Det-Tronics FlexVu® Explosion-Proof Universal Display Unit Model UD20 Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instructions FlexVu® Explosion-Proof Universal Display Unit Model UD20 4.1 Rev: 6/15 95-8620 Table Of Contents APPLICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 HART Communication . . . . . . . . . . . . . . . . . . . . 1 Magnetic Switches . . . . . . . . . . . . . . . . . . . . . . . 2 Device Enclosure . . . . . . . . . . . . . . . . . . . . . . . . 2 Device Display . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Output Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 IMPORTANT SAFETY NOTES . . . . . . . . . . . . . . . . . 3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Identification of Vapor(s) to be Detected . . . . . . . 3 Identification of Detector Mounting Locations . . . 4 Device Mounting Orientation . . . . . . . . . . . . . . . 4 Sensor Separation . . . . . . . . . . . . . . . . . . . . . . . 4 WIRING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Power Supply Requirements . . . . . . . . . . . . . . . 4 Wiring Cable Requirements . . . . . . . . . . . . . . . . 5 Wiring Procedure . . . . . . . . . . . . . . . . . . . . . . . . 5 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 General Calibration Information . . . . . . . . . . . . . 9 Calibration Procedure . . . . . . . . . . . . . . . . . . . . 10 Input Trim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 DEVICE REPAIR AND RETURN . . . . . . . . . . . . . . 14 ORDERING INFORMATION . . . . . . . . . . . . . . . . . . 14 APPENDIX A —HART COMMUNICATION WITH GT3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 APPENDIX B — hart communication WITH PIRECL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 INSTRUCTIONS FlexVu® Explosion-Proof Universal Display Unit Model UD20 Important Be sure to read and understand the entire instruction manual before installing or operating the gas detection system. The Model UD20 is to be used with GT3000 or PIRECL gas detectors to provide early warning of the presence of a gas mixture. Proper device installation, operation, and maintenance is required to ensure safe and effective operation. If this equipment is used in a manner not specified in this manual, safety protection may be impaired. Application The FlexVu® Model UD20 Universal Display Unit is designed to be used with the GT3000 Toxic Gas Detector or PIRECL Combustible Gas Detector. The UD20 functions as a pass through device, reading the linear 4-20 mA dc output signal from the gas detector. The detected gas concentration is indicated on an alpha-numeric display. The UD20 automatically detects the gas type and operating range of the sensor to which it is connected via the HART signal. All electronics are enclosed in an explosion-proof aluminum or stainless steel housing. The display unit is used with a single gas detector, which may be either directly coupled to the display unit, or remotely located using a sensor termination box. The UD20 features non‑intrusive calibration, which can be performed by using a handheld magnet to activate internal magnetic switches on the faceplate. Calibration can also be performed by activating the gas detector’s internal magnetic reed switch with a magnet. 4.1 © Detector Electronics Corporation 2016 Description The UD20 provides display and control capabilities for the GT3000 or PIRECL detector, and uses the following I/O: Signal Inputs: 4-20 mA loop with HART, from the gas detector User Inputs: Magnetic switches (4), on the display panel Visible Outputs: LCD display Signal Pass Through: 4-20 mA loop with HART, from the gas detector HART communication The UD20 is a HART primary master and communicates with the gas detector, which is a HART slave device. Upon power-up, the UD20 requests the gas type, measurement range, and unit of measure from the gas detector. During normal operation, the UD20 continually polls the gas detector for status information. Because the UD20 is a primary master, a secondary master can also be used to communicate with the gas detector at the same time. If an Asset Management System (AMS) is used, it should be configured as a secondary master to avoid communication conflicts with the UD20. Rev. 6/15 95-8620 magnetic switches Gas Level Indication Four internal magnetic switches provide a non‑intrusive user interface that allows adjustment of configuration parameters and calibration in the field without the use of a HART handheld device. The switches are labeled as follows: The UD20 display shows “100+” to indicate an overrange condition. For an under-range condition, the gas level indication is 0%. Actual gas level can be determined using the UD20 menu: “PROCESS VARS” > ANALOG INPUT XX.XX mA. Special States Cancel, Escape Enter, Select Special states other than normal operation and warmup require user action for activation. These include Calibration, Display / Sensor Configuration, Display / Sensor Status, and Fault indication. Refer to Appendix A (UD20 with GT3000) or Appendix B (UD20 with PIRECL) for details regarding the HART menu structure. Previous Write Protect The write protect function in the sensor can be accessed by the UD20. Next Logging The UD20 has no logging capability of its own, but reads the logs that are stored in the gas detector and displays them on the LCD screen. If an AMS is used, it communicates directly with the gas detector and not with the UD20. Device Enclosure The housing for the UD20 consists of a multi‑port aluminum or stainless steel explosion‑proof junction box with a clear viewing window. Available conduit entry sizes include 3/4” NPT and M25. UD20 Device display GT3000 or PIRECL GAS DETECTOR The UD20 is provided with a 160 x 100 dot matrix LCD display. See Figure 1. During normal operation, the UD20’s LCD display indicates the gas type and units measured. The UD20 communicates with the sensor and allows easy access to various operating parameters. Refer to Appendix A or B for details. CONTROL SYSTEM Operating and history logs that are stored in the gas detector and displayed by the UD20 include: • Calibration • Sensor running hours • Sensor temperature • Events PREVIOUS Output Test CANCEL / ESCAPE The UD20 can start and stop a 4-20 mA Loop Test for checking correct operation of system output devices. ENTER / SELECT NEXT B2426 Figure 1—Faceplate of UD20 4.1 2 95-8620 Fault Conditions When used with PIRECL faulty or incomplete configuration will render warranty and product certifications invalid. When the UD20 is connected to the PIRECL, it automatically programs the PIRECL for “User Defined Fault Mode” upon power-up and sets the following default output levels: Calibration = 3.8 mA Warm-up, General Fault, Blocked Optics = 3.0 mA CAUTION The device contains no user serviceable components. Service or repair should never be attempted by the user. Device repair should be performed only by the manufacturer or trained service personnel. This is done to ensure that the loop current stays high enough to keep the UD20 powered. The UD20 is a loop powered device and requires a minimum of 2.6 mA to operate properly. Setting the fault level to a value other than default is not recommended. Liabilities The manufacturer’s warranty for this product is void, and all liability for proper function of the Display is irrevocably transferred to the owner or operator in the event that the device is serviced or repaired by personnel not employed or authorized by Detector Electronics Corporation, or if the device is used in a manner not conforming to its intended use. If the UD20 is unable to change the mode, an “ANALOG CODE FLT” fault is annunciated. Refer to the PIRECL instruction manual, number 958526, for complete information regarding fault detection and fault modes. Caution Observe precautions for handling electrostatic sensitive devices. Zero Drift Condition Zero Drift Faults are not specifically indicated by the UD20. When the process variable drops between 3.9 and 3.1 mA, no fault is indicated by the UD20. When the process variable reaches 3.0 mA, the UD20 will display a FAULT. If a FAULT condition is indicated, the user can interrogate UD20 and PIRECL fault conditions within the UD20 menu. If no specific fault is indicated, use the menu to determine the PIRECL process variable. If that value is ≤3.9 mA, the PIRECL is experiencing zero drift and should be re-calibrated. caution Unused entries must be closed with suitably certified blanking elements upon installation. Installation note The UD20 housing must be electrically connected to earth ground. A dedicated earth ground terminal is provided on the UD20. important safety notes Install gas detectors in the appropriate locations following the guidelines below. The UD20 should be located where it can be easily viewed by personnel. The GT3000 can be attached directly to the UD20 housing, or it can be mounted remotely using a Det‑Tronics STB termination box (see the “Sensor Separation” section for additional information). The Model PIRECL should be mounted as described in the PIRECL instruction manual, number 95-8526. Detectors must always be installed per local installation practices. CAUTION The wiring procedures in this manual are intended to ensure proper functioning of the device under normal conditions. However, because of the many variations in wiring codes and regulations, total compliance to these ordinances cannot be guaranteed. Be certain that all wiring complies with the NEC as well as all local ordinances. If in doubt, consult the authority having jurisdiction before wiring the system. Installation must be done by a properly trained person. Always define the following application details: CAUTION This product has been tested and approved for use in hazardous areas. However, it must be properly installed and used only under the conditions specified within this manual and the specific approval certificates. Any device modification, improper installation, or use in a 4.1 Identification of vapor(s) to be detected It is necessary to always identify the vapor(s) of interest at the job site. In addition, the fire hazard properties of the vapor, such as vapor density, flashpoint, and vapor pressure should be identified and used to assist in selecting the optimum detector mounting location within the area. 3 95-8620 Device Mounting Orientation The GT3000 detector must be mounted in a vertical position only, with the sensor pointing down. See Figure 2. For details on mounting the PIRECL detector, refer to manual number 95-8526. Important The GT3000 must be oriented with the LEDs facing forward so they are easily visible to personnel within the area. To ensure correct orientation (the LEDs are not visible when power is off), position the GND lug on the left hand side and the calibration notch to the front. Note that the LEDs are located directly above the calibration notch. GND LUG GREEN LED SENSOR SEPARATION CALIBRATION NOTCH Det-Tronics sensor termination boxes (Model STBs) enable the installation of the GT3000 separately from the UD20 Universal Display Unit. Two‑conductor shielded cable is required to prevent possible nuisance EMI/RFI. B2436 Figure 2—Correct Mounting Orientation Identification of DETECTOR mounting locations The PIRECL contains its own termination box for sensor separation from the UD20. Identification of the most likely leak sources and leak accumulation areas is typically the first step in identifying the best detector mounting locations. In addition, identification of air current / wind patterns within the protected area is useful in predicting gas leak dispersion behavior. This information should be used to identify optimum sensor installation points. The maximum cable length between the termination box and the UD20 is 2000 ft. WIRING power SUPPLY requirements If the vapor of interest is lighter than air, place the sensor above the potential gas leak. Place the sensor close to the floor for gases that are heavier than air. Note that air currents may cause a gas that is slightly heavier than air to rise under some conditions. Heated gases may also exhibit the same phenomenon. Calculate the total gas detection system power consumption rate in watts from cold start-up. Select a power supply with adequate capability for the calculated load. Ensure that the selected power supply provides regulated and filtered 24 Vdc output power for the entire system. If a back-up power system is required, a float‑type battery charging system is recommended. If an existing source of 24 Vdc power is being utilized, verify that system requirements are met. The most effective number and placement of detectors varies depending on the conditions at the job site. The individual designing the installation must often rely on experience and common sense to determine the detector quantity and best locations to adequately protect the area. Note that it is typically advantageous to locate detectors where they are accessible for maintenance. Locations near excessive heat or vibration sources should be avoided if possible. NOTE The UD20 and GT3000/PIRECL communicate using HART protocol, which requires a power supply with low noise levels for proper operation. (For detailed information regarding power supply specifications, refer to the HART Communication Foundation’s document “FSK Physical Layer Specification” HCF_SPEC-54.) Final suitability of possible gas detector locations should be verified by a job site survey. For locations with expected periodic temperature conditions from –20°C to –40°C, a certified heater jacket system must be applied, and the installation should be accepted by the local authority having jurisdiction. 4.1 4 95-8620 Wiring cable requirements junction box entries, plugs, & fittings Always use proper cabling type and diameter for input power as well as output signal wiring. 22 to 14 AWG shielded stranded copper wire is recommended. Always install a properly sized, master power fuse or breaker on the system power circuit. WARNING All entries must contain appropriately rated plugs or fittings. It is required that each plug or fitting be wrench-tightened to an appropriate installation torque and meet the minimum thread engagement requirements per the applicable local standards, codes, and practices in order to retain the defined ratings. PTFE sealant or equivalent should be used on NPT threads. For ambient temperatures below –10°C, use field wiring suitable for the expected conditions. For ambient temperatures above +60°C, use field wiring and cable glands suitable for 15°C above the maximum expected conditions. Wiring Procedure note The use of shielded cable in conduit or shielded armored cable is required. In applications where the wiring is installed in conduit, dedicated conduit is recommended. Avoid low frequency, high voltage, and non-signaling conductors to prevent nuisance EMI problems. UD20 with GT3000 Refer to Figures 3 and 4 for wiring illustrations. UD20 with PIRECL No separate power lines are required for the UD20. The device can be powered off the PIRECL detector in a two-wire configuration. Figures 5 through 8 illustrate the Eclipse wired to a UD20 with the 4-20 mA output shown in various wiring schemes. caution The use of proper conduit installation techniques, breathers, glands, and seals is required to prevent water ingress and/or maintain the explosion-proof rating. A 250 ohm, 3 watt HART resistor must be installed. The current loop resistance must not exceed 440 ohms for correct functioning of the detector mA output and HART signal. Figure 9 shows the UD20 with the optional Bartec SelfRegulating Heating Cable System. HAZARDOUS LOCATION UD20 DISPLAY UNIT + + J2-6 24 VDC SEE NOTES 2 AND 3 – J2-5 SHIELD J2-4 POWER LOOP NON-HAZARDOUS LOCATION – SEE NOTE 5 RED BLACK GREEN J2-3 + J2-2 – J2-1 SHIELD SENSOR LOOP SEE NOTE 1 J2 GT3000 GAS DETECTOR NOTE 1 CONNECT THE GREEN SENSOR LEAD TO THE CHASSIS GROUND LUG ON THE INSIDE BOTTOM OF THE UD20 DISPLAY UNIT ENCLOSURE. NOTE 2 250 OHM, 3 WATT RESISTOR REQUIRED (INCLUDED). NOTE 3 EXTERNAL HART COMMUNICATION DEVICES CAN BE CONNECTED ACROSS THE 250 OHM RESISTOR, ACROSS J2-5 AND J2-6, OR ACROSS J2-2 AND J2-3. NOTE 4 JUNCTION BOX MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. NOTE 5 GROUND THE SHIELD AT THE POWER SOURCE END ONLY. B2406 Figure 3—Basic Wiring for Explosion Proof UD20 with GT3000 4.1 5 95-8620 HAZARDOUS LOCATION NON-HAZARDOUS LOCATION UD20 DISPLAY UNIT + J2-6 + – J2-5 – SHIELD J2-4 24 VDC POWER LOOP SEE NOTES 2 AND 3 SENSOR TERMINATION BOX SEE NOTE 1 GREEN RED + J2-2 – J2-1 SHIELD SEE NOTE 1 AC K D BLACK SENSOR LOOP RED BL RE J2-3 BLACK J2 GT3000 GAS DETECTOR NOTE 1 GROUND THE SHIELD AT THE POWER SOURCE END ONLY. NOTE 2 250 OHM, 3 WATT RESISTOR REQUIRED (INCLUDED). NOTE 3 EXTERNAL HART COMMUNICATION DEVICES CAN BE CONNECTED ACROSS THE 250 OHM RESISTOR, ACROSS J2-5 AND J2-6, OR ACROSS J2-2 AND J2-3. NOTE 4 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. B2483 Figure 4—Basic Wiring for Explosion Proof UD20 and GT3000 with Terminal Box PIRECL POINTWATCH ECLIPSE ANALOG INPUT / POWER V+ 250Ω 3 WATT SIGNAL V– UD20 DISPLAY UNIT J2-6 POWER + J2-5 POWER – J2-4 SHIELD J2-3 SENSOR + J2-2 J2-1 1 24 VDC – 2 24 VDC + 3 CALIBRATE 4 24 VDC – 5 24 VDC + 6 + 4-20 MA 7 – 4-20 MA SENSOR – SHIELD NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM. B2688 Figure 5—PIRECL Detector Wired to UD20 (Non-Isolated 4-20 mA Output, Sinking, 4-Wire) 4.1 6 95-8620 PIRECL POINTWATCH ECLIPSE ANALOG INPUT / POWER V+ 1 24 VDC – 2 24 VDC + 3 CALIBRATE 4 24 VDC – 5 24 VDC + 6 + 4-20 MA 7 – 4-20 MA SIGNAL 250Ω 3 WATT UD20 DISPLAY UNIT V– J2-6 POWER + J2-5 POWER – J2-4 SHIELD J2-3 SENSOR + J2-2 J2-1 SENSOR – SHIELD NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM. B2689 Figure 6—PIRECL Detector Wired to UD20 (Non-Isolated 4-20 mA Output, Sourcing, 4-Wire) POWER SUPPLY V+ ANALOG INPUT / POWER PIRECL POINTWATCH ECLIPSE V– V+ 250Ω 3 WATT SIGNAL V– UD20 DISPLAY UNIT J2-6 POWER + J2-5 POWER – J2-4 SHIELD J2-3 SENSOR + J2-2 J2-1 1 24 VDC – 2 24 VDC + 3 CALIBRATE 4 24 VDC – 5 24 VDC + 6 + 4-20 MA 7 – 4-20 MA SENSOR – SHIELD NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM. B2690 Figure 7—PIRECL Detector Wired to UD20 (Isolated 4-20 mA Output, Sinking, 4-Wire) 4.1 7 95-8620 POWER SUPPLY V+ PIRECL POINTWATCH ECLIPSE ANALOG INPUT / POWER V– V+ SIGNAL 250Ω 3 WATT UD20 DISPLAY UNIT V– J2-6 POWER + J2-5 POWER – J2-4 SHIELD J2-3 SENSOR + J2-2 J2-1 1 24 VDC – 2 24 VDC + 3 CALIBRATE 4 24 VDC – 5 24 VDC + 6 + 4-20 MA 7 – 4-20 MA SENSOR – SHIELD NOTE 1 JUNCTION BOXES MUST BE ELECTRICALLY CONNECTED TO EARTH GROUND. NOTE 2 THE TOTAL LOOP RESISTANCE = 250 OHMS MINIMUM, 440 OHMS MAXIMUM. B2691 Figure 8—PIRECL Detector Wired to UD20 (Isolated 4-20 mA Output, Sourcing, 4-Wire) NON-HAZARDOUS LOCATION CLASS I, DIV. 2 HAZARDOUS LOCATION HEATER BARTEC HEATER POWER SUPPLY JUNCTION BOX HEATER CABLE TEMP CONTROL LIMIT MAXIMUM = 20°C HEATER JACKET GAS DETECTOR GAS POWER SUPPLY DETECTOR UD20 GAS DETECTOR CONTROL A2693 Figure 9—UD20 with Bartec Self-Regulating Heating Cable System 4.1 8 95-8620 Table 1— GT3000 LEDs During Calibration Device Status Green LED Yellow LED Normal Steady On Off Waiting for Zero Off Steady On Waiting for Gas Off Blinking Waiting for Span Off Blinking Remove Cal Gas Off Off Normal Steady On Off Calibration The calibration process is automatic, with the exception of gas delivery. LEDs on the gas detector and messages on the UD20 faceplate guide the operator when to apply and remove the calibration gas. See Tables 1 and 2. The calibration process can be initiated using the magnetic switch on the gas detector or using the magnetic switches on the UD20 faceplate. All sensors, including oxygen, should be in clean air (20.9% oxygen) when the calibration sequence is initiated. note The UD20 allows the operator to adjust the calibration gas concentration within the range of 30 to 90% full scale. The default value for all gas sensors except oxygen is 50% full scale. Oxygen sensors use a default value of 20.9%. Once calibration is initiated, the process proceeds automatically. The yellow LED on the gas detector and the digital display on the UD20 are used to inform the operator of the progress of the calibration procedure, and also signal when to apply and when to remove the calibration gas. The calibration can be aborted after zero calibration by activating the magnetic switch on the gas detector or navigating the UD20 menu. General Calibration Information All GT3000 and PIRECL Gas Detectors provide a twopoint calibration — zero and span. Table 2—PIRECL LEDs and Current Output Status During Normal Calibration Procedure 4.1 Description Indicating LED (on-board/PIRTB) Operator Action Normal-ready to calibrate steady green/off Purge with clean air if required Initiate Calibration steady red/on-steady Apply Magnet for 2 seconds min. Zero Calibration complete flashing red/on-flashing Apply Calibration Gas to device Span Calibration in progress flashing red/on-flashing Continue cal gas flow Span Calibration complete off/on-steady Remove Calibration Gas Output Returns to Normal steady green/off Calibration Completed Normal Operation steady green/off None 9 95-8620 If the calibration sequence is aborted or not completed successfully, the detector reverts back to the previous calibration values and signals a calibration fault. If a successful calibration cannot be performed, the calibration fault can be cleared by activating the magnetic switch on the gas detector for one second. Table 3— GT3000 LEDs and 4-20 mA Output During Various Status Conditions For help assessing when a fault has occurred see Tables 3 and 4. The calibration process can fail for the following causes: • Zero is out of range • Span is out of range Function Green LED Yellow LED Analog 4-20 mA Signal Warm-up Off Steady On 3.5 Normal Operation Steady On Off 4-20 Fault Condition Off Steady On 3.5 Calibration • Time-Out. The time and date of calibration events are logged in the GT3000’s non‑volatile memory along with the calibration outcome. Possible calibration scenarios include the following: See Table 1 3.8 Calibration Procedure • Aborted Calibration 1. Clean air must be present at the sensor prior to initiating calibration. The use of bottled air is recommended. • Failed Calibration 2. Calibration can be initiated in one of two ways: • Successful Calibration A. Hold the calibration magnet against the designated location on the sensor module (see Figures 10 and 11) until the green LED turns off and the yellow LED turns on steady (approximately one second). NOTE The calibration procedure must be completed within a ten minute period. If the calibration is not completed, a calibration fault will be generated and the previous calibration data will be used. B. Initiate calibration via the magnetic switches on the UD20. Select Main Menu > Device Cal > Calibration > Execute. NOTE To ensure reliable protection, it is important to check and calibrate the detection system on a regularly scheduled basis. The frequency of these checks is determined by the requirements of the particular installation – typically 30, 60, or 90 day intervals, depending on the ambient conditions. The UD20 indicates “Waiting for Zero.” The transmitter immediately begins taking zero readings. MULTICOLOR LED HART COMMUNICATION PORT (COVER INSTALLED) PLACE CALIBRATION MAGNET HERE TO ACTIVATE INTERNAL REED SWITCH LED EARTH GND LUG PLACE MAGNET HERE TO ACTIVATE INTERNAL MAGNETIC SWITCH WEATHER BAFFLE CALIBRATION MAGNET A2452 C2058 Figure 10—Location of Magnetic Switch on GT3000 Detector 4.1 CALIBRATION NOZZLE Figure 11— Location of Magnetic Switch on PointWatch Eclipse 10 95-8620 Specifications Table 4— PIRECL LEDs and 4-20 mA Output During Various Status Conditions Function Green LED Yellow LED Analog 4-20 mA Signal Warm-up Off Steady On 3.0 Normal Operation Steady On Off 4-20 Fault Condition Off Steady On 3.0 Blocked Optics Off Steady On 3.0 Calibration See Table 2 UD20 Universal display unit operating voltage— 24 Vdc nominal, operating range is 19 to 30 Vdc. UD20 maximum loop voltage drop (Vd) 5.0 volts. Maximum loop supply 30.0 Vdc. operating power— 0.8 watt maximum. operating current— A minimum of 3.5 mA. CURRENT OUTPUT— Linear 4-20 mA (with HART output directly from detector). 250 ohm, 3 watt termination resistor required (included). 3.8 3. When the zero calibration is complete, the yellow LED changes from steady to blinking and the UD20 indicates “Waiting for Gas.” Apply the calibration gas to the sensor. When gas is detected, the UD20 indicates “Waiting for Span.” Operating temperature— –20°C to +70°C. 4. When the yellow LED turns off and the UD20 indicates “Remove Cal Gas”, remove the calibration gas. The gas level at the sensor gradually returns to zero. The green LED turns on steady to indicate that the device has returned to normal operation, using the new calibration data. Humidity Range— 5 to 95% RH (Det-Tronics verified). Storage Temperature— –40°C to +70°C. sensor compatibility— The UD20 can be used with any of the Det‑Tronics GT3000 line of gas detectors, or the Det-Tronics Model PIRECL hydrocarbon gas detector. Input Trim Wiring terminals— Terminals can handle wire sized from 22 to 14 AWG. The input current loop of the UD20 is trimmed at the factory. However, it can also be trimmed in the field using the following automated process. Conduit Entries— 3/4” NPT or M25. (Five conduit entries standard.) Navigate through the menu to “Input Loop Cal.” Main Menu Process Vars Display Status Device Status Display Setup Device Setup Device Cal Device Test ENCLOSURE MATERIAL— Epoxy coated aluminum or 316 stainless steel. SHIPPING WEIGHT— Aluminum: 4.15 pounds (1.88 kilograms). Stainless steel: 10.5 pounds (4.76 kilograms). Display Setup Contrast Control Input Loop Cal DIMENSIONS— See Figures 12 and 13. electro-magnetic compatibility— EMC Directive 2004/108/EC EN55011 (Emissions) EN50270 (Immunity) Upon entering Input Loop Cal, the UD20 commands the detector to output 4 mA, and then automatically calibrates its own input. The UD20 then commands the detector to output 20 mA, and subsequently calibrates its own input. 4.1 WARRANTY— 12 months from date of installation or 18 months from date of shipment, whichever occurs first. 11 95-8620 6.48 (16.5) 3.46 (8.8) 4.7 (11.9) 5.2 (13.2) 5.86 (14.9) 11.28 (28.7) R0.175 A2442 Figure 12—Dimensions of UD20 with GT3000 in Inches (Centimeters) 4.1 12 95-8620 UD20 CERTIFICATIONS— FM: Class I, Div. 1, Groups B, C & D; Class I, Div. 2, Groups B, C & D (T4); Class I, Zone 1/2 AEx d IIC T6; Class II/III, Div. 1/2, Groups E, F & G. Tamb –20°C to +70°C NEMA/Type 4X, IP66 Conduit seal not required. 5.86 (14.9) 5.2 (13.2) 2.7 (6.9) Performance verified in accordance with: ANSI/ISA-92.00.01 FM 6340 ANSI/ISA 60079-29-1 FM 6320 4.7 (11.9) 3.46 (8.8) CSA: CSA 08 2162793. Class I, Div. 1, Groups B, C & D; (Tamb = –20°C to +70°C). Class II/III, Div. 1/2, Groups E, F & G; (Tamb = –20°C to +70°C). Class I, Div. 2, Groups A, B, C & D (T4); (Tamb = –40°C to +70°C) with Bartec Self-Regulating Heating Cable System. (Install per Figure 9.) Type 4X, IP66. Conduit seal not required. FM ® APPROVED 3.77 (9.6) 1.28 (3.3) ATEX: 0539 II 2 G Ex db IIC T6 Tamb –20°C to +70°C FM08ATEX0044X IP66 Performance verified in accordance with: EN 60079-29-1. FM APPROVED C2281 Figure 13—Dimensions of Model STB Termination F M Box in Inches (Centimeters) APPROVED IECEx: Ex db IIC T6 Tamb –20°C to +70°C IECEx FMG 08.0010X IP66 Performance verified in accordance with: IEC 60079-29-1. INMETRO UL-BR 15.0469X Ex d IIC T6 Gb IP66 T6 (Tamb –20°C to +70°C) All cable entry devices shall be Brazil certified in the type of explosion protection, flameproof enclosure “d,” suitable for the conditions of use and correctly installed, with an ingress protection rating of IP66. A screw or cover lock is provided for a secondary means of fastening the cover. 4.1 13 95-8620 DEVICE REPAIR AND RETURN ORDERING INFORMATION Prior to returning devices, contact the nearest local Detector Electronics office so that a Return Material Identification (RMI) number can be assigned. A written statement describing the malfunction must accompany the returned device or component to assist and expedite finding the root cause of the failure. Sensor module, Transmitter module, and termination boxes (if used) must be ordered separately. Refer to the UD20 Model Matrix for ordering details. Replacement Parts/accessories Pack the unit properly. Always use sufficient packing material. Where applicable, use an antistatic bag as protection from electrostatic discharge. Part Number Description 009700-001 Magnetic Tool 009793-002 Electronics Module 010253-001 250 ohm Terminating Resistor Kit 101197-001* Stop Plug, 3/4" NPT, AL 101197-004* Stop Plug, 3/4" NPT, SS 101197-005 Stop Plug, M25, AL, IP66 101197-003 Stop Plug, M25, SS, IP66 010816-001 Stop Plug, 20PK, 3/4" NPT, AL, EX D 010817-001 Stop Plug, 20PK, 3/4" NPT, SS, EX D 010818-001 Stop Plug, 20PK, M25, AL, IP66, EXDE 010819-001 Stop Plug, 20PK, M25, SS, IP66, EXDE 104190-001 Stop Plug, M25, AL, INMETRO 104190-002 Stop Plug, 3/4” NPT, AL, INMETRO 104190-003 Stop Plug, M25, SS, INMETRO 104190-004 Stop Plug, 3/4” NPT, SS, INMETRO 102804-001 Reducer, M25 to M20, AL 102804-003 Reducer, M25 to M20, SS 103922-001 475 Field Communicator 005003-001 Lubriplate grease, 1 oz. 107427-040 O-ring - Viton 104371-001 Bartec Self-Regulating Heating Cable System note Inadequate packaging that ultimately causes damage to the returned device during shipment will result in a service charge to repair the damage incurred during shipment. Return all equipment transportation prepaid to the factory in Minneapolis. NOTE It is highly recommended that a complete spare be kept on hand for field replacement to ensure continuous protection. * NEMA 4/IP66 rating requires addition of non‑hardening thread sealant or Teflon tape. UD20 Model Matrix MODEL UD20 DESCRIPTION Universal Display Unit TYPE MATERIAL A Aluminum S Stainless Steel (316) TYPE THREAD TYPE 5M 5 PORT, METRIC M25 5N 5 PORT, 3/4” NPT TYPE 26 OUTPUTS Pass-Through 4-20 mA, HART TYPE APPROVAL W FM/CSA/ATEX/IECEx B INMETRO (Brazil) TYPE 2 4.1 14 CLASSIFICATION (Div/Zone) Ex d (Flameproof) 95-8620 Appendix A HART Communication – UD20 with GT3000 MENU STRUCTURE This section displays the menu trees for the UD20 connected to the GT3000. The menu tree shows the primary commands and options available when using menu selections. 4.1 15 95-8620 4.1 16 95-8620 PROCESS VARS DISPLAY STATUS DEVICE STATUS DISPLAY SETUP DEVICE SETUP DEVICE CAL DEVICE TEST        MAIN MENU             SELF TEST RESPONSE TEST LOOP TEST D/A TRIM DEVICE TEST   CAL GAS CONC XX.XX CALIBRATION  CHANGE SNSR TYPE DEVICE CAL HART OPTION RTC WRITE PROTECT DEVICE SETUP CONTRAST CONTROL INPUT LOOP CAL DISPLAY SETUP GENERAL INFO FAULT/STATUS TX INFO SENSOR INFO HISTORY DEBUG MENU DEVICE STATUS DISPLAY INFO FAULT/STATUS DEBUG MENU DISPLAY STATUS XXXX X.XX XX.XX MA X.XX X.XX Y/N PROCESS VARS GAS NAME GAS VALUE ANALOG INPUT URV LRV FAULT DISPLAY INFO XX XX XX XX XX XX ZERO TRIM GAIN TRIM D/A TRIM SET 4-20 MA  LOOP TEST EXECUTE ABORT CALIBRATION CHANGE STATE CHANGE PASSWORD WRITE PROTECT XXXXX WRITE PROTECT SECONDS MINUTES HOURS DAY MONTH YEAR TAG XXXXX DESCRIPTOR XXXXX DATE DD/MMM/YYYY MESSAGE XXXXX FINAL ASSY NUM XXXX HART OPTION HART ERRORS LCD TEST ANALOG INPUT RTC XXXXX Y/N Y/N FAULT ANY FAULT CAL ACTIVE WARM UP INPUT HART STATUS ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ADC FAULT ANALOG CODE FLT FLASH FAULT RAM FAULT WDT FAULT 3V FAULT INPUT LOOP FAULT SET 4-20 MA 3.5 MA 4 MA 6 MA 8 MA 10 MA 12 MA 14 MA 16 MA 18 MA 20 MA XX.XX MA XXXXX DEBUG MENU OP MODE FAULT STATUS FAULT/STATUS F/W REV XX.XX SERIAL NUMBER XXXX TEMPERATURE XX.XX C FACTORY CONFIG GENERAL INFO  XX.XX XX.XX XXXXX XXXXX XXXXX XXXXX XX.XX C XXXXX XXXXX Y/N Y/N Y/N Y/N MODBUS ERRORSXXXX DEBUG MENU TX HISTORY  SENSOR HISTORY  CALIBRATION LOG  EVENT LOG  HISTORY SENSITIVITY XXX.XX GAS TYPE XXXXX H/W REV XX.XX F/W REV XX.XX SERIAL NUMBER XXXXX REV XXXXX USL XX.XX LSL XX.XX RUNNING HOURS XXXXX CAL POINT ZERO XX.XX CAL POINT SPAN XX.XX PPM HOURS XXXXX TEMPERATURE XX.XX C SENSOR INFO RTC H/W REV F/W REV SERIAL NUMBER UNIVERSAL REV FLD DEV REV RUNNING HOURS TEMPERATURE TX INFO OP MODE CAL STATE TX STATUS TX FAULT SENSOR STATUS SENSOR FAULT FAULT/STATUS MANUFACTURER  MODEL FGP_TX TAG XXXXX DESCRIPTOR XXXXX DATE DD/MMM/YYYY MESSAGE XXXXX FINAL ASSY NUM XXXXX DEVICE ID XXXXX RTC XX XX XX XX XX XX XXXX XX.XX C XXXXX XX.XX C XXXXX EVENT DATE TIME CAL ID DATE TIME ZERO SPAN TX STATUS ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF CAL FAULT EE FAULT ADC FAULT ADC CNTR FAULT 3V FAULT ZERO DRIFT FAULT FLASH CRC FAULT RAM FAULT TEMP FAULT COMM FAULT ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF SENSOR FAULT SENSOR FAULT WARM UP CAL ACTIVE CAL SWITCH WRITE PROTECT EOL SENSOR CHANGE CONFIG SENSOR STATUS EE FAULT ON/OFF ADC FAULT ON/OFF DAC FAULT ON/OFF LOW VOLTAGE ON/OFF FLASH CRC FAULT ON/OFF RAM FAULT ON/OFF TEMP FAULT ON/OFF SOFTWARE FAULT ON/OFF EE SAFETY FAULT ON/OFF ZERO DRIFT FAULT ON/OFF SENSOR MISMATCH ON/OFF SENSOR FAULT ON/OFF TX FAULT TX FAULT ON/OFF WARM UP ON/OFF CHANGE CONFIG ON/OFF MULTI DROP ON/OFF WRITE PROTECT ON/OFF SELF TEST ON/OFF RESPONSE TEST ON/OFF CURRENT FIXED ON/OFF LOOP TEST ON/OFF FACTORY MODE ON/OFF SNSR ASSY REMOVE ON/OFF UD20 GT3000 March 1, 2013 1.26 Detector Electronics XXXXX DD/MMM/YYYY HH:MM:SS EVENT LOG XXX DD/MMM/YYYY HH:MM:SS XX.XX XX.XX CALIBRATION LOG RUNNING HOURS MAX TEMP MAX TEMP TIME MIN TEMP MIN TEMP TIME SENSOR HISTORY RUNNING HOURS XXXXX MAX TEMP XX.XX C MAX TEMP TIME XXXXX MIN TEMP XX.XX C MIN TEMP TIME XXXXX TX HISTORY SECONDS MINUTES HOURS DAY MONTH YEAR 6901 West 110th Street Minneapolis, MN 55438 USA Appendix B HART Communication – UD20 with PIRECL MENU STRUCTURE This section displays the menu trees for the UD20 connected to the PIRECL. The menu tree shows the primary commands and options available when using menu selections. 4.1 17 95-8620 4.1 18 95-8620 PROCESS VARS DISPLAY STATUS DEVICE STATUS DISPLAY SETUP DEVICE SETUP DEVICE CAL DEVICE TEST MAIN MENU                SELF TEST RESPONSE TEST LOOP TEST D/A TRIM        XX.XX   DD/MMM/YYYY DEVICE TEST CAL GAS CONC CAL GAS TYPE CALIBRATION CAL DATE DEVICE CAL DEVICE OPTION ALARM SETTING HART OPTION RS485 WRITE PROTECT DEVICE SETUP CONTRAST CONTROL INPUT LOOP CAL DISPLAY SETUP GENERAL INFO FAULT/STATUS DEVICE INFO SENSOR INFO HISTORY DEVICE STATUS DISPLAY INFO FAULT/STATUS DEBUG MENU DISPLAY STATUS XXXX X.XX XX.XX MA X.XX X.XX Y/N PROCESS VARS GAS NAME GAS VALUE ANALOG INPUT URV LRV FAULT   XX.XX XX.XX XX.XX XX.XX   XX.XX Y/N XX.XX Y/N   XXX ENABLED ZERO TRIM GAIN TRIM SET 4-20 MA EXECUTE ABORT D/A TRIM LOOP TEST  CALIBRATION SAME AS MEASURE GAS METHANE PROPANE CAL GAS TYPE CHANGE STATE CHANGE PASSWORD WRITE PROTECT WRITE PROTECT BAUD RATE PARITY POLL ADDRESS RS485 TAG XXXXX DESCRIPTOR XXXXX DATE DD/MMM/YYYY MESSAGE XXXXX FINAL ASSY NUM XXXXX HART OPTION RST LATCH ALARMS HGH ALARM LEVEL HGH ALARM LATCH LOW ALARM LEVEL LOW ALARM LATCH ALARM SETTING GAS TYPE UNIT OF MEASURE URV LRV USL LSL ANALOG CODE VAL ANALOG FLT CODE DEVICE OPTION PARITY SET 4-20 MA 3.5 MA 4 MA 6 MA 8 MA 10 MA 12 MA 14 MA 16 MA 18 MA 20 MA NONE EVEN ODD 1200 2400 4800 9600 19.2K BAUD RATE ECLIPSE PIR9400 USER DEFINED ANALOG FAULT CODE XX.XX XX.XX XX.XX XX.XX ANALOG CODE VAL WARM UP BLOCKED OPTIC CALIBRATION FAULT %LFL PPM VOL % UNIT OF MEASURE METHANE ETHANE PROPANE ETHYLENE PROPYLENE BUTANE SPECIAL GAS TYPE XXXXX XXXXX Y/N Y/N DEVICE HISTORY CALIBRATION LOG EVENT LOG HISTORY ACTIVE REFERENCE RATIO ABSORPTION TEMPERATURE VOL % USL LSL SPAN FACTOR    XX.XX XX.XX XX.XX XX.XX XX.XXC XX.XX XX.XX XX.XX XX.XX SENSOR INFO XXXXX XXXXX XXXXX XXXXX DEVICE INFO SERIAL NUMBER UNIVERSAL REV FLD DEV REV S/W REV OP MODE CAL STATE FAULT STATUS FAULT/STATUS MANUFACTURER  MODEL ECLIPSE TAG XXXXX DESCRIPTOR XXXXX DATE DD/MMM/YYYY MESSAGE XXXXX FINAL ASSY NUM XXXXX DEVICE ID XXXXX GENERAL INFO EVENT HOURS XXXXX XX.XX C XXXX XX.XX C XXXX XX.XX C XXXX XX.XX C XXXX EVENT LOG XXXXX XXXXX XXXXX XXXXX CALIBRATION LOG RUNNING HOURS MAX TEMP MAX TEMP TIME MIN TEMP MIN TEMP TIME MAX RESET TEMP MAX RESET TIME MIN RESET TEMP MIN RESET TIME CAL ID HOURS ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF DEVICE HISTORY CAL ACTIVE WARM UP LOW ALARM HIGH ALARM HART SELF TEST CHANGE CONFIG STATUS CAL FAULT DIRTY OPTICS LAMP FAULT START CAL FAULT EE 1 FAULT EE 2 FAULT REF ADC SAT ACTIVE ADC SAT 24V FAULT 12V FAULT 5V FAULT ZERO DRIFT FAULT FLASH CRC FAULT RAM FAULT FAULT 6901 WEST 110TH STREET MINNEAPOLIS, MN 55438 USA FAULT ANY FAULT CAL ACTIVE WARM UP INPUT HART STATUS ADC FAULT ANALOG CODE FLT FLASH FAULT RAM FAULT WDT FAULT 3V FAULT INPUT LOOP FAULT ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF UD20 PIRECL March 1, 2013 1.26 Detector Electronics XX.XX MA XXXXX DEBUG MENU XXXXX Y/N Y/N FAULT/STATUS HART ERRORS LCD TEST ANALOG INPUT OP MODE FAULT STATUS XX.XX XXXX XX.XX C DISPLAY INFO F/W REV SERIAL NUMBER TEMPERATURE FACTORY CONFIG 95-8620 FlexSonic ® Acoustic Leak Detector X3301 Multispectrum IR Flame Detector ­Corporate Office 6901 West 110th Street Minneapolis, MN 55438 USA www.det-tronics.com PointWatch Eclipse ® IR Combustible Gas Detector Phone: 952.946.6491 Toll-free: 800.765.3473 Fax: 952.829.8750 det-tronics@det-tronics.com FlexVu ® Universal Display with GT3000 Toxic Gas Detector Eagle Quantum Premier ® Safety System All trademarks are the property of their respective owners. © 2016 Detector Electronics Corporation. All rights reserved. Det-Tronics manufacturing system is certified to ISO 9001— the world’s most recognized quality management standard.