 minutes, then rinse with
deionized water.
Dirty Glass pH Bulb
Electrode in Beaker
(4) Rinse Electrode
(5) Refill Electrode with KCl
Hach KS410
Solution
Clean Gel Electrolyte in
Hach IntelliCAL PHC101
pH Electrode
Dirty Glass pH Bulb
Rinse Electrode
Dirty Glass pH Bulb
Hach Electrode
Cleaning Solution
12. Sulfide Precipitation
Sulfide and silver ions can react to form a dark precipitation inside of refillable pH electrodes.
This precipitate can block the reference junction. Soak the pH electrode in thiourea solution
(example: Hach KS410 Solution) for a few minutes to remove the blockage.
Ceramic Pin Junction
Blocked by Sulfide
Precipitation
Ceramic Pin Junction
Blocked by Sulfide
Precipitation
Ceramic Pin Junction
Cleared of Sulfide
Precipitation by Hach
KS410 Solution
Ceramic Pin Junction
Cleared of Sulfide
Precipitation by Hach
KS410 Solution
Electrode in Beaker
13. KCl Crystallization Inside the Electrode
Hach KS410
Solution
The most common electrolyte in Hach pH electrodes is 3 molar KCl. In some instances, such
as low temperature, the KCl may crystallize inside the electrode. If the KCl crystals block the
ion exchange with the reference element the electrode may read incorrect pH values.
Soak the electrode in a warm (45ºC) KCl filling solution or buffer. Allow the electrode to cool
down to room temperature. This should cause the KCl crystals to dissolve.
KCl Crystals Blocking
Reference Junction
Electrode in Beaker
Hach Solutions for Your pH Electrodes
2965249
Electrode Cleaning Solution (500 mL)
2756549
Fill or Storage Solution, 3M KCl (500 mL)
2964449
Electrode Rinse Solution, Non-Ionic Surfactant (500 mL)
2841700
Fill Solution, 3M KCl Saturated w/ AgCl (30 mL)
2975149
Acid Electrode Cleaning Solution (500 mL)
2965026
Fill Solution, 2.44M Viscous KCl (59 mL)
For more information on Hach’s pH electrodes, please visit us at: hach.com/HQdGuide-pH
To order any needed pH electrode solutions, please call Hach at: 800-227-4224
LIT2781 Rev 1
C14 Printed in U.S.A.
©Hach Company, 2014. All rights reserved.