Yokogawa Model ZR22S, ZR402G Separate type Explosion-proof Zirconia Oxygen Analyzer User's Manual
User’s
Manual
Model ZR22S, ZR402G
Separate type Explosion-proof
Zirconia Oxygen Analyzer
IM 11M13A01-02E
IM 11M13A01-02E
6th Edition
i
Introduction
Thank you for purchasing the ZR22S, ZR402G Separate type Explosion-proof Zirconia Oxygen
Analyzer.
Please read the following respective documents before installing and using the ZR22S, ZR402G
Separate type Explosion-proof Zirconia Oxygen Analyzer.
The related documents are as follows.
General Specifications
Contents
Model ZR22S and ZR202S Explosion-proof Direct In Situ
Zirconia Oxygen Analyzers
Document number
Note
GS 11M13A01-01E
* the “E” in the document number is the language code.
User’s Manual
Contents
Model ZR22S, ZR402G Separate type Explosion-proof
Zirconia Oxygen Analyzer
Model ZR22A, ZR202A
Heater Assembly
Model EXAxt ZR Series
HART Protocol
Document number
IM 11M13A01-02E
Note
(This manual)
IM 11M12A01-21E
IM 11M12A01-51E
* the “E” in the document number is the language code.
An exclusive User’s Manual might be attached to the products whose suffix codes or option
codes contain the code “Z” (made to customers' specifications). Please read it along with this
manual.
The EXAxt ZR Separate type Explosion-proof Zirconia Oxygen Analyzer has been developed for
combustion control in various industrial processes. This analyzer basically consists of a detector
and a converter. You can select between several versions, based upon your application.
Optional accessories are also available to improve measurement accuracy and automate
calibration. An optimal control system can be realized by adding appropriate options.
This instruction manual refers to almost all of the equipment related to the EXAxt ZR.
You may skip any section(s) on the equipment which is not included in your system.
Regarding the HART communication protocol, refer to IM 11M12A01-51E.
IM 11M12A01-51E has been published as " Model EXAxt ZR Series HART Protocol" .
The all-in-one version (with sensor and analyzer integrated in one body) is described in
IM 11M12A01-04E
Models and descriptions in this manual
Model
Product Name
Description in this manual
Specification Installation
Operation Maintenance CMPL
ZR22S
General-purpose detector
ZR22S
High temperature detector
(0.15 m)
ZR402G
Converter
ZO21R
Probe protector
ZO21P
High temperature probe
adapter
Media No. IM 11M13A01-02E
6th Edition : June 2017 (YK)
All Rights Reserved Copyright © 2005, Yokogawa Electric Corporation
IM 11M13A01-02E
ii
ZA8F
ZR40H
Flow setting unit
(for manual calibration use)
Automatic Calibration unit
-
Auxiliary ejector assembly for
high temperature use
(Part No. E7046EC, E7046EN)
-
Calibration gas unit case
(Part No. E7044KF)
-
Check valve
(Part No. K9292DN, K9292DS)
Standard gas unit
ZO21S
CMPL : Customer Maintenance Parts List
This manual consists of twelve chapters. Please refer to the reference chapters for installation,
operation and maintenance.
Table of Contents
Chapter
Outline
Relates to
Installation Operation Maintenance
1. Overview
Equipment models and system configuration
examples
B
C
B
2. Specifications
Standard specification, model code (or
part number), dimension drawing for each
equipment
A
B
B
3. Installation
Installation method for each equipment
A
C
4. Piping
Examples of piping in two standard system
configurations
A
C
5. Wiring
Wiring procedures such as “Power supply
wiring”, “output signal wiring” or others
A
C
6. Components and
Their Functions
Major parts and function are described in this
manual
C
7. Startup
B
B
Basic procedure to start operation of EXAxt
ZR.
Chapter 7 enables you to operate the
equipment immediately.
A
C
8. Detailed Data Setting
Details of key operations and displays
B
C
9. Calibration
Describes the calibration procedure required in
the course of operation.
B
C
10. Other Functions
Other functions described
B
C
11. Inspection and
Maintenance
How to conduct maintenance of EXAxt ZR and
procedures for replacement of deteriorated
parts
B
A
12. Troubleshooting
This chapter describes measures to be taken
when an abnormal condition occurs.
C
A
CMPL (parts list)
User replaceable parts list
C
B
A : Read and completely understand before operating the equipment.
B : Read before operating the equipment, and refer to it whenever necessary.
C : Recommended to read it at least once.
IM 11M13A01-02E
iii
n ATEX Documentation
This procedure is only applicable to the countries in European Union.
GB
DK
SK
CZ
I
LT
E
LV
NL
EST
PL
SF
SLO
P
H
F
BG
D
RO
S
M
GR
IM 11M13A01-02E
iv
n Precautions in Handling Explosion-proof Zirconia Oxygen Analyzer
The explosion-proof zirconia oxygen analyzer (Model ZR22S) are designed as explosion-proof
instruments.
When using either of these instruments in an explosion-susceptible hazardous area, note the
following and observe the given precautions:
Use only the supplied, the explosion-proof zirconia oxygen analyzer (Model ZR22S) and
accessories, or any explosion-proof certification may be invalidated.
For the details, refer to the system configurations in the manual.
CAUTION
Only trained persons use this instrument in industrial locations.
Explosion-proof Approval followings:
ZR22S-A (ATEX); Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
ZR22S-B (FM); Class I, Division 1, Groups B, C and D, Class II/III, Division 1,
Groups E, F and G, T2
ZR22S-C (CSA); Class I, Division 1, Groups B, C and D, Class II/III, Division 1,
Groups E, F and G, T2
ZR22S-D (IECEx); Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
n For the safe use of this equipment
WARNING
EXAxtZR is very heavy. Handle it with care. Do not drop it. Handle safely to avoid injury.
Connect the power supply cord only after confirming that the supply voltage matches the rating of
this equipment. In addition, confirm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line
for maintenance or other reasons, protect yourself from potential poisoning by using a protective
mask or ventilating the area well.
CAUTION
Requirements for explosion-proof use:
The ZR22S is connected to a ZR402G or AV550G*1 that is mounted in a non-hazardous area.
The ambient temperature is in the range -20 to +60°C. The surface temperature of the ZR22S is
not over the temperature class T2 (300°C)*2.
*1: Refer to IM 11M12D01-01E
*2: The terminal box temperature does not exceed 150ºC
Oxygen concentration of sample/reference /calibration gas shall not exceed that found in normal
air, typically 21 vol%.
IM 11M13A01-02E
v
NOTE
The cell (sensor) at the tip of the detector is made of ceramic (zirconia element). Do not drop
the detector or subject it to pressure stress.
• Do NOT allow the sensor (probe tip) to make contact with anything when installing the
detector.
• Avoid any water dropping directly on the probe (sensor) of the detector when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that there
is no leakage of the gas. If there is any leakage of the gas, the moisture drawn from the
sample gas may damage the sensor.
• The detector (especially at the tip) becomes very hot. Be sure to handle it with gloves.
(1) About This Manual
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without
permission.
• This manual explains the functions contained in this product, but does not warrant that those will
suit the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, should any errors or omissions come to the attention of the user, please contact the
nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may not be changed on any
change of specification, construction and parts when the change does not affect the functions or
performance of the product.
• If the product is used in a manner not specified in this manual, the safety of this product may be
impaired.
NOTE
This instrument is tested and certificated as explosion-proof type. Please note that the
construction of the instrument, installation, external wiring, maintenance or repair is strictly
restricted, and non-observation or negligence of this restriction would result in dangerous
condition.
(2) Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protection and
safety of personnel, product and system containing the product.
IM 11M13A01-02E
vi
(3) The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that the operator must follow the instructions laid out in this manual in
order to avoid the risk of personnel injury electric shock, or fatalities. The manual describes what
special care the operator must exercise to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
NOTE
This symbol draws attention to information essential for understanding the operation and
functions.
Protective Ground Terminal
Function Ground Terminal (Do not use this terminal as the protective
ground terminal.)
Alternating current
• Special descriptions in this manual
This manual indicates operation keys, displays and drawings on the product as follows:
• Operation keys, Enclosed in [ ], displays on the panel “ “.
(Ex. [MODE] key)
(Ex. selection display
(Ex. data display
“ BASE “)
“ 102” lit, “ 102” flashing)
• Drawing for flashing
Indicated in light print.
(Flashing)
(lit)
IM 11M13A01-02E
vii
n NOTICE
l Specification check
When the instrument arrives, unpack the package with care and check that the instrument has
not been damaged during transportation. In addition, please check that the specification matches
the order, and required accessories are not missing. Specifications can be checked by the model
codes on the nameplate. Refer to Chapter 2 Specifications for the list of model codes.
l Details on operation parameters
When the EXAxt ZR Separate type Oxygen Analyzer arrives at the user site, it will operate based
on the operation parameters (initial data) set before shipping from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting analysis.
Where necessary, set the instrument parameters for appropriate operation.
For details of setting data, refer to chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the changed
setting data.
l Product Disposal:
The instrument should be disposed of in accordance with local and national legislation/
regulations.
n Trademark Acknowledgments
• All other company and product names mentioned in this user’s manual are trademarks or
registered trademarks of their respective companies.
• We do not use TM or ® mark to indicate those trademarks or registered trademarks in this
user’s manual.
IM 11M13A01-02E
viii
CE marking products
u
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
n Identification Tag
This manual and the identification tag attached on packing box are essential parts of the product.
Keep them together in a safe place for future reference.
n Users
This product is designed to be used by a person with specialized knowledge.
l How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office. Do not
dispose them as domestic household waste.
Battery type: Manganese dioxide lithium battery
Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in
DIRECTIVE 2006/66/EC.
IM 11M13A01-02E
ix
u
After-Sales Warranty
l Do not modify the product.
l Yokogawa warrants the product for the period stated in the pre-purchase
quotation. Yokogawa shall conduct defined warranty service based on its
standard.
l During the warranty period, for repair under warranty carry or send the
product to the local sales representative or service office. Yokogawa will
replace or repair any damaged parts and return the product to you.
• Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
• If we replace the product with a new one, we won’t provide you with a repair report.
l In the following cases, customer will be charged repair fee regardless of
warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa did
not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
l Yokogawa does not warrant conformance with the specific application at the
user site. Yokogawa will not bear direct/indirect responsibility for damage due
to a specific application.
l Yokogawa will not bear responsibility when the user configures the product
into systems or resells the product.
l Maintenance service and supplying repair parts will be covered for five years
after the production ends. For repair for this product, please contact the
nearest sales office described in this instruction manual.
IM 11M13A01-02E
Blank Page
Toc-1
Model ZR22S, ZR402G
Separate type Explosion-proof
Zirconia Oxygen Analyzer
IM 11M13A01-02E 6th Edition
CONTENTS
Introduction...............................................................................................................i
u
CE marking products.................................................................................viii
u
After-Sales Warranty....................................................................................ix
1. Overview..................................................................................................... 1-1
1.1
1.2
2.
< EXAxt ZR > System Configuration................................................................ 1-1
1.1.1
System 1............................................................................................. 1-1
1.1.2
System 2............................................................................................. 1-2
< EXAxt ZR > System Components................................................................. 1-3
1.2.1
System Components.......................................................................... 1-3
1.2.2
Detectors and Accessories................................................................. 1-3
Specifications............................................................................................ 2-1
2.1
General Specifications...................................................................................... 2-1
2.2
Separate type Explosion-proof Detector and Related Equipment............... 2-3
2.3
2.4
2.5
2.2.1
ZR22S Separate type Explosion-proof Detector................................ 2-3
2.2.2
ZO21R Probe Protector...................................................................... 2-8
Separate type Explosion-proof Detector for High Temperature and Related
Equipment........................................................................................................... 2-9
2.3.1
ZR22S (0.15 m) Separate type Explosion-proof Detector for High
Temperature........................................................................................ 2-9
2.3.2
ZO21P High Temperature Probe Adapter ....................................... 2-10
ZR402G Separate type Converter.................................................................. 2-12
2.4.1
Standard Specification...................................................................... 2-12
2.4.2
Functions.......................................................................................... 2-13
ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit................. 2-17
2.5.1
ZA8F Flow Setting Unit..................................................................... 2-17
2.5.2
ZR40H Automatic Calibration Unit.................................................... 2-19
2.6
ZO21S Standard Gas Unit............................................................................... 2-21
2.7
Other Equipments............................................................................................ 2-22
2.7.1
Ejector Assembly for High Temperature (E7046EC, E7046EN)...... 2-22
2.7.2
Stop Valve (L9852CB, G7016XH).................................................... 2-23
2.7.3
Check Valve (K9292DN, K9292DS)................................................. 2-24
2.7.4
Air Set................................................................................................ 2-25
2.7.5
Cylinder Pressure Reducing Valve (G7013XF, G7014XF)............... 2-26
2.7.6
ZR22A Heater Assembly.................................................................. 2-27
IM 11M13A01-02E
Toc-2
3. Installation.................................................................................................. 3-1
3.1
3.2
3.3
3.4
3.5
3.6
Installation of General-purpose Detector........................................................ 3-2
3.1.1
Installation Location............................................................................ 3-2
3.1.2
ATEX Flameproof Type....................................................................... 3-2
3.1.3
FM Explosion-proof Type.................................................................... 3-3
3.1.4
CSA Explosion-proof Type.................................................................. 3-4
3.1.5
IECEx Flameproof Type...................................................................... 3-5
3.1.6
Probe Insertion Hole........................................................................... 3-7
3.1.7
Installation of the Detector.................................................................. 3-7
3.1.8
Installation of Probe Protector (ZO21R)............................................. 3-8
Installation of High Temperature Detector (Model ZR22S--015)................ 3-8
3.2.1
Installation Location............................................................................ 3-8
3.2.2
Usage of High Temperature Probe Adapter
(Model ZO21P)................................................................................... 3-9
3.2.3
Probe Insertion Hole......................................................................... 3-10
3.2.4
Mounting of High Temperature Detector........................................... 3-11
Installation of ZR402G Converter................................................................... 3-12
3.3.1
Installation Location.......................................................................... 3-12
3.3.2
Mounting of the Converter................................................................ 3-12
Installation of ZA8F Flow Setting Unit........................................................... 3-14
3.4.1
Installation Location.......................................................................... 3-14
3.4.2
Mounting of ZA8F Flow Setting Unit................................................. 3-14
Installation of ZR40H Automatic Calibration Unit........................................ 3-15
3.5.1
Installation Location.......................................................................... 3-15
3.5.2
Mounting of ZR40H Automatic Calibration Unit................................ 3-15
Insulation Resistance Test.............................................................................. 3-17
4. Piping.......................................................................................................... 4-1
4.1
4.2
Piping for a System Using Flow Setting Unit for Manual Calibration ......... 4-2
4.1.1
Parts Required for Piping in a System Using Flow Setting Units for
Manual Calibration.............................................................................. 4-3
4.1.2
Piping for the Calibration Gas Inlet..................................................... 4-4
4.1.3
Piping for the Reference Gas Inlet...................................................... 4-4
4.1.4
Piping for the Reference Gas Outlet................................................... 4-4
4.1.5
Piping to the High Temperature Probe Adapter.................................. 4-4
Piping for a System to Perform Automatic Calibration................................. 4-6
4.2.1
Parts Required for Piping in a System to Perform Automatic Calibration
4-7
4.2.2
Piping for the Calibration Gases......................................................... 4-8
4.2.3
Piping for the Reference Gas............................................................. 4-8
4.2.4
Piping to the High Temperature Probe Adapter.................................. 4-9
4.2.5
Piping for Blow back........................................................................... 4-9
5. Wiring.......................................................................................................... 5-1
5.1
General................................................................................................................ 5-1
IM 11M13A01-02E
Toc-3
5.2
5.3
5.4
5.5
5.6
5.7
5.8
6.
5.1.1
Terminals for the External Wiring in the Converter............................. 5-2
5.1.2
Wiring.................................................................................................. 5-3
5.1.3
Mounting of Cable Gland.................................................................... 5-4
Wiring for Detector Output................................................................................ 5-5
5.2.1
Cable Specifications........................................................................... 5-5
5.2.2
Connection to the Detector................................................................. 5-6
5.2.3
Connection to the Converter............................................................... 5-6
Wiring for Power to Detector Heater................................................................ 5-7
5.3.1
Cable Specifications........................................................................... 5-7
5.3.2
Connection to Detector....................................................................... 5-7
5.3.3
Connection to Converter..................................................................... 5-8
Wiring for Analog Output.................................................................................. 5-9
5.4.1
Cable Specifications........................................................................... 5-9
5.4.2
Wiring Procedure................................................................................ 5-9
Power and Grounding Wiring........................................................................... 5-9
5.5.1
Power Wiring....................................................................................... 5-9
5.5.2
Grounding Wiring.............................................................................. 5-10
Contact Output Wiring..................................................................................... 5-10
5.6.1
Cable Specifications......................................................................... 5-10
5.6.2
Wiring Procedure.............................................................................. 5-11
Wiring for ZR40H Automatic Calibration Unit............................................... 5-11
5.7.1
Cable Specifications......................................................................... 5-11
5.7.2
Wiring Procedure.............................................................................. 5-12
Contact Input Wiring........................................................................................ 5-12
5.8.1
Cable Specifications......................................................................... 5-12
5.8.2
Wiring Procedure.............................................................................. 5-12
Components and Their Functions........................................................... 6-1
6.1
ZR22S Detector.................................................................................................. 6-1
6.1.1
General-purpose Explosion-proof Detector (except for ZR22S--015).6-1
6.1.2
High Temperature Detector (ZR22S--015)...................................... 6-2
6.2
ZR402G Converter............................................................................................. 6-3
6.3
Touchpanel Switch Operations........................................................................ 6-4
6.4
6.3.1
Basic Panel and Switch...................................................................... 6-4
6.3.2
Display Configuration (for Oxygen Analyzer)..................................... 6-5
6.3.3
Display Functions................................................................................ 6-6
6.3.4
Entering Numeric and Text Data......................................................... 6-6
ZA8F Flow Setting Unit, ZR40H Automatic Calibration Unit......................... 6-8
7. Startup........................................................................................................ 7-1
7.1
Startup Procedure.............................................................................................. 7-1
7.2
Checking Piping and Wiring Connections...................................................... 7-1
7.3
Checking Valve Setup........................................................................................ 7-2
IM 11M13A01-02E
Toc-4
7.4
Supplying Power to the Converter................................................................... 7-2
7.5
Confirmation of Converter Type Setting......................................................... 7-3
7.6
Confirmation of Detector Type Setting............................................................ 7-4
7.7
Selection of Sample Gas................................................................................... 7-4
7.8
Output Range Setting........................................................................................ 7-4
7.9
Setting Display Item........................................................................................... 7-5
7.10
Checking Current Loop..................................................................................... 7-6
7.11
Checking Contact I/O......................................................................................... 7-7
7.12
8.
7.11.1
Checking Contact Outputs.................................................................. 7-7
7.11.2
Checking Calibration Contact Outputs............................................... 7-8
7.11.3
Checking Contact Inputs..................................................................... 7-8
Calibration.......................................................................................................... 7-9
7.12.1
Calibration Setup................................................................................ 7-9
7.12.2
Manual Calibration............................................................................ 7-10
Detailed Data Setting................................................................................ 8-1
8.1
8.2
8.3
8.4
8.5
8.6
Current Output Setting...................................................................................... 8-1
8.1.1
Setting Minimum Current (4 mA) and Maximum Current (20 mA)..... 8-1
8.1.2
Input Ranges....................................................................................... 8-1
8.1.3
Setting Output Smoothing Coefficient................................................ 8-2
8.1.4
Selection of Output Mode................................................................... 8-2
8.1.5
Default Values..................................................................................... 8-2
Output Hold Setting........................................................................................... 8-3
8.2.1
Definition of Equipment Status........................................................... 8-3
8.2.2
Preference Order of Output Hold Value.............................................. 8-4
8.2.3
Output Hold Procedure....................................................................... 8-4
8.2.4
Default Values..................................................................................... 8-5
Oxygen Concentration Alarms Setting........................................................... 8-6
8.3.1
Setting the Alarm Values..................................................................... 8-6
8.3.2
Alarm Output Actions.......................................................................... 8-6
8.3.3
Alarm Setting Procedure..................................................................... 8-7
8.3.4
Default Values..................................................................................... 8-8
Contact Output Setting...................................................................................... 8-8
8.4.1
Contact Output ................................................................................... 8-8
8.4.2
Setting Procedure............................................................................... 8-9
8.4.3
Default Values................................................................................... 8-11
Contact Input Setting....................................................................................... 8-12
8.5.1
Contact Input Functions.................................................................... 8-12
8.5.2
Setting Procedure............................................................................. 8-13
8.5.3
Default Values................................................................................... 8-13
Other Settings.................................................................................................. 8-14
8.6.1
Setting the Date-and-Time................................................................ 8-14
IM 11M13A01-02E
Toc-5
8.6.2
Setting Periods over which Average Values Are Calculated and Periods
over which Maximum and Minimum Values Are Monitored............. 8-14
8.6.3
Setting Fuels..................................................................................... 8-15
8.6.4
Setting Purging................................................................................. 8-20
8.6.5
Setting Passwords............................................................................ 8-21
9. Calibration.................................................................................................. 9-1
9.1
9.2
10.
Calibration Briefs............................................................................................... 9-1
9.1.1
Principle of Measurement................................................................... 9-1
9.1.2
Calibration Gas................................................................................... 9-2
9.1.3
Compensation..................................................................................... 9-3
9.1.4
Characteristic Data from a Sensor Measured During Calibration...... 9-4
Calibration Procedures..................................................................................... 9-5
9.2.1
Calibration Setting............................................................................... 9-5
9.2.2
Default Values..................................................................................... 9-8
9.2.3
Calibration........................................................................................... 9-9
Other Functions....................................................................................... 10-1
10.1
Detailed-data Display....................................................................................... 10-1
10.1.1
Cell Voltage....................................................................................... 10-1
10.1.2
Thermocouple Voltage...................................................................... 10-2
10.1.3
Cold Junction Resistance (C.J. Voltage).......................................... 10-2
10.1.4
Cell’s Internal Resistance................................................................. 10-2
10.1.5
Software Revision............................................................................. 10-2
10.1.6
Span gas and Zero gas Correction Ratios....................................... 10-2
10.1.7
Cell Response Time.......................................................................... 10-3
10.1.8
Robustness of a Cell......................................................................... 10-3
10.1.9
Cell Temperature............................................................................... 10-3
10.1.10 C. J. Temperature............................................................................. 10-4
10.1.11 Maximum Oxygen Concentration..................................................... 10-4
10.1.12 Minimum Oxygen Concentration...................................................... 10-4
10.1.13 Average Oxygen Concentration....................................................... 10-4
10.1.14 Heater On-Time Ratio....................................................................... 10-4
10.1.15 Time.................................................................................................. 10-4
10.1.16 History of Calibration Time................................................................ 10-4
10.1.17 Power Supply Voltage....................................................................... 10-4
10.1.18 Power Frequency.............................................................................. 10-4
10.2
10.3
Trend Graph...................................................................................................... 10-5
10.2.1
Setting Display Items........................................................................ 10-5
10.2.2
Sampling Interval.............................................................................. 10-5
10.2.3
Setting Upper and Lower Limit Values on Graph............................. 10-6
10.2.4
Default Setting.................................................................................. 10-6
Other Display-related Functions.................................................................... 10-7
10.3.1
Auto-Return Time.............................................................................. 10-7
IM 11M13A01-02E
Toc-6
10.4
10.3.3
Language Selection.......................................................................... 10-7
Blow Back......................................................................................................... 10-8
10.4.1
Mode of Blow back............................................................................ 10-8
10.4.2
Operation of Blow back..................................................................... 10-8
10.4.3
Setting Output Hold Time and Blow back Time................................ 10-9
10.4.4
Setting Interval, Start Date, and Start Time...................................... 10-9
10.4.5
Default Setting................................................................................10-10
Parameter Initialization..................................................................................10-10
10.6
Reset................................................................................................................10-13
10.7
Handling of the ZO21S Standard Gas Unit..................................................10-14
10.7.1
Standard Gas Unit Component Identification.................................10-14
10.7.2
Installing Gas Cylinders.................................................................. 10-15
10.7.3
Calibration Gas Flow.......................................................................10-15
Methods of Operating Valves in the ZA8F Flow Setting Unit....................10-18
10.8.1
Preparation Before Calibration....................................................... 10-18
10.8.2
Operating the Span Gas Flow Setting Valve.................................. 10-18
10.8.3
Operating the Zero Gas Flow Setting Valve...................................10-18
10.8.4
Operation After Calibration............................................................. 10-19
Inspection and Maintenance.................................................................. 11-1
11.1
11.2
11.3
12.
Entering Tag Name........................................................................... 10-7
10.5
10.8
11.
10.3.2
Inspection and Maintenance of the Detector................................................ 11-2
11.1.1
Cleaning the Calibration Gas Tube................................................... 11-2
11.1.2
Replacing the Sensor Assembly....................................................... 11-2
11.1.3
Replacement of the Heater Unit (Assembly).................................... 11-5
11.1.4
Replacement of Flame Arrestor Assembly....................................... 11-8
11.1.5
Replacement of O-ring...................................................................... 11-8
11.1.6
Cleaning the High Temperature Probe Adapter................................ 11-9
11.1.7
Stopping and Re-starting Operation................................................. 11-9
Inspection and Maintenance of the Converter........................................... 11-10
11.2.1
Replacing Fuses............................................................................. 11-10
11.2.2
Cleaning...........................................................................................11-11
11.2.3
Adjust LCD screen contrast.............................................................11-11
Replacing Flowmeter in ZR40H Automatic Calibration Unit..................... 11-12
Troubleshooting...................................................................................... 12-1
12.1
12.2
12.3
Displays and Remedies When Errors Occur................................................ 12-1
12.1.1
Error Types........................................................................................ 12-1
12.1.2
Remedies When an Error Occurs..................................................... 12-2
Displays and Remedies When Alarms are Generated................................. 12-5
12.2.1
Alarm Types...................................................................................... 12-5
12.2.2
Remedies When Alarms are Generated........................................... 12-6
Countermeasures When Measured Value Shows Error............................ 12-11
12.3.1
Measured Value Higher Than True Value....................................... 12-11
IM 11M13A01-02E
Toc-7
12.3.2
Measured Value Lower Than True Value.......................................12-12
12.3.3
Measurements Sometimes Show Abnormal Values......................12-12
Customer Maintenance Parts List ......................................CMPL 11M13A01-02E
Customer Maintenance Parts List ......................................CMPL 11M12C01-01E
Customer Maintenance Parts List ......................................CMPL 11M12A01-11E
Customer Maintenance Parts List ......................................CMPL 11M03B01-10E
Customer Maintenance Parts List ......................................CMPL 11M03B01-05E
Customer Maintenance Parts List ..........................................CMPL 11M3D1-01E
Revision Information................................................................................................i
IM 11M13A01-02E
Blank Page
1-1
<1. Overview>
1. Overview
The EXAxtZR Separate type Explosion-proof Zirconia Oxygen Analyzer is used to monitor
and control the oxygen concentration in combustion gases, in boilers and industrial furnaces,
for wide application in industries which consume considerable energy–such as steel, electric
power, oil and petrochemical, ceramics, paper and pulp, food, or textiles, as well as incinerators
and medium/small boilers. It can help conserve energy in these industries. The EXAxtZR also
contributes to preservation of the earth’s environment in preventing global warming and air
pollution by controlling complete combustion to reduce CO2, SOx and NOx.
ZR22S Separate type Explosion-proof Detector uses a high-reliability Zirconia sensor, and its
heater assembly can be replaced on site. The detector is mounted, for example, on the wall of a
flue and can measure the gases directly.
For use in combustion gases at temperatures up to 1400°C, choose the general-purpose 0.15
m long detector, which is combined with ZO21P, the high temperature probe protector. The
converter is equipped with an LCD touch screen which has various setting displays, a calibration
display, oxygen concentration trend display, with easier operation and improvement of display
functions. The converter is equipped with various standard functions such as measurement
and calculation as well as maintenance functions including self-test. Analyzer calibration can
also be fully automated ─ and ZR40H, an automatic calibration unit, is available. Choose the
detector version which best suits your needs so that an optimal combustion control system can
be obtained.
Some examples of typical system configurations are illustrated below:
1.1
< EXAxt ZR > System Configuration
The system configuration should be determined by the conditions; e.g. whether calibration is
to be automated, and whether flammable gas is present and requires safety precautions. The
system configuration can be classified into three basic patterns as follows:
1.1.1
System 1
This system is for monitoring and controlling oxygen concentration in the combustion gases of
a large-size boiler or heating furnace. Clean (dry) air (21%O2) is used as the reference gas and
the span gas for calibration. Zero gas is fed in from a cylinder during calibration. The gas flow is
controlled by the ZA8F flow setting unit (for manual valve operation).
Hazardous Area
ZR402G Converter
ZR22S Separate type Explosion-proof
Zirconia Oxygen Analyzer, Detector
Stop valve
or
Check valve
EXA ZR402G
*1
Reference gas
Non-hazardous Area
Flowmeter Needle
valve
Air Set
Calibration gas
Instrument air
Span gas (*)
ZA8F
Flow Setting Unit
* Calibration gas unit same as for zero gas.
Pressure
reducing
valve
~ 100 to 240 V AC
Contact input
Analog output,
contact output
Digital output
(HART)
Zero gas
cylinder
Calibration gas
unit case
F01-1E.ai
Figure 1.1 System configuration 1
IM 11M13A01-02E
1.1.2
1-2
<1. Overview>
System 2
This example, System 2, represents typical applications in large boilers and heating furnaces,
where there is a need to monitor and control oxygen concentration. The reference gas and
calibration-time span gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder.
System 2 uses the ZR40H automatic calibration unit, with auto-switching of the calibration gas.
A “combustible gas detected” contact input turns off power to the heater. There’s also contact
output from the converter that can be used to operate a purge gas valve to supply air to the
sensor.
Hazardous Area
ZR22S Separate type Explosion-proof
Zirconia Oxygen Analyzer, Detector
Non-hazardous Area
ZR402G Converter
EXA ZR402G
Check valve
*1
Signal
(6-core shield cable)
~ 100 to 240 V AC
Heater (2-core cable)
Reference gas
Flowmeter Needle
valve
Air Set
Calibration gas
Contact input
Analog output, contact output
Digital output (HART)
Instrument air
ZR40H
Automatic Calibration Unit
Pressure
regulating
valve
*2
Zero gas cylinder
Calibration gas
unit case
F01-2E.ai
*1: Shield cable;Use shielded signal cables, and connect the shield to the FG terminal of the converter.
*2: When a zirconia oxygen analyzer is used, 100% N2 gas cannot be used as the zero gas.
Use approximately 1% of O2 gas (N2-based).
Figure 1.2 System configuration 2
IM 11M13A01-02E
1.2
1.2.1
1-3
<1. Overview>
< EXAxt ZR > System Components
System Components
Separate type
System configuration
System 1 System 2
System Components
ZR22S Separate type Explosion-proof Zirconia Oxygen Analyzers, Detector
ZR402G Separate type Zirconia Oxygen Analyzer, Converter
ZO21P High Temperature Probe Adapter for separate type Zirconia Oxygen Analyzer
E7046EC, E7046EN Ejector Assembly for High Temperature
ZO21R Probe Protector for Zirconia Oxygen Analyzers
ZO21S Standard Gas Unit
ZA8F Flow Setting Unit for manual calibration
ZR40H Automatic Calibration Unit for Separate type Analyzer
L9852CB, G7016XH Stop Valve for Calibration gas line
( )
K9292DN, K9292DS Check Valve for Calibration gas line
( )
G7003XF/K9473XK, G7004XF/K9473XG Air Set
G7013XF, G7014XF Pressure Reducing Valve for Gas Cylinder
ZR22A, Heater Assembly (Spare Parts for ZR22S)
: Items required for the above system example
: To be selected depending on each application. For details, refer to corresponding chapter.
( ) : Select either
1.2.2
Mounting
Detectors and Accessories
Sample gas temperature 0 to 700°C
Insertion
General-purpose detector
length
Sample gas temperature 700 to 1400°C
Application
High temperature detector
Heating
furnace
Sample
outlet
Absorption
structure
Detector
(ZR22S)
Boiler
Heating
furnace
2m
or less
Probe Protector
Gas Flow (ZO21R)
Sample inlet
For
Detector pulverized
(ZR22S) coal boiler
F01-4E.ai
High temperature
detector
Sample inlet
F01-5E.ai
Probe material and Temperature:
SUS310S; 800°C, SiC; 1400°C
Mounting: Vertical downwards
Insertion length: 0.5, 0.6, 0.7, 0.8,
0.9, 1.0, 1.5 m
When duct pressure is atmospheric
or negative, attach ejector assembly.
F01-3E.ai
Horizontal
to
vertical
High temperature
probe adapter
ZO21P
Application
with gas
flow velocity
10 m/sec or
more
Ejector assembly for high
temperature (E7046EC, E7046EN)
Needle
valve
Inlet
Pressure gauge
Ejector
Blow
F01-6E.ai
IM 11M13A01-02E
Blank Page
2-1
<2. Specifications>
2.
Specifications
This chapter describes the specifications for the following:
ZR22S
Separate type explosion-proof detector (See Subsection 2.2.1)
ZO21R
Probe protector (See Subsection 2.2.2)
ZR22S (0.15 m) Separate type explosion-proof detector for high temperature
(See Subsection 2.3.1)
ZO21P
High temperature probe adapter (See Subsection 2.3.2)
ZR402G
Separate type converter (See Section 2.4)
ZA8F
Flow setting unit (See Subsection 2.5.1)
ZR40H
Automatic calibration unit (See Subsection 2.5.2)
ZO21S
Standard gas unit (See Section 2.6)
Other Equipments (See Section 2.7)
CAUTION
Requirements for explosion-proof use:
The ZR22S is connected to a ZR402G or AV550G*1 that is mounted in a non-hazardous area.
The ambient temperature is in the range -20 to + 60°C. The surface temperature of the ZR22S is
not over the temperature class T2 (300°C)*2.
*1: Refer to IM 11M12D01-01E
*2: The terminal box temperature does not exceed 150°C.
Oxygen concentration of sample/reference/calibration gas shall not exceed that found in normal
air, typically 21 vol%.
2.1
General Specifications
Standard Specifications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases. May not be applicable corrosive gas such
as ammonia and chlorine is present — Contact with YOKOGAWA and
its agency.)
Measurement System: Zirconia system
Measurement range: 0.01 to 100 vol%O2
Output Signal:
4 to 20 mA DC (maximum load resistance 550 Ω)
Setting Range: Any setting in the range of 0 to 5 through 0 to 100 vol%O2 (in 1 vol%O2),
or partial range
Digital Communication (HART): 250 to 550 Ω, depending on number of field devices connected
to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range:
0 to 100 vol%O2
Warm-up Time:
Repeatability:
Approx. 20 min.
±0.5% Maximum value of set range; Range from 0 to 5 vol%O2 or more
and less than 0 to 25 vol%O2
±1% Maximum value of set range; Range from 0 to 25 vol%O2 or more
and up to 0 to 100 vol%O2
IM 11M13A01-02E
2-2
<2. Specifications>
Linearity:
(Excluding standard gas tolerance)
(Use oxygen of known concentration (within the measuring range) as the
zero and span calibration gases.)
±1% Maximum value of set range; Range from 0 to 5 vol%O2 to 0 to
25 vol%O2. (Sample gas pressure: within ±4.9 kPa)
±3% Maximum value of set range; Range from 0 to 25 vol%O2 or more
and less than 0 to 50 vol%O2. (Sample gas pressure: within
±0.49 kPa)
±5% Maximum value of set range; Range from 0 to 50 vol%O2 to 0 to
100vol%O2. (Sample gas pressure: within ±0.49 kPa)
Drift: (Excluding the first two weeks in use)
Both zero and span ±2% Maximum value of range setting/month
Response Time : Response of 90% within 5 seconds. (Measured after gas is introduced
from calibration gas inlet and analog output starts changing.)
Safety, EMC, and RoHS conformity standards of ZR22S and ZR402G
Installation altitude based on IEC 61010:
2000 m or less
Category based on IEC 61010:
II (Note)
Pollution degree based on IEC 61010:
2 (Note)
Not Installation category, called over-voltage category, specifies impulse withstand voltage. Category II is for electrical
equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce
dielectric strength. Degree 2 is the normal indoor environment.
Safety: Conforms to EN 61010-1, EN61010-2-030,
CAN/CSA-C22.2 No. 61010.1 certified,
UL Std. No. 61010-1 certified
EMC: Conforms to EN 61326-1*, Class A, Table 2,
EN 61326-2-3, EN 61000-3-2
*: Influence of immunity environment (Criteria A ): ±20% of F. S.
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
EN61326-1 Class A
Korea Electromagnetic Conformity Standard
RoHS:
EN 50581
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
IM 11M13A01-02E
2.2
2-3
<2. Specifications>
Separate type Explosion-proof Detector and
Related Equipment
Separate type Explosion-proof detector ZR22S can be used in combination with the probe
protector ZO21R (see Subsection 2.2.2).
2.2.1
ZR22S Separate type Explosion-proof Detector
Flameproof Type
ATEX Flameproof: ZR22S-A
Applicable Standard:
EN 60079-0: 2012+A11: 2013, EN 60079-1: 2014
EN 60079-31: 2014
Certificate Number:
KEMA 04ATEX2156 X
Type of protection: Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
Equipment Group:
II
Category:
2GD
Temperature class for Ex “db”: T2
The maximum surface temperature for Ex “tb”: T300°C
Degree of protection of enclosure: IP66
NAME PLATE
MODEL : Specified model code
SUFFIX : Specified suffix code
STYLE : Style code
AMB. TEMP : Ambient temperature
NO. : Serial No. and year of production*1
Yokogawa Electric Corporation : The manufacturer name
Tokyo 180-8750 JAPAN : The manufacturer address*2
No. KEMA 04ATEX2156 X
Ex db IIB+H2 T2 Gb,
Ex tb IIIC T300°C Db
The country of origin
*1: The third to seventh figure from the last shows the year of production.
e.g. 27D327560 2005.02
The year of production
*2: "180-8750" is a zip code which represents the following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
F02-1.1E.ai
IM 11M13A01-02E
2-4
<2. Specifications>
FM Explosion-proof:
ZR22S-B
Applicable Standard: FM3600 1998, FM3615 1989, FM3810 2005,
ANSI/NEMA 250 1991
Type of protection:
Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating:
NEMA 4X
Temperature Class:
T2
CSA Explosion-proof: ZR22S-C
Applicable Standard: C22.2 No.0-M1991, C22.2 No.0.4-04,
C22.2 No.0.5-1982, C22.2 No.25-1966,
C22.2 No.30-M1986, C22.2 No.94-M91,
C22.2-No.61010-1-04
Certificate Number:
1649642
Type of protection:
Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure:
Type 4X
Temperature Class:
T2
IECEx Flameproof: ZR22S-D
Applicable Standard: IEC 60079-0: 2011, IEC 60079-1: 2014, IEC 60079-31: 2013
Certificate Number:
IECEx KEM 06.0006X
Type of protection:
Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
Temperature class for Ex “db”: T2
The maximum surface temperature for Ex “tb”: T300°C
Degree of protection of enclosure: IP66
NAME PLATE
MODEL : Specified model code
SUFFIX : Specified suffix code
STYLE : Style code
AMB. TEMP : Ambient temperature
NO. : Serial No. and year of production*1
Yokogawa Electric Corporation : The manufacturer name
Tokyo 180-8750 JAPAN : The manufacturer address*2
No. IECEx KEM 06.0006X
Ex db IIB+H2 T2 Gb,
Ex tb IIIC T300°C Db
The country of origin.
*1: The third to seventh figure from the last shows the year of production.
e.g. 27D327560 2005.02
*2: "180-8750" is a zip code which represents the following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
F02-2E.ai
IM 11M13A01-02E
<2. Specifications>
2-5
Sample Gas Temperature:0 to 700°C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the
temperature is greater than 600°C or greater.
700 to 1400°C (with High Temperature Probe Adapter)
For high temperature sample gas, apply 0.15 m length probe and
High Temperature Probe Adapter ZO21P.
Sample Gas Pressure: -5 to +5 kPa
For 0.15 m probe, -0.5 to +5 kPa.
No pressure fluctuation in the furnace should be allowed.
Oxygen concentration of sample gas:
For explosion-proof use, not more than that found in normal air, typically 21 vol%
Probe Length:
0.15, 0.4, 0.7, 1.0, 1.5, 2.0 m
Probe Material:
316 SS (JIS)
Ambient Temperature: -20 to +60°C (-20 to +150°C on the terminal box surface)
Reference Gas System: Instrument Air
Instrument Air System:
Pressure; 50 kPa + the pressure inside the furnace
(It is recommended to use air which has been dehumidified by cooling to
dew point -20°C or less, and dust or oil mist are removed.)
Consumption; Approx. 1 NL/min
Oxygen concentration of calibration gas:
For explosion-proof use, not more than that found in normal air, typically 21 vol%
Wetted Material: 316L SS (JIS), Zirconia, 304 SS (JIS) (flange), Hastelloy B,
(Inconel 600, 601)
Construction: Heater and thermocouple replaceable construction.
Equivalent to NEMA 4X/IP66 (Achieved when pipes are installed at
calibration gas and reference gas inlets and pipe is installed so that
reference gas can be exhausted to clean atmosphere. Excluding probe
top. And achieved when the cable entry is completely sealed with a cable gland.)
Terminal Box Case: Material; Aluminum alloy
Terminal Box Paint Color: Case; Mint green (Munsell 5.6BG3.3/2.9)
Cover; Mint green (Munsell 5.6BG3.3/2.9)
Finish:
Polyurethane corrosion-resistant coating
Gas Connection:
Rc1/4 or 1/4 FNPT
Wiring Connection:
ATEX; M20 by 1.5 mm or 1/2 NPT select one type (2 pieces)
FM; 1/2 NPT (2 pieces)
CSA; 1/2 NPT (2 pieces)
IECEx; M20 by 1.5 mm or 1/2 NPT select one type (2 pieces)
Installation:
Flange mounting
Probe Mounting Angle: Horizontal to vertically downward.
Weight: Insertion length of 0.4 m: Approx. 13kg (ANSI 150 4)
Insertion length of 0.7 m: Approx. 14 kg (ANSI 150 4)
Insertion length of 1.0 m: Approx. 15 kg (ANSI 150 4)
Insertion length of 1.5 m: Approx. 17 kg (ANSI 150 4)
Insertion length of 2.0 m: Approx. 19 kg (ANSI 150 4)
Available Converter: ZR402G, AV550G
CAUTION
The ZR22S must be used in conjunction with a ZR402G or AV550G. If used with a converter
other than a ZR402G or AV550G, the ZR22S does not operate as an explosion-proof equipment.
IM 11M13A01-02E
2-6
<2. Specifications>
Model and Codes
Model
Suffix code
ZR22S
-------------------------------
- - - - - - - - - - - Separate type Explosion-proof Zirconia Oxygen Analyzer, Detector
-A
-B
-C
-D
-Q
-R
-------------------------------------------------------------
ATEX certified flameproof
(*11)
FM certified explosion-proof
CSA certified explosion-proof
IECEx certified flameproof
(*11)
EAC with PA certified explosion-proof (*13)
EAC certified explosion-proof
(*13)
-------------------------------------------------------------
0.15 m (for high temperature use)
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
---------------------
Stainless steel
Stainless steel with Inconel calibration gas tube
---------------------------------------------------------------------------------------------------------------------------------------------
ANSI Class 150 2 RF SUS304 (JIS)
(*10)
ANSI Class 150 3 RF SUS304 (JIS)
ANSI Class 150 4 RF SUS304 (JIS)
(*10)
DIN PN10 DN50 A SUS304 (JIS)
DIN PN10 DN80 A SUS304 (JIS)
DIN PN10 DN100 A SUS304 (JIS)
JIS 5K 65 FF SUS304 (JIS)
JIS 10K 65 FF SUS304 (JIS)
JIS 10K 80 FF SUS304 (JIS)
JIS 10K 100 FF SUS304 (JIS)
JIS 5K 32 FF SUS304 (JIS) (for high temperature use)
JPI Class 150 4 RF SUS304 (JIS)
JPI Class 150 3 RF SUS304 (JIS)
Westinghouse
Explosion-proof
Approval (*12)
-015
-040
-070
-100
-150
-200
Length
Wetted material
-S
-C
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-Q
-R
-S
-W
Flange
(*2)
-E
Reference gas
(*1)
- - - - - - - - - - - M20 x1.5 mm
- - - - - - - - - - - 1/2 NPT
-E
Instruction manual
(*7)
(*3)
(*8)
- - - - - - - - - - - Rc1/4
- - - - - - - - - - - 1/4 NPT (Female)
-M
-T
Connection box thread
(*9)
- - - - - - - - - - - English
-A - - - - - - - - - - - Always -A
─
/C
Valves /CV
/SV
Tag plates /SCT
/PT
Options
*12
*13
Description
- - - - - - - - - - - External connection (Instrument air)
-R
-T
Gas thread
*1
*2
*3
*4
*5
*6
*7
*8
*9
*10
*11
Option code
Inconel bolt
Check valve
Stop valve
Stainless steel tag plate
Printed tag plate
(*4)
(*5)
(*5)
(*6)
(*6)
Used with the ZO21P High Temperature Probe Adapter. Select flange (-Q).
The thickness of the flange depends on its dimensions*3 The thickness of the flange depends on its dimensions.
The flange thickness does not conform to JIS specification.
Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700ºC).
Specify either /CV or /SV option code.
Specify either /SCT or /PT option code.
Recommended if sample gas contains corrosive gas like chlorine.
Piping for reference gas must be installed to supply reference gas constantly at a specified flow rate.
When selecting code -B (FM certified explosion-proof) or -C (CSA certified explosion-proof), select code -T (1/2 NPT).
Confirm inside diameter of pipe attached to customer’s flange in case that -A or -E is selected.
The cable entry devices (cable glands etc.) and blind plugs shall be in type of protection Ex “db” or Ex “tb”, suitable for the
conditions of use and correctly installed. They shall provide a degree of ingress protection of at least IP66.
When using ZR22S as CE marking compliance product, select -A (ATEX certified flameploof).
“-Q” is the explosion-proof type of EAC with Pattern Approval for Russia. “-R” is the explosion-proof type of EAC for Kazakhstan
and Belarus.
Standard Accessory
Item
Allen wrench
Parts No.
Q'ty
L9827AB
1
Description
For lock screw
IM 11M13A01-02E
2-7
<2. Specifications>
External Dimensions
ZR22S Separate type Explosion-proof Zirconia Oxygen Analyzer, Detectors
24 ±3
L
Unit : mm
311 ±3
85 ±3
Ø 50.8
Ø 87
±3
±3
t
L (m)
0.15
0.4
0.7
1.0
1.5
2.0
tolerance (mm)
±4
±5
±7
±8
±10
±12
156
95 ±3
±3
Rc1/4 or 1/4 NPT (F)
Reference gas inlet
2-M20 × 1.5 or 2-1/2 NPT (F)
Cable connection port
48
Rc1/4 or 1/4 NPT (F)
Calibration gas inlet
Flange
±3
Ø124
±3
25 ±2
Ø 52 ±3
Rc1/4 or 1/4 NPT (F)
Reference gas outlet
n-ØC
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JIS 5K 32 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
A
152.4
190.5
228.6
165
200
220
155
175
185
210
115
229
190
155
B
120.6
152.4
190.5
125
160
180
130
140
150
175
90
190.5
152.4
127
n
4
4
8
4
8
8
4
4
8
8
4
8
4
4
C
19
19
19
18
18
18
15
19
19
19
15
19
19
11.5
t
19
24
24
18
20
20
14
18
18
18
5
24
24
14
ØB
ØA
FLANGE
F02-3E.ai
IM 11M13A01-02E
2-8
<2. Specifications>
2.2.2
ZO21R Probe Protector
Used when sample gas flow velocity is approx. 10 m/sec or more and dust particles wears the
detector in cases such as pulverized coal boiler of fluidized bed furnace (or burner) to protect the
detector from wearing by dust particles.
Insertion Length:
1.05, 1.55, 2.05 m.
Flange: JIS 5K 65A FF equivalent. ANSI Class 150 4 FF (without serration)
equivalent. However, flange thickness is different.
Material:
316 SS (JIS), 304 SS (JIS) (Flange)
Weight: 1.05 m; Approx. 6/10 kg (JIS/ANSI),
1.55 m; Approx. 9/13 kg (JIS/ANSI),
2.05 m; Approx. 12/16 kg (JIS/ANSI)
Installation: Bolts, nuts, and washers are provided for detector, probe protector and
process-side flange.
Model and Codes
Model
ZO21R
Suffix code
-L
Insertion
length
Flange ( *1)
-100
-150
-200
----------------------------
1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)
-J
-A
-------------------
JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304
----------
Style B
*1 Thickness of flange depends on dimensions of flange.
External Dimension
Unit: mm
Flange <1>
(with bolts, nuts and washers)
Gasket (Thickness 3.0)
Washer (M12)
Mounting nut (M12)
316 SS (JIS)
Gas flow
ØB
ØA
Ø60.5
Description
Probe Protector(0 to 700°C)
*B
Style code
Option code
----------
D
t
ØB
l (Insert length)
C
l=1050,1550,2050
Dimensions of holes on opposing surface
F02-4E.ai
Flange<1>
JIS 5K 65 FF 304 SS (JIS)
ANSI Class 150 4 FF 304 SS (JIS)
A
B
C
t
D
155
130
228.6
190.5
4 - Ø15
5
40
8 - Ø19
12
50
IM 11M13A01-02E
2.3
2-9
<2. Specifications>
Separate type Explosion-proof Detector for
High Temperature and Related Equipment
2.3.1
ZR22S (0.15 m) Separate type Explosion-proof
Detector for High Temperature
Standard Specifications
Construction:
Water-resistant
Probe length:
0.15 m
Terminal box:
Aluminum alloy
Probe material: Probe material in contact with gas: 316 SS (JIS), 316L SS (JIS) (Probe)
304 SS (JIS) (Flange), Zirconia (Sensor), Hastelloy B, (Inconel 600, 601)
Weight:
Approx. 6 kg
Installation: Flange mounting (When using high temperature detector, high
temperature probe adapter ZO21P is necessary.)
Flange standard:
JIS 5K 32 FF equivalent (thickness varies)
Mounting angle: Any angle between horizontal and vertical downward (high temperature
probe is fitted with an adapter)
Reference gas and calibration gas piping connection: Rc1/4 or 1/4 NPT female
Ambient temperature: -20 to +60°C (-20 to +150°C on the case surface)
Sample gas temperature: 0 to 700°C (temperature at the measuring point of the sampling gas.)
When sample gas is 700°C to 1400°C, the high temperature probe
adapter is used.
Temperature of the high temperature probe adapter shall not exceed
300°C to protect the gasket and avoid the bolts seizing together.
Sample gas pressure: -0.5 to +5 kPa. When using in the range of 0 to 25 vol%O2 or more,
the sample gas pressure should be in the range of -0.5 to +0.5 kPa.
(Where the sample gas pressure for the high temperature probe is
negative, an auxiliary ejector assembly is necessary.)
Refer to Subsection 2.2.1. for the explosion-proof and other specifications.
Model and Code
Refer to “Model and Codes” in page 2-6.
External Dimensions
Refer to the Figure in page 2-7.
IM 11M13A01-02E
2-10
<2. Specifications>
2.3.2
ZO21P High Temperature Probe Adapter
Measuring O2 in the high temperature gases (exceeds 700°C) requires a explosion-proof
detector ZR22S of 0.15 m length and a high temperature probe adapter.
Sample gas temperature:
0 to 1400°C (when using SiC probe)
0 to 800°C (when using 310S SS probe)
Sample gas pressure: -0.5 to +5 kPa. When using in the range of 0 to 25 vol%O2 or more,
the sample gas pressure should be in the range of -0.5 to +0.5 kPa.
(Where the sample gas pressure for the high temperature probe is
negative, an auxiliary ejector assembly is necessary.)
Insertion length:
0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.5 m
Material in Contact with Gas: 316 SS (JIS), SiC or 310S SS, 304 SS (JIS) (flange)
Probe Material:
SiC, 310S SS (JIS)
Installation:
Flange mounting (FF type or RF type)
Probe Mounting Angle: Vertically downward within ± 5°. Where the probe material is 310S SS,
horizontal mounting is available.
Construction:
Non-explosion-proof. Rainproof construction
Weight(example): Insertion length of 1.0 m: approx. 5.3 kg (JIS) / approx. 11.3 kg (ANSI)
Insertion length of 1.5 m: approx. 5.8 kg (JIS) / approx. 11.8 kg (ANSI)
Model and Codes
Model
ZO21P
Material
Insertion
length
Flange
Style code
Option
Suffix code
Option code
-H
-A
-B
-050
-060
-070
-080
-090
-100
-150
-J
-N
-M
-L
-A
-R
-Q
-T
-S
-E
*B
Description
-------
High Temperature Probe Adapter
-------------
SiC
SUS310S
------------------------------------
0.5 m
0.6 m
0.7 m
0.8 m
0.9 m
1.0 m
1.5 m
-------------------------------------------------------------
JIS 5K 50 FF SUS304 (JIS)
JIS 10K 65 FF SUS304 (JIS)
JIS 10K 80 FF SUS304 (JIS)
JIS 10K 100 FF SUS304 (JIS)
ANSI Class 150 4 RF SUS304 (JIS)
ANSI Class 150 2 1/2 RF SUS304 (JIS)
ANSI Class 150 3 RF SUS304 (JIS)
JPI Class 150 3 RF SUS304 (JIS)
JPI Class 150 4 RF SUS304 (JIS)
DIN PN10 DN50 A SUS304 (JIS)
-------
Style B
/EJ1
/EJ2
/SCT
Ejector Assy with E7046EC
Ejector Assy with E7046EN
Stainless steel tag plate
Note:The Insertion length 0.15 m of the ZR22S should be specified.
IM 11M13A01-02E
<2. Specifications>
2-11
External Dimension
Unit: mm
Approx. 351
Sample gas outlet
180
Flange (Thickness 5)
JIS 5K 32 FF equivalent
Ø124±3
Ø60.5
110
ØA
Flange <1>
Approx. 48
85±3
t
Flange provided
by customer
Detector (ZR22S)
Ø52 over
95±3
Reference gas inlet Rc1/4 or 1/4 NPT
48±3
High temperature
Probe SiC pipe
25±3
Approx. 100
Gasket (Thickness 1.5)
Ø115
Ø60.5
170
Approx. 215
Rc1/2(Note2)
L (Insertion length) (Note1)
Ø30
Pipe hole (2- M20, 2-1/2 NPT)
Calibration gas inlet Rc1/4 or 1/4 NPT
ØA
C
ØB
(Note 1) L = 0.5、0.6、0.7、0.8、0.9、1.0、1.5 (m)
(Note 2) Sample gas outlet
(if the sample gas pressure is negative, connect the ejector assembly.)
F02-5E.ai
A
B
C
t
JIS 5K 50 FF 304 SS (JIS)
<1> Flange
130
105
4 - Ø15
14
JIS 10K 65 FF 304 SS (JIS)
175
140
4 - Ø19
18
JIS 10K 80 FF 304 SS (JIS)
185
150
8 - Ø19
18
JIS 10K 100 FF 304 SS (JIS)
210
175
8 - Ø19
18
ANSI Class 150 2 1/2 RF 304 SS (JIS)
177.8
139.7
4 - Ø19
22.4
ANSI Class 150 3 RF 304 SS (JIS)
190.5
152.4
4 - Ø19
24
ANSI Class 150 4 RF 304 SS (JIS)
228.5
190.5
8 - Ø19
24
JPI Class 150 3 RF 304 SS (JIS)
190
152.4
4 - Ø19
24
JPI Class 150 4 RF 304 SS (JIS)
229
190.5
8 - Ø19
24
DIN PN10 DN50 A 304 SS (JIS)
165
126
4 - Ø18
18
IM 11M13A01-02E
2.4
2-12
<2. Specifications>
ZR402G Separate type Converter
CAUTION
Converter (Model ZR402G) must not be located in hazardous area.
2.4.1
Standard Specification
The ZR402G Separate type Converter can be controlled by LCD touchscreen on the converter.
Display:
LCD display of size 320 by 240 dot with touchscreen.
Output Signal:
4 to 20 mA DC, two points (maximum load resistance 550 Ω)
Contact Output Signal: Four points (one is fail-safe, normally open)
Contact Input:
Two points
Auto-calibration Output: Two points (for dedicated automatic calibration unit)
Ambient Temperature: -20 to +55°C
Storage Temperature: -30 to +70°C
Ambient Humidity:
0 to 95%RH (Non-condensing)
Power Supply Voltage: Ratings; 100 to 240 V AC, Acceptable range; 85 to 264 V AC
Power Supply Frequency: Ratings; 50/60 Hz, Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
Maximum Distance between Detector and Converter:
Conductor two-way resistance must be
10 Ω or less (when a 1.25 mm2 cable or equivalent is used, 300 m or
less.)
Construction: Outdoor installation, equivalent to NEMA 4X/IP66 (with conduit holes
completely sealed with a cable gland)
Wiring Connection:
G1/2, Pg13.5, M20 by 1.5 mm, 1/2 NPT (with plug), eight holes
Installation:
Panel, wall or 2-inch pipe mounting
Case:
Aluminum alloy
Paint Color:
Door: Silver gray (Munsell 3.2PB7.4/1.2)
Case: Silver gray (Munsell 3.2PB7.4/1.2)
Finish:
Polyurethane corrosion-resistance coating
Weight:
Approx. 6 kg
IM 11M13A01-02E
2.4.2
2-13
<2. Specifications>
Functions
Display Functions:
Value Display; Displays values of the measured oxygen concentration, etc
Graph Display; Displays trends of measured oxygen concentration
Data Display; Displays various useful data for maintenance, such as cell temperature,
reference junction temperature, maximum/minimum oxygen
concentration, or the like
Status Message; Indicates an alarm or error occurrence with flashing of the
corresponding icon. Indicates status such as warming-up, calibrating,or
the like by icons.
Alarm, Error Display; Displays alarms such as “Abnormal oxygen concentration” or
errors such as “Abnormal cell e.m.f.” when any such status occurs.
Calibration Functions:
Automatic calibration; Requires the ZR40H Automatic Calibration Unit. It calibrates
automatically at specified intervals.
Semi-auto Calibration; Requires the Automatic Calibration Unit. Input calibration direction
on the touchscreen or contact, then it calibrates automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration gas in
operation interactively with an LCD touchscreen.
Blow back Function:
Output through the contact in the set period and time. Auto/Semi_Auto selectable.
Maintenance Functions:
Can operate updated data settings in daily operation and checking. Display data settings,
calibration data settings, blow back data settings, current output loop check, contact
input/output check.
Setup Functions:
Initial settings suit for the plant conditions when installing the converter.
Equipment settings, current output data settings, alarm data settings, contact data
settings, other settings.
Self-diagnosis:
This function diagnoses conditions of the converter or the detector and indicates when
any abnormal condition occurs.
Password Functions:
Enter your password to operate the analyzer excepting data display. Individual
passwords can be set for maintenance and setup.
IM 11M13A01-02E
2-14
<2. Specifications>
Display and setting content:
Measuring Related Items: Oxygen concentration (vol%O2), output current value (mA), air ratio,
moisture quantity (in hot gases) (vol% H2O)
Display Items:
Cell temperature (°C), thermocouple reference junction temperature
(°C), maximum/minimum/average oxygen concentration (vol% O2), cell
e.m.f. (mV), cell internal resistance (Ω), cell condition (in four grades),
heater on-time rate (%), calibration record (ten times), time (year/month/
day, hour/minute)
Calibration Setting Items: Span gas concentration (vol%O2), zero gas concentration (vol%O2),
calibration mode (automatic, semi-automatic, manual), calibration type
and method (zero-span calibration, zero calibration only, span calibration
only), stabilization time (min.sec), calibration time (min.sec), calibration
interval (day/hour), starting time (year/month/day, hour/minute)
Equipment Related Items: Sample (Measuring) gas selection
Output Related Items: Analog output/output mode selection, output conditions when warmingup/maintenance/calibrating (during blow back)/abnormal, 4 mA/20 mA
point oxygen concentration (vol%O2), time constant.
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol%O2),
oxygen concentration low alarm/low-low alarm limit values (vol%O2),
oxygen concentration alarm hysteresis (vol%O2), oxygen concentration
alarm detection, alarm delay (seconds)
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 to 4
(abnormal, high-high alarm, high alarm, low alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration gas
pressure decrease, temperature high alarm, blow back, flameout gas
detection, calibration coefficient (correction ratio) alarm, stabilization
timeout)
Converter Output: Two points mA analog output (4 to 20 mA DC (maximum load resistance
of 550 Ω)) and one mA digital output point (HART) (minimum load
resistance of 250 Ω).
Range: Any setting between 0 to 5 through 0 to 100 vol%O2 in 1 vol%O2, or
partial range is available (Maximum range value/minimum range
value 1.3 or more).For the log output, the minimum range value is fixed
at 0.1 vol%O2.
4 to 20 mA DC linear or log can be selected. Input/output isolation.
Output damping: 0 to 255 seconds.Hold / non-hold selection, preset value setting possible
with hold.
Contact Output: Four points, contact capacity 30 V DC 3 A, 250 V AC 3 A (resistive load)
Three of the output points can be selected to either normally energized
or normally de-energized status.
Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9
vol%O2) can be added to high/low alarms.
The following functions are programmable for contact outputs.
(1) Abnormal, (2) High-high alarm, (3) High alarm, (4) Low-low alarm,
(5) Low alarm, (6) Maintenance, (7) Calibration, (8) Range switching
answer-back, (9) Warm-up, (10) Calibration gas pressure decrease
(answer-back of contact input), (11) Temperature high alarm,
(12) Blow back start, (13) Flameout gas detection (answer-back of
contact input), (14) Calibration coefficient (correction ratio) alarm,
(15) Startup power stabilization timeout alarm
Contact output 4 is set to normally operated, and fixed error status.
IM 11M13A01-02E
2-15
<2. Specifications>
Two points (voltage-free)
Contact Input:
The following functions are programmable for contact inputs:
(1) Calibration gas pressure decrease alarm, (2) Range switching,
(3) External calibration start, (4) Process alarm (if this signal is received,
the heater power turns off), (5) Blow back start
Contact capacity:
Off-state leakage current; 3 mA or less
Self-diagnosis: Abnormal cell, abnormal cell temperature (low/high), abnormal
calibration, defective A/D converter, defective digital circuit
Calibration: Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (All are operated
interactively with an LCD touchscreen). Either zero or span can be skipped.
Zero calibration gas concentration setting range: 0.3 to 100 vol%O2 (in increments of
0.01 vol%O2 in smallest units).
Span calibration gas concentration setting range: 4.5 to 100 vol%O2 (in increments of
0.01 vol%O2 in smallest units).
Use nitrogen-balanced mixed gas containing 0 to 10 vol%O2 scale of
oxygen, and 80 to 100 vol%O2 scale of oxygen for standard zero gas
and standard span gas respectively.
Calibration interval; date/time setting: maximum 255 days
Model and Codes
Suffix code
Option code
ZR402G
Model
----------------
------------
Separate type Zirconia Oxygen Analyzer, Converter
Converter
thread
-P
-G
-M
-T
---------------------------------------------
G1/2
Pg13.5
M20x1.5
1/2NPT
--------------------------------------------------------
Japanese
English
German
French
Chinese
----------------------------------
Japanese
English
Chinese
------------
Always -A
/H
/SCT
/PT
/C2
Hood (*2)
Stainless steel tag plate (*1)
Printed tag plate (*1)
Failure alarm down-scale:
Output status at CPU failure and hardware error is 3.6 mA or less (*3)
Failure alarm up-scale:
Output status at CPU failure and hardware error is 21.0 mA or more (*3)
EAC with PA (*4)
EAC (*4)
-J
-E
-G
-F
-C
Display
-J
-E
-C
Instruction manual
-A
—
Options
Tag plates
NAMUR NE43
compliant
Standard
*1
*2
*3
*4
/C3
/EQ
/ER
Description
Specify either /SCT or /PT option code.
Sun shield hood is still effective even if scratched.
Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
“/EQ” is EAC with Pattern Approval for Russia. “/ER” is EAC for Kazakhstan and Belarus.
Standard Accessories
Item
Fuse
Parts No.
Q’ty
A1113EF
1
3.15A
For pipe, panel or wall mounting
Bracket for mounting
F9554AL
1
Screw for Bracket
F9123GF
1
Description
IM 11M13A01-02E
2-16
<2. Specifications>
External Dimensions
Unit: mm
1 to 6 (Panel Thickness)
2B mounting pipe
4 - Ø6 holes
for Wall mounting
100
126.5
280
EXA ZR402G
20.2
(*1/2NPT)
8
40
228
40 40
10
57.3
36
54.7
136.3
111
36 23
8-G1/2, *8-1/2NPT etc
(Wiring connection)
*: 1/2NPT with plug
38 24 14 38
( for wall mounting)
274 0
126.5
+2
4 - Ø6 holes
4-R8 to R10
or
4-C5 to C8
+2
190
Wall mounting
183 0
Panel Cut-out
F02-6E.ai
With sun shield hood (option code /H)
243
123 ±3
±4
155.5 ±3
39 ±2
55 ±2
205.5
±4
94.5
±4
±3
ZR402G
251.5
Material of Hood : Aluminum
F02-7E.ai
IM 11M13A01-02E
2-17
<2. Specifications>
2.5
ZA8F Flow Setting Unit and ZR40H
Automatic Calibration Unit
2.5.1
ZA8F Flow Setting Unit
This flow setting unit is applied to the reference gas and the calibration gas in a system
configuration (System 1). Used when instrument air is provided.
This unit consists of a flowmeter and flow control valves to control the flow rates of calibration gas
and reference gas.
Standard Specifications
Construction:
Dust-proof and rainproof construction
Case Material:
SPCC (Cold rolled steel sheet)
FIowmeter Scale:
Calibration gas; 0.1 to 1.0 L/min.
Reference gas; 0.1 to 1.0 L/min.
Painting:
Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections:
Rc1/4 or 1/4FNPT
Reference Gas Pressure: Clean air supply of sample gas pressure plus approx. 50 kPaG
(or sample gas pressure plus approx. 150 kPaG when a check valve
is used). Pressure at inlet of the Flow Setting Unit.(Maximum 300 kPaG)
Reference Gas Consumption:
Approx. 1.5 L/min
Calibration Gas (zero gas, span gas) Consumption:
Approx. 0.7 L/min (at calibration time only)
Weight:
Approx. 2.3 kg
NOTE
Use instrument air for span calibration gas, if no instrument air is available, contact YOKOGAWA.
Model and Codes
Model
Suffix code
Option code
ZA8F
--------
---------
Flow setting unit
Joint
-J
-A
-----------------
Rc 1/4
With 1/4 NPT adapter
---------
Style C
Style code
*C
Description
IM 11M13A01-02E
2-18
<2. Specifications>
External Dimensions
180
ø6 Hole
Unit : mm
140
7
REFERENCE CHECK
REFERENCE SPAN
235.8
JIS 50A (60.5mm)
mounting pipe
ZERO
222.8
Calibration gas outlet
Span gas inlet
Reference gas outlet
Zero gas inlet
32
Piping connection port A
REF
OUT
70
35
20
AIR
IN
CHECK
OUT
35
35
8
SPAN
IN
ZERO
IN
35
35
20
Model
Piping connection port A
ZA8F-J*C
5 - Rc1/4
ZA8F-A*C
5 - 1/4 NPT
Instrument air inlet
Weight : Approx. 2.3 kg
PIPNG INSIDE THE FLOW SETTING UNIT
CHECK
OUT
REF
OUT
Flowmeter
Flowmeter
AIR IN
ZERO
GAS IN
SPAN
GAS IN
Instrument air
Approx. 1.5 L/min.
Air Set
Air pressure ;
without check valve ; sample gas pressure + approx.50 kPaG
with check valve ; sample gas pressure + approx.150 kPaG
F02-8E.ai
IM 11M13A01-02E
2-19
<2. Specifications>
2.5.2
ZR40H Automatic Calibration Unit
CAUTION
Automatic Calibration Unit (Model ZR40H) must not be located in hazardous area.
This automatic calibration unit is applied to supply specified flow of reference gas and calibration
gas during automatic calibration to the detector in a system configuration (System 2).
• Specifications
Used when auto calibration is required for the separate type and instrument air is provided.
The solenoid valves are provided as standard.
Construction:
Dust-proof and rainproof construction:
NEMA4X/IP67-only for case coating solenoid valve (excluding
flowmeter)
Mounting:
2-inch pipe or wall mounting, no vibration
Materials: Body;
Piping;
Flowmeter;
Bracket;
Aluminum alloy,
316 SS (JIS), 304 SS (JIS),
MA (Metha acrylate resin).
304 SS (JIS)
Finish:
Polyurethane corrosion-resistance coating, mint green (Munsell
5.6BG3.3/2.9)
Piping Connection:
Refer to Model and Codes
Power Supply:
24 V DC (from ZR402G), Power consumption; Approx.1.3 W
Reference Gas Pressure: Sample gas pressure plus approx. 150 kPa (690 kPa max.), (Pressure
at inlet of the Automatic Calibration Unit)
Air Consumption:
Approx. 1.5 L/min
Weight:
Approx. 3.5 kg
Ambient Temperature: -20 to +55°C, no condensation or freezing
Ambient Humidity:
0 to 95%RH
Storage Temperature: -30 to +65°C
• Model and Codes
Model
Suffix code
Option code
ZR40H
-------------
--------
Automatic calibration unit for ZR402G
Gas piping
connection
-R
-T
---------------
Rc 1/4
1/4 NPT
-----------------------------
Pipe connection (G1/2)
Pg 13.5
20 mm (M20 x 1.5)
1/2 NPT
--------
Always -A
Wiring
connection
—
-P
-G
-M
-T
-A
Description
IM 11M13A01-02E
2-20
<2. Specifications>
External Dimensions
Unit: mm
2B pipe mounting example
Wiring inlet ; 2-G1/2,Pg13.5,M20X1.5 or 1/2NPT(Female)
*1 with four ISO M6 screws can wall-mount
54
41.2
90
*1
4 - Ø6.5
41.2
116.5
71.5
140
12
26
(wiring inlet is at same position on rear)
49.5
Connection port
Flowmeter
Needle valve
OCK
250
223
Terminal box
Setting Valve for
reference gas
Setting Valve for
calibration gas
2B mounting pipe
Calibration gas outlet
Rc1/4 or 1/4 NPT(Female)
42
46
102
16
58 MAX
Zero gas inlet
Rc1/4 or 1/4 NPT(Female)
26
40
30
Reference gas outlet
Rc1/4 or 1/4 NPT(Female)
47.5
25
Reference gas inlet
Rc1/4 or 1/4 NPT(Female)
F02-9E.ai
Piping Diagram
ZR402G Converter
AC-Z
AC-S
AC-C
REF
OUT
CHECK
OUT
ZR40H Automatic
Calibration Unit
Zero
Flowmeter
Flowmeter
EV1
*
Span
EV2
EV1
Solenoid valve
EV2
Solenoid valve
AIR IN
ZERO GAS IN
Instrument air Approx. 1.5 L/min
*Needle valve is supplied as an accessory with flowmeter
F02-10E.ai
IM 11M13A01-02E
<2. Specifications>
2.6
2-21
ZO21S Standard Gas Unit
CAUTION
Standard Gas Unit (Model ZO21S) must not be located in hazardous area.
This is a handy unit to supply zero gas and span gas to the detector as calibration gas. It is used
in combination with the detector only during calibration.
The ZO21S does not conform to CE marking.
Standard Specifications
Function:
Portable unit for calibration gas supply consisting of span gas (air) pump, zero
gas cylinder with sealed inlet, flow rate checker and flow rate needle valve.
Sealed Zero Gas Cylinders (6 provided): E7050BA
Capacity:
1l
Filled pressure:
Approx. 686 kPaG (at 35°C)
Composition:
0.95 to 1.0 vol%O2 + N2 balance
Power Supply:
100, 110, 115, 200, 220, 240 V AC ±10%, 50/60 Hz
Power Consumption: Max. 5 VA
Case material:
SPCC (cold rolled steel sheet)
Paint Color: Mainframe; Munsell 2.0 GY3.1/0.5 equivalent
Cover;
Munsell 2.8 GY6.4/0.9 equivalent
Piping:
Ø6 х Ø4 mm flexible tube connection
Weight:
Approx. 3 kg
Model and Codes
Model
Suffix code
Option code
ZO21S
-----------
-------------
Standard gas unit
Power
supply
-2
-3
-4
-5
-7
-8
-------------------------------------------------------------------------
200 V AC 50/60 Hz
220 V AC 50/60 Hz
240 V AC 50/60 Hz
100 V AC 50/60 Hz
110 V AC 50/60 Hz
115 V AC 50/60 Hz
-------------------------
Japanese version
English version
-------------
Style A
Panel
-J
-E
*A
Style code
Description
1600
228
92
External Dimensions
354
Zero gas cylinder (6 cylinder): E7050BA
253
Unit: mm
Flow checker
Span gas valve
Zero gas valve
Gas outlet
F02-11E.ai
IM 11M13A01-02E
2.7
2-22
<2. Specifications>
Other Equipments
2.7.1
Ejector Assembly for High Temperature (E7046EC,
E7046EN)
This ejector assembly is used where pressure of sample gas for high temperature detector is
negative. This ejector assembly consists of an ejector, a pressure gauge assembly and a needle
valve.
Standard Specifications
Needle Valve
Connection:
Rc1/4 or 1/4 FNPT
Material:
316 SS (JIS)
Pressure Gauge Assembly
Material in Contact with Gas: 316 SS (JIS)
Case Material:
Aluminum alloy (Paint color; black)
Connection:
G3/8 x R1/4 or 1/4 FNPT
Scale:
0 to 100 kPaG
Ejector
Ejector Inlet Air Pressure: 29 to 69 kPaG
Air Consumption:
Approx. 30 to 40 L/min
Suction gas flow rate: 3 to 7 L/min
Connection to Blow:
Rc1/4, 304 SS (JIS)
Tube Connection:
Ø 6 / Ø 4 mm or 1/4 inch copper tube (stainless tube)
(Note) Pipe and connections are not provided.
Part No.
Description
E7046EC
Needle valve; Rc1/4, Ejector; Ø6 / Ø4 TUBE joint, Pressure gauge; R1/4
E7046EN
Needle valve; 1/4 FNPT, Ejector; 1/4 TUBE joint, Pressure gauge; 1/4 NPT
Dimension
Pressure gauge assembly
39
Pressure gauge
Instrument air inlet
<1>
40
Approx. 88
Ø43
Full open
height
Connector (Tee) is not
included in ejector assembly.
<3>
<2>
Nozzle (Note1)
R1/2
Approx. 70
38
Ejector
Needle valve
Approx. 67
Detector
20
Tee
Blow Rc1/4
(Note1) The connecter of ejector is a dedicated
connecter with nozzle function.
<1> Rc1/4 or 1/4 FNPT
<2> Ø6/Ø4 or 1/4 inch copper tube (stainless)
with ejector to connect
<3> R1/4 or 1/4 NPT
F02-12E.ai
IM 11M13A01-02E
<2. Specifications>
Needle valve
Pressure gauge
Po
Qa
Air source
2-23
L
Ejector
Pg
Qg
Graph 1
Po (kPa)
200
100
Graph 2
Qa (l/min)
40
P= 0.5
0
30
5
10
15
0
L (m)
Pressure setting characteristics
Graph 3
Pg (kPa)
40
P (kPa)
60
80
Air consumption characteristics
-1.0
Qg (l/min)
8
-0.5
4
0
0
Graph 4
Gas pressure: 0 kPa
Sample Gas
Po (kPa)
P
(kPa)
Pg (kPa)
Qa (l/min)
Qg (l/min)
L
(m)
Gas pressure : -150 Pa
40
60
80
P (kPa)
Suction pressure characteristics
40
60
: Pressure setting
: Drive pressure (at the ejector entrance)
: Suction pressure
: Air consumption
: Suction flow
: Distance between the ejector and the pressure
gauge
80
P (kPa)
Suction flow characteristics
F02-13E.ai
< Pressure setting for the ejector for high temperature use >
Pressure supply for the ejector should be set so that the suction flow of the sample gas becomes
approximately 5 L/min.
To set this, proceed as follows:
(1) In Graph 4, draw a horizontal line from the 5 L/min point on the vertical axis (Suction flow:
Qg) toward the gas pressure line to be used, to find the point of intersection. Draw a line
vertically down from the point of intersection to the axis to find the drive pressure, P (at the
ejector entrance).
(2) In Graph 1, determine Po (pressure setting) from L (the distance between the ejector and
the pressure gauge).
(3) Open the needle valve to supply air for the ejector to the pressure gauge until it indicates the
pressure setting, Po.
NOTE
Qg (the suction flow) may require change according to the conditions of use. Refer to Subsection
3.2.2 and Subsection 4.1.4 for details.
Graph explanation
1) Graph 1 is to compensate for pressure loss in piping between the ejector and the pressure
gauge, and find Po (pressure setting).
2) Graph 2 shows correlation between P (drive pressure) and Qa (air consumption).
3) Graph 3 shows correlation between P (drive pressure) and Pg (suction pressure; when the
sample gas inlet of the ejector is closed).
4) Graph 4 shows correlation between P (drive pressure) and Qg (suction flow) for each gas
pressure.
2.7.2
Stop Valve (L9852CB, G7016XH)
This valve mounted on the calibration gas line in the system using ZA8F flow setting unit for
manual calibration.
IM 11M13A01-02E
<2. Specifications>
2-24
Standard Specifications
Material:
Connection:
Weight:
316 SS (JIS)
Rc 1/4 or 1/4 FNPT
Approx. 200 g
Part No.
Description
L9852CB
Joint: Rc 1/4, Material: 316 SS (JIS)
G7016XH
Joint: 1/4 FNPT, Material: 316 SS (JIS)
Unit : mm
Ø48
58
(Full open length)
Analyzer
Approx. 100
40
L9852CB 2-Rc1/4
G7016XH 2-1/4NPT
G7209XA 2-Rc1/4
K9470ZN 2-1/4NPT
F02-14.1E.ai
2.7.3
Check Valve (K9292DN, K9292DS)
This valve is mounted on the calibration gas line (directly connected to the detector).
This valve prevents the sample gas from entering the calibration gas line. Although it functions as
a stop valve, operation is easier than a stop valve as it does not require opening/closing at each
calibration.
Screw a check valve, instead of a stop valve into the calibration gas inlet of the detector.
IM 11M13A01-02E
<2. Specifications>
2-25
Standard Specifications
Material:
Connection:
Pressure:
Weight:
304 SS (JIS)
Rc1/4 or 1/4 FNPT
150 kPaG or more and 350 kPaG or less
Approx. 90 g
Part No.
Description
K9292DN
Joint: Rc 1/4, Material: 304 SS (JIS)
K9292DS
Joint: 1/4 FNPT, Material: 304 SS (JIS)
K9292DN : Rc 1/4(A),R 1/4(B)
K9292DS : 1/4 FNPT(A),1/4 NPT(Male)(B)
A
Unit: mm
B
Approx. 54
Approx. 19
F02-15E.ai
2.7.4
Air Set
This set is used to lower the pressure when instrument air is used as the reference and span
gases.
Standard Specifications
• G7003XF, K9473XK
Primary Pressure:
Max. 1 MPaG
Secondary Pressure: 0.02 to 0.2 MPaG
Connection:
Rc1/4 or 1/4 FNPT (with joint adapter)
Weight:
Approx.1 kg
Part No.
Description
G7003XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XK
Joint: 1/4 FNPT (with joint adapter), Material: Zinc alloy, Adapter: 316 SS (JIS)
• G7004XF, K9473XG
Primary Pressure:
Max. 1 MPaG
Secondary Pressure: 0.02 to 0.5 MPaG
Connection:
Weight:
Rc1/4 or 1/4 FNPT with joint adapter
Approx. 1 kg
Part No.
Description
G7004XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XG
Joint: 1/4 FNPT (with joint adapter), Material: Zinc alloy, Adapter: 316 SS (JIS)
IM 11M13A01-02E
<2. Specifications>
2-26
External Dimensions
Unit : mm
View A
Panel cut dimensions
Horizontal
mounting
22 ø15
Vertical
mounting
40
+0.5
2-ø2.2-0
40
2-ø6.5
max. 55
2-ø6 screw depth 8
Max. 210
Ø74
Panel (Horizontal mounting)
Secondary
pressure
gauge
Panel (Vertical mounting)
Secondary
A
88
Primary
G7003XF, G7004XF: Rc 1/4
K9473XK, K9473XG: 1/4 FNPT
(with joint adapter)
Approx. 122
2.7.5
F02-16E.ai
Cylinder Pressure Reducing Valve (G7013XF,
G7014XF)
This pressure reducing valve is used with the zero gas cylinders.
Standard Specifications
Primary Pressure:
Max. 14.8 MPaG
Secondary Pressure: 0 to 0.4 MPaG
Connection:
Inlet; W22 14 threads, right hand screw
Outlet; Rc1/4 or 1/4 FNPT
Material:
Brass body
IM 11M13A01-02E
2-27
<2. Specifications>
Unit : mm
Approx.112
Secondary pressure gauge
Primary pressure gauge
Reducing valve handle
W22 (Right hand screw)
Inlet
ACH
O
IH
TAK
Stop valve
Secondary
safety valve
* Outlet
Primary safety valve
Approx. 59
Approx. 82
Approx. 174
Approx. 163
Part No.
G7013XF
G7014XF
2.7.6
* Outlet
Rc1/4
1/4 NPT female with adapter
F02-17E.ai
ZR22A Heater Assembly
Model and Codes
Style:2
Model
Suffix code
Option code
Description
ZR22A
---------
-----------
Heater Assembly for ZR22S
Length (*1)
-015
-040
-070
-100
-150
-200
-------------------------------------------------------------
0.15 m
0.4 m
0.7 m
1m
1.5 m
2m
---------------------
with Jig (*2)
None
-----------
Always-A
Jig for change
—
-A
-N
-A
*1 Suffix code of length should be selected as same as ZR22S installed.
*2 Jig part no. is K9470BX to order as a parts after purchase.
(Note) The heater is made of ceramic, do not drop or subject it to pressure stress.
Yokogawa shall not guarantee the heater assembly after its replacement.
IM 11M13A01-02E
<2. Specifications>
2-28
(K9470BX)
K9470BX
Jig for change
∅21.7
External Dimensions
Unit : mm
30
∅43
L ± 12
Model & Codes
L
Weight (kg)
ZR22A-015
302
Approx. 0.5
ZR22A-040
552
Approx. 0.8
ZR22A-070
852
Approx. 1.2
ZR22A-100
1152
Approx. 1.6
ZR22A-150
1652
Approx. 2.2
ZR22A-200
2152
Approx. 2.8
F02-18E.ai
IM 11M13A01-02E
<2. Specifications>
2-29
IM 11M13A01-02E
3-1
<3. Installation>
3. Installation
This chapter describes installation of the following equipment:
Section 3.1
Section 3.2
Section 3.3
Section 3.4
Section 3.5
General-purpose Detector (except ZR22S--015)
High Temperature Detector (ZR22S--015)
ZR402G Converter
ZA8F Flow Setting Unit
ZR40H Automatic Calibration Unit
CAUTION
ZR402G Converter and ZR40H Automatic Calibration Unit must not be located in hazardous area.
CAUTION
Requirement for explosion-proof use:
The ZR22S is connected to a ZR402G or AV550G*1 that is mounted in a non-hazardous area.
The ambient temperature is in the range -20 to +60°C. The surface temperature of the ZR22S is
not over the temperature class T2 (300°C)*2.
*1: Refer to IM 11M12D01-01E
*2: The terminal box temperature does not exceed 150°C.
Oxygen concentration of sample/reference/calibration gas shall not exceed that found in normal
air, typically 21 vol%.
CAUTION
• When connecting an AV550G, the AV550G should be installed in a non-hazardous area.
• When using options like automatic calibration or blow back, if the corresponding solenoid valves
are to be mounted in a hazardous location then be sure to use explosion-proof solenoid valves
and appropriate wiring in explosion-proof conduit.
Hazardous area
Non-hazardous area
Averaging Converter (AV550G)
Detector
Signal
Analog outputs:
Averaged and
individual outputs
Heater
Solenoid valve
Contact output
Calibration
Contact
contact outputs output
Contact input
Power supply
Calibration gas line
Flowmeter
Needle valve
Reference
gas line
Span gas cylinder
(Instrument air)
Solenoid valve
Needle valve
Flowmeter
Instrument air
Pressure reducing valve
Stop valve
Air set
Zero gas cylinder
F03-1E.ai
Figure 3.1
Connection with AV550G
Regarding the AV550G installation procedure, refer to IM 11M12D01-01E.
IM 11M13A01-02E
3-2
<3. Installation>
3.1
Installation of General-purpose Detector
3.1.1
Installation Location
The following should be taken into consideration when installing the detector:
(1) Easy and safe access to the detector for checking and maintenance work.
(2) An ambient temperature of not more than 60°C, and the terminal box should not be affected
by radiant heat.
(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The sample gas satisfies the specifications described in Chapter 2.
(6) No sample gas pressure fluctuations.
3.1.2
ATEX Flameproof Type
ZR22S-A Detector for use in hazardous locations:
Note 1: Applicable Standard: EN 60079-0: 2012+A11: 2013, EN 60079-1: 2014,
EN 60079-31: 2014
Certificate Number:
KEMA 04ATEX2156 X
The symbol “X” placed after certificate number indicates that the equipment is
subjected to special conditions for safe use. Refer to Note 6.
Type of protection:
Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
Equipment Group:
II
Category: 2GD
Temperature class for Ex “db”: T2
The maximum surface temperature for Ex “tb”: T300°C
Degree of protection of enclosure: IP66
Note 2: Wiring
• All wiring shall comply with local installation requirement.
Note 3: Operation
• Keep to “WARNING” on the Detector.
WARNING: DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE
IS PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD-READ
USER’S MANUAL (IM11M13A01-02)
FOR INSTALLATION AND SAFE USE, READ IM11M13A01-02
USE AT LEAST 150°C HEAT RESISTANT CABLES & CABLE
GLANDS
• Take care not to generate mechanical sparking when accessing to the detector
and peripheral devices in hazardous location
Note 4: Maintenance and Repair
• The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void the
certification.
IM 11M13A01-02E
3-3
<3. Installation>
Note 5: Cable Entry
• The threaded type of cable entry is marked beside the cable entry according to
the following markings.
Threaded type : Marking
M20 x 1.5 :
M
1/2 NPT :
N
• The cable entry devices (cable glands etc.) and blind plugs shall be in type of
protection Ex “db” or Ex “tb”, suitable for the conditions of use and correctly
installed. They shall provide a degree of ingress protection of at least IP66.
Note 6: Special conditions for safe use
Electrostatic charge may cause an explosion hazard. Avoid any actions that
cause the generation of electrostatic charge, such as rubbing with a dry cloth.
If it is mounted in the area where the use of Ex “tb” apparatus is required, it
shall be installed in such a way that the risk from electrostatic discharges and
propagating brush discharges caused by rapid flow of dust is avoided.
Note 7: Special fastener
Hexagon socket head cap screws below are special fastener according to
EN60079-0: 2012+A11: 2013 (see the figure below). Material property classes
of them are A2-50 or better.
Special fastener
3.1.3
Figure 3.2
Special fastener
FM Explosion-proof Type
ZR22S-B Detector for use in hazardous locations:
Note 1: Applicable Standard: FM3600 1998,FM3615 1989, FM3810 2005, ANSI/
NEMA 250 1991
Type of protection:
Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating:
NEMA 4X
Temperature Class:
T2
Note 2: Wiring
• All wiring shall comply with National Electrical Code ANSI/NEPA 70 and Local
Electrical Code.
• In hazardous location, wiring shall be in conduits as shown in the figure.
WARNING: SEAL ALL CONDUITS WITHIN 18 INCHES OF THE ENCLOSURE.
IM 11M13A01-02E
<3. Installation>
Hazardous Locations Division 1
18 inches (475 mm) MAX.
ZR22S Detector
Figure 3.3
Conduit
3-4
Non-hazardous Location Equipment
ZR402G or AV550G
Sealing Fitting
F03-2E.ai
Wiring of ZR22S-B
Note 3: Operation
• Keep the “WARNING” label to the Detector.
WARNING: OPEN CIRCUIT BEFORE REMOVING COVER. USE AT LEAST
150°C HEAT RESISTANT CABLES.
• Take care not to generate mechanical sparking when accessing to the detector and
peripheral devices in hazardous location
Note 4: Maintenance and Repair
• The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void Factory
Mutual Explosion-proof Approval.
3.1.4
CSA Explosion-proof Type
ZR22S-C Detector for use in hazardous locations:
Note 1: Applicable Standard: C22.2 No.0-M1991, C22.2 No.0.4-04,
C22.2 No.0.5-1982, C22.2 No.25-1966,
C22.2 No.30-M1986, C22.2 No.94-M91,
C22.2-No.61010-1-04
Certificate Number: 1649642
Type of protection: Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure: Type 4X
Temperature Class: T2
Note 2: Wiring
• All wiring shall comply with Canadian Electrical Code Part 1 and Local Electrical Code.
• In hazardous location, wiring shall be in conduits as shown in the figure.
WARNING: SEAL ALL CONDUITS WITHIN 50 cm OF THE ENCLOSURE.
UN SELLE DOIT ÊTRE INSTALLÊ ÁMOINS DE 50 cm
DU BÎTIER.
Hazardous Locations Division 1
50 cm MAX.
ZR22S Detector
Figure 3.4
Conduit
Sealing Fitting
Non-hazardous Location Equipment
ZR402G or AV550G
F03-3E.ai
Wiring of ZR22S-C
IM 11M13A01-02E
3-5
<3. Installation>
Note 3: Operation
• Keep the “WARNING” label to the Detector.
WARNING: OPEN CIRCUIT BEFORE REMOVING COVER. REFER
TO IM 11M13A01-02E. USE AT LEAST 150ºC HEAT
RESISTANT CABLES.
OUVRIR LE CIRCUIT AVANT D’ENLEVER LE COUVERCLE.
UTILISEZ DES CÁBLES RÉSISTANTS Á 150ºC
MINIMUM. VEUILLEZ VOUS RÉFÉRER AU IM 11M13A01-02E.
• Take care not to generate mechanical sparking when accessing to the detector and
peripheral devices in hazardous location
Note 4: Maintenance and Repair
• The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void Canadian
Standards Explosion-proof Certification.
3.1.5
IECEx Flameproof Type
ZR22S-D Detector for use in hazardous locations:
Note 1: Applicable Standard: IEC 60079-0: 2011, IEC 60079-1: 2014, IEC 60079-31:
2013
Certificate Number:
IECEx KEM 06.0006X
The symbol “X” placed after certificate number indicates that the equipment is
subjected to special conditions for safe use. Refer to Note 6.
Type of protection:
Ex db IIB+H2 T2 Gb, Ex tb IIIC T300°C Db
Temperature class for Ex “db”: T2
The maximum surface temperature for Ex “tb”: T300°C
Degree of protection of enclosure: IP66
Note 2: Wiring
• All wiring shall comply with local installation requirement.
Note 3: Operation
• Keep to “WARNING” on the Detector.
WARNING: DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE
IS PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD-READ
USER’S MANUAL (IM11M13A01-02)
FOR INSTALLATION AND SAFE USE, READ IM11M13A01-02
USE AT LEAST 150°C HEAT RESISTANT CABLES & CABLE
GLANDS
• Take care not to generate mechanical sparking when accessing to the analyzer
and peripheral devices in hazardous area.
Note 4: Maintenance and Repair
• The analyzer modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void the
certification.
IM 11M13A01-02E
<3. Installation>
3-6
Note 5: Cable Entry
• The threaded type of cable entry is marked beside the cable entry according to
the following markings.
Threaded type : Marking
M20 × 1.5 :
M
1/2 NPT :
N
•
In case of ANSI 1/2 NPT plug, ANSI hexagonal wrench should be applied to
screw in.
• The cable entry devices (cable glands etc.) and blind plugs shall be in type of
protection Ex “db” or Ex “tb”, suitable for the conditions of use and correctly
installed. They shall provide a degree of ingress protection of at least IP66.
Note 6: Special conditions for safe use
Electrostatic charge may cause an explosion hazard. Avoid any actions that
cause the generation of electrostatic charge, such as rubbing with a dry cloth.
If it is mounted in the area where the use of Ex “tb” apparatus is required, it
shall be installed in such a way that the risk from electrostatic discharges and
propagating brush discharges caused by rapid flow of dust is avoided.
Note 7: Special fastener
Hexagon socket head cap screws below are special fastener according to IEC
60079-0: 2011 (see the figure below). Material property classes of them are A250 or better.
Special fastener
Figure 3.5
Special fastener
IM 11M13A01-02E
3.1.6
3-7
<3. Installation>
Probe Insertion Hole
Includes those analyzers equipped with a probe supporter and probe protector.
When preparing the probe insertion hole, the following should be taken into consideration:
CAUTION
• The outside dimension of detector may vary depending on its options. Use a pipe that is
large enough for the detector. Refer to Figure 3.6 for the dimensions.
• If the detector is mounted horizontally, the calibration gas inlet and reference gas inlet
should face downwards.
• The sensor (zirconia cell) at the tip of the detector may deteriorate due to thermal shock if
water drops are allowed to fall on it, as it is always at high temperature.
(1) Do not mount the probe with the tip higher than the probe base.
(2) The detector probe should be mounted at right angles to the sample gas flow or the probe
tip should point downstream.
Bounds of the probe
insertion hole location
(vertical)
Flange matches
the detector size
100 mm
*1
*1
Type
Standard
With probe protector
Outside diameter of detector
52 mm in diameter
60.5 mm in diameter
(horizontal)
100 mm
Four-hole flange
Figure 3.6
3.1.7
Eight-hole flange
F03-4E.ai
Example of forming probe insertion hole
Installation of the Detector
CAUTION
• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop the
detector, as impact will damage it.
• A gasket should be used between the flanges to prevent gas leakage. The gasket material
should be heatproof and corrosion-proof, suited to the characteristics of the sample gas.
The following should be taken into consideration when mounting the general-purpose detector:
(1) Make sure that the cell mounting screws (four bolts) at the tip of the detector are not loose.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the reference gas
inlet should face downward.
IM 11M13A01-02E
3-8
<3. Installation>
3.1.8
Installation of Probe Protector (ZO21R)
The detector is used with a probe protector to prevent the sensor from being worn by dust
particles when there is a high concentration of dust and gas flow exceeds 10 m/sec (pulverized
coal boiler or fluidized-bed furnace).
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector
in the probe insertion hole. The probe protector should be installed so that the notch is
downstream of the sample gas flow.
(2) Make sure that the sensor assembly mounting screws (four bolts) at the tip of the detector
are not loose.
(3) When the detector mounted horizontally, the reference gas and calibration gas inlet should
face downward.
Unit: mm
1050, 1550, 2050
Gasket (t1.5)
Direction of the sample gas flow
Probe top
Mount the protector so that the notch is
on the downstream side of the gas flow.
Figure 3.7
3.2
F03-5E.ai
Mounting of detector with a probe protector
Installation of High Temperature Detector
(Model ZR22S--015)
3.2.1
Installation Location
This detector is used with the High Temperature Probe Adapter (Model ZO21P) when the
temperature of sample gas exceeds 700°C, or when it is required due to maintenance spaces.
The following should be taken into consideration when installing the detector:
(1) Easy and safe access to the detector for checking and maintenance work.
(2) Ambient temperature of not more than 60°C and the terminal box should not be exposed to
radiant heat.
(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The sample gas should satisfy the specifications described in Chapter 2.
CAUTION
Be sure to read Subsection 3.1.2 to 3.1.5 where the important information on writing is provided.
IM 11M13A01-02E
3.2.2
3-9
<3. Installation>
Usage of High Temperature Probe Adapter
(Model ZO21P)
During analysis, the surface temperature of the probe adapter should be within the range from
more than the dew point of the sample gas and 300°C or less to prevent ejector clogging, gasket
deterioration or bolt scoring.
Where the dew point of the sample gas is not known, keep within the range of more than 200°C
to less than 300°C.
The temperature shall be measured at the probe in the probe adapter and the surface of the blind
flange at the opposite side.
When the surface temperature is not within the above range, the following measures can be
taken to change the temperature.
CAUTION
When the surface temperature of ZR22S or ZO21P exceeds temperature class T2(300°C), this
system shall not satisfy explosion-proof requirement for outside of the furnace.
(1) When the furnace pressure is negative, lower the pressure setting to reduce induction flow
of the sample gas.
Refer to Subsection 2.7.1, Ejector Assembly for High Temperature, for the setting of
induction flow. When you reduce induction flow, ensure that the ejector inducts air when the
furnace pressure fluctuates.
(2) When the furnace pressure is positive, close the needle valve for the sample gas outlet to
reduce the exhaust gas flow. Refer to Subsection 4.1.5, Piping to the High Temperature
Probe Adapter.
(3) When the probe adapter is surrounded by a heat insulator, remove the heat insulator.
Ensure that the temperature of the probe adapter does not fall below the dew point of the
gas in winter.
(4) To prevent temperature rises due to radiant heat, insert heat insulator between the wall of
the furnace and the probe adapter.
(5) To prevent temperature rises from thermal conduction, place the mounting flange as far
from the wall of the furnace as possible.
CAUTION
When a heater is used in hazardous area, then be sure to use appropriate explosion-proof heater
and wiring conduit.
IM 11M13A01-02E
3-10
<3. Installation>
<When the surface temperature is less than 200°C or below the dew point of the sample
gas>
(1) When the furnace pressure is negative, raise the pressure setting to increase induction flow
of the sample gas.
Refer to Subsection 2.7.1, Ejector Assembly for High Temperature, for the setting of
induction flow.
If there is much dust in the gas, the ejector may become clogged as induction flow increases.
(2) When the furnace pressure is positive, open the needle valve of the sample gas outlet to
increase the gas flow.
Refer to Subsection 4.1.5, Piping to the High Temperature Probe Adapter.
(3) Warm the probe adapter. Refer to Subsection 4.1.5, Piping to the High Temperature Probe
Adapter.
(4) When the surface temperature is still less than 200°C or below the dew point of the sample
gas, even if the above measures have been taken, warm the probe adapter using a heat
source such as steam.
3.2.3
Probe Insertion Hole
A high temperature detector consists of a ZR22S--015 Detector and ZO21P High Temperature
Probe Adapter. When forming the probe insertion hole, the following should be taken into
consideration:
(1) If the probe is made of silicon carbide (SiC), the probe hole should be formed so that the
probe is mounted vertically (within ± 5° tilt).
(2) In the case where the probe is made of stainless steel and the probe adapter is to be
mounted horizontally, the probe hole should be formed so that the probe tip is not higher
than the probe base.
Figure 3.8 illustrates examples of the probe insertion hole.
JIS 5K 50 FF (equivalent) or
ANSI Class 150 4 RF
(equivalent) flange
100mm
A space of 52 mm or more
in diameter, and long enough
for the length of the probe
The allowable margin for probe
tilt is within ± 5°.
An SiC probe shall be mounted vertically.
Figure 3.8
100mm
A space of 52 mm or
more in diameter, and
long enough for the length JIS 5K 50 FF (equivalent)
of the probe
or ANSI Class 150 4 RF
(equivalent) flange
Never mount the probe
with its tip higher than
the base
Horizontal mounting is used with a SUS probe.
F03-6E.ai
Examples of the probe insertion hole
IM 11M13A01-02E
3-11
<3. Installation>
3.2.4
Mounting of High Temperature Detector
CAUTION
• Ceramic (zirconia) is used in the sensor (cell) portion on the detector probe tip. Care should
be taken not to drop the detector during installation.
• The same applies to the high temperature probe adapter with a probe made of silicon
carbide (SiC).
• A gasket should be used on the flange surface to prevent gas leakage. The gasket material
should be selected depending on the characteristics of the measurement gas. It should be
heatproof and corrosion-proof. The parts, which should be supplied by the user, are listed in
Table 3.1.
Table 3.1 Accessories for mounting high temperature probe adapter
Mounting flange specification
JIS 5K 50 FF (equivalent)
ANSI Class 150 4RF (equivalent)
Parts name
Q’ty
Note
Gasket
1
Heatproof and corrosion-proof
Bolt (M12 by 50)
4
Nut (M12)
4
Washer (for M12)
8
Gasket
1
Bolt (M16 by 60)
8
Nut (M16)
8
Washer (for M16)
16
Heatproof and corrosion-proof
A high temperature detector should be mounted as follows:
(1) It is recommended to mount the detector vertically. When it is impossible due to the physical
arrangements and the detector is mounted horizontally, ensure that the probe tip be placed
no higher than the probe base.
(2) When mounting a high temperature probe adapter, be sure to insert a gasket between the
flanges to prevent gas leakage. When the furnace pressure is negative, ensure that there is
no leakage of air into the detector.
(3) When mounting the detector in a position other than vertical, the cable inlet should face downward.
(4) When installing the detector in a low-temperature location such as in the open air, cover the
probe adapter including the ejector with a heat insulator (e.g. ceramic wool) to keep it warm
and to prevent condensation of drain on the ejector.
Wiring connection
Mount the probe so that
the probe tip is not higher
than the probe base.
Wiring connection
When using an SiC probe, mount the probe vertically
with an allowable tilt of ±5°.
F03-7E.ai
Figure 3.9
Mounting of the High Temperature Detector
IM 11M13A01-02E
3.3
3-12
<3. Installation>
Installation of ZR402G Converter
CAUTION
Converter (Model ZR402G) must not be located in hazardous area.
3.3.1
Installation Location
The following should be taken into consideration when installing the converter:
(1) Readability of the indicated values of oxygen concentration or messages on the converter
display. Easy and safe access to the converter for operating keys on the panel and for
checking and maintenance work.
(2) An ambient temperature of not more than 55 °C and little change in temperature
(recommended within 15 °C in a day).
(3) The normal ambient humidity (recommended between 40 to 75% RH) and without any
corrosive gases.
(4) No vibration.
(5) Near to the detector.
(6) Not in direct rays of the sun. If the sun shines on the converter, prepare the hood (/ H) or
other appropriate sunshade.
(7) Non-hazardous location.
3.3.2
Mounting of the Converter
The converter can be mounted on a pipe (nominal JIS 50A: O.D. 60.5 mm), a wall or a panel. The
converter can be mounted at an angle to the vertical, however, it is recommended to mount it on
a vertical plane.
Mount the converter as follows.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the converter. (Converter weighs approximately 6 kg.)
(2) Mount the converter on the pipe. Fix it firmly on the pipe in the procedure described in Figure 3.9.
Mounting procedure
1. Put four bolts in the holes on the fitting.
2. Clamp the pipe with the fitting and a bracket,
with the four bolts passing through the bracket holes.
3. Secure the fitting and the bracket tightly to the pipe with
four washers and nuts.
A pipe to be mounted (nominal JIS 50A)
Bracket
*Fitting
*Washer
Nut
*Bolt
* Note: These fittings are attached to the converter
when it is delivered.
Figure 3.10
Bolt
Washer
F03-8E.ai
Pipe Mounting
IM 11M13A01-02E
3-13
<3. Installation>
(1) Drill mounting holes through the wall as shown in Figure 3.11.
Unit: mm
Four holes 6 mm
in diameter
for M5 screws
126.5
190
F03-10E.ai
F03-9E.ai
Figure 3.11 Mounting holes
Figure 3.12 Wall Mounting
(2) Mount the converter. Secure the converter on the wall using four screws.
Note: For wall mounting, the bracket and bolts are not used.
(1) Cut out the panel according to Figure 3.13.
Unit: mm
+2
274 0
Washer
Bolt
+2
183 0
Figure 3.13 Panel cutout sizes
F03-11E.ai
Clamp screw
Fitting
F03-12E.ai
Figure 3.14 Panel mounting
(2) Remove the fitting from the converter by loosening the four screws.
(3) Insert the converter case into the cutout hole of the panel.
(4) Attach the mounting fitting which is once removed in step (2) again to the converter.
(5) Firmly fix the converter to the panel. Fully tighten the two clamp screws to hold the panel with
the fitting.
IM 11M13A01-02E
3.4
3.4.1
3-14
<3. Installation>
Installation of ZA8F Flow Setting Unit
Installation Location
The following should be taken into consideration:
(1) Easy access to the unit for checking and maintenance work.
(2) Near to the detector or the converter
(3) No corrosive gas.
(4) An ambient temperature of not more than 55 °C and little changes of temperature.
(5) No vibration.
(6) Little exposure to rays of the sun or rain.
3.4.2
Mounting of ZA8F Flow Setting Unit
The flow setting unit can be mounted either on a pipe (nominal JIS 50A) or on a wall. It should be
positioned vertically so that the flowmeter works correctly.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the flow setting unit. (The unit weighs approximately 2 to 3.5 kg.)
(2) Mount the flow setting unit on the pipe by tightening the nuts with the U-bolt so that the metal
fitting is firmly attached to the pipe.
F03-13E.ai
Figure 3.15 Pipe Mounting
(1) Make a hole in the wall as illustrated in Figure 3.16.
(2) Mount the flow setting unit. Remove the pipe mounting parts from the mount fittings of the
flow setting unit and attach the unit securely on the wall with four screws.
Unit : mm
223
140
4 - Ø6.5 hole, or M5 screw
F03-14E.ai
Figure 3.16 Mounting holes
F03-15E.ai
Figure 3.17 Wall mounting
IM 11M13A01-02E
3.5
<3. Installation>
3-15
Installation of ZR40H Automatic Calibration
Unit
CAUTION
Automatic Calibration Unit (Model ZR40H) must not be located in hazardous area.
3.5.1
Installation Location
The following should be taken into consideration:
(1) Easy access to the unit for checking and maintenance work.
(2) Near to the detector or the converter
(3) No corrosive gas.
(4) An ambient temperature of not more than 55 °C and little change of temperature.
(5) No vibration.
(6) Little exposure to rays of the sun or rain.
(7) Non-hazardous location.
3.5.2
Mounting of ZR40H Automatic Calibration Unit
The automatic calibration unit can be mounted either on a pipe (nominal JIS 50 A ) or on a wall.
It should be positioned vertically so that the flowmeter works correctly.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
of automatic calibration unit. (The unit weights approximately 3.5 kg.)
(2) Mount the automatic calibration unit on the pipe by tightening the nuts with the U-bolt so that
the metal fitting is firmly attached to the pipe.
F03-16E.ai
Figure 3.18 Pipe Mounting
IM 11M13A01-02E
<3. Installation>
3-16
(1) Make a hole in the wall as illustrated in Figure 3.19.
Unit : mm
223
140
4 - Ø6.5 hole, or M6 screw
F03-17E.ai
Figure 3.19
Mounting holes
(2) Mount the automatic calibration unit. Remove the U-bolt from the automatic calibration unit
and attach the unit on the wall with four screws. When setting it with M5 bolts, use washers.
4-Ø6.5
F03-18E.ai
Figure 3.20
Wall Mounting
IM 11M13A01-02E
3.6
3-17
<3. Installation>
Insulation Resistance Test
Even if the testing voltage is not so great that it causes dielectric breakdown, testing may cause
deterioration in insulation and a possible safety hazard. Therefore, conduct this test only when it
is necessary.
The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied for as
short a time as practicable to confirm that insulation resistance is 20 MΩ or more.
Remove wiring from the converter and the detector.
1. Remove the jumper plate located between terminal G and the protective grounding terminal.
2. Connect crossover wiring between L and N.
3. Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to the
crossover wiring, and (-) terminal to ground.
4. Turn the insulation resistance tester ON and measure the insulation resistance.
5. After testing, remove the tester and connect a 100 kΩ resistance between the crossover
wiring and ground, to discharge for over 1 min. During discharge, do not touch the terminal.
6. Testing between the heater terminal and ground, contact output terminal and ground, analog
output/input terminal and ground can be conducted in the same manner.
7. Although contact input terminals are isolated, insulation resistance test cannot be conducted
because the breakdown voltage of the surge-preventing arrestor between the terminal and
ground is low.
8. After conducting all the tests, replace the jumper plate as it was.
Insulation
resistance tester
-
ZR402G Converter
ZR22S Detector
+
crossover wiring
1
FG
12
FG
2
3
4
5
6
7
8
9
10
11
AO1 AO1 AO2 AO2 CJ
CJ
TC
TC CELL CELL
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
13
14
15
16
17
18
19
20
21
22
DI-1 DI-2 DI-C
AI
AI
AC- AC- ACFG
FG
(+)
(-) ZERO SPAN COM
CELL CELL
(+)
(-)
TC
(+)
TC
(-)
CJ
(+)
CJ
(-)
Contact
input 2
Contact
input 1
Solenoid valve
for automatic
calibration
Span gas
Solenoid valve for automatic
calibration Zero gas
Automatic Calibration Unit ZR40H
23
24
25
26
27
28
29
30
31
32
33
DO-1 DO-1 DO-2 DO-2 DO-3 DO-3 DO-4 DO-4 HTR HTR
L
34
N
35
G
36
HTR
HTR
Remove
jumper plate
Remove wiring to connect external circuits
and Connect crossover wiring
crossover wiring
Insulation
resistance tester
-
+
Insulation
resistance tester
-
+
F03-19E.ai
IM 11M13A01-02E
Blank Page
<4. Piping>
4-1
4. Piping
This chapter describes piping procedures based on two typical system configurations for EXAxt
ZR Separate type Explosion-proof Zirconia Oxygen Analyzer.
• Ensure that each check valve, stop valve and joint used for piping do not allow leakage.
Especially, if there is any leakage of the calibration gas from pipes and joints, it may cause
clogging of the pipes or incorrect calibration.
• Be sure to conduct leakage test after piping.
• Basically, apply instrument air (dehumidified by cooling to the dew point -20°C or lower, and
removing any dust, oil mist and the like) for the reference gas.
CAUTION
Do not loosen or remove any Flame Arrestor of gas inlet/outlet during piping.
The detector modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void ATEX Certification, FM Approval, CSA
Certification and IECEx Certification.
Flame Arrestor
F04E.ai
IM 11M13A01-02E
4.1
4-2
<4. Piping>
Piping for a System Using Flow Setting Unit
for Manual Calibration
The piping for a system using flow setting units for manual calibration is shown in Figure 4.1.
Hazardous Area
ZR402G Converter
ZR22S Separate type Explosion-proof
Zirconia Oxygen Analyzer, Detector
Stop valve
or
Check valve
Reference gas
Non-hazardous Area
EXA ZR402G
Flowmeter Needle
valve
Air Set
Calibration gas
Instrument air
Span gas (*)
ZA8F
Flow Setting Unit
* Calibration gas unit same as for zero gas.
Pressure
reducing
valve
~ 100 to 240 V AC
Contact input
Analog output,
contact output
Digital output
(HART)
Zero gas
cylinder
Calibration gas
unit case
F04-1E.ai
Figure 4.1 Typical Piping for System Using Flow Setting Unit for Manual Calibration
The following outlines some points to note regarding the piping for this system.
• Connect a stop valve or check valve to the nipple at the calibration gas inlet of the detector.
• When a high temperature detector is used and the sample gas pressure is negative,
connect an auxiliary ejector to the sample gas exhaust hole of the high temperature probe
adapter (see Subsection 4.1.5, Figure 4.3).
• When a high temperature detector is used and the pressure of the measured gas is 0.5
kPa or higher, it is recommended that a needle valve (throttle) be used in the sample gas
exhaust of the high temperature probe adapter (see Subsection 4.1.5, Figure 4.4).
CAUTION
This is for lowering the sample gas temperature below 700°C. If the gas temperature is high and
the pressure is also significantly high, the sampled gas temperature may not reduced below
700ºC when reaching the detector.
On the other hand, if the sample gas temperature is lowered too much, condensation may be
produced in the high temperature probe adapter. During wintertime, it is recommended that
the high temperature probe adapter should be protected with an insulating material to prevent
condensation (see Figure 4.5).
For the usage of the high temperature probe adapter, refer to Subsection 3.2.2.
• When a high temperature detector is used and blow back is required to eliminate dust
accumulating in the probe of the high temperature probe adapter, air supply piping for purging
should be installed.
IM 11M13A01-02E
4-3
<4. Piping>
CAUTION
If a sample gas contains much dust (e.g., in recovery boilers or cement kilns), the probe is
more likely to become clogged. To eliminate this dust accumulation using air pressure, piping is
generally installed from an air source only when cleaning is performed. Some cases, however,
may need a permanent installation of the blow back piping. See Subsection 4.2.5 for the
installation of blow back piping.
4.1.1
Parts Required for Piping in a System Using Flow
Setting Units for Manual Calibration
Referring to Table 4.1, check that the parts required for your system are ready.
Table 4.1
Detector
Piping location
General-purpose
detector
Calibration gas inlet
Parts Name
Remark
Stop valve or check
valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Provided by YOKOGAWA (K9292DN or K9292DS)
Nipple *
R1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
General parts
Pressure reducing valve Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection R1/4 or 1/4 NPT
Reference gas inlet
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection R1/4 or 1/4 NPT
High temperature
detector (0.15 m)
Calibration gas inlet
General parts
General parts
Stop valve or check
valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Provided by YOKOGAWA (K9292DN or K9292DS
Nipple *
R1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
General parts
Pressure reducing valve Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection R1/8 or 1/8 NPT
Reference gas inlet
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection R1/4 or 1/4 NPT
Sample gas outlet
General parts
Ejector assembly *
Recommended by YOKOGAWA
(E7046EC or E7046EN)
T-shaped joint of the
same diameter *
Rc1/4 or 1/4 NPT
General parts
Needle valve *
Rc1/4 or 1/4 NPT
General parts
Reducing nipple *
R1/2 to R1/4 or R1/2 to 1/4 NPT General parts
Note: Parts with marking * are used when required. General parts can be found on the local market.
IM 11M13A01-02E
4.1.2
4-4
<4. Piping>
Piping for the Calibration Gas Inlet
This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and
between the ZA8F flow setting unit and the ZR22S detector.
The cylinder should be placed in a calibration gas unit case or the like to avoid any direct sunlight
or radiant heat so that the gas cylinder temperature does not exceed 40°C. Mount a pressure
reducing valve (specified by YOKOGAWA) on the cylinder.
Mount a check valve or stop valve (specified by YOKOGAWA) on the nipple (found on the local
market) at the calibration gas inlet of the detector as illustrated in Figure 4.2.
(The check valve or the stop valve may have been mounted on the detector when shipped.)
Connect the flow setting unit and the detector to a stainless steel pipe 6 mm (O.D.) × 4 mm (I.D.)
or larger (or nominal size 1/4 inch).
Check valve
Stop valve
Nipple
Tube
connection
joint
Piping for the Calibration gas inlet,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe
Piping for the Reference gas inlet,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe
Piping for the Reference gas outlet,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe
F04-2E.ai
Figure 4.2 Piping for the Calibration Gas Inlet
4.1.3
Piping for the Reference Gas Inlet
Reference gas piping is required between the air source (instrument air) and the ZA8F flow
setting unit, and between the ZA8F flow setting unit and the ZR22S detector.
Insert the air set next to the flow setting unit in the piping between the air source and the flow
setting unit.
Use a 6 mm (O.D.) × 4 mm (I.D.) or larger (or nominal size 1/4 inch) stainless steel pipe between
the flow setting unit and the detector.
4.1.4
Piping for the Reference Gas Outlet
If the ZR22S is exposed to rain or water splash, connect the pipe outlet on downward.
4.1.5
Piping to the High Temperature Probe Adapter
• The sample gas should be at a temperature below 700°C before reaching the detector
sensor. If the gas is under negative pressure, it should be fed to the detector by suction.
• For usage of the high temperature probe adapter when using high temperature detector,
refer to Subsection 3.2.2.
• If the sample gas is under negative pressure, connect the ejector assembly (E7046EC/
E7046EN) as illustrated in Figure 4.3. Mount the pressure gauge as close as possible to the
ejector. However, if the ambient temperature is too high, mount the gauge in a location with
a temperature below 40°C.
IM 11M13A01-02E
4-5
<4. Piping>
Pressure gauge
Ejector
Ejector assembly for
high temperature
Detector
High temperature
probe adapter
F04-3E.ai
Figure 4.3 Mounting the ejector assembly
If the temperature of the sample gas is high and its pressure exceeds 0.49 kPa, the temperature
of the sample gas at the detector may not be below 700°C.
In such a case, connect a needle valve (found on the local market) through a nipple (also found
on the local market) to the sample gas outlet (Rc1/2) of the probe adapter so that the sample gas
exhaust volume is restricted.
In cases where condensation is likely to occur in the probe adapter when the sample gas is
cooled, protect the probe adapter with an insulating material as illustrated in Figure 4.5.
Reducing nipple
(R1/2-R1/4 or R1/2-1/4 NPT)
Needle valve
F04-4E.ai
Figure 4.4 Mounting the needle valve for restricting exhaust flow of the sample gas
Cover flange
Sample gas outlet
High temperature probe adapter
Detector
Insulating material
Probe
F04-5E.ai
Figure 4.5 Preventing to condensation
IM 11M13A01-02E
4.2
4-6
<4. Piping>
Piping for a System to Perform Automatic
Calibration
The piping for a system to perform automatic calibration is shown in Figure 4.6.
The piping is basically the same as that of a system using flow setting unit for manual calibration.
Refer to Section 4.1.
Hazardous Area
ZR22S Separate type Explosion-proof
Zirconia Oxygen Analyzer, Detector
Non-hazardous Area
ZR402G Converter
EXA ZR402G
Check valve
Signal
(6-core shield cable)
~ 100 to 240 V AC
Heater (2-core cable)
Reference gas
Flowmeter Needle
valve
Air Set
Calibration gas
Contact input
Analog output, contact output
Digital output (HART)
Instrument air
ZR40H
Automatic Calibration Unit
Pressure
reducing
valve
Calibration gas
unit case
Zero gas cylinder
F04-6E.ai
Figure 4.6 Typical Piping for a System to perform Automatic Calibration
IM 11M13A01-02E
4.2.1
4-7
<4. Piping>
Parts Required for Piping in a System to Perform
Automatic Calibration
Referring to Table 4.2, check that the parts required for your system are ready.
Table 4.2
Detector
Piping location
General-purpose
detector
Calibration gas inlet
Parts Name
Remark
Check valve
Recommended by YOKOGAWA
(K9292DN or K9292DS)
Nipple *
R1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
General parts
Pressure reducing valve Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection R1/4 or 1/4 NPT
Reference gas inlet
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection R1/4 or 1/4 NPT
High temperature
detector (0.15 m)
Calibration gas inlet
General parts
General parts
Check valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Provided by YOKOGAWA (K9292DN or K9292DS
Nipple *
R1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
General parts
Pressure reducing valve Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection R1/8 or 1/8 NPT
Reference gas inlet
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection R1/4 or 1/4 NPT
Sample gas outlet
General parts
Ejector assembly *
Recommended by YOKOGAWA
(E7046EC or E7046EN)
T-shaped joint of the
same diameter *
Rc1/4 or 1/4 NPT
General parts
Needle valve *
Rc1/4 or 1/4 NPT
General parts
Reducing nipple *
R1/2 to R1/4 or R1/2 to 1/4 NPT General parts
Note: Parts with marking * are used when required. General parts can be found on the local market.
IM 11M13A01-02E
4-8
<4. Piping>
4.2.2
Piping for the Calibration Gases
The piping for the calibration gases should be installed between the calibration gas cylinders (or
instrument air source) and the detectors with the ZR40H automatic calibration unit.
Figure 4.7 shows the ZR40H Automatic Calibration Unit piping diagram.
Adjust secondary pressure of both the air set and the zero gas pressure reducing valve so that
these two pressures are approximately the same. The flow rate of zero and span gases (normally
instrument air) are set by a single needle valve.
After installation and wiring, check the calibration contact output (see Subsection 7.11.2), and
adjust zero gas pressure reducing valve and calibration gas needle valve so that zero gas flow is
within the permitted range. Next check span gas calibration contact output and adjust air set so
that span gas flow is within the permitted range.
REF
OUT
CHECK
OUT
flowmeter
*Needle valve
flowmeter
*
Solenoid valve
ZERO GAS IN
AIR IN
To zero gas pressure
reducing valve
Instrument air
Approx.1.5 L/min.
Air set
*: Needle valve is supplied as an accessory with flowmeter
F04-7E.ai
Figure 4.7 ZR40H Automatic Calibration Unit piping diagram
If the contact input to the converter is used for the blow back function, prepare blow back piping
according to Subsection 4.2.5.
NOTE
Blow back function means the function to get rid of dust inside a probe in a high temperature
probe adapter by using compressed air, when a high temperature detector is used.
4.2.3
Piping for the Reference Gas
The piping for the reference gas should be installed between the air source (instrument air) and
the detector through flowmeters, needle valves of the ZR40H Automatic Calibration Unit.
•
se stainless steel pipes with 6 O.D. x 4 I.D. mm (or nominal 1/4 inch) or larger inside
U
diameter for the piping for the reference gas.
IM 11M13A01-02E
4.2.4
4-9
<4. Piping>
Piping to the High Temperature Probe Adapter
The piping to the high temperature probe adapter is required when a high temperature detector is
used.
This piping is the same as the one in the system using flow setting unit for manual calibration.
See Subsection 4.1.5.
4.2.5
Piping for Blow back
This piping is required when the blow back function is carried out. The piping described below
provides automatic blow back operation when the “ blow back start “ command is entered to the
converter.
Solenoid valve
Pressure reduction valve
Pipe junction
Blow pipe
F04-8E.ai
Figure 4.8 Blow back Piping
CAUTION
When mounting solenoid valves (e.g. used with options like automatic calibration or blow back) in
a hazardous area, then be sure to use explosion-proof solenoid valves and appropriate wiring in
explosion-proof conduit.
The following parts are required for blow back piping.
• Blow pipe (to be prepared as illustrated in Figure 4.11.)
• Two-way solenoid valve: “ Open “ when electric current is on. (Found on the local market)
• Air set (Yokogawa recommended: G7003XF / K9473XK or G7004XF / K9473XG)
IM 11M13A01-02E
<4. Piping>
4-10
Manufacture the blow pipe as illustrated in Figure 4.9, and mount it on the high temperature
probe adapter.
Unit : mm
Φ
90
4-Φ9
Blind flange of the high temperature probe adapter
8 ~ 10
Rc1/4
Φ73
Welded
Approximately
200
8 (O.D.) by 6 (I.D.) Stainless steel pipe
F04-9E.ai
Figure 4.9 Blow pipe Construction
IM 11M13A01-02E
5-1
<5. Wiring>
5. Wiring
In this Chapter, the wiring necessary for connection to the EXAxt ZR Separate type Explosionproof Zirconia Oxygen Analyzer is described.
5.1
General
CAUTION
• NEVER supply current to the converter or any other device constituting a power circuit in
combination with the converter, until all wiring is completed.
• This product complies with CE marking.
Where compliance with CE marking is necessary, the following wiring procedure is
necessary.
1. Install an external switch or circuit breaker to the power supply of the converter.
2. Use an external switch or circuit breaker rated 5 A and conforming with IEC 947-1 or IEC
947-3.
3. It is recommended that the external switch or circuit breaker be mounted in the same room
as the converter.
4. The external switch or circuit breaker should be installed within the reach of the operator,
and marked as the power supply switch of this equipment.
Wiring procedure
Wiring should be performed according to the following procedure:
1. Be sure to connect the shield line to FG terminal of the converter.
2. The outer sheath of the signal line should be stripped to a length of 50 mm or less.
The most outer sheath of the power cable should be stripped to a length of 20 mm or less.
3. Signals may be affected by noise if signal lines, power cable and heater cable are located in
the same conduit. When using conduit, signal lines should be installed in a separate conduit
from power and heater cables.
4. Install metal blind plug(s) in unused cable connection gland(s) of the converter.
5. Metal conduit should be grounded.
6. The following cables are used for wiring:
Table 5.1 Cable specifications
Terminal name of converter
Name
CELL+, CELLHTR TC+, HTR TCCJ+, CJ-
Detector signal
HEATER
Detector heater
L, N
Power supply
AO-1+, AO-1-, AO-2+, AO-2- Analog output
Need for shields
Number of cores
О
6
2
2 or 3 *
О
2 or 4
DO-1, DO-2, DO-3, DO-4
Contact output
AC-Z, AC-S, AC-C
Automatic calibration unit
3
DI-1, DI-2, DI-C
Contact input
3
Note *:
2 to 8
When the case is used for protective grounding, use a 2-wire cable.
IM 11M13A01-02E
5-2
<5. Wiring>
WARNING
Cables that withstand temperatures at least 80°C should be used for wiring.
CAUTION
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3) grounding
(the grounding resistance is 100 Ω or less).
• Special consideration of cable length should be taken for the HART communication, For the
detail, refer to Subsection 1.1.2 of the IM 11M12A01-51E “ Communication Line Requirements “.
5.1.1
Terminals for the External Wiring in the Converter
Open the front door and remove the terminal covering plate to gain access to the converter
external wiring terminals (see Figure 5.1).
CAUTION
After wiring necessary cable to the converter terminals, be sure to fix the terminal covering plate
with two screws again.
Front door
Terminals are visible
when the terminal cover is removed.
Cable inlet
F05-1E.ai
Figure 5.1 Terminals for external wiring in the converter
IM 11M13A01-02E
5-3
<5. Wiring>
5.1.2
Wiring
Connect the following wiring to the converter. It requires a maximum of seven wiring connections
as shown below.
(1) Detector output (connects the converter with the detector.)
(2) Detector heater power (connects the converter with the detector.)
(3) Analog output signal
(4) Power and ground
(5) Contact output
(6) Operation of the solenoid valve of automatic calibration unit
(7) Contact input
ZR22S Separate type Explosion-proof
Zirconia Oxygen Analyzer, Detector
ZR402G Separate type
Zirconia Oxygen Analyzer, Converter
Analog
output 1
4-20 mA DC
Digital output
Analog
output 2
4-20 mA DC
1
FG
12
FG
2
3
4
5
6
7
8
9
10
11
AO-1 AO-1 AO-2 AO-2 CJ
CJ
TC
TC CELL CELL
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
13
14
15
16
17
18
19
20
21
22
DI-1 DI-2 DI-C AI
AI
AC- AC- AC- FG
FG
(+)
(-) ZERO SPAN COM
Contact
input 2
Solenoid valve for automatic
calibration Zero gas
Automatic Calibration Unit ZR40H
23
24
25
26
27
28
29
30
31
32
33
DO-1 DO-1 DO-2 DO-2 DO-3 DO-3 DO-4 DO-4 HTR HTR
L
Contact
output 1
Contact
output 2
TC
(+)
TC
(-)
CJ
(+)
CJ
(-)
Solenoid valve
for automatic
calibration
Span gas
Contact input 1
*1
CELL CELL
(+)
(-)
Contact
output 3
Contact
output 4
34
N
35
G
36
FG
HTR HTR
*1
100-240V AC
50/60 Hz
*1
F05-2E.ai
*1 The ground wiring of the converter should be connected to either the protective ground terminal in the equipment or
the ground terminal of the converter case.
Ground to earth, ground resistance: 100 Ω or less.
Figure 5.2 Wiring connection to the converter
IM 11M13A01-02E
5.1.3
<5. Wiring>
5-4
Mounting of Cable Gland
For each cable connection opening of the converter, mount a conduit that matches the thread
size, or a cable gland.
EXA ZR402G
8-G1/2, 8-1/2NPT
or the like
(Wiring connection)
Ground terminal (M4)
Adaptor for 1/2 NPT thread
F05-3E.ai
Figure 5.3 Cable gland mounting
IM 11M13A01-02E
5.2
5-5
<5. Wiring>
Wiring for Detector Output
This wiring enables the converter to receive cell output from the detector, output from a
thermocouple and a reference junction compensation signal. Install wires that allow for 10 Ω of
loop resistance or less. Keep detector wiring away from power wiring.
CAUTION
Be sure to read Subsection 3.1.2 to 3.1.5 where the important information on writing is provided.
(1) Surface temperature of the detector terminal box: 75°C or less
Hazardous area
Non-hazardous area
Detector
Converter
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
CJ(+)
TC(-)
CJ(-)
CJ(-)
CJ(+)
FG
Shielded cables
(2) Surface temperature of the detector terminal box: exceeding 75°C
Detector
Hazardous area
Terminal box
Non-hazardous area
Converter
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
TC(-)
CJ(+)
CJ(+)
CJ(-)
CJ(-)
FG
Heat-resistant wiring
Shielded cables
F05-4E.ai
Figure 5.4 Wiring for detector output
CAUTION
If shielded cables cannot be used between the detector and the terminal box, for example,
when heat-resistant wiring is used, locate the detector and the terminal box as close together as
possible.
5.2.1
Cable Specifications
During operation of the process, the terminal box may get quite hot and may reach temperatures
of up to 150°C. The cable may be exposed to even higher temperatures, so be sure to use
suitably heat-resistant cable.
Basically, a cable (six core) that withstand temperatures of at least 80°C is used for this wiring.
When the ambient temperature of the detector exceeds 75°C, install a terminal box, and connect
with the detector using six-piece 600V silicon rubber insulated glass braided wire.
IM 11M13A01-02E
5-6
<5. Wiring>
5.2.2
Connection to the Detector
To connect cables to the detector, proceed as follows:
(1) Mount conduits of the specified thread size or cable glands to the wiring connections of the detector.
The detector may need to be removed in future for maintenance, so be sure to allow
sufficient cable length.
(2) If the ambient temperature at the location of wire installation is 75 to 150°C, be sure to use a
flexible metallic wire conduit. If a non-shielded “ 600 V silicon rubber insulated glass braided
wire “ is used, keep the wire away from noise sources to avoid noise interference.
(3) Figure 5.5 shows the layout of the detector terminals.
TC +
TC -
CELL +
CELL CELL
(+)
(-)
TC
(+)
TC
(-)
CJ
(+)
HTR
1
CELL
2 3
TC
4 5
7
CJ
H T R
CJ
(-)
HTR
6
8
To converter,
To ground
or
terminal box
To converter,
or
terminal box
F05-5E.ai
Figure 5.5 Detector terminals
The sizes of the terminal screw threads are M3.5 except for the M4 on grounding terminal.
Each wire in the cable should be terminated in the corresponding size of crimp terminal (*1)
respectively.
*1: If the surface temperature of the detector terminal box installation site exceeds 60°C, use a “ bare crimp-on terminal”.
(4) Except when “600 V silicon rubber insulated glass braided wire” is used, connect the cable
shield to the FG terminal of the converter.
5.2.3
Connection to the Converter
To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals.
(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal
box. For wiring between the terminal box and the converter, basically use a cable that
withstand temperatures of at least 80°C, rather than individual wires.
NOTE
The above is to prevent moisture or corrosive gas from entering the converter and to ground the
detector without fail.
IM 11M13A01-02E
<5. Wiring>
5.3
Wiring for Power to Detector Heater
5-7
This wiring provides electric power from the converter to the heater for heating the sensor in the
detector.
CAUTION
Refer to section 3.1.
(1) Surface temperature of the detector terminal box: 75 °C or less
Hazardous area
Non-hazardous area
Detector
Converter
HTR 7
HTR 8
HEATER
(2) Surface temperature of the detector terminal box: exceeding 75 °C
Hazardous area
Non-hazardous area
Terminal box
Detector
HTR 7
HTR 8
Converter
HEATER
Heat-resistant wiring
F05-6E.ai
Figure 5.6 Wiring for power to detector heater
5.3.1
Cable Specifications
Basically, cables (2 cores) that withstand temperatures of at least 80°C are used for this wiring.
When the ambient temperature of the detector exceeds 75°C, install a terminal box, and connect
to the detector using six-piece 600 V silicon rubber insulated glass braided wires.
5.3.2
Connection to Detector
When connecting the cable to the detector, proceed as follows:
(1) Mount cable glands or conduits of the specified thread size to the wiring connections of the
detector. The detector may need to be removed in future for maintenance, so be sure to
allow sufficient cable length.
(2) If the ambient temperature at the location of wire installation is 75 to 150°C, be sure to use a
flexible metallic conduit for the wire. If a non-shielded “ 600 V silicon rubber insulated glass
braided wire” is used, keep the wire away from noise sources to avoid noise interference.
(3) The size of the terminal screw threads is M3.5. Each cable should be terminated in the
corresponding size crimp-on terminals contact (*1) respectively.
*1 If the surface temperature of the detector terminal box installation site exceeds 60 °C, use a “ bare crimp-on terminal”.
IM 11M13A01-02E
<5. Wiring>
5-8
CAUTION
• Before opening the detector cover, loosen the lock screw. If the screw is not loosened first,
the screw will damage the cover, and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid gouging the
thread.
• Notice when closing the cover of the detector
After screwing the cover in the detector body, secure it with the lock screw.
Lock screw
Detector cover
F05-7E.ai
Figure 5.7 Detector
5.3.3
Connection to Converter
To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated in
the corresponding size crimp-on terminals.
(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal
box. For wiring between the terminal box and the converter, basically use a cable that
withstand temperatures of at least 80°C, rather than wire.
CAUTION
The above is to prevent moisture or corrosive gas from entering the converter.
Where the ambient environment of the detector and the converter is well-maintained, it is
permissible to connect the wiring from the detector directly to the converter with protection by
conduits.
WARNING
This wiring is to carry power for the heater. Be careful to wire the correct terminals, and be
careful not to ground or short circuit terminals when wiring, as otherwise the instrument may be
damaged.
IM 11M13A01-02E
5.4
5-9
<5. Wiring>
Wiring for Analog Output
This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.
Maintain the load resistance including the wiring resistance at 550 Ω or less.
ZR402G
Receiver 1
Converter
+
–
AO-1(+)
AO-1(-)
AO-2(+)
AO-2(-)
FG
Receiver 2
Shielded cable
+
–
F05-8E.ai
Figure 5.8 Wiring for analog output
5.4.1
Cable Specifications
For this wiring, use a 2-core or a 4-core shielded cable.
5.4.2
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals. Ensure that the cable shield is connected
to the FG terminal of the converter.
(2) Be sure to connect “+” and “-” polarities correctly.
5.5
Power and Grounding Wiring
This wiring supplies power to the converter and grounds the converter/detector.
ZR402G
Converter
Grounding to the ground terminal
on the converter case
ZR22S
Detector
L N G
Converter case
Jumper plate
FG terminal Lock washer
Crimp-on terminal of
the ground wire
100 - 240V AC
50/60Hz
Ground
F05-9E.ai
Figure 5.9 Power and Grounding wiring
5.5.1
Power Wiring
Connect the power wiring to the L and N terminals of the converter. Proceed as follows:
(1) Use a 2-core or a 3-core cable.
(2) The size of converter terminal screw threads is M4. Each wire in the cable should be
terminated corresponding to crimp-on terminals.
IM 11M13A01-02E
5.5.2
<5. Wiring>
5-10
Grounding Wiring
The ground wiring of the detector should be connected to the ground terminal of the detector
case. The ground wiring of the converter should be connected to either the ground terminal of the
converter case or the protective ground terminal in the equipment. The grounding terminals of the
detector are of size of M4 (inside) and M5 (outside), and the grounding terminals of the converter
are of size M4. Proceed as follows:
(1) Keep ground resistance to 100 Ω or less.
(2) When the ambient temperature of the wiring installation is 75 to 150°C for the wiring of the
detector, use wiring material with sufficient heat resistance.
(3) When connecting the ground wiring to the ground terminal of the converter case, be sure
that the lock washer is in contact with the case surface (see Figure 5.9.).
(4) Ensure that the jumper plate is connected between the G terminal and the protective ground
terminal of the converter.
(5) No intermediate parts are used for the internal ground terminal of the detector. Use crimping
terminal for connection to the internal ground terminal in order to avoid corrosion by high
contact potentials.
(6) In order to prevent the earthing conductor from loosening, the conductor must be secured
to the terminal, tightening the screw with torque of approx. 1.2 N•m (for M4) or 2.0 N•m (for
M5).
(7) Care must be taken not to twist the conductor.
5.6
Contact Output Wiring
Contact outputs 1 to 3 can be freely assigned to “low limit alarm”, “high limit alarm”, etc. user
selectable, but the assignment of contact output 4 is fixed (“error output”). And the action (contact
closed on error output) also cannot be changed.
When using these contact outputs, install the wiring as follows:
Converter
Terminal box
Annunciator or the like
DO-1
DO-1
#1 Output
DO-2
DO-2
#2 Output
DO-3
DO-3
#3 Output
DO-4
DO-4
#4 Output
F05-10E.ai
Figure 5.10 Contact output wiring
5.6.1
Cable Specifications
Number of wire in cable varies depending on the number of contact used.
IM 11M13A01-02E
5.6.2
5-11
<5. Wiring>
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals.
(2) The capacities of the contact output relay are 30V DC 3 A, 250V AC 3 A. Connect a load
(e.g. pilot lamp and annunciator) within these limits.
5.7
Wiring for ZR40H Automatic Calibration Unit
This wiring is for operating the solenoid valve for the zero gas and the span gas in the ZR40H
Automatic Calibration Unit, in a system where the calibration gas flow rate is automatically
controlled (e.g. System configuration 2). When installing this wiring, proceed as follows:
Span gas solenoid valve
Wiring inlet
2-G1/2, Pg13.5, M20 or 1/2NPT
Zero gas solenoid valve
ZERO
COMMON
SPAN
COMMON SPAN
ZERO
to ZR402G terminal "AC-COM"
to ZR402G terminal "AC-SPAN"
to ZR402G terminal "AC-ZERO"
Terminal screw : M4
Use a 3-core cable or equivalent
Figure 5.11 Automatic Calibration Unit
5.7.1
F05-11E.ai
Cable Specifications
Use a 3-core cable for the above wiring.
IM 11M13A01-02E
<5. Wiring>
5.7.2
5-12
Wiring Procedure
M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals. M4 screws are used for the terminals of the
solenoid valve as well.
ZR402G
Converter
ZR40H Automatic
Calibration unit
AC-Z
Zero
AC-S
AC-C
Span
F05-12E.ai
Figure 5.12 Wiring for Automatic Calibration Unit
5.8
Contact Input Wiring
The converter can execute specified function when receiving contact signals.
To use these contact signals, proceed wiring as follows:
ZR402G
Converter
Terminal box
DI-1
Contact input 1
DI-2
DI-C
Contact input 2
F05-13E.ai
Figure 5.13 Contact Input Wiring
5.8.1
Cable Specifications
Use 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine which
cable to use.
5.8.2
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals.
(2) The ON/OFF level of this contact input is identified by the resistance. Connect a contact
input that satisfies the specifications in Table 5.2.
Table 5.2 Identification of Contact Input ON/OFF
Resistance
Closed
Open
200 Ω or less
100 kΩ or more
IM 11M13A01-02E
6.
6-1
<6. Components and Their Functions>
Components and Their Functions
In this Chapter, the names and functions of components are described for the major equipment of
the EXAxt ZR Separate type Explosion-proof Zirconia Oxygen Analyzer.
6.1
ZR22S Detector
6.1.1
General-purpose Explosion-proof Detector (except for
ZR22S--015)
Probe
This part is inserted in the furnace.
Select length from 0.15, 0.4,
0.7, 1.0, 1.5 or 2.0 m.
Contact
Flange
Used to mount the detector.
Select from JIS, ANSI or DIN
standard models.
Flame arrestor assembly
mounting screw
Calibration gas
pipe opening
Metal O-ring
Pipe support
U-shaped pipe
Bolt
Probe
Screw
Sensor (cell)
Filter
Washer
Flame arrestor assembly
F06-1E.ai
Figure 6.1 General-purpose Explosion-proof Detector (Standard type)
IM 11M13A01-02E
6.1.2
6-2
<6. Components and Their Functions>
High Temperature Detector (ZR22S--015)
Sample gas outlet
When a sample gas pressure is
negative, connect the auxiliary
ejector assembly.
When the sample gas is high
temperature and high pressure,
and does not fall below 700°C,
connect a pressure control valve
(e.g. a needle valve).
Separate type
High Temperature Detector (ZR22S--015)
When the temperature of the sample gas is
between 700°C and 1400°C, mount this
detector with a ZO21P probe adapter.
Flange
Selectable from JIS,
ANSI or DIN standards
High Temperature Probe Adapter
(ZO21P)
The probe is made of either SUS310S
or silicon carbide (SiC).
Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0
or 1.5 m.
When using an SiC probe, mount it
vertically downward.
F06-2E.ai
Figure 6.2 High Temperature Detector
IM 11M13A01-02E
6.2
6-3
<6. Components and Their Functions>
ZR402G Converter
Complete Operation Display
■ Interactive operations along with operation display.
■ A variety of display modes - enabling you to select
the operation mode freely.
■ Back-lit LCD allows viewing even in the darkness.
■ Error codes and details of errors can be checked in
the field without the need to refer to appropriate
instruction manual.
■ Password for security
Touch Screen Display
● Example of basic display
This data display provides for interactive operation.
● Example of trend display
— displays data changes
Self-testing suggests countermeasures for
problems
If a problem occurs, the liquid-crystal display will
provide an error code and description of the
problem. This enables prompt and appropriate
corrective action to be taken.
During calibration, you can check the stabilized,
display data while viewing oxygen trend data,thus
providing highly reliable calibration.
Error code
Reason for error
● Example of setting data display
Error 1
Cell failure
Error 2
Abnormal heater temperature
Error 3
Defective A/D converter
Error 4
Faulty EEPROM
Alarm 1
Abnormal oxygen concentration
mA-output setup
Alarm 2
Abnormal moisture
Alarms setup
Alarm 3
Abnormal mixing ratio
Contact setup
Alarm 6
Abnormal zero calibration ratio
Others
Alarm 7
Abnormal span calibration ratio
Alarm 8
Stabilization time over
Commissioning
Basic setup
Enter
● One-touch interactive display operation
● User-friendly design providing easy operation
Disp3_1.ai
without having to use the instruction manual
F06-3E.ai
Figure 6.3 Converter
IM 11M13A01-02E
6.3
6-4
<6. Components and Their Functions>
Touchpanel Switch Operations
6.3.1
Basic Panel and Switch
The converter uses a touchpanel switch which can be operated by just touching the panel
display. Figure 6.4 shows the Basic panel display. The switches that appear in the switch display
area vary depending on the panel display, allowing all switch operations. Table 6.1 shows the
switch functions.
Tag name display area
Tag:
Primary value
Secondary value
Tertiary value
21.0
% O2
17.43mA -Output1
17.43mA -Output2
Switch display area
Alarm and error
display area
F06-4E.ai
Figure 6.4 Basic Panel Display
Tag name display area: Displays the set tag name (Refer to Subsection 10.3.2, Entering Tag
Name).
Primary to tertiary value: Displays the selected item. (Refer to Section 7.9, Setting Display Item.)
Switch display area: Displays switches and functions selected according to the panel display.
Alarm and error display area: Displays an error if an alarm or error occurs. If you touch this area,
the details of the error or alarm are then displayed.
Table 6.1 Switches and Their Functions
Home key:
Returns to the Execution/Setup display.
Enter
Enter key:
Enters the input value and sets up the
selected item.
Reject key:
Moves back to the previous display.
Setup key:
Used to enter the Execution/Setup display.
Cursor key:
Moves the cursor down.
Detailed-data key:
Displays the analog input value.
Graph display key:
Displays a trend graph.
Cursor:
Points the cursor at the currently
selected item.
Alarm:
Displayed if an alarm arises.
Error:
Displayed if an error occurs.
T06-1E.ai
IM 11M13A01-02E
6.3.2
6-5
<6. Components and Their Functions>
Display Configuration (for Oxygen Analyzer)
Figure 6.5 shows the configuration. A password the displays positioned below enables display
“Execution/Setup” to be protected. If a password has not been set, press the [Enter] key to
proceed to the next panel display. The [Home] key enables you to return to Execution/Setup from
any panel display..
[Basic panel display]
[Detailed-data display]
[Execution/Setup display]
[Trend Graph display]
[Password Entry display]
Calibration execution
Returns to the
Execution/Setup
display.
Blow back execution
Returns to the
Execution/Setup
display.
Reset panel display
Returns to the
Execution/Setup
display.
Maintenance panel
display
Display setup
Display item
Trend graph
Auto return time
Language
Calibration setup
Mode
Points
Zero gas conc.
Span gas conc.
Timing
Blow back setup
mA-output loop check
Contact check
Returns to the
Execution/Setup
display.
Commissioning
display
Basic setup
mA-output setup
mA-output1
mA-output2
Set presets
Alarms setup
Hysteresis
Contact delay
Setpoints
Contact setup
Output contact 1
Output contact 2
Output contact 3
Input contacts
Others
Clock
Averaging
Fuel setup
Purging
Unit setup
Passwords
Defaults
Returns to the
Execution/Setup
display.
F06-5E.ai
Figure 6.5 Display Configuration
IM 11M13A01-02E
6-6
<6. Components and Their Functions>
6.3.3
Display Functions
Individual panel displays in the display configuration provide the following functions:
(1) Basic panel display: Displays the values measured in three selected items. (see Section
7.9, “Setting Display Item”).
(2) Execution/Setup display: Selects the calibration, maintenance and setup items.
(3) Detailed-data display: This allows you to view such detailed data as the cell electromotive
force and cell temperature. (see Section 10.1, “Detailed-data
Display”, later in this manual).
(4) Trend Graph display: Displays a trend graph. (see Section 10.2, “Trend Graph”, later in this
manual).
(5) Calibration execution: Makes zero and span calibrations. (see Chapter 9, “Calibration”, and
the associated sections later in this manual).
(6) Blow back execution: Executes a “Blow back setup”. (see Section 10.4, “Blow Back”, later in
this manual).
(7) Reset panel display: If an error arises, you can restart the equipment from this display. (for
more details, see Section 10.6, “Reset”, later in this manual).
(8) Maintenance panel display: Sets the data for equipment maintenance or makes a loop
check.
(9) Commissioning display: Sets up the operation data. (For details, see Chapter 8, “Detailed
Data Setting”, and the associated sections later in this manual.)
6.3.4
Entering Numeric and Text Data
This section sets out how to enter numeric and text data. If only numeric values are entered,
a numeric-data entry display as in Figure 6.6 then appears. Press the numeral keys to enter
numeric values. If those values include a decimal point as in Figure 6.6, the decimal point need
not be entered because the decimal point position is already fixed, so just enter 00098.
New value:
0
000.98 % O2
7
8
9
4
5
6
1
2
3
Enter
F06-6E.ai
Figure 6.6 Numeric-data Entry Display
To enter a password (in combination with text data, numeric values and codes), the alphabetic
character entry panel display first appears. If you press any numeral key [0 - 9], the current
display then changes to the numeric-value entry panel display, enabling you to enter numeric
values. If you press the [other] key, the current display then changes to the code-entry display,
enabling you to enter codes. These displays alternate between the three. Figure 6.7 shows the
relationship between these three displays. Three alphabetic characters and three codes are
assigned for each individual switch. If the alphabetic character key is pressed and held, three
characters appear in turn. Move the cursor to the desired character and release the key to enter
it. If an incorrect character is entered, move the cursor to re-enter the characters. The following
shows an example of entering “abc%123.”
IM 11M13A01-02E
<6. Components and Their Functions>
Operation
Press the [ABC] key once.
Press and hold the [ABC] key.
6-7
Display
A_
AA
B
C
Release the [ABC] key when the character B
appears in the cursor position.
Enter the character C in the same manner
as above.
Press the [other] key.
AB_
Press and hold the [$%&] key and enter “%.”
Then press the [0-9] key.
ABC%_
Enter the numeric characters 1, 2 and 3 in turn.
Press the [Enter] key to complete the entry.
ABC%123_
ABC_
Siki06E.ai
Enter your password
0-9
A
B C
D
E F
G
H I
other
J
K L
M
N O
P
Q R
Space
- $
S
T U
V
W X
Y
Z @
Enter your password
Enter
Enter your password
A-Z
7
8
9
A-Z
!
0 #
$
% &
'
( )
other
4
5
6
0-9
*
+ ,
–
. /
:
; <
0
1
2
3
>
? @
[
\ ]
^
_ `
Enter
[
{
}
Enter
F06-7E.ai
Figure 6.7 Text Entry Display
IM 11M13A01-02E
6.4
6-8
<6. Components and Their Functions>
ZA8F Flow Setting Unit, ZR40H Automatic
Calibration Unit
Reference gas flow
setting valve
Span gas flow
setting valve
Zero gas flow
setting valve
Flowmeter for
reference gas
Flowmeter for
calibration gas
F06-8E.ai
Figure 6.8 ZA8F Flow Setting Unit
Flowmeter for
calibration gas
Flowmeter for
reference gas
Reference gas flow
setting valve
Terminal Box
Made in Japan
REF.OUT CAL.OUT SPAN IN ZERO IN
Calibration gas flow
setting valve
F06-9E.ai
Figure 6.9 ZR40H Automatic Calibration Unit
IM 11M13A01-02E
7-1
<7. Startup>
7. Startup
The following describes the minimum operating requirements — from supplying power to the
converter to analog output confirmation to manual calibration.
System tuning by the HART communicator, refer to IM11M12A01-51E “ HART Communication
Protocol ”.
7.1
Startup Procedure
The startup procedure is as follows:
CAUTION
If you connect Model ZO21D* detectors, then you need to change the detector parameters.
Before connecting power, refer to: Section 7.6 Confirmation of Detector Type Setting.
* The ZO21D is not a explosion-proof detector.
Check settings
Check piping & wiring
Check valve type setting
If connecting ZO21D
Briefly power on
Set detector to ZO21D
(front panel)
If detector is ZR22
Power OFF
Permanent power wiring
Start set parameters
Warmup, then calibrate
F07-1E.ai
Figure 7.1 Startup Procedure
7.2
Checking Piping and Wiring Connections
Check that the piping and wiring connections have been properly completed in accordance with
Chapter 4, “Piping,” and Chapter 5, “Wiring.”
IM 11M13A01-02E
7-2
<7. Startup>
7.3
Checking Valve Setup
Set up valves and associated components used in the analyzer system as follows:
(1) If a stop valve is used in the detector’s calibration gas inlet, fully close this valve.
(2) If instrument air is used as the reference gas, adjust the air-set secondary pressure so that
an air pressure equals sample gas pressure plus approx. 50 kPa (or sample gas pressure
plus approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa) is
obtained. Turn the reference gas flow setting valve in the flow setting unit to obtain a flow of
800 to 1000 mL/min. (Turning the valve shaft counterclockwise increases the rate of flow.
Before turning the valve shaft, if the valve has a lock nut, first loosen the lock nut.) After
completing the valve setup, be sure to tighten the lock nut.
NOTE
The calibration gas flow setting is described later. Fully close the needle valve in the flow setting
unit.
7.4
Supplying Power to the Converter
CAUTION
To avoid temperature changes around the sensor, it is recommended that (rather than turning it
on and off) power be continuously supplied to the Oxygen Analyzer if it is used in an application
where it is used periodically.
It is also recommended to flow a span gas (instrument air) beforehand.
Supply power to the converter. A display as in Figure 7.2, which indicates the detector’s sensor
temperature, then appears. As the heat in the sensor increases, the temperature gradually rises
to 750°C. This takes about 20 minutes after the power is turned on, depending somewhat on the
ambient temperature and the sample gas temperature.
After the sensor temperature has stabilized at 750°C, the converter is in measurement mode.
The display panel then displays the oxygen concentration as in Figure 7.3. This is called the
Basic panel display.
Tag:
Tag:
319
21.0
% O2
°C
4.00mA -Output1
4.00mA -Output2
Hold
Warmup
17.43mA -Output1
17.43mA -Output2
F07-2E.ai
Figure 7.2 Display During Warm-up
F07-3E.ai
Figure 7.3 Measurement Mode Display
IM 11M13A01-02E
7.5
7-3
<7. Startup>
Confirmation of Converter Type Setting
This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
Note that if the converter type setting is changed, the operating data that have been set are then
initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display
(Figure 7.4).
(2) Use the [▼] key to select “Commissioning” and press the [Enter] key.
(3) In the Passwords display, enter the [Enter] key. If the password is to be set again, enter the
new password (for details, see Subsection 8.6.5, “Setting Passwords” later in this manual).
(4) The Commissioning display shown in Figure 7.5 appears. Select “Basic setup” and press
the [Enter] key.
(5) The Basic setup display shown in Figure 7.6 then appears. Confirm the currently set
converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the
converter was set for high temperature humidity use before shipment.
(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.7
then appears.
(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.
(8) If the type of converter is changed after setting the operating data, those data are then
initialized and the default settings remain. Reset the operating data to meet the new type of
equipment.
Execution/Setup
Execution
Calibration
Blow back
Reset
Setup
Maintenance
Commissioning
Commissioning
Basic setup
mA-output setup
Alarms setup
Contact setup
Others
Enter
F07-4E.ai
Figure 7.4 “Execution/Setup” Display
Enter
F07-5E.ai
Figure 7.5 “Commissioning” Display
Basic setup
Basic setup
This instrument is for
Oxygen
Please note:changing this
Basic setting will return ALL
other setting to Default!!
This instrument is for
Oxygen
Humidity
Please note:changing
this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Detector is ZR22
Select measure gas: Wet
F07-6E.ai
Figure 7.6 “Basic Setup” Display
Select measure gas: Wet
Enter
Enter
F07-7E.ai
Figure 7.7 Equipment Setup
IM 11M13A01-02E
7.6
7-4
<7. Startup>
Confirmation of Detector Type Setting
Check that the detector in Figure 7.6 is the one for this equipment.
WARNING
● If this converter is to be used in conjunction with the ZO21D*, the power requirements are
limited to 115 V AC or less, 50 or 60 Hz (it cannot be used with a 125 V or greater, or in the
EEC).
● If detector settings are to be changed, first disconnect the wiring connections between the
detector and the converter. Then change detector settings appropriately.
* The ZO21D is not a explosion-proof detector.
7.7
7.8
Selection of Sample Gas
Combustion gases contain moisture created by burning hydrogen in the fuel. If this moisture is
removed, the oxygen concentration might be higher than before. You can select whether the
oxygen concentration in a wet gas is to be measured directly, or compensated for its dry-gas
value before use. Select the “Select measure gas: Wet” in Figure 7.6 to select either wet or dry
gas.
Output Range Setting
This section sets forth analog output range settings. For details, consult Section 8.1, “Current
Output Setting”, later in this manual.
Minimum Current (4 mA) and Maximum Current (20 mA) Settings
To set the minimum and maximum current settings, follow these steps:
(1) Select the “Commissioning” in the Execution/Setup display.
(2) Select the “mA-output setup” in the Commissioning display; the display shown in Figure 7.8
then appears.
(3) Select the “mA-output 1” in the mA-outputs display. The “mA-output 1 range” display shown
in Figure 7.9 then appears.
(4) In the display shown in Figure 7.9, select “Min. oxygen conc” and press the [Enter] key to
display the numeric-data entry display. Enter the oxygen concentration at a 4 mA output;
enter [010] for a ten-percent oxygen concentration measurement.
(5) Also in Figure 7.9, select “Max. oxygen conc” at a 20 mA output. Enter the appropriate
maximum oxygen concentration (at the 20 mA output) in the same manner as in step 4
above.
(6) Set the “mA-output 2” in the same manner as in the appropriate steps above.
mA-outputs
mA-output1 range
Both outputs are 4-20mA
mA-output1
mA-output2
Set presets
Enter
F07-8E.ai
Figure 7.8 Setting “mA-outputs”
Parameter: Oxygen
Set range
Min. oxygen conc:
0
%O2
Max. oxygen conc:
25
%O2
Output damping:
0s
Output mode: Linear
Enter
F07-9E.ai
Figure 7.9 Setting “mA-output1 range”
IM 11M13A01-02E
7.9
7-5
<7. Startup>
Setting Display Item
This section briefly describes the display item settings shown in Figure 7.9, “Basic Panel Display.”
Tag:
Primary value
Secondary value
Tertiary value
21.0
%O2
17.43mA -Output1
17.43mA -Output2
F07-10E.ai
Figure 7.10 Basic Panel Display
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select the “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” in the Maintenance panel display (Figure 7.11).
The Display setup display (Figure 7.12) then appears.
(3) In the above Display setup display, select the “Display item”. The Display item display
(Figure 7.13) then appears. From this display, select the “Primary value” and press the
[Enter] key to display the “Display item” selection display (Figure 7.14).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.1, Display Items, enabling the selection of display items in individual display
areas.
Maintenance
Display setup
Display setup
Calibration setup
Blow back setup
mA-output loop check
Contact check
Display item
Trend graph
Auto return time:
0min
Language: English
Enter
F07-11E.ai
Figure 7.11 “Maintenance” Display
Enter
F07-12E.ai
Figure 7.12 Display Setup
Display item
Display item
Primary value:
Oxygen
Secondary value:
mA-output1
Tertiary value:
mA-output2
Tag name:
Enter
F07-13E.ai
Figure 7.13 “Display Item” Display
Primary value:
Oxygen
Oxygen
Item of output1 damping
Secondary
value:
Item of output2
damping
mA-output1
Tertiary value:
mA-output2
Tag name:
Enter
F07-14E.ai
Figure 7.14 Display Item Selection
IM 11M13A01-02E
7-6
<7. Startup>
Table 7.1 Display Items
Item
Primary value
Secondary and
tertiary values
Oxygen concentration
Oxygen concentration during measurement
Air ratio
Current computed air ratio
Moisture quantity
Moisture quantity (%H2O) in the exhaust gas
Display
Output 1 item
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Output 2 item
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Current output 1
Output current value from analog output 1
Current output 2
Output current value from analog output 2
*1: If an analog output damping constant is set, the oxygen concentration display then includes these settings.
About the Air ratio:
“Air ratio” is defined as the ratio of (the amount of air theoretically required to completely burn all
the fuel) to (the amount of air actually supplied).
For this equipment, the air ratio will be obtained in a simplified way by measuring the oxygen
concentration in the exhaust gas. The air ratio may be expressed mathematically by:
m = { 1 / ( 21- Oxygen concentration ) } 21
If you use the air ratio data for estimating the combustion efficiency, etc., check that no air is
leaking in beforehand and that the measured value has not been affected by any interference
gas (CH4, CO, H2, etc.).
About moisture quantity:
The moisture quantity in the exhaust gas is calculated based on the parameters of the fuel setting
(refer to Subsection 8.6.3, “Setting Fuels”, later in this manual). The moisture content may be
expressed mathematically by:
Moisture quantity = {(water vapor content per fuel unit quantity) + (water content in air)}/ total
amount of exhaust gas
= ( Gw + 1.61 Z Ao m ) / ( X + Ao m )
where,
Gw = Water vapor content in exhaust gas, m3/kg (m3)
Z
= Ambient absolute humidity, kg/kg
Ao = Ideal air amount, m3/kg (m3)
m
= Air ratio
X
= Fuel coefficient, Nm3/kg or m3/m3
For details on each parameter, refer to Subsection 8.6.3, “Setting Fuels”.
7.10 Checking Current Loop
The set current can be output as an analog output.
(1) Press the [Setup] key on the Basic panel display to display the Execution/Setup display.
Then select the “Maintenance” in the Execution/Setup display.
(2) Select the “mA-output loop check” in the Maintenance panel display to display the “mA output loop check” display, enabling you to check the “mA-output 1” and the “mA-output 2”.
Select the desired output terminal for current-loop checking (see Figure 7.15).
IM 11M13A01-02E
7-7
<7. Startup>
(3) At the time of entering the numeric-data entry display, the output current will change to 4 mA
(default value). If the desired current is entered, the corresponding output will be provided.
mA - output loop check
mA - output 1
mA - output 2
Enter
F07-15E.ai
Figure 7.15 “mA - output loop check” Panel Display
7.11 Checking Contact I/O
Conduct the contact input and output checking as well as operational checking of the solenoid
valves for automatic calibration.
7.11.1
Checking Contact Outputs
To check the contact output, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” in that display.
(2) Select “Contact check” then “Output contacts” in the Maintenance panel display to display
the Output contacts display (see Figure 7.17).
(3) In this display, select the desired “output contact” for checking. The display, which enables the
closing and opening of contacts, then appears. Use the display to conduct continuity checking.
Contact check
Output contacts
Output contacts
Output contact 1 : Open
Calibration contacts
Output contact 2 : Open
Input contacts
Output contact 3 : Open
Output contact 4 : Open
Enter
F07-16E.ai
Figure 7.16 “Contact Check” Display
Enter
F07-17E.ai
Figure 7.17 “Output contacts” Check Panel Display
CAUTION
• If you conduct an open-close check for “Output contact 4”, Error 1 or Error 2 will occur.
This is because the built-in heater power of the detector, which is connected to “Output contact
4”, is turned off during the above check. So, if the above error occurs, reset the equipment
or turn the power off and then back on to restart (refer to Section 10.6, “Reset”, later in this
manual).
IM 11M13A01-02E
7-8
<7. Startup>
7.11.2
Checking Calibration Contact Outputs
The calibration contacts are used for solenoid valve drive signals for the ZR40H Automatic
Calibration Unit. When using the ZR40H Automatic Calibration Unit, use the calibration contact
output to check that the wiring connections have been properly completed and check equipment
operation.
(1) Referring to Subsection 7.11.1, display the contact check display.
(2) Select the “Calibration contacts” to display the panel display as Figure 7.18 shows.
(3) Open the “Zero gas contact” and the “Span gas contact”. This will help check the automatic
calibration unit and wiring connections.
Calibration contacts
Zero gas contact : Open
Span gas contact : Open
Enter
F07-18E.ai
Figure 7.18 “Calibration contacts” Check Panel Display
NOTE
“Open” and “Closed” displayed on the Calibration contacts display indicate actions of drive
contacts and are opposite to the valve open and close actions. If “Open” is displayed on the
Calibration contacts display, no calibration gas flows. If “Closed” is displayed on that display,
calibration gas flows.
7.11.3
Checking Contact Inputs
(1) Referring to Subsection 7.11.1, display the contact check display.
(2) Display the “Input contacts” check display as Figure 7.19 shows. The “Open” or “Closed”
Input contact in the display shows the current contact input terminal status and the display
changes according to the contact status. Using this enables you to check that the wiring
connections have been properly completed.
Input contacts
Input contact 1 : Open
Input contact 2 : Open
Enter
F07-19E.ai
Figure 7.19 “Input contacts” Check Panel Display
IM 11M13A01-02E
7-9
<7. Startup>
7.12 Calibration
To calibrate this instrument, the procedure is to measure zero gas and span gas and set the
instrument to read the known concentrations. The procedure for both zero and span calibration,
or for either zero or span calibration, can be performed manually from the touch display, or can
be performed semi-automatically using contact signal inputs to start calibration, (allowing preset
calibration and stabilization times), or it can be performed automatically at preset intervals.
There are three types of calibration procedures available:
(1) Manual calibration conducting zero and span calibrations, or either of these calibrations in
turn.
(2) Semi-automatic calibration which uses the touchpanel or a contact input signal and
conducts calibration operations based on a preset calibration and stabilization times.
(3) Automatic calibration conducted at preset intervals.
Manual calibration needs the ZA8F Flow Setting Unit to allow manual supply of the calibration
gases. Semi-automatic and automatic calibrations need the ZR40H Automatic Calibration Unit
to allow automatic supply of the calibration gases. The following sections set forth the manual
calibration procedures. For details on semi-automatic and automatic calibrations, consult
Chapter 9, Calibration, later in this manual.
7.12.1
Calibration Setup
Mode Setting
For the mode setting, do the following:
Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” in the Execution/Setup display to display the Maintenance panel display.
Then select “Calibration setup” to display the Calibration setup display as Figure 7.20 shows.
Select Mode in this panel, and then select “Manual”, “Semi-Auto” or “Auto”.
Calibration setup
Mode: Manual
Points: Both
Zero gas conc:
1.00%
Span gas conc: 21.00%
Timing
Enter
F07-20E.ai
Figure 7.20 “Calibration Setup” Display
Calibration Setting Procedures
Select “Points” (calibration procedure) in the Calibration setup display to display the “Both (Span
– Zero”, “Span”, “Zero” selection display. In this display, select “Both (Span – Zero)”.
Calibration Gas Concentration Setting
(1) Zero gas concentration
If zero-gas concentration is selected, the Numeric-data Entry display then appears.
Use this display to enter an oxygen concentration value for the zero-gas calibration;
if the oxygen concentration is 0.98 vol%O2, enter 00098.
IM 11M13A01-02E
7-10
<7. Startup>
(2) Span gas concentration
With “Span gas conc” selected in the Calibration setup display, display the Numeric - data
Entry display and enter an oxygen concentration value for the span gas calibration; If
instrument air is used, enter 02100 for a 21 vol%O2 value.
When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas),
use a portable oxygen analyzer to measure the actual oxygen concentration, and then enter
it.
CAUTION
If instrument air is used for the span gas, dehumidify the air by cooling to a dew point of - 20°C
and remove any oil mist or dust.
7.12.2
Manual Calibration
Preliminary
Before performing manual calibration, be sure that the ZA8F Flow Setting Unit zero gas flow
setting valve is fully closed. Open the zero gas cylinder pressure reducing valve so that the
secondary pressure equals sample gas plus approx. 50 kPa (or sample gas pressure plus
approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa).
This applies even if you are using the ZR40H Automatic Calibration Unit.
Calibration Procedures
This manual assumes that the instrument air is the same as the reference gas used for the span
gas. Follow the steps below to conduct manual calibration:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Calibration” in the Execution/Setup display. In doing so, the Calibration display
as in Figure 7.21 appears.
Calibration
Manual calibration
Check the span gas
concentration value
Manual calibration
Next
Cancel calibration
Change value:
21.00%
Enter
F07-21E.ai
Figure 7.21 “Calibration” Display
Enter
F07-22E.ai
Figure 7.22 “Manual Calibration” Display
(2) Press the [Enter] key to select “Span calibration”. The Manual calibration display shown
in Figure 7.22 then appears. Check that the oxygen concentration for the span gas in this
display coincides with the oxygen concentration in the calibration gas actually used. If the
check results are assumed to be OK, select “Next” in the Manual calibration display.
IM 11M13A01-02E
7-11
<7. Startup>
(3) Follow the display message in Figure 7.23 to turn on span gas flow. Open the span gas flow
valve for the Flow Setting Unit by loosening the valve lock nut and slowly turning the valve
shaft counterclockwise to flow the span gas at 600 ± 60 ml/min. Use the calibration gas
flowmeter to check the flow.
Manual calibration
Manual calibration
Open span gas valve.
21.00%
Set flow span gas to
600ml/min.
0.5min./div
Valve opened
Cancel calibration
1.00%
21.0 %O2
Enter
F07-23E.ai
CAL. TIME
Enter
F07-24E.ai
Figure 7.23 Span Gas Flow Display
Figure 7.24 Span Gas Trend Graph
(for Manual Calibration)
(4) If “Valve opened” is selected as in Figure 7.23 , an oxygen concentration trend graph (with
the oxygen concentration being measured) appears (see Figure 7.24). The CAL.TIME in
the bottom area of the panel flashes. Observe the trend graph and wait until the measured
value stabilizes in the vicinity of 21% on the graph. At this point, calibration has not yet been
executed yet, so even if the measured value is above or below 21%, no problem occurs.
(5) After the measured value has stabilized, press the [Enter] key to display the “Span
calibration complete” display shown in Figure 7.25. At this point, the measured value is
corrected to equal the span gas concentration setting. Close the span gas flow valve. The
valve lock nut should be tightened completely so that the span gas does not leak.
Manual calibration
Manual calibration
Span calibration
Close the span gas valve.
Check the zero gas
concentration value
Next
Cancel calibration
Change value:
1.00%
Zero calibration
End
Enter
F07-25E.ai
Enter
Figure 7.25 Span Calibration Complete
Figure 7.26 Zero Gas Concentration
Zero Calibration Start Display
Check Display
(in Manual Calibration)
F07-26E.ai
(6) Select “Zero calibration” as in Figure 7.25 to display the zero gas concentration check
display (Manual calibration) . Check that the zero gas oxygen concentration value and the
calibration gas oxygen concentration value agree. Then select “Next” as in Figure 7.26.
IM 11M13A01-02E
7-12
<7. Startup>
(7) Follow the instructions in the display as in Figure 7.27 to turn on the zero gas flow. To do
this, open the zero gas flow valve for the Flow Setting Unit and adjust that valve to obtain a
flow of 600 ± 60 ml/min.
(The valve should be adjusted by loosening its lock nut and slowly turning the valve shaft
counterclockwise. Use the calibration gas flowmeter to check the flow.)
Manual calibration
Manual calibration
Open zero gas valve.
21.00%
Set flow zero gas to
600ml/min.
0.5min./div
Valve opened
Cancel calibration
1.00%
Enter
Figure 7.27 Zero Gas Flow Display
(for Manual calibration)
F07-27E.ai
21.0%O2
CAL. TIME
Enter
F07-28E.ai
Figure 7.28 Manual Calibration, Trend Graph
(8) If “Valve opened” is selected as in Figure 7.27 , an oxygen concentration trend graph (with
the oxygen concentration being measured) appears (see Figure 7.28). The CAL.TIME in
the bottom area of the panel flashes. Observe the trend graph and wait until the measured
value stabilizes in the vicinity of the zero gas concentration on the graph. At this point, no
calibration has been executed yet, so even if the measured value is above or below the zero
gas concentration value, no problem occurs.
(9) After the measured value has stabilized, press the [Enter] key to display the “Zero
calibration complete” display shown in Figure 7.29. At this point, the measured value is
corrected to equal the zero gas concentration setting. Close the zero gas flow valve. The
valve lock nut should be tightened completely so that the zero gas does not leak.
Manual calibration
Zero calibration
Close the zero gas valve.
Span calibration
End
Enter
F07-29E.ai
Figure 7.29 Zero Calibration Complete Display
(10) Select “End” in the display as shown in Figure 7.29. An oxygen concentration trend graph
(with the oxygen concentration being measured) appears and HOLD TIME then flashes.
This time is referred to as the output-stabilization time. If the HOLD TIME has been set
with the output hold setting, the analog output remains held (refer to Section 8.2, Output
Hold Setting, later in this manual). Manual calibration is completed when the preset hold
(output stabilization) time elapses. This hold (output stabilization) time is set to 10 minutes
at the factory before shipment. If you press the [Enter] or [Reject] key within the hold (output
stabilization) time, manual calibration is then completed.
IM 11M13A01-02E
8-1
<8. Detailed Data Setting>
8.
Detailed Data Setting
8.1
Current Output Setting
This section describes setting of the analog output range.
8.1.1
Setting Minimum Current (4 mA) and Maximum
Current (20 mA)
To set the minimum and maximum currents, proceed as follows:
(1) Select the “Commissioning” in the Execution/Setup display.
(2) Select the “mA-output setup” in the Commissioning display.
(3) Select the “mA-output1” in the mA-outputs display.
(4) Select the “Min. oxygen conc” in the mA-output1 range display and press the [Enter] key.
The numeric-data entry display then appears. Enter the oxygen concentration for the
minimum current (4 mA); for example, enter “010” for 10 vol%O2.
(5) Select the “Max. oxygen conc” in the mA-output1 range display and enter the oxygen
concentration for the maximum current (20 mA) in the same manner as in step 4 above.
(6) Set the “mA-output2” in the same way as the setting procedure for mA-output1 given above.
8.1.2
Input Ranges
The range low and high values are restricted as follows:
The range min. O2 concentration value (corresponding to 4 mA output) can be set to either 0
vol%O2 or in the range of 6 to 76 vol%O2.
The range max. O2 concentration value (corresponding to 20 mA output) can be set to any value
in the range of 5 to 100 vol%O2, however the range max. setting must be at least 1.3 times the
range min. setting.
If you do not observe this restriction, the measurement will be invalid, and any previous valid
value will be used. The gray area in figure represents the valid setting range.
Setting example 1
If the range minimum (corresponding to 4 mA output) is set to 10 vol%O2 then range maximum
(corresponding to 20 mA output) must be at least 13 vol%O2.
IM 11M13A01-02E
8-2
<8. Detailed Data Setting>
Setting example 2
Maximum oxygen concentration, vol%O2 (for a maximum current of 20 mA)
If the range minimum (corresponding to 4 mA output) is set to 75 vol%O2 then range maximum
(corresponding to 20 mA output) must be at least 75x1.3=98 vol%O2 (rounding decimal part up).
95
85
75
Ranges over which oxygen
concentrations can be set
65
55
45
Outside ranges
35
25
15
5
5
15
25
35
45
55
Minimum oxygen concentration, vol%O2 (for a minimum current of 4 mA)
Minimum-Maximum setting range of oxygen concentration
8.1.3
65
75
F08E.ai
Setting Output Smoothing Coefficient
When the oxygen concentration in the sample gas fluctuates rapidly, if the measured value is
used for control this can lead to problems with undesirable frequent ON/OFF switching. You can
set a smoothing time constant of between 0 and 255 seconds to reduce the effect. Select the
appropriate output damping constant from the numeric-data entry display. To set 30 seconds,
enter 030.
8.1.4
Selection of Output Mode
You can select whether the relationship between the sample oxygen concentration and the
analog output signal be linear or logarithmic. Press the [Enter] key in the output mode display. A
linear/ logarithmic selection display then appears. Select the desired mode.
NOTE
If you select an output mode of “logarithmic” then regardless of range setting the minimum output
value becomes 0.1 vol%O2 fixed.
8.1.5
Default Values
When the analyzer is delivered or reset to defaults, the output current default settings by as
shown in Table 8.1.
Table 8.1 Output Current Default Values
Item
Default setting
Min. oxygen concentration
0 vol%O2
Max. oxygen concentration
25 vol%O2
Output damping constant
0 (seconds)
Output mode
Linear
IM 11M13A01-02E
8.2
8-3
<8. Detailed Data Setting>
Output Hold Setting
The “output hold” functions hold an analog output signal at a preset value during the equipment’s
warm-up time or calibration or if an error arises. Outputs 1 and 2 can not be set individually. Table
8.2 shows the analog outputs that can be retained and the individual states.
Table 8.2
Equipment status
Output
During warm-up
hold values available
4 mA
20 mA
Without hold feature
Retains output from just before
occurrence
Set value (2.4 to 21.6 mA)
○: The output hold functions are available.
Under
maintenance
Under calibration
During blow back
On Error
occurrence(*1)
○
○
○
○
○
○
○
○
○
○
○
○
*1: The output hold functions on error occurrence are unavailable when option code “/C2” or “/C3”(NAMER NE 43 compliant) is
specified.
8.2.1
Definition of Equipment Status
(1) During warm-up
“During warm-up” is the time required after applying power until sensor temperature stabilizes at
750 ºC and the instrument is in the measurement mode.
(2) Under Maintenance
“Under maintenance” is the time that starts when you move to the Execution/Setup display by
touching the [Setup] key on the Basic panel display and ends when you return to the Basic panel
display. It includes when you operates keys on lower level menu displays of the Execution/Setup
display.
(3) Under Calibration (see Chapter 9, Calibration, later in this manual)
For manual Calibration, “Under calibration” is the time that starts when you move to the Manual
calibration display (Figure 7.22) from the Calibration display, lasts while you are operating keys
for performing calibration manually, and ends when you press the End key and after a preset hold
time has elapsed.
For semi-automatic calibration, “Under calibration” is the time required from entering calibration
instructions to perform a, either by using the touchpanel or by a contact input, calibration until the
hold (output stabilization) time elapses.
For automatic calibration, “Under calibration” is the time required, after performing an appropriate
calibration until the hold (output stabilization) time elapses.
(4) During Blow back (see Section 10.4, Blow Back, later in this manual)
During semi-automatic blow back:
“During semi-automatic blow back” is the time required after pressing the [Blow back start] key,
by using the touchpanel or entering a blow back start instruction by using a contact input, until the
blow back time and hold (output stabilization) time elapse.
During automatic blow back:
“During automatic blow back” is the time required after reaching the blow back start time until the
blow back time and hold (output stabilization) time elapse.
(5) On Error occurrence
This is the time at which any of Errors 1 to 4 occurs.
IM 11M13A01-02E
8-4
<8. Detailed Data Setting>
8.2.2
Preference Order of Output Hold Value
The output hold value takes the following preference order:
Preference order (high)
On error occurrence
Under calibration or during blow back
Under maintenance
During warm-up
Siki08-1E.ai
For example, if the output current is set to 4 mA during maintenance, and no output-hold output
for during calibration is preset, the output is held at 4 mA during the maintenance display.
However, the output hold is released at the time of starting the calibration, and the output will be
again held at 4 mA after completing the calibration and when the hold (output stabilization) time
elapses.
8.2.3
Output Hold Procedure
To set the output hold, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Commissioning” in the Execution/Setup display. Next, select the “mA-output
setup” and “Set presets”, then the “mA-outputs presets” display as shown in Figure 8.1.
mA-outputs presets
mA-outputs presets
Warm up:
4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Preset value: 4 . 0 mA
Cal.blowback: Hold
Preset value: 4 . 0 mA
Error:
Preset
Preset value: 3 . 4 mA
Warm up:
4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Non-Hold
Preset value: Hold
4.0mA
Preset
Cal.blowback: Hold
Preset value: 4 . 0 mA
Error:
Preset
Preset value: 3 . 4 mA
Enter
F08-1E.ai
Figure 8.1 “mA-outputs presets” Display
Enter
F08-2E.ai
Figure 8.2 Selecting Maintenance
(2) From this display (Figure 8.1), select the desired display. Figure 8.2 shows an example
of selecting “Maintenance”. Select the desired output status.
(3) If a preset value is selected, set the corresponding output current. If you select a preset
value just below Maintenance on the screen, the numeric-data entry display will appear.
Enter the current value you want. To set 10 mA, type in 010 and press the [Enter] key to
complete the setting. The setting range is from 2.4 to 21.6 mA.
NOTE
“Error” of mA-outputs presets is not displayed when option code “/C2” or “/C3” (NAMUR NE 43
compliant) is specified.
IM 11M13A01-02E
8-5
<8. Detailed Data Setting>
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as shown in
Table 8.3.
Table 8.3 Output Hold Default Values
Status
Output hold setting
Preset value
During warm-up
4 mA
4 mA
Under maintenance
Holds output at value just before maintenance started
4 mA
Under calibration or blow back
Holds output at value just before starting calibration or
blow back
4 mA
On Error occurrence
Holds output at a preset value
3.4 mA
IM 11M13A01-02E
8.3
8-6
<8. Detailed Data Setting>
Oxygen Concentration Alarms Setting
The analyzer enables the setting of four alarms — high-high, high, low, and low-low alarms —
depending upon the oxygen concentration. The following section sets out the alarm operations
and setting procedures.
8.3.1
Setting the Alarm Values
(1) High-high and high alarm values
If high-high and high alarm values are set to ON, then alarms occur if measured valves exceed
the alarm set values. The oxygen alarm set values can be set in the range of 0 to 100 vol%O2.
(2) Low and low-low alarm values
If low-low and low alarm values are set, then alarms occur if measured values fall below the
alarm set values. The oxygen alarm set values can be set in the range of 0 to 100 vol%O2.
8.3.2
Alarm Output Actions
If the measured values of the oxygen concentration fluctuate between normal (steady-state)
values and the alarm setting, alarm outputs may be frequently issued and canceled.
To avoid this, set the alarm delay and hysteresis for alarm canceling under the alarm output
conditions, as Figure 8.3 shows.
When a delay time is set, an alarm will not be issued so quickly even if the measured value differs
from the steady-state and enters the alarm setpoint range.
If the measured value remains within the alarm setpoint range for a certain period of time (for the
preset delay time), an alarm will be issued. On the other hand, there will be a similar delay each
time the measured value returns to the steady-state from the alarm setpoint range (canceling the
alarm status).
If hysteresis is set, alarms will be canceled when the measured value is less than or greater than
the preset hysteresis values.
If both the delay time and hysteresis are set, an alarm will be issued if the measured value is in
the alarm setpoint range and the delay time has elapsed.
For the alarm to be reset (canceled), the measured value must be beyond the preset hysteresis
value and the preset delay time must have elapsed.
Refer to Figure 8.3 for any further alarm output actions. The delayed time and hysteresis settings
are common to all alarm points.
Alarm range
A
B
C
D
7.5%
High limit alarm setpoint
Hysteresis
2.0%
5.5%
Oxygen concentration
Delay time:
Delay time:
5 seconds
5 seconds
Alarm output ON
OFF
Delay time:
5 seconds
F08-3E.ai
Figure 8.3 Alarm Output Action
IM 11M13A01-02E
8-7
<8. Detailed Data Setting>
In the example in Figure 8.3, the high limit alarm point is set to 7.5 vol%O2, the delay time is set to
five seconds, and hysteresis is set to 2 vol%O2.
Alarm output actions in each section in this figure are as follows:
A. Although the oxygen concentration value exceeds the high limit alarm setpoint, it falls below
the high limit alarm setpoint before the preset delay time of five seconds elapses. So, no
alarm is issued.
B. The oxygen concentration value exceeds the high limit alarm setpoint and the delay time
elapses during that measurement. So, an alarm is issued.
C. Although the oxygen concentration value falls below the hysteresis set value, the value rises
again and exceeds the hysteresis set value before the preset delay time elapses. So, the
alarm is not canceled.
D. The oxygen concentration value falls below the hysteresis set value and the preset delay
time elapses, so the alarm is canceled.
8.3.3
Alarm Setting Procedure
To set the alarm setpoints, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning display then
appears.
(3) Select the “Alarms setup” in the Commissioning display. The Alarms setup display shown in
Figure 8.4 then appears.
• To set the hysteresis, proceed to the following steps.
(4) Select “Hysteresis” in the Alarms setup display. The numeric-data entry display then
appears. Enter the desired hysteresis value as a percent of oxygen concentration. To set 2.5
vol%O2, enter “0025.” The hysteresis setting can be in the range of 0 to 9.9 vol%O2.
• To set the delay time, proceed as per the following steps.
(5) Select the “Contact delay” in the Alarms setup display. The numeric-data entry display then
appears. Enter the desired delay time, in seconds. To set three seconds, enter “003.” The
delay time setting can be in the range of 0 to 255 seconds.
• To set the alarm point, proceed to the following steps.
(6) Select the “Setpoints” in the Alarms setup display. The Oxygen alarms display then appears,
as shown in Figure 8.5.
(7) When you select “High alarm” in the Oxygen alarms display, the “OFF” or “ON” selection
display then appears. If you select “ON,” the High alarm will then be enabled (enable/
disable).
(8) To set the High alarm values select “Set value” just below the High alarm. The numeric-data
entry display then appears. Enter the alarm set value (percent of oxygen concentration). If
you want to set the alarm value to 10 vol%O2, enter “010.”
(9) Set the other alarm settings in the same manner as in the steps above.
NOTE
No alarm is issued when alarm is set to “OFF” (disabled). To use the alarm functions, be sure to
set the alarms “ON”.
IM 11M13A01-02E
Oxygen alarms
Alarms setup
Parameter:
Hysteresis:
Oxygen
0.1%O2
Contact delay:
3s
Setpoints
Enter
F08-4E.ai
Figure 8.4 “Alarms Setup” Display
8.3.4
8-8
<8. Detailed Data Setting>
High High alarm: OFF
Set value: 1 0 0 . 0 % O2
High alarm:
ON
Set value: 1 0 0 . 0 % O2
Low alarm:
OFF
Set value:
0 . 0 % O2
Low Low alarm: OFF
Set value:
0 . 0 % O2
Enter
F08-5E.ai
Figure 8.5 “Oxygen Alarms” Display
Default Values
When the analyzer is delivered, or if data are initialized, the default alarm set values are as
shown in Table 8.4.
Table 8.4 Alarm Setting Default Values
Item
Setting range
Hysteresis
0 to 9.9 vol%O2
0.1 vol%O2
Delay time
0 to 255 seconds
3 seconds
High-high limit alarm
OFF
High-high limit alarm setpoint
0 to 100 vol%O2
High limit alarm
100 vol%O2
OFF
High limit alarm setpoints
0 to 100 vol%O2
Low limit alarm
100 vol%O2
OFF
Low limit alarm setpoint
0 to 100 vol%O2
0 vol%O2
0 to 100 vol%O2
0 vol%O2
Low-low limit alarm
OFF
Low-low limit alarm setpoint
8.4
Default setting
Contact Output Setting
8.4.1
Contact Output
Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For details,
see Section 2.1, General Specifications) The operation modes of each contact output are as
follows. For Output contacts 1 to 3 you can select open or closed contact when the contact is
“operated”. Default is closed. For Output contact 4, contact is closed.
When power fails, Output contacts 1 to 3 are open, and 4 is closed.
Table 8.5
When no power is applied
to this equipment
Display
State when contact “ operated ”
Output contact 1
Open (deenergized) or closed (energized) selectable.
Open
Output contact 2
Open (deenergized) or closed (energized) selectable.
Open
Output contact 3
Open (deenergized) or closed (energized) selectable.
Output contact 4
Closed (deenergized) only
Open
Closed
IM 11M13A01-02E
8-9
<8. Detailed Data Setting>
8.4.2
Setting Procedure
To set the contact output, follow these steps.
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select the “Commissioning” in the Execution/Setup display. The Commissioning display
then appears.
(3) Select the “Contact setup” in the Commissioning display. The Contact setup display shown
in Figure 8.6 then appears.
(4) Select the desired Output contact. This section shows an example where “Output contact 1”
is selected (see Figure 8.7).
(5) Each set item and the selected items are briefly described in Table 8.6. The following
describes an example of setting where Output contact 1 is closed during calibration.
(6) Select “Others” in the Output contact 1 display. The “Contact1 Others” display shown in
Figure 8.8 then appears. Select “Calibration” in the Contact1 Others display.
(7) The ON or OFF selection display then appears. Select “ON” herein.
(8) Press the [Reject] key to go back to the previous display.
(9) Move the pointer to “During power-off the contact is open and in condition it is Open” and
press the [Enter] key. The “OFF” or “ON” selection display then appears. If you select “OFF,”
this means “Open” in normal conditions and “Closed” when the contact output is on.
CAUTION
• The Output contact 4 is fixed as “ close in power ON”, which cannot be changed by setting.
Output contact 1
Contact setup
Output contact 1
Output contact 2
Output contact 3
Alarms
Others
During power-off the
contact is open and
in condition it is Open
lnput contacts
Enter
Enter
F08-6E.ai
Figure 8.6 “Contact Setup” Display
Contact1
F08-7E.ai
Figure 8.7 “Output Contact 1” Display
Others
Warm up:
Range change:
Calibration:
Maintenance:
Blow back:
Temp.input high:
Cal.gas press.low:
Process up set:
ON
OFF
OFF
ON
OFF
OFF
OFF
OFF
Enter
F08-8E.ai
Figure 8.8 “Contact1 Others” Display
IM 11M13A01-02E
8-10
<8. Detailed Data Setting>
Table 8.6 Contact Output Settings
Item to be selected
Alarm
and
Error
settings
Other
settings
Brief description
High-high limit alarm
If “High-High alarm ON” is selected, contact output occurs when the
high-high limit is issued. To do this, it is required, in Alarms setup, that
the high-high alarm be set on beforehand (see Section 8.3).
High limit alarm
If “High alarm ON” is selected, contact output occurs when the high
limit alarm is provided. To do this, it is required, in Alarms setup, that
the high limit alarm be set on beforehand (see Section 8.3).
Low limit alarm
If “Low alarm ON” is selected, contact output occurs when the low
limit alarm is provided. To do this, it is required, in Alarms setup, that
the low limit alarm be set on beforehand (see Section 8.3).
Low-low limit alarm
If “Low-Low alarm ON” is selected, contact output occurs when the
low-low limit alarm is issued. To do this, it is required, in Alarms setup,
that the low-low alarm be set on beforehand (see Section 8.3).
Calibration coefficient
(correction ratio) alarm
If Calibration correction ratio alarm is ON (enabled), then when a
zero calibration correction ratio alarm (Alarm 6) or span calibration
correction ratio alarm (Alarm 7) occurs then calibration correction
ratio alarm contact output occurs (see Subsection 12.2.1)
Startup power stabilization
timeout alarm
If set ON then contact output occurs when startup power stabilization
timeout alarm (Alarm 8) occurs (see Subsection12.2.1)
Error
If “Error ON” is selected, contact output occurs when an error is
issued. (See Subsection 12.1.1).
Warm-up
If “Warm-up ON” is selected, contact output occurs during warm-up.
For the definition of Warm-up (see Subsection 8.2.1).
Output range change
If “Range change ON” is selected, contact output occurs (“answerback signal to a range change signal”) while a range change signal is
applied to a contact input. To do this, it is required, in “Input contacts”
setup, that the range change be selected beforehand. For more on
this (see Subsection 8.5.2).
Calibration
If “Calibration ON” is selected, contact output occurs during
calibration. For the definition of Under calibration (see Subsection
8.2.1).
Maintenance
If “Maintenance ON” is selected, contact output occurs during
maintenance. For the definition of Under maintenance (see
Subsection 8.2.1).
Blow back
If “Blow back ON” is selected, contact output occurs during blow
back. For the definition of During blow back (see Subsection 8.2.1).
High limit temperature alarm Not supported by the oxygen analyzer.
Calibration gas press. low
If “Cal. gas press. low ON” is selected, contact output occurs
(“answer-back signal to a calibration gas low pressure signal”)when a
calibration gas low pressure signal is applied to the contact input. To
do this, it is required, in “Input contacts” setup, that “Cal. gas press.
low” be selected beforehand. For more on this (see Subsection
8.5.2).
Process upset
If “Process upset” is selected, contact output occurs (“answer-back
signal to a process upset signal) when the process upset signal
is applied to the contact input. To do this, it is required, in “Input
contacts” setup, that “Process up set” be selected beforehand (see
Subsection 8.5.2).
Note: To provide an alarm with a contact output, be sure to make an alarm setting.
When using contact output as an answer-back signal for an input contact, be sure to make a contact input settings.
IM 11M13A01-02E
<8. Detailed Data Setting>
8.4.3
8-11
Default Values
When the analyzer is delivered, or if data are initialized, alarm and other setting defaults are as
shown in Table 8.7.
Table 8.7 Contact Output Default Settings
Item to be selected
Output
contact 1
Output
contact 2
Output
contact 3
High-high limit alarm
Alarm
settings
—
High limit alarm
ON
—
Low limit alarm
ON
—
Low-low limit alarm
—
Calibration coefficient (correction ratio) alarm
—
Startup power stabilization timeout alarm
—
Error
ON
Warm-up
ON
—
Output range change
—
Calibration
Other
settings
Output
contact 4
ON
Maintenance
—
ON
—
Blow back
—
High limit temperature alarm
—
Calibration gas press. low
(default)
(default)
(default)
Process upset
Operating contact status
—
—
Open
Closed
Closed
Closed
(fixed)
Note: Blank boxes in the above table indicate that the default is “disabled“.
IM 11M13A01-02E
<8. Detailed Data Setting>
8.5
Contact Input Setting
8.5.1
Contact Input Functions
8-12
The converter “Input contacts” execute set functions by accepting a remote dry-contact (“voltagefree contact”) signal. Table 8.8 shows the functions executed by a remote contact signal.
Table 8.8 Contact Input Functions
Item
Function
Calibration gas pressure low
Contact input disables Semi-automatic or Automatic Calibration.
Measuring range change
While Contact signal is ON, range of mA-Output 1 is switched to 0-25 vol%O2
and “Range” is displayed on the screen. See Figure 8.9.
Calibration start
Contact input starts Semi-Automatic Calibration. Calibration Mode setting must
be [Semi_Auto] or [Auto]. Contact signal must be applied for at least 1 sec.
Even if input signal continues to be applied, calibration is not repeated unless
contact input is released then reapplied.
Process upset
If the Contact signal is ON, heater power will be switched off. (A one- to
11-second time interval single-output signal is available as a contact signal.) If
this operation starts, the sensor temperature decreases and an error occurs.
To restore it to normal, turn the power off and then back on, or reset the
analyzer.
(Combustible gas detection)
Blow back start
Tag:
Contact input starts Blow back. Contact signal must be applied for at least 1
sec. Even if input signal continues to be applied, Blow back is not repeated
unless contact input is released then reapplied. (Refer to Section 10.4, Blow
Back.)
21.0
%O2
7.35mA -Output1
7.35mA -Output2
Range
F08-9E.ai
Figure 8.9 Changing Measuring Range with Contact Input
CAUTION
1. Measurement range switching function by an external contact input is available for analog
output1 only.
2. When making a semi-automatic calibration, be sure to set the semi-automatic or automatic
mode using the Calibration setup display.
3. When carrying out blow back, be sure to set “Blow back” in the “Contact setup”.
4. When the combustible gas detection signal is sent to the contact input, the converter will cut
the power supply to the heater of the detector. As a result, the heater temperature becomes
low and Error 1 or Error 2 happens.
IM 11M13A01-02E
8.5.2
8-13
<8. Detailed Data Setting>
Setting Procedure
This setting example shows how to set “When Input contact 1 Open, Start Semi_Auto
Calibration”. Refer to Section 7.11.3.
1) From the Basic panel display touch [Setup] key, and the Execution/Setup display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Contact setup” then “Input contacts”.
4) Select “Input 1” . Function Selection window (Fig 8.11) appears.
5) Select “Calibration start”.
6) For Input contact 1 select (operate if) “Closed”. (You can select “Open” or “Closed”).
7) Change Closed to Open.
Input contacts
Input contacts
Input1: No function
Input1: No
No function
function
Cal.
gas
Input1 contact
is press.
active low
at:
Range change
Closed start
Calibration
up set
Input2: Process
No function
Blow back start
Input2 contact is active at:
Closed
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
F08-10E.ai
Figure 8.10 “Input Contacts” Display
8.5.3
Enter
F08-11E.ai
Figure 8.11 “Input Contacts” Display
Default Values
All contact inputs are set to “No function”(disabled) prior factory shipment or after data
initialization..
IM 11M13A01-02E
8-14
<8. Detailed Data Setting>
8.6
Other Settings
8.6.1
Setting the Date-and-Time
The following describes how to set the date-and-time. Automatic calibration or blow back works
following this setting.
Proceed as follows:
1) If you touch the [Setup] key in the Basic panel display, the Execution/Setup display appears.
2) Select the “Commissioning” and the Commissioning display appears.
3) Select the “Others” and the display of Fig 8.12 appears.
4) Select the “Clock” and the display of Fig 8.13 appears.
5) Select the “Set date” and the numerical entry display appears. To enter June 21. 2004 ,
enter 210604. Touch the [Enter] key and you revert to the display of Fig 8.13.
6) Select the “Set time”. Enter the time in 24-hour format. To set 2:30 pm, enter 1430 on the
numerical display and touch enter. Touch the [Enter] key, the clock starts from 00 seconds..
Others
Clock
Clock
Averaging
Fuel setup
Purging
Unit setup
Passwords
Defaults
Set date: 1 3 / 0 7 / 0 0
Set time: 1 4 : 3 9
1 3 J ul. 2 0 0 0
14:39:55
Enter
F08-12E.ai
Figure 8.12 Other Settings
8.6.2
Enter
F08-13E.ai
Figure 8.13 Clock Display
Setting Periods over which Average Values Are
Calculated and Periods over which Maximum and
Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maximum and
minimum values under measurement (see Section 10.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration average
values are calculated and maximum and minimum values are monitored.
Procedure
1) From the Basic panel display touch [Setup] key, and the Execution/Setup display appears.
2) Select the “Commissioning” and the Commissioning display appears.
3) Select the “Others” then Averaging. The display of Figure 8.14 appears.
4) Select the “Set period over which average is calculated” and numerical entry display
appears. To set 3 hours, enter 003. Entry range is 1 to 255 hours.
5) Select the “Set period over which maximum and minimum is stored” and numerical entry
display appears. To set 48 hours, enter 048. Entry range is 1 to 255 hours.
IM 11M13A01-02E
8-15
<8. Detailed Data Setting>
Averaging
Set period over which
average is calculated:
1h
Set period over which
maximum and minimum is
stored:
24h
Enter
F08-14E.ai
Figure 8.14 S
etting Average Value Calculation Periods and
Maximum and Minimum Value Monitoring Periods
Default Values
“Set period over which average is calculated” is set to 1hr, and “Set period over which maximum
and minimum is stored” is set to 24 hrs prior factory shipment or after data initialization.. .
8.6.3
Setting Fuels
Input Parameters
The analyzer calculates the moisture content contained in exhaust gases. The following sets
forth the fuel parameters necessary for calculation and their entries. The moisture quantity may
be mathematically expressed by:
(water vapor caused by combustion and water vapor contained in the exhaust gas)
Moisture quantity =
+ (water vapor contained in air for combustion)
actual exhaust gas (including water vapor) per fuel
x 100
= (Gw + Gw1)/G x 100
=
Gw + (1.61 x Z x m x Ao)
Go + Gw + [(m - 1) Ao + (1.61 x Z x m x Ao)]
Gw + (1.61 x Z x m x Ao )
X + Ao x m
x 100
x 100
............ Equation 1
............ Equation 2
where,
Ao : Theoretical amount of air per unit quantity of fuel, m3/kg (or m3/m3) ... 2 in Table 8.9
G: Actual amount of exhaust gas (including water vapor) per unit quantity of fuel, m3/kg (or m3/m3)
Gw : Water vapor contained in exhaust gas per unit quantity of fuel (by hydrogen
and moisture content in fuel), m3/kg (or m3/m3)
............ 1 in Table 8.9
Gw1: Water vapor contained in exhaust gas per unit quantity of fuel (moisture content
in air), m3/kg (or m3/m3)
Go: Theoretical amount of dry exhaust gas per unit quantity of fuel, m3/kg (or m3/m3)
m: Air ratio
X : Fuel coefficient determined depending on low calorific power of fuel, m3/kg (or m3/m3) ....... 3 in Table 8.9
Z : Absolute humidity of the atmosphere, kg/kg ....... Figure 8.16
Siki08-2E.ai
Fill in the boxes with fuel parameters in Equation 2 above to calculate the moisture content.
Use Ao, Gw and X shown in Table 8.9. If there are no appropriate fuel data in Table 8.9, use the
following equations for calculation. Find the value of “Z” in Equations 1 and 2 using Japanese
Industrial Standards JIS B 8222. If a precise measurement is not required, obtain the value of “Z”
using a graph for the absolute humidity indicated by a dry and wet bulb hygrometer.
IM 11M13A01-02E
8-16
<8. Detailed Data Setting>
For liquid fuel
Amount of water vapor in exhaust gas (Gw) = (1/100) {1.24 (9h + w)} (m 3 /kg)
3
Theoretical amount of air (Ao) = 12.38 x (Hl/10000) – 1.36
(m /kg)
Low calorific power = Hl
X value = (3.37 / 10000) x Hx – 2.55 (m 3 /kg)
where,
Hl: low calorific power of fuel
h: Hydrogen in fuel (weight percentage)
w: Moisture content in fuel (weight percentage)
Hx: Same as numeric value of Hl
For gas fuel
Amount of water vapor in exhaust gas (Gw) = (1/100) {(h2) + 1/2 ∑y (Cx Hy) + wv} (m 3 /m3 )
Theoretical amount of air (Ao) = 11.2 x (Hl/10000)
3
3
(m /m )
Low calorific power = Hl
X value = (1.05 / 10000) x Hx
where,
3
3
(m /m )
Hl: low calorific power of fuel
CxHy: Each hydrocarbon in fuel (weight percentage)
h2: Hydrogen in fuel (weight percentage)
wv: Moisture content in fuel (weight percentage)
Hx: Same as numeric value of Hl
For solid fuel
3
Amount of water vapor in exhaust gas (Gw) = (1/100) {1.24 (9h + w)} (m /kg)
Theoretical amount of air (Ao) = 1.01 x (Hl / 1000) + 0.56
(m 3 /kg)
Low calorific power = Hl = Hh – 25 (9h + w) (kJ/kg)
X value = 1.11 - (0.106 / 1000 ) x Hx
where,
3
3
(m /m )
w: Total moisture content in use (weight percentage)
h: Hydrogen content (weight percentage)
The average hydrogen content of coal mined in Japan, which is a dry ash-free type, is
5.7 percent. Accordingly, "h" may be expressed mathematically by:
h = 5.7 [{100 – (w + a)} / 100] x (100 – w) / (100 – w1)
where,
a: Ash content (%)
w1: Moisture content (%), analyzed on a constant humidity basis
Hh: Higher calorific power of fuel (kJ/kg)
Hl: Low calorific power of fuel (kJ/kg)
Hx: Same numeric value of Hl
F08-15E.ai
Figure 8.15 Calculation Formula
IM 11M13A01-02E
<8. Detailed Data Setting>
8-17
40
39
0.046
38
0.044
0.042
37
0.040
36
0.038
35
0.036
34
0.034
33
0.032
32
0.030
31
0.028
30
Wet-bulb
temperature, °C
29
0.026 Absolute
28
0.024
27
26
0.022
25
0.020
24
0.018
22
20
0.016
18
0.014
16
0.012
14
12
4
2
0
8
6
humidity, kg/kg
0.010
10
0.008
0.006
0.004
-2
0.002
0
2
4
6
8
10
12 14 16
18 20
22 24 26 28 30 32 34 36
Dry-bulb temperature, °C
38 40
0.000
F08-16E.ai
Figure 8.16 Absolute Humidity of Air
IM 11M13A01-02E
8-18
<8. Detailed Data Setting>
Table 8.9 Fuel Data
• For liquid fuel
Fuel
properties Specific
weight
kg/l
Type
Chemical component
(weight percentage)
C
H
O
N
S
w
Theoretical
Amount of combustion
amount
X
gas m3/kg
of air for
value
combustion
Lower
CO2 H2O SO2 N2 Total
m3/kg
order
Calorific
power kJ/kg
Ash Higher
content order
Kerosene
0.78~
85.7 14.0
0.83
—
—
0.5
0.0
0.0
46465 43535
11.4
1.59 1.56 0.00 9.02 12.17 0.96
Light oil
0.81~
85.6 13.2
0.84
—
—
1.2
0.0
0.0
45879 43032
11.2
1.59 1.47 0.00 8.87 11.93 0.91
0.85~
85.9 12.0 0.7
0.88
0.5
0.5
0.3 0.05 45544 42739
10.9
1.60 1.34 0.00 8.61 11.55 0.89
0.83~
84.6 11.8 0.7
0.89
0.5
2.0
0.4 0.05 45125 42320
10.8
1.58 1.32 0.01 8.53 11.44 0.86
0.90~
84.5 11.3 0.4
0.93
0.4
3.0
0.5 0.05 43827 41274
10.7
1.58 1.27 0.02 8.44 11.31 0.77
0.93~
86.1 10.9 0.5
0.95
0.4
1.5
0.5
0.1
43952 41441
10.7
1.61 1.22 0.01 8.43 11.27 0.79
0.94~
84.4 10.7 0.5
0.96
0.4
3.5
0.5
0.1
43116 40646
10.5
1.58 1.20 0.02 8.32 11.12 0.72
0.92~
86.1 10.9 0.5
1.00
0.4
1.5
0.6
0.1
43660 41190
10.7
1.61 1.22 0.01 8.43 11.27 0.77
0.94~
83.0 10.5 0.5
0.97
0.4
3.5
2.0
0.1
43032 40604
10.3
1.55 1.18 0.02 8.18 10.93 0.72
Heavy No.1
oil
A class
1
No.2
B
Heavy oil
class 2
No.1
Heavy No.2
oil
C class
3
No.3
No.4
2
1
• For gas fuel
Fuel
properties Specific
Type
weight
kg/m3
CO
H2
CO2 CH2 CmHn O2
Higher
order
20428 18209
0.544
9.0 50.5 2.6 25.9 3.9
0.1
8.0
Blast furnace
gas
1.369
25.0 2.0 20.0
—
—
53.0
Natural gas
0.796
2.0 88.4 3.2
0.1
4.2
—
—
combustion
Lower
m3/m3
CO2 H2O
order
N2
Coke oven
gas
—
Combustion
product,
m3/ m3
Theoretical
Calorific
amount
power kJ/m3
of air for
Chemical component
(weight percentage)
3
N2
Total
X
value
4.455
0.45 1.10 3.60 5.15 0.46
3349
0.603
0.45 0.02 1.01 1.48 0.08
37883 34074
9.015
0.98 1.88 7.17 10.03 0.86
3391
Propane
2.030
C3H8 90%, C4H10 10%
102055 93976
24.63
3.10 4.10 19.5 26.7 2.36
Butane
2.530
C3H8 10%, C4H10 90%
125496 115868
30.37
3.90 4.90 24.0 32.8 2.91
(Gases)
Oxygen
(Molecular Formula)
1.43
O2
—
—
—
—
—
—
—
—
—
—
—
—
—
Nitrogen
1.25
N2
—
—
—
Hydrogen
0.09
H2
12767 10758
2.390
—
1.0
1.89 2.89 0.27
Carbon
monoxide
1.25
CO
12642 12642
2.390
1.0
—
1.89 2.89 0.32
Carbon dioxide
1.96
CO2
—
—
—
Methane
0.72
CH4
39750 35820
—
—
9.570
1.0
2.0
7.57 10.6 0.90
Ethane
1.34
C2H6
69638 63744
16.74
2.0
3.0
13.2 18.2 1.60
Ethylene
1.25
C2H4
62991 59060
14.35
2.0
2.0
11.4 15.4 1.48
Propane
1.97
C3H8
99070 91255
23.91
3.0
4.0
18.9 25.9 2.29
Butane
2.59
C4H10
128452 118623
31.09
4.0
5.0
24.6 33.6 2.98
2
1
—
—
—
3
IM 11M13A01-02E
8-19
<8. Detailed Data Setting>
Procedure
To make a fuel setting, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select the “Commissioning” in the Execution/Setup display. The Commissioning display
then appears.
(3) Select the “Others” in that display and then the Fuel setup shown in Figure 8.17.
(4) Choose the “Theoretical air quantity required” and the “Contents of moisture in exhaust gas”
in turn. The numeric-data entry display then appears. Enter numeric data using the numeric
keys.
(5) Choose “more” in the Fuel setup display. The Fuel setup shown in Figure 8.18 then appears.
(6) Set the numeric data to the “Value of the X coefficient” and then to the “Absolute humidity of
the atmosphere”.
(7) Choose “finished” to return to the display shown in Figure 8.17.
Fuel setup
Fuel setup
Value of the X coefficient:
1.00
Absolute humidity of the
atmosphere:
0. 1 0 0 0 kg/kg
finished.
See Manual for details!
Contents of moisture in
exhaust gas: 1. 0 0 m3/kg
Theoretical air quantity
required:
1. 0 0 m3/kg
more.....
Enter
F08-17E.ai
Figure 8.17 Fuel Setup
Enter
F08-18E.ai
Figure 8.18 Fuel Setup
Default Values
When the analyzer is delivered, or if data are initialized, default, parameter settings are as shown
in Table 8.10.
Table 8.10 Default Settings for Fuel Values
Item
Default setting
Amount of water vapor in exhaust gas
0.00 m3/kg (m3)
Theoretical amount of air
1.00 m3/kg (m3)
X value
1.00
Absolute humidity of the atmosphere
0.0 kg/kg
IM 11M13A01-02E
8-20
<8. Detailed Data Setting>
8.6.4
Setting Purging
Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100 °C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
Tag:
30
°C
4mA -Output1
4mA -Output2
Purge
Hold
F08-19E.ai
Figure 8.19 Display during Purging
Procedure
Set the purging time as follows:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select the “Commissioning” in the Execution/Setup display. The Commissioning display
then appears.
(3) Select the “Others” in that display and the Others display then appears, as shown in Figure
8.20.
(4) Select the “Purging”. The purging time setting display appears, as shown in Figure 8.21.
(5) Point to the “Purging time” and press the [Enter] key. Then the display for selecting purging
time appears.
(6) Enter the desired numeric value from the numeric-data entry display.
The allowable input ranges from 0 to 60 minutes.
Purging
Others
Clock
Averaging
Fuel setup
Purging
Unit setup
Passwords
Defaults
Purging time: 0 min
Enter
F08-20E.ai
Figure 8.20 Other Settings
Enter
F08-21E.ai
Figure 8.21 Purging Display
IM 11M13A01-02E
8-21
<8. Detailed Data Setting>
8.6.5
Setting Passwords
Unauthorized access to lower level menu displays from the Execution/Setup display can be
protected by passwords. You can set separate passwords for Calibration, Blow back, and
Maintenance and for Commissioning
Proceed as follows:
1) From the Basic panel display touch the [Setup] key, and the Execution/Setup display
appears.
2) Select the “Commissioning” , and the Commissioning display appears.
3) Select the “Others” then the “Passwords”. The display shown in Figure 8.22 appears.
First set the password for Calibration, Blow back, and Maintenance.
4) Select “Calibration, Blow back, and Maintenance”.
5) A text entry display appears. Enter password as up to 8 alphanumeric characters.
6) You can enter a password for Commissioning by the same procedure.
7) Record the passwords and look after them carefully.
Passwords
Calibration,Blow back
and Maintenance:
Commissioning:
Enter
F08-22E.ai
Figure 8.22 Passwords Display
The passwords are not set as shipped from factory. If you reset initialize the parameters, and
password settings are deleted.
If you forget a password, select “Commissioning” in the Execution/Setup display, and enter
“MOON”. By doing so, you can enter the Commissioning display only. Then display the
Passwords and verify the set passwords.
IM 11M13A01-02E
Blank Page
9-1
<9. Calibration>
9. Calibration
9.1
Calibration Briefs
9.1.1
Principle of Measurement
This section sets forth the principles of measurement with a zirconia oxygen analyzer before
detailing calibration.
A solid electrolyte such as zirconia allows the conductivity of oxygen ions at high temperatures.
Therefore, when a zirconia-plated element with platinum electrodes on both sides is heated up
in contact with gases having different oxygen partial pressures on each side, the element shows
the action of the concentration cell. In other words, the electrode in contact with a gas with a
higher oxygen partial pressure acts as a negative electrode. As the gas comes in contact with the
zirconia element in this negative electrode, oxygen molecules in the gas acquire electrons and
become ions. Moving in the zirconia element, they eventually arrive at the positive electrode on
the opposite side.
There, the electrons are released and the ions return to the oxygen molecules. This reaction is
indicated as follows:
Negative electrode: O2 + 4e
Positive electrode: 2 O 2-
2 O 2O2 + 4 e
The electromotive force E (mV) between the two electrodes, generated by the reaction, is
governed by Nernst’s equation as follows:
E = -RT/nF ln Px/Pa........................................Equation (1)
where,
R: Gas constant
T: Absolute temperature
n: 4
F: Faraday’s constant
Px: O
xygen concentration in a gas in contact with the positive zirconia
electrode (%)
Pa: Oxygen concentration in a gas in contact with the negative zirconia
electrode (%)
Assuming the zirconia element is heated up to 750°C, then we obtain equation (2) below.
E = −50.74 log Px ...................................... Equation (2)
Pa
Siki09E.ai
With this analyzer, the sensor (zirconia element) is heated up to 750°C, so Equation (2) is valid.
At that point, the relationship as in Figure 9.1 is effected between the oxygen concentration of the
sample gas in contact with the positive electrode and the electromotive force of the sensor (cell),
where a comparison gas of air is used on the negative electrode side.
IM 11M13A01-02E
9-2
<9. Calibration>
120
100
0.51 vol% O2,81.92mV(Zero origin of calibration)
80
Cell
voltage
(mV)
60
40
20
21.0 vol% O2, 0mV
(Span origin of calibration)
0
-20
-40
0.1
0.5
1
5
10
21.0
50
100
Oxygen concentration (vol % O2)
F09-1E.ai
Figure 9.1 Oxygen concentration in a Sample Gas vs Cell Voltage (21 vol%O2 Equivalent)
The measurement principles of a zirconia oxygen analyzer have been described above.
However, the relationship between oxygen concentration and the electromotive force of a cell is
only theoretical. Usually, in practice, a sensor shows a slight deviation from the theoretical value.
This is the reason why calibration is necessary. To meet this requirement, an analyzer calibration
is conducted so that a calibration curve is obtained, which corrects the deviation from the
theoretical cell electromotive force.
9.1.2
Calibration Gas
A gas with a known oxygen concentration is used for calibration. Normal calibration is performed
using two different gases: a zero gas of low oxygen concentration and a span gas of high oxygen
concentration. In some cases, only one of the gases needs to be used for calibration. However,
even if only one of the gases is normally used, calibration using both gases should be done at
least once.
The zero gas normally used has an oxygen concentration of 0.95 to 1.0 vol%O2 with a balance of
nitrogen gas (N2). The span gas widely used is clean air (at a dew-point temperature below -20
°C and free of oily mist or dust, as in instrument air).
For best accuracy, as the span gas use oxygen whose concentration is near the top of the
measurement range, in a nitrogen mixture.
IM 11M13A01-02E
9-3
<9. Calibration>
9.1.3
Compensation
The deviation of a measured value from the theoretical cell electromotive force is checked by the
method in Figure 9.2 or 9.3.
Figure 9.2 shows a two-point calibration using two gases: zero and span. Cell electromotive
forces for a span gas with an oxygen concentration p1 and a zero gas with an oxygen
concentration p2 are measured while determining the calibration curve passing between these
two points. The oxygen concentration of the sample gas is determined from this calibration
curve. In addition, the calibration curve corrected by calibration is compared with the theoretical
calibration curve for determining the zero correction ratio represented by B/A x 100 (%) on the
basis of A, B and C shown in Figure 9.2 and a span correction ratio of C/A x 100 (%). If the zero
correction ratio exceeds the range of 100±30% or the span correction ratio becomes larger than
0±18%, calibration of the sensor becomes impossible.
81.92
Zero origin
ez
e2
Cell
electromotive
force, mV
Calibration curve
before correction
e1
es
B
A
Corrected calibration
curve (theoretical
calibration curve)
C
0
21.0
p1
Span gas concentration
Span origin
p2
0.51
Zero gas concentration
Oxygen concentration (vol%O2)
Zero correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%
Span correction ratio = (C/A) x 100 (%) Correctable range: 0 ± 18%
F09-2E.ai
Figure 9.2 Calculation of a Two-point Calibration Curve and Correction Ratios using Zero and
Span Gases
Figure 9.3 shows a one-point calibration using only a span gas. In this case, only the cell
electromotive force for a span gas with oxygen concentration p1 is measured. The cell
electromotive force for the zero gas is carried over from a previous measurement to obtain the
calibration curve. The principle of calibration using only a span gas also applies to the one-point
calibration method using a zero gas only.
81.92
Zero origin
ez
Cell
electromotive
force, mV
e1
es
Calibration curve before
correction
Previous
zero gas data
B
A
Corrected calibration curve
(theoretical calibration curve)
C
0
21.0
p1
Span gas concentration
0.51
Span origin
Oxygen concentration (vol%O2)
Zero correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%
Span correction ratio = (C/A) x 100 (%) Correctable range: 0 ± 18%
F09-3E.ai
Figure 9.3 Calculation of a One-point Calibration Curve and Correction Ratios using a Span Gas
IM 11M13A01-02E
9.1.4
9-4
<9. Calibration>
Characteristic Data from a Sensor Measured During
Calibration
During calibration, calibration data and sensor status data (listed below) are acquired.
However, if the calibration is not properly conducted (an error occurs in automatic or semiautomatic calibration), these data are not collected in the current calibration.
These data can be observed by selecting the [Detailed-data] key from the Basic panel display.
For an explanation and the operating procedures of individual data, consult Subsection 10.1.1,
Detailed-data Display.
(1) Record of span correction ratio
Recorded the past ten span correction ratios.
(2) Record of zero correction ratio
Recorded the past ten zero correction ratios.
(3) Response time
You can monitor the response time provided that a two-point calibration has been done in semiautomatic or automatic calibration.
(4) Cell’s internal resistance
The cell’s internal resistance gradually increases as the cell (sensor) deteriorates. You can
monitor the values measured during the latest calibration. However, these values include the
cell’s internal resistance and other wiring connection resistance. So, the cell’s degrading cannot
be estimated from these values only.
When only a span calibration has been made, these values will not be measured, and previously
measured values will remain.
(5) Robustness of a cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
in a number on four levels.
IM 11M13A01-02E
9-5
<9. Calibration>
9.2
Calibration Procedures
NOTE
Calibration should be made under normal operating conditions (if the probe is connected to a
furnace, the analyzer will undergo calibration under the operating conditions of the furnace).
To make a precise calibration, conduct both zero and span calibrations.
9.2.1
Calibration Setting
The following sets forth the required calibration settings:
Mode
There are three calibration modes available:
(1) Manual calibration which allows zero and span calibrations or either one manually in turn;
(2) Semi-automatic calibration which lets calibration start with the touchpanel or a contact input,
and undergoes a series of calibration operations following preset calibration interval and
stabilization time.
(3) Automatic calibration which is carried out automatically following preset calibration interval.
Calibrations are limited by the following mode selection:
• When Manual calibration is selected:
Manual calibration only can be conducted. (This mode does not allow semi-automatic calibration
with a contact input nor automatic calibration even when its start-up time has reached.)
• When Semi-automatic calibration is selected:
This mode enables manual and semi-automatic calibrations to be conducted. (The mode,
however, does not allow automatic calibration even when its start-up time has reached.)
• When Automatic calibration is selected:
This calibration can be conducted in any mode.
To execute this calibration, follow these steps:
(1) Select the [Setup] key from the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” from the Execution/Setup display.
(2) Select “Calibration setup” from the Maintenance display. Then select “Mode” from the
Calibration setup display (see Figure 9.4).
Now you can select “Manual”, “Semi_Auto”, or “Auto” calibration.
IM 11M13A01-02E
9-6
<9. Calibration>
Calibration Procedure
Select both span and zero calibrations or span calibration only or zero calibration only.
Usually select span and zero calibrations.
Select “Points” from the Calibration setup display and then you can select “Both,” “Span” or
“Zero” (see Figure 9.5).
Calibration setup
Calibration setup
Mode:
Mode: Manual
Points: Semi_Auto
Auto
Zero gas conc:
1.00%
Manual
Points: Both
Span
Zero gas
conc:
1.00%
Zero
Span gas conc: 21 . 0 0 %
Timing
Span gas conc: 2 1 . 0 0 %
Timing
Enter
Enter
F09-4E.ai
Figure 9.4 Calibration Setup display
F09-5E.ai
Figure 9.5 Calibration Setup display
Zero gas Concentration
Set the oxygen concentration for zero calibration. Enter the oxygen concentration for the zero
gas in the cylinder used in the following procedures:
Select “Zero gas conc” from the Calibration setup display. The numeric-data entry display then
appears. Enter the desired oxygen concentration for the zero calibration. (The zero gas set
ranges from 0.3 to 100 vol%O2.)
Enter 00098 for an oxygen concentration of 0.98 vol%O2 .
Span gas Concentration
Set the oxygen concentration for span calibration. If instrument air is used as the span gas, enter
21%O2.
Select “Span gas conc” from the Calibration setup display. Enter the desired span gas oxygen
concentration from the numeric-data entry display.
(The span gas set ranges from 4.5 to 100 vol%O2.)
Enter 02100 for an oxygen concentration of 21 vol%O2.
Instrument air is here defined as dry air with a dew-point temperature of no higher than -20°C. If
the dew-point temperature is higher than -20°C, use a hand-held oxygen analyzer to measure
the actual oxygen concentration.
When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas), use a
hand-held oxygen analyzer to measure the actual oxygen concentration, and then enter it.
CAUTION
(1) When the instrument air supply is to be used as span gas, cool it to -20°C below dew point
to remove moisture, oil mist and dust from the air.
(2) If you do not do this to purify the air, then the accuracy of the calibration may be affected.
IM 11M13A01-02E
9-7
<9. Calibration>
Setting Calibration Time
• When the calibration mode is in manual:
First set the “Hold time” (output stabilization time). This indicates the time required from the
end of calibration to entering a measurement again. This time, after calibration, the sample gas
enters the sensor to set the time until the output returns to normal. The output remains held
after completing the calibration operation until the hold (output stabilization) time elapses. The
calibration time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. For more
details, consult Section 8.2, Output Hold Setting.
• When the calibration mode is in semi-automatic:
Set the hold (output stabilization) time and calibration time. The calibration time is the time
required from starting the flow of the calibration gas to reading out the measured value. The
set calibration time is effective in conducting both zero and span calibrations. The calibrationtime set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. Figure 9.6 shows the
relationship between the calibration time and hold (output stabilization) time.
Calibration start
(contact or switch input)
Span calibration
(span gas valve open)
Calibration time
Zero calibration
(zero gas valve open)
Calibration time
Analog output status
Hold (output
stabilization) time
Analog output remains held
(when output remains held)
F09-6E.ai
Figure 9.6 Calibration and Hold Time Settings
• When the calibration mode is in automatic:
In addition to the above Hold (output stabilization) time and Calibration time, set the Interval, Start
date, and Start time.
Interval means the calibration intervals ranging from 000 days, 00 hours to 255 days, 23 hours.
Set the first calibration day and the start-calibration time to the “Start date” and “Start time”
respectively. For example, to start the first calibration at 1:30 p.m. on March 25, 2001, enter
25/03/01 to the start date and 13 hours, 30 minutes to the start time, following the steps below:
(1) Select the “Calibration timing” display. A panel display as shown in Figure 9.7 appears.
(2) Select each item for the calibration to display the numeric-data entry display. Enter the
desired numeric values for the calibration.
Calibration timing
Hold time: 1 0 min 0 0 s
Cal time: 1 0 min 0 0 s
lnterval:
30d 00h
Start date: 0 1 / 0 1 / 0 0
Start time: 0 0 : 0 0
Enter
F09-7E.ai
Figure 9.7 Calibration Timing Display
IM 11M13A01-02E
9-8
<9. Calibration>
NOTE
When setting calibration timing requirements, bear the following precautions in mind:
(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus
calibration time, the second calibration start time will conflict with the first calibration. In such
a case, the second calibration will not be conducted. (When both zero and span calibrations
are to be performed, the calibration time is double that required for a single (zero or span)
calibration.)
(2) For the same reason, if the calibration start time conflicts with manual calibration or semiautomatic calibration, the current calibration will not be conducted.
(3) If the calibration time conflicts with maintenance service or blow back operations, calibration
will start after completing the maintenance service or blow back operations (see Subsection
8.2.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be
conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
9.2.2
Default Values
When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as
shown in Table 9.1.
Table 9.1 Default Settings for Calibration
Item
Default Setting
Calibration mode
Manual
Calibration procedure
Span - zero
Zero gas (oxygen) concentration
1.00%
Span gas (oxygen) concentration
21.00%
Hold (output stabilization) time
10 minutes, 00 seconds
Calibration time
10 minutes, 00 seconds
Calibration interval
30 days, 00 hours
Start date
01 / 01 / 00
Start time
0:00
IM 11M13A01-02E
9.2.3
9-9
<9. Calibration>
Calibration
NOTE
1) Perform calibration under normal working conditions (e.g. continuous operation with sensor
mounted on furnace).
2) Perform both Span and Zero calibration for best resultant accuracy.
3) When instrument air is used for the span calibration, remove the moisture from the
instrument air at a dew-point temperature of -20°C and also remove any oily mist and dust
from that air.
4) If dehumidifying is not enough, or if foul air is used, the measurement accuracy will be
adversely affected.
Manual Calibration
For manual calibration, consult Section 7.12, Calibration, earlier in this manual.
Semi-automatic Calibration
To start calibration, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select the “Calibration” in the Execution/Setup display. The Calibration display shown
in Figure 9.8 appears.
(2) Select the “Semi-auto calibration” to display the Semi-auto calibration display shown in
Figure 9.9.
(3) Select the “Start calibration”. The display shown in Figure 9.10 appears, and then start
calibration.
Calibration
Semi-auto calibration
Check settings
Zero gas conc:
1.00%
Span gas conc: 2 1 . 0 0 %
Cal time: 1 0 min 0 0 s
Go to setup to change value.
Start calibration
Cancel calibration
Manual calibration
Semi-auto calibration
Enter
Enter
F09-8E.ai
Figure 9.8 Calibration Display
F09-9E.ai
Figure 9.9 Semi-auto Calibration Display
Semi-auto calibration
2 1 . 0 0%
0.5min./div
1 . 0 0%
2 1 . 0 % O2
CAL. TIME
F09-10E.ai
Figure 9.10 Semi-auto Calibration Display
IM 11M13A01-02E
9-10
<9. Calibration>
To start calibration using a contact input, follow these steps:
(1) Make sure that Calibration start has been selected in the “Input contacts” display (see
Section 8.5, earlier in this manual).
(2) Apply an input contact to start calibration.
To stop calibration midway, follow these steps:
(1) Press the [Reject] key. If this key is pressed midway during calibration, the calibration will
stop and the hold (output stabilization) time will be set up.
(2) Press the [Reject] key once again to return to the Basic panel display and the analyzer will
be in normal measurement.
Automatic Calibration
No execution operations are required for automatic calibration. Automatic calibration starts in
accordance with a preset start day and time. Calibration is then executed at preset intervals.
NOTE
Before starting Semi-automatic calibration or Automatic calibration, operate the calibration gas
solenoid valves and adjust calibration gas flow to 600 ± 60 ml/min.
IM 11M13A01-02E
10-1
<10. Other Functions>
10. Other Functions
10.1 Detailed-data Display
Press the [Detailed-data] key on the Basic panel display to view the detailed operation data as
shown in Figure 10.1.
Pressing the [▼] or [▲] key, you can advance the page or go back to your desired page.
• Detailed-data display
There are nine panel displays for viewing detailed data. The following briefly describes each data
item.
Tag:
Tag:
Cell voltage:
0 . 6 mV
4 2 . 1 mV
Thermo voltage:
C.J.resistance: 1 1 8 1 . 4 Ω
44Ω
Cell resistance:
0.24
Soft.rev.:
Span gas ratio:
0.0%
Zero gas ratio:
100.0%
Response time:
0s
Cell robustness: life> 1year
Cell temperature:
750 °C
C.J.temperature:
47 °C
Hold
F10-1aE.ai
Hold
F10-1bE.ai
Figure 10.1 Detailed-data Display
10.1.1
Cell Voltage
The cell (sensor) voltage will be an index to determine the amount of degradation of the
sensor. The cell voltage corresponds to the oxygen concentration currently being measured.
If the indicated voltage approximates the ideal value (corresponding to the measured oxygen
concentration), the sensor will be assumed to be normal.
The ideal value of the cell voltage (E), when the oxygen concentration measurement temperature
is controlled at 750°C, may be expressed mathematically by:
E = -50.74 log (Px/Pa) [mV]
where, Px: Oxygen concentration in the sample gas
Pa: Oxygen concentration in the reference gas, (21%O2)
Table 10.1 shows oxygen concentration versus cell voltage.
Table 10.1 Oxygen Concentration Vs. Cell Voltage, (cell temperature: 750 ºC)
%O2
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
mv
117.83
102.56
93.62
87.28
82.36
78.35
74.95
72.01
69.41
%O2
1
2
3
4
5
6
7
8
9
mv
67.09
51.82
42.88
36.54
31.62
27.61
24.21
21.27
18.67
%O2
10
21.0
30
40
50
60
70
80
90
mv
16.35
0
-7.86
-14.2
-19.2
-23.1
-26.5
-29.5
-32.1
%O2
100
mv
-34.4
IM 11M13A01-02E
10-2
<10. Other Functions>
10.1.2 Thermocouple Voltage
The cell temperature is measured with a Type K (chromel-alumel) thermocouple.
The thermocouple cold junction is located in the detector terminal box. The cell temperature and
the thermocouple voltage (including the voltage corresponding to the cold junction temperature)
are displayed.
10.1.3 Cold Junction Resistance (C.J. Voltage)
The ZR22 Detector measures the cold junction temperature using an RTD (Pt 1000).
(The earlier model of ZO21D uses transistors to measure the cold junction temperature.)
If detector “ZR22” is selected in the Basic setup display, the RTD resistance values will be
displayed. If the ZO21D is selected, the transistor voltage will be displayed.
10.1.4
Cell’s Internal Resistance
A new cell (sensor) indicates its internal resistance of 200 Ω maximum. As the cell degrades, so
will the cell’s internal resistance increase. The degradation of the cell cannot be evaluated just by
changes in cell’s internal resistance, however. Those changes in the cell’s internal resistance are
just a guide to the extent the cell is degrading. The updated values obtained during the calibration
are displayed.
10.1.5
Software Revision
The revision (number) of the software installed is displayed.
10.1.6
Span gas and Zero gas Correction Ratios
These are used to check for degradation of the sensor (cell). If the correction ratio is beyond the
limits as shown in Figure 10.2, the sensor should no longer be used.
These ratios can be found by calculating the data as shown below.
81.92
Zero origin
ez
Cell
electromotive
force, mV
Calibration curve
before correction
Previous
zero gas data
B
A
e1
Corrected calibration curve
(theoretical calibration curve)
es
C
0
21.0
p1
(Span gas concentration)
Oxygen concentration (vol%O2)
Zero gas correction ratio = (B/A) x 100 (%)
Span gas correction ratio = (C/A) x 100 (%)
0.51
Span origin
F10-2E.ai
Correctable range: 100 ± 30%
Correctable range: 0 ± 18%
Figure 10.2 Span gas and Zero gas Correction Ratios
IM 11M13A01-02E
10-3
<10. Other Functions>
10.1.7
Cell Response Time
The cell’s response time is obtained in the procedure shown in Figure 10.3. If only either a zero or
span calibration has been carried out, the response time will not be measured just as it will not be
measured in manual calibration.
Five minutes maximum
Response time
mA
90%
100%
10 % of analog
output span
Time
Start calibration
Calibration
complete
The response time is obtained after the corrected calibration curve has been found. The response time
is calculated, starting at the point corresponding to 10 % of the analog output up to the point at 90 % of
the analog output span. That is, this response time is a 10 to 90 % response.
F10-3E.ai
Figure 10.3 Typical response time characteristic
10.1.8
Robustness of a Cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
as one of four time periods during which the cell may still be used:
(1) more than a year
(2) more than six months
(3) more than three months
(4) less than one month
The above four time periods are tentative and only used for preventive maintenance, not for
warranty of the performance.
This cell’s robustness can be found by a total evaluation of data involving the response time,
the cell’s internal resistance, and calibration factor. However, if a zero or span calibration was not
made, the response time cannot be measured. In such a case, the response time is not used as
a factor in evaluating the cell’s robustness.
10.1.9
Cell Temperature
This displays the cell (sensor) temperature, which is determined from the thermocouple
electromotive force (e.m.f.) and cold junction temperature. Normally it is 750°C.
IM 11M13A01-02E
10-4
<10. Other Functions>
10.1.10 C. J. Temperature
This indicates the detector terminal box temperature, which compensates for the cold junction
temperature for a thermocouple measuring the cell temperature. When the ZR22 Detector is
used, the maximum C. J. temperature will be 150°C. If the terminal box temperature exceeds
this, take measures to shield the terminal box from heat radiation.
The maximum C. J. temperature varies depending on the type of detector.
10.1.11 Maximum Oxygen Concentration
The maximum oxygen concentration and the time of its occurrence during the period specified
in the Averaging display are displayed. After the present monitoring interval has elapsed, the
maximum oxygen concentration that has been displayed so far will be cleared and a new
maximum oxygen concentration will be displayed. If the setup period of time is changed, the
current maximum oxygen concentration will be displayed (for more details, see Subsection 8.6.2
earlier in this manual).
10.1.12 Minimum Oxygen Concentration
The minimum oxygen concentration and the time of its occurrence during the period specified in the
Averaging display are displayed. If the setup period elapses, the minimum oxygen concentration
that has been displayed so far will be cleared and a new minimum oxygen concentration will be
displayed. If the setup period of time is changed, the current minimum oxygen concentration will be
displayed (for more details, see Subsection 8.6.2 earlier in this manual).
10.1.13 Average Oxygen Concentration
The average oxygen concentration during the periods over which average values are calculated
is displayed. If the setup period elapses, the average oxygen concentration that has been
displayed so far will be cleared and a new average oxygen concentration will be displayed. If the
setup period of time is changed, the current average oxygen concentration will be displayed (for
more details, see Subsection 8.6.2 earlier in this manual).
10.1.14 Heater On-Time Ratio
The probe sensor is heated to and maintained at 750°C. When the sample gas temperature is
high, the amount of heater ON-time decreases.
10.1.15 Time
The current date and time are displayed. These are backed up with built-in batteries, so the clock
continue the run even if the power is switched off.
10.1.16 History of Calibration Time
The calibration-conducted dates and times, and span gas and zero gas correction ratios for the
past ten calibrations are stored in memory.
10.1.17 Power Supply Voltage
For the temperature control for the heater of the detector to work best, you should set the power
supply voltage and frequency appropriately, as the control parameters are based on this. Set the
AC supply voltage to “Low” if supply is 140 V AC or less, and to “High” if it is 180 V AC or more.
10.1.18 Power Frequency
Set the AC supply frequency setting appropriately — “Low” for 50 Hz, and “High” for 60 Hz.
IM 11M13A01-02E
<10. Other Functions>
10-5
10.2 Trend Graph
Press the [Graph display] key in the Basic panel display to switch to the graph display.
This will help grasp the measured-value trend. Touching anywhere on the graph display will
return to the Basic panel display. To set the Trend graph display, follow the steps in Subsection
10.2.1.
10.2.1
Setting Display Items
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select the “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” in the Maintenance display.
(3) Select the “Trend graph” in the Display setup display. The Trend graph display shown in
Figure 10.4 appears.
(4) Select the “Parameter: Oxygen” in the Trend graph display. Then select the desired display
item shown in Table 10.2.
Table 10.2 Trend Graph Display Items
Selected item
Description
Oxygen concentration
Oxygen concentration graph under measurement
Output 1
If this equipment is for the oxygen analyzer, the trend graph will be an
oxygen concentration graph.
Output 2
If this equipment is for the oxygen analyzer, the trend graph will be an
oxygen concentration graph.
Trend graph
Parameter:
Oxygen
Sample interval:
Scale Oxygen
Upper limit: 2 5 . 0
Lower limit:
0.0
30s
% O2
% O2
Enter
F10-4E.ai
Figure 10.4 Trend Graph Display
10.2.2
Sampling Interval
To plot a graph, set the sampling period for the measurement data.
This graph allows the plotting of 60 data items on one division on the time axis. So, if you set
a ten-second sampling interval, one division corresponds to 600 seconds (Figure 10.5). The
allowable sampling intervals range from 1 to 30 seconds. If you set a one-second sampling
interval, the axis of the abscissas then corresponds to five minutes.
If you set it to 30 seconds, the axis of the abscissas then corresponds to 150 minutes.
IM 11M13A01-02E
10-6
<10. Other Functions>
Preset upper limit
25.0 %O2
Preset lower limit
0.0 %O2
10 min./div
12.3 %O2
Time per division,
calculated by the set
sampling interval
Currently measured
value
60 data
Sampling period
F10-5E.ai
Figure 10.5 Plotting Graph for Sampling Interval
10.2.3 Setting Upper and Lower Limit Values on Graph
Set upper and lower limit values on the graph in the following procedure:
Press “Upper limit” in the Trend graph display. The numeric-data entry key appears.
Enter the upper limit value. Also enter the lower limit value in the same way. The allowable
settings for both upper limit and lower limit values range from 0 to 100%O2.
10.2.4
Default Setting
When the analyzer is delivered, or if data are initialized, the set data are by default, as shown in
Table 10.3.
Table 10.3 Default Values for Graph Setting
Item Default
Value
Parameter
Oxygen concentration
Sample interval
30 seconds
Upper limit
25%O2
Lower limit
0%O2
NOTE
If a rapid change in the measured value occurs during sampling, no sampled data are plotted on
the graph. Use the graph indication tentatively. Check the output current for accurate data.
IM 11M13A01-02E
10-7
<10. Other Functions>
10.3
Other Display-related Functions
10.3.1
Auto-Return Time
On the Execution/Setup display or lower level menu displays, if no keys are touched for a preset
time, the Auto return time, then the display will automatically revert to the Basic panel display. The
“Auto return time” can be set in the range 0 to 255 minutes. If it is set to 0, then the display does
not automatically revert. By default, the “Auto return time” is set to 0 (zero).
Set the Auto return time in the following procedure:
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display,
then select “Maintenance”. Select “Display setup”, then “Auto return time”.
(2) The numerical entry display appears for you to enter the desired Auto return time.
To set an Auto return time of one hour, enter 060.
(3) If you set 0, then the Auto return function does not operate.
10.3.2
Entering Tag Name
You can attach a desired tag name to the equipment. To attach it, follow these steps:
(1) Select the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select the “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” in the Maintenance display.
(3) Select the “Display item” in the Display setup display. The display shown in Figure 10.6 then
appears.
(4) Select the “Tag name” in the Display item display. The text-data entry display then appears.
(5) Enter up to 12 alphanumeric characters including codes for the desired tag name.
10.3.3
Language Selection
You can select a display language from among English, Japanese, German and French.
The display language is set to the one specified in the purchase order when the analyzer is
shipped from the factory.
To select the language you want, follow these steps:
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select the “Maintenance”, then “Display setup”.
(3) Select the “Language” and the dropdown selection (Figure 10.7) allows you to select the
desired language.
Display setup
Display item
Primary value:
Oxygen
Secondary value:
mA-ouput1
Tertiary value:
mA-output2
Tag name:
Display item
Trend graph
Auto return time:
0 min
Language: English
Deutsch
Francias
Enter
F10-6E.ai
Figure 10.6 Display Item Display
Enter
F10-7E.ai
Figure 10.7 Selecting the Display Language
IM 11M13A01-02E
<10. Other Functions>
10-8
10.4 Blow Back
This section explains the parameter settings for performing blow back.
10.4.1
Mode of Blow back
There are three modes of blow back operation: no function, semi-automatic, and automatic.
Blow back is not performed when the mode is set to “No function”. In “Semi_Auto” mode, blow
back can be started by key operation on the display or by a contact input signal, and then
sequentially performed at a preset blow back time and hold time. In “Auto” mode, blow back is
automatically performed at preset intervals. The following restrictions apply:
When “No function” is selected, blow back is not performed.
When “Semi_Auto” is selected, semi-automatic blow back can be performed. (Blow back does
not start at Auto blow back start time.)
When “Auto” is selected, blow back can be performed in either “Auto” or “Semi_Auto” mode.
(1) From the Basic panel display, touch [Setup] key; on the Execution/Setup display which
appears, select “Maintenance”.
(2) On the Maintenance display, select “Blow back setup” and the Mode selection pull down
allows you to select between “No function”, “Semi_Auto” and “Auto” (see Figure 10.8).
Blow back setup
Mode: No function
Semi_Auto
Auto
Enter
F10-8E.ai
Figure 10.8 Blow back setup Display
10.4.2 Operation of Blow back
Figure 10.9 shows a timing chart for the operation of blow back. To execute blow back with a
contact input, use a contact input with an ON-time period of one to 11 seconds.
Once blow back starts, a contact output repeatedly opens and closes at an interval of
approximately 10 seconds during the preset blow back time. After the blow back time elapses,
the analog output remains held at the preset status until the hold time elapses (refer to Section
8.2, earlier in this manual).
As the hold (output stabilization) time, set the time until the sample gas is returned to the sensor
and output returns to the normal operating conditions, after completing blow back operations.
ON-time period of 1 to 11 seconds for the contact input
Blow back start
(with contact input or
touchpanel operation)
Contact output
(blow back switch)
Blow back time
ON
(for 10
seconds)
ON
(for 10
seconds)
OFF
(for 10 seconds)
Analog output
ON
(for 10
seconds)
Hold time
Analog output remains held
(with HOLD feature specified)
F10-9E.ai
Figure 10.9 Operation of Blow back
IM 11M13A01-02E
10-9
<10. Other Functions>
10.4.3 Setting Output Hold Time and Blow back Time
If the blow back mode is in “No function”, the output “Hold time” and “Blow back time” are not
displayed. If you select “Hold time”, the numeric-data entry display appears. Enter the desired
“Hold time” (output stabilization time) from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
When you select “Blow back time”, the numeric-data entry display appears. Enter the desired
“Blow back time” from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
10.4.4
Setting Interval, Start Date, and Start Time
The “Interval” is the time to execute blow back. Display the numeric-data entry panel display to
set the desired interval from 000 days, 00 hours to 255 days, 59 hours.
For the “Start date” and “Start time”, set the date when the blow back is first executed and the
time when to start the blow back, respectively. If you want to execute the first blow back, for
example, at 4:00 p.m. on March 25, 2001, enter 25/ 03/ 01 for the Start date and 16:00 for the
Start time.
Blow back setup
Mode:
Auto
Hold time: 1 0 min 0 0 s
Blow back time:
1 0 min 0 0 s
Interval:
30d 00h
Start date:
01/01/00
Start time:
00:00
Enter
F10-10E.ai
Figure 10.10 Blow back Setup Display
In the Blow back setup display shown in Figure 10.10, if you choose “Mode: No function” or
“Semi_Auto”, the Interval, Start date, and Start time for these are not displayed.
NOTE
• If the blow back is executed with a contact input, it must be preset in the Contact input
setting (for more details, see Section 8.5, earlier in this manual).
• In Section 8.4, Contact Output Setting, earlier in this manual, set the contact used as the
blow back switch beforehand.
• Do not set any other function for the contact used as the blow back switch.
Otherwise, blow back may be activated when the contact is closed by any other function.
• No blow back is executed during calibration or maintenance service. If automatic blow back
reaches the preset start time during calibration or maintenance service, blow back will be
executed after completing the calibration or maintenance service and after the equipment
returns to the measurement mode.
• If automatic blow back reaches the preset start time during semi-automatic blow back, the
current automatic blow back will not be executed.
• If you set the blow back interval at 000 days, 00 hours, only the first blow back is then
executed. No subsequent blow backs will be executed.
• If a past date is set for the Start time, no blow back will be executed.
IM 11M13A01-02E
10-10
<10. Other Functions>
10.4.5
Default Setting
When the analyzer is delivered, or if data are initialized, the blow back settings are by default, as
shown in Table 10.4.
Table 10.4 Blow back Default Setting
Item
Default setting
Mode
No function (“invalid”)
Hold time
10 minutes, 00 seconds
Blow back time
10 minutes, 00 seconds
Interval
30 days, 00 hours
Start date
01/ 01/00
Start time
0:00
10.5 Parameter Initialization
Parameter settings can be initialized to the factory default settings. Initialization can be done
for all parameters or for individual parameters. The parameters that can be initialized and their
defaults are listed in Table 10.5.
To initialize the set data, follow these steps:
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Commissioning”, next “Others” then “Defaults”. A display like Figure 10.11. appears.
(3) Select the desired item to be initialized then a display like Figure 10.12 appears.
Select “Defaults start” then initialization starts.
Defaults
Display
Calibration
Blow back
mA-outputs
Alarms
Contacts
Others
All data
Defaults
Defaults start
Cancel
Enter
F10-11E.ai
Figure 10.11 Defaults Display
Display
Enter
F10-12E.ai
Figure 10.12 Initialization Start Display
CAUTION
• Do not turn off the power during initialization. Otherwise, initialization will not be performed
properly.
IM 11M13A01-02E
10-11
<10. Other Functions>
Table 10.5 Initialization Items and Default Values
Item
Equipment
selection
Initialization Parameter
Not initialized
Detector
ZR22
Sample gas
Display item
Displayed data
Trend graph
Wet gas
1st display item
Oxygen concentration
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Parameter
Oxygen concentration
Sampling interval
30 seconds
Upper limit (graph)
25 vol%O2
Lower limit (graph)
Calibration data
Blow back
Default setting
Type of equipment
0 vol%O2
Automatic return time
0 min.
Language
Not initialized
Calibration setting
Blow back setting
mA-output 1
mA-output 2
Current output
data
Output hold setting
Mode
Manual
Calibration procedure
Span - zero
Zero gas concentration
1.00 vol%O2
Span gas concentration
21.00 vol%O2
Output hold time
10 min., 00 sec. (Semi_Auto)
Calibration time
10 min., 00 sec. (Semi_Auto)
Interval
30 days, 00 hr. (Auto)
Start date
01/01/00 (Auto)
Start time
0:00 (Auto)
Mode
No function (invalid)
(Output) hold time
10 min., 00 sec.
Blow back time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
0:00
Parameter
Oxygen concentration
Min. oxygen concentration
0 vol%O2
Max. oxygen concentration
25 vol%O2
Output damping constant
0 sec.
Output mode
Linear
Warm-up
4 mA
Set value of Warm-up
4 mA
Maintenance
Previous value held
Set value of Maintenance
4 mA
Calibration, blow back
Previous value held
Set value of Calibration, blow
back
4 mA
Error
Preset value held
Set value of Error
3.4 mA
Continue to the next page
IM 11M13A01-02E
10-12
<10. Other Functions>
Item
Initialization Parameter
Default setting
Oxygen
concentration
Parameter
Alarm setting
Alarm data
Alarm set value
Output contact 1
Hysteresis
0.1 vol%O2
Delayed action of alarm contact
3 seconds
High-high alarm
OFF
Alarm value of High-high alarm
100 vol%O2
High limit alarm
OFF
Alarm value of High limit alarm
100 vol%O2
Low limit alarm
OFF
Alarm value of Low limit alarm
0 vol%O2
Low-low alarm
OFF
Alarm value of Low-low alarm
0 vol%O2
Alarm
OFF
Other settings
Warm-up
ON
Output range now being switched
OFF
Now calibrating
OFF
Now maintenance servicing
ON
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Open
Alarm
Output contact 2
Other settings
Contacts
OFF
Warm-up
OFF
Output range switching
OFF
Now calibrating
ON
Now maintenance servicing
OFF
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Alarm
Output contact 3
Closed
High-high alarm
OFF
High limit alarm
ON
Low limit alarm
ON
Low-low alarm
OFF
Calibration coefficient (correction
ratio) alarm
OFF
Startup power stabilization timeout
OFF
Error
OFF
Other settings
OFF
Contact output
Alarm
Output contact 4
Input contact 1
Input contact 2
Closed
Error
ON
Other settings
OFF
Contact output
Closed (Fixed)
Function
OFF
Action
Closed
Continue to the next page
IM 11M13A01-02E
10-13
<10. Other Functions>
Item
Initialization Parameter
Average value/
Maximum and minimum values
Other data
Fuel set value
Default setting
Average-value calculation
One hour
Intervals over which max. and min.
values are monitored
24 hours
Amount of water vapor in exhaust gas
0.00 m3/kg (m3)
Theoretical amount of air
1.00 m3/kg (m3)
X value
1.00
Absolute humidity of atmosphere
0.0 kg/kg
Password
Deleted
10.6 Reset
Resetting enables the equipment to restart. If the equipment is reset, the power is turned off
and then back on. In practical use, the power remains on, and the equipment is restarted under
program control. Resetting will be possible in the following conditions:
(1) Error 1 – if the cell voltage is defective
(2) Error 2 – if a temperature alarm occurs
(3) Error 3 – if the A/D converter is defective
(4) Error 4 – if an EEPROM write error occurs
For details on error occurrence, consult Chapter 12, Troubleshooting, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater.
To cancel the error, reset the equipment following the steps below, or turn the power off and then
back on.
NOTE
Make sure that before resetting or restarting the power that there is no problem with the detector
or converter.
If a problem arises again after the resetting, turn the power off and troubleshoot the problem by
consulting the Troubleshooting chapter later in this manual.
To reset the equipment, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Choose “Reset”. The Reset display shown in Figure 10.13 appears.
(3) Choose “Start reset” and then press the [Enter] key to reset the equipment and the
equipment will then be in its warm-up state.
Reset
Start reset
Cancel reset
Enter
F10-13E.ai
Figure 10.13 Reset Display
IM 11M13A01-02E
<10. Other Functions>
10-14
10.7 Handling of the ZO21S Standard Gas Unit
The following describes how to flow zero and span gases using the ZO21S Standard Gas Unit.
Operate the ZO21S Standard Gas Unit, according to the procedures that follow.
10.7.1
Standard Gas Unit Component Identification
Carrying case
Flow checker
Checks the zero and
span gas flow.
Span gas valve
Controls the span gas (air) flow.
Zero gas valve regulator
Cover screws (six pcs.)
Tube connection
Pump
Supplies span gas (air)
Gas cylinder
Contains the zero gas. A gas
of 7 Nl is charged to 700 kPa
Zero gas valve
Clamp
Attaches to the gas
cylinder for use.
Clamps the gas cylinder.
Power cord
Applies the power to operate the
pump to supply the span gas.
F10-14E.ai
Figure 10.14 Standard Gas Unit Component Identification
IM 11M13A01-02E
10.7.2
<10. Other Functions>
10-15
Installing Gas Cylinders
Each ZO21S Standard Gas Unit comes with six zero gas cylinders including a spare. Each
gas cylinder contains 7-liters of gas with a 0.95 to 1.0 vol%O2 (concentration varies with each
cylinder) and nitrogen, at a pressure of 700 kPaG (at 35°C).
The operating details and handling precautions are also printed on the product. Please read them
beforehand.
To install the gas cylinder, follow these steps:
(1) Attach the zero gas valves onto the gas cylinder. First, turn the valve regulator of the zero
gas valves counterclockwise to completely retract the needle at the top from the gasket
surface. Maintaining the valve in this position, screw the valve mounting into the mouthpiece
of the gas cylinder. (If screw connection is proper, you can turn the screw manually. Do not
use any tool.) When the gasket comes in contact with the mouthpiece of the gas cylinder
and you can no longer turn it manually, tighten the lock nut with a wrench.
(2) Remove the carrying case from the standard gas unit. The case is attached to the unit with
six screws. So, loosen the screws and lift them off.
(3) Slide the gas cylinder through the hole in the back of the unit and connect the tube (the
piping in the unit) to the valve connections. Insert each tube at least 10 mm to prevent
leakage, and secure it using a tube clamp.
(4) Attach the gas cylinder to the case. Extend the valve regulator of the zero gas valves
through the hole in the front panel of the unit and secure the bottom of the cylinder with the
clamp.
(5) Take note of the oxygen concentration of the sealed gas indicated on the gas cylinder and
replace the carrying case. Enter the oxygen concentration of the sealed gas, following
instructions on the converter display. Also check that no piping is disconnected.
Thus, the work of installing a gas cylinder is completed. However, gases in the cylinders cannot
immediately flow out after these procedures. To discharge the gases, it is necessary for the
needle in the zero gas valves to puncture a hole in the gas cylinder.
For this operation, see Subsection 10.7.3.
10.7.3
Calibration Gas Flow
(1) To operate the standard gas unit, place it on a nearly horizontal surface in order to allow the
flow check to indicate the precise flow rate. In addition, a power supply for driving the span
gas (air) supply pump is required near the unit (the length of the power cord attached to the
unit is 2 m). Select a suitable location for the unit near the installation site of the converter.
(2) Connect the tube connector port of the standard gas unit to the calibration gas inlet of the
detector, using a polyethylene resin tube with an outside diameter of 6 mm. Be careful to
prevent gas leakage.
(3) Fully open the stop valve mounted on the calibration gas inlet of the detector.
(4) Enter the oxygen concentration of the sealed gas (noted from the cylinder) into the
converter. Also check that the oxygen concentration of the span gas is correctly set
(21 vol%O2 for clean air). When using the ZO21S Standard Gas Unit (for use of the
atmospheric air as a span gas), use a hand-held oxygen analyzer to measure the actual
oxygen concentration, and then enter it.
IM 11M13A01-02E
10-16
<10. Other Functions>
The standard gas unit is used only when manual calibration is employed. Therefore, the timing
for flowing span gas (air) is included in the manual calibration procedure described in Subsection
10.7.2. For operation of the converter, see Section 7.12, earlier in this manual.
(1) When the message “Open span gas valve.” is displayed on the converter display during
calibration, plug the power cord into the power supply socket to start the pump of the
standard gas unit.
(2) Next, adjust the flow rate to 600 ± 60 ml/min using the span gas valve “AIR” (the flow check
ball stops floating on the green line when the valve is slowly opened). To rotate the valve
shaft, loosen the lock nut and turn it using a flat-blade screwdriver. Turning the valve shaft
counterclockwise increases the flow rate.
(3) After adjusting the flow rate, tighten the valve lock nut.
(4) Select “Valve opened” (to start calibration) in the Manual calibration display shown in
Figure 10.15. Check the Trend graph display to see that the measured value is stabilized.
Then press the [Enter] key. The Manual calibration display shown in Figure 10.16 appears.
Disconnect the power cord to stop the pump.
Manual calibration
Manual calibration
Open span gas valve.
Span calibration
Set flow span gas to
600ml/min.
Close the span gas valve.
Valve opened
Cancel calibration
Zero calibration
End
Enter
Enter
F10-15E.ai
Figure 10.15 Manual Calibration Display
F10-16E.ai
Figure 10.16 Manual Calibration Display
Cause a zero gas to flow according to the Manual calibration display shown in Figure 10.17.
(1) Use the needle of the zero gas valve “ CHECK GAS “ to puncture a hole in the gas cylinder
installed as described in Subsection 10.7.2. Fully clockwise turn the valve regulator by hand.
(2) Next, adjust the flow rate to 600 ± 60 ml/min (the flow check ball stops floating on the green
line when the valve is slowly opened). Turn the regulator of the zero gas valves back slowly
counterclockwise. At that time, the flow rate also decreases as the inner pressure of the gas
cylinder decreases. Therefore, monitor the flow check and, when the ball’s position changes
greatly, readjust the valve.
(3) Select “Valve opened” (to start calibration) from the Manual calibration display. Check the
Trend graph display to see that the measured value is stabilized. Then press the [Enter] key.
The Manual calibration display shown in Figure 10.18 appears. Then stop the zero gas flow
immediately. Turn the zero gas valve regulator fully clockwise. If this valve regulator is not
properly adjusted, the needle valve will not close completely and a cylinder gas may leak.
IM 11M13A01-02E
10-17
<10. Other Functions>
Manual calibration
Manual calibration
Open zero gas valve.
Zero calibration
Set flow zero gas to
600ml/min.
Close the zero gas valve.
Valve opened
Span calibration
Cancel calibration
End
Enter
Enter
F10-17E.ai
Figure 10.17 Manual Calibration Display
F10-18E.ai
Figure 10.18 Zero gas Calibration Complete
(in Manual Calibration Display)
NOTE
Be sure not to terminate a calibration in progress because of a shortage of gas in the cylinder.
Each gas cylinder is operable for nine minutes or more provided the gas is discharged at the
specified rate. Therefore, if your calibration time is estimated at four minutes, you can operate
the zero-point calibration twice.
(1) Fully close the stop valve mounted on the calibration gas inlet of the detector.
(2) Remove the tube connecting the detector to the standard gas unit.
WARNING
Store the standard gas unit with the gas cylinder mounted where the ambient temperature does
not exceed 40°C. Otherwise, the gas cylinder may explode. Store the spare gas cylinders under
the same condition.
IM 11M13A01-02E
<10. Other Functions>
10-18
10.8 Methods of Operating Valves in the ZA8F
Flow Setting Unit
The ZA8F Flow Setting Unit is used as a calibration device for a system conforming to System 1.
Calibration in such a system is to be manually operated. So, you have to operate the valve of the
Flow Setting Unit each time calibration is made (starting and stopping the calibration gas flow and
adjusting the flow rate). This applies even if you are using the ZR40H Automatic Calibration Unit.
For operation of the converter, see Section 7.12, earlier in this manual.
10.8.1
Preparation Before Calibration
To operate the ZA8F Flow Setting Unit, prepare for calibration as follows:
(1) Check for a complete closing of the zero gas flow setting valve in the unit and open the
reducing valve for the zero gas cylinder until the secondary pressure equals sample gas
pressure plus approx. 50 kPa (or sample gas pressure plus approx. 150 kPa when a check
valve is used, maximum pressure rating is 300 kPa).
(2) Check that the oxygen concentration of the zero gas and span gas (instrument air 21
vol%O2) in the cylinder is set in the converter.
10.8.2
Operating the Span Gas Flow Setting Valve
The following description is given assuming that instrument air, the same as the reference gas, is
used as the span gas.
(1) When the display shown in Figure 10.15 appears during calibration, open the span gas
flow setting valve of the flow setting unit and adjust the flow rate to 600 ± 60 ml/min. Turn
the valve slowly counterclockwise after loosening the lock nut if the valve has a lock nut. To
check the flow rate, use the calibration flowmeter. If the sample gas pressure is extremely
high, adjust the sample gas pressure to obtain pressures (listed in Table 10.6) ±10%.
Table 10.6
Sample gas pressure (kPa)
50
100
150
200
250
Flow rate (mL/min)
500
430
380
350
320
(2) Adjust the flow rate and select “Valve opened” in the Manual calibration display.
Check the Trend graph display to see that the measured value is stabilized. Then press the
[Enter] key. The Manual calibration display shown in Figure 10.16 appears.
(3) Close the span gas flow setting valve to stop the span gas (air) flow. If the valve has a lock
nut, be sure to tighten the lock nut to prevent any leakage of the span gas into the sensor
during measurement.
10.8.3
Operating the Zero Gas Flow Setting Valve
Operate the zero gas flow setting valve during zero-point calibration in the following procedures:
(1) When the display shown in Figure 10.19 appears during calibration, open the zero gas
flow setting valve of the flow setting unit and adjust the flow rate to 600 ± 60 ml/ min. To
rotate the valve shaft, if the valve has a lock nut loosen the lock nut and slowly turn it
counterclockwise. To check the flow rate, monitor the calibration gas flowmeter.
If the sample gas pressure is extremely high, adjust the sample gas pressure to obtain
pressures (listed in Table 10.6) ±10%.
IM 11M13A01-02E
<10. Other Functions>
10-19
Manual calibration
Open zero gas valve.
Set flow zero gas to
600ml/min.
Valve opened
Cancel calibration
Enter
F10-19E.ai
Figure 10.19 Manual Calibration Display
(2) Adjust the flow rate and select “Valve opened” in the Manual calibration display.
Check the Trend graph display to see that the measured value is stabilized. Then press the
[Enter] key. The Manual calibration display shown in Figure 10.20 appears.
Manual calibration
Zero calibration
Close the zero gas valve.
Span calibration
End
Enter
F10-20E.ai
Figure 10.20 Zero-point Calibration Complete (in Manual Calibration)
(3) Close the zero gas flow setting valve to stop the zero gas flow. If the valve has a lock nut,
be sure to tighten the lock nut to prevent the any leakage of the zero gas into the sensor
because the valve may become loose during measurement.
10.8.4
Operation After Calibration
No special operation of the instrument is needed after calibration. However, it is recommended
that the pressure reducing valve for the zero gas cylinders be closed because calibration is not
required so often.
IM 11M13A01-02E
Blank Page
11-1
<11. Inspection and Maintenance>
11. Inspection and Maintenance
This chapter describes the inspection and maintenance procedures for the EXAxt ZR Explosionproof Zirconia Oxygen Analyzer to maintain its measuring performance and normal operating
conditions.
WARNING
When checking the detector, carefully observe the following:
• The instrument modification or parts replacement by other than authorized representation
of Yokogawa Electric Corporation is prohibited and will void ATEX flameproof Certification,
Factory Mutual Explosion-proof approval and Canadian Standards Explosion-proof
Certification.
• Do NOT touch the probe if it has been in operation immediately just before being checked.
(The sensor at the tip of the probe heats up to 750°C during operation. If you touch it, you
will get burned.)
CAUTION
• Do not subject the probe to shock or cool it rapidly.
The sensor is made of ceramic (zirconia). If the detector is dropped or bumped into
something, the sensor may be damaged and no longer work.
• Do not reuse a metal O-ring to seal the cell assembly. If you replace the cell or remove it
from the probe for checking, be sure to replace the metal O-ring. Otherwise, the furnace gas
may leak, and then the leaking corrosive gas will cause the built-in heater or thermocouple
to go open circuit, or the detector may corrode.
• Before opening or closing the terminal box, first remove dust, sand, or the like from the
terminal box cover.
IM 11M13A01-02E
11-2
<11. Inspection and Maintenance>
11.1 Inspection and Maintenance of the Detector
11.1.1
Cleaning the Calibration Gas Tube
The calibration gas, supplied through the calibration gas inlet of the terminal box into the detector,
flows through the tube and comes out at the tip of the probe. The tube might become clogged
with dust from the sample gas. If you become aware of clogging, such as when a higher pressure
is required to achieve a specified flow rate (600 ± 60 ml/min), clean the calibration gas tube.
To clean the tube, follow these steps:
(1) Remove the detector from the installation assembly. Remove the Flame Arrestor Assembly
from Sensor Assembly (See Figure 11.2)
(2) Following Subsection 11.1.2, later in this manual, remove the four bolts (and associated
spring washers) that tighten the sensor assembly, and the pipe support as well as the
U-shaped pipe.
(3) Use a rod 2 to 2.5 mm in diameter to clean the calibration gas tube inside the probe.
In doing this, keep air flowing from the calibration gas inlet at about 600 ml/min and insert
the rod into the tube (3-mm inside diameter). However, be careful not to insert the rod
deeper than 40 cm for a general-purpose detector, or 15 cm for a high temperature detector.
(4) Clean the U-shaped pipe. The pipe can be rinsed with water. However, it should be dried out
thoroughly before reassembly.
(5) Restore all components you removed for cleaning. Follow Subsection 11.1.2 to restore all
components in their original positions. Be sure to replace the O-ring(s) with new ones.
Exploded view of components
Rod
(with outside diameter
of 2 to 2.5 mm)
Calibration gas tube
F11-1E.ai
Figure 11.1 Cleaning the Calibration Gas Tube
11.1.2
Replacing the Sensor Assembly
The performance of the sensor (cell) deteriorates as its surface becomes soiled during operation.
Therefore, you have to replace the sensor when its life expectancy expires, for example, when
it can no longer satisfy a zero correction ratio of 100±30% or a span correction ratio of 0±18%.
In addition, the sensor assembly is to be replaced if it becomes damaged and can no longer
operate during measurement.
If the sensor becomes no longer operable (for example, due to breakage), investigate the cause
and remedy the problem as much as possible to prevent recurrence.
IM 11M13A01-02E
11-3
<11. Inspection and Maintenance>
CAUTION
• If the sensor assembly is to be replaced, allow enough time for the detector to cool down
from its high temperature. Otherwise, you may get burned.
• If the cell assembly is to be replaced, be sure to replace the metal O-ring and the contact
together. Additionally, even in a case where the cell is not replaced, if the contact becomes
deformed and cannot make complete contact with the cell, replace the contact.
• If there is any corroded or discolored area in the metal O-ring groove in which the contact
is embedded, sand the groove with sandpaper or use a metal brush, and then sand further
with a higher grade of sandpaper (no. 1500 or so), or use an appropriate metal brush
to eliminate any sharp protrusions on the groove. The contact’s resistance should be
minimized.
• Use sensor assemblies manufactured in or after Sept. 2000: the serial number on the side
of the sensor assembly should be 0J000 or later (for example: 0K123, 1AA01 etc)
Sensor
Flame arrestor
assembly
Bolts (four)
Metal O-ring
U-shaped pipe
support
Probe
Contact
Filter
U-shaped pipe
Washers
(four)
1/8 turn – tighten bolts 1/8 turn
(approximately 45°) each
F11-2E.ai
Figure 11.2 Exploded View of Sensor Assembly
CAUTION
Optional Inconel bolts have a high coefficient of expansion. If excess torque is applied while
the bolts are being tightened, abnormal strain or bolt breakage may result. So, tighten the bolts
following the instructions given above.
IM 11M13A01-02E
<11. Inspection and Maintenance>
11-4
1. Identifying parts to be replaced
In order not to lose or damage disassembled parts, identify the parts to be replaced from
among all the parts in the sensor assembly. Normally, replace the sensor, metal O-ring and
contact together at the same time. If required, also replace the U-shaped pipe, bolts, filter and
associated spring washers.
2. Removal procedures
(1) Remove the flame arrestor assembly with a special wrench (part no. K9471UX).
(2) Remove the four bolts and associated washers from the tip of the detector probe.
(3) Remove the U-shaped pipe support together with the U-shaped pipe. Remove filter also.
(4) Pull the sensor assembly toward you while turning it clockwise. Also, remove the metal
O-ring between the assembly and the probe. Remove filter also.
(When replacing the assembly, be careful not to scratch or dent the tip of the probe with
which the metal O-ring comes in contact (the surface with which the sensor flange also
comes in contact). Otherwise, the sample gas will not be sealed.)
(5) Use tweezers to pull the contact out of the groove in the tip of the probe.
(6) Clean the sensor assembly, especially the metal O-ring contact surface to remove any
contaminants adhering to that part. If you can use any of the parts from among those
removed, also clean them up to remove any contaminants adhering to them.
(Once the metal O-ring has been used, it cannot be reused. So, be sure to replace it.)
3. Part assembly procedure
(1) First, install the contact. Being careful not to cause irregularities in the pitch of the coil spirals
(i.e., not to bend the coil out of shape), place it in the ringed groove properly so that it forms
a solid contact.
Groove in which the contact (E7042BS) is placed
F11-3E.ai
Figure 11.3 Installing the Contact
(2) Next, make sure that the O-ring groove on the flange surface of the sensor is clean. Install
the metal O-ring in that O-ring groove, and then insert the sensor in the probe while turning
it clockwise. After inserting it until the metal O-ring comes in contact with the probe’s O-ring
contact surface, properly align the U-shaped-pipe insertion holes with the bolt openings.
(3) Attach the U-shaped pipe to its support with filter, then fully insert the U-shaped pipe and its
support into the probe.
(4) Coat the threads of the four bolts with anti-seize grease and then screw them in along with
the washers. First, tighten the four bolts uniformly by hand, and then use a torque wrench
to tighten all areas of the metal O-ring uniformly, that is, to make sure the sensor flange is
perfectly horizontal to the O-ring’s working face in the probe.
This is done by tightening first one bolt and then its opposing bolt each 1/8 turn, and then
one of the other bolts followed by its opposing bolt, each also 1/8 turn. This continues
in rotating fashion until they are all fully tightened with the torque wrench preset to
approximately 5.9 N • m. If they are not uniformly tightened, the sensor or heater may be
damaged.
Replacement of the sensor assembly is now complete. Attach and fix the flame arrestor
assembly. Install the detector and restart operation. Calibrate the instrument before making
a measurement.
IM 11M13A01-02E
11.1.3
<11. Inspection and Maintenance>
11-5
Replacement of the Heater Unit (Assembly)
This Subsection describes the replacement procedure for the heater unit.
The sensor or ceramic heater-furnace core internal structure is subject to fracturing, so do NOT
subject it to strong vibrations or shock. Additionally, the heater unit reaches high temperatures
and is subjected to high voltages. So, maintenance services should be performed after the power
is off and the heater unit temperature has returned to normal room temperature.
For details, refer to IM11M12A01-21E “ Heater Assembly ”.
NOTE
If the heater strut assembly cannot be removed because a screw has fused to its thread, one of
our service representatives can fix it.
IM 11M13A01-02E
11-6
<11. Inspection and Maintenance>
14
A
13
11
10
12
A
22
8
9
22
7
4
5
3
6
22
2
1
21
TC +
View A - A
15
23
TC -
CELL +
26
24
16
27
25
1
17
12
18
19
CELL
2 3
TC
4 5
CJ
6
28
31
29
30
20
13
32
7
H T R
8
F11-4E.ai
Figure 11.4 Exploded View of Detector
IM 11M13A01-02E
<11. Inspection and Maintenance>
11-7
Replacement of heater strut assembly
Refer to Figure 11.4 as an aid in the following discussion.
Remove the U-shaped pipe support 4 , the U-shaped pipe 5 , the Filter and the Sensor (Cell)
6 according to Subsection 11.1.2 Replacing the Sensor Assembly. Open the terminal box 14
and remove the three terminal connections – CELL +, TC + and TC -. Before disconnect the
HTR terminals, remove the terminal block screw 26 . Keeping the other terminal remaining to be
connected. Disconnect the two HTR connections.(These terminals have no polarity.)
Remove the four bolts 10 and terminal box
not get caught in the terminal box.
14
with care so that the already disconnected wire will
Loosen Screw 17 until the plate 15 of heater strut assembly 21 can be removed. There’s no need
to remove O-ring 16 which prevents Screw 17 from coming out. Pull out connector 12 .
Loosen and remove the screw for the heater assembly fixation 8 with a special wrench (part no.
K9470BX or equivalent) and then remove the heater strut assembly 21 from the detector 22 .
To reassemble the heater strut assembly, reverse the above procedure:
Insert the heater strut assembly 21 into the detector 22 , while inserting the calibration pipe in the
detector 22 into the heater section in the heater strut assembly 21 as well as in the bracket hole.
Coat the screw for the heater assembly fixation 8 with grease (NEVER-SEEZ: G7067ZA) and
tighten the screw for the heater assembly fixation 8 with a special wrench (part no. K9470BX or
equivalent) with a tightening torque of 12N·m ±10 percent.
Next, to install the O-rings 20 on the calibration gas and reference gas pipes, disassemble the
connector 12 in the following procedure:
First, remove the screw 23 and then remove the plate 15 and two caps 18 . If the O-ring 20
remains in the hole, pull them out from the back. Pass the heater and thermocouple lead wire
through the connector 12 . Also, pass the calibration gas and reference gas pipes through the
opening of the connector 12 . If the O-ring 20 fails, replace it with a new one.
Push the two caps 18 into the associated opening of the connector 12 . Insert the plate 15 ,
aligning it with the groove of the cap 18 , and tighten it with the screw 23 . If you attempt to insert
the calibration gas and reference gas pipes into the connector 12 without disassembling the
connector 12 , the O-ring may be damaged. Tighten Screw 17 to the plate 15 of heater strut
assembly 21 until connector 12 can’t move.
Reassemble in reverse order to the above disassembly procedure.
The two wires with ceramic insulators from the heater strut assembly are heater wires, and the
single-core shielded wire is the cell signal + terminal; for the two-core shielded cable, the semitranslucent rubber-sheathed wire is the thermocouple + terminal, and the other wire is the −
terminal. (If the wires are marked, match the markings with those on the terminal board).
When installing the sensor (cell)
6
, replace the metal O-ring
7
with a new one.
IM 11M13A01-02E
11-8
<11. Inspection and Maintenance>
11.1.4
Replacement of Flame Arrestor Assembly
If it takes longer for the analyzer to return to read the concentration of a sample gas after
calibration, the flame arrestor may have become clogged. Inspect the flame arrestor and if
necessary, clean or replace it.
Set the flame arrestor assembly 1 in place using a special pin spanner (with a pin 4.5 mm in
diameter: part no. K9471UX or equivalent). If a flame arrestor assembly that has already been
replaced once is used again, apply grease (NEVER-SEEZ: G7067ZA) to the threads of the flame
arrestor assembly.
If the flame arrestor assembly is clogged with dust, replace it with new one or wash it.
In case of the ATEX flameproof model (MS Code: ZR22S-A-...) or IECEx flameproof model (MS
Code: ZR22S-D-...), the flame arrestor assembly 1 is bonded to the detector 22 with an ceramic
adhesive. To remove the flame arrestor assembly 1 , crack the hardened adhesive on the joint
by tapping it with a flat blade screwdriver and a hammer or appropriate tools. After reattaching the
flame arrestor assembly 1 , apply a small amount of ceramic adhesive (part no. G7018ZA), with
a diameter not exceeding 10 mm, to the joint part. Be careful not to allow the ceramic adhesive
to enter between the female and the male screws. Before applying, stir the ceramic adhesive
thoroughly. The ceramic adhesive should be stored in a cool, dark place and has a shelf life of 6
months from the date of shipment.
Apply adhesive here
Flame Arrestor Assembly
Detector
Detector
Flame Arrestor Assembly
F11-5E.ai
Figure 11.5 Removal of Flame Arrestor Assembly
11.1.5
Replacement of O-ring
The detector uses three different types of O-rings
Part No.
13 , 19 , and 20 , two O-rings of each type.
Description
7
K9470BJ
Metal O-ring
13
K9470ZS
O-ring with grease
19 20
K9470ZP
Two pairs of O-rings with grease
IM 11M13A01-02E
11-9
<11. Inspection and Maintenance>
11.1.6
Cleaning the High Temperature Probe Adapter
CAUTION
• Do NOT subject the probe of the High Temperature Probe Adapter (ZO21P) to shock.
This probe uses silicon carbide (SiC) which may become damaged if it is subjected to a
strong shock or thermal shock.
The high temperature detector is structured so that the gas to be measured is directed toward
the detector with the high temperature probe adapter. Therefore, if the probe or the sample gas
outlet clogs, a precise measurement is no longer possible because of no gas flow. If you use the
high temperature detector, you have to inspect it periodically and, if any part of it is significantly
clogged with dust, clean it.
Dust found sticking to the probe should be blown off. If any dust still remains after the blowing,
clean it with a metal rod, etc., inserted. In addition, if dust is found on the auxiliary ejector
assembly or needle valve (choke) at the sample gas outlet, remove these parts from the high
temperature probe adapter and then clean them. To remove dust, blow air on them or rinse them
with water.
11.1.7
Stopping and Re-starting Operation
When operation is stopped, take care of the followings so that the sensor of the detector cannot
become unused.
CAUTION
When operating an instrument such as boiler or industrial furnace is stopped with the zirconia
oxygen analyzer operation, moisture can condensate on the sensor portion and dusts may stick
to it.
If operation is restarted in this condition, the sensor which is heated up to 750°C may become
permanently contaminated. The dust can greatly degrade sensor performance. If a large amount
of water is condensed, the sensor can be broken and never re-useful.
To prevent the above nonconformity, take the following action when stopping operation.
(1) If possible, keep on supplying the power to converter and flowing reference gas to the
sensor.
If impossible to do the above, remove the detector.
(2) If impossible to supply the power or remove the detector, at least keep on flowing air at 600
mL/min into the calibration gas pipe.
When restarting operation, be sure to flow air, for 5-10 minutes, at 600 mL/min into the calibration
gas pipe before supplying the power to converter.
IM 11M13A01-02E
<11. Inspection and Maintenance>
11-10
11.2 Inspection and Maintenance of the
Converter
The converter does not require routine inspection and maintenance. If the converter does not
work properly, in most cases it probably comes from problems or other causes.
A dirty touchpanel should be wiped off with a soft dry cloth.
11.2.1
Replacing Fuses
The converter incorporates a fuse, as indicated in Figure 11.6. If the fuse blows out, replace it in
the following procedure.
NOTE
• If a replaced fuse blows out immediately, there may be a problem in the circuit.
Go over the circuit completely to find out why the fuse has blown.
• This fuse is for protecting the main power supply circuit and does not provide overcurrent
protection for the heater temperature control circuit. For overcurrent protection circuitry,
refer to Subsection 12.1.2, Error 2: Heater Temperature Failure.
Fuse
F11-6E.ai
Figure 11.6 Location of Fuse in the Converter
IM 11M13A01-02E
11-11
<11. Inspection and Maintenance>
To replace the fuse, follow these steps:
(1) Turn off the power to the converter for safe replacement.
(2) Remove the fuse from its holder. With the appropriate flat-blade screwdriver that just fits the
holder cap slot (Figure 11.7), turn the fuse holder cap 90° counterclockwise.
By doing so, you can remove the fuse together with the cap.
(3) Check the rating of the fuse and that it satisfies the following:
Maximum rated voltage: 250 V
Maximum rated current: 3.15 A
Type: Time-lag fuse
Standards: UL-, CSA- or VDE-approved
Part number: A1113EF
Place a new, properly rated fuse in the holder together with the cap, and push and turn the cap
clockwise 90° with the screwdriver to complete installation of the fuse.
Cap
Flat-blade
screwdriver
Socket
Fuse
F11-7E.ai
Figure 11.7 Removing the Fuse
11.2.2
Cleaning
Use a soft dry cloth to clean any part of the converter during inspection and maintenance.
11.2.3
Adjust LCD screen contrast
An LCD is built in the ZR402G converter. The contrast of this LCD is affected by its ambient
temperature. For this reason, the LCD is shipped, after adjusting the contrast so as to become
the most suitable in a room temperature(20-30°C). However, when display on the LCD is hard
to see, adjust the LCD contrast by change the resistance of the variable resistor; its position is
shown in Figure 11.8
Variable resistor
F11-8E.ai
Figure 11.8
IM 11M13A01-02E
<11. Inspection and Maintenance>
11-12
11.3 Replacing Flowmeter in ZR40H Automatic
Calibration Unit
(1) Remove piping and wiring, and remove the ZR40H from the 2B pipe or wall mounting.
(2) Remove four M6 bolts between brackets and remove piping extension
(3) Remove bolts holding flowmeter, and replace it. A white back plate (to make the float easy
to see) is attached. The end of the pin holding down the back plate must be on the bracket
side.
(4) Replace piping, and fix M6 bolts between brackets. *1
*1 : When disassembling and reassembling, mark original positions, and tighten an extra 5-10° when reassembling.
After tightening, do a liquid leakage test.
Flowmeter with back plate
Bolt fixes bracket
COM ON S N
Z RO
Connecter
F11-9E.ai
Figure 11.9 Flowmeter replacement
C
D
Two screws fix flowmeter
Connect piping pairs A-A', B-B', C-C', D-D'
A'
A
C'
B'
D'
B
F11-10E.ai
Figure 11.10 Fixing Flowmeter
IM 11M13A01-02E
12-1
<12. Troubleshooting>
12. Troubleshooting
This chapter describes errors and alarms detected by the self-diagnostic function of the
converter. It also explains inspections and remedies when other problems occur.
12.1 Displays and Remedies When Errors Occur
12.1.1
Error Types
An error occurs when an abnormality is detected in the detector or the converter, e.g., in the cell
(sensor), detector heater, or internal circuits of the converter.
If an error occurs, the converter performs the following:
(1) Stops the supply of power to the heater in the detector to insure system safety.
(2) Causes an error indication in the display to start blinking to notify of an error generation
(Figure 12.1).
(3) Sends a contact output if the error is set up for Contact output setting for that contact
(refer to Section 8.4, Contact Output Setting).
(4) Changes the analog output status to the one set in Output hold setting.
(refer to Section 8.2, Output Hold Setting).
When the display shown in Figure 12.1 appears, pressing the error indication brings up a
description of the error (Figure 12.2). The content of errors that are displayed include those
shown in Table 12.1.
Tag:
Tag:
21.0
Error 2 :Heater temp.
Alarm 1 1 : T.C. voltage
%O2
17.42mA -Output1
17.42mA -Output2
Hold
Hold
F12-1E.ai
Figure 12.1
F12-2E.ai
Figure 12.2
Table 12.1 Types of Errors and Reasons for Occurrence
Error No.
Error Type
Occurrence Conditions
Error 1
Cell voltage failure
The cell (sensor) voltage signal input to the converter falls below -50 mV.
Error 2
Heater temperature failure
The heater temperature does not rise during warm-up, or it falls below
730 ºC or exceeds 780°C after warm-up is completed.
Or this occurs if the TC+, TC- thermocouple terminals wired to detector
with reverse (wrong) polarity.
Error 3
A/D converter failure
The A/D converter fails in the internal electrical circuit in the converter.
Error 4
Memory failure
Data properly are not written into memory in the internal electrical circuit
in the converter.
IM 11M13A01-02E
12-2
<12. Troubleshooting>
12.1.2
Remedies When an Error Occurs
Error 1: Cell Voltage Failure
Error 1 occurs when the cell (sensor) voltage input to the converter falls below -50 mV
(corresponding to about 200 vol%O2). The following are considered to be the causes for the cell
voltage falling below -50 mV:
(1) Poor contact in terminal connections between the converter and detector.
(2) Breakage in wiring cable between the converter and the detector.
(3) Damage or deterioration of the sensor assembly.
(4) Continuity failure between the sensor assembly electrode and the contact.
(5) Wiring failure inside the detector.
(6) Abnormality in electrical circuits inside the converter.
Turn off power to the converter.
Is there any breakage
or poor contact in the wiring to the
converter terminals?
Yes
Replace the damaged
portions.
Examine the wiring connection to converter terminals
"CELL+" and "CELL-". Also, check the wiring
connection in the repeater terminal box if it is used.
No
Is there any breakage
or poor contact in the wiring to the
detector terminals?
Yes
Replace the damaged
portions.
Examine the wiring connection status to detector
terminals 1 and 2. Also, check that terminals and
cable conductors are not corroded.
No
Is the sensor extremely
dirty, corroded or broken?
Yes
Replace the sensor
assembly.
See Subsection 11.1.2
No
Is there any
disconnection or poor continuity
in the wiring between the
detector and
converter?
Yes
Replace the wiring cable.
No
End. Carry out calibration.
Remove the wiring conductors from detector
terminals 1 and 2 and short out the removed
conductors. Measure the resistance of these
wiring conductors on the converter side.
The resistance value is normal if it indicates
10 Ω or less.
Turn on the power to the converter and temporarily
place the analyzer in the operating status.
No
Is an error indicated
on the display?
Remove the sensor assembly from the detector
and check for the presence of corrosion that may
cause a poor contact between the electrode and
the contact, also, check for dirt. A sensor assembly
in which no abnormality is found may be used again.
However, be sure to use a new metal O-ring and
contact even in such a case.
Yes
Replace the sensor
assembly and temporarily
place the analyzer in the
operating status.
See Subsection 11.1.2
The analyzer operates normally.
No
Yes
A failure in the detector or
the converter is suspected.
Contact Yokogawa.
F12E.ai
IM 11M13A01-02E
<12. Troubleshooting>
12-3
Error 2: Heater Temperature Failure
This error occurs if the detector heater temperature does not rise during warm-up, or if the
temperature falls below 730°C, or exceeds 780°C after warm-up is completed. In addition, when
Error 2 occurs, Alarm 10 (cold junction temperature alarm) or Alarm 11 (thermocouple voltage
alarm) may be generated at the same time. Be sure to press the error indication to get the error
description and confirm whether or not these alarms are being generated simultaneously.
If Alarm 10 is generated simultaneously, a failure in the cold junction system on the detector
terminal block is suspected. In this case, follow the procedure according to troubleshooting for
Alarm 10 in Subsection 12.2.2.
If Alarm 11 is generated simultaneously, a failure in the thermocouple system located in the
detector heater is suspected. In this case, follow the procedure according to troubleshooting for
Alarm 11 in Subsection 12.2.2.
If this failure occurs immediately after the power is supplied, the polarity at thermocouple input
connection (TC+, TC-) on the converter may be reversed. Check the connection from the
detector.
Causes considered for cases where Error 2 occurs independently are shown below.
(1) Faulty heater in the detector (heater wire breakage).
(2) Faulty thermocouple in the detector.
(3) Faulty cold junction sensor located at the detector terminal block.
(4) Failure in electrical circuits inside the converter.
(5) Heater temperature control overcurrent limiting triggered.
(6) TC+ ,TC- thermocouple terminals wired to detector with reverse (wrong) polarity.
Overcurrent protection is triggered if there are problems in the heater wiring. When the protective
circuit is triggered, the internal fuse blows and the heater is disconnected, resulting in Error 2
(heater temperature failure).
(1) Turn off power to the converter.
(2) Remove the cable from terminals 7 and 8 of the detector and measure the resistance
value between these terminals. The heater unit is normal if the resistance is lower than
about 90 Ω. If the resistance value is higher, failure of the heater unit is suspected. In
this case, replace the heater unit (refer to Subsection 11.1.3, Replacement of the Heater
Unit (Assembly)). In addition, check that the wiring resistance between the converter and
detector is 10 Ω or less.
(3) Ensure that TC+ terminal (terminal 3 in detector) is connected to converter TC+ terminal,
and TC- terminal (terminal 4) is connected to converter TC- terminal.
(4) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance value
between these terminals. The thermocouple is considered normal if the resistance value is
5 Ω or less. If the value is higher than 5 Ω, it may indicate that the thermocouple wire has
broken or is about to break. In this case, replace the heater unit (refer to Subsection 11.1.3,
Replacement of the Heater Unit (Assembly)). Also, check that the wiring resistance between
the converter and detector is 10 Ω or less.
(5) Even if items (2) to (4) are normal, the heater overcurrent protection fuse may have blown.
Check for wiring problems such as the following:
1) Heater terminals shorted.
2) Heater terminal(s) shorted to ground.
3) Heater terminals shorted to power supply.
If the internal fuse blows, this cannot be replaced by the user. Contact your Yokogawa service
representative.
IM 11M13A01-02E
12-4
<12. Troubleshooting>
NOTE
• Measure the thermocouple resistance value after the temperature difference between the
detector tip and the ambient atmosphere has decreased to 50°C or less. If the thermocouple
voltage is large, accurate measurement cannot be achieved.
Error 3: A/D Converter Failure/Error 4: Writing-to-memory Failure
• A/D Converter Failure
It is suspected that a failure has occurred in the A/D converter mounted in the electrical circuits
inside the converter.
• Writing-to-memory Failure
It is suspected that a failure has occurred in an operation writing to the memory (EEPROM)
mounted in the electrical circuits inside the converter.
Turn off the power to the converter once and then restart the converter. If the converter
operates normally after restarting, an error might have occurred due to a temporary drop in the
voltage (falling below 85 V, the least amount of voltage required to operate the converter) or a
malfunction of the electrical circuits affected by noise. Check whether or not there is a failure in
the power supply system or whether the converter and detector are securely grounded.
If the error occurs again after restarting, a failure in the electrical circuits is suspected.
Consult the service personnel at Yokogawa Electric Corporation.
IM 11M13A01-02E
12-5
<12. Troubleshooting>
12.2 Displays and Remedies When Alarms are
Generated
12.2.1
Alarm Types
When an alarm is generated, the alarm indication blinks in the display to notify of the alarm
(Figure 12.3). Pressing the alarm indication displays a description of the alarm.
Alarms include those shown in Table 12.2.
Tag:
Tag:
21.0
Alarm 6 :Zero conc.ratio
%O2
17.42mA -Output1
17.42mA -Output2
F12-3E.ai
Figure 12.3
F12-4E.ai
Figure 12.4
Table 12.2 Types of Alarms and Reasons for Occurrence
Alarm No.
Alarm Type
Occurrence Conditions
Alarm 1
Oxygen concentration
alarm
Measured oxygen concentration value exceeds or falls below the
preset alarm limits. (refer to Section 8.3, Oxygen Concentration Alarms
Setting).
Alarm 6
Zero calibration coefficient
(correction ratio) alarm
In automatic or semi-automatic calibration, zero correction ratio
is outside the range of 100±30%. (refer to Subsection 9.1.3,
Compensation).
Alarm 7
Span calibration coefficient
(correction ratio) alarm
Span correction ratio is outside the range of 0±18% (refer to Subsection
9.1.3, Compensation).
Alarm 8
EMF stabilization time-up
alarm
In automatic or semi-automatic calibration, cell voltage does not stabilize
after the calibration time is up.
Alarm 10
Cold junction temperature
alarm
Temperature of the cold junction placed in the detector terminal box falls
below -25°C or exceeds 155°C.
Alarm 11
Thermocouple voltage
alarm
Generated when thermocouple voltage exceeds 42.1 mV (about
1020°C) or falls below -5 mV (about -170°C).
Alarm 13
Battery low alarm
Internal battery needs replacement.
If an alarm is generated, actions such as turning off the heater power are not carried out. The
alarm is cancelled when the cause of the alarm is eliminated. However, Alarm 10 and/or Alarm
11 may be generated concurrently with Error 2 (heater temperature failure). In this case, the
operation when the error occurs has priority.
If the power to the converter is turned off after an alarm is generated and the converter is restarted
before the cause of the alarm has been eliminated, the alarm will be generated again. However,
Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is executed.
IM 11M13A01-02E
12-6
<12. Troubleshooting>
12.2.2
Remedies When Alarms are Generated
Alarm 1: Oxygen Concentration Alarm
This alarm is generated when a measured value exceeds an alarm set point or falls below it.
For details on the oxygen concentration alarm, see Section 8.3, Oxygen Concentration Alarms
Setting, in the chapter on operation.
Alarm 6: Zero Calibration Coefficient (Correction Ratio) Alarm
In automatic or semi-automatic calibration, this alarm is generated when the zero correction ratio
is out of the range of 100±30% (refer to Subsection 9.1.3, Compensation).
The following are possible causes of this alarm.
(1) The zero gas oxygen concentration does not agree with the value of the zero gas
concentration set in Calibration setup. Otherwise, the span gas is used as the zero gas.
(2) The flow rate of zero gas is out of the specified flow (600±60 ml/min).
(3) The sensor assembly is damaged and so cell voltage is not normal.
(1) Check that the following have been set up correctly. If not, correct them. Then, recalibrate.
a. Check the preset zero gas concentration in the Calibration setup display. The displayed
concentration value has agreed with the concentration of the zero gas actually used.
b. The piping for calibration gases has been constructed so that the zero gas does not leak.
(2) If the alarm is not generated during the recalibration, improper calibration conditions are
considered as the cause of the alarm generated in the previous calibration. In this case, no
particular restoration is necessary.
(3) If the alarm is generated again during the recalibration, deterioration of or damage to the
sensor assembly is considered as the cause of the alarm. It is necessary to replace the
sensor (cell) with a new one. Before replacement, carry out the following.
Check the cell voltages when the zero gas and span gas are introduced.
a. Display the Detailed-data display by pressing the [Detailed-data] key in the Basic panel
display.
b. When the [▼] key is pressed once, the cell voltage should be indicated on the top line
(Figure 12.5).
c. Check whether or not the displayed cell voltage is very different from the theoretical value
at the oxygen concentration of the relevant gas. See Table 12.3 for the theoretical cell
voltages. Although the tolerance to the theoretical value cannot be generally specified, a
reasonable one may be approximately ±10 mV.
Table 12.3 Oxygen Concentration and Cell Voltage
Oxygen concentration (vol%O2)
Cell voltage(mV)
1%
67.1
21%
0
IM 11M13A01-02E
12-7
<12. Troubleshooting>
(4) Check whether the deterioration of or damage to the sensor assembly that caused the
alarm has occurred suddenly during the current calibration in the following procedure:
a. Call up the Detailed-data Display.
b. Use the [Page Scroll] key to check Calibration data (Figure 12.6). The span and zero
correction ratios of the last ten calibration can be checked here. By checking these data,
whether the sensor deterioration has occurred suddenly or gradually can be determined.
Tag:
Tag:
Cell voltage:
0 . 6 mV
4 2 . 1 mV
Thermo voltage:
C.J.resistance: 1 1 8 1 . 4 Ω
44Ω
Cell resistance:
0.24
Soft.rev.:
Calibration data:
1: 2 0 0 0 / 07 / 07
Span gas ratio:
Zero gas ratio:
2: 2 0 0 0 / 07 / 07
Span gas ratio:
Zero gas ratio:
Hold
F12-5E.ai
Figure 12.5 Detailed-data Display
1 4 : 57
-2.9%
70.3%
14:54
-2.9%
89.7%
Hold
F12-6E.ai
Figure 12.6 Calibration History
(5) If the sensor assembly has deteriorated suddenly, the check valve that prevents moisture in
the furnace from entering into the calibration pipes may have malfunctioned. If the furnace
gas flows into calibration lines, the gas is cooled and thus condensation develops and
accumulates in the pipe. During calibration the condensate is carried with the calibration gas
and blow onto the sensor assembly, whereby the cell is cooled quickly. This results in the
failed sensor assembly.
(6) If the sensor assembly has deteriorated gradually, check the condition of the sensor
assembly following the procedure below.
a. Use the [Page Scroll] key to check Cell Resistance. It should be 200 Ω or less if the cell
(sensor) is new. On the other hand, if the cell (sensor) is approaching the end of its service
life, it will be 3 to 10 kΩ.
b. Use the [Page Scroll] key to check Cell robustness. It should say “life > 1 year” if the cell
(sensor) is in good condition.
Tag:
Span gas ratio:
-2.9%
Zero gas ratio:
70.3%
Response time:
0s
Cell robustness: life< 1 month
Cell temperature:
1 0 6 7 °C
C.J.temperature:
4 7 °C
Hold
F12-7E.ai
Figure 12.7
IM 11M13A01-02E
12-8
<12. Troubleshooting>
Alarm 7: Span Calibration Coefficient (Correction Ratio) Alarm
In automatic or semi-automatic calibration, this alarm is generated when the span correction ratio
is out of the range of 0±18% (refer to Subsection 9.1.3, Compensation).
The following are suspected as the cause:
(1) The oxygen concentration of the span gas does not agree with the value of the span gas set
in Calibration setup.
(2) The flow of the span gas is out of the specified flow value (600±60 ml/min).
(3) The sensor assembly is damaged and the cell voltage is abnormal.
(1) Confirm the following and carry out calibration again: If the items are not within their proper
states, correct them.
a. If the display “Span gas conc” is selected in Calibration setup, the set point should agree
with the concentration of span gas actually used.
b. The calibration gas tubing should be constructed so that the span gas does not leak.
(2) If no alarm is generated as a result of carrying out re-calibration, it is suspected that
improper calibration conditions were the cause of the alarm in the preceding calibration.
In this case, no specific restoration is necessary.
(3) If an alarm is generated again as a result of carrying out re-calibration, deterioration of or
damage to the sensor assembly is suspected as the cause of the alarm. Replacement
of the sensor (cell) with a new one is necessary. However, before replacement, carry
out the procedure described in step (3) and later of in Subsection 12.2.2, Alarm 6: Zero Calibration Coefficient (Correction
Ratio) Alarm.
Alarm 8: EMF Stabilization Time-up Alarm
This alarm is generated if the sensor (cell) voltage has not stabilized even after the calibration
time is up for the reason that the calibration gas (zero gas or span gas) has not filled the sensor
assembly of the detector.
The following are suspected as the cause:
(1) The flow of the calibration gas is less than normal (a specified flow of 600±60 ml/min).
(2) The length or thickness of the calibration gas tubing has been changed (lengthened or
thickened).
(3) The sample gas flows toward the tip of the probe.
(4) The sensor (cell) response has deteriorated.
(1) Carry out calibration by passing the calibration gas at the specified flow (600±60 ml/min)
after checking that there is no leakage in the tubing.
(2) If calibration is carried out normally, perform a steady operation without changing the
conditions. If the error occurs again, check whether or not the reason is applicable to the
following and then replace the sensor assembly.
• A lot of dust and the like may be sticking to the tip of the detector probe. If dust is found,clean
the probe (see Subsection 11.1.1).
In addition, if an error occurs in calibration even after the sensor assembly is replaced, the
influence of sample gas flow may be suspected. Do not let the sample gas flow toward the
tip of the detector probe, for example, by changing the mounting position of the detector.
IM 11M13A01-02E
12-9
<12. Troubleshooting>
Alarm 10: Cold Junction Temperature Alarm
This alarm is generated when the temperature of the cold junction located at the terminal block of
the detector falls below -25°C or exceeds 155°C. Check the following:
Display “C.J.temperature” in the Detailed-data display. If “C.J.temperature” is indicated as 200°C
or -50°C, the following can be considered.
(1) Breakage of the cold junction signal wires between the converter and the detector occurred,
or the cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the cold junction signal wiring are shorted out in the
wiring extension or at the connection terminals.
(3) A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
(4) A failure of the electrical circuits inside the converter occurred.
If “C.J.temperature” exceeds 150°C or falls below -20°C, the following can be considered.
(1) The temperature of the detector terminal block is out of the operating temperature range
(-20 to 150°C).
(2) A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
(3) A failure of the electrical circuits inside the converter occurred.
Before proceeding to the following troubleshooting procedure, examine whether or not the
temperature of the detector terminal block is out of the operating temperature range.
The operating temperature range varies with the type of detector. If the detector terminal block
is out of its operating temperature range, take the measure to lower the temperature, such as
situating it so that it is not subjected to radiant heat.
The case where the Model ZR22 Detector is used:
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 5 and 6 of the detector and measure the resistance
between these terminals. If the resistance value is out of the range of 1 to 1.6 kΩ, the cold
junction temperature sensor is considered to be faulty. Replace that temperature sensor
with a new one.
(3) If the resistance value is within the above range, the cold junction temperature sensor
seems to be normal. Check whether or not the cable is broken or shorted out, and whether
the cable is securely connected to the terminals. Also, check that the resistance of the wiring
between the converter and detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa Electric Corporation.
IM 11M13A01-02E
<12. Troubleshooting>
12-10
Alarm 11: Thermocouple Voltage Alarm
This alarm is generated when the e.m.f. (voltage) of the thermocouple falls below -5 mV (about
-170°C) or exceeds 42.1 mV (about 1020°C). Whenever Alarm 11 is generated, Error 2 (heater
temperature failure) occurs.
The following are suspected as the cause:
(1) Breakage of the heater thermocouple signal wire between the converter and the detector
occurred, or the cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the heater thermocouple signal wiring are shorted out in
the wiring extension or at the connection terminals.
(3) A failure of the thermocouple at the detector heater assembly occurred.
(4) A failure of the electrical circuits inside the converter occurred.
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance
between these terminals. If the resistance value is 5 Ω or less, the thermocouple seems to
be normal. If it is higher than 5 Ω, it may indicate the possibility that the thermocouple has
broken or is about to break. In this case, replace the heater assembly (refer to Subsection
11.1.3, Replacement of the Heater Unit (Assembly)).
CAUTION
• Measure the thermocouple resistance value after the difference between the detector tip
temperature and ambient temperature falls to 50°C or less. If the thermocouple voltage is
large, accurate measurement cannot be achieved.
(3) If the thermocouple is normal, check whether or not the wiring cable is broken or shorted
out, and also whether the wiring cable is securely connected to the terminals.
Also check that the wiring resistance between the converter and the detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa.
Alarm 13: Battery Low Alarm
An internal battery is used as backup for the clock. After this alarm occurs, removing power from
the instrument may cause the clock to stop but should not affect stored parameters. The internal
clock is used for blow back scheduling; if you use this then after a battery alarm occurs (until the
battery is replaced) be sure to check/correct the date and time every time you turn on the power.
When the battery low alarm occurs, remember that the battery cannot be replaced by the user.
Contact your Yokogawa service representative.
NOTE
Battery life varies with environmental conditions.
* If power is applied to the instrument continuously, then the battery should not run down, and
life is typically about ten years. However the battery will be used during the time interval
between shipment from the factory and installation.
* If power is not applied to the instrument, at normal room temperatures of 20 to 25°C then
battery life is typically 5 years, and outside this range but within the range -30 to +70°C then
battery life is typically 1 year.
IM 11M13A01-02E
<12. Troubleshooting>
12-11
12.3 Countermeasures When Measured Value
Shows Error
The causes that the measured value shows an abnormal value is not always due to instrument
failures. There are rather many cases where the causes are those that sample gas itself is
in abnormal state or external causes exist, which disturb the instrument operation. In this
section, causes of and measures against the cases where measured values show the following
phenomena will be described.
(1) The measured value is higher than the true value.
(2) The measured value is lower than the true value.
(3) The measured value sometimes shows abnormal values.
12.3.1
Measured Value Higher Than True Value
(1) The sample gas pressure becomes higher.
The measured oxygen concentration value X (vol%O2) is expressed as shown below,
when the sample gas pressure is higher than that in calibration by ∆p (kPa).
X=Y [ 1+ (∆p/101.30) ]
where Y: Measured oxygen concentration value at the same pressure as in calibration
(vol%O2).
Where an increment of the measured value by pressure change cannot be neglected,
measures must be taken.
Investigate the following points to perform improvement available in each process.
• Is improvement in facility’s aspect available so that pressure change does not occur?
• Is performing calibration available under the average sample gas pressure
(internal pressure of a furnace)?
(2) Moisture content in a reference gas changes (increases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
in the air may cause an error in measured oxygen concentration value (vol%O2).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
(3) Calibration gas (span gas) is mixing into the detector due to leakage.
If the span gas is mixing into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
higher than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
confirming that they are in fully closed states. In addition, check the tubing joints for leakage.
(4) The reference gas is mixing into the sample gas and vice versa.
Since the difference between oxygen partial pressures on the sensor anode and cathode
sides becomes smaller, the measured value shows a higher value.
An error which does not appear as the Error 1 may occur in the sensor. Sample gas and/or
the reference gas may be leaking. Visually inspect the sensor. If any crack is found, replace
the sensor assembly with a new one.
NOTE
Data such as cell robustness displayed in the Detailed-data display should also be used for
deciding sensor quality as references.
IM 11M13A01-02E
12.3.2
<12. Troubleshooting>
12-12
Measured Value Lower Than True Value
(1) The sample gas pressure becomes lower.
Where an increment of the measured value due to pressure change cannot be neglected,
take measures referring to Subsection 12.3.1 (1).
(2) Moisture content in a reference gas changes (decreases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
content in the air may cause an error in measured oxygen concentration value (vol%O2).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
(3) Calibration gas (zero gas) is mixed into the detector due to leakage.
If the zero gas is mixed into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
lower than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
confirming that they are in a fully closed situation.
(4) Combustible components exist in the sample gas. If combustible components exist in the
sample gas, they burn in the sensor and thus oxygen concentration decreases.
Check that there are no combustible components.
(5) Temperature of the detector cell reaches 750°C or more.
12.3.3
Measurements Sometimes Show Abnormal Values
(1) Noise may be mixing in with the converter from the detector output wiring.
Check whether the converter and detector are securely grounded.
Check whether or not the signal wiring is laid along other power cords.
(2) The converter may be affected by noise from the power supply.
Check whether or not the converter power is supplied from the same outlet, switch, or
breaker as other power machines and equipment.
(3) Poor wiring contact.
If there is poor contact in the wiring, the sensor voltage or thermocouple e.m.f. (voltage) may
vary due to vibration or other factors.
Check whether or not there are loose points in the wiring connections or loose crimping
(caulking) at the crimp-on terminal lugs.
(4) There may be a crack in the sensor or leakage at the sensor-mounting portion.
If the indication of concentration varies in synchronization with the pressure change in the
furnace, check whether or not there is a crack in the sensor or whether the sensor flange is
sticking tightly to the probe-attaching face with the metal O-ring squeezed.
(5) There may be leakage in the calibration gas tubing.
In the case of a negative furnace inner pressure, if the indication of concentration varies with
the pressure change in the furnace, check whether or not there is leakage in the calibration
gas tubing.
IM 11M13A01-02E
Customer
Maintenance
Parts List
Model ZR22S
Zirconia Oxygen Analyzer, Detector
(Separate type Explosion-proof)
1
ZR22A
Heater Assembly
7
2
3
6
5
10
4
Item
1
2
3
4
5
6
7
8
9
10
11
9
8
11
Part No.
Qty
K9477EA
--G7109YC
K9470BK
E7042DW
1
4
--K9470ZF
K9470ZG
E7042BR
K9470BM
1
K9473AN
--ZR01A01-01
ZR01A01-02
ZR01A01-05
1
1
ZR01A01-10
E7042BS
K9470BJ
E7042AY
--K9470ZK
K9470ZL
4
1
1
1
1
1
1
ZR22S.ai
Description
Flame Arrestor
Bolt
(M5x12, SUS316 stainless steel)
(M5x12, inconel) for Option code "/C"
Washer (SUS316 stainless steel)
Bolts and Washers
G7109YC x 4 + E7042DW x 4
K9470BK x 4 + E7042DW x 4 for Option code "/C"
Plate
Pipe
Pipe for Option code "/C"
Cell Assembly
1 piece (E7042UD)
2 pieces
5 pieces
10 pieces
Contact
Metal O-ring
Filter
Calibration Tube Assembly
Cal. Gas Tube Assembly
Cal. Gas Tube Assembly for Option code "/C"
All Rights Reserved, Copyright © 2005, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M13A01-02E
1st Edition : Apr. 2005 (YK)
4th Edition : Feb. 2016 (YK)
Customer
Maintenance
Parts List
Model ZR402G
Zirconia Oxygen Analyzer/High Temperature
Humidity Analyzer,Converter
1
Hood for ZR402G
ZR402G
2
Item
Part No.
Qty
Description
1
A1113EF
1
Fuse (3.15A)
2
K9471UF
1
Hood
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M12C01-01E
1st Edition : Jan. 2000 (YK)
3rd Edition : Feb. 2011 (YK)
Customer
Maintenance
Parts List
Model ZR40H
Automatic Calibration Unit
for Separate type Zirconia Oxygen/Humidity
Analyzer (ZR22G + ZR402G)
1
LOCK
ZERO
LOCK
COMMON SPAN ZERO
AUTO CALIBRATION UNIT
MODEL
SUFFIX
ZR40H
STYLE
S1
SUPPLY 690kPa MAX.
AMB.TEMP-20 TO 558C
USED WITHZR402G
NO.
Made in Japan
REF.OUT
CAL.OUT
SPAN IN
ZERO IN
Item
Part No.
Qty
Description
1
K9473XC
1
Flowmeter
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M12A01-11E
1st Edition : Dec. 2000 (YK)
4th Edition : Mar. 2013(YK)
Customer
Maintenance
Parts List
Model ZO21P-H
Zirconia Oxygen Analyzer
High Temperature Probe Adaptor
4
5,6
7
3
8,9,10
11
ZR22G
2
1
Item
Part No.
Qty
Description
Item
Part No.
Qty
Description
1
K9292TP
E7046CF
K9292TQ
E7046CG
E7046CH
1
1
1
1
1
Probe (SiC, L=0.5m)
Probe (SiC, L=0.6m)
Probe (SiC, L=0.7m)
Probe (SiC, L=0.8m)
Probe (SiC, L=0.9m)
2
E7046FA
E7046FE
E7046FK
E7046FD
E7046FC
1
1
1
1
1
Probe Adaptor (for JIS 5K-50-FF)
Probe Adaptor (for ANSI CLASS150-4-RF)
Probe Adaptor (for DIN PN10-DN50-A)
Probe Adaptor (for JIS 10K-100-FF)
Probe Adaptor (for JIS 10K-80-FF)
E7046AL
E7046BB
K9292TV
E7046CR
K9292TW
1
1
1
1
1
Probe (SiC, L=1.0m)
Probe (SiC, L=1.5m)
Probe (SUS, L=0.5m)
Probe (SUS, L=0.6m)
Probe (SUS, L=0.7m)
E7046FB
E7046FG
E7046FF
E7046FJ
E7046FH
1
1
1
1
1
Probe Adaptor (for JIS 10K-65-FF)
Probe Adaptor (for ANSI CLASS150-3-RF)
Probe Adaptor (for ANSI CLASS150-2 1/2-RF)
Probe Adaptor (for JPI CLASS150-4-RF)
Probe Adaptor (for JPI CLASS150-3-RF)
E7046CS
E7046CT
E7046AP
E7046AQ
1
1
1
1
Probe (SUS, L=0.8m
Probe (SUS, L=0.9m
Probe (SUS, L=1.0m
Probe (SUS, L=1.5m
3
4
5
6
7
E7046FQ
E7046FN
Y9825NU
Y9800WU
G7073XL
1
1
4
8
1
Gasket
Plate
Bolt
Washer
Gasket
8
9
10
11
Y9630RU
Y9121BU
Y9120WU
Y9801BU
4
4
4
4
Bolt
Nut
Washer
Nut
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M03B01-10E
1st Edition : Dec. 2000 (YK)
8th Edition : Sep. 2012 (YK)
Customer
Maintenance
Parts List
E7046EC/E7046EN
Zirconia Oxygen Analyzer
Auxiliary Ejector Assembly (for Model ZO21P-H)
1
3
2
4
5
6
7
Item
Part No.
Qty
1
E7046EC
1
Auxiliary Ejector Assembly, Connection Rc1/4
E7046EN
1
Auxiliary Ejector Assembly, Connection 1/4 NPT
L9852CB
1
Needle Valve, Connection Rc1/4
G7016XH
1
Needle Valve, Connection 1/4NPT
E7046EK
1
Pressure Gauge, Connection Rc1/4
E7046EV
1
Pressure Gauge, Connection 1/4NPT
E7046ED
1
Ejector, Connection Rc1/4
E7046EP
1
Ejector, Connection 1/4NPT
E7046EF
1
Nozzle Assembly, Tube Connection Ø6хØ4 tube
E7046ER
1
Nozzle Assembly, Tube Connection 1/4 inch tube
6
G7031XA
1
Tee, Connection Rc1/4
7
E7046EJ
1
Reducing nipple, Connection R1/4
2
3
4
5
Description
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M03B01-05E
1st Edition : Dec. 2000 (YK)
8th Edition : Mar. 2013 (YK)
Customer
Maintenance
Parts List
Model ZO21S
Zirconia Oxygen Analyzer/ High Temperature
Humidity Analyzer, Standard Gas Unit
1
2
3
Item
Part No.
Qty
Description
1
———
1
Pump (see Table 1)
2
E7050BA
1
Zero Gas Cylinder (x6 pcs)
3
E7050BJ
1
Needle Valve
Table 1
Power
Pump
AC 100 V
110
115
E7050AU
AC 200 V
220
240
E7050AV
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M3D1-01E
1st Edition : Jan. 2000 (YK)
4th Edition : Mar. 2011 (YK)
i
Revision Information
Title
: Model ZR22S, ZR402G Separate type Explosion-proof Zirconia Oxygen Analyzer
Manual No. : IM 11M13A01-02E
June 2017/6th Edition
Added RoHS compliance statement with some revision on pages i, iv, vii,2-1, 2-2, 2-4, 2-5, 2-6, 3-1,
3-2, 3-3, 3-5, 3-6.
Apr. 2016/5th Edition
Addition the related documents, etc.
CMPL 11M13A01-02E revised to 4th edition.
Aug. 2015/4th Edition
Revised section
2.1.1 “ZR22S Separate type Explosion-proof Detector “ NAME PLATE:
Addition of the maximum surface temperature for dust-proof.
“General Specifications”: Added to “Standard Specifications”
2.7.2 “Stop Valve”: Changed of the weight and external dimensions.
2.7.3 “Check Valve”: Changed of the weight.
Nov. 2014/3rd Edition
Revised and Corrected over all
Jul. 2006/2nd Edition
Introduction
Warning about Explosion-proof: Deleted description.
Introduction
Explosion-proof Approval: Added description.
2.2.1 “ZR22S Separate type Explosion-proof Detector”: Added description.
2.2.1 “ZR22 Separate type Explosion-proof Detector”: Added description.
2.5
Model and Suffix Code: Added one suffix code.
2.7.5 “Cylinder Regulator value (Part No. G7013XF or G7014XF)”: Change drawing.
3.1.5 “IECEx Flameproof Type”
8.2.3 “Output Hold Setting”: Changed value in Figures 8.2 and 8.3.
8.2.4 “Default Values”: Changed descriptions in Table 8.2.
8.6.4 “Setting Purging”: Added item.
8.6.5 “Setting Passwords”: Changed Section number.
10.3 “Operational Data Initialization”: Changed value in Table 10.5.
12.2.2 Alarm 6: Changed descriptions, Alarm 7: Changed descriptions.
CMPL 11M12A01-02E: Changed part numbers
Apr. 2005/1st Edition
Newly published
IM 11M13A01-02E
Blank Page