Endress + Hauser Operating Instructions Stamolys CA71CL Analyzer for Photometric Determination of Free Available and Total Chlorine
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Operating Instructions
Stamolys CA71CL
Analyzer for Photometric Determination of Free Available and
Total Chlorine
BA354C/07/en/10.06
51512312
Valid as of:
Software version 5.9
Brief overview
1
3
2
12
4
30
3
38
E K
RS 232
15
120
326
60
GY 59
+17 V WH 56
39
I2 +
I2 -
40
NC
NO
COM
Reserve
38
fx
55
Mess 2
Meas 2
37
n.b.
n.c.
54
+24V
36
I1 + 30
0V
35
I1 -
Status GN 53
31
29
n.b.
n.c.
n.b.
n.c.
NC
NO
27
26
E
K
25
24
23
Ø 26
Ø 21
47
125
Ø 26
360
Kanal 2
Channel 2
22
21
20
19
18
17
16
15
ALARM 2
Kanal 1
Channel 1
COM
28
Analog Out
0/4-20mA
Schirm
Screen
keine Probe
No sample
OPTION
68
V 1 BU 67
0 V BN 66
YE 57
ALARM 2
69
V2
71
BK 58
-17 V
14
13
12
ALARM 1
V3
73
0V fx
MP5 MP6
0V
ALARM 1
74
72
Schirm
Screen BN 63
n.b.
WH 62
n.c.
n.b.
YE 61
n.c.
STÖRUNG
ERROR
GN 64
70
11
F
Kalibrierung
Messung
Spülung
Warten
PARAMETRIEREN
F
Messbereich
Anfang
Messbereich
Ende
Grenzwert
GW 1
Grenzwert
GW 2
Zeitpunkt
1. Messung
Messintervall
1. Kalibrierung
KONFIGURIEREN
F
Codenummer
Photometer
Werkseinstellung
Maßeinheit
Kalibrierfaktor
c-Offset
Verdünnung
AUTO
0V RELAIS / Relay:
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!
Terminals voltage to ground even when the unit is switched off !!!
7
L2
6
N
580
648
243
165
7
119
78
8
YE 76
Photometer
Schirm/
Screen
BK 65
M 0.2 A
V4
MESSEN
555
524
9
79
BK 78
RD 77
A
F5
Schirm
Screen 75
LECKF.
LEAK S.
60
200
80
SCHRITTMOTOR
STEP MOTOR
140
100 OR
15
6
Photometer -17V
Photometer +17V
250
160 F4 M 0.2 A
330
10
M CE
9
15
5
L1
4
PE
AC Motor
T 0.1A
F2
NETZ
MAINS
3
Elektronik
Electronics
T 0.5A
F1
L
2
N
1
PE
Zeitpunkt
Kalibrierintervall
155
Ø 21
5
110
Ø 27
163
11
AC MOTOR
M
Probe
ansaugen
131
V3
161
134
160
161
160
144
200
V2
V1
207
436
130
A
B
SPRACHAUSWAHL
F
Deutsch
FEHLER
LESEN
F
Fehlermeldungen
SERVICE
F
Pumpen
und Ventile
English
Français
Suomi
Polski
Italiano
A
154
B
141
Espanol
V1
143
121
170/171
F
R1
R2
B
C
Messwerte
130
R
121
Frequenz
135
135
DATENSPEICHER
V2
Ausgabe
C
Löschen
142
123
122
120
D
E
F
a0006651
A
→ ä8
Overview of analyzer: most important modules
Æ
B
→ ä9
→ ä 14
Dimensions, installation conditions
Installation instructions and installation examples
Æ
C
→ ä 15
→ ä 18
Terminal assignment
Signals, switching contacts
Æ
D
→
→
→
→
ä 24
ä 29
ä 50
ä 31
Operation: configuration
Calibration
Operating matrix
Commissioning
Æ
E
→ ä 35, → ä 58
→ ä 36
→ ä 41, → ä 54
Maintenance schedule
Replacing consumables and wear parts
Accessories
Æ
F
→ ä 43
→ ä 45, → ä 54
Troubleshooting
Spare parts
Æ
→ ä 48
Technical data
Endress+Hauser
Stamolys CA71CL
Table of contents
1
Safety instructions . . . . . . . . . . . . . . . . 4
1.1
1.2
1.3
1.4
1.5
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation, commissioning and operation . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Notes on safety icons and icons . . . . . . . . . . . . . . . .
2
Identification . . . . . . . . . . . . . . . . . . . . 6
2.1
2.2
2.3
Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Certificates and approvals . . . . . . . . . . . . . . . . . . . . 7
3
Installation . . . . . . . . . . . . . . . . . . . . . . 8
3.1
3.2
3.3
3.4
3.5
3.6
The analyzer at a glance . . . . . . . . . . . . . . . . . . . . . 8
Incoming acceptance, transport, storage . . . . . . . . . . 9
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 9
Installation instructions . . . . . . . . . . . . . . . . . . . . . 12
Installation examples . . . . . . . . . . . . . . . . . . . . . . . 14
Post-installation check . . . . . . . . . . . . . . . . . . . . . . 14
4
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1
4.2
4.3
4.4
4.5
Electrical connection . . . . . . . . . . . . . . . . . . . . . . .
Signal connections . . . . . . . . . . . . . . . . . . . . . . . . .
Switching contacts . . . . . . . . . . . . . . . . . . . . . . . .
Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . .
Post-connection check . . . . . . . . . . . . . . . . . . . . . .
5
Operation . . . . . . . . . . . . . . . . . . . . . . 22
5.1
5.2
5.3
Display and operating elements . . . . . . . . . . . . . . . 22
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6
Commissioning. . . . . . . . . . . . . . . . . . 31
6.1
6.2
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7
Maintenance. . . . . . . . . . . . . . . . . . . . 35
7.1
7.2
7.3
7.4
7.5
7.6
7.7
7.8
Maintenance schedule . . . . . . . . . . . . . . . . . . . . . .
Replacing reagents . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing pump hoses . . . . . . . . . . . . . . . . . . . . . .
Replacing valve hoses . . . . . . . . . . . . . . . . . . . . . .
Replacing the static mixer . . . . . . . . . . . . . . . . . . .
Replacing the photometer cell . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Placing out of service . . . . . . . . . . . . . . . . . . . . . . .
8
Accessories. . . . . . . . . . . . . . . . . . . . . 41
8.1
8.2
8.3
8.4
Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . .
Reagents, cleaners, standard solutions . . . . . . . . . .
Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . .
Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Endress+Hauser
4
4
4
4
5
15
18
19
20
21
8.5
8.6
Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Additional accessories . . . . . . . . . . . . . . . . . . . . . . 42
9
Troubleshooting . . . . . . . . . . . . . . . . . 43
9.1
9.2
9.3
9.4
9.5
Troubleshooting instructions . . . . . . . . . . . . . . . . .
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software history . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10
Technical data . . . . . . . . . . . . . . . . . . . 48
10.1
10.2
10.3
10.4
10.5
10.6
10.7
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
11
Appendix . . . . . . . . . . . . . . . . . . . . . . . 50
11.1
11.2
11.3
11.4
Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . .
Ordering forms . . . . . . . . . . . . . . . . . . . . . . . . . . .
Analyzer settings . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance schedule . . . . . . . . . . . . . . . . . . . . . .
43
45
46
47
47
48
48
48
48
49
49
49
50
54
56
58
Index . . . . . . . . . . . . . . . . . . . . . . . . . . 61
35
36
36
38
39
39
40
40
41
41
41
41
3
Safety instructions
Stamolys CA71CL
1
Safety instructions
1.1
Designated use
The analyzer is a compact photometric analytical system.
It is designed for monitoring free available or total chlorine in water.
CA71 is particularly suitable for the following applications:
• Drinking water treatment
• Monitoring wastewater outlets
• Monitoring swimming pool water
• Process water treatment
Any use other than that described here compromises the safety of persons and the entire measuring
system and is, therefore, not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2
Installation, commissioning and operation
Please note the following:
• Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
The technical personnel must be authorized to perform such tasks by the owner-operator.
• The electrical connection may only be established by authorized electrical technicians.
• Technical personnel must have read and understood these Operating Instructions and must
follow the instructions they contain.
• Before commissioning the entire measuring point, check that all the connections are correct.
Ensure that electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning. Mark
the damaged product as defective.
• Faults at the measuring point may only be rectified by authorized and specially trained personnel.
• If faults cannot be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out directly at the
manufacturer’s or by the service organization.
1.3
Operational safety
The analyzer has been designed to state-of-the-art technology, has been tested and left the factory
in a condition in which it is safe to operate.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety regulations:
• Installation instructions
• Local prevailing standards and regulations
1.4
Return
If the analyzer requires repair, please send it cleaned to your sales center.
Please use the original packaging when returning the device.
Please enclose the completed "Declaration of Contamination and Cleaning" form (copy the second
last page of these Operating Instructions) with the packaging and also the shipping documents. The
unit cannot be repaired if this declaration is not duly completed!
4
Endress+Hauser
Stamolys CA71CL
Safety instructions
1.5
#
"
!
Endress+Hauser
Notes on safety icons and icons
Warning!
This symbol alerts you to hazards. They could cause serious injuries as well as damage to the
instrument if ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
5
Identification
Stamolys CA71CL
2
Identification
2.1
Device designation
2.1.1
Nameplate
Compare the order code on the nameplate (on the analyzer) to the product structure and your order.
You can read the following information from the
nameplate:
• Order code (device version)
• Serial number
• Measuring range
• Outputs and communication
• Power supply
• Degree of protection
• (Permitted) ambient temperature
a0006011
Fig. 1:
2.1.2
Nameplate
Product structure
Measuring range
A
B
C
D
Y
0.01 to 1.00 mg/l free available chlorine
0.01 to 1.00 mg/l total chlorine
0.10 to 10.00 mg/l free available chlorine
0.10 to 10.00 mg/l total chlorine
Special version as per customer specifications
Sample transfer
1
2
From one measuring point (one-channel version)
From two measuring points (two-channel version)
Power supply
0
1
2
3
230 V AC / 50 Hz
115 V AC / 60 Hz
115 V AC / 50 Hz
230 V AC / 60 Hz
Collecting vessel (for up to 3 analyzers)
A
B
C
D
Without collecting vessel
Collecting vessel without level monitoring
Collecting vessel with level monitoring (one-channel version only)
Two collecting vessels without level monitoring (two-channel version)
Housing
1
2
3
8
Open design
GFR
Stainless steel (1.4301 / AISI 304) + vent.
Stainless steel (1.4301 / AISI 304) without vent.
Output
A
B
0/4 to 20 mA, RS 232
Profibus DP
Reagents
1
2
3
CA71CL-
6
To be ordered separately
One set inactive
Three sets inactive
complete order code
Endress+Hauser
Stamolys CA71CL
Identification
2.2
!
Scope of delivery
Note!
Reagents must be ordered separately for version CA71XX-XXXXXX1.
Inactive reagents form part of the scope of delivery for all other versions. They must be mixed before
use. Please refer to the guidelines enclosed for this purpose.
The scope of delivery comprises:
• 1 analyzer with power supply plug
• 1 cleaning injector
• 1 can of silicone spray
• 1 Noprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)
• 1 C-Flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)
• 1 C-Flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)
• 2 hose connectors of each:
– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)
– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)
– 6.4 mm x 3.2 mm (0.25 inch x 0.13 inch)
• 2 T-hose connectors of each:
– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)
– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)
• 1 interference suppressor for the current output
• 4 edge covers (with GFR housing only)
• 1 roll of Teflon tape
• 1 quality certificate
• 1 set of Operating Instructions (English)
2.3
Certificates and approvals
2.3.1
4 mark
Declaration of conformity
The product meets the legal requirements of the harmonized European standards. The
manufacturer confirms compliance with the standards by affixing the 4 mark.
2.3.2
Manufacturer's certificate
Quality certificate
With this certificate, the manufacturer confirms that all technical regulations are met and specified
tests have been carried out individually for your product.
Endress+Hauser
7
Installation
Stamolys CA71CL
3
Installation
3.1
The analyzer at a glance
a0006582
Fig. 2:
1
2
3
4
5
1)
Analyzer (cabinet version, without hoses)
Reagent pump (P2), supply from canister
Display
Serial interface RS 232
Photometer cell
Static mixer
6
7
8
9
10
Valve V2
Outlet for sample or reagent mixture
Valve V1
Channel switchover: top channel 1, bottom
channel 21)
Sample pump P1, supply see below
Only for two-channel version
Supply to sample pump:
• Valve V1
– Hose at front: sample supply
– Hose at rear: supply from valve V2 (cleaner or standard)
• Valve V2
– Hose at front: supply from canister with standard
– Hose at rear: supply from canister with cleaner
8
Endress+Hauser
Stamolys CA71CL
Installation
3.2
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents.
Keep the damaged products until the matter has been settled.
• Check that the scope of delivery is complete and matches your order and the shipping documents.
• Pack the device in such a way as to protect it reliably against impact and moisture for storage and
transportation. The original packaging provides optimum protection. In addition, the permitted
ambient conditions must be observed (see "Technical data").
• If you have any queries, please contact your supplier or your sales center.
3.3
Installation conditions
3.3.1
Design, dimensions
Version with stainless steel housing
Detail A
3(0.12)
30(1.18)
15(0.59)
38(1.5)
110(4.33)
Ø 27(1.06)
163(6.42)
9(0.35)
15(0.59)
250 (9.84)
330 (13.0)
160 (6.30)
140 (5.51)
15(0.59)
100(3.94)
360 (14.2)
436 (17.2)
580 (22.8)
Ø 21(0.83)
Ø 26(1.02)
648 (25.5)
Ø 26 (1.02)
47(1.85)
Ø 21(0.83)
78(3.07)
119(4.69)
165 (6.50)
125(4.92)
555 (21.8)
326 (12.8)
524 (20.6)
200 (7.87)
A
243 (9.57)
120(4.72)
60(2.36)
207 (8.15)
mm(inch)
a0001361
Fig. 3:
Endress+Hauser
Stainless steel version
9
Installation
Stamolys CA71CL
Version with GFR housing
250 (9.84)
120 (4.72)
Ø 16(0.63)
Ø 26(1.02)
163 (6.42)
160 (6.30)
140 (5.51)
100(3.94)
330 (13.0)
648 (25.5)
Ø 26(1.02)
Ø 21
(0.83)
119/4.69
Ø 21(0.83)
78(3.07)
47(1.85)
165 (6.50)
555 (21.8)
200 (7.87)
125 (4.92)
7-11/
0.28 - 0.43
(M 6)
243 (9.57)
120
(4.72)
60 (2.36)
Ø 26 (1.02)
326 (12.8)
Ø 16 (0.63)
360 (14.2)
436 (17.2)
Ø 27 (1.06)
120(4.72)
465 (18.3)
mm(inch)
500 (19.7)
a0001354-en
Fig. 4:
GFR version
Open version
381 (15.0)
351 (13.8)
197 (7.76)
Ø6 (0.24)
192 (7.56)
19 (0.75)
175 (6.89)
562 (22.1)
540 (21.3)
RS 232
325 (12.8)
9 (0.35)
8 (0.31)
mm (inch)
a0001356
Fig. 5:
10
Version without housing (open design)
Endress+Hauser
Stamolys CA71CL
Installation
!
Note!
For the reagents, you need an additional base at max. 35 cm (13.8 inch) under the pumps in the
open version. The dimensions of the reagent bottles are as follows: 90 x 90 x 215 mm (3.54 x 3.54
x 8.46 inch). Depending on the version, the number of bottles varies between 2 and 5.
Furthermore, in these versions the outlet pipe has to be installed to the right beside the analyzer.
Please refer to the sheet enclosed with the Operating Instructions.
The outlet pipe must be secured to a wall in such a way that the incline of the outlet pipes from the
photometer is between 5 and 10 % . You must extend the outlet pipes where necessary.
3.3.2
Connecting the sample line
145 (5.71)
10 (0.39)
140 (5.51)
120 (4.72)
79 (3.11)
48(1.89)
17 (0.67)
70(2.76) 42(1.65)
130 (5.12)
4 x Ø 6.6 (0.26)
60 (2.36)
27 (1.06)
3 x Ø 28/20 (1.1/0.79)
115 (4.53)
Ø 34/25 (1.34/0.98)
4 x Ø 5.5/10.4 (0.22/0.41)
Ø 56 (2.2)
*
80(3.15)
100 (3.9)
60(2.36)
Ø 50H7 (1.97H0.28)
*
M20x1.5
Ø 24 (0.94)
*
mm (inch)
70 (2.76)
a0001363
Fig. 6:
1
2
3
4
5
Collecting vessel at side wall (optional)
Ventilation
Sample supply from sample conditioning
Collecting vessel
Electrical connections
Analyzer sample inlet
a0001346
Fig. 7:
*
6
7
8
Dimensions of collecting vessel and base plate
* Dimensions variable, as freely adjustable
Sampling for analyzer
Sample overflow
Analyzer outlet
One-channel version
Collecting vessel (at analyzer outer side, with and without level monitoring)
Connection
Hose ID 3.2 mm (0.13")
Customer collecting vessel
Connection
Max. distance from collecting vessel to analyzer
Max. height difference from collecting vessel to analyzer
Hose ID 1.6 mm (0.06")
1 m (3.28 ft)
0.5 m (1.64 ft)
Two-channel version
• One or two collecting vessels (with or without level monitoring) form part of the scope of delivery,
depending on the version.
• In the two-channel version, level monitoring is only possible for one channel.
• One collecting vessel can be mounted on the housing. A second vessel would have to be made
available near the device.
Endress+Hauser
11
Installation
Stamolys CA71CL
Adjusting the level monitoring
You have to adjust conductive level monitoring depending on whether you want to connect one,
two or three analyzers to this single collecting vessel.
1.
Depending on the application, mount the right adjusting pin (→ å 8, → å 9, item 2) or do
not use any adjusting pin.
2.
Pull the marked pipe (item 3) downwards in accordance with your application (1, 2 or 3
analyzers) to set an optimum collecting volume.
1
1
2
2
3
3
a0001367
Fig. 8:
1
2
3
3.4
One analyzer
M 3x12
M 3x35 (adjusting pin)
Marking 1
1
3
a0001366
Fig. 9:
1
2
3
Two analyzers
M 3x12
M 3x20 (adjusting pin)
Marking 2
a0001365
Fig. 10:
1
3
Three analyzers
M 3x12
Marking 3
Installation instructions
To install the analyzer at the intended location, proceed as follows:
12
1.
Mount the analyzer and secure it to a wall using screws (Ø6 mm / 0.24").
For the installation dimensions, please see the previous chapter.
2.
Using a spirit level, check that the cabinet is hanging or standing straight. Only in this way can
any air bubbles present escape from the cell.
3.
Fit the edge covers (with GFR housing only).
4.
Lay the drain pipe for the reaction products. Where possible, use solid pipes (PVC or PE,
internal diameter 1" with 3% incline).
5.
Insert the valve hoses accordingly. These were partly removed from the valves for
transportation. This prevents the hoses from sticking or pressure being applied to a point in the
hose over an extended period.
→ å 11
Endress+Hauser
Stamolys CA71CL
Installation
1
3
2
V3
V2
V1
6
7
5
a0006362
Fig. 11:
Valves and valve hoses
V1-2 Valves
Channel switchover
1
To sample pump
2
Connecting hose to valve 1, at rear
3
4
5
6
7
Outlet hose
Hose, valve 2 at front, standard
Hose, valve 2 at rear, cleaner
Hose, valve 1 at front, sample
Secure the hose boxes in the appropriate pump holders (→ å 12):
Sample pump left, reagent pump right. The direction of flow of the sample and reagent must
be counterclockwise.
6.
R3
R2
R1
a0002206
Fig. 12:
R1
R2
R3
7.
Reagent pump, top view
Reagent 1
Reagent 2 (if present)
Reagent 3 (if present)
Connect the sample supply.
! Note!
The sample can be obtained as follows:
– Directly or via a reversible flow filter or a cross-flow filter by means of a small pump (rating
approx. 300 ml/min), suitable for clear media, e.g. in the outlet of a sewage treatment plant
– From microfiltration; this is practical for media containing flocculants, e.g. in an activated
sludge basin
– Sample conditioning using ultrafiltration for contaminated media, e.g. from the primary
settling tank.
For questions regarding sample conditioning and its automation, please contact Service or the
sales center responsible for your region.
8.
Endress+Hauser
Connect the tubes from canisters containing reagents, standards and cleaning agents to the
following nozzles:
Canister
Tube designation (mark)
Sample
Reagent 1
Reagent 2
Cleaner
P
CL-1
CL-2
R
13
Installation
Stamolys CA71CL
3.5
Installation examples
3.5.1
CAT 430 or ultrafiltration provided by customer and two CA 71
units
• Permeate can contain air bubbles (CAT430)
or is free of bubbles (ultrafiltration provided by
customer)
• Distance between the analyzers as short as
possible: sample line between T-section and
second analyzer (, → å 13, item 2) shorter
than 1.5 m
• Cross-section of sample line ID 3.2 - 4 mm
• Only one collecting vessel required
5
CA 71
1
! Note!
Make sure sufficient sample is always available
for both analyzers. Bear this in mind when
selecting maintenance intervals for CAT430 and
when setting the buffer volume at the collecting
vessel.
3.5.2
4
3
2
a0001728
Fig. 13:
1
2
3
4
5
Installation example
Sample from CAT430
Sample line
Collecting vessel overflow
T-section
Collecting vessel
CAT 411, CAT 430 and two CA 71 units (two-channel version)
• Permeate can contain air bubbles (are
removed by the sample collector)
• Distance between the analyzers as short as
possible: sample line between T-section and
second analyzer (, → å 14, item 3) shorter
than 1.5 m
• Cross-section of sample line ID 3.2 - 4 mm
• One collecting vessel each (without level
monitoring) for CAT411 or CAT430
1
6
2
CA 71
3
CA 71
4
5
a0001729
Fig. 14:
! Note!
Make sure sufficient sample is always available
for both analyzers. Bear this in mind when
selecting maintenance intervals for CAT 411
and CAT 430.
3.6
CA 71
1
2
3
4
5
6
Installation example
Sample from CAT430
Sample from CAT411
Collecting vessels
Collecting vessel overflow
Sample lines
T-sections
Post-installation check
• After installation, check that all connections are secure and leaktight.
• Ensure that the hoses cannot be removed without effort.
• Check all hoses for damage.
14
Endress+Hauser
Stamolys CA71CL
Wiring
#
4
Wiring
4.1
Electrical connection
Warning!
• The electrical connection may only be established by an electrical technician.
• The electrical technician must have read and understood these Operating Instructions and must
follow the instructions they contain.
• Before beginning the connection work, ensure that no voltage is applied to the cables.
4.1.1
"
Quick wiring guide
Caution!
• You have to fold out the frame of the analyzer to reach the terminal block.
• Beforehand, release the hoses from the outlet pipe. Otherwise there is a danger of flooding.
• Connect the hoses again after folding the frame back in.
Fold out the frame as follows:
1.
Release the two bottom Allen screws (AF 6) by 3-4 rotations (→ å 15, item 1).
2.
Unscrew the two upper Allen screws completely so that the frame folds out. In this way, you
reach the terminal block (item 2).
2
M
E K
RS 232
1
a0001370
Fig. 15:
1
2
!
Endress+Hauser
Folding out the frame
Allen screws AF 6
Terminal block
Note!
The device does not have a power supply switch. Therefore, it is advantageous to have a fused
socket near the device.
15
Wiring
Stamolys CA71CL
4.1.2
Caution!
The following diagram shows an example of the connection compartment sticker (→ å 16). The
terminal assignment and cable colors can deviate from the actual assignment and colors!
Only use the terminal assignment of the stick-on label in the device (→ å 17) to connect your
analyzer!
BK 78
RD 77
YE 76
74
V3
69
73
V2
68
72
V 1 BU 67
71
0 V BN 66
Schirm
Shield BN 63
n.b.
WH 62
n.c.
n.b.
YE 61
n.c.
60
GY 59
-17 V
YE 57
Schirm
Shield
keine Probe
No sample
40
Schirm
Shield
39
I2 +
+17 V
56
Reserve
38
I2 -
fx
55
Mess 2
Meas 2
37
Schirm
Shield
n.b.
n.c.
WH 54
Status GN 53
NC
NO
COM
n.b.
n.c.
n.b.
n.c.
31
+24V
36
I1 + 30
0V
35
I1 -
29
NC
NO
27
26
25
24
23
21
20
19
18
17
16
15
ALARM 2
70
BK 58
Kanal 2
Channel 2
22
14
13
12
ALARM 1
SCHRITTMOTOR
STEP MOTOR
OR 79
V4
LECKF.
LEAK S.
Schirm
Screen 75
0V
Kanal 1
Channel 1
COM
28
Analog Out
0/4-20mA
OPTION
GN 64
80
0V fx
MP5 MP6
Photometer
Schirm
Shield
BK 65
M 0.2 A
STÖRUNG
ERROR
F5
ALARM 2
Photometer -17V
Photometer +17V
F4 M 0.2 A
ALARM 1
"
Terminal assignment
11
0V RELAIS / Relay:
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!
BU
7
L2
YE
6
N
RD
5
L1
GN/YE
4
PE
AC MOTOR
Terminals voltage to ground even when the unit is switched off !!!
AC Motor
T 0.1A
F2
Elektronik
Electronics
T 0.5A
F1
NETZ
MAINS
3
L
2
N
1
PE
a0001369
Fig. 16:
Example of connection stick-on label
3
2
60
GY 59
0V
-17 V
+17 V
fx
n.b.
n.c.
0V fx
MP5 MP6
BK 58
YE 57
56
55
WH 54
Status GN 53
Schirm
Shield
keine Probe
No sample
Reserve
Mess 2
Meas 2
Kanal 1
Channel 1
COM
28
Analog Out
0/4-20mA
40
39
Schirm
Shield
NC
NO
I2 +
38
I2 -
37
Schirm
Shield
COM
31
+24V
36
I1 + 30
0V
35
I1 -
29
n.b.
n.c.
n.b.
n.c.
NC
NO
27
26
25
24
23
Kanal 2
Channel 2
22
21
16
15
20
14
19
13
18
17
12
Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!
Terminals voltage to ground even when the unit is switched off !!!
ALARM 2
GN 64
Schirm
Shield BN 63
n.b.
WH 62
n.c.
n.b.
YE 61
n.c.
11
0V RELAIS / Relay:
BU
7
L2
YE
6
N
RD
5
L1
GN/YE
4
PE
AC MOTOR
68
0 V BN 66
OPTION
70
69
V2
V 1 BU 67
Photometer
Schirm
Shield
BK 65
M 0.2 A
V3
ALARM 2
V4
71
ALARM 1
74
73
72
ALARM 1
F5
Schirm
Screen 75
LECKF.
LEAK S.
YE 76
SCHRITTMOTOR
STEP MOTOR
80
BK 78
RD 77
STÖRUNG
ERROR
Photometer -17V
Photometer +17V
F4 M 0.2 A
OR 79
AC Motor
T 0.1A
F2
Elektronik
Electronics
T 0.5A
F1
NETZ
MAINS
3
L
2
N
1
PE
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
1
a0001372
Fig. 17:
1
2
3
16
Analyzer from above (open/folded out)
Connection compartment sticker
Board with terminal block
Rear of analyzer
Endress+Hauser
Stamolys CA71CL
Wiring
Function
Mains
Alarm value 1, channel 1
Alarm value 2, channel 1
Alarm value 1, channel 2
Alarm value 2, channel 2
Fault
Unassigned
Analog output 1
0/4 to 20 mA
Analog output 2
0/4 to 20 mA
Sample conditioning
remote control
Designation
Terminal
One-channel version
Terminal
Two-channel version
L
3
3
N
2
2
PE1)
1
1
COM
25
25
NC
24
24
NO
23
23
COM
28
28
NC
27
27
NO
26
26
COM
–
13
NC
–
12
NO
–
11
COM
–
16
NC
–
15
NO
–
14
COM
19
19
NC
18
18
NO
17
17
COM
22
22
NC
21
21
NO
20
20
+
36
36
–
35
35
Shield
37
37
+
–
39
–
–
38
Shield
–
40
Input
57
57
0V
53
53
Input
–
55
0V
–
53
Channel switchover
1)
!
Endress+Hauser
Brass screw with bolt at top right in the connection compartment (marked with
*)
Note!
• Alarm values 1 and 2 do not need to be connected if the process control system sets its own alarm
values at the analog output.
• When using a sample conditioning system: connect terminals 57 and 53 on CA 71 to the
corresponding terminals of the sample conditioning system. For the allocation of these terminals,
please see the Operating Instructions of the sample conditioning system.
• If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement (sample not
ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.
17
Wiring
Stamolys CA71CL
4.2
Signal connections
4.2.1
Shielding of the analog outputs
The interference suppressor dampens interference from electromagnetic effects on control, power
and signal lines.
After connecting the data transfer cables, clip the interference suppressor (in scope of delivery) onto
the cable cores (not onto the outer insulation of the cable!). You have to route the cable screen
outside around the interference suppressor and connect it to PE (brass screw with bolt) at the top
right in the connection compartment (→ å 18).
3
4
1
2
a0001746
Fig. 18:
1
2
3
4
!
Interference protection of the signal cable
Cable shield (at PE in connection compartment!)
Signal cable
Interference suppressor
Cable cores of the signal cable
Note!
With the two-channel version (data cables at analog output 1 and analog output 2), route the cable
cores of the two data cables through the interference suppressor.
4.2.2
Connection
Signal inputs
One-channel version
Designation
Function
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
AV 1
Alarm value 1 exceeded or undershot
AV 2
Alarm value 2 exceeded or undershot
Fault
Error message, can be retrieved using operation menu
Measurement end
Displays "Measurement finished" (5 s)
l-1 channel 1
0 or 4 mA = start of measuring range
20 mA = end of measuring range
Signal outputs
Analog output
18
Endress+Hauser
Stamolys CA71CL
Wiring
4.2.3
Two-channel version
Connection
Signal inputs
Signal outputs
Designation
Function
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
AV 1 - 1
Alarm value 1, channel 1 exceeded or undershot
AV 2 - 1
Alarm value 2, channel 1 exceeded or undershot
AV 1 - 2
Alarm value 1, channel 2 exceeded or undershot
AV 2 - 2
Alarm value 2, channel 2 exceeded or undershot
Fault
Error message, can be retrieved using operation menu
Channel 1/2 or measuring
end1)
Displays active channel
Displays "Measurement finished" (5 s)
l-1 channel 1
0 or 4 mA = start of measuring range
20 mA = end of measuring range
l-2 channel 2
0 or 4 mA = start of measuring range
20 mA = end of measuring range
Meas. 2
0 V = channel 1
24 V = channel 2
Analog output
Channel selection
1)
Can be selected alternatively
4.3
Switching contacts
One-channel version
Terminal connection for
condition met
Terminal connection for
condition not met
Terminal connection for
power off
AV 1
A: R:
2525
--
2324
A:
R:
25
25
-
24
23
25
-
24
AV 2
A: R:
2828
--
2627
A:
R:
28
28
-
27
26
28
-
27
Fault
A: R:
1919
--
1718
A:
R:
19
19
-
18
17
19
-
18
2216
13
---
2014
11
22
16
13
-
21
15
12
22
16
13
-
21
15
12
Unassigned
Two-channel version
Terminal connection for
condition fulfilled
Endress+Hauser
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1 - 1
A:
R:
25
25
-
23
24
A:
R:
25
25
-
24
23
25
-
24
AV 1 - 2
A:
R:
13
13
-
11
12
A:
R:
13
13
-
12
11
13
-
12
AV 2 - 1
A:
R:
28
28
-
26
27
A:
R:
28
28
-
27
26
28
-
27
AV 2 - 2
A:
R:
16
16
-
14
15
A:
R:
16
16
-
15
14
16
-
15
Fault
A:
R:
19
19
-
17
18
A:
R:
19
19
-
18
17
19
-
18
19
Wiring
Stamolys CA71CL
Terminal connection for
condition fulfilled
Channel 1/2
measurement end
A:
R:
22
22
-
20
21
Terminal connection for
condition not fulfilled
A:
R:
22
22
-
Terminal connection for
power off
21
20
22
-
21
A = NO configured
R = NC configured
!
Note!
Condition met means:
• AV 1: concentration > alarm value 1
• AV 2: concentration > alarm value 2
• Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
4.4
Serial interface
RS 232 of CA 71
COM 1/2 at PC
SUB-D, nine-pin
Function
Function
SUB-D, nine-pin
3
TxD
RxD
2
2
RxD
TxD
3
8
CTS
RTS
7
CTS
8
GND
5
5
Software protocol:
Output format:
GND
9600, N, 8, 1
ASCII
The results (measured value+engineering unit) are output in the "Data memory-Measured values"
menu.
The calibration results (measured value+engineering unit) are output in the "Data
memory-Calibration factors" menu.
!
Note!
• A null modem cable is required (not a crossed one).
• The interface does not have to be configured at the analyzer.
The following commands can be sent from the PC in order to read out data:
• "D" = Data memory-Measured values
• "C" = Data memory-Calibration factors
• "S" = Setup (parameter entry, configuration...)
• "F" = Frequency (current)
20
Endress+Hauser
Stamolys CA71CL
Wiring
4.5
Post-connection check
Carry out the following checks after electrical connection:
Device status and specifications
Note
Are the analyzer or cables damaged externally?
Visual inspection
Electrical connection
Note
Does the supply voltage correspond to the data on the nameplate?
230 V AC / 50 Hz115 V AC / 60 Hz
Are current outputs shielded and connected?
Do the mounted cables have adequate strain relief?
Is the cable type route completely isolated?
Guide power supply and signal lines
separately over the entire travel distance.
Separate cable ducts are ideal.
Cable run without loops and crossovers?
Are power supply and signal lines connected correctly according to wiring
diagram?
Are all screw terminals tightened?
Are all cable entries mounted, tightened and leaktight?
Interference suppressor at the analog output?
Current output simulation
See procedure below
Current output simulation:
1.
Hold both arrow keys down (see "Display and operating elements" chapter) and connect the
analyzer to the power supply. Wait until the display shows "0 mA".
2.
At your PLC, PCS or data logger, check whether the same current value is applied.
3.
Press the
key. This allows you to browse to the next current values (4, 12, 20 mA,
depending on the setting).
4.
Check that the current values in question are also present at your PLC, PCS or data logger.
5.
If the values are not present:
a. Check the terminal assignment for analog output 1 or 2.
F
b. Disconnect the analog outputs from your PLC, DCS or data logger and repeat steps 1-4 but
this time measure the current values at the terminals of the analyzer and not at the PLC,
DSC or data logger.
If these current values are correct, please check your PLC, DCS, the data logger and/or the
electrical cables.
Endress+Hauser
21
Operation
Stamolys CA71CL
5
Operation
5.1
Display and operating elements
1
2
M CE
E
K
4
RS 232
3
a0001679
Fig. 19:
1
2
3
4
Display and operating elements
LED display (measured value)
LC display (measured value + status)
Serial interface RS 232
Operating keys with indicator LEDs
5.2
Local operation
The operating keys and the integrated indicator LEDs have the following functions:
Key
H
K
V
W
F
J
1)
22
Key function
Indicator LED function
– “Auto measuring” option
– Back to the main menu from all sub-menus
Alarm value 1 exceeded
– Backwards in the sub-menu
horizontal, see Appendix
Alarm value 2 exceeded
– Backwards in the main menu (vertical)
– Increase value
Measuring range exceeded
– Forwards in the main menu (vertical)
– Reduce value
Measuring range undershot
– Select item
– Accept value, move forward in submenu (horizontal)1))
Error message must be called up
– Selection in the sub-menu
– Select sign +/-
Cleaning interval running
The number of decimal places can be changed by pressing the F key and the V or W key simultaneously.
Endress+Hauser
Stamolys CA71CL
Operation
5.2.1
Main menu
H
Access the main menu by holding down the
key until AUTO MEASURING is displayed.
The main menu items and information on them are explained in the following table.
Menu item
Display
Info
AUTO MEASURING
Calibration, measurement, rinsing time-controlled actions
PARAMETER ENTRY
Default settings for measuring ranges, alarm values,
calibration, rinsing
CONFIGURATION
Basic settings such as parameters, engineering units,
assignment of analog outputs and alarm values, date,
time, offset values.
LANGUAGE
For selecting the menu language
ERROR DISPLAY
Displays error messages
SERVICE
Foe switching valves and pumps manually
DATA MEMORY 1
Last 1024 measured values for channel 1
DATA MEMORY 2
(two-channel version only)
Last 1024 measured values for channel 2
5.2.2
AUTO MEASURING
The actions "calibration", "measuring" and "rinsing" are triggered by time-control.
The settings for these actions are made in the "PARAMETER ENTRY" menu.
The action in question is displayed on the LC display. The most recently registered concentration
value is displayed until the end of the next measurement.
Otherwise, "wait" is displayed if:
• The time of the first measurement has not yet been reached or
• The measuring interval has not yet elapsed
!
Endress+Hauser
Note!
"Measuring" flashes when the analyzer is ready for the next measurement but has not yet received
the enable signal from the sample collector or the sample conditioning unit.
23
Operation
Stamolys CA71CL
5.2.3
!
CONFIGURATION
Note!
Some settings that can be made in this menu affect the values in the PARAMETER ENTRY menu.
In view of this, run through the CONFIGURATION menu first during initial commissioning.
Menu item
Range of adjustment
(factory settings in bold)
Code number
03
Input 03.
If an incorrect code is entered the program exits the
sub-menu.
Photometer
Depending on specification
CL-A
CL-B
CL-C
CL-D
The setting displays the parameter that is being measured.
This is defined by the product specification and set in this
menu item at the factory. Do not change the value.
Otherwise, you will receive an "Incorrect photometer"
error message.
Default settings
Yes / no
If "yes" is selected, all the settings are reset to the factory
settings.
You must set the current date and time (scroll through
the menu, third-last point).
In the event of a reset, the date for the 1st calibration and
the 1st flush is set to the following day.
Measuring unit
mg/l
The measuring unit selected depends on the type of
photometer. This setting also affects the scope of the
measuring range.
Calibration factor
0.10 to 10.00
The calibration factor is the ratio of the measured
concentration of the calibration standard to the
predefined concentration of the standard (see
"PARAMETER ENTRY", calibration solution). The
deviation results from factors such as reagent aging, aging
of constructive components, etc.
The calibration factor compensates for these effects. CA
71 checks the registered calibration factor logically. If the
factor is outside of the error tolerance, the calibration is
automatically repeated. If the value is still outside the
tolerance range after repeating calibration, an error
message appears and the analyzer continues to work with
the most recently registered, logically correct factor.
The last 100 calibration factors are stored in the memory
with the date and time and can be retrieved by pressing
the J key.
The calibration factor can be changed manually.
Optional dilution modules are taken into consideration in
the calibration factor.
Concentration offset
0.00 to 50.0 mg/l
The offset gives users the option of adapting to an
external comparison measurement.
(Change the sign with the J key.)
0.10 to 10.00
1.00
If the sample is diluted externally between taking the
sample and the analyzer, the dilution factor has to be
entered here.
If an optional dilution module is used, the factor remains
1. The dilution is taken into consideration by means of
the calibration factor.
20 to 300 s
80 s
Dosing time for sample or standard. During this time, the
entire system is flushed with sample or standard, so when
the reagent is added there is definitely only fresh sample
in the mixer. If sufficient sample is available, or in the
event of an external collecting vessel with a long supply
line, select the highest possible value.
Dilution
Delay to sample
24
Display
Info
Endress+Hauser
Stamolys CA71CL
Menu item
Operation
Range of adjustment
(factory settings in bold)
Display
Info
Selection for the measuring range scope for channel 1. If
the concentration measuring range is 0 to x mg/l, either
4 mA or 0 mA corresponds to 0 mg/l. The end of the
measuring range is the same in both cases at 20 mA.
Analog output 1
Analog output 2
Only two-channel version!Selection for the measuring
range scope for channel 2. The scopes of measuring range
are independent of each other for channel 1 and channel
2 and are determined by the start of measuring range
setting (channel 1 / channel 2) or end of measuring range
setting (channel 1 / channel 2) in the PARAMETER
ENTRY menu.
Alarm value AV 1-1
Setting as to whether the contact for alarm value 1,
channel 1 works as an NO or NC contact.
Alarm value AV 2-1
Setting as to whether the contact for alarm value 2,
channel 1 works as an NO or NC contact.
0 to 20 mA / 4 to 20 mA
NO
NC
Alarm value AV 1-2
!
Note!
Changes only take effect after a reset
(power off/on)!
Two-channel version only!
Setting as to whether the contact for alarm value 1,
channel 2 works as an NO or NC contact.
Alarm value AV 2-2
Two-channel version only!
Setting as to whether the contact for alarm value 2,
channel 2 works as an NO or NC contact.
Error contact
Setting as to whether the contact for the fault message
works as an NO or NC contact.
Current date/time
01.01.96 00:00 to 31.12.95 23:59
Setting the system clock.
Format DD.MM.YY hh:mm.
Calibrate offset
Yes / no
Frequency offset1)
Pressing the J key starts blank value measurement for
compensating the reagent's inherent color.
Frequency offset
– 5000 to +5000
0
Manually changing the frequency offset.1)
1)
Redetermine the frequency offset every time the reagents are changed or the photometer is replaced. To do so, instead of sample connect deionized water to
the sample inlet to determine the frequency offset (blank value). The value is generally between 0 and 10 Hz.
Endress+Hauser
25
Operation
Stamolys CA71CL
5.2.4
Menu item
Measuring range
start 1
Measuring range
start 2
Measuring range
end 1
Measuring range
end 2
PARAMETER ENTRY
Range of adjustment
(factory settings in bold)
CL-A: 0.010 to 1.000 mg/l / 0 mg/l
CL-B: 0.010 to 1.000 mg/l / 0 mg/l
CL-C: 0.10 to 10.00 mg/l / 0 mg/l
CL-D: 0.10 to 10.00 mg/l / 0 mg/l
CL-A: 0.010 to 1.000 mg/l / 1 mg/l
CL-B: 0.010 to 1.000 mg/l / 1 mg/l
CL-C: 0.10 to 10.00 mg/l / 10 mg/l
CL-D: 0.10 to 10.00 mg/l / 10 mg/l
Display
Info
The specified concentration is allocated a value of 0 or 4
mA at analog output 11) .
Only two-channel version!The specified concentration
is allocated a value of 0 or 4 mA at analog output 2.
The specified concentration is allocated a value of 20 mA
at analog output 1.
Only two-channel version!The specified concentration
is allocated a value of 20 mA at analog output 2.
Alarm value
AV 1 - 1
CL-A: 0.010 to 1.00 mg/l / 0.5 mg/l
CL-B: 0.010 to 1.00 mg/l / 0.5 mg/l
CL-C: 0.10 to 10.00 mg/l / 5 mg/l
CL-D: 0.10 to 10.00 mg/l / 5 mg/l
Concentration threshold value for limit relay 1, channel 1
(switching hysteresis 2% of alarm value).
Alarm value
AV 2 - 1
CL-A: 0.010 to 1.000 mg/l / 1 mg/l
CL-B: 0.010 to 1.000 mg/l / 1 mg/l
CL-C: 0.10 to 10.00 mg/l / 10 mg/l
CL-D: 0.10 to 10.00 mg/l / 10 mg/l
Concentration threshold value for limit relay 2, channel 1
(switching hysteresis 2% of alarm value).
Alarm value
AV 1 - 2
CL-A: 0.010 to 1.00 mg/l / 0.5 mg/l
CL-B: 0.010 to 1.00 mg/l / 0.5 mg/l
CL-C: 0.10 to 10.00 mg/l / 5 mg/l
CL-D: 0.10 to 10.00 mg/l / 5 mg/l
Only two-channel version!Concentration threshold
value limit relay 1, channel 2 (switching hysteresis 2% of
alarm value).
Alarm value
AV 2 - 2
CL-A: 0.010 to 1.000 mg/l / 1 mg/l
CL-B: 0.010 to 1.000 mg/l / 1 mg/l
CL-C: 0.10 to 10.00 mg/l / 10 mg/l
CL-D: 0.10 to 10.00 mg/l / 10 mg/l
Only two-channel version!Concentration threshold
value limit relay 2, channel 2 (switching hysteresis 2% of
alarm value).
Time
1. measurement
01.01.96 00:00 to 31.12.95 23:59
Date format DD.MM.YY, time hh.mm. After each change
the instrument does not wait for the measuring interval. If
the measurement is to start immediately, set the time in
the past.
Measuring interval
2 to 120 minutes
10
Time between two measurements.
If the setting is 2 minutes, the measurements take place
without any pauses.
Frequency of
measurement
Channel 1
0 to 9
12)
Two-channel version only!
Number of measurements at channel 1 before
switching to channel 2.
Frequency of
measurement
Channel 2
0 to 9
1
Two-channel version only!
Number of measurements at channel 2 before
switching to channel 1.
Time
1. calibration
01.01.96 00:00 to 31.12.95 23:59
Calibration interval
0 to 720 h
0h
Calibration solution
CL-A: 0.010 to 1.000 mg/l / 0.01 mg/l
CL-B: 0.010 to 1.000 mg/l / 0.01 mg/l
CL-C: 0.10 to 10.00 mg/l / 0.01 mg/l
CL-D: 0.10 to 10.00 mg/l / 0.01 mg/l
26
"
Caution!
Automatic calibration is not possible for chlorine since
the standard only lives for a few hours.
For this reason, do not change the factory setting in these
menus!
The value of 0 h for the calibration interval means that
calibration does not take place.
Endress+Hauser
Stamolys CA71CL
Menu item
Operation
Range of adjustment
(factory settings in bold)
Display
Info
Time of 1st flushing (DD.MM.YY, time hh.mm). After
each change the instrument does not wait for the flushing
interval. If the flushing is to start immediately, set the
time in the past.
1. flushing
01.01.96 00:00 to 31.12.95 23:59
Flushing interval
0 to 720 h
48 h
Time between two flushings.
The "0 h" setting means that cleaning no longer takes
place.
Flushing hold on
0 to 60 s
60 s
Dwell time of flushing solution in the
pump-mixer-photometer line.
Recommended: 30 to 60 s.
– Set the time to 4:00 to track the effects of flushing in
the curve.
– If you have started flushing manually, you should
redefine the time of the 1st flushing because the
interval is calculated starting from the last flushing.
1)
Option 0 or 4 mA: see "CONFIGURATION".
2)
All channels set to 0 means that the channel is selected externally. All channels set to 1 means alternating, beginning with channel 1.
!
Note!
• Carry out flushing with standard cleaners approx. 3-4 hours before the next calibration.
• Flushing with special cleaners (e.g. hydrochloric acid or ammonia) has a lasting effect on
calibration. Therefore, carry out this cleaning 3-4 hours before calibration. Then rinse with
deionized water afterwards.
5.2.5
LANGUAGE
The following languages are available:
• Deutsch
• English
• Français
• Espanol
• Suomi
• Polski
• Italiano
5.2.6
!
Endress+Hauser
ERROR DISPLAY
Note!
• This menu is a "Read-Only-Menu".
• You can find the individual error messages, their meaning and solutions to problems in the
"Troubleshooting instructions" chapter.
• If at least one error message is present, the signal output is set to "fault".
• Causes of faults are queried with every measurement. If an error which occurred previously no
longer exists, the message is automatically deleted. If this does not happen automatically, error
messages can be deleted by quickly switching the analyzer off and back on again.
27
Operation
Stamolys CA71CL
5.2.7
SERVICE
Menu item
Display
Info
"Virtual switching board"
Various valve and pump combinations can be selected.
The setting options are:
– Valve 1: P (sample) or S (standard)
– Valve 2: S (standard) or R (cleaner)
– Valve 3 (two-channel version only): 1 (channel 1) or
2 (channel 2)
– Valve 4 (to empty the cell, thereby achieving better
cleaning and minimizing memory effects):
z (closed) or a (open)
– Pump 1 (sample or standard/cleaner) and pump 2
(reagents): s (stop) or g (go)
– Mixture:
The reagent and sample pumps can be switched on
together so that they run in the same ratio as for filling
the sample-reagent mixture in measurement mode.
s (stop) and g (go).
Pumps and valves
!
Note!
G can only be selected if P1 and P2 are "Off".
The following valve combinations are possible: (applies to
one-channel and two-channel version, whereby, in the
latter, the user chooses between channel 1 and 2 by
means of valve 3)
– V1: P, V2: S
Passage for the sample. This combination is
automatically reset on leaving the service menu.
– V1: S, V2: S
Passage for standard solution
– V1: S, V2: R
Passage for cleaner
Signal frequency
5.2.8
!
DATA MEMORY-measured values
Note!
Two menus - "DATA MEMORY 1" (measured values) and "DATA MEMORY 2" (calibration data,
see the following section) - are only available in the two-channel version. In the one-channel
version there is only one menu "DATA MEMORY".
Menu item
28
Signal frequency of the photometer
Display
Info
Measured values
The data memory contains the last 1024 concentration
measured values with date and time. If no values are
available, "Empty set" appears. Browse through the data
sets with the V and W keys.
Serial output
You can output all data sets (in ASCII-format) via the
serial interface. For this, the PC must be configured as
follows: 9600, N, 8, 1.
To send data, the PC must send the ASCII character 68
("d").
Clear data
This deletes all data sets.
Endress+Hauser
Stamolys CA71CL
Operation
5.2.9
!
DATA MEMORY-calibration data
Note!
You get to the data memory calibration data by means of the CONFIGURATION menu. In the
"Calibration factor" menu item, press the
key.
J
Menu item
Display
Info
Calibration factor
This data memory contains the last 100 calibration factors
with date and time. If no values are available, "Empty set"
appears. Browse through the data sets with the V and
W keys.
Serial output
only possible via PC!
You can output all data sets (in ASCII-format) via the
serial interface. For this, the PC must be configured as
follows: 9600, N, 8, 1.
To send data, the PC must send the ASCII character 67
("c").
No display
Clear data
This deletes all data sets.
5.3
Calibration
5.3.1
Standard calibration data
The signal strength is processed device-internally as a frequency.
The following table provides an overview of the standard calibration data.
!
Note!
Compare these values to your own data.
After changes in the CONFIGURATION menu and in the event of software updates, you can check
and, if necessary, change the calibration data in the sub-menu.
Measuring range
Endress+Hauser
Concentration [mg/l]
Frequency [Hz]
Free available or total chlorine, lower
measuring range CL-A+B
0.010 to 1.000 mg/l
0.000
0.100
0.200
0.300
0.400
0.500
0.600
0.700
0.800
1.000
0
313
759
1156
1641
2058
2469
2889
3241
3820
Free available or total chlorine, upper
measuring range CL-C+D
0.10 to 10.00 mg/l
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
10.00
0
278
487
649
795
924
1019
1115
1203
1350
29
Operation
Stamolys CA71CL
5.3.2
Calibration example
Since automatic calibration is not possible, you must compute the calibration factor and enter it at
the analyzer.
1.
Take some of the sample out of the collecting vessel to determine the chlorine in the
laboratory.
2.
Make a note of the current measured value of the analyzer.
3.
Perform the laboratory measurement simultaneously.
4.
Calculate the calibration factor (= laboratory value: analyzer measured value).
This value has to be entered at the analyzer.
5.
Hold H down until AUTO MEASURING appears.
6.
Use W to move through the CONFIGURATION menu and press F.
7.
Use F to go to the "Calibration factor" item.
8.
Select the item with F.
9.
Enter the computed calibration factor with the aid of the W and V keys.
10. Press F to accept the value and then press H twice to return to the main menu.
11. Press
F again. This takes you back to the measuring mode.
The new calibration factor is accepted and the analyzer measured value and laboratory measured
value must match.
!
30
Note!
If the measured value and the laboratory value do not match after entering the factor, check the
calculation and correct the factor entered, if necessary.
Endress+Hauser
Stamolys CA71CL
Commissioning
#
!
6
Commissioning
6.1
Function check
Warning!
• Check that all connections have been made correctly. Check, in particular, that all hose
connections are secure so that no leaks occur.
• Ensure that the line voltage corresponds to the voltage specified on the nameplate.
6.2
Switching on
6.2.1
Dry commissioning
Note!
• If possible, let the analyzer warm up in standby mode before commissioning ("Auto measuring"
display). The time can be defined via the "1. measurement" option in the PARAMETER ENTRY
menu.
• At the start of measurement with a cold analyzer, the first measuring results will be incorrect. The
reaction is temperature-dependent and if the temperature is too low the predefined reaction time
is insufficient for a complete reaction. For this reason, never carry out calibration with a cold
analyzer. Wait at least two hours before carrying out calibration.
When the analyzer has been configured and calibrated, the measuring cycle starts automatically.
Configuration is no longer necessary.
To perform initial commissioning or to readjust the device parameters, proceed as follows:
1.
Plug in the analyzer's plug with grounding contact into the socket.
2.
Press the H key until AUTO MEASURING is displayed.
3.
Select the CONFIGURATION menu and configure up to and including the "Current
date/time".
With H you can return to the main menu.
4.
Now work through the PARAMETER ENTRY and SERVICE menus.
With H you can return to the main menu.
5.
Select CONFIGURATION again and use F to go to the "Calibrate offset" menu item.
6.
Connect a vessel containing distilled water to the "Sample" connection and start the frequency
offset (J key). The value determined is displayed and saved.
7.
Then reconnect the sample line.
With H you can return to the main menu.
The analyzer starts automatically (triggered by a control signal or integrated timer) with the
"Calibration", "Measurement", and "Flushing" procedures in accordance with the device parameters
that you have set (1st calibration, 1st measurement, 1st flushing times and the respective intervals
control the procedure).
Settings for
Function
Measurement
Flushing (sample)
Delay to sample
Stabilization
1. measurement (basis)
Fill mixture
Reaction
2nd measurement (plateau)
Flushing (sample)
Duration [s]
3 x 15
20 to 999
8
Range of adjustment
CONFIGURATION / "Delay to sample"
SERVICE / "Pumps and valves"
30
see Technical
data
30
Endress+Hauser
31
Commissioning
Stamolys CA71CL
Function
Cleaning
6.2.2
Pump cleaning solution
Allow to react
Pump cleaning solution
Duration [s]
½ flushing
duration
5
½ flushing
duration
Range of adjustment
PARAMETER ENTRY / "Flushing hold on"
Wet commissioning
Wet commissioning is different to dry commissioning in that in wet commissioning you fill the
reagent lines before the automatic cycles are started.
Proceed as follows:
!
1.
Plug in the analyzer's plug with grounding contact into the socket.
2.
Press the H key until AUTO MEASURING is displayed.
3.
Select the SERVICE menu.
4.
Switch the P2 reagent pump "on" (select P2 with F and set to "g" with V) and leave it running
until you can tell that there are reagents at the T-hose connector. After this, switch P2 "off" (s)
again with W.
5.
Now switch the valves to passage for standard (select V1: S, V2: S; with F or K switch to
"S" with V) and then switch the P1 sample pump "on". Leave the pump running until you can
tell that there is standard at the T-hose connector. Switch P1 back "off".
6.
Now switch the valves to passage for cleaning agent (select V1: S, V2: R; with F or K, switch
to "R" or "S" with V) and then switch the P1 sample pump "on". Leave the pump running until
you can tell that there is cleaning agent at the T-hose connector. Switch P1 back "off".
7.
Now switch the valves to passage for sample (select V1: P, V2: S; with F or K, switch to "P"
or "S" with V) and then switch the P1 sample pump "on". If you notice that sample is present
at the T-hose connector, let the pump run for another 2 minutes. This removes any standard
or cleaner residues.
Then switch P1 "off".
Note!
For the two-channel version, valve V3 must also be set for switchover between channel 1 and
channel 2.
8.
Now proceed as with dry commissioning (from step 2).
6.2.3
Using a dilution module
You can extend the measuring range of your analyzer with the sample dilution module.
How it works:
• The analyzer determines the absorption of the sample at the specified wavelength.
It does not matter whether the sample was diluted previously or not.
• The characteristic of the photometer (--> standard calibration data) refers to the concentrations
of the diluted sample and not to those of the (undiluted) original sample.
• The dilution factor is not taken into account until the measured value is calculated.
• The concentration information you specify for the measuring range and the alarm values in
PARAMETER ENTRY refer to the concentrations of the original sample.
!
32
Note!
You achieve maximum measuring accuracy if you also dilute the standard with the dilution module.
In the event of a 1:10 dilution, you have to use a 10 mg/l standard solution instead of a 1.0 mg/l
standard solution (without a dilution module) for example.
Balance changes to the dilution ratio (e.g. through hose aging) with the calibration. The calibration
factor determined takes into account all deviations from the ideal state through reagents, the
photometer and dilution.
Endress+Hauser
Stamolys CA71CL
Commissioning
Installing the dilution module
The dilution module is supplied ready to use.
Installation:
"
1.
Release the sample pump hose from the hose connector to valve 1 (→ å 20, item 4).
2.
Connect the sample hose of the dilution module (black/black or yellow/blue, depending on
the version) to the hose connector that is now free.
3.
Connect the T-section (item 1) of the dilution module to the input of the sample pump (hose
box K1). Use the original sample pump hose for this purpose.
4.
Fit the dilution module onto the free hose box positions of the sample pump.
5.
Connect an outflow hose (O) to the free outlet of the T-piece (item 1).
Route this hose in the device in such a way that is guided at least 20 cm upwards and then into
the open outlet. This prevents the hose from running dry.
Caution!
Do not connect the outflow hoses of the dilution module and the photometer.
a0006622
Fig. 20:
K1
K2
K3
A
O
P
W
1
2
3
4
Endress+Hauser
Dilution module
Hose box of the original sample pump
Hose box for additional sample (standard)
Hose box for dilution water
Sample supply to photometer
Overflow (in outlet of analyzer)
Sample hose (sample or standard)
Dilution water inlet (to be provided by the client)
T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)
Static mixer 122-012
T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)
Hose connector 3.2 x 1.6 mm (0.13 x 0.06 inch)
33
Commissioning
Stamolys CA71CL
Changing parameters
!
Note!
The following section only lists the parameters where you have to make changes to the standard
settings.
Start with CONFIGURATION.
CONFIGURATION
Delay to sample
Increase the value of the standard setting by 40 seconds since additional sample volume is
drawn in.
PARAMETER ENTRY
Measuring range start 1 /
measuring range start 2
Measuring range end 1 /
measuring range end 2
Enter values that refer to the concentration of the undiluted sample.
If you are using a dilution module with the factor 10, for example, multiply the settings to
date (without the dilution module) by 10.
Alarm values AV
1-1 / 1-2 / 2-1 / 2-2
Calibration solution
34
Enter the original concentration (undiluted) of the standard here.
Endress+Hauser
Stamolys CA71CL
Maintenance
7
"
!
Maintenance
Caution!
You may not carry out any procedures not listed in the following chapters yourself.
This work may only be carried out by Service.
Note!
At the end of any maintenance work, you must perform an F-offset measurement and a calibration.
In this way, you adapt the analyzer to the altered conditions.
7.1
Maintenance schedule
All the maintenance tasks that have to be carried out during normal operation of the analyzer are
explained below. If you are using a sample conditioning unit, e.g. CAT430, coordinate the
maintenance work required for it with that of the analyzer. For this, read the maintenance chapter
in the Operating Instructions in question.
Period of time
Task
Note
Weekly
– Check and note down the calibration factor
(for service purposes)
– Move the valve hoses and spray with silicon
spray (extends the operating time).
CONFIGURATION
Every 2 weeks
– Replace reagents and redetermine the
calibration factor
– see chapter "Replacing reagents"
and CONFIGURATION
– You must release the box of the sample
pump for this purpose.
Monthly
– Flush sample line hose system with
pressurized water (disposable injector),
check and
replace reagents if necessary.
– Flush the sample line hose system with
12.5% bleaching lye (sodium hypochloride)
and rinse thoroughly with water1)
– see chapter "Replacing reagents"
– You must release the box of the sample
pump.
– Connect the disposable injector instead of
the sample supply.
– SERVICE
V1: S, P1: g, P2: s, V2: S
Attach solution to sampling connection.
#
Warning!
Caustic.Wear protective gloves and goggles.
Beware of reagent splashing.
– Spray pump hoses and roller heads with
silicone spray
– Check sample collector for fouling and clean
it if necessary
Every 3 months
– Cleaning drain lines:
Flush all hoses with 10% ammonia solution
and then rinse with sample for 30 minutes at
least
– Rotate pump hoses
Every 6 months
– Replace pump hoses
– Replace valve hoses
1)
!
Endress+Hauser
– see chapter "Replacing pump hoses"
With standard cleaner for CA71SI-A
Note!
Prior to working on the reagent hoses, the hoses must be disconnected from the canisters in order
to prevent contamination of the reagents.
35
Maintenance
Stamolys CA71CL
7.2
#
Replacing reagents
Warning!
• There is a danger of crushing at the door and pump heads.
• Refer to the warning instructions in the safety data sheets when handling reagents. Wear
protective clothing, gloves and goggles.
• Make sure the workplace is well ventilated when you work with chlorine bleach and/or
ammonia. If you feel unwell, consult a physician immediately.
• If reagents come into contact with the skin or eyes, carefully rinse with copious amounts of water
and consult a physician immediately.
• Never add water to reagents. Reagents containing acids may splash and heat may build up.
If you store the reagents correctly (in the dark, not over 20 °C) they will keep (unopened) for a
minimum of 12 weeks from the date of production (batch number). When this period of time has
expired, the reagents must be checked and replaced where necessary. A cool, dark storage area is
required to achieve the maximum shelf life.
The reagents must be replaced when:
• The reagents have been contaminated by sample (see "Troubleshooting instructions")
• The reagents are too old
• The reagents have been spoilt by incorrect storage conditions or environmental influences.
7.2.1
Checking reagents
1.
Produce a standard (1 mg/l Cl2).
2.
Check the concentration of the standard in the laboratory. Adapt the values (PARAMETER
ENTRY, calibration solution") or replace the standard solution.
3.
Mix 10 ml standard solution and 5 ml of each reagent in a beaker. You must replace the
reagents if there is no visible coloration after max. ten minutes.
The finished mixture must be free of particles.
The reagents must be replaced if no visual coloration (blue) occurs after max. 2 minutes or if the
mixture is not free of particles.
7.2.2
1.
Carefully remove the hoses from the canisters and wipe them with a dry (paper) towel. Wear
protective gloves when doing this.
2.
Switch on the reagent pump for about 5 seconds.
3.
Flush the reagent hose with plenty of distilled water (see SERVICE).
For this purpose, place a beaker with distilled water into the device.
4.
Replace the reagent canisters and feed the hoses into the new canisters.
5.
Fill the reagent hose with the new reagents (SERVICE). Switch all pumps to "g" for this
purpose. Do not switch the pumps off ("s") until no more air bubbles are to be seen in the hoses.
6.
Determine the reagent blank value by using deionized water as a sample (see
"CONFIGURATION").
The value determined is saved by the analyzer as the frequency offset.
7.
Then carry out a calibration (see chapter "Calibration").
7.3
#
Replacing reagents
Replacing pump hoses
Warning!
When removing hoses from the hose nozzles, beware of reagent splashing.
For this reason, wear protective clothing, gloves and goggles.
The peristaltic pumps used for the analyzer convey the medium in a combination of vacuum and
displacement pump. The pump rate is dependent on the elasticity of the pump hoses. Elasticity
36
Endress+Hauser
Stamolys CA71CL
Maintenance
decreases and the pump rate drops as mechanical stress increases. Wear depends on mechanical
stress (measuring interval, pump pressure). Up to a certain degree, the wear effect can be
compensated by calibration. If the impairment in hose elasticity is too great, the pump rate is no
longer reproducible and this leads to incorrect measurements. This is why it is necessary to replace
the hoses.
Removing the old hoses
1.
Remove the reagent hoses from the reagent canisters in order to prevent contamination of the
reagents.
2.
Flush the old hoses with water first and then with air to empty them (see SERVICE).
3.
Remove the hoses from the nipples of the hose boxes (→ å 21, item 5).
4.
Release the hose box(es) (1 to 3 per pump):
– Press against the lower retainer (item 3).
– You can now remove the hose box along with the pump hose.
– Remove the old hose from the case any dispose of it.
– Clean the hose box and the roller head (item 1) with water.
1
2
5
3
1
2
3
4
5
Roller head
Upper retainer of hose box
Lower retainer of hose box
Guide at pump hose
Nipple with guide
4
a0001652
Fig. 21:
Reagent pump
Installing the new hoses
1.
Fit the new hose on the hose box.
2.
First pull the hose downwards at both ends and then push the guide on the hose into that of
the nipple on the hose box. Make sure it is seated correctly.
3.
First place the hose box into the upper retainer (item 2) of the pump and then press the case
into the lower retainer (item 3). Make sure the hose boxes are in the correct order in the pump
(→ å 22).
R3
R2
R1
a0002206
Fig. 22:
R1
R2
R3
4.
Endress+Hauser
Reagent pump, top view
Reagent 1
Reagent 2 (if present)
Reagent 3 (if present)
Spray the new pump hoses, the hose boxes and the roller heads with silicone spray.
37
Maintenance
Stamolys CA71CL
"
5.
Reconnect the reagent hoses to the canisters.
6.
After installation, refill the hoses with sample, standard or cleaner (SERVICE)1).
7.
Determine the frequency offset (CONFIGURATION) and carry out calibration ("Calibration").
Caution!
Make sure you connect the new pump hoses to the correct connections at the T-hose connector.
For the order numbers of the pump hoses, see the chapter "Troubleshooting"/"Spare parts".
Setting the contact pressure of the pump
If the pump hoses cannot be filled free of bubbles, adjust the adjuster screw for the contact pressure
of the pump:
1.
Release the adjuster screw (→ å 23, item 3) to the extent that no more sample is pumped.
2.
Tighten the screw just to the point that sample is pumped.
3.
Tighten the screw another complete rotation.
3
a0001651
Fig. 23:
3
"
Sample pump
Adjuster screw for the contact pressure
Caution!
Adjust the contact pressure of the reagent hoses so that no sample is pumped into the reagent.
Otherwise the reagent becomes immediately unusable.
For this reason, only perform tests with distilled water.
7.4
Replacing valve hoses
To replace the hoses, proceed as follows:
1.
Flush the old hoses with water first and then with air to empty them (see SERVICE).
2.
Remove the hoses from the valves:
a. You can disconnect the front hoses directly because the valves are open when de-energized
b. To remove the back hoses, press the black button on the valve and disconnect the hoses.
1)
38
3.
Spray the new hoses and valves with silicone spray before inserting them.
4.
Reverse the sequence of operations to install the new hoses.
Ensure that the hoses are connected correctly (→ å 24).
5.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
6.
Determine the frequency offset (CONFIGURATION) and carry out calibration ("Calibration").
Make sure that the hoses are bubble-free when filled. If not, check the sample flow rate (see below).
Endress+Hauser
Stamolys CA71CL
Maintenance
1
3
2
V3
7
V4
V2
V1
6
5
4
a0001730
Fig. 24:
V1-4
V3
1
2
3
4
5
6
7
7.5
Valves and valve hoses
Valves 1, 2 and 4 (if present)
Channel switching
Sample
To the pump
Y-piece, hose connection to valve 1, at rear
Outlet hose
Hose valve 2 at front, standard
Hose valve 2 at rear, cleaner
Hose valve 1 at front, sample
Replacing the static mixer
Proceed as follows to replace the mixer (see "Troubleshooting/spare parts"):
1.
Rinse first with water and then with air (see SERVICE).
2.
Unscrew the four screws on the photometer housing and remove it.
3.
Disconnect the mixer from the photometer and from the T-section under the photometer
housing or release the mixer from the retainer.
4.
Remove the old mixer and insert the new one.
5.
Connect the new mixer to the photometer and the T-section again.
6.
Attach the photometer housing and screw it down.
7.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
8.
Carry out a calibration (PARAMETER ENTRY).
7.6
"
Replacing the photometer cell
Caution!
Using electronic assemblies
Electronic assemblies are sensitive to electrostatic discharge (ESD). For this reason, discharge
yourself at a protective earth before handling the assemblies.
Removing the old cell
Endress+Hauser
1.
Rinse first with water and then with air (see SERVICE).
2.
Switch the analyzer off.
3.
Unscrew the four screws on the photometer housing and remove it.
4.
Release the nuts from the guide screws and remove the photometer completely.
5.
Unscrew the four screws on the side of the photometer on which there is no ribbon cable.
6.
Take the photometer's electronics apart.
7.
Take out the cell and remove the hoses.
39
Maintenance
Stamolys CA71CL
"
Caution!
Do not under any circumstances touch the optical window of the cell with your fingers!
Otherwise, traces of grease remain on the optical surfaces. This can lead to corrupted measured
values.
Installing the new cell
1.
Insert the new cell.
2.
Connect the cell to the hoses such that the sample is fed in from below.
3.
Secure the hoses with the supplied cable connectors to stop the cell from slipping.
4.
Reassemble the photometer and tighten the screws and nuts.
5.
Attach the photometer housing and screw it down.
6.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
7.
Carry out an f-offset measurement (CONFIGURATION) and a calibration ("Calibration").
7.7
"
Cleaning
Caution!
When cleaning, ensure that you do not damage the nameplate on the analyzer. Do not use any
solvent-based cleaning agents.
To clean the analyzer housing, proceed as follows:
• Stainless steel housing (stainless steel SS 1.4301 (AISI 304)):
With a lint-free cloth and Glittol RG 10.51
• GFR housing:
With a damp cloth or with tenside-based (alkaline) cleaning agent.
7.8
Placing out of service
You must place the analyzer out of service before shipping or before extended breaks in operation
(more than 5 days).
"
Caution!
Before placing the instrument out of service, thoroughly rinse all the lines with clean water.
Proceed as follows:
!
40
1.
Remove the reagent and standard hoses from the canisters and immerse them in a tank
containing clean water.
2.
Switch valve 1 to "Standard" and switch pumps 1 and 2 on for one minute (SERVICE menu).
3.
Remove the hoses from the water and allow the pumps to run until the hoses are completely
dry.
4.
If you are using a continuous sample supply:
Disconnect the sample line.
5.
Flush the sampling hoses with clean water and then with compressed air to empty the hoses
completely.
6.
Remove the valve hoses from the valves.
7.
Release the pump hoses by removing the hose box from the holder from below.
Note!
Store open reagent and standard containers in the refrigerator. Bear the shelf life in mind.
Endress+Hauser
Stamolys CA71CL
Accessories
8
!
Accessories
Note!
The following section provides you with information on the accessories which could be supplied at
the time this documentation was issued.
Please contact your local Service Center for accessories that are not listed here.
8.1
Collecting vessel
– For sampling from pressurized systems
– Results in an unpressurized continuous sample stream
• Collecting vessel without level monitoring; Order No. 51512088
• Collecting vessel with level monitoring (conductive); Order No. 51512089
• Level monitoring retrofit kit; Order No. 71023419
8.2
Reagents, cleaners, standard solutions
• Reagent set for free available chlorine, active, 1 l each of CL1+CL2; Order No. CAY543-V10AAE
• Reagent set for free available chlorine, inactive, 1 l each of CL1+CL2; Order No.
CAY543-V10AAH
• Reagent set for total chlorine, active, 1 l each of CL1+CL2; Order No. CAY546-V10AAE
• Reagent set for total chlorine, inactive, 1 l each of CL1+CL2; Order No. CAY546-V10AAH
• Cleaner, 1 l; Order No. CAY544-V10AAE
!
Note!
Please note the disposal instructions in the safety data sheets of the reagents!
8.3
Maintenance kit
• Maintenance kit CAV 740:
– 1 set of pump hoses, yellow/blue
– 1 set of pump hoses, black/black
– 1 pack of hose connectors each
Order No. CAV 740-1A
• Maintenance kit CAV740:
– 1 set of pump hoses, violet/white
– 1 set of pump hoses, black/black
– 1 pack of hose connectors each
Order No. CAV740-4A
8.4
Hose cleaner
• Alkaline cleaner, 100 ml (3.4 fl.oz.); Order No. CAY746-V01AAE
• Acidic cleaner, 100 ml (3.4 fl.oz.); Order No. CAY747-V01AAE
8.5
Dilution module
Dilution module
1 set of pump hoses, 2 hose boxes, 1 set of hose connectors, static mixer
• Dilution 1:3
Order No. C-A030228-10
• Dilution 1:10
Order No. C-A030228-11
Endress+Hauser
41
Accessories
Stamolys CA71CL
8.6
Additional accessories
• Interference suppressor for control, power and signal lines
Order No. 51512800
• Silicone spray,
Order No. 51504155
• Valve set, 2 pieces, for two-channel version,
Order No. 51512234
• Upgrade kit for upgrading from one-channel to two-channel version,
Order No. 51512640
42
Endress+Hauser
Stamolys CA71CL
Troubleshooting
9
Troubleshooting
9.1
Troubleshooting instructions
Although the analyzer is not very prone to faults due to its simple design, problems at the measuring
point can, of course, not be completely ruled out.
Possible errors, their causes and their possible remedies are thus listed below.
9.1.1
Error messages
Error message
Possible cause
Tests and / or corrective measures
Cell dirty
Not enough light intensity at
receiver, e.g. due to sedimented
particles
– Flush with 12.5% bleaching lye (monthly maintenance)
– When using a CAT 430:
Check the filter.
Wrong cell
Wrong cell
Check the setting in the CONFIGURATION menu, "Photometer".
No sample
Establish sample transfer.
Level measurement defective
Check level measurement at the sample collector.
Leak at canisters or hoses
Replace defective components and clean and dry the analyzer or those components affected by the
leak.
Photometer defective
Inform E+H Service
Electrical connection
Check all electrical connections.
Fuse defective
Replace fuse F4 or F5 (semi-delay 0.2 A)
No sample
Spillage error
No measuring signal
9.1.2
Error
Measured values always
the same
Measured values
inaccurate
Endress+Hauser
System errors/process errors without messages
Possible cause
Tests and / or corrective measures
Reagents contaminated
or aged
Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not change color
after max. 10 min, replace the reagents.
No sample, no reagents
Make sure sample and reagents are supplied, check level monitor and clean if necessary
System blocked
Flush with 12.5% bleaching lye (monthly maintenance)
Incorrect concentration
of standard
Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER ENTRY",
"Calibration solution") or replace the standard.
Reagents contaminated
or aged
Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not change color
after max. 10 min, replace the reagents.
Reagent blank value too
high
After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,
"Offset calibration")
Wrong dimension
Check the setting in the CONFIGURATION menu, "Measuring unit" (e.g. ion instead of element).
Wrong cell
Check the setting in the CONFIGURATION menu, "Photometer"
Sample suction time too
short
Increase the suction time (CONFIGURATION, "Delay to sample")
Detect interfering substances (see Technical Information, "Measuring principle"),
Matrix effects
(substances that interfere possibly use sample conditioning
with the photometric
method)
Filter life too long
Take a counter sample at the analyzer inlet and check the concentration in the laboratory. If there is no
deviation to the analyzer measured values, replace the ultrafiltration modules or reversible flow filters more
frequently.
System blocked or
contaminated
Flush with 12.5% bleaching lye (monthly maintenance)
Dosing
Replace pump hoses.
Cell dirty
Flush with 12.5% bleaching lye first and then with 5% hydrochloric acid
43
Troubleshooting
Stamolys CA71CL
Error
Possible cause
Tests and / or corrective measures
Counter sample in the
laboratory delivers
deviating measured
values
Sample aging
Shorten time between sampling and analysis.
Wrong size for analog
output
Check the setting (CONFIGURATION, "Analog output 1" or " 2").
Incorrect measuring
range
Adjust measuring range (PARAMETER ENTRY, "Measuring range")
Background noise
Check electrical cables for interference from strong sources of induction.
No power
Check the electrical connection and ensure power supply.
Fuse
Replace fuse F1 (slow-blow 0.5 A)
Measured value transfer
defective
Analyzer will not switch
on
Analyzer is running but Initialization failed
display reading is garbled
or off
Switch the analyzer off and after approx. 30 seconds back on.
Leak
See "Spillage error" error message
Leak sensor bypassed
Interrupt the contact between the two leak sensors (terminals 67-66)
Fuse
Check all fuses and replace if necessary.
Pump defective
Service
Leak at photometer
Service
Time of 1st
measurement not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired
Change parameter setting.
Time of 1st flushing not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired or 0
h
Change parameter setting.
Device or discharge
blocked
Remove blockage.
Service
Water hardness
Lime deposits can be removed by flushing with 5% hydrochloric acid.
If necessary, dose EDTA in sample stream to prevent precipitation (EDTA not with CA71HA).
Not enough sample
conditioning
Shorten the cleaning intervals of sample conditioning.
Pumps fail to run
Measurement does not
start
"Measurement" display
flashes
Flushing does not start
Leak at photometer
Blockage, deposits in
device
44
Endress+Hauser
Stamolys CA71CL
Troubleshooting
9.2
Spare parts
9.2.1
Overview for
155
131
V3
160
B
121
160
144
200
154
V1
143
161
V2
V1
A
130
161
134
V2
130
141
170/171
S
R
R1
R2
121
B
135
135
C
142
123
122
120
a0001375
Fig. 25:
Sample input, two-channel version
Standard and reagent canisters
Collecting vessel
Sample
A
B
C
P
9.2.2
Item
Endress+Hauser
Overview of components and spare parts
R
R1
R2
S
Cleaner canister
Reagent 1 canister
Reagent 2 canister
Standard solution canister
Spare parts for sample and reagent transportation
Spare part
Order number
120
Norprene hose, 1.6 mm
51504116
121
C-Flex hose, 3.2 mm (permeate inlet and overflow with sample collector.)
51504114
122
C-Flex hose ID 6.4 mm
51504115
123
C-Flex hose, 1.5 mm
51512535
130
Hose connector 1.6 mm x 1.6 mm (10 pieces)
51506495
131
T-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)
51506490
134
Y-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)
51512096
135
Connection nipples for sample collector (10 piece)
51512099
136
T-hose connector 3.2 mm x 3.2 mm x 3.2 mm (10 pieces)
51516166
140
Mixer set, complete (2 pieces)
51512101
141
Collecting tank
51512102
142
Outlet pipe with hose nipple (2 pieces)
51512104
143
Leak sensor for collecting tank
51512103
154
Valve, complete
51512100
155
Valve set for two-channel version
51512234
45
Troubleshooting
Stamolys CA71CL
Item
Order number
160
Roller head with holder for hose pump
51512085
161
Hose box for pump
51512086
170
Collecting vessel with level monitoring
51512089
171
Collecting vessel without level monitoring
51512088
9.2.3
Analyzer spare parts
Item
Spare part
130-133
Maintenance kit CAV740 (CA71CL-C+D):
– 1 set of pump hoses, blue/yellow (item 113, 12 piece 51506434)
– 1 set of pump hoses, black/black (12 piece 51506497)
– 1 pack of hose connectors each
CAV740-1A
Maintenance kit CAV740 (CA71CL-A+B):
– 1 set of pump hoses, violet/white (item 113, 12 piece 51508945)
– 1 set of pump hoses black/black (12 piece 51506497)
– 1 pack of hose connectors each
CAV740-4A
144
200
9.3
46
Spare part
Order number
Photometer cell
– CL-A+B
– CL-C+D
51513566
51505776
Photometer
– CL-A+B
– CL-C+D
51513230
51513231
Software history
Date
Version Changes in the software
Documentation
08/2006
6.0
Extension
• New parameter FE-D
BA360C/07/xx/07.06
06/2006
5.9
Extension
• New parameter NO-D
• Stepping motor position is also reset in a reset
• New stepping motor setting for CU-A/B (50 rpm)
• Spanish replaces Dutch and Hungarian
BA355C/07/xx/07.06
09/2005
5.8
Extension
• New parameter PH-E
• Altered factory settings for SI-A and AM-B
• Characteristics for FE-B, CU-A and CR-A extended
• Measuring range limitation increased to 50000 for all
parameters and engineering units
• Calibration value can be adjusted to end of measuring range
BA356C/07/xx/07.06
06/2005
5.7
Improvement
• Stack overflow eliminated
• FE-C valve control corrected
• RTC properly dealt with when setting/reading the time
-
07/2004
5.6
Extension
• Range of adjustment for delay to sample to 999s
• Range of adjustment for cleaning duration to 300s
• Range of adjustment for frequencies to 5800 Hz
• Factory settings accepted as specified
• Altered measuring cycles for AM-D, AL-A (like hardness)
BA353C/07/xx/09.04
BA357C/07/xx/10.04
Endress+Hauser
Stamolys CA71CL
Troubleshooting
Date
Version Changes in the software
Documentation
05/2004
5.5
Extension/improvement
• Longer initialization
• Altered measuring cycle for SI
• Shortened reaction time, AM-C (90 s)
• SI-A: 0 Hz signal error only after 90 s
• Reason for crashing during frequency display in SERVICE
eliminated
• Restart after power failure
BA364C/07/xx/06.04
03/2004
5.4
Extension/improvement
• New process: HA
• Menu extension: Cell drain time
• Calibration valve off during reaction
• Hungarian and Spanish no longer available
• New hardness engineering units
• New standard parameters/factory settings
• Stepping motor activation corrected
BA361C/07/xx/02.04
08/2003
5.3
Improvement
• Fault elimination for measuring range switchover, HA
-
08/2003
5.2
Extension/improvement
• Factory settings for SI changed
• CL characteristic changed
• New startup message
• Fault elimination for output for current output, two-channel
-
05/2003
5.1
Extension/improvement
• New parameter CL
• Memory correction
• Time for "Delay to sample" is used for zero measurement
• Factory settings for MN, SI, HA changed
BA354C/07/07.03
12/2002
5.0
Original software
BA353C/07/xx/01.03
BA355C/07/xx/01.03
BA356C/07/xx/01.03
BA357C/07/xx/01.03
BA358C/07/xx/01.03
BA359C/07/xx/01.03
BA360C/07/xx/01.03
BA361C/07/xx/01.03
BA362C/07/xx/01.03
BA363C/07/xx/01.03
BA364C/07/xx/07.03
9.4
Return
If the analyzer requires repair, please send it cleaned to your sales center.
Please use the original packaging when returning the device.
Please enclose the completed "Declaration of Contamination and Cleaning" form (copy the second
last page of these Operating Instructions) with the packaging and also the shipping documents. The
unit cannot be repaired if this declaration is not duly completed!
9.5
Disposal
Electronic components are used in the product. For this reason, the product must be disposed of as
electronic waste.
Please observe local regulations.
Endress+Hauser
47
Technical data
Stamolys CA71CL
10
Technical data
10.1
Input
Measured variable
Free available or total Cl2 [mg/l]
Measuring range
• CL-A+B
0.01 to 1.00 mg/l
• CL-C+D
0.10 to 10.00 mg/l
Wave length
555 nm
Reference wave length
880 nm (only CL-C+D)
10.2
Output
Output signal
0/4 to 20 mA
Signal on alarm
Contacts: 2 limit contactors (per channel), 1 collection alarm contact
Optionally: end of measurement (with two channel version, display of channel
no. available)
Load
Max. 500 Ω
Data interface
RS 232 C
Load capacity
230 V / 115 V AC max. 2 A, 30 V DC max. 1 A
10.3
Power supply
Supply voltage
115 V AC / 230 V AC ±10%, 50/60 Hz
Power consumption
Approx. 50 VA
Current consumption
Approx. 0.2 A at 230 V
Approx. 0.5 A at 115 V
Fuses
1 x slow-blow 0.5 A for analyzer electronics
2 x semi-delay 0.2 A for photometer
10.4
Performance characteristics
Time between two measurements tmeas = reaction time + rinse time + waiting time + post rinse time + filling
time+delay to sample + reject reagent (min. waiting time = 0 min)
48
Maximum measured error
±2 % of end of measuring range
Measuring interval
tmeas ... 120 min
Reaction time
• 1 minute
Sample requirement
• 15 ml (0.51 fl.oz.) / measurement
Reagent requirement
CL-A/B: 2 x 0.56 ml (0.019 fl.oz.)
CL-C/D: 2 x 0.30 ml (0.01 fl.oz.)
2 x 2.6 l (0.69 gal.) (CL-A/B) or
2 x 1.3 l (0.34 gal.) (CL-C/D) reagents per month
Calibration interval
0 to 720 h
Flushing interval
0 to 720 h
Rinse time
Selectable from 20 to 300 s (standard = 60 s)
Post rinse time
30 s
Endress+Hauser
Stamolys CA71CL
Technical data
Filling time
20 s
Maintenance interval
6 months (typically)
Servicing requirements
15 minutes / week (typically)
10.5
Ambient temperature range
5 to 40 °C (40 to 100 °F), severe fluctuations must be avoided
Humidity
Below the condensation limit, installation in usual, clean rooms
Outdoor installation only possible with protective devices (customer-supplied)
Degree of protection
IP 43
10.6
Process
Sample temperature
5 to 40 °C (40 to 100 °F)
Sample flow rate
Min. 5 ml/min (0.0013 gal/min)
Consistency of the sample
Low solid content (dry matter < 50 mg/l (ppm))
Sample inlet
Unpressurized
10.7
Endress+Hauser
Environment
Mechanical construction
Design, dimensions
See chapter "Installation"
Weight
GFR housing
Stainless steel housing
Without housing
Approx. 28 kg (62 lbs)
Approx. 33 kg (73 lbs)
Approx. 25 kg (55 lbs)
Materials
Housing
Front windows
Endless hose
Pump hose
Valves
Stainless steel 1.4301 (AISI 304) or GFR
Polycarbonate
C-Flex®, Norprene®
Tygon®, Viton®
Tygon®, silicone
49
Appendix
Stamolys CA71CL
11
Appendix
11.1
Operating matrix
One-channel version
a0001907-en
50
Endress+Hauser
Stamolys CA71CL
Appendix
a0001908-en
Endress+Hauser
51
Appendix
Stamolys CA71CL
Two-channel version
a0001909-en
52
Endress+Hauser
Stamolys CA71CL
Appendix
a0001910-en
Endress+Hauser
53
Appendix
Stamolys CA71CL
11.2
Ordering forms
11.2.1
Reagents and accessories
To Fax No:
Fax for ordering reagents
To (address of your sales center)
From (billing address)
Company:
Re:
Street:
Zip code / Town:
Fax / Telephone:
Address for delivery (if not the above address)
Company / Name:
Street / Zip code / Town:
Quantity
Quantity
Quantity
Article
Designation
CAV740-1A
Maintenance kit CA 7X
• 1 set of pump hoses, yellow/blue (12 piece)
• 1 set of pump hoses, black/black (12 piece)
• 1 pack of hose connectors each, item 130, 131, 132, 133, 134 (10 pieces each)
• C-Flex hose for valves 1.40 m
Article
Designation
CAV740-4A
Maintenance kit CA 7X
• 1 set of pump hoses, violet/white (12 piece)
• 1 set of pump hoses, black/black (12 piece)
• 1 pack of hose connectors each, item 130, 131, 132, 133, 134 (10 pieces each)
Order No.
Designation
CAY543-V10AAE
Reagent set, free available chlorine, active, 1 l CL1+CL2 each
CAY543-V10AAH
Reagent set, free available chlorine, inactive, 1 l CL1+CL2 each
CAY546-V10AAE
Reagent set, total chlorine, active, 1 l CL1+CL2 each
CAY546-V10AAH
Reagent set, total chlorine, inactive, 1 l CL1+CL2 each
CAY544-V10AAE
Cleaner, 1 l
Chemicals for ultrafiltration
Quantity
Order No.
Designation
CAY746-V01AAE
Alkaline cleaner P3-Ultrasil 130, 100 ml
CAY746-V10AAE
Alkaline cleaner P3-Ultrasil 130, 1 l
CAY746-V50AAE
Alkaline cleaner P3-Ultrasil T130, 5 l
CAY747-V01AAE
Acidic cleaner P3-Ultrasil 130, 100 ml
CAY747-V10AAE
Acidic cleaner P3-Ultrasil 130, 1 l
CAY747-V50AAE
Acidic cleaner P3-Ultrasil 130, 5 l
Place
Date
Earliest delivery one week after order received. Delivery is unpaid.
54
Stamp/Signature
Endress+Hauser
Stamolys CA71CL
Appendix
11.2.2
Ordering wear parts
To Fax No:
Fax for ordering wear parts
To (address of your sales center)
From (billing address)
Company:
Re:
Street:
Zip code / Town:
Fax / Telephone:
Address for delivery (if not the above address)
Company / Name:
Street / Zip code / Town:
Quantity
Item
110
1)
Pieces/pack.
Designation
Order number
12
Tygon pump hose, yellow/blue
51506434
111
12
Tygon pump hose, black/black
51506437
113
12
Tygon pump hose, violet/white
51508945
120
15 m
Norprene hose, ID 1.6 mm
51504116
121
7.5 m
C-Flex hose, ID 3.2 mm (permeate inlet and overflow with sample collector.)
51504114
122
7.5 m
C-Flex hose, ID 6.4 mm
51504115
123
1 m (3.28 ft)
C-Flex hose, ID 1.5 mm
51512535
130
10
Hose connector, 1.6 mm x 1.6 mm
51506495
131
10
T-hose connector, 1.6 mm x 1.6 mm x 1.6 mm
51506490
132
10
Hose connector 3.2 mm x 3.2 mm
51506491
10
T-hose connector 6.4 mm x 6.4 mm x 6.4 mm
51506493
10
Hose connector 6.4 mm x 6.4 mm
51506494
133
10
Hose connector 3.2 mm x 6.4 mm
51506492
134
10
Y-hose connector 1.6 mm x 1.6 mm x 1.6 mm
51512096
135
10
Connection nipples for sample collector (10 piece)
51512099
144
1
Photometer cell
– CL-A+B
– CL-C+D
51513566
51505776
155
1
Valve set for two-channel version
51512234
160
1
Roller head with holder for hose pump
51512085
161
1
Hose box for pump
51512086
170
1
Collecting vessel with level monitoring
51512089
171
1
Collecting vessel without level monitoring
51512088
200
1
Photometer type1):
1
Silicone spray
51504155
1
Cleaning injector
51503943
1
CA71 kit, complete valve, for one-channel version
51512100
2
CA71 kit, complete valve set, for two-channel version
51512234
1
CA71 kit, level monitoring, retrofitting for collecting vessel
71023419
Please take the designation and the Order No. from the "Troubleshooting/Spare parts" section and enter them here!
Place
Date
Earliest delivery one week after order received. Delivery is unpaid.
Endress+Hauser
Stamp/Signature
55
Appendix
Stamolys CA71CL
11.3
Analyzer settings
Place:
Type:
Analyzer serial No.:
Photometer serial No.:
Software version:
Date:
Photometer type:
Measuring unit:
Calibration factor:
C-offset:
• mg/l
• μg/l
Dilution:
Delay to sample:
s
Analog output:
• 0-20 mA
• 4-20 mA
AV 1:
• NC
• NO
AV 2:
• NC
• NO
Error contact:
• NC
• NO
Frequency offset:
Hz
Zero frequency base line: (dist. water without reagent)
Hz
Measuring range start:
• mg/l
• μg/l
Measuring range end:
• mg/l
• μg/l
AV 1:
• mg/l
• μg/l
AV 2:
• mg/l
• μg/l
1. measurement:
Measuring interval:
Min
1. calibration:
Calibration interval:
h
Calibration solution:
• mg/l
• μg/l
1. flushing:
Flushing interval:
h
Flushing hold on:
s
56
Endress+Hauser
Stamolys CA71CL
Appendix
Submenu
Error mask:
MB >:
MBE:
Post rinsing:
Filling time:
Reaction time:
RPM:
K floating mean:
N:
Points
C1:
mg/l / μg/l
F 1:
Hz
C2:
mg/l / μg/l
F 2:
Hz
C3:
mg/l / μg/l
F 3:
Hz
C4:
mg/l / μg/l
F 4:
Hz
C5:
mg/l / μg/l
F 5:
Hz
C6:
mg/l / μg/l
F 6:
Hz
C7:
mg/l / μg/l
F 7:
Hz
C8:
mg/l / μg/l
F 8:
Hz
C9:
mg/l / μg/l
F 9:
Hz
C10:
mg/l / μg/l
F 10:
Hz
Date:
Endress+Hauser
Service technician:
57
Appendix
Stamolys CA71CL
11.4
Maintenance schedule
a0001911-en
58
Endress+Hauser
Stamolys CA71CL
Endress+Hauser
Appendix
59
Appendix
60
Stamolys CA71CL
Endress+Hauser
Stamolys CA71CL
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Alarm values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–26
Analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Analyzer at a glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Auto measuring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 8
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . 9, 12
Interference suppressor . . . . . . . . . . . . . . . . . . . . . . . . 18, 42
C
L
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Calibration factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Calibration interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Check
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Post-installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cleaners . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 31
Concentration offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Connection
Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Level monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
D
DATA MEMORY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32, 41
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical technician . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ERROR DISPLAY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
15
49
27
43
F
Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Flushing interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Frequency offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
G
GFR housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
H
Housing
GFR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Stainless steel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Without. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
I
M
Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu
Auto measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Data memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23
35
41
35
24
49
23
24
28
27
27
23
26
28
39
N
Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Notes on safety conventions and icons . . . . . . . . . . . . . . . . . 5
O
Offset
Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Open version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 22
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Overview of analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
P
PARAMETER ENTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . 48
Photometer cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Placing out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Post-connection check. . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Post-installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Product structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Pumps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Icons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Endress+Hauser
61
Stamolys CA71CL
Q
Quality certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
R
Reaching the terminal block. . . . . . . . . . . . . . . . . . . . . . . . 15
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36, 41
Replacing
Photometer cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Pump hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Static mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Return. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 47
S
Safety conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Serial interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Shielding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Signal connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Stainless steel housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Standard solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Static mixer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Switching contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48–49
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
V
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
62
Endress+Hauser
www.endress.com/worldwide
BA354C/07/en/10.06
Printed in Germany / FM+SGML 6.0 / DT