Det-Tronics FlexVu® Explosion-Proof Universal Display Unit Model UD20 Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instructions
FlexVu® Explosion-Proof
Universal Display Unit
Model UD20
4.1
Rev: 6/15
95-8620
Table Of Contents
APPLICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
HART Communication . . . . . . . . . . . . . . . . . . . . 1
Magnetic Switches . . . . . . . . . . . . . . . . . . . . . . . 2
Device Enclosure . . . . . . . . . . . . . . . . . . . . . . . . 2
Device Display . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Output Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
IMPORTANT SAFETY NOTES . . . . . . . . . . . . . . . . . 3
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Identification of Vapor(s) to be Detected . . . . . . . 3
Identification of Detector Mounting Locations . . . 4
Device Mounting Orientation . . . . . . . . . . . . . . . 4
Sensor Separation . . . . . . . . . . . . . . . . . . . . . . . 4
WIRING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Power Supply Requirements . . . . . . . . . . . . . . . 4
Wiring Cable Requirements . . . . . . . . . . . . . . . . 5
Wiring Procedure . . . . . . . . . . . . . . . . . . . . . . . . 5
CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General Calibration Information . . . . . . . . . . . . . 9
Calibration Procedure . . . . . . . . . . . . . . . . . . . . 10
Input Trim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
DEVICE REPAIR AND RETURN . . . . . . . . . . . . . . 14
ORDERING INFORMATION . . . . . . . . . . . . . . . . . . 14
APPENDIX A —HART COMMUNICATION
WITH GT3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
APPENDIX B — hart communication
WITH PIRECL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
INSTRUCTIONS
FlexVu® Explosion-Proof
Universal Display Unit
Model UD20
Important
Be sure to read and understand the entire
instruction manual before installing or operating
the gas detection system. The Model UD20 is to
be used with GT3000 or PIRECL gas detectors
to provide early warning of the presence of a gas
mixture. Proper device installation, operation,
and maintenance is required to ensure safe and
effective operation. If this equipment is used in
a manner not specified in this manual, safety
protection may be impaired.
Application
The FlexVu® Model UD20 Universal Display Unit is
designed to be used with the GT3000 Toxic Gas
Detector or PIRECL Combustible Gas Detector.
The UD20 functions as a pass through device, reading
the linear 4-20 mA dc output signal from the gas
detector. The detected gas concentration is indicated
on an alpha-numeric display. The UD20 automatically
detects the gas type and operating range of the sensor
to which it is connected via the HART signal.
All electronics are enclosed in an explosion-proof
aluminum or stainless steel housing. The display unit
is used with a single gas detector, which may be either
directly coupled to the display unit, or remotely located
using a sensor termination box. The UD20 features
non‑intrusive calibration, which can be performed by
using a handheld magnet to activate internal magnetic
switches on the faceplate. Calibration can also be
performed by activating the gas detector’s internal
magnetic reed switch with a magnet.
4.1 © Detector Electronics Corporation 2016
Description
The UD20 provides display and control capabilities for the
GT3000 or PIRECL detector, and uses the following I/O:
Signal Inputs:
4-20 mA loop with HART,
from the gas detector
User Inputs:
Magnetic switches (4),
on the display panel
Visible Outputs:
LCD display
Signal Pass Through: 4-20 mA loop with HART,
from the gas detector
HART communication
The UD20 is a HART primary master and communicates
with the gas detector, which is a HART slave device.
Upon power-up, the UD20 requests the gas type,
measurement range, and unit of measure from the gas
detector. During normal operation, the UD20 continually
polls the gas detector for status information. Because
the UD20 is a primary master, a secondary master can
also be used to communicate with the gas detector at
the same time. If an Asset Management System (AMS)
is used, it should be configured as a secondary master
to avoid communication conflicts with the UD20.
Rev. 6/15
95-8620
magnetic switches
Gas Level Indication
Four internal magnetic switches provide a non‑intrusive
user interface that allows adjustment of configuration
parameters and calibration in the field without the use
of a HART handheld device. The switches are labeled
as follows:
The UD20 display shows “100+” to indicate an overrange condition. For an under-range condition, the
gas level indication is 0%. Actual gas level can be
determined using the UD20 menu: “PROCESS VARS”
> ANALOG INPUT XX.XX mA.
Special States
Cancel, Escape
Enter, Select
Special states other than normal operation and warmup require user action for activation. These include
Calibration, Display / Sensor Configuration, Display /
Sensor Status, and Fault indication. Refer to Appendix
A (UD20 with GT3000) or Appendix B (UD20 with
PIRECL) for details regarding the HART menu structure.
Previous
Write Protect
The write protect function in the sensor can be accessed
by the UD20.
Next
Logging
The UD20 has no logging capability of its own, but
reads the logs that are stored in the gas detector and
displays them on the LCD screen. If an AMS is used,
it communicates directly with the gas detector and not
with the UD20.
Device Enclosure
The housing for the UD20 consists of a multi‑port
aluminum or stainless steel explosion‑proof junction
box with a clear viewing window. Available conduit
entry sizes include 3/4” NPT and M25.
UD20
Device display
GT3000 or PIRECL
GAS
DETECTOR
The UD20 is provided with a 160 x 100 dot matrix
LCD display. See Figure 1. During normal operation,
the UD20’s LCD display indicates the gas type and
units measured. The UD20 communicates with the
sensor and allows easy access to various operating
parameters. Refer to Appendix A or B for details.
CONTROL
SYSTEM
Operating and history logs that are stored in the gas
detector and displayed by the UD20 include:
• Calibration
• Sensor running hours
• Sensor temperature
• Events
PREVIOUS
Output Test
CANCEL / ESCAPE
The UD20 can start and stop a 4-20 mA Loop Test for
checking correct operation of system output devices.
ENTER / SELECT
NEXT
B2426
Figure 1—Faceplate of UD20
4.1
2
95-8620
Fault Conditions When used with PIRECL
faulty or incomplete configuration will render
warranty and product certifications invalid.
When the UD20 is connected to the PIRECL, it
automatically programs the PIRECL for “User Defined
Fault Mode” upon power-up and sets the following
default output levels:
Calibration = 3.8 mA
Warm-up, General Fault, Blocked Optics = 3.0 mA
CAUTION
The device contains no user serviceable
components. Service or repair should never be
attempted by the user. Device repair should be
performed only by the manufacturer or trained
service personnel.
This is done to ensure that the loop current stays high
enough to keep the UD20 powered. The UD20 is a loop
powered device and requires a minimum of 2.6 mA to
operate properly. Setting the fault level to a value other
than default is not recommended.
Liabilities
The manufacturer’s warranty for this product is
void, and all liability for proper function of the
Display is irrevocably transferred to the owner or
operator in the event that the device is serviced or
repaired by personnel not employed or authorized
by Detector Electronics Corporation, or if the
device is used in a manner not conforming to its
intended use.
If the UD20 is unable to change the mode, an “ANALOG
CODE FLT” fault is annunciated.
Refer to the PIRECL instruction manual, number 958526, for complete information regarding fault detection
and fault modes.
Caution
Observe precautions for handling electrostatic
sensitive devices.
Zero Drift Condition
Zero Drift Faults are not specifically indicated by the
UD20. When the process variable drops between 3.9
and 3.1 mA, no fault is indicated by the UD20. When the
process variable reaches 3.0 mA, the UD20 will display
a FAULT. If a FAULT condition is indicated, the user can
interrogate UD20 and PIRECL fault conditions within
the UD20 menu. If no specific fault is indicated, use the
menu to determine the PIRECL process variable. If that
value is ≤3.9 mA, the PIRECL is experiencing zero drift
and should be re-calibrated.
caution
Unused entries must be closed with suitably
certified blanking elements upon installation.
Installation
note
The UD20 housing must be electrically connected
to earth ground. A dedicated earth ground
terminal is provided on the UD20.
important safety notes
Install gas detectors in the appropriate locations
following the guidelines below. The UD20 should be
located where it can be easily viewed by personnel. The
GT3000 can be attached directly to the UD20 housing,
or it can be mounted remotely using a Det‑Tronics STB
termination box (see the “Sensor Separation” section
for additional information). The Model PIRECL should
be mounted as described in the PIRECL instruction
manual, number 95-8526. Detectors must always be
installed per local installation practices.
CAUTION
The wiring procedures in this manual are intended
to ensure proper functioning of the device under
normal conditions. However, because of the
many variations in wiring codes and regulations,
total compliance to these ordinances cannot be
guaranteed. Be certain that all wiring complies
with the NEC as well as all local ordinances. If
in doubt, consult the authority having jurisdiction
before wiring the system. Installation must be
done by a properly trained person.
Always define the following application details:
CAUTION
This product has been tested and approved
for use in hazardous areas. However, it must
be properly installed and used only under the
conditions specified within this manual and
the specific approval certificates. Any device
modification, improper installation, or use in a
4.1
Identification of vapor(s) to be detected
It is necessary to always identify the vapor(s) of interest
at the job site. In addition, the fire hazard properties
of the vapor, such as vapor density, flashpoint, and
vapor pressure should be identified and used to assist
in selecting the optimum detector mounting location
within the area.
3
95-8620
Device Mounting Orientation
The GT3000 detector must be mounted in a vertical
position only, with the sensor pointing down. See Figure
2. For details on mounting the PIRECL detector, refer to
manual number 95-8526.
Important
The GT3000 must be oriented with the LEDs
facing forward so they are easily visible to
personnel within the area. To ensure correct
orientation (the LEDs are not visible when power
is off), position the GND lug on the left hand side
and the calibration notch to the front. Note that the
LEDs are located directly above the calibration
notch.
GND LUG
GREEN LED
SENSOR SEPARATION
CALIBRATION NOTCH
Det-Tronics sensor termination boxes (Model STBs)
enable the installation of the GT3000 separately from
the UD20 Universal Display Unit. Two‑conductor
shielded cable is required to prevent possible nuisance
EMI/RFI.
B2436
Figure 2—Correct Mounting Orientation
Identification of DETECTOR mounting
locations
The PIRECL contains its own termination box for sensor
separation from the UD20.
Identification of the most likely leak sources and
leak accumulation areas is typically the first step in
identifying the best detector mounting locations. In
addition, identification of air current / wind patterns
within the protected area is useful in predicting gas leak
dispersion behavior. This information should be used to
identify optimum sensor installation points.
The maximum cable length between the termination
box and the UD20 is 2000 ft.
WIRING
power SUPPLY requirements
If the vapor of interest is lighter than air, place the sensor
above the potential gas leak. Place the sensor close to
the floor for gases that are heavier than air. Note that air
currents may cause a gas that is slightly heavier than
air to rise under some conditions. Heated gases may
also exhibit the same phenomenon.
Calculate the total gas detection system power
consumption rate in watts from cold start-up. Select a
power supply with adequate capability for the calculated
load. Ensure that the selected power supply provides
regulated and filtered 24 Vdc output power for the
entire system. If a back-up power system is required,
a float‑type battery charging system is recommended.
If an existing source of 24 Vdc power is being utilized,
verify that system requirements are met.
The most effective number and placement of detectors
varies depending on the conditions at the job site. The
individual designing the installation must often rely on
experience and common sense to determine the detector
quantity and best locations to adequately protect the
area. Note that it is typically advantageous to locate
detectors where they are accessible for maintenance.
Locations near excessive heat or vibration sources
should be avoided if possible.
NOTE
The UD20 and GT3000/PIRECL communicate
using HART protocol, which requires a power
supply with low noise levels for proper operation.
(For detailed information regarding power supply
specifications, refer to the HART Communication
Foundation’s document “FSK Physical Layer
Specification” HCF_SPEC-54.)
Final suitability of possible gas detector locations
should be verified by a job site survey.
For locations with expected periodic temperature
conditions from –20°C to –40°C, a certified heater jacket
system must be applied, and the installation should be
accepted by the local authority having jurisdiction.
4.1
4
95-8620
Wiring cable requirements
junction box entries, plugs, & fittings
Always use proper cabling type and diameter for input
power as well as output signal wiring. 22 to 14 AWG
shielded stranded copper wire is recommended.
Always install a properly sized, master power fuse or
breaker on the system power circuit.
WARNING
All entries must contain appropriately rated
plugs or fittings. It is required that each plug or
fitting be wrench-tightened to an appropriate
installation torque and meet the minimum thread
engagement requirements per the applicable
local standards, codes, and practices in order
to retain the defined ratings. PTFE sealant or
equivalent should be used on NPT threads.
For ambient temperatures below –10°C, use field wiring
suitable for the expected conditions. For ambient
temperatures above +60°C, use field wiring and cable
glands suitable for 15°C above the maximum expected
conditions.
Wiring Procedure
note
The use of shielded cable in conduit or shielded
armored cable is required. In applications where
the wiring is installed in conduit, dedicated
conduit is recommended. Avoid low frequency,
high voltage, and non-signaling conductors to
prevent nuisance EMI problems.
UD20 with GT3000
Refer to Figures 3 and 4 for wiring illustrations.
UD20 with PIRECL
No separate power lines are required for the UD20.
The device can be powered off the PIRECL detector in
a two-wire configuration. Figures 5 through 8 illustrate
the Eclipse wired to a UD20 with the 4-20 mA output
shown in various wiring schemes.
caution
The use of proper conduit installation techniques,
breathers, glands, and seals is required to prevent
water ingress and/or maintain the explosion-proof
rating.
A 250 ohm, 3 watt HART resistor must be installed.
The current loop resistance must not exceed 440 ohms
for correct functioning of the detector mA output and
HART signal.
Figure 9 shows the UD20 with the optional Bartec SelfRegulating Heating Cable System.
HAZARDOUS LOCATION
UD20 DISPLAY UNIT
+
+
J2-6
24 VDC
SEE NOTES 2 AND 3
–
J2-5
SHIELD
J2-4
POWER
LOOP
NON-HAZARDOUS LOCATION
–
SEE NOTE 5
RED
BLACK
GREEN
J2-3
+
J2-2
–
J2-1
SHIELD
SENSOR
LOOP
SEE NOTE 1
J2
GT3000
GAS DETECTOR
NOTE 1
CONNECT THE GREEN SENSOR LEAD TO THE CHASSIS GROUND LUG
ON THE INSIDE BOTTOM OF THE UD20 DISPLAY UNIT ENCLOSURE.
NOTE 2
250 OHM, 3 WATT RESISTOR REQUIRED (INCLUDED).
NOTE 3
EXTERNAL HART COMMUNICATION DEVICES CAN BE CONNECTED ACROSS
THE 250 OHM RESISTOR, ACROSS J2-5 AND J2-6, OR ACROSS J2-2 AND J2-3.
NOTE 4
JUNCTION BOX MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
NOTE 5
GROUND THE SHIELD AT THE POWER SOURCE END ONLY.
B2406
Figure 3—Basic Wiring for Explosion Proof UD20 with GT3000
4.1
5
95-8620
HAZARDOUS LOCATION
NON-HAZARDOUS LOCATION
UD20 DISPLAY UNIT
+
J2-6
+
–
J2-5
–
SHIELD
J2-4
24 VDC
POWER
LOOP
SEE NOTES 2 AND 3
SENSOR TERMINATION BOX
SEE NOTE 1
GREEN
RED
+
J2-2
–
J2-1
SHIELD
SEE NOTE 1
AC
K
D
BLACK
SENSOR
LOOP
RED
BL
RE
J2-3
BLACK
J2
GT3000
GAS DETECTOR
NOTE 1
GROUND THE SHIELD AT THE POWER SOURCE END ONLY.
NOTE 2
250 OHM, 3 WATT RESISTOR REQUIRED (INCLUDED).
NOTE 3
EXTERNAL HART COMMUNICATION DEVICES CAN BE CONNECTED ACROSS
THE 250 OHM RESISTOR, ACROSS J2-5 AND J2-6, OR ACROSS J2-2 AND J2-3.
NOTE 4
JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
B2483
Figure 4—Basic Wiring for Explosion Proof UD20 and GT3000 with Terminal Box
PIRECL
POINTWATCH ECLIPSE
ANALOG INPUT / POWER
V+
250Ω
3 WATT
SIGNAL
V–
UD20
DISPLAY UNIT
J2-6
POWER +
J2-5
POWER –
J2-4
SHIELD
J2-3
SENSOR +
J2-2
J2-1
1
24 VDC –
2
24 VDC +
3
CALIBRATE
4
24 VDC –
5
24 VDC +
6
+ 4-20 MA
7
– 4-20 MA
SENSOR –
SHIELD
NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM.
B2688
Figure 5—PIRECL Detector Wired to UD20 (Non-Isolated 4-20 mA Output, Sinking, 4-Wire)
4.1
6
95-8620
PIRECL
POINTWATCH ECLIPSE
ANALOG INPUT / POWER
V+
1
24 VDC –
2
24 VDC +
3
CALIBRATE
4
24 VDC –
5
24 VDC +
6
+ 4-20 MA
7
– 4-20 MA
SIGNAL
250Ω
3 WATT
UD20
DISPLAY UNIT
V–
J2-6
POWER +
J2-5
POWER –
J2-4
SHIELD
J2-3
SENSOR +
J2-2
J2-1
SENSOR –
SHIELD
NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM.
B2689
Figure 6—PIRECL Detector Wired to UD20 (Non-Isolated 4-20 mA Output, Sourcing, 4-Wire)
POWER SUPPLY
V+
ANALOG INPUT / POWER
PIRECL
POINTWATCH ECLIPSE
V–
V+
250Ω
3 WATT
SIGNAL
V–
UD20
DISPLAY UNIT
J2-6
POWER +
J2-5
POWER –
J2-4
SHIELD
J2-3
SENSOR +
J2-2
J2-1
1
24 VDC –
2
24 VDC +
3
CALIBRATE
4
24 VDC –
5
24 VDC +
6
+ 4-20 MA
7
– 4-20 MA
SENSOR –
SHIELD
NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM.
B2690
Figure 7—PIRECL Detector Wired to UD20 (Isolated 4-20 mA Output, Sinking, 4-Wire)
4.1
7
95-8620
POWER SUPPLY
V+
PIRECL
POINTWATCH ECLIPSE
ANALOG INPUT / POWER
V–
V+
SIGNAL
250Ω
3 WATT
UD20
DISPLAY UNIT
V–
J2-6
POWER +
J2-5
POWER –
J2-4
SHIELD
J2-3
SENSOR +
J2-2
J2-1
1
24 VDC –
2
24 VDC +
3
CALIBRATE
4
24 VDC –
5
24 VDC +
6
+ 4-20 MA
7
– 4-20 MA
SENSOR –
SHIELD
NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND.
NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM.
B2691
Figure 8—PIRECL Detector Wired to UD20 (Isolated 4-20 mA Output, Sourcing, 4-Wire)
NON-HAZARDOUS LOCATION
CLASS I, DIV. 2 HAZARDOUS LOCATION
HEATER
BARTEC HEATER
POWER SUPPLY
JUNCTION BOX
HEATER CABLE
TEMP CONTROL LIMIT
MAXIMUM = 20°C
HEATER JACKET
GAS DETECTOR
GAS
POWER SUPPLY
DETECTOR
UD20
GAS DETECTOR
CONTROL
A2693
Figure 9—UD20 with Bartec Self-Regulating Heating Cable System
4.1
8
95-8620
Table 1— GT3000 LEDs During Calibration
Device Status
Green LED
Yellow LED
Normal
Steady On
Off
Waiting for Zero
Off
Steady On
Waiting for Gas
Off
Blinking
Waiting for Span
Off
Blinking
Remove Cal Gas
Off
Off
Normal
Steady On
Off
Calibration
The calibration process is automatic, with the exception
of gas delivery. LEDs on the gas detector and messages
on the UD20 faceplate guide the operator when to apply
and remove the calibration gas. See Tables 1 and 2.
The calibration process can be initiated using the
magnetic switch on the gas detector or using the
magnetic switches on the UD20 faceplate. All sensors,
including oxygen, should be in clean air (20.9% oxygen)
when the calibration sequence is initiated.
note
The UD20 allows the operator to adjust the
calibration gas concentration within the range of
30 to 90% full scale. The default value for all gas
sensors except oxygen is 50% full scale. Oxygen
sensors use a default value of 20.9%.
Once calibration is initiated, the process proceeds
automatically. The yellow LED on the gas detector and
the digital display on the UD20 are used to inform the
operator of the progress of the calibration procedure,
and also signal when to apply and when to remove the
calibration gas.
The calibration can be aborted after zero calibration by
activating the magnetic switch on the gas detector or
navigating the UD20 menu.
General Calibration Information
All GT3000 and PIRECL Gas Detectors provide a twopoint calibration — zero and span.
Table 2—PIRECL LEDs and Current Output Status During Normal Calibration Procedure
4.1
Description
Indicating LED
(on-board/PIRTB)
Operator Action
Normal-ready to calibrate
steady green/off
Purge with clean air if required
Initiate Calibration
steady red/on-steady
Apply Magnet for 2 seconds min.
Zero Calibration complete
flashing red/on-flashing
Apply Calibration Gas to device
Span Calibration in progress
flashing red/on-flashing
Continue cal gas flow
Span Calibration complete
off/on-steady
Remove Calibration Gas
Output Returns to Normal
steady green/off
Calibration Completed
Normal Operation
steady green/off
None
9
95-8620
If the calibration sequence is aborted or not completed
successfully, the detector reverts back to the previous
calibration values and signals a calibration fault. If
a successful calibration cannot be performed, the
calibration fault can be cleared by activating the
magnetic switch on the gas detector for one second.
Table 3— GT3000 LEDs and 4-20 mA Output
During Various Status Conditions
For help assessing when a fault has occurred see
Tables 3 and 4.
The calibration process can fail for the following causes:
• Zero is out of range
• Span is out of range
Function
Green
LED
Yellow
LED
Analog
4-20 mA
Signal
Warm-up
Off
Steady On
3.5
Normal
Operation
Steady On
Off
4-20
Fault
Condition
Off
Steady On
3.5
Calibration
• Time-Out.
The time and date of calibration events are logged
in the GT3000’s non‑volatile memory along with the
calibration outcome. Possible calibration scenarios
include the following:
See Table 1
3.8
Calibration Procedure
• Aborted Calibration
1. Clean air must be present at the sensor prior
to initiating calibration. The use of bottled air is
recommended.
• Failed Calibration
2. Calibration can be initiated in one of two ways:
• Successful Calibration
A. Hold the calibration magnet against the
designated location on the sensor module
(see Figures 10 and 11) until the green LED
turns off and the yellow LED turns on steady
(approximately one second).
NOTE
The calibration procedure must be completed
within a ten minute period. If the calibration is not
completed, a calibration fault will be generated
and the previous calibration data will be used.
B. Initiate calibration via the magnetic switches on
the UD20. Select Main Menu > Device Cal >
Calibration > Execute.
NOTE
To ensure reliable protection, it is important to
check and calibrate the detection system on a
regularly scheduled basis. The frequency of these
checks is determined by the requirements of the
particular installation – typically 30, 60, or 90 day
intervals, depending on the ambient conditions.
The UD20 indicates “Waiting for Zero.” The
transmitter immediately begins taking zero readings.
MULTICOLOR LED
HART COMMUNICATION PORT
(COVER INSTALLED)
PLACE CALIBRATION MAGNET
HERE TO ACTIVATE INTERNAL
REED SWITCH
LED
EARTH GND LUG
PLACE MAGNET HERE
TO ACTIVATE INTERNAL
MAGNETIC SWITCH
WEATHER BAFFLE
CALIBRATION MAGNET
A2452
C2058
Figure 10—Location of Magnetic Switch on GT3000 Detector
4.1
CALIBRATION NOZZLE
Figure 11— Location of Magnetic Switch on PointWatch Eclipse
10
95-8620
Specifications
Table 4— PIRECL LEDs and 4-20 mA Output
During Various Status Conditions
Function
Green
LED
Yellow
LED
Analog
4-20 mA
Signal
Warm-up
Off
Steady On
3.0
Normal
Operation
Steady On
Off
4-20
Fault
Condition
Off
Steady On
3.0
Blocked
Optics
Off
Steady On
3.0
Calibration
See Table 2
UD20 Universal display unit
operating voltage—
24 Vdc nominal, operating range is 19 to 30 Vdc.
UD20 maximum loop voltage drop (Vd) 5.0 volts.
Maximum loop supply 30.0 Vdc.
operating power—
0.8 watt maximum.
operating current—
A minimum of 3.5 mA.
CURRENT OUTPUT—
Linear 4-20 mA
(with HART output directly from detector).
250 ohm, 3 watt termination resistor required (included).
3.8
3. When the zero calibration is complete, the yellow
LED changes from steady to blinking and the UD20
indicates “Waiting for Gas.” Apply the calibration
gas to the sensor. When gas is detected, the UD20
indicates “Waiting for Span.”
Operating temperature—
–20°C to +70°C.
4. When the yellow LED turns off and the UD20
indicates “Remove Cal Gas”, remove the calibration
gas. The gas level at the sensor gradually returns to
zero. The green LED turns on steady to indicate that
the device has returned to normal operation, using
the new calibration data.
Humidity Range—
5 to 95% RH (Det-Tronics verified).
Storage Temperature—
–40°C to +70°C.
sensor compatibility—
The UD20 can be used with any of the Det‑Tronics
GT3000 line of gas detectors, or the Det-Tronics Model
PIRECL hydrocarbon gas detector.
Input Trim
Wiring terminals—
Terminals can handle wire sized from 22 to 14 AWG.
The input current loop of the UD20 is trimmed at the
factory. However, it can also be trimmed in the field
using the following automated process.
Conduit Entries—
3/4” NPT or M25. (Five conduit entries standard.)
Navigate through the menu to “Input Loop Cal.”
Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Device Test
ENCLOSURE MATERIAL—
Epoxy coated aluminum or 316 stainless steel.
SHIPPING WEIGHT—
Aluminum:
4.15 pounds (1.88 kilograms).
Stainless steel: 10.5 pounds (4.76 kilograms).
Display Setup
Contrast Control
Input Loop Cal
DIMENSIONS—
See Figures 12 and 13.
electro-magnetic compatibility—
EMC Directive 2004/108/EC
EN55011 (Emissions)
EN50270 (Immunity)
Upon entering Input Loop Cal, the UD20 commands
the detector to output 4 mA, and then automatically
calibrates its own input. The UD20 then commands the
detector to output 20 mA, and subsequently calibrates
its own input.
4.1
WARRANTY—
12 months from date of installation or 18 months from
date of shipment, whichever occurs first.
11
95-8620
6.48
(16.5)
3.46
(8.8)
4.7
(11.9)
5.2
(13.2)
5.86
(14.9)
11.28
(28.7)
R0.175
A2442
Figure 12—Dimensions of UD20 with GT3000 in Inches (Centimeters)
4.1
12
95-8620
UD20 CERTIFICATIONS—
FM:
Class I, Div. 1, Groups B, C & D;
Class I, Div. 2, Groups B, C & D (T4);
Class I, Zone 1/2 AEx d IIC T6;
Class II/III, Div. 1/2, Groups E, F & G.
Tamb –20°C to +70°C
NEMA/Type 4X, IP66
Conduit seal not required.
5.86
(14.9)
5.2
(13.2)
2.7
(6.9)
Performance verified in accordance with:
ANSI/ISA-92.00.01
FM 6340
ANSI/ISA 60079-29-1
FM 6320
4.7
(11.9)
3.46
(8.8)
CSA:
CSA 08 2162793.
Class I, Div. 1, Groups B, C & D;
(Tamb = –20°C to +70°C).
Class II/III, Div. 1/2, Groups E, F & G;
(Tamb = –20°C to +70°C).
Class I, Div. 2, Groups A, B, C & D (T4);
(Tamb = –40°C to +70°C) with
Bartec Self-Regulating Heating
Cable System. (Install per Figure 9.)
Type 4X, IP66.
Conduit seal not required.
FM
®
APPROVED
3.77
(9.6)
1.28
(3.3)
ATEX:
0539
II 2 G
Ex db IIC T6
Tamb –20°C to +70°C
FM08ATEX0044X
IP66
Performance verified in accordance with:
EN 60079-29-1.
FM
APPROVED
C2281
Figure 13—Dimensions of Model STB Termination
F M Box
in Inches (Centimeters)
APPROVED
IECEx:
Ex db IIC T6
Tamb –20°C to +70°C
IECEx FMG 08.0010X
IP66
Performance verified in accordance with:
IEC 60079-29-1.
INMETRO
UL-BR 15.0469X
Ex d IIC T6 Gb IP66
T6 (Tamb –20°C to +70°C)
All cable entry devices shall be Brazil certified in the
type of explosion protection, flameproof enclosure “d,”
suitable for the conditions of use and correctly installed,
with an ingress protection rating of IP66.
A screw or cover lock is provided for a secondary
means of fastening the cover.
4.1
13
95-8620
DEVICE REPAIR AND RETURN
ORDERING INFORMATION
Prior to returning devices, contact the nearest local
Detector Electronics office so that a Return Material
Identification (RMI) number can be assigned. A
written statement describing the malfunction must
accompany the returned device or component to
assist and expedite finding the root cause of the
failure.
Sensor module, Transmitter module, and termination
boxes (if used) must be ordered separately.
Refer to the UD20 Model Matrix for ordering details.
Replacement Parts/accessories
Pack the unit properly. Always use sufficient packing
material. Where applicable, use an antistatic bag as
protection from electrostatic discharge.
Part Number Description
009700-001 Magnetic Tool
009793-002 Electronics Module
010253-001
250 ohm Terminating Resistor Kit
101197-001* Stop Plug, 3/4" NPT, AL
101197-004* Stop Plug, 3/4" NPT, SS
101197-005
Stop Plug, M25, AL, IP66
101197-003
Stop Plug, M25, SS, IP66
010816-001
Stop Plug, 20PK, 3/4" NPT, AL, EX D
010817-001
Stop Plug, 20PK, 3/4" NPT, SS, EX D
010818-001
Stop Plug, 20PK, M25, AL, IP66, EXDE
010819-001
Stop Plug, 20PK, M25, SS, IP66, EXDE
104190-001
Stop Plug, M25, AL, INMETRO
104190-002
Stop Plug, 3/4” NPT, AL, INMETRO
104190-003
Stop Plug, M25, SS, INMETRO
104190-004
Stop Plug, 3/4” NPT, SS, INMETRO
102804-001 Reducer, M25 to M20, AL
102804-003 Reducer, M25 to M20, SS
103922-001
475 Field Communicator
005003-001 Lubriplate grease, 1 oz.
107427-040
O-ring - Viton
104371-001
Bartec Self-Regulating Heating
Cable System
note
Inadequate packaging that ultimately causes
damage to the returned device during shipment
will result in a service charge to repair the damage
incurred during shipment.
Return all equipment transportation prepaid to the
factory in Minneapolis.
NOTE
It is highly recommended that a complete spare
be kept on hand for field replacement to ensure
continuous protection.
* NEMA 4/IP66 rating requires addition of non‑hardening
thread sealant or Teflon tape.
UD20 Model Matrix
MODEL
UD20
DESCRIPTION
Universal Display Unit
TYPE
MATERIAL
A
Aluminum
S
Stainless Steel (316)
TYPE
THREAD TYPE
5M
5 PORT, METRIC M25
5N
5 PORT, 3/4” NPT
TYPE
26
OUTPUTS
Pass-Through 4-20 mA, HART
TYPE
APPROVAL
W
FM/CSA/ATEX/IECEx
B
INMETRO (Brazil)
TYPE
2
4.1
14
CLASSIFICATION (Div/Zone)
Ex d (Flameproof)
95-8620
Appendix A
HART Communication – UD20 with GT3000
MENU STRUCTURE
This section displays the menu trees for the UD20 connected to the GT3000. The menu tree shows the primary
commands and options available when using menu selections.
4.1
15
95-8620
4.1
16
95-8620
PROCESS VARS
DISPLAY STATUS
DEVICE STATUS
DISPLAY SETUP
DEVICE SETUP
DEVICE CAL
DEVICE TEST
MAIN MENU
SELF TEST
RESPONSE TEST
LOOP TEST
D/A TRIM
DEVICE TEST
CAL GAS CONC
XX.XX
CALIBRATION
CHANGE SNSR TYPE
DEVICE CAL
HART OPTION
RTC
WRITE PROTECT
DEVICE SETUP
CONTRAST CONTROL
INPUT LOOP CAL
DISPLAY SETUP
GENERAL INFO
FAULT/STATUS
TX INFO
SENSOR INFO
HISTORY
DEBUG MENU
DEVICE STATUS
DISPLAY INFO
FAULT/STATUS
DEBUG MENU
DISPLAY STATUS
XXXX
X.XX
XX.XX MA
X.XX
X.XX
Y/N
PROCESS VARS
GAS NAME
GAS VALUE
ANALOG INPUT
URV
LRV
FAULT
DISPLAY INFO
XX
XX
XX
XX
XX
XX
ZERO TRIM
GAIN TRIM
D/A TRIM
SET 4-20 MA
LOOP TEST
EXECUTE
ABORT
CALIBRATION
CHANGE STATE
CHANGE PASSWORD
WRITE PROTECT
XXXXX
WRITE PROTECT
SECONDS
MINUTES
HOURS
DAY
MONTH
YEAR
TAG
XXXXX
DESCRIPTOR
XXXXX
DATE
DD/MMM/YYYY
MESSAGE
XXXXX
FINAL ASSY NUM XXXX
HART OPTION
HART ERRORS
LCD TEST
ANALOG INPUT
RTC
XXXXX
Y/N
Y/N
FAULT
ANY FAULT
CAL ACTIVE
WARM UP
INPUT HART
STATUS
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ADC FAULT
ANALOG CODE FLT
FLASH FAULT
RAM FAULT
WDT FAULT
3V FAULT
INPUT LOOP FAULT
SET 4-20 MA
3.5 MA
4 MA
6 MA
8 MA
10 MA
12 MA
14 MA
16 MA
18 MA
20 MA
XX.XX MA
XXXXX
DEBUG MENU
OP MODE
FAULT
STATUS
FAULT/STATUS
F/W REV
XX.XX
SERIAL NUMBER XXXX
TEMPERATURE XX.XX C
FACTORY CONFIG
GENERAL INFO
XX.XX
XX.XX
XXXXX
XXXXX
XXXXX
XXXXX
XX.XX C
XXXXX
XXXXX
Y/N
Y/N
Y/N
Y/N
MODBUS ERRORSXXXX
DEBUG MENU
TX HISTORY
SENSOR HISTORY
CALIBRATION LOG
EVENT LOG
HISTORY
SENSITIVITY
XXX.XX
GAS TYPE
XXXXX
H/W REV
XX.XX
F/W REV
XX.XX
SERIAL NUMBER XXXXX
REV
XXXXX
USL
XX.XX
LSL
XX.XX
RUNNING HOURS XXXXX
CAL POINT ZERO XX.XX
CAL POINT SPAN XX.XX
PPM HOURS
XXXXX
TEMPERATURE XX.XX C
SENSOR INFO
RTC
H/W REV
F/W REV
SERIAL NUMBER
UNIVERSAL REV
FLD DEV REV
RUNNING HOURS
TEMPERATURE
TX INFO
OP MODE
CAL STATE
TX STATUS
TX FAULT
SENSOR STATUS
SENSOR FAULT
FAULT/STATUS
MANUFACTURER
MODEL
FGP_TX
TAG
XXXXX
DESCRIPTOR
XXXXX
DATE
DD/MMM/YYYY
MESSAGE
XXXXX
FINAL ASSY NUM XXXXX
DEVICE ID
XXXXX
RTC
XX
XX
XX
XX
XX
XX
XXXX
XX.XX C
XXXXX
XX.XX C
XXXXX
EVENT
DATE
TIME
CAL ID
DATE
TIME
ZERO
SPAN
TX STATUS
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
CAL FAULT
EE FAULT
ADC FAULT
ADC CNTR FAULT
3V FAULT
ZERO DRIFT FAULT
FLASH CRC FAULT
RAM FAULT
TEMP FAULT
COMM FAULT
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
SENSOR FAULT
SENSOR FAULT
WARM UP
CAL ACTIVE
CAL SWITCH
WRITE PROTECT
EOL SENSOR
CHANGE CONFIG
SENSOR STATUS
EE FAULT
ON/OFF
ADC FAULT
ON/OFF
DAC FAULT
ON/OFF
LOW VOLTAGE
ON/OFF
FLASH CRC FAULT ON/OFF
RAM FAULT
ON/OFF
TEMP FAULT
ON/OFF
SOFTWARE FAULT ON/OFF
EE SAFETY FAULT ON/OFF
ZERO DRIFT FAULT ON/OFF
SENSOR MISMATCH ON/OFF
SENSOR FAULT
ON/OFF
TX FAULT
TX FAULT
ON/OFF
WARM UP
ON/OFF
CHANGE CONFIG ON/OFF
MULTI DROP
ON/OFF
WRITE PROTECT ON/OFF
SELF TEST
ON/OFF
RESPONSE TEST ON/OFF
CURRENT FIXED ON/OFF
LOOP TEST
ON/OFF
FACTORY MODE ON/OFF
SNSR ASSY REMOVE ON/OFF
UD20
GT3000
March 1, 2013
1.26
Detector Electronics
XXXXX
DD/MMM/YYYY
HH:MM:SS
EVENT LOG
XXX
DD/MMM/YYYY
HH:MM:SS
XX.XX
XX.XX
CALIBRATION LOG
RUNNING HOURS
MAX TEMP
MAX TEMP TIME
MIN TEMP
MIN TEMP TIME
SENSOR HISTORY
RUNNING HOURS XXXXX
MAX TEMP
XX.XX C
MAX TEMP TIME XXXXX
MIN TEMP
XX.XX C
MIN TEMP TIME XXXXX
TX HISTORY
SECONDS
MINUTES
HOURS
DAY
MONTH
YEAR
6901 West 110th Street
Minneapolis, MN 55438
USA
Appendix B
HART Communication – UD20 with PIRECL
MENU STRUCTURE
This section displays the menu trees for the UD20 connected to the PIRECL. The menu tree shows the primary
commands and options available when using menu selections.
4.1
17
95-8620
4.1
18
95-8620
PROCESS VARS
DISPLAY STATUS
DEVICE STATUS
DISPLAY SETUP
DEVICE SETUP
DEVICE CAL
DEVICE TEST
MAIN MENU
SELF TEST
RESPONSE TEST
LOOP TEST
D/A TRIM
XX.XX
DD/MMM/YYYY
DEVICE TEST
CAL GAS CONC
CAL GAS TYPE
CALIBRATION
CAL DATE
DEVICE CAL
DEVICE OPTION
ALARM SETTING
HART OPTION
RS485
WRITE PROTECT
DEVICE SETUP
CONTRAST CONTROL
INPUT LOOP CAL
DISPLAY SETUP
GENERAL INFO
FAULT/STATUS
DEVICE INFO
SENSOR INFO
HISTORY
DEVICE STATUS
DISPLAY INFO
FAULT/STATUS
DEBUG MENU
DISPLAY STATUS
XXXX
X.XX
XX.XX MA
X.XX
X.XX
Y/N
PROCESS VARS
GAS NAME
GAS VALUE
ANALOG INPUT
URV
LRV
FAULT
XX.XX
XX.XX
XX.XX
XX.XX
XX.XX
Y/N
XX.XX
Y/N
XXX
ENABLED
ZERO TRIM
GAIN TRIM
SET 4-20 MA
EXECUTE
ABORT
D/A TRIM
LOOP TEST
CALIBRATION
SAME AS MEASURE GAS
METHANE
PROPANE
CAL GAS TYPE
CHANGE STATE
CHANGE PASSWORD
WRITE PROTECT
WRITE PROTECT
BAUD RATE
PARITY
POLL ADDRESS
RS485
TAG
XXXXX
DESCRIPTOR
XXXXX
DATE
DD/MMM/YYYY
MESSAGE
XXXXX
FINAL ASSY NUM
XXXXX
HART OPTION
RST LATCH ALARMS
HGH ALARM LEVEL
HGH ALARM LATCH
LOW ALARM LEVEL
LOW ALARM LATCH
ALARM SETTING
GAS TYPE
UNIT OF MEASURE
URV
LRV
USL
LSL
ANALOG CODE VAL
ANALOG FLT CODE
DEVICE OPTION
PARITY
SET 4-20 MA
3.5 MA
4 MA
6 MA
8 MA
10 MA
12 MA
14 MA
16 MA
18 MA
20 MA
NONE
EVEN
ODD
1200
2400
4800
9600
19.2K
BAUD RATE
ECLIPSE
PIR9400
USER DEFINED
ANALOG FAULT CODE
XX.XX
XX.XX
XX.XX
XX.XX
ANALOG CODE VAL
WARM UP
BLOCKED OPTIC
CALIBRATION
FAULT
%LFL
PPM
VOL %
UNIT OF MEASURE
METHANE
ETHANE
PROPANE
ETHYLENE
PROPYLENE
BUTANE
SPECIAL
GAS TYPE
XXXXX
XXXXX
Y/N
Y/N
DEVICE HISTORY
CALIBRATION LOG
EVENT LOG
HISTORY
ACTIVE
REFERENCE
RATIO
ABSORPTION
TEMPERATURE
VOL %
USL
LSL
SPAN FACTOR
XX.XX
XX.XX
XX.XX
XX.XX
XX.XXC
XX.XX
XX.XX
XX.XX
XX.XX
SENSOR INFO
XXXXX
XXXXX
XXXXX
XXXXX
DEVICE INFO
SERIAL NUMBER
UNIVERSAL REV
FLD DEV REV
S/W REV
OP MODE
CAL STATE
FAULT
STATUS
FAULT/STATUS
MANUFACTURER
MODEL
ECLIPSE
TAG
XXXXX
DESCRIPTOR
XXXXX
DATE
DD/MMM/YYYY
MESSAGE
XXXXX
FINAL ASSY NUM
XXXXX
DEVICE ID
XXXXX
GENERAL INFO
EVENT
HOURS
XXXXX
XX.XX C
XXXX
XX.XX C
XXXX
XX.XX C
XXXX
XX.XX C
XXXX
EVENT LOG
XXXXX
XXXXX
XXXXX
XXXXX
CALIBRATION LOG
RUNNING HOURS
MAX TEMP
MAX TEMP TIME
MIN TEMP
MIN TEMP TIME
MAX RESET TEMP
MAX RESET TIME
MIN RESET TEMP
MIN RESET TIME
CAL ID
HOURS
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
DEVICE HISTORY
CAL ACTIVE
WARM UP
LOW ALARM
HIGH ALARM
HART SELF TEST
CHANGE CONFIG
STATUS
CAL FAULT
DIRTY OPTICS
LAMP FAULT
START CAL FAULT
EE 1 FAULT
EE 2 FAULT
REF ADC SAT
ACTIVE ADC SAT
24V FAULT
12V FAULT
5V FAULT
ZERO DRIFT FAULT
FLASH CRC FAULT
RAM FAULT
FAULT
6901 WEST 110TH STREET
MINNEAPOLIS, MN 55438
USA
FAULT
ANY FAULT
CAL ACTIVE
WARM UP
INPUT HART
STATUS
ADC FAULT
ANALOG CODE FLT
FLASH FAULT
RAM FAULT
WDT FAULT
3V FAULT
INPUT LOOP FAULT
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
UD20
PIRECL
March 1, 2013
1.26
Detector Electronics
XX.XX MA
XXXXX
DEBUG MENU
XXXXX
Y/N
Y/N
FAULT/STATUS
HART ERRORS
LCD TEST
ANALOG INPUT
OP MODE
FAULT
STATUS
XX.XX
XXXX
XX.XX C
DISPLAY INFO
F/W REV
SERIAL NUMBER
TEMPERATURE
FACTORY CONFIG
95-8620
FlexSonic ® Acoustic
Leak Detector
X3301 Multispectrum
IR Flame Detector
Corporate Office
6901 West 110th Street
Minneapolis, MN 55438 USA
www.det-tronics.com
PointWatch Eclipse ® IR
Combustible Gas Detector
Phone: 952.946.6491
Toll-free: 800.765.3473
Fax: 952.829.8750
det-tronics@det-tronics.com
FlexVu ® Universal Display
with GT3000 Toxic Gas Detector
Eagle Quantum
Premier ®
Safety System
All trademarks are the property of their respective owners.
© 2016 Detector Electronics Corporation. All rights reserved.
Det-Tronics manufacturing system is certified to ISO 9001—
the world’s most recognized quality management standard.