Emerson 56 Advanced Dual-Input Analyzer Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
56 Instruction Manual
LIQ_MAN_56
May 2014
56 Advanced Dual-Input Analyzer
This page left blank intentionally
Essential Instructions
Read this page before proceeding
Your instrument purchase from Rosemount Analytical, Inc. is one of the finest available for your particular application.
These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in operating and
maintaining the instrument. To ensure their continued operation to the design specifications, personnel should read this
manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If
this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be
impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal
injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount Analytical representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes.
Only connect the product to electrical sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All instrument enclosures must be closed and protective covers must be in place unless qualified personnel are performing maintenance.
WARNING
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
•Installation and servicing of this product may expose personnel to dangerous voltages.
•Main power wired to separate power source must be disconnected before servicing.
•Do not operate or energize instrument with case open!
•Signal wiring connected in this box must be rated at least 240 V for European mains operation.
•Non-metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and jumper wires.
•Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with
Type 4X or IP66 conduit plugs to maintain the ingress protection rating (NEMA 4X)
•Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other
applicable national or local codes.
•Operate only with front panel fastened and in place.
•Safety and performance require that this instrument be connected and properly grounded through a three-wire power
source.
•Proper use and configuration is the responsibility of the user.
CAUTION
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interference. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class
A computing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may
cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
CAUTION
This product is not intended for use in the light industrial, residential or commercial environments per
the instrument’s certification to EN61326-1:2006.
Section i: Quick Start Guide
1. Refer to Section 2.0 for mechanical installation instructions.
2.
Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the
sensor instruction sheet for additional details. Make current output, alarm relay and
power connections
3. Once connections are secured and verified, apply power to the analyzer.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol
appears, consult the manual for appropriate action.
4. When the analyzer is powered up for the first time, Time/Date and Quick Start screens
appear. Quick Start operating tips are as follows:
a. Window screens will appear. The field with the focus will appear with dark blue backlighting. The field with focus can be edited by press ENTER/MENU.
b. The Time and Date screen to set the real-time clock will appear. Accept the displayed
time by pressing ENTER on Time and date OK or press the down key to Change the
time and date.
c. The first Quick Start screen appears. Choose the desired language by pressing
ENTER/MENU to edit the active field and scrolling to the language of choice. Press
ENTER/MENU and press the down arrow to highlight NEXT.
d. The Navigation Rules for operating the keypad will be displayed.
e. Choose the measurement for Sensor 1 (and Sensor 2) and proceed to the remaining
Quick Start steps.
f. Keypad operation guidelines will appear to guide the user how operate the user interface.
g. NOTE: To edit a field with backlit focus, press ENTER/MENU. To scroll up or down, use
the keys to above or below the ENTER key. To move the cursor left or right, use the
keys to the left or right of the ENTER key. To edit a numeric value including decimal
points, use the alphanumeric keypad then press ENTER.
h. NOTE: Press ENTER to store a setting or value. Press EXIT to leave without storing
changes. Pressing EXIT during Quick Start returns the display to the initial start-up
screen (select language). To proceed to the next Quick Start step, use the right key or
the down key to highlight NEXT. Press ENTER.
5.
After the last step, the main display appears. The current outputs are assigned to default
values before probes are wired to the analyzer. After the last step, the main display appears. The outputs are assigned to
default values.
6. To change output, and all settings, press ENTER/MENU from the live screen. Using the
down and right arrow keys, select one of the following menus and navigate the screen of
choice.
7. To return the analyzer to the default settings, choose Reset under the Menu selection
screen.
About This Document
This manual contains instructions for installation and operation of the 56 Advanced Dual-Input
Analyzer. The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
A
08/11
This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect
any changes in the product offering.
B
11/12
Add new feature - configuration transfer via USB. Add new section for
existing features - PID control and TPC relay activation, Non-Incendive Field
Wiring drawings.
C
11/13
Add sec. 9 - Advanced Relay Functions. Add procedure for Software Update
and Configuration Transfer via USB (for units manufactured October 2012
and later). Add procedure for Data Logger and Event Logger Download to
USB. Add new sections for HART and Profibus Communications. Add
Appendix 1 – HART and Device and Appendix 2 - HART Status Bits.
SAFETY MESSAGES
Procedures and instructions in this section may require special precautions to ensure the
safety of the personnel performing the operations. Information that raises potential safety
issues is indicated by a warning symbol (
). This symbol identifies a potential hazard.
CAUTION
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Table of Contents
May 2014
Contents
Section 1: Description and Specifications
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
1.11
1.12
1.13
Features and Applications ................................................................................1
Enhanced Features ..........................................................................................2
Specifications-General ....................................................................................3
Contacting Conductivity .................................................................................7
Toroidal Conductivity ......................................................................................8
pH/ORP ...........................................................................................................6
Flow ................................................................................................................9
4-20mA Current Input .....................................................................................9
Chlorine.........................................................................................................10
Dissolved Oxygen .........................................................................................12
Dissolved Ozone ...........................................................................................12
Turbidity........................................................................................................13
Ordering Information.....................................................................................14
Section 2: Installation
2.1 Unpacking and Inspection .............................................................................15
2.2 Installation.....................................................................................................15
Section: 3 Wiring
3.1
3.2
3.3
3.4
General .........................................................................................................21
Preparing Conduit Openings..........................................................................22
Preparing Sensor Cable ..................................................................................22
Power, Output, Alarms and Sensor Connections ............................................22
Section 4: Display and Operation
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
User Interface ................................................................................................31
Instrument Keypad ........................................................................................31
Main Display ..................................................................................................32
Menu System.................................................................................................32
USB Data Port ................................................................................................33
56 Data Logger and Event Logger Download Procedure .................................33
Software Upgrade .........................................................................................35
Configuration Transfer ...................................................................................35
Section 5: Programming the Analyzer - Basics
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
Table of Contents
General .........................................................................................................37
Changing the Startup Settings ......................................................................37
Programming Temperature ..........................................................................38
Configuring and Ranging the Current Outputs...............................................38
Setting a Security Code .................................................................................39
Security Access ..............................................................................................39
Using Hold ....................................................................................................40
Resetting Factory Defaults – Reset Analyzer ..................................................40
Programming Alarm Relays............................................................................40
i
Table of Contents
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Section 6: Programming - Measurements
6.1 Programming Measurements – Introduction ................................................43
6.2 pH .................................................................................................................43
6.3 ORP ..............................................................................................................44
6.4 Contacting Conductivity ...............................................................................45
6.5 Toroidal Conductivity ....................................................................................46
6.6 Chlorine.........................................................................................................47
6.7 Oxygen..........................................................................................................49
6.8 Ozone ...........................................................................................................50
6.9 Turbidity .......................................................................................................51
6.10 Flow ..............................................................................................................52
6.11 Current Input ................................................................................................52
Section 7: PID Control
7.1 Introduction .................................................................................................55
7.2 PID Setup.......................................................................................................59
Section 8: Time Proportional Control
8.1 Introduction .................................................................................................63
8.2 TPC Setup......................................................................................................63
Section 9: Alarm Relay Functions
9.1 General..........................................................................................................67
9.2 High/Low Concentration Alarm......................................................................67
9.3 Delay Timer: ..................................................................................................68
9.4 Bleed and Feed ..............................................................................................70
9.5 Totalizer Based Relay Activation .....................................................................71
9.6 Interval Timer ................................................................................................72
9.7 Date and Time Activation ..............................................................................74
Section 10: Calibration
10.1 Calibration – Introduction .............................................................................75
10.2 pH Calibration ...............................................................................................75
10.3 ORP Calibration .............................................................................................76
10.4 Contacting Conductivity Calibration .............................................................77
10.5 Toroidal Conductivity Calibration ..................................................................78
10.6 Chlorine Calibration ......................................................................................79
10.7 Oxygen Calibration .......................................................................................82
10.8 Ozone Calibration .........................................................................................84
10.9 Calibrating Temperature................................................................................85
10.10 Turbidity .....................................................................................................85
10.11 Pulse Flow ...................................................................................................86
ii
Table of Contents
56 Analyzer Instruction Manual
Table of Contents
LIQ_MAN_51-56
May 2014
Section 11: HART® Communications
11.1 Introduction ..................................................................................................87
11.2 Physical Installation and Configuration ..........................................................88
11.3 Measurements Available via HART .................................................................89
11.4 Diagnostics Available via HART ......................................................................90
11.5 HART Hosts ...................................................................................................91
11.6 Wireless Communication using the 56 ...........................................................94
11.7 Field Device Specification (FDS) .....................................................................94
Section 12: Profibus Communications
12.1 General ..........................................................................................................95
12.2 Profibus Features ...........................................................................................95
12.3 Profibus Communications..............................................................................96
12.4 Data Transmission .......................................................................................100
12.5 Installation and Wiring.................................................................................111
Section 13: Maintenance
13.1 Overview .....................................................................................................113
13.2 Analyzer Maintenance..................................................................................113
13.3 USB Port ......................................................................................................113
Section 14: Return of Material
14.1 General........................................................................................................115
14.2 Warranty Repair ..........................................................................................115
14.3 Non-Warranty Repair ..................................................................................115
HART Appendix 1
HART Appendix 2
Table of Contents
................................................................................................117
................................................................................................121
iii
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
This page left blank intentionally
iv
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
May 2014
Section 1: Description and Specifications
1.1
Features and Applications
This multi-parameter unit serves industrial, commercial and municipal applications with the
widest range of liquid measurement inputs and digital communications available.
The 56 advanced dual-input analyzer supports continuous measurement of liquid analytical inputs from one or two sensors. The modular design allows signal input boards to be field replaced,
making configuration changes easy. The high resolution full-color display gives unsurpassed
visibility and functionality for liquid analytical instrumentation.
DUAL INPUT INSTRUMENT – single or dual measurement of pH/ORP, Resistivity/Conductivity, % Concentration, Total Dissolved Solids, Total Chlorine, Free Chlorine, Monochloramine,
Dissolved Oxygen, Dissolved Ozone, Turbidity, Pulse Flow, Temperature, and 4-20mA input
from any device.
FULL COLOR DISPLAY: The high resolution full-color display allows at-a-glance viewing of
process readings – indoors or outdoors. Six additional process variables or diagnostic parameters are displayed for quick determination of process or sensor condition. The contrast of backlit display can be adjusted and the main screen can be customized to meet user requirements.
DIGITAL COMMUNICATIONS: HART® version 5 and 7 digital communications are available on
the 56. An optional Profibus® DP digital communications board is available for Profibus installations. 56 HART units communicate with the 475 HART hand-held communicator and HART
hosts such as AMS Intelligent Device Manager. 56 Profibus units are fully compatible with
Profibus DP networks and Class 1 or Class 2 masters. HART and Profibus DP configured units
will support any single or dual measurement configurations of the 56.
MENUS: Easily-managed window screens for easy navigation to local configuration and routine
calibration. Quick Start and all menu screens are available in multiple locally displayed languages. Alpha-numeric keypad allows easy entries during configuration and calibration.
QUICK START PROGRAMMING: Popular Quick Start screens appear the first time the unit is
powered. The instrument auto-recognizes each measurement input type and prompts the
user to configure each sensor loop in a few quick steps for immediate commissioning.
USER HELP SCREENS: A complete user guide and troubleshooting manual is embedded in
the instrument’s memory and easily accessed via the INFO key on the local display. Detailed
instructions and troubleshooting tips in multiple languages are intended to provide adequate guidance to resolve most problems on site.
HAZARDOUS AREA APPROVALS AND SAFETY APPROVALS: None.
ENCLOSURE: The instrument enclosure fits standard DIN panel cutouts. The versatile
enclosure design supports panel-mount, pipe-mount, and surface/wall-mount installations.
No Enclosure ratings – None.
SECURITY ACCESS CODES: Two levels of security access are available. Program one access
code for routine maintenance and hold of current outputs; program another access code for
all configuration menus and functions.
Description and Specifications
1
Section 1: Description and Specifications
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
DIAGNOSTICS: The analyzer continuously monitors itself and the sensor(s) for fault and warning conditions. A display banner flashes red to indicate a Fault condition and yellow for a
Warning condition to visually alert field personnel. Details and troubleshooting information
for any specific fault or warning can be readily accessed by pressing the INFO key.
LOCAL LANGUAGES: Rosemount Analytical extends its worldwide reach by offering nine
menu languages – English, French, German, Italian, Spanish, Portuguese, Chinese, Russian
and Polish. Every unit includes user programming menus; calibration routines; faults and
warnings; and user help screens in all nine languages.
CURRENT OUTPUTS: Every unit includes four 4-20 mA or 0-20 mA electrically isolated current outputs giving the ability to transmit the measurement value and the temperature for
both sensors. Users have wide latitude to assign any measurement value or live diagnostic to
any current output for reporting. Output dampening can be enabled with time constants
from 0 to 999 seconds. HART digital communications transmitted via current output 1 is
standard on all units (option code –HT).
1.2
Enhanced Features
PROCESS TRENDING GRAPHS: High-resolution color graphs of measurement data can be displayed on-screen to pinpoint process disruptions or measurement problems and to estimate
probe maintenance frequency. The analyzer gives the user the ability to zoom in to a specific
narrow timeframe of process measurements for detailed on-screen evaluation.
DATA LOGGER AND EVENT LOGGER: Extensive onboard data storage captures measurement
data from both channels every 30 seconds for 30 days for on-screen display or local upload to
a USB 2.0 memory device. 300 significant analyzer events are recorded including start-up
time, calibrations, hold outputs, configurations, alarms, power interruptions, faults, and
more. All process data and events are time/date stamped.
USB 2.0 DATA TRANSFER PORT: A USB port is built-in to allow local data transfer of process
data and events using a standard USB memory device. Cleanly formatted EXCEL data is useful for evaluation of process data on a computer and identification of critical alarm or fault
events.
PID CONTROL: Proportional, Integral and Derivative settings allow the analog current outputs to adjust a control device that has continuous adjustability by acting on process measurements or temperature. PID is typically used on modulating control devices such as
automated control valves or variable volume pumps. Any current output can be programmed for PID functions.
ALARM RELAY CAPABILITIES: Four Single Pole Double Throw alarm relays are fully assignable and programmable to trigger alarms upon reaching measurement or diagnostics setpoints or fault conditions. Further relay settings include TPC, synchronized interval timers
and four specialized timer functions described below. All relays are independently activated.
Failsafe operation and programming of relay default state (normally open or normally closed)
is software selectable.
TIMER FUNCTIONS: Basic TPC (Time Proportional Control) settings are available. Interval
timers set relays by interval time, on-time and recovery time for discrete on/off control devices based on measurement inputs. In addition, four real-time clock relay functions are implemented including: bleed and feed, day and time interval timers, delay timer and a flow
totalizer. These advanced timer features support a number of specialized applications that
normally require dedicated timer control devices or DCS programming.
2
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
May 2014
WIRELESS THUM ADAPTOR COMPATIBLE: Enable wireless transmissions of process variables and diagnostics from hard-to-reach locations where it is impractical to run wires for
current outputs. When commissioned with the THUM Adaptor, 56 HART® units can communicate on Emerson wireless networks using HART 7 wireless protocol.
SMART-ENABLED PH: Rosemount Analytical’s SMART pH capability can eliminate field calibration of pH probes through automatic upload of calibration data and history – fully calibrating
the pH loop. pH probe changes are literally plug and play using SMART pH sensors with VP
cables connections.
ADVANCED FUNCTIONS: Several specialty measurements are supported including: high reference impedance pH sensors, Ion Selective Electrode measurements, pH loop calibration by
entering pH slope and reference offset, Isopotential point for pH, inferred pH determination
using dual contacting conductivity inputs, differential conductivity, differential flow, totalized
flow, current input from any 4-20mA source, dual range calibration for chlorine sensors, programmable polarizing voltage for amperometric oxygen sensors and software selectable normally
open or normally closed alarm relays – to name a few.
1.3
Specifications - General
Case: Polycarbonate. Type 4X, IP66.
NOTE:
To ensure a water-tight seal, tighten all four front panel screws to 6 in-lbs of torque.
Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132mm)
Conduit openings: Accepts (6) PG13.5 or 1/2 in. conduit fittings
Display: Large 3.75 x 2.2 in. (95.3 x 55.9mm) high resolution color LCD displays large
process variables and user-definable display of diagnostic parameters. Calibration, programming and information screens display clear, easy-to-read characters. The color display is
back-lit and backlighting intensity is user adjustable. Measurement character height: (.5")
13mm. Main display can be customized to meet user requirements.
Ambient temperature and humidity: -10 to 60°C, (14 to 140°F) RH 5 to 95% (non-condensing). For Turbidity only: 0 to 55°C (32 to 131°F). RH 5 to 95% (non-condensing).
NOTE:
The analyzer is operable from -5 to 55°C (-23 to 131°F) with some degradation in display response or performance. Above 60°C, the following components will progressively and automatically shut down: display,
USB communications port, current outputs, alarm relays, main circuit board.
WARNING
Always remove USB memory device at ambient temp above 60°C. Do not access USB port if
combustible atmosphere is present.
Storage temperature: -20 to 60°C, (-4 to 140°F)
Power: Code -02: 20 to 30 VDC. 20 W
Code –03: 85 to 264 VAC, 47.5 to 65.0 Hz, 20 W
Real time clock back-up: 24 hours.
Description and Specifications
3
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Hazardous Location Approvals:
Options for CSA: 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, HT and DP
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4A
Tamb= 50°C
Enclosure Type 4X
See Non-Incendive Field Wiring drawing 1400668. Evaluated to the ANSI/UL Standards.
The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the product has been
evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S.
respectively.
NOTE: Single-input Turbidity configurations (models 56-02-27-38 or -HT, 56-03-27-38 or -HT) and
dual-input Turbidity only configurations (56-02-27-37 or -HT, 56-03-27-37 -HT) are CSA approved class
I Div. 2 for hazardous area installation.
Options for FM: -02, 03, 20, 21, 22, 23, 24, 25, 26, 27, 30, 31, 32, 33, 34, 35, 36, 37,
38, HT and DP.
Class I, Division 2, Groups A, B, C, & D
Class Il & lll, Division 2, Groups E, F, & G
T4A Tamb= 50°C
IP66
See Non-Incendive Field Wiring drawing 1400667.
NOTE: Single-input Turbidity configurations (models 56-02-27-38 or -HT, 56-03-27-38 or -HT) and
dual-input Turbidity only configurations (56-02-27-37 or -HT, 56-03-27-37 or -HT) are FM approved
class I Div. 2 for hazardous area installation.
Ordinary Locations (only with -UL ordering option):
Options for UL: -02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, HT
and DP.
LISTED
Pollution Degree 2: Normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by condensation must
be expected.
Altitude: for use up to 2000 meter (6562 ft.)
RFI/EMI: EN61326-1:2006
LVD: EN-61010-1:2001
Input: One or two isolated sensor inputs. Measurement choices of pH/ORP, resistivity/conductivity/ TDS, % concentration, ratio conductivity, total and free chlorine, monochloramine, dissolved oxygen, dissolved ozone, turbidity, pulse flow, temperature and raw 4-20mA input.
For contacting conductivity measurements, temperature element must be a Pt1000 RTD.
For other measurements (except ORP, flow and turbidity), use either a PT100 RTD, PT1000
RTD, or 22k NTC (D.O. only).
4
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
May 2014
Outputs: Four 4-20 mA or 0-20 mA isolated current outputs. Fully scalable. Max Load: 550
Ohms. Output 1 superimposes the HART digital signal. Outputs can be programmed for
PID control. Output dampening can be enabled with time constants from 0 to 999 seconds.
HART digital communications transmitted via current output 1 is standard on all units
(option code –HT).
Alarms: Four alarm relays for process measurement(s) or temperature. Any relay can be programmed for any measurement, timer, TPC or fault alarm operation, instead of a process
alarm. When selected, a fault alarm will activate the relay when a sensor or analyzer fault occurs.
Each relay can be configured independently. Alarm logic (high or low activation or USP*) and
deadband are user-programmable.
*USP alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the
limit. conductivity/resistivity measurement only)
Relays: Form C, SPDT, epoxy sealed
Maximum Relay Current
Power Input
Resistive
28 VDC 5.0 A
5.0 A
115 VAC 5.0 A
5.0 A
230 VAC 5.0 A
5.0 A
Inductive load: 1/8 HP motor (max.), 115/240 VAC
Terminal Connections Rating:
Power connector (-02 24VDC power supply and -03 85-264VAC power supply): 24-12 AWG
wire size.
Signal board terminal blocks: 26-16 AWG wire size.
Current output connectors: 26-16 AWG wire size.
Alarm relay terminal blocks: 24-12 AWG wire size.
Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 3 lbs/4 lbs
(1.5 kg/2.0 kg)
Description and Specifications
5
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
1.4
LIQ_MAN_51-56
Contacting Conductivity (Codes -20 and -30)
Measures conductivity in the range 0 to 600,000 µS/cm (600mS/cm). Measurement choices are
conductivity, resistivity, total dissolved solids, salinity, and % concentration. Temperature compensation can be disabled, allowing the analyzer to display raw conductivity.
NOTE:
When two contacting conductivity sensors are used, The 56 can derive an inferred pH value. Inferred
pH is calculated pH, not directly measured pH. Inferred pH is calculated from straight and cation conductivity. It is applicable only if the alkalizing agent is NaOH or NH3 and the major contaminant is NaCl.
It is strictly an application for power plants.
Performance Specifications - Analyzer
Measurement Range: see table below
Solution temperature compensation: manual slope (X%/°C), high purity water (dilute
sodium chloride), and cation conductivity (dilute hydrochloric acid).
Salinity: uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying conductivity at 25ºC by 0.65
Five percent concentration curves: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, 0-25% or 9699.7% H2SO4. The conductivity concentration algorithms for these solutions are fully temperature compensated.
Four temperature compensation options: manual slope (X%/°C), high purity water (neutral
salt), cation conductivity (dilute hydrochloric acid) and raw.
Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading
Recommended Sensors for Contacting Conductivity:
All Rosemount Analytical ENDURANCE 400 series conductivity sensors (Pt 1000 RTD) and
410VP 4-electrode high-range conductivity sensor.
PERFORMANCE SPECIFICATIONS
Recommended Range – Contacting Conductivity
Cell
0.01S/cm
Constant
0.01
0.1mS/cm
1.0mS/cm
10mS/cm
100mS/cm
1000mS/cm
0.01mS/cm to 200mS/cm
0.1
100mS/cm
1000mS/cm
200mS/cm to 6000mS/cm
0.1mS/cm to 2000mS/cm
1.0
10mS/cm
2000mS/cm to 60mS/cm
1 mS/cm to 20mS/cm
20mS/cm to 600mS/cm
4-electrode
2 mS/cm to 1400mS/cm
Temperature Specifications:
Temperature range
0-200ºC
Temperature Accuracy,
Pt-1000, 0-50 ºC
± 0.1ºC
Cell Constant Linearity
±0.6% of reading in recommended range
+2 to -10% of reading outside high recommended range
±5% of reading outside low recommended range
Temperature Accuracy,
Pt-1000, Temp. > 50 ºC
6
± 0.5ºC
±4% of reading in recommended range
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
1.5
May 2014
Toroidal Conductivity (Codes -21 and -31)
Measures conductivity in the range of 1 (one) µS/cm to 2,000,000 µS/cm (2 S/cm). Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration.
Temperature compensation can be disabled, allowing the analyzer to display raw conductivity.
For more information concerning the use and operation of the toroidal conductivity sensors,
refer to the product data sheets.
Performance Specifications- Analyzer
Measurement Range: see table below
Repeatability: ±0.25% ±5 µS/cm after zero cal
Salinity: uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying conductivity at 25ºC by 0.65
Five percent concentration curves: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, 0-25% or 9699.7% H2SO4. The conductivity concentration algorithms for these solutions are fully temperature compensated. For other solutions, the analyzer accepts as many as five data points
and fits
either a linear (two points) or a quadratic function (three or more points) to the data. Reference temperature and linear temperature slope may also be adjusted for optimum results.
Three temperature compensation options: manual slope (X%/°C), neutral salt (dilute
sodium chloride) and raw.
Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading
Recommended Sensors:
All Rosemount Analytical submersion/immersion and flow-through toroidal sensors.
Loop Performance (Following Calibration)
Temperature range
-25 to 210ºC
(-13 to 410ºF)
226: ±1% of reading ±5 S/cm in recommended range
Temperature Accuracy,
Pt-100, -25 to 50 ºC
± 0.5ºC
225 & 228: ±1% of reading ±10µS/cm in
recommended range
222, 242: ±4% of reading in recommended range
Temperature Accuracy,
Pt-100,. 50 to 210ºC
± 1ºC
225, 226 & 228: ±5% of reading outside high
recommended range
226: ±5µS/cm outside low recommended range
225 & 228: ±15µS/cm outside low recommended
range
PERFORMANCE SPECIFICATIONS Recommended Range - Toroidal Conductivity
Model
1mS/cm
10mS/cm
226
100mS/cm
1000mS/cm
10mS/cm
100mS/cm
5mS/cm to 500mS/cm
1000mS/cm
2000mS/cm
500mS/cm to 2000mS/cm
225 & 228
15mS/cm to 1500mS/cm
242
222
(1in & 2in)
Description and Specifications
1500mS/cm to 2000mS/cm
100mS/cm to 2000mS/cm
500mS/cm to 2000mS/cm
7
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
1.6
LIQ_MAN_51-56
pH/ORP (Codes -22 and -32)
For use with any standard pH or ORP sensors. Measurement choices are pH, ORP, Redox, Ammonia, Fluoride or custom ISE. The automatic buffer recognition feature uses stored buffer pH values
and their temperature curves for the most common buffer standards available worldwide. The analyzer will recognize the pH value of the buffer being measured and perform a self stabilization
check on the sensor before completing the calibration. Manual or automatic temperature compensation is menu selectable. Change in process pH due to temperature can be compensated
using a programmable temperature coefficient. For more information concerning the use and operation of the pH or ORP sensors, refer to sensor product data sheets. The 56 can also derive an
inferred pH value. Inferred pH can be derived and displayed when two contacting conductivity
sensors are used. (56-0X-20-30-XX)
Performance Specifications (pH input) - Analyzer
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ ºC
Solution temperature correction: pure water, high pH (dilute base), Ammonia and custom
Buffer recognition: NIST (including non-NIST pH 7.01 buffer), DIN 19267, Ingold, Merck, and
Fisher
Input filter: Time constant 1 - 999 sec, default 4 sec.
Response time: 5 seconds to 95% of final reading
Recommended Sensors for pH:
Compatible with standard pH sensors with and without integral preamps. Supports Smart
pH sensors from Rosemount Analytical (includes Smart integral preamps).
General purpose and high
performance pH 396PVP,
3900VP and 3300HT sensors
Performance Specifications (ORP input) - Analyzer
Measurement Range [ORP]: -1500 to +1500 mV
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV / ºC
Input filter: Time constant 1 - 999 sec, default 4 sec.
Response time: : 5 seconds to 95% of final reading
Recommended Sensors for ORP:
Compatible with standard ORP sensors with and without integral preamps.
NOTE:
Some older sensor preamps may not be compatible with the 56 (contact the factory for details).
8
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
1.7
May 2014
Flow (Code -23 and -33)
For use with most pulse signal flow sensors, the 56 user-selectable units of measurement include flow rates in GPM (gallons per minute), GPH (gallons per hour), cu ft/min (cubic feet per
min), cu ft/hour (cubic feet per hour), LPM (liters per minute), LPH (liters per hour), or m3/hr
(cubic meters per hour), and velocity in ft/sec or m/sec. When configured to measure flow, the
unit also acts as a totalizer in the chosen unit (gallons, liters, or cubic meters). Dual flow instruments can be configured as a % recovery, flow difference, flow ratio, or total (combined) flow.
Performance Specifications - Analyzer
Frequency Range: 3 to 1000 Hz
Flow Rate: 0 - 99,999 GPM, LPM, m3/hr, GPH, LPH, cu ft/min, cu ft/hr.
Totalized Flow: 0 – 9,999,999,999,999 Gallons or m3, 0 – 999, 999,999,999 cu ft.
Accuracy: 0.5%
Input filter: Time constant 0-999 sec., default 5 sec.
1.8
4-20mA Current Input (Codes -23 and -33)
For use with any transmitter or external device that transmits 4-20mA or 0-20mA current
outputs. Typical uses are for temperature compensation of live measurements (except ORP,
turbidity and flow) and for continuous pressure input for continuous measurement of % oxygen
gas. External input of atmospheric pressure for oxygen measurement allows continuous partial
pressure compensation while the 56 enclosure is completely sealed.
Externally sourced current input is also useful for calibration of new or existing sensors that
require temperature measurement or atmospheric pressure inputs. In addition to live continuous compensation of live measurements, the current input board can also be used simply to
display and trend the measured temperature or the calculated partial pressure from the external device. This feature leverages the large display variables on the 56 as a convenience for technicians. Temperature can be displayed in degrees C or degrees F. Partial pressure can be
displayed in inches Hg, mm Hg, atm (atmospheres), kPa (kiloPascals), bar or mbar. The current input board serves as a power supply for loop-powered devices that do not actively
power their 4-20mA output signals.
Performance Specifications
Measurement Range *[mA]: 0-20 or 4-20
Accuracy: ±0.03mA
Input filter: Time constant 0-999 sec., default 5 sec.
*Current input not to exceed 22mA
Description and Specifications
9
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
1.9
LIQ_MAN_51-56
Chlorine (Code -24 and -34)
Free and Total Chlorine
The 56 is compatible with the 499ACL-01 free chlorine sensor and the 499ACL-02 total chlorine sensor. The 499ACL-02 sensor must be used with the TCL total chlorine sample conditioning system. The 56 fully compensates free and total chlorine readings for changes in
membrane permeability caused by temperature changes.
For free chlorine measurements, both automatic and manual pH corrections are available.
For automatic pH correction, select code -32 and an appropriate pH sensor. For more information concerning the use and operation of the amperometric chlorine sensors and the TCL
measurement system, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100µA
Automatic pH correction (requires Code -32): 6.0 to 10.0 pH
Temperature compensation: Automatic or manual (0-50°C).
Input filter: Time constant 1 - 999 sec, default 5 sec.
Response time: 6 seconds to 95% of final reading
Recommended Sensors
Chlorine: 499ACL-01 Free Chlorine or 499ACL-02 Total Chlorine
pH: The following pH sensor is recommended for automatic pH correction of free chlorine
readings: 3900
Monochloramine
The 56 is compatible with the 499A CL-03 Monochloramine sensor. The 56 fully
compensates readings for changes in membrane permeability caused by temperature
changes. Because monochloramine measurement is not affected by pH of the process, no
pH sensor or correction is required. For more information concerning the use and operation
of the amperometric chlorine sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100µA
Temperature compensation: Automatic or manual (0-50°C).
Input filter: Time constant 1 - 999 sec, default 5 sec.
Response time: 6 seconds to 95% of final reading
Recommended Sensors
Rosemount Analytical 499ACL-03 Monochloramine sensor
10
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
May 2014
pH-Independent Free Chlorine
The 56 is compatible with the 498CL-01 pH-independent free chlorine sensor. The 498CL-01
sensor is intended for the continuous determination of free chlorine (hypochlorous acid plus
hypochlorite ion) in water. The primary application is measuring chlorine in drinking water. The
sensor requires no acid pre-treatment, nor is an auxiliary pH sensor required for pH correction.
The 56 fully compensates free chlorine readings for changes in membrane permeability caused
by temperature. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100µA
pH independent
Temperature compensation: Automatic (via RTD) or manual (0-50°C).
Input filter: Time constant 1 - 999 sec, default 5 sec.
Response time: 6 seconds to 95% of final reading
Recommended Sensors
Rosemount Analytical 498CL-01 pH independent free chlorine sensor
Chlorine sensors with Variopol
connection and cable connection
498CL-01
Description and Specifications
11
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
1.10
LIQ_MAN_51-56
Dissolved Oxygen (Codes -25 and -35)
The 56 is compatible with the 499ADO, 499ATrDO, Hx438, Gx438 and BX438 dissolved
oxygen sensors and the 4000 percent oxygen gas sensor. The 56 displays dissolved oxygen in
ppm, mg/L, ppb, µg/L, % saturation, % O2 in gas, ppm O2 in gas. The analyzer fully
compensates oxygen readings for changes in membrane permeability caused by
temperature changes. An atmospheric pressure sensor is included on all dissolved oxygen
signal boards to allow automatic atmospheric pressure determination during air calibration.
Calibration can be corrected for process salinity if removing the sensor from the process liquid
is impractical. The analyzer can be calibrated against a standard instrument. For more
information on the use of amperometric oxygen sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor (when O2 <1.00 ppm); 0.1%
Input Range: 0nA – 100µA
Temperature Compensation: Automatic or manual (0-50°C).
Input filter: Time constant 1 - 999 sec, default 5 sec
Response time: 6 seconds to 95% of final reading
Recommended Sensor
Rosemount Analytical amperometric membrane and steam-sterilizable sensors listed above
1.11
Dissolved Ozone (Code -26 and -36)
The 56 is compatible with the 499AOZ sensor. The 56 fully compensates ozone readings for
changes in membrane permeability caused by temperature changes. For more information
concerning the use and operation of the amperometric ozone sensors, refer to the product
data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100µA
Temperature Compensation: Automatic or manual (0-35°C)
Input filter: Time constant 1 - 999 sec, default 5 sec.
Response time: 6 seconds to 95% of final reading
Recommended Sensor
Rosemount Analytical 499A OZ ozone sensor.
Dissolved Oxygen 499ADO
sensor with Variopol
connection
12
Dissolved Ozone 499AOZ
sensors with Polysulfone
body Variopol
connection and cable
connection
Description and Specifications
56 Analyzer Instruction Manual
Section 1: Description and Specifications
LIQ_MAN_51-56
1.12
May 2014
Turbidity (Codes -27 and -37)
The 56 instrument is available in single and dual turbidity configurations for the Clarity II turbidimeter. It is intended for the determination of turbidity in filtered drinking water. The other
components of the Clarity II turbidimeter – sensor(s), debubbler/measuring chamber(s), and
cable for each sensor must be ordered separately or as a complete system with the 56.
The 56 turbidity instrument accepts inputs from both USEPA 180.1 and ISO 7027compliant sensors. Four fully programmable relays with timers are included.
Note: the 56 Turbidity must be used with Clarity II sensor, sensor cable and debubbler.
Performance Specifications - Analyzer
Units: Turbidity (NTU, FTU, or FNU); total suspended solids (mg/L, ppm, or no units)
Display resolution-turbidity: 4 digits; decimal point moves from x.xxx to xxx.x
Display resolution-TSS: 4 digits; decimal point moves from x.xxx to xxxx
Calibration methods: User-prepared standard, commercially prepared standard, or grab
sample. For total suspended solids user must provide a linear calibration equation.
Inputs: Choice of single or dual input, EPA 180.1 or ISO 7027 sensors.
Field wiring terminals: Removable terminal blocks for sensor connection.
Accuracy after calibration at 20.0 NTU:
0-1 NTU ±2% of reading or 0.015 NTU, whichever is greater. 0-20 NTU: ±2% of reading.
Description and Specifications
13
Section 1: Description and Specifications
56 Analyzer Instruction Manual
May 2014
1.13
LIQ_MAN_51-56
Ordering Information
The 56 Analyzer offers single or dual sensor input with an unrestricted choice of dual
measurement combinations. Measurements capabilities include pH/ORP, Resistivity/
Conductivity, % Concentration, Total Chlorine, Free Chlorine, Monochloramine, Dissolved
Oxygen, Dissolved Ozone, Turbidity, Pulse Flow, Temperature, and 4-20mA input.
The device includes two isolated inputs, nine local languages, four 4-20mA current outputs,
removable connectors for power and current outputs, and four solid plugs for closure of
openings. HART digital communications is included at no additional charge. Profibus digital
communications is optional.
56 Advanced Dual-Input Analyzer
Level 1
02
03
POWER
24 VDC with four alarm relays
85-265 VAC switching, 50/60 Hz with four alarm relays
Level 2
20
21
22
23
24
25
26
27
MEASUREMENT 1
Contacting Conductivity
Toroidal Conductivity
pH/ORP
Flow/Current Input
Chlorine
Dissolved Oxygen
Ozone
Turbidity
Level 3 MEASUREMENT 2
30
Contacting Conductivity
31
Toroidal Conductivity
32
pH/ORP/ISE
Flow/Current Input
33
34
Chlorine
35
Dissolved Oxygen
Ozone
36
37
Turbidity
38
None
Level 4
COMMUNICATIONS
HT
HART® digital communication
DP
Profibus DP digital communication
14
Ordering Information
56 Analyzer Instruction Manual
Section 2: Installation
LIQ_MAN_51-56
May 2014
Section 2.0 – Installation
2.1
Unpacking and Inspection
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
Save the box. If there is no apparent damage, unpack the container. Be sure all items shown
on the packing list are present. If items are missing, notify Rosemount Analytical
immediately.
2.2
Installation
2.2.1
General Information
1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in
areas of extreme temperatures is not recommended unless a sunshield is used. Make
sure to note the Ambient temperature specifications in section 1. The analyzer cannot
be operated in ambient (shaded) conditions greater than 60°C.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency
interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be
sure there is easy access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. See Figures 2-1 and 2-2.
5. Install cable gland fittings and plugs as needed to properly seal the analyzer on all six enclosure openings. The USB port cover must be fully installed on the front cover to ensure
proper analyzer sealing.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol
appears, consult the manual for appropriate action.
Installation
15
Section 2: Installation
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Fig. 2-1 56 Panel Mounting Installation dimensions
MILLIMETER
INCH
16
Installation
56 Analyzer Instruction Manual
Section 2: Installation
LIQ_MAN_51-56
May 2014
Fig. 2-2 56 Pipe and Wall Mounting Installation dimensions
MILLIMETER
INCH
Shown with Mounting Kit PN 23820-00
Installation
17
Section 2: Installation
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Fig. 2-3 FM Non-incendive field wiring installation for the 56-27-37 Analyzer
18
Installation
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 2: Installation
May 2014
Fig. 2-4 CSA Non-incendive field wiring installation
Installation
19
Section 2: Installation
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Fig. 2-4 CSA Non-incendive field wiring installation
20
Installation
56 Analyzer Instruction Manual
Section 3: Wiring
LIQ_MAN_51-56
May 2014
Section 3.0 Wiring
3.1
General
The 56 is easy to wire. It includes removable connectors and slide-out signal input boards
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring
locations.
3.1.1
Removable connectors and signal input boards
The 56 uses removable signal input boards and communication boards for ease of wiring and
installation. Each of the signal input boards can be partially or completely removed from the
enclosure for wiring. The 56 has three slots for placement of up to two signal input boards
and one communication board
3.1.2
Slot 1-Left
Slot 2 – Center
Slot 3 – Right
Profi board
Signal Board 1
Signal Board 2
Signal input boards
Slots 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board following the lead locations marked on the board. After wiring the sensor leads to
the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten the cable gland nut to secure the cable and
ensure a sealed enclosure.
NOTE:
For the purpose of replacing factory-installed signal input boards, Rosemount® Analytical Inc. is the sole
supplier.
3.1.3
Digital communications
®
HART® digital communications is standard on 56. HART versions 5 and 7 are available on the
56 and can be switched using the local keypad. A Profibus DP communication board is available as options for 56 communication with a host. HART communications supports Bell 202
digital communications over an analog 4-20mA current output. Profibus DP is an open communications protocol which operates over a dedicated digital line to the host.
3.1.4
Alarm relays
Four alarm relays are supplied with the switching power supply (85 to 264VAC, -03 order
code) and the 24VDC power supply (20-30VDC, -02 order code). All relays can be used for
process measurement(s) or temperature. Any relay can be configured as a fault alarm instead of
a process alarm. Each relay can be configured independently and each can be programmed
as an interval timer, typically used to activate pumps or control valves. As process alarms,
alarm logic (high or low activation or USP*) and deadband are user-programmable. Customerdefined failsafe operation is supported as a programmable menu function to allow all relays
to be energized or not-energized as a default condition upon powering the analyzer. The
USP* alarm can be programmed to activate when the conductivity is within a user-selectable
percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is installed.
Wiring
21
Section 3: Wiring
56 Analyzer Instruction Manual
May 2014
3.2
LIQ_MAN_51-56
Preparing Conduit Openings
There are six conduit openings in all configurations of 56. (Note that four of the openings will
be fitted with plugs upon shipment.)
Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with Type 4X or IP66 conduit plugs.
NOTE:
Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer.
3.3
Preparing Sensor Cable
The 56 is intended for use with all Rosemount Analytical sensors. Refer to the sensor
installation instructions for details on preparing sensor cables.
3.4
Power, Output, and Sensor Connections
3.4.1
Power wiring
Two Power Supplies are offered for the 56:
a.
24VDC (20 – 30V) Power Supply (-02 ordering code)
b.
85 – 265 VAC Switching Power Supply (-03 ordering code)
AC mains leads and 24VDC leads are wired to the Power Supply board which is mounted vertically
on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board. Wire the power leads to the Power Supply board using the lead markings on the board.
The grounding plate is connected to the earth terminal of the -03 (85-265VAC) power supply. The green colored screws on the grounding plate are intended for connection to some
sensors to minimize radio frequency interference. The green screws are not intended to be
used for safety purposes.
3.4.2
Current output wiring
All instruments are shipped with four 4-20mA current outputs. Wiring locations for the outputs are on the Main board which is mounted on the hinged door of the instrument. Wire the
output leads to the correct position on the Main board using the lead markings (+/positive, /negative) on the board. Male mating connectors are provided with each unit.
3.4.3
Alarm relay wiring
Four alarm relays are supplied with the switching power supply (85 to 265VAC, -03 order
code) and the 24VDC power supply (20-30VDC, -02 order code). Wire the relay leads on each
of the independent relays to the correct position on the power supply board using the
printed lead markings (NO/Normally Open, NC/Normally Closed, or Com/Common) on the
board.
22
Wiring
56 Analyzer Instruction Manual
Section 3: Wiring
LIQ_MAN_51-56
3.4.4
May 2014
Sensor wiring to signal boards
Wire the correct sensor leads to the measurement board using the lead locations marked directly on the board. After wiring the sensor leads to the signal board, carefully slide the wired
board fully into the enclosure slot and take up the excess sensor cable through the cable gland.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded
metal conduit. Connect the shield to earth ground. AC wiring should be 14 gauge or greater.
Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the
switch or breaker near the analyzer and label it as the disconnecting device for the analyzer.
Keep sensor and output signal wiring separate from power wiring. Do not run sensor and
power wiring in the same conduit or close together in a cable tray.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol appears, consult the manual for appropriate action.
Wiring
23
Section 3: Wiring
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
FIGURE 3-1 Power Wiring for 56 24VDC Power Supply (-02 ordering code) PN 24365-00
FIGURE 3-2 Power Wiring for 56 85-264 VAC Power Supply (-03 ordering code)
PN 24358-00
24
Wiring
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 3: Wiring
May 2014
FIGURE 3-3 Output Wiring for 56 Main PCB PN 24308-00
Wiring
25
Section 3: Wiring
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Figure 3-4 56 Recommended Wire Entry Points
Figure 3-5 56 Recommended Wire Entry and THUM Adaptor Installation
26
Wiring
56 Analyzer Instruction Manual
Section 3: Wiring
LIQ_MAN_51-56
May 2014
Figure 3-6 Contacting Conductivity signal board and Sensor cable leads
Figure 3-7 Toroidal Conductivity Signal board and Sensor cable leads
Wiring
27
Section 3: Wiring
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Figure 3-8 pH/ORP/ISE signal board and Sensor cable leads
Figure 3-9 Amperometric signal (Chlorine, Oxygen, Ozone) board and Sensor cable leads
28
Wiring
56 Analyzer Instruction Manual
Section 3: Wiring
LIQ_MAN_51-56
May 2014
Figure 3-10 Turbidity signal board with plug-in Sensor connection
Figure 3-11 Flow/Current Input signal board and Sensor cable leads
Wiring
29
Section 3: Wiring
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
This page left blank intentionally
30
Wiring
56 Analyzer Instruction Manual
Section 4: Display and Operation
LIQ_MAN_51-56
May 2014
Section 4.0 Display and Operation
4.1
User Interface
The 56 has a large display which shows two live
measurement readouts in large digits and up to
six additional process variables or diagnostic
parameters concurrently. The display is back-lit
and the format can be customized to meet user
requirements. The ENTER/MENU key allows access
to Calibration, Hold (of current outputs),
Programming, Display, Data and HART® functions.
In addition, a dedicated INFO key is available to
provide access to useful diagnostic and
instrument information regarding installed
sensor(s) and any problematic conditions. The
display flashes a red banner to indicate a Fault
condition and a yellow banner for a Warning
condition. Help screens are displayed for fault and
warning conditions to guide the user in troubleshooting. During calibration and programming,
key presses guide the user step-by-step through procedures. An alpha-numeric keypad similar
to a cell phone keypad is available to allow the user to enter data during programming and
calibration or lengthy tags to describe process points, sensors, or instrumentation.
4.2
Instrument Keypad
There are three Function keys, four Navigation keys and an alpha-numeric keypad on the
instrument keypad.
Function keys
The ENTER/MENU key is used to access menus for programming and calibrating the instrument as well as retrieving stored data. Eight top-level menu items appear when pressing the
ENTER/MENU key from the main display of live readings:
• Calibrate: calibrate attached sensors and analog outputs.
• Program: Program outputs, relays, measurement, temperature, and security codes.
• Hold: Suspend current outputs.
• Display Setup: Program graphic trend display, brightness, main display format, tags,
language, and warnings.
• Data storage and retrieval: Enable data and event storage, download data, and vie
events.
• HART or Profibus: Program HART and Profibus communication parameters.
• Time and Date: Set and view real-time clock settings.
• Reset: Reset all instrument settings, calibration settings or current outputs to factory defaults.
Calibrate
Data storage and retrieval
Program
HART
Hold
Time and Date
Display setup
Reset
Software upgrade
Display and Operation
31
Section 4: Display and Operation
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
The ENTER/MENU key is also used to enter selections or enable programming and calibration steps.
The EXIT key returns to the previous menu level.
The INFO key provides detailed instructions and explanations during programming and calibrating procedures. It also provides troubleshooting tips for all faults and warnings that may occur
during calibration or continuous operation in process.
Navigation Keys
The four Navigation keys arranged around the ENTER/MENU key operate in an intuitive manner similar to the navigation keys on a computer keyboard. During menu operation, these keys
are used to move the highlighted screen selection to another adjacent screen item. During tag
entry, the left key is used to delete entries during active alpha-numeric character entry.
Alpha-numeric Keypad
The alpha-numeric keypad has 12 keys as outlined below.
• Nine keys are alpha-numeric
• One key is a dedicated “1” key
• One key is a dedicated “0” key
• One key is a dedicated “.” (decimal point) key
The alpha-numeric keypad operates the same as entries on a mobile phone. The nine alphanumeric keys have multiple characters that can be entered for tag entries or during programming and calibration steps. Character selections are made by pressing the key multiple times to
toggle to characters that are available on each key.
4.3
Main Display
The 56 displays one or two primary measurement values, up to six secondary measurement values, fault and warning banner, alarm relay flags, and a digital communications icon.
Process Measurements:
Two process variables are displayed if two signal boards are installed. One process variable
and process temperature is displayed if one signal board is installed with one sensor. The
Upper display area shows the Sensor 1 process reading. The Center display area shows the
Sensor 2 process reading. For dual conductivity, the Upper and Center display areas can be
assigned to different process variables as follows:
For single input configurations, the Upper display area shows the live process variable and
the Center display area can be assigned to Temperature or blank.
Secondary Values:
Up to six secondary values are shown in six display quadrants at the bottom half of the
screen. All six secondary value positions can be programmed by the user to any display parameter available. Possible secondary values include:
4.4
Menu System
The 56 menu system is similar to a computer. Pressing the ENTER/MENU key at any time
opens the top-level menu including Calibration, Hold, Programming, Display, Data and HART
32
Display and Operation
56 Analyzer Instruction Manual
Section 4: Display and Operation
LIQ_MAN_51-56
May 2014
functions. To find a menu item, use the directional Navigation keys to highlight a menu item.
Press ENTER/MENU and simply direct the cursor to the desired operation and follow the
screen prompts. Pressing the BACK screen control available on some menu screens will revert to the immediate previous menu screen. Pressing the EXIT key will return to the previous hierarchical menu level.
Fault and Warning banner:
If the analyzer detects a problem with itself or the sensor the word Fault banner (red) and/or
Warning banner (yellow) will appear at the bottom of the main display. A fault requires immediate attention. A warning indicates a problematic condition or an impending failure. For detailed troubleshooting assistance, press INFO.
4.5
USB Data Port
The 56 menu system is similar to a computer. A USB 2.0 data port is accessible on the front
panel of the 56 instrument. The USB data port can be used for download of measurement
data and events using a USB memory device. It can also be used to download and upload
complete analyzer configurations to copy all programmed settings to another 56 analyzer.
NOTE: only 56 units which display the “Transfer Configurations” tab under the Data Storage
and Retrieval menu are capable of downloading and uploading analyzer configurations.
The USB data port is easily accessed by inserting a coin in the vertical slot of the cover and
rotating counterclockwise one quarter turn to remove the cover and NEMA seal.
Caution: not all USB memory devices will physically fit into the 56 data port. After removing
the USB cover and seal, make sure that the USB memory device can be easily and fully
inserted into the USB data port without any mechanical conflict with the USB data port
flange. The USB communications port is protected by a NEMA-rated seal and cover. Do not
remove the cover during cleaning of the analyzer housing. Never remove the USB port cover
when the instrument is operated in a hazardous rated area.
NOTE: the data logger and event logger are disabled by default setting upon initial startup
from the factory. Always enable the data logger and event logger under the “Data Storage
and Retrieval” menu to initiate internal recording of process data and event data.
4.6
56 Data Logger and Event Logger Download
Procedure
4.6.1
Description
The 56 analyzer supports download of stored data logger and event logger data at the device. The download process is performed using a USB 2.0 flash drive memory device inserted
into the USB data port on the front panel of the 56 analyzer. The data can be uploaded to a
PC for viewing in preformatted EXCEL tables.
4.6.2
56 Data Logger Download Procedure
1. From the main menu, select Data storage and retrieval.
2. Select the Download tab.
3. Select “Download measurement” data.
Display and Operation
33
Section 4: Display and Operation
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
4. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting
a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to
access the data port.
5. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB insertion connector is keyed.
6. The Download screen will report Earliest data available and Latest data available that can
be downloaded from the internal data logger file.
7. With a USB 2.0 flash drive properly inserted, choice Selected Range (default) or All Data.
8. If Selected Range is chosen, define the Start Date and End Date in the screen fields. Note
that the Start Date and End Date reported will default to the current data that is recognized by the analyzer.
9. If All Data is selected, all stored date (up to 30 days) will be downloaded.
10. Select START and press ENTER.
11. A Data download information and instruction screen is displayed while data logger files
are being downloaded. The download process will be completed in a few minutes.
12. When the data logger download is complete, a confirmation screen is display with the
reported range of dates that were downloaded to the USB flash drive.
13. Carefully remove the USB 2.0 flash drive from the USB port.
14. Insert the NEMA cap into the 56 USB front display opening.
15. You may exit the Download completed screen by selecting BACK.
16. To view data logger files, insert the USB flash drive into a computer. On the designated
drive associated with the USB flash drive, individual data logger files for each day can be
opened as EXCEL formatted files in the root directory of the USB drive. Note that the
data codes are assigned file names based on the analyzer’s recognized dates.
4.6.3
56 Event Logger Download Procedure
1. From the main menu, select Data storage and retrieval.
2. Select the Download tab.
3. Select “Download events”.
4. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting
a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to
access the data port.
5. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB insertion connector is keyed.
6. A Download events information screen is briefly displayed while events files are being
downloaded. The download process may require a few minutes to complete.
7. When the events download is complete, a confirmation screen appears.
8. Carefully remove the USB 2.0 flash drive from the USB port.
9. Insert the NEMA cap into the 56 USB front display opening.
10. You may exit the Download completed screen by selecting BACK.
11. To view event logger files, insert the USB flash drive into a computer. On the designated
drive associated with the USB device, the single event logger file can be opened as an
EXCEL formatted file in the root directory of the USB drive. Note that the data codes are
assigned file names based on the analyzer’s recognized dates.
34
Display and Operation
56 Analyzer Instruction Manual
Section 4: Display and Operation
LIQ_MAN_51-56
May 2014
4.7
Software Upgrade
4.7.1
Description
All 56 advanced analyzers with serial number J12- or later (October 2012) allow software upgrades using the device’s USB data port. To download and install software upgrades, access
Rosemount Analytical’s Software Downloads site or click HERE.
4.7.2
Software upgrade checklist
Before upgrading, please confirm:
• Serial Number date codes of the 56 analyzer is J12 (October 2012) or later.
• Confirm that Safe Area requirements are met before installing software
• USB 2.0 flash drive memory device is available
4.7.3
Software upgrade procedure
1. Download the 56 software upgrade file to a computer. Access software HERE. Save the
file to the main root directory of a USB flash drive memory stick. Only one 56 software
upgrade file should be present at the root directory.
2. Save all user settings before upgrading the software by copying the analyzer’s configuration to a flash drive. Access this procedure at the local 56 device menu location:
MENU/Data storage and retrieval/Transfer Configuration.
3. Download the Data Logger measurement data and the Events before upgrading the software. Access these procedures at 56 menu location: MENU/Data storage and
retrieval/Download/Download measurement data and at /Download/Download Events.
4. With the 56 analyzer powered up and the front enclosure panel completely closed, remove the NEMA cap from the front display by inserting a coin into the cap’s vertical slot
and rotating counterclockwise. Remove the USB cap and seal to access the data port.
5. Carefully insert a USB 2.0 flash drive into the USB port on the 56. The USB insertion
connector is keyed. Note that some USB devices will not fit into the USB port due to
mechanical restrictions.
6. Press the ENTER/MENU key. Select Software upgrade in the main menu.
7. Select NEXT. Press ENTER/MENU to start software upgrade. The current software version and the new software version are reported on the screen.
8. Select NEXT. Press ENTER/MENU. “Software upgrade in process” and a progress bar will
appear. The process may require up to 5 minutes to complete.
9. The Time and Date screen will appear. Enter local real time clock and date information.
Press ENTER/MENU.
10. The main screen will appear reporting live process values. Programming of settings and
calibration may be required.
4.8
Configuration Transfer
4.8.1
Description
All 56 unit with serial number J12- or later (October 2012) will support the function of Configuration Transfer from one 56 analyzer to another, or multiple 56 analyzers. The transfer is
Display and Operation
35
Section 4: Display and Operation
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
done by using a version 2.0 USB flash drive (USB memory stick) for downloading the existing
configuration and uploading to another instrument.
4.8.2
Transfer Configuration Process User Notes
1. Configuration Transfer can only be performed between instruments of identical configuration. The analyzers must have the same signal boards installed.
2. Several text files downloaded to the flash drive can be saved to a PC and used later.
3. Only one set of configuration files will be stored on the flash drive. The files that are
downloaded will over-write any existing files stored in the root directory of the flash
drive.
4.8.3
Transfer Configuration Procedure - User Notes
1. Confirm all user settings before transferring configuration.
2. From the main menu, select Data storage and retrieval.
3. Select the Transfer Configuration tab
4. Select “Copy analyzer configuration to the flash drive”.
5. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting
a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to
access the data port.
6. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB insertion connector is keyed.
7. Select Copy data. Press ENTER.
8. An information screen appears warning users that any configuration files that exist in the
root directly of the flash drive will be over-written upon configuration transfer. If No is
selected (default setting), existing configuration files will not be overwritten. Select Yes
to transfer the configuration file to the flash drive.
9. The configuration file will be transferred (downloaded) in about 20 seconds. A screen
will report that the configuration file has been transfused.
10. Carefully remove the USB 2.0 flash drive from the USB port.
11. Insert the NEMA cap into the 56 USB front display opening.
12. Follow steps 2-3 above on the instrument that will receive the copied configuration.
13. On the 56 analyzer instrument that will receive the copied configuration, remove the
NEMA cap from the front display.
14. Carefully insert the USB 2.0 flash drive containing the configuration file into the USB
port. .
15. Select “Copy configuration from the flash drive to the analyzer”. Press ENTER.
16. You have two options for configuration transfer. Select one of the following and press
ENTER:
a.
Copy configuration data only to the analyzer.
b. Copy configuration and calibration data to the analyzer.
17. Select Copy data. Press ENTER.
18. The configuration file will be transferred (uploaded) in about 20 seconds. A screen will
report that the configuration file has been transfused.
19. Carefully remove the USB 2.0 flash drive from the USB port.
20. Insert the NEMA cap into the 56 USB front display opening.
36
Display and Operation
56 Analyzer Instruction Manual
Section 5: Programming Basics
LIQ_MAN_51-56
May 2014
Section 5.0. Programming the Analyzer - Basics
5.1
General
Typical programming steps include the following listed procedures. Each of these programming
functions are easily and quickly accomplished using the intuitive menu systems.
• Changing the measurement type, measurement units and temperature units.
• Choose temperature units and manual or automatic temperature compensation mode
• Configure and assign values to the current outputs
• Set a security code for two levels of security access
• Accessing menu functions using a security code
• Enabling and disabling Hold mode for current outputs
• Choosing the frequency of the AC power (needed for optimum noise rejection)
• Resetting all factory defaults, calibration data only, or current output settings only
5.2
Changing Startup Settings
To change the measurement type, measurement units, or temperature units that were initially
entered in Quick Start, choose the Reset function or access the Program menus for sensor 1 or
sensor 2. The following choices for specific measurement type, measurement units are available for each sensor measurement board.
TABLE 5-1. Measurements and Measurement Units
Signal board
Available measurements
Measurements units:
pH/ORP (-22, -32)
pH, ORP, Redox, Ammonia, Fluoride,
Custom ISE
pH, mV (ORP)
%, ppm, mg/L, ppb, µg/L, (ISE)
Contacting conductivity
(-20, -30)
Toroidal conductivity
(-21, -31)
Conductivity, Resistivity, TDS, Salinity,
NaOH (0-12%), HCl (0-15%), Low H2SO4,
High H2SO4, NaCl (0-20%),
Custom Curve
Conductivity, Resistivity, TDS, Salinity,
NaOH (0-12%), HCl (0-15%), Low H2SO4,
High H2SO4, NaCl (0-20%),
Custom Curve
µS/cm, mS/cm, S/cm
% (concentration)
µS/cm, mS/cm, S/cm
% (concentration)
Chlorine
(-24, -34)
Free Chlorine, pH Independ. Free Cl, Total
Chlorine, Monochloramine
ppm, mg/L
Oxygen
(-25, -35)
Oxygen (ppm), Trace Oxygen (ppb),
Percent Oxygen in gas, Salinity
ppm, mg/L, ppb, µg/L % Sat, Partial
Pressure, % Oxygen In Gas, ppm
Oxygen In Gas
Ozone (-26, -36)
Temperature (all)
Ozone
Programming Basics
Temperature
ppm, mg/L, ppb, µg/L
°C. ºF
37
Section 5: Programming Basics
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
To change the measurement type,
measurement units, or temperature
units, access the Reset screens by pressing ENTER/MENU from the main screen.
To change the measurement type,
measurement units, or temperature
units, access the Program screens by
pressing ENTER/MENU from the main
screen.
5.3
Programming Temperature
Most liquid analytical measurements (except ORP) require temperature compensation. The
56 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature
correction is off, the 56 uses the temperature entered by the user in all temperature correction calculations.
To select automatic or manual temp
compensation, set the manual reference
temperature, and to program temperature units as °C or °F, access the Temperature screens by pressing ENTER/MENU
from the main screen.
5.4
Configuring and Ranging the Current Outputs
The 56 accepts inputs from two sensors
and has four analog current outputs. Ranging the outputs means assigning values to
the low (0 or 4 mA) and high (20 mA) outputs. This section provides a guide for configuring and ranging the outputs. ALWAYS
CONFIGURE THE OUTPUTS FIRST.
To configure the outputs, access the Outputs screen by pressing ENTER/MENU from
the main screen.
38
Programming Basics
56 Analyzer Instruction Manual
Section 5: Programming Basics
LIQ_MAN_51-56
5.5
May 2014
Setting a Security Code
The security codes prevent accidental or unwanted changes to program settings, displays,
and calibration. The 56 has two levels of security code to control access and use of the
instrument to different types of users.
The two levels of security are:
• All: This is the Supervisory security
level. It allows access to all menu functions, including Programming, Calibration, Hold and Display.
• Calibration/Hold: This is the operator
or technician level menu. It allows access to only calibration and Hold of the
current outputs.
To set security codes, access the Security screen by pressing ENTER/MENU from the main
screen. Upon entry of the proper code, the following security screen will appear.
5.6
Security Access
When entering the correct access code for the Calibration/Hold security level, the Calibration and Hold menus are accessible. This allows operators or technicians to perform routine
maintenance. This security level does not allow access to the Program or Display menus.
When entering the correct access code for All security level, the user has access to all menu
functions, including Programming, Calibration, Hold and Display.
The 56 menus use a security code, access
the Security screen by pressing
ENTER/MENU from the main screen.
If a security code is currently programmed, the follow security screen will
appear. Enter the code.
1. If a security code has been
programmed, selecting the
Calibrate, Hold, Program or
Display top menu items causes
the security access screen to appear.
2. Enter the three-digit security code for the appropriate security level.
3. If the entry is correct, the appropriate menu screen appears. If the entry is incorrect,
the Invalid Code screen appears. The Enter Security Code screen reappears after 2
seconds.
Programming Basics
39
Section 5: Programming Basics
56 Analyzer Instruction Manual
May 2014
5.7
LIQ_MAN_51-56
Using Hold
The analyzer output is always proportional to measured value. To prevent improper operation
of systems or pumps that are controlled directly by the current output, place the analyzer in
hold before removing the sensor for
calibration and maintenance. Be sure to
remove the analyzer from hold once
calibration is complete. During hold, all
outputs remain at the last value. Once in
hold, all current outputs remain on
Hold indefinitely.
To hold the outputs and alarm relays,
access the Hold screen by pressing
ENTER/MENU from the main screen.
5.8
Resetting Factory Default Settings
This section describes how to restore factory calibration and default values. The process also
clears all fault messages and returns the display to the first Quick Start screen. The 56
offers three options for resetting factory
defaults.
a. reset all settings to factory defaults
b. reset sensor calibration data only
c. reset analog output settings only
To reset to factory defaults, reset
calibration data only or reset analog
outputs only, access the Reset screen
by pressing ENTER/MENU from the
main screen.
5.9
Programming Alarm Relays
The 56 24VDC (-02 order code) and the
AC switching power supply (-03 order
code) provide four alarm relays for process
measurement or temperature. Each alarm
can be configured as a fault alarm instead
of a process alarm. Also, each relay can be
programmed independently and each can
be programmed as an interval timer or
one of four advanced timer functions. This
section describes how to configure alarm
relays, simulate relay activation, and
synchronize timers for the four alarm relays.
This section provides details to program the following alarm features. To program the alarm
relays, access the Program screen by pressing ENTER/MENU from the main screen and then
select the Relay tab and the Configure relay control.
40
Programming Basics
56 Analyzer Instruction Manual
Section 5: Programming Basics
LIQ_MAN_51-56
May 2014
The following relay functions can be programmed to any relay from the Configure Relay
screen:
1. assign a relay
2. define a relay function
3. assign a Measurement
4. set relay logic
5. enter setpoints
6. set deadband
7. set normal state
8. set USP Safety level (contacting conductivity)
To program these relay functions, access the Configure Relay screen by pressing
ENTER/MENU from the main Relay programming screen.
1. To assign a relay, highlight the desired Relay 1-4 and press ENTER/MENU.
2. To define a relay function, select
from Setpoint, Interval Timer, TPC,
Bleed and Feed, Water Meter, Delay
timer, Date and Time, Fault or None
and press ENTER/MENU.
3. To assign a measurement to a
specific relay, select the desired
measurement or temperature input
and press ENTER/MENU.
4. To set relay logic to activate alarms
at a High reading or a Low reading, select high
or low and press ENTER/MENU.
5. To enter setpoints for relays, enter the desired value for the process measurement
or temperature at which to activate an alarm event and press ENTER/MENU.
6. To set deadband as a measurement value, enter the change in the process value needed
after the relay deactivates to return to normal (and thereby preventing
repeated alarm activation) and press ENTER/MENU.
7. To set the Normal alarm condition, select Open or Closed and press ENTER/MENU.
Program the normal state to define the desired alarm default state to normally
open or normally closed upon power up.
8. To set USP Safety, enter the percentage below the limit at which to activate the alarm
and press ENTER/MENU. NOTE: USP Safety only appears if a contacting
conductivity board is installed.
Programming Basics
41
Section 5: Programming Basics
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
This section provides details to simulate
relay action. To simulate relays, access
the Program screen by pressing
ENTER/MENU from the main screen and
then select the Relay tab.
To simulate alarm relay conditions,
access the Simulate Relay Action screen
by pressing ENTER/MENU from the main
Relay programming screen.
Alarm relays can be manually set for the
purposes of checking devices such as
valves or pumps. Under the Alarms
Settings menu, this screen will appear to
allow manual forced activation of the
alarm relays. Select the desired alarm
condition to simulate.
42
Programming Basics
56 Analyzer Instruction Manual
Section 6: Programming Measurements
LIQ_MAN_51-56
May 2014
Section 6.0 Programming - Measurements
6.1
Programming Measurements – Introduction
The 56 automatically recognizes each installed measurement board upon first power-up and
each time the analyzer is powered. Completion of Quick Start screens upon first power
up enable measurements, but additional steps may be required to program the analyzer for
the desired measurement application. This section covers the following programming and
configuration functions:
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pH-see Sec. 6.2)
3. Enabling manual temperature correction and entering a reference temperature
(all sections)
4. Enabling sample temperature correction and entering temperature correction
slope (selected sections)
5. Defining measurement display resolution (pH and amperometric)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see Sec’s 6.4, 6.5)
9. Entering a cell constant for a contacting or toroidal sensor (see Sec’s 6.4, 6.5)
10. Entering a temperature element/RTD offset or temperature slope (conductivitysee Sec’s 6.4)
11. Creating an application-specific concentration curve (conductivity-see Sec’s 6.4, 6.5)
12. Enabling automatic pH correction for free chlorine measurement (Sec. 6.6.1)
To fully configure the analyzer for each installed measurement board, you may use the following:
1. Reset Analyzer function to reset factory defaults and configure the measurement board
to the desired measurement. Follow the Reset Analyzer menu (Fig. 5-5) to
reconfigure the analyzer to display new measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following configuration and programming guidelines for the applicable measurement.
6.2
pH Measurement Programming
The section describes how to configure the 56 analyzer for pH measurements. The following programming and configuration functions are covered.
1. Measurement type: pH Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
2. Preamp location: Analyzer Identify preamp location
3. Filter: 4 sec Override the default input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high reference impedance
5. Sensor wiring scheme: Normal or Reference to Ground
6. Resolution: 0.01pH Select 0.01pH or 0.1pH for pH display resolution
7. Enabling pH sensor diagnostics
Programming Measurements
43
Section 6: Programming Measurements
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
To configure the pH measurement board, access the Program screen by pressing
ENTER/MENU from the main screen and then select the Measurement tab.
1. To Select a Measurement type, select
from: pH, ORP, Redox, Ammonia,
Fluoride, and Custom ISE and press
ENTER/MENU.
2. To program the Preamp location, select Analyzer or Sensor/JBox, and
press ENTER/MENU.
3. To Override the default input
filter, enter 0-999 seconds and press
ENTER/MENU.
4. To program Reference Impedance. Select Low or High and press ENTER/MENU.
5. To choose the wiring scheme, Select Normal or Reference to Ground and press
ENTER/MENU.
6. To program the display resolution, Select 0.01 pH or 0.1 pH and press ENTER/MENU.
7. To enable pH sensor diagnostics, select the “pH diagnostic setup” tab under Programming. Select Sensor 1 or Sensor 2. Select NEXT. Select On under sensor diagnostics to
enable pH diagnostics.
NOTE: pH sensor diagnostics must be enabled to include diagnostic values such as Glass Impedance and Reference Impedance in EXCEL data log sheets after data download to USB.
Enabling pH sensor diagnostics also allows assignment of Glass Impedance and Reference
Impedance to the two-dimensional on-screen process graph accessible under “Display
Setup/View Graph”.
6.3
ORP Measurement Programming
The section describes how to configure the 56 analyzer for ORP measurements.
The following programming and configuration functions are covered:
1. Measurement type: pH Select pH,
ORP, Redox, Ammonia, Fluoride,
Custom ISE
2. Preamp location: Analyzer Identify
preamp location
3. Filter: 4 sec Override the default
input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high
reference impedance
5. Sensor wiring scheme: Normal or
Reference to Ground
To configure the ORP measurement board, access the Program screen by pressing
ENTER/MENU from the main screen and then select the Measurement tab.
1. To Select a Measurement type, select ORP and press ENTER/MENU.
2. To program the Preamp
location, select Analyzer or
Sensor/JBox, and press ENTER/MENU.
3. To Override the default input
filter, enter 0-999 seconds and press ENTER/MENU.
44
Programming Measurements
56 Analyzer Instruction Manual
Section 6: Programming Measurements
LIQ_MAN_51-56
May 2014
4. To program Reference Impedance. Select Low or High and press ENTER/MENU.
5. To choose the wiring scheme, Select Normal or Reference to Ground and press
ENTER/MENU.
6.4
Contacting Conductivity
Measurement Programming
The section describes how to configure the 56 analyzer for conductivity measurements using
contacting conductivity sensors. The following programming and configuration functions are
covered.
1. Measure: Conductivity, Select Conductivity, Resistivity, TDS. Salinity or % conc
2. Type: 2-Electrode Select 2-Electrode or 4-Electrode type sensors
3. Cell K: 1.00000/cm Enter the cell Constant for the sensor
4. Measurement units
5. Filter: 2 sec Override the default input filter, enter 0-999 seconds
6. Range: Auto Select measurement Auto-range or specific range
7. Temp Comp: Slope Select Temp
Comp: Slope, Neutral Salt, Cation or
Raw
8. Slope: 2.00%/°C Enter the linear temperature coefficient
9. Ref Temp: 25.0°C Enter the Reference temp
10. Cal Factor: default=0.95000/cm
Enter the Cal Factor for 4-Electrode
sensors from the sensor tag
To configure the Contacting conductivity measurement board, access the Program screen by
pressing ENTER/MENU from the main screen and then select the Measurement tab.
1. To program a Measurement type, select Conductivity Select Conductivity, Resistivity,
TDS. Salinity or % conc. and press ENTER/MENU.
2. To program a sensor type, select 2Electrode Select 2-Electrode or 4Electrode type sensors and press
ENTER/MENU.
3. To program the Cell constant, enter
the exact cell constant value expressed as 1.XXXXX/cm the for the
sensor and press ENTER/MENU.
4. To program Measurement units, select uS/cm or mS/cm and press
ENTER/MENU.
5. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
6. To program the measurement range, select a specific range appropriate for your process
and press ENTER/MENU.
7. To program the Temp Comp method, choose Slope, Neutral Salt, Cation or Raw
and press ENTER/MENU.
Programming Measurements
45
Section 6: Programming Measurements
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
8. To change the Temperature compensation Slope, enter the linear temperature
coefficient expressed as X.XX%/°C and press ENTER/MENU.
9. To program the Reference Temperature for Manual temperature compensation (not
from probe RTD), enter the Reference temp expressed as XX.X°C and press
ENTER/MENU.
10. To program the Cal Factor for 4-Electrode sensors, enter the value shown on the sensor
tag, expressed as X.XXXXX/cm and press ENTER/MENU.,
6.5
Toroidal Conductivity
Measurement Programming
The section describes how to configure the 56 analyzer for conductivity measurements using inductive/toroidal sensors. The following programming and configuration functions are covered:
1. Measure: Conductivity Select Conductivity, Resistivity, TDS. Salinity or % conc
2. Sensor model: 228 Select sensor type
3. Measurement units
4. Cell K: 3.00000/cm Enter the cell Constant for the sensor
5. Temp Comp: Slope Select Temp Comp: Slope, Neutral Salt, Cation or Raw
6. Slope: 2.00%/°C Enter the linear temperature coefficient
7. Ref Temp: 25.0°C Enter the Reference temp
8. Filter: 2 sec Override the default
input filter, enter 0-999 seconds
9. Range: Auto Select measurement
Auto-range or specific range
To configure the Contacting conductivity
measurement board, access the Program
screen by pressing ENTER/MENU from
the main screen and then select the
Measurement tab.
1. To program a Measurement type,
select Conductivity Select
Conductivity, Resistivity, TDS.
Salinity or % conc. and press
ENTER/MENU.
2. To program the sensor model, select
228 or other toroidal model number
and press ENTER/MENU.
3. To program Measurement units,
select uS/cm or mS/cm and press
ENTER/MENU.
4. To program the Cell constant, enter the exact cell constant value expressed as
3.XXXXX/cm the for the sensor and press ENTER/MENU.
5. To program the Temp Comp method, choose Slope, Neutral Salt, Cation or Raw
and press ENTER/MENU.
6. To change the Temperature compensation Slope, enter the linear temperature
coefficient expressed as X.XX%/°C and press ENTER/MENU.
46
Programming Measurements
56 Analyzer Instruction Manual
Section 6: Programming Measurements
LIQ_MAN_51-56
May 2014
7. To program the Reference Temperature for Manual temperature compensation
(not from probe RTD), enter the Reference temp expressed as XX.X°C and press
ENTER/MENU.
8. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
9. To program the measurement range, select a specific range appropriate for your process
and press ENTER/MENU.
6.6
Chlorine Measurement Programming
With a Chlorine measurement board installed, The 56 can measure any of four variants
of Chlorine:
•
Free Chlorine
•
Total Chlorine
•
Monochloramine
•
pH-independent Free Chlorine
The section describes how to configure the 56 analyzer for Chlorine measurements.
6.6.1
Free chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Free Chlorine measurement using amperometric chlorine sensors. The following programming and configuration
functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl,
Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display resolution 0.01 or 0.001
4. Free Cl Correct: Live Select Live/Continuous pH correction or Manual
5. Manual pH: 7.00 pH For Manual pH
correction, enter the pH value
6. Filter: 5 sec Override the default input filter, enter 0-999 seconds
7. Dual Slope Calibration: Enable or Disable
1. To program the Measurement type, select Free Chlorine,
pH Ind Free Cl., Total Cl, or Monochloramine and press ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press
ENTER/MENU.
3. To program the Measurement Resolution: Select 0.01 or 0.001 and press
ENTER/MENU.
4. To program Free Cl Correction, select Live/Continuous pH correction
or Manual and press ENTER/MENU
Programming Measurements
47
Section 6: Programming Measurements
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
5. To program for Manual pH correction, enter the pH value and press ENTER/MENU.
6. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
7. To use Dual Slope Calibration, select Enable or Disable and press ENTER/MENU.
6.6.2
Total chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Total Chlorine measurement using amperometric chlorine sensors. The following programming and configuration
functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl,
Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5 sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable or Disable
1. To program the Measurement type, select Free Chlorine, pH Ind Free Cl., Total Cl, or
Monochloramine and press
ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press
ENTER/MENU.
3. To program the Measurement Resolution: Select 0.01 or 0.001 and press
ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press
ENTER/MENU.
5. To use Dual Slope Calibration, select Enable or Disable and press ENTER/MENU.
6.6.3
Monochloramine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Monochloramine
measurement using amperometric chlorine sensors. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl,
Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5 sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable or
Disable
48
Programming Measurements
56 Analyzer Instruction Manual
Section 6: Programming Measurements
LIQ_MAN_51-56
May 2014
1. To program the Measurement type, select Free Chlorine, pH Ind Free Cl., Total Cl, or
Monochloramine and press ENTER/MENU.
2. To program the Measurement Units: select ppm mg/L and press ENTER/MENU.
3. To program the Measurement Resolution: Select 0.01 or 0.001 and press
ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press
ENTER/MENU.
5. To use Dual Slope Calibration, select
Enable or Disable and press
ENTER/MENU.
6.6.4
pH-independent free chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Free Chlorine measurements using the pH-independent free chlorine sensor, 498CL-01, manufactured by Rosemount Analytical. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl,
Monochloramine
2. Units: ppm Select units ppm
or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable
or Disable
1. To program the Measurement type, select Free Chlorine,
pH Ind Free Cl., Total Cl, or Monochloramine and press ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press
ENTER/MENU.
3. To program the Measurement
Resolution: Select 0.01 or 0.001 and
press ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press
ENTER/MENU.
5. To use Dual Slope Calibration, select
Enable or Disable and press ENTER/MENU.
6.7
Oxygen Measurement Programming
This section describes how to configure the 56 analyzer for dissolved and gaseous oxygen
measurement using amperometric oxygen sensors. The following programming and configuration functions are covered:
Programming Measurements
49
Section 6: Programming Measurements
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
1. Sensor type: Select Water/Waste, Trace. BioRx, BioRx-Other, Brew, %O2 In Gas
2. Measure type: Select Concentration, % Saturation, Partial Pressure, Oxygen in Gas
3. Units: ppm Select ppm, mg/L, ppb, µg/L, % Sat, %O2-Gas, ppm Oxygen-Gas
4. Pressure Units: bar Select
pressure units: mm Hg, in Hg,. Atm,
kPa, mbar, bar
5. Use Press: At Air Cal Select atmospheric pressure source – internal or
mA Input
6. Salinity: 00.0‰ Enter Salinity as ‰
7. Filter: 5sec Override the default
input filter, enter 0-999 seconds
8. Partial Press: mmHg Select mm Hg,
in Hg. atm, kPa, mbar or bar for Partial pressure
1. To program Sensor type, Select Water/Waste, Trace. BioRx, BioRx-Other, Brew,
or %O2 In Gas and press ENTER/MENU.
2. To program Measure type: Select
Concentration, % Saturation, Partial
Pressure, or Oxygen in Gas and press
ENTER/MENU.
3. To program measurement Units,
Select ppm, mg/L, ppb, g/L, % Sat,
%O2-Gas, or ppm Oxygen-Gas and
press ENTER/MENU.
4. To program Pressure Units: Select
pressure units: mm Hg, in Hg,. Atm,
kPa, mbar, or bar and press
ENTER/MENU.
5. To program which atmospheric Pressure source to use during Air Cal, Select internal or
mA Input and press ENTER/MENU.
6. To program Salinity, Enter the known Salinity value as ‰ and press ENTER/MENU.
7. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
8. To program Partial Press: Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial
pressure and press ENTER/MENU.
6.8
Ozone Measurement Programming
This section describes how to configure the 56 analyzer for ozone measurement using amperometric ozone sensors. The following programming and configuration functions are covered:
1. Units: ppm Select ppm, mg/L, ppb, µg/L
2. Resolution: 0.001 Select display resolution 0.01 or 0.001
3. Filter: 5sec Override the default input filter, enter 0-999 seconds
50
Programming Measurements
56 Analyzer Instruction Manual
Section 6: Programming Measurements
LIQ_MAN_51-56
May 2014
1. To program measurement Units, Select ppm, mg/L, ppb, µg/L, % Sat, %O2-Gas, or ppm
Oxygen-Gas and press ENTER/MENU.
2. To program the Measurement Resolution: Select 0.01 or 0.001 and press ENTER/MENU.
3. To Override the default input
filter, enter 0-999 seconds and press ENTER/MENU.
6.9
Turbidity Measurement Programming
This section describes how to configure the 56 analyzer for Turbidity measurements. The
following programming and configuration functions are covered.
1. Measurement type: Turbidity Select
Turbidity or TSS calculation
(estimated TSS)
2. Sensor type: Select EPA or ISO
3. Measurement units: NTU, FTU, FNU
4. Filter: 20 sec Override the default
input filter, enter 0-999 seconds
5. Bubble Rejection: On Intelligent
software algorithm to eliminate
erroneous readings caused by
bubble accumulation in the sample
1. To program the Measurement type,
Select Turbidity or TSS
calculation (estimated TSS)
and press ENTER/MENU.
2. To program the Sensor type:
Select EPA or ISO and press
ENTER/MENU.
3. To program Measurement units:
NTU, FTU, FNU and press
ENTER/MENU.
4. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
5. To program Bubble Rejection, select On or Off and press ENTER/MENU.
Programming Measurements
51
Section 6: Programming Measurements
56 Analyzer Instruction Manual
May 2014
6.10
LIQ_MAN_51-56
Flow Measurement Programming
This section describes how to configure the 56 analyzer for flow measurement when used
with a compatible pulse flow sensor. The following programming and configuration
functions are covered:
To program pulse flow, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each
programming routine.
1. Measurement type Pulse
Flow Select Pulse Flow or mA
Current Input
2. Measurement units: GPH Select
GPM, GPH, cu ft/min, cu ft/hour,
LPM, L/hour, m3/hr. or
3. Filter: 5sec Override the default
input filter, enter 0-999 seconds
1. To program Measurement type, Select Pulse Flow or mA Current Input and press
ENTER/MENU.
2. To program Measurement units: Select GPM, GPH, cu ft/min, cu ft/hour, LPM, L/hour, or
m3/hr. and press ENTER/MENU.
3. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
6.11
Current Input Programming
This section describes how to configure the 56 analyzer for current input measurement when
wired to an external device that transmits 4-20mA or 0-20mA analog current output. The following programming and configuration functions are covered.
1. Measurement type mA input Override the default (Flow) and select mA current input
2. mA Input Temperature Select Temperature, Pressure, Flow or Other
3. Measurement units: degC Select measurement units based on selected input device type
4. Input Range: 4-20mA Select 4-20mA
or 0-20mA
5. Low Value: 0.000oC Enter
the low measurement value to assign
to 4mA
6. High Value: 100.0oC Enter
the high measurement value to assign to 20mA
7. Filter: 05 sec Override the
default input filter, enter 0-999
seconds
52
Programming Measurements
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 6: Programming Measurements
May 2014
1. To Override the default the
Measurement type (Flow)
select mA current input and press
ENTER/MENU.
2. To program the mA Input type,
Select Temperature, Pressure, Flow
or Other and press ENTER/MENU.
3. To program Measurement units,
Select measurement units based on
selected input device type and press
ENTER/MENU.
4. To program the Input Range: Select 4-20mA or 0-20mA and press ENTER/MENU.
5. To program the Low input Value, enter the low measurement value to assign to 4mA
and press ENTER/MENU.
6. To program the High input Value, enter the high measurement value to assign to 20mA
and press ENTER/MENU.
7. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
Programming Measurements
53
Section 6: Programming Measurements
May 2014
LIQ_MAN_51-56
This page left blank intentionally
54
56 Analyzer Instruction Manual
56 Analyzer Instruction Manual
Section 7: PID Control
LIQ_MAN_51-56
May 2014
Section 7.0 PID Control
7.1
Introduction
7.1.1
Measurement and Set Point (Feedback Control)
The 56 controller is given two items of information: measurement and set point. The controller reacts to the difference in value of these two signals and produces an analog output
signal to eliminate that difference. As long as the difference exists, the controller will try to
eliminate it with the output signal. When measurement and set point are equal, the condition of the controller is static and its output is unchanged. Any deviation of measurement
from set point will cause the controller to react by changing its output signal.
7.1.2
Proportional Mode
The simplest control is proportional control. In this control function, the error from set point,
divided by the control range, is multiplied by the Gain constant to produce the output.
The control range is the percent of the analog output span (the difference between the 4 (or
0) mA and 20 mA settings) through which the measured variable must move to change the
output from minimum to maximum.
The smaller the Gain, the less the controller reacts to changes in the measured variable. The
larger the Gain, the more the controller reacts to changes in the measured variable.
The proportional control output is given by the expression below. As can be seen, the overall
gain is determined by the control range chosen (URV and LRV) and the Gain:
Proportional Output (%) = Gain * (PV – SP) * 100 / (URV, upper range value and LRV, lower
range value).
7.1.2.1
Direct Acting Control Action
Direct acting control action increases the control output as the measured variable increases
above the setpoint. The LRV is usually set to the setpoint value, so that the control output is
0% at the setpoint, and the URV is greater than the setpoint so that the 100% control output
is at a higher measurement value. The Gain parameter can then be adjusted to produce the
desired gain.
Fig. 7-1 Direct Acting Control
PID Control
55
Section 7: PID Control
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Example of direct acting control: Lower the pH of a solution at 10 pH by adding acid to
control it at 8 pH with the Gain parameter assumed to be 1.0. The higher the measured pH,
the more acid is required to lower the pH toward the setpoint, but as the pH approaches the
setpoint less acid is required:
Fig. 7-2 Example of Direct Acting Control
7.1.2.2
Reverse Acting Control Action
Reverse acting control action, decreases the control output as the measured variable increases toward the setpoint. The LRV is usually set to the setpoint value, so that the control
output is 0% at the setpoint, and the URV is less than the setpoint value so that the 100% control output is at a lower measurement value. The Gain can then be adjusted to produce the
desired gain.
Fig. 7-3 Reverse Acting Control
Example of reverse acting control: Add base to a solution at 8.0 pH, to control the pH to
10.0 pH with an assumed Gain parameter of 1.0. The lower the measured pH, the more
base is required to raise the pH toward the setpoint, but as the pH approaches the setpoint
less base is required:
56
PID Control
56 Analyzer Instruction Manual
Section 7: PID Control
LIQ_MAN_51-56
May 2014
Fig. 7-4 Example of Reverse Acting Control
7.1.3
Proportional Bias
Most processes require that the measured variable be held at the set point. The proportional
mode alone will not automatically do this, if an output greater than 0% is needed to keep the
PV at setpoint. At setpoint, the control output is 0%, and if a non-zero control output is
needed to keep the PV at the setpoint, proportional alone will only stabilize the measured
variable at some offset (deviation) from the desired setpoint.
Bias is used to provide a constant control output at the setpoint to maintain PV at the setpoint. The effect of Bias is expressed as follows:
Proportional Output (%) = [Gain * (PV – SP) * 100 / (URV – LRV)] + BIAS
Fig. 7-5 Direct Acting Control with Bias
PID Control
57
Section 7: PID Control
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Fig. 7-5 Reverse Acting Control with Bias
7.1.4
Proportional Plus Integral (Reset)
For the automatic elimination of deviation, Integral mode, also referred to as Reset, is used.
The proportional function is modified by the addition of automatic reset, rather than a constant Bias value. With the reset mode, the controller continues to change its output until the
deviation between measurement and set point is eliminated.
The action of the reset mode depends on the overall gain. The rate at which it changes the
controller output is based on the overall gain band size and the reset time (I). The reset time
is the time required for the reset mode to repeat the proportional action once. It is expressed
as seconds per repeat, adjustable from 0-3,000 seconds.
The reset mode repeats the proportional action as long as an offset from the set point exists.
Reset action is cumulative. The longer the offset exists, the more the output signal is increased. If the PV overshoots the setpoint, the reset action will decrease. When the measurement reaches the setpoint and the proportional control action becomes zero, there will
be an accumulated integral control action to keep the process at the setpoint.
The controller configured with reset continues to change until there is no offset. If the offset
persists, the reset action eventually drives the controller output to its 100% limit - a condition
known as "reset windup".
Once the controller is "wound up", the deviation must be eliminated or redirected before the
controller can unwind and resume control of the measured variable. The integral time can be
cleared and the "windup" condition quickly eliminated by manually overriding the 56 analog
output using the manual mode to reduce the control output and then setting the reset time
to 0 seconds to make integral control action 0%. The reset time can then be changed to
avoid reset windup.
The proportional plus integral control output is given below. Note that the larger the reset
time (I), the slower the integral response will be:
% Output = [Gain × 100 / (URV – LRV)] × [(PV – SP) + 1/I Σ (PVt – SP) Δt]
58
PID Control
56 Analyzer Instruction Manual
Section 7: PID Control
LIQ_MAN_51-56
7.1.5
May 2014
Derivative Mode (Rate)
Derivative mode provides a 3rd control mode, which responds to the rate of change of the
Proportional control output, multiplied by the Derivative parameter D which has units of
seconds. The contribution of the derivative response is given below:
% Output = [Gain × 100 / (URV – LRV)] × D × [(PVt – SP) − (PVt-1 – SP)] / Δt
The purpose of derivative action is to provide a quick control response to changes in the
measured parameter. In general, it is not often used in concentration control, and in fact, it
has been estimated that 90 to 95% of all control applications use only Proportional plus
Integral control. Any noise in the measurement causes problems with derivative action.
Temperature measurements tend to be less noisy than other measurements, and derivative
action is most often used for temperature control.
7.1.6
Process Characterization and Tuning
Control loops are tuned by the choice of the control range and the selection of the control
parameters. How these parameters are chosen should depend on how the process responds
to controller output. The process response is characterized by certain behaviors, which are
due to such factors as mixing and reaction time, response time of the process to control
output changes, and the characteristics of the final operator, i.e. control values, pumps,
heaters, etc. With these characteristics known, initial control settings can be developed.
A good reference to PID control is provided by the book, “Control Loop Foundation—Batch
and Continuous Processes”, by Terrence Blevins and Mark Nixon, International Society of
Automation, Research Triangle Park, NC, © 2011.
A guide to tuning control is provided by the book, “Good Tuning: A Pocket Guide”, by
Gregory K. McMillan, International Society of Automation, Research Triangle Park, NC, ©
2005.
7.2
PID Setup
7.2.1
PID Control
The 56 current ¬outputs (one or all four) can be programmed for PID control. PID control
can be applied to any of the measurements provided by the sensor boards, such as pH, conductivity, and concentrations. In addition, PID control can be applied to temperature and
any measurement input to the 56 using the flow/4-20mA board.
The output signal of PID control is used with a final control element, which can vary is output
from 0 to 100% in response to the control signal. Final control elements can include control
valves, pumps or heaters.
7.2.2
Selecting PID Control
Select PID control, the analog output to be used, and the measurement and range from the
main analog output setup window:
PID Control
59
Section 7: PID Control
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Basic Definitions
• Output – Select the analog output (1 through 4) to be configured for PID control.
• Analog/PID/Simulate – Choose PID
• Assign – Select the Measurement to be controlled. Note: This measurement can also
be a 4-20 mA signal input brought in by the flow/ 4-20 mA board.
• Range – Select either 0-20mA or 4-20mA range, depending on the signal range used by
the final control element, e.g. a pump or valve.
• Select Next to go to the PID Setup parameters.
7.2.3
PID Setup Parameters
The PID control setup window contains the PID control tuning parameters.
Also note that the upper portion of the screen shows the measurement chosen for control
(PV), and the control output in mA and % Output. This makes it possible to observe the primary variable (PV) and the control output, in terms of percent and milliamps, while tuning
PID control.
PID Control Parameters: Basic Definitions
• Setpoint – Select the desired setpoint.
• URV – The value of PV (in the above example, 14.00 pH) at which the control will be 20
mA or 100% output.
• LRV -- The value of PV (in the above example 0.00 pH) at which the control will be 0 or 4
mA or 0% output.
60
PID Control
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 7: PID Control
May 2014
Note: If you want the control output to increase as PV (in this case pH) increases, URV should be
greater than LRV. This is direct acting control action. Examples of direct acting control are the
addition of acid to decrease pH and adding water to a solution to decrease the concentration.
If you want the control output to decrease as PV increases, i.e. reverse acting control action, the
URV should be less than LRV. Examples of reverse acting control are adding caustic to increase
pH and adding a concentrated solution to water to make a solution of lower concentration.
• Bias (range: 0 to 100%; default 0%) – Bias is a fixed control output which allows the control output to be greater than zero when the measurement (PV) is at setpoint. It is used
in proportional only control to prevent cycling resulting from the control output going
to zero at the setpoint.
• Gain (range: 0.0 to 1000.0; default 0.0) – In proportional (P) only control, the output is
directly proportional to the difference between the process variable (PV) and the setpoint divided by the output span (URV – LRV). Gain is a factor which multiplies the proportional output to meet the requirements of the process being controlled. Using Gain
values less than 1 reduce the proportional output while Gain values greater than 1 increase the proportional output.
• Integral (Reset, I) (range: 0 to 3,000 seconds; default 0 seconds) – Integral repeats the
proportional action in a time period given by the reset time (I). The reset time is given
in seconds per repeat and is adjustable from 0 to 3,000 seconds. Integral control acts as
an automatic bias which increases or decreases the overall control output in response to
the error (PV – SP) to keep the PV at the setpoint.
• Derivative (Rate, D) (range: 0 to 3,000 seconds; default 0 seconds) –Derivative action,
gives an immediate control output in response to changes in the proportional output
with time (derivative). The amount of increase or decrease depends on the rate of
change of the error. The rate constant (D) allows the user to adjust the amount derivative control contributes to the control signal. Smaller values reduce the effect of derivative control.
• Mode: Mode has two settings, Auto (Automatic) and Man (Manual). In the Auto mode
the control output is controlled automatically by the PID algorithm. In manual mode,
the control output can be set to a constant value; this is useful during transmitter calibration or servicing.
• Value (Manual) (range: 0 100%) – When the Manual mode is chosen; this control appears on the screen and allows you to write the constant control output value in the Manual mode.
Using the Next button lead to the final PID control window:
7.2.4
Transport Time
Transport Time makes it possible to apply PID control action to a process flowing in a pipe for
a short period of time (run time), and then hold the control output fixed to allow the treated
sample time to mix and travel to the pH or other analytical sensor (transport time). If properly tuned the PV should reach the setpoint after successive time periods.
It is best used when raw sample pH (or concentration) remains relatively steady for long periods of time, as is the case for samples flowing from a large body of water. It should not be
used where process upsets are possible because the delay in applying control will make recovery from the upset slow, and can result in overshoot after the incoming sample has returned to a normal range.
PID Control
61
Section 7: PID Control
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Transport Time Parameters: Basic Definitions
• Transport Time (On/Off) – Turns the Transport Time feature on or off.
• Transport Time (range 1 to 600 seconds) – When Transport Time is turned On; a
control appears at the right of it, which allows the value of the Transport Time to be
enter. This will be the time period that PID control output is held constant, while the
treated sample travels to the sensor.
• Run Time (range 0 to 60 seconds) – This is the time period that PID control action is
automatic. It always must be a shorter time than Transport Time.
Fault: Basic Definitions
1. Fault – When a measurement fault occurs (either sensor or transmitter) the control
output can be setup to continue providing a live control output or the output can be
set to a fixed value.
If the live reading is used during a fault condition, the control output could be based on
an erroneous measurement, which might cause problems. Using a fixed value for
control output during a fault condition can ensure that the control output goes to an
acceptable value.
2. Fault Current – If a fixed value on fault is chosen, this parameter selects the output.
The control output on fault can be set to a value to prevent a major upset or an unsafe
condition.
Note: If a fixed fault current output is chosen and PID control Mode is set to Manual, the
Manual output value will override the Fault Current value.
62
Time Proportional Control
56 Analyzer Instruction Manual
Section 8: Time Proportional Control
LIQ_MAN_51-56
May 2014
Section 8.0 Time Proportional Control
8.1
Introduction
8.1.1
Time Proportional Control
Time Proportional Control is more commonly known as Duty Cycle or Pulse Width Modulation. It applies PID control to the activation of a relay rather than using an analog output.
The TPC output is defined as the percent of time that a relay is on (% On Time), during a user
selected time period (Time Period). As the control output increases the on time increases:
Fig. 8-1 TPC Periods and On Time
The proportional, integral, and derivative are defined the same as analog PID control, but use
% On Time instead of % Output:
Proportional On Time (%) = [Gain × (PV – SP) x 100 / (URV – LRV)] + BIAS
Proportional and Integral Control:
% On Time = [Gain × 100 / (URV – LRV)] + [(PV – SP) + 1/I Σ (PVt – SP) Δt]
Derivative Mode:
% On Time = [Gain × 100 / (URV – LRV)] × D × [(PVt – SP) − (PVt-1 – SP)] / Δt
As with analog PID control, TPC can be direct acting (URV > LRV) or reverse acting (LRV >
URV). For more detail see Section X.1, PID Control Introduction.
8.2
TPC Setup
8.2.1
Selecting TPC
Select TPC control, the relay to be used, and the measurement to be controlled from the
main relay setup window:
Time Proportional Control
63
Section 8: Time Proportional Control
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Basic Definitions
• Relay – Select the relay (1 through 4) to be use for TPC control.
• Type – Choose TPC
• Assign – Select the Measurement to be controlled. Note: This measurement can also
be a 4-20 mA signal input brought in by the flow/ 4-20 mA board.
• Select Next to go to the PID Setup parameters
8.2.2
TPC Setup Parameters
The TPC setup window contains the PID control tuning parameters.
Also note that the upper portion of the screen shows the relay number and the value of the
measurement assigned to it, the % On Time for the relay, and the current state of the relay,
i.e. On or Off. This makes it possible to observe the primary variable (PV), the % On Time,
and the relay state, while tuning time proportional control.
PID Control Parameters: Basic Definitions
• Setpoint – Select the desired setpoint.
• URV – The value of PV (in the above example, 14.00 pH) at which the control will be
100% On Time.
• LRV – The value of PV (in the above example 0.00 pH) at which the control will be 0% On
Time.
Note: If you want the control output to increase as PV (in this case pH) increases, URV should be
greater than LRV. This is direct acting control action. Examples of direct acting control are the
addition of acid to decrease pH and adding water to a solution to decrease the concentration.
64
PID Control
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 8: Time Proportional Control
May 2014
If you want the control output to decrease as PV increases, i.e. reverse acting control action, the
URV should be less than LRV. Examples of reverse acting control are adding caustic to increase
pH and adding a concentrated solution to water to make a solution of lower concentration.
• Time Period (range: 10 to 3,000 seconds; default 30) -- The time period for each cycle of
TPC.
• Bias (range: 0 to 100%; default 0%) – Bias is a fixed control output which allows the % on
time to be greater than zero when the measurement (PV) is at setpoint. It is used in proportional only control to prevent cycling resulting from the % on time going to zero at the
setpoint.
• Relay default (Close, Open, None) – Select the relay action during a fault condition.
• Gain (range: 0.0 to 1000.0; default 0.0) – In proportional (P) only control, the output is
directly proportional to the difference between the process variable (PV) and the setpoint
divided by the output span (URV – LRV). Gain is a factor which multiplies the proportional
output to meet the requirements of the process being controlled. Using Gain values less
than 1 reduce the proportional output while Gain values greater than 1 increase the
proportional output.
• Integral (Reset, I) (range: 0 to 3,000 seconds; default 0 seconds) – Integral repeats the
proportional action in a time period given by the reset time (I). The reset time is given in
seconds per repeat and is adjustable from 0 to 3,000 seconds. Integral control acts as an
automatic bias which increases or decreases the overall control output in response to the
error (PV – SP) to keep the PV at the setpoint.
• Derivative (Rate, D) (range: 0 to 3,000 seconds; default 0 seconds) –Derivative action,
gives an immediate control output in response to changes in the proportional output with
time (derivative). The amount of increase or decrease depends on the rate of change of
the error. The rate constant (D) allows the user to adjust the amount derivative control
contributes to the control signal. Smaller values reduce the effect of derivative control.
• Mode: Mode has two settings, Auto (Automatic) and Man (Manual). In the Auto mode
the control output is controlled automatically by the PID algorithm. In manual mode, the
control output can be set to a constant value; this is useful during transmitter calibration
or servicing.
• % On Time (Manual) (range: 0 100%) – When the Manual mode is chosen; this control
appears on the screen and allows you to write the constant On Time value in the Manual
mode.
PID Control
65
Section 8: Time Proportional Control
May 2014
LIQ_MAN_51-56
This page left blank intentionally
66
56 Analyzer Instruction Manual
56 Analyzer Instruction Manual
Section 9: Alarm Relay Functions
LIQ_MAN_51-56
May 2014
Section 9.0 Alarm Relay Functions
9.1
General
An alarm is a relay that closes a set of contact points (a switch) inside the analyzer. In doing
so, the relay closes an electrical circuit and turns on a device wired to the contacts. The 56
Advanced Analyzer has four alarm relays and seven relay control functions. The relays are
turned on and off by the analyzer based on the control points, setpoints or control
parameters that you program into the analyzer through the keypad. See Section 9.2
through 9.7 to program the alarm relay functions. Each relay functions section includes a
description, a figure detailing its operation, a step-by-step setup procedure, and a table or
default and programmable limit settings.
The 56 has the following relay control functions:
Table 9-1. Alarm Relay Functions
Relay Control Functions
Common applications
Section
High/Low Concentration Alarm
measurement setpoint control
Sec. 9.2
Delay Timer
chemical mixing and neutralization
Sec. 9.3
Bleed and Feed
blowdown and chemical addition
Sec. 9.4
Totalizer Relay Activation
chemical dosing in reactors
Sec. 9.5
Interval Timer
periodic probe cleaning
Sec. 9.6
Date and Time Activation
seawater-cooled condensers
Sec. 9.7
9.2
High/Low Concentration Alarm
9.2.1
Description
High/Low concentration alarms are setpoint alarms with adjustable deadband. These operate as simple on/off alarms used for applications requiring discrete on/off control of pumps
and valves. Typical applications include demineralizer bed regeneration and blowdown in
boilers and cooling towers. Any active device variable in the 56 analyzer can be programmed
as a high/low concentration alarm including the primary or secondary variables, temperature, raw values and diagnostics.
A schematic of the high/low concentration (setpoint) alarm operation is shown below.
Figure 9-1. High/Low Concentration Alarm operation
PV
high alarm
setpoint
alarm activates
1.5
deadband (0.2)
1.3
alarm deactivates
time
high alarm logic
Alarm Relay Functions
67
Section 9: Alarm Relay Functions
56 Analyzer Instruction Manual
May 2014
9.2.2
LIQ_MAN_51-56
Setup
Access high/low concentration (setpoint) alarms by pressing ENTER/MENU from the main
screen and then Program/Relays/Configure Relay. From the main relay programming screen,
program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Setpoint as the relay type and press ENTER/MENU.
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to
the designated relay and press ENTER/MENU.
4. Logic: Set High for a high setpoint or Low for a low setpoint and press ENTER/MENU.
5. Setpoint: Enter the desired setpoints value. Press ENTER/MENU.
6. Deadband: To set deadband as a measurement value, enter the change in the process
value needed after the relay deactivates to return to normal (and thereby preventing repeated alarm activation). Press ENTER/MENU.
7. Select NEXT. Press ENTER/MENU to advance to the next setup screen.
8. Normal state: Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
Table 9-2. Defaults and programmable limits
Relay Function
Limits and Selections
Default
Setpoint
NA
NA
Logic
Low/High
High
Setpoint
*
*
Deadband
*
0.000
On time
0 to 999.9 min
0 min
Delay time
0 to 999.9 min
0 min
Normal state
Close/Open
Open
* See Appendix 1 – HART and Device Variables
9.3
Delay Timer
9.3.1
Description
Delay Timer is a concentration control scheme which delays live measurement after chemical
addition using (one or all four of) of the 56 alarm relays. This ensures sufficient mixing time
in a vessel or recirculation loop before live sensor measurement, preventing unmixed readings that might cause overshooting. Relay On time and Delay times are field-programmable.
Typical applications that would utilize the Delay Timer are: concentration control in vessels,
pH adjustments for neutralization and endpoint control for oxidation-reduction reactions.
68
Alarm Relay Functions
56 Analyzer Instruction Manual
Section 9: Alarm Relay Functions
LIQ_MAN_51-56
May 2014
A schematic of the Delay Timer operation is shown:
Figure 9-2. Delay Timer Alarm operation
PV
measurement
high alarm
setpoint
deadband
time
on time
9.3.2
delay time
Setup
Access Delay Timer by pressing ENTER/MENU from the main screen and then Program/Relays/
Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Delay Timer as the relay type and press ENTER/MENU.
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to
the designated relay and press ENTER/MENU.
4. Logic: Set High for high reading setpoint logic or Low relay logic for low reading setpoint
logic and press ENTER/MENU.
5. Setpoint: Enter the desired setpoints value. This will activate an alarm event when the
process measurement reaches the entered setpoint value. Press ENTER/MENU. See
Table 9-2 for entry limits.
6. Deadband: To set deadband as a measurement value, enter the change in the process
value needed after the relay deactivates to return to normal (and thereby preventing repeated alarm activation). Press ENTER/MENU.
7. Select NEXT Press ENTER/MENU to advance to the next setup screen.
8. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The assigned measurement value will be on hold during this time.
9. Delay time: Enter the time in minutes (X.X min) to take the assigned measurement off
hold after the relay is re-energized to begin reporting live values.
10. Normal state: Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
Alarm Relay Functions
69
Section 9: Alarm Relay Functions
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Table 9-3. Defaults and programmable limits
Relay Function
Limits and Selections
Default
Delay Timer
NA
NA
Logic
Low/High
High
Setpoint
*
*
Deadband
*
0.000
Normal state
Close/Open
Open
* See Appendix 1 – HART and Device Variables
9.4
Bleed and Feed
9.4.1
Description:
A bleed and feed timer is typically used to replace chemicals lost during blowdown. It
involves two or more relays. The bleed relay is a normal setpoint alarm relay. Once the bleed
relay deactivates, one or more feed relays activate for a percentage of the time the bleed
relay was on.
Bleed and Feed supports continuous monitoring of blow-down water conductivity to
determine the point of excessive conductivity. At a programmable maximum concentration
value, dumping (bleeding) of the excessively dirty blow-down water is triggered.
Subsequently, pumping (feeding) of additional make-up water chemicals is enabled to
account for lost blow-down water. Through level control, make-up water is added in
proportion to the volume of blowdown material lost through dumping and evaporation.
A schematic of the Bleed and Feed timer operation is shown:
Figure 9-3. Bleed and feed timer alarm operation
on
bleed on
feed on
AL1
(bleed)
AL2
(feed)
off
feed time = X% of bleed time
time
delay
9.4.2
Setup
Access Bleed and Feed Timers by pressing ENTER/MENU from the main screen and then Program/Relays/Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign relay 1 for Bleed and Feed and press ENTER/MENU.
2. Type: Select Bleed and Feed as the relay type and press ENTER/MENU.
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to
the designated relay and press ENTER/MENU.
4. Logic: Set High for high setpoint or Low for low setpoint and press ENTER/MENU.
70
Alarm Relay Functions
56 Analyzer Instruction Manual
Section 9: Alarm Relay Functions
LIQ_MAN_51-56
May 2014
5. Setpoint: Enter the desired setpoints value. This will activate an alarm event when the
process measurement reaches the entered setpoint value. Press ENTER/MENU. See Table
9-3 for entry limits.
6. Select NEXT Press ENTER/MENU to advance to the next setup screen.
7. Deadband: To set deadband as a measurement value, enter the change in the process
value needed after the relay deactivates to return to normal (and thereby preventing repeated alarm activation). Press ENTER/MENU.
8. Normal state: Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
9. Select NEXT Press ENTER/MENU to advance to the next setup screen: Configure feed
relay.
10. Feed relay: Assign relay 2, 3, or 4 as a feed relay and press ENTER/MENU.
11. Linked to bleed relay: No entry required. The relay originally programmed as the Bleed
relay is displayed.
12. Delay time: Enter the time in minutes (X.X min) after the Bleed time is activated before
triggering this feed relay.
13. Feed time equals: Enter the percent of time that the Bleed timer is on (X.X%) to activate
this feed relay (for pumping make-up water chemicals).
Relay Function
Limits and Selections
Default
Bleed and Feed
NA
NA
Logic
Low/High
High
Setpoint
*
*
Deadband
*
0.000
Normal state
Close/Open
Open
Feed relay
1, 2, 3, 4
not assigned
Linked to bleed relay
1, 2, 3, 4
1
Delay time
0 - 999.9 min
1.0 min
Feed time equals
0 - 999.9% of bleed time
10.0% of bleed time
Normal state
Close/Open
Open
* See Appendix 1 – HART and Device Variables
9.5
Totalizer Based Relay Activation
9.5.1
Description:
A totalizer based timer feeds chemicals for a preset period every time a programmed volume
of liquid has been added to or removed from a vessel. The relay energizes when the volume
has been reached and remains energized for a fixed time. The process repeats once the volume has been reached again.
Totalizer Based Relay Activation triggers a relay at user-defined intervals based on accumulated totalized flow. The scheme uses pulse inputs from a flow meter or 4-20mA current
input(s) from a flow transmitter to calculate total flow (as volume).
Alarm Relay Functions
71
Section 9: Alarm Relay Functions
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
A typical application for totalized flow relay activation is controlling chemical dosing in reactors.
A schematic of the Totalizer timer operation is shown:
Figure 9-4. Totalizer alarm operation
relay on
time
X volume
accumulated
X volume
accumulated
time
9.5.2
Setup
Access Totalizer Timers by pressing ENTER/MENU from the main screen and then Program/Relays/ Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Totalizer timer as the relay type and press ENTER/MENU.
3. Assign: Assign Pulse flow S1 (sensor 1) or S2 (sensor 2) as the measurement input and
press ENTER/MENU.
4. Active relay after: Enter accumulated volume (XX.XXXX) and units of measurement
(gal, thousand gal, million Gal, trillion Gal).
5. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The
assigned measurement value will be on hold during this time.
6. Normal state: Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to
normally open or normally closed upon power up.
Table 9-5. Defaults and programmable limits
Relay Function
Limits and Selections
Default
Totalizer Based Timer
NA
NA
Activate relay after
0 to 99.9990
10
units
Gal, Liters, cuft,m3 accumulated
E3gal (x1000 gal)
accumulated
On time
0 to 999.9 min
0 min
Normal state
Close/Open
Open
9.6
Interval Timer
9.6.1
Description:
The interval timer may be used for periodic sensor cleaning or periodic process adjustment.
The cycle begins at the Interval time when the switch is turned on. When the Interval time
72
Alarm Relay Functions
56 Analyzer Instruction Manual
Section 9: Alarm Relay Functions
LIQ_MAN_51-56
May 2014
has expired, the analyzer activates hold mode on the assigned measurement and the relay is
energized for the On time period.
A schematic of the Interval timer operation is shown:
Figure 9-5. interval timer Alarm operation
on time
hold
recovery
time
time
interval time
9.6.2
Setup
Access Interval timer by pressing ENTER/MENU from the main screen and then Program/Relays/ Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Interval timer as the relay type and press ENTER/MENU.
3. Interval time: Enter the time in hours (XX.X hours) between complete interval cycles.
4. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The assigned measurement value will be on hold during this time.
5. Recovery time: Enter the time in minutes (XX min) before the process is restored and
live measurements can resume.
6. *Hold while active: Select which sensors outputs should be on hold (S1, S2 or both)
during the interval timer activation time. Press ENTER/MENU.
7. Select NEXT Press ENTER/MENU to advance to the next setup screen.
8. Normal state: Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to
normally open or normally closed upon power up.
*56 units with software ver. 2.1X and greater allow override of interval timer to ensure that all relays and outputs are held if desired.
Table 9-6. Defaults and programmable limits
Alarm Relay Functions
Relay Function
Limits and Selections
Default
Interval timer
NA
NA
Interval time
0 to 999.9 hr
24.0 hr
On time
0 to 999.9 sec
10 sec
Recovery time
0 to 999 sec
60 sec
Hold while active
0 to 999 sec
0 sec
Normal state
Sensor 1, Sensor 1, both
Sensor 1
73
Section 9: Alarm Relay Functions
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
9.7
Date and Time Activation
9.7.1
Description:
This relay feature allows programming of 1 to 4 relays to activate on an assigned day of the
week and time of day or night for an assigned interval. They function like sprinkler timers.
The programmable timeframe cycle is two weeks.
An example application for Date and Time Activation is daily chlorine dosing in seawatercooled condensers.
The Date and Time relay setup screen is shown:
Figure 9-6. Date and Time alarm operation
Live display
Outputs
Relays
Measure Temperature
pH diagnostic setup
Security
Enter duration in minutes.
Week 1
Start
Sun
Mon
Tues
Wed
Thurs
Fri
Sat
hh
hh
hh
hh
hh
hh
hh
mm
mm
mm
mm
mm
mm
mm
Duration
NEXT
BACK
Fault/warning banner
9.7.2
Setup
Access Date and Time timer by pressing ENTER/MENU from the main screen and then
Program/Relays/Configure Relay. From the main relay programming screen, program this
feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Date and time timer as the relay type and press ENTER/MENU.
3. Select NEXT Press ENTER/MENU to advance to the next setup screen.
4. The Week 1 calendar will appear. Program the start relay activate time by entering
day(s) of week, hour(s) of day and minutes for each hour. Enter the duration of time in
minutes (XX min) for relay activation. Up to four relays can be simultaneously energized for any programmed times
5. Select NEXT Press ENTER/MENU to advance to the next setup screen.
6. The Week 2 calendar will appear. Repeat the programming entries for week 2 in the
same manner as week 1. Up to four relays can be simultaneously energized for any programmed times.
CAUTION
Date and Time timer operation depends on accurate setup of the internal real time clock. Continuous powered operation of the 56 analyzer is recommended to preserve programmed Date and Time
timer clock settings.
74
Alarm Relay Functions
56 Analyzer Instruction Manual
Section 10: Calibration
LIQ_MAN_51-56
May 2014
Section 10.0 Calibration
10.1
Calibration – Introduction
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a
calibrated laboratory instrument, or standardizing to some known reference (such as a
commercial buffer). The auto-recognition feature of the analyzer will enable the appropriate
calibration screens to allow calibration for any single sensor configuration or dual sensor
configuration of the analyzer. Completion of Quick Start upon first power up enables live
measurements but does not ensure accurate readings in the lab or in process. Calibration
should be performed with each attached sensor to ensure accurate, repeatable readings.
This section covers the following programming and configuration functions:
1. Auto buffer cal for pH (pH Cal - Sec.7.2)
2. Manual buffer cal for pH (pH Cal - Sec.7.2)
3. Set calibration stabilization criteria for pH (pH Cal - Sec.7.2)
4. Standardization calibration (1-point) for pH, ORP and Redox (pH Cal - Sec.7.2 and 7.3)
5. Entering the cell constant of a conductivity sensor (Conductivity Cal - Sec. 7.4 and 7.5)
6. Calibrating the sensor in a conductivity standard Conductivity Cal - Sec. 7.4 and 7.5)
7. Calibrating the analyzer to a laboratory instrument (Contacting Conductivity Cal - Sec.7.4)
8. Zeroing an chlorine, oxygen or ozone sensor (Amperometric Cal - Sec’s 7.6, 7.7, 7.8)
9. Calibrating an oxygen sensor in air (Oxygen Cal - Sec’s 7.6)
10. Calibrating the sensor to a sample of known concentration (Amperometric Cal - Sec’s
7.6, 7.7, 7.8)
11. Enter a manual reference temperature for temperature compensation of the
process measurement
10.2
pH Calibration
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto
calibration instead of manual calibration. Auto calibration avoids common pitfalls and
reduces errors. The analyzer recognizes the buffers and uses temperature-corrected pH
values in the calibration. Once the 56 successfully completes the calibration, it calculates and
displays the calibration slope and offset. The slope is reported as the slope at 25°C.
To calibrate the pH loop with a connected pH sensor, access the Calibration screen by
pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press
ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen
will appear with detailed instructions and information.
The following calibration routines are covered:
1. Auto Calibration - pH 2 point buffer
calibration with auto buffer
recognition
2. Manual Calibration - pH 2 point buffer
calibration with manual buffer value
entry
3.
Calibration
Standardization - pH 1 point buffer
calibration with manual buffer value
entry
75
Section 10: Calibration
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
4. Entering A Known Slope Value - pH
Slope calibration with manual entry
of known slope value
1. To Auto Calibrate the pH loop using
2 point buffer calibration with auto
buffer recognition, select Auto Buffer
and follow the step-by-step procedures displayed on-screen.
2. To Manual Calibrate the pH loop
using 2 point buffer calibration with
manual buffer value entry, select Manual Buffer and follow the step-by-step procedures
displayed on-screen.
3. To Standardization Calibrate the pH loop using 1 point buffer calibration with manual
buffer value entry, select Standardize and follow the step-by-step procedures displayed onscreen.
4. To Calibrate the pH loop using with manual entry of a Known Slope Value and Reference offset value, select Slope/Offset and follow the step-by-step procedures displayed on-screen.
10.3
ORP Calibration
For process control, it is often important to make the measured ORP agree with the ORP of a
standard solution. During calibration, the measured ORP is made equal to the ORP of a
standard solution at a single point.
To calibrate the ORP loop with a connected ORP sensor, access the Calibration screen by
pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press
ENTER/MENU. Press INFO at any time to
learn more about this procedure. A yellow
screen will appear with detailed
instructions and information.
The following calibration routine is
covered:
1. To Standardization the ORP loop
using 1 point buffer calibration with
manual buffer value entry, follow the
step-by-step procedures displayed
on-screen.
76
Calibration
56 Analyzer Instruction Manual
Section 10: Calibration
LIQ_MAN_51-56
10.4
May 2014
Contacting Conductivity Calibration
New conductivity sensors rarely need calibration. The cell constant printed on the label is
sufficiently accurate for most applications.
CALIBRATING AN IN-SERVICE CONDUCTIVITY SENSOR
After a conductivity sensor has been in service for a period of time, recalibration may be necessary. There are three ways to calibrate a sensor.
a.
Use a standard instrument and sensor to measure the conductivity of the process stream. It is
not necessary to remove the sensor from the process piping. The temperature correction
used by the standard instrument may not exactly match the temperature correction used
by the 56. To avoid errors, turn off temperature correction in both the analyzer and the
standard instrument.
b. Place the sensor in a solution of known conductivity and make the analyzer reading
match the conductivity of the standard solution. Use this method if the sensor can be
easily removed from the process piping and a standard is available. Be careful using
standard solutions having conductivity less than 100 µS/cm. Low conductivity standards
are highly susceptible to atmospheric contamination. Avoid calibrating sensors with
0.01/cm cell constants against conductivity standards having conductivity greater than
100 µS/cm. The resistance of these solutions may be too low for an accurate
measurement. Calibrate sensors with 0.01/cm cell constant using method c.
c.
To calibrate a 0.01/cm sensor, check it against a standard instrument and 0.01/cm
sensor while both sensors are measuring water having a conductivity between 5 and 10
µS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide, saturate the
sample with air before making the measurements. To ensure adequate flow past the
sensor during calibration, take the sample downstream from the sensor. For best
results, use a flow-through standard cell. If the process temperature is much different
from ambient, keep connecting lines short and insulate the flow cell.
To calibrate the conductivity loop with a connected contacting conductivity sensor, access
the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure.
A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zero the analyzer with the sensor attached
2. In Process Cal Standardize the sensor
to a known conductivity
3. Cell K: 1.00000/cm Enter the cell
Constant for the sensor
4. Meter Cal Calibrate the analyzer to a
lab conductivity instrument
5. Cal Factor: 0.95000/cm Enter the Cal
Factor for 4-Electrode sensors from
the sensor tag
Calibration
77
Section 10: Calibration
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step procedures
displayed on-screen.
2. To perform an In-Process Calibration
of the conductivity loop by
Standardizing the sensor to
a known conductivity, follow
the step-by-step procedures
displayed on-screen.
3. To calibrate the conductivity loop by entering a Cell constant, Enter the cell Constant for
the sensor and follow the step-by-step procedures displayed on-screen.
4. To Meter Cal Calibrate the analyzer to a lab conductivity instrument, follow the step-bystep procedures displayed on-screen.
5. To enter the Cal Factor to support calibration of a 4-Electrode sensors, enter the Cal
Factor for the 4-Electrode sensors from the sensor tag and follow the step-by-step
procedures displayed on-screen.
10.5
Toroidal Conductivity Calibration
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a conductivity standard). This section contains procedures for the first time use and for routine
calibration of the 56 analyzer.
To calibrate the conductivity loop with a connected contacting conductivity sensor, access
the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2
Measurement and press ENTER/MENU. Press INFO at any time to learn more about this
procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zero the analyzer with the
sensor attached
2. In Process Cal Standardize the sensor
to a known conductivity
3. Cell K: 1.00000/cm Enter the cell
Constant for the sensor
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step procedures displayed onscreen.
2. To perform an In-Process
Calibration of the conductivity loop
by Standardizing the sensor to a
known conductivity, follow the stepby-step procedures displayed onscreen.
3. To calibrate the conductivity loop by
entering a Cell constant, Enter the
cell Constant for the sensor and follow the step-by-step procedures displayed on-screen.
78
Calibration
56 Analyzer Instruction Manual
LIQ_MAN_51-56
10.6
Section 10: Calibration
May 2014
Calibration —Chlorine
With a Chlorine measurement board and the appropriate sensor, the 56 can measure any of
four variants of Chlorine:
• Free Chlorine
• Total Chlorine
• Monochloramine
• pH-independent Free Chlorine
The section describes how to calibrate any compatible amperometric chlorine sensor. The following calibration routines are covered in the family of supported Chlorine sensors:
• Air Cal
• Zero Cal
• In Process Cal
10.6.1
Calibration — Free Chlorine
A free chlorine sensor generates a current directly proportional to the concentration of free
chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no
chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale
standard). The zero calibration is necessary because chlorine sensors, even when no chlorine
is in the sample, generate a small current called the residual current. The analyzer
compensates for the residual current by subtracting it from the measured current before
converting the result to a chlorine value. New sensors require zeroing before being placed in
service, and sensors should be zeroed whenever the electrolyte solution is replaced.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU
from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at
any time to learn more about this procedure. A yellow screen will appear with detailed
instructions and information.
The following calibration routines are
covered:
1. Zero Cal Zeroing the sensor in
solution with zero free chlorine
2. Grab Cal Standardizing to a
sample of known free chlorine
concentration
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step procedures displayed onscreen.
2. To perform a Grab Calibration by
Standardizing the sensor, follow the
step-by-step procedures displayed
on-screen.
Calibration
79
Section 10: Calibration
May 2014
10.6.2
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Calibration — Total Chlorine
Total chlorine is the sum of free and combined chlorine. The continuous determination of total
chlorine requires two steps. First, the sample flows into a conditioning system (TCL) where a
pump continuously adds acetic acid and potassium iodide to the sample. The acid lowers the pH,
which allows total chlorine in the sample to quantitatively oxidize the iodide in the reagent to iodine. In the second step, the treated sample flows to the sensor. The sensor is a membrane-covered amperometric sensor, whose output is proportional to the concentration of iodine.
Because the concentration of iodine is proportional to the concentration of total chlorine, the analyzer can be calibrated to read total chlorine. Because the sensor really measures iodine, calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a
solution containing a known amount of iodine (full-scale standard). The Zero calibration is necessary because the sensor, even when no iodine is present, generates a small current called the
residual current. The analyzer compensates for the residual current by subtracting it from the
measured current before converting the result to a total chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process
Calibration is to establish the slope of the calibration curve. Because stable total chlorine standards do not exist, the sensor must be calibrated against a test run on a grab
sample of the process liquid. Several manufacturers offer portable test kits for this purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the
main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to
learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution
with zero total chlorine
2. Grab Cal Standardizing to a
sample of known total chlorine concentration
1. To Zero Calibrate the analyzer with the
sensor attached, follow the step-by-step
procedures
displayed on-screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-bystep procedures
displayed on-screen.
80
Calibration
56 Analyzer Instruction Manual
LIQ_MAN_51-56
10.6.3
Section 10: Calibration
May 2014
Calibration — Monochloramine
A monochloramine sensor generates a current directly proportional to the concentration
of monochloramine in the sample. Calibrating the sensor requires exposing it to a solution
containing no monochloramine (zero standard) and to a solution containing a known amount of
monochloramine (full-scale standard). The Zero calibration is necessary because monochloramine
sensors, even when no monochloramine is in the sample, generate a small current called the
residual or zero current. The analyzer compensates for the residual current by
subtracting it from the measured current before converting the result to a monochloramine
value. New sensors require zeroing before being placed in service, and sensors should be zeroed
whenever the electrolyte solution is replaced. The best zero standard is deionized water.
The purpose of the In Process calibration is to establish the slope of the calibration curve.
Because stable monochloramine standards do not exist, the sensor must be calibrated against
a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits
for this purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the
main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to
learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution
with zero total chlorine
2. Grab Cal Standardizing to a sample of
known monochloramine concentration
1. To Zero Calibrate the analyzer with the
sensor attached, follow the step-by-step
procedures displayed on-screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-bystep procedures displayed on-screen.
Calibration
81
Section 10: Calibration
56 Analyzer Instruction Manual
May 2014
10.6.4
LIQ_MAN_51-56
pH-Independent Free Chlorine Measurement
A free chlorine sensor generates a current directly proportional to the concentration of free
chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no
chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale
standard). The zero calibration is necessary because chlorine sensors, even when no chlorine is
in the sample, generate a small current called the residual current. The
analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed
in service, and sensors should be zeroed whenever the electrolyte solution is replaced.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the
main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to
learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1.
Zero Cal Zeroing the sensor in solution with zero total chlorine
2. Grab Cal Standardizing to a sample
of known chlorine concentration
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step
procedures displayed on-screen.
2. To perform a Grab Calibration by
Standardizing the sensor, follow the
step-by-step procedures displayed
on-screen.
10.7
Calibration — Oxygen
Oxygen sensors generate a current directly proportional to the concentration of dissolved
oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no
oxygen (zero standard) and to a solution containing a known amount of oxygen (full-scale
standard). The Zero Calibration is necessary because oxygen sensors, even when no oxygen
is present in the sample, generate a small current called the residual current. The analyzer
compensates for the residual current by subtracting it from the measured current before
converting the result to a dissolved oxygen value. New sensors require zeroing before being
placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced.
The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen
can also be used. The 499A TrDO sensor, used for the determination of trace (ppb) oxygen
levels, has very low residual current and does not normally require zeroing. The residual
current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen. The purpose of the In
Process Calibration is to establish the slope of the calibration curve. Because the solubility of
82
Calibration
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 10: Calibration
May 2014
atmospheric oxygen in water as a function of temperature and barometric pressure is well
known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated
water is difficult to prepare and use, so the universal practice is to use air for calibration. From the
point of view of the oxygen sensor, air and air-saturated water are identical. The equivalence
comes about because the sensor really measures the chemical potential of oxygen. Chemical
potential is the force that causes oxygen molecules to diffuse from the sample into the sensor
where they can be measured. It is also the force that causes oxygen molecules in air to dissolve
in water and to continue to dissolve until the water is saturated with oxygen. Once the water is
saturated, the chemical potential of oxygen in the two phases (air and water) is the same.
Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules
diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends
on the difference in chemical potential between oxygen in the sensor and oxygen in the sample.
An electrochemical reaction, which destroys any oxygen molecules entering the sensor, keeps the
concentration (and the chemical potential) of oxygen inside the sensor equal to zero. Therefore,
the chemical potential of oxygen in the sample alone determines the diffusion rate and the
sensor current. When the sensor is calibrated, the chemical potential of oxygen in the standard
determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is
immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the
calculation of solubility in common units (like ppm DO) simpler, it is convenient to use watersaturated air for calibration. Automatic air calibration is standard. The user simply exposes the
sensor to water-saturated air. The analyzer monitors the sensor current. When the current is
stable, the analyzer stores the current and measures the temperature using a temperature
element inside the oxygen sensor. The user must enter the barometric pressure. From the
temperature the analyzer calculates the saturation vapor pressure of water. Next, it calculates
the pressure of dry air by subtracting the vapor pressure from the barometric pressure. Using
the fact that dry air always contains 20.95% oxygen, the analyzer calculates the partial pressure
of oxygen. Once the analyzer knows the partial pressure of oxygen, it uses the Bunsen
coefficient to calculate the equilibrium solubility of atmospheric oxygen in water at the
prevailing temperature. At 25°C and 760 mmHg, the equilibrium solubility is 8.24 ppm. Often it
is too difficult or messy to remove the sensor from the process liquid for calibration. In this case,
the sensor can be calibrated against a measurement made with a portable laboratory
instrument. The laboratory instrument typically uses a membrane-covered amperometric sensor
that has been calibrated against water-saturated air.
To calibrate the oxygen sensor, access the Calibration screen by pressing ENTER/MENU from
the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any
time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in a
medium with zero oxygen
2. Air Cal Calibrating the sensor
in a water-saturated air sample
3. In Process Cal Standardizing to
a sample of known oxygen
concentration
4. Sen@ 25°C:2500µA/ppm Entering a
known slope value for sensor response.
Calibration
83
Section 10: Calibration
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step procedures displayed onscreen.
2. To Air Cal Calibrating the sensor in a
water-saturated air sample, follow
the step-by-step procedures displayed on-screen.
3. To perform an In-Process Calibration
by Standardizing the sensor, follow
the step-by-step procedures displayed on-screen.
4. To calibrate the oxygen sensor by manually Entering a known slope value for sensor response, follow the step-by-step procedures displayed on-screen.
10.8
Calibration — Ozone
An ozone sensor generates a current directly proportional to the concentration of ozone in
the sample. Calibrating the sensor requires exposing it to a solution containing no ozone
(zero standard) and to a solution containing a known amount of ozone (full-scale standard).
The Zero Calibration is necessary because ozone sensors, even when no ozone is in the
sample, generate a small current called the residual or zero current. The analyzer
compensates for the residual current by subtracting it from the measured current before
converting the result to an ozone value. New sensors require zeroing before being placed in
service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best
zero standard is deionized water. The purpose of the In Process Calibration is to establish the
slope of the calibration curve. Because stable ozone standards do not exist, the sensor must
be calibrated against a test run on a grab sample of the process liquid. Several manufacturers
offer portable test kits for this purpose.
To calibrate the ozone sensor, access the Calibration screen by pressing ENTER/MENU from the
main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to
learn more about this procedure. A yellow
screen will appear with detailed
instructions and information.
The following calibration routines are
covered:
1. Zero Cal Zeroing the sensor in
solution with zero total chlorine
2. Grab Cal Standardizing to a sample of
known ozone concentration
1. To Zero Calibrate the analyzer with
the sensor attached, follow the stepby-step procedures
displayed on-screen.
2. To perform a Grab Calibration by
Standardizing the sensor, follow the
step-by-step procedures displayed
on-screen.
84
Calibration
56 Analyzer Instruction Manual
LIQ_MAN_51-56
10.9
Section 10: Calibration
May 2014
Calibrating Temperature
Most liquid analytical measurements require temperature compensation (except ORP). The
56 performs temperature compensation automatically by applying internal temperature
correction algorithms. Temperature correction can also be turned off. If temperature
correction is off, the 56 uses the manual temperature entered by the user in all temperature
correction calculations.
To calibrate temperature, access the Calibration screen by pressing ENTER/MENU from the
main screen, select S1 or S2 Temperature and press ENTER/MENU. Press INFO at any time to
learn more about this procedure. A
yellow screen will appear with detailed
instructions and information.
The following calibration routine is
covered:
1. Temperature with manual
temperature entry
1. To Calibrate Temperature, follow the
step-by-step procedures
displayed on-screen.
10.10
Turbidity
This section describes how to calibrate the turbidity sensor against a user-prepared standard
as a 2-point calibration with deionized water, against a 20 NTU user-prepared standard as a
single point calibration, and against a grab sample using a reference turbidimeter.
To calibrate the turbidity sensor, access the Calibration screen by pressing ENTER/MENU from
the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time
to learn more about this procedure. A yellow screen will appear with detailed instructions and
information.
The following calibration routine is covered:
1. Slope Calibration Slope cal with pure
water and a standard of known turbidity
2. Standardize Calibration
Standardizing the sensor to a known
turbidity
3. Grab Calibration Standardizing the
sensor to a known turbidity based on a
reference turbidimeter
Calibration
85
Section 10: Calibration
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
1. To calibrate the turbidity loop using
Slope Calibration with pure water
and a standard of known turbidity,
follow the step-by-step procedures
displayed on-screen.
2. To calibrate the turbidity loop using
Standardize Calibration by Standardizing the sensor to a known turbidity,
follow the step-by-step procedures
displayed on-screen.
3. To calibrate the turbidity loop using Grab Calibration by Standardizing the sensor to a
known turbidity based on a reference turbidimeter, follow the step-by-step procedures
displayed on-screen.
10.11
Pulse flow
A variety of pulse flow sensors can be wired to the Flow signal input board to measure flow volume, total volume and flow difference (if 2 Flow signal boards are installed). The 56 Flow signal
board will support flow sensors that are self-driven (powered by the rotation of the impeller
paddle-wheel).
To calibrate the pulse flow sensor, access the Calibration screen by pressing ENTER/MENU from
the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time
to learn more about this procedure. A yellow screen will appear with detailed instructions and
information.
The following calibration routine is covered:
1. K Factor A constant value representing pulses/Gal of flow
2. Frequency/Velocity & Pipe
Alternate cal method – requires manual entry of frequency (Hz) per velocity and Pipe diameter used
3. In process Calibration based on
known volume per unit of time
4. Totalizer Control User settings to
stop, restart and reset total volume
meter
1. To enter a K Factor constant value
representing pulses/Gal
of flow, follow the step-by-step
procedures displayed on-screen.
2. To calibrate pulse flow
Frequency/Velocity & Pipe as an
alternate cal method, follow
the step-by-step procedures
displayed on-screen.
3. To In-process Calibration the pulse flow sensor based on known volume per unit of time,
follow the step-by-step procedures displayed on-screen.
4. To stop, restart and reset Totalizer Control total volume meter, follow the step-by-step
procedures displayed on-screen.
86
Calibration
56 Analyzer Instruction Manual
Section 10: Calibration
LIQ_MAN_51-56
May 2014
Section 11.0 HART® Communications
11.1
Introduction
The 56 can communicate with a HART host using HART Revision 5 or HART Revision 7. The
revision of HART used by the 56 can be selected using the keypad/display or a HART master
such as AMS or the 475 Handheld Communicator. The default version of HART is Revision 5.
Since some HART hosts cannot accommodate HART 7, the choice of HART Revision should
be based on the capabilities of the host, and should be chosen as a first step in configuration.
When HART 5 is chosen, the Device Revision of the 56 is Device Revision 1; when HART 7 is
chosen the Device Revision is Revision 2, or higher, for later revisions of the 56. The Device
Revision of the DD (Device Description) and install files for AMS and DeltaV used should be
the same as the Device Revision of the 56.
HART 5 Device Identification (56 Revision 1):
Manufacturer Name: Rosemount Analytical Inc.
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 86 (0x0056)
HART Protocol Revision: 5.1
Device Revision: 1
Capabilities: Supports all signal boards except the turbidity and flow/4-20mA input signal
boards.
HART 7 Device Identification (56 Revision 2):
Manufacturer Name: Rosemount Analytical Inc.
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 11862 (0x2E56)
HART Protocol Revision: 7.1
Device Revision: 2
Capabilities: Supports all signal boards.
HART 7 Device Identification (56 Revision 3):
Manufacturer Name: Rosemount Analytical Inc.
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 11862 (0x2E56)
HART Protocol Revision: 7.1
Device Revision: 3
Capabilities: Supports all signal boards and the complete set of parameters for standardized
PID with PID transport time feature and TPC control.
HART Communications
87
Section 11: HART Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
11.2
Physical Installation and Configuration
11.2.1
HART Wiring and Output Configuration
HART communications is superimposed on Analog Output 1 for all of the measurements and
parameters of the 56. The 4-20 mA current of Analog Output 1 can be configured by the keypad display to be powered by the 56 (Output 1 power: Internal), or by an external 24 VDC
power supply, or an I/O that provides power (Output 1 power: External).
11.2.2
HART Multidrop (Bus) Configuration
The HART Polling Address should be left at its default value of “0”, unless the 56 is used in a
Multidrop configuration with up to 14 other transmitters. When the Polling Address is
greater than “0”, the 4-20 mA output is held at 4 mA or below, and does not change in response to changes in the measurement in HART 5.
In HART 7, Loop Current Mode should be set to “On” to hold the current output to a minimum value. In both HART 5 and HART 7, Output Power should be set to “External” so that an
external 24 VDC power supply can be used to power the multidrop bus.
11.2.3
HART Configuration
To access the HART Configuration screens, press the “HART” button in the Main Menu. The
following controls are available:
HART Configuration Screen 1: Basic Definitions
• Host HART mode – toggles between HART version 5 and HART version 7. If the HART
host being used can accommodate HART 7, HART 7 should be chosen due to its larger
feature set. If the host can only use HART 5, then HART 5 must be chosen.
NOTE: If the 56 is connected to a HART host and the HART version is changed, the host
will likely detect the transmitter as a new transmitter with a different device revision
number.
• Tag – The traditional 8 character HART tag number
• Long tag – HART tag number of up to 32 characters (HART 7 only).
88
HART Communications
56 Analyzer Instruction Manual
Section 11: HART Communications
LIQ_MAN_51-56
May 2014
• Polling address – Choose “0” unless Multidrop is being used. If Multidrop is being
used, each transmitter should have its own polling address of from 1 to 15.
• Loop current mode – Set Output 1 current to a minimum value for multidrop applications (HART 7 only).
• Output 1 Power – Select “Internal” to power Output 1 with the 56. Select “External”
to power the current loop with an external power supply, e.g. a host I/O that provides
power (source) to the transmitter (sink).
HART Configuration Screen 2 Basic Definitions
• Burst command:
Off – Turns burst mode off
Cmd 1 – Bursts the Primary Value
Cmd 2 – Bursts Loop Current + % of range of the Primary Value
Cmd 3 – Bursts Dynamic Variables (PV, SV, TV, & QV) + Loop Current
Cmd 9 – Bursts up to 8 Device Variables with time stamp and status (HART 7 only)
Cmd 33 – Bursts 4 Device Variables
Cmd 48 – Bursts Additional Transmitter Status Bits (HART 7 only)
Cmd 93 – Bursts Trend Data (HART 7 only)
• Find device cmd – Setting Find Device to “On”, enables the 56 to be indentified by the
host. The transmitter returns identity information including device type, revision level,
and device ID.
• Response preambles – Preambles synchronize the receiver to the incoming data. Response preambles are the number of bytes of preambles (2 to 20) sent by the 56 at the
start of a response packet.
11.3
Measurements Available via HART
A number of live measurements are made available by HART in addition to the main measurements such as pH or Conductivity. All of these measurements are called Device Variables,
HART Communications
89
Section 11: HART Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
which can be mapped to the Dynamic Variables PV, SV, TV, and QV for regular reading by the
typical HART host.
The 56 assigns the Dynamic Variables PV, SV, TV, and QV to Analog Outputs 1, 2, 3, and 4
respectively. Conversely, measurements assigned to Outputs 1 through 4, will automatically
be assigned to PV through QV.
Each measurement board will have its own set of Device Variables, based on the secondary
measurements used in making the main measurement. Appendix 1 shows the Device
Variable for the each sensor boards used in the 56, and the Dynamic Variables, which they
can be mapped to.
11.4
Diagnostics Available via HART
11.4.1
Status Information -- Device Status Bits
Bit 0
Primary Variable out of Limits:
This bit is set when PV is out of its limits.
Bit 1
Non-primary Variable out of Limits:
This bit is set when non PV is out of its limits.
Bit 2
Loop Current Saturated:
This bit is set when the Analog Output 1 current is less than 1.0 mA or greater than
22.0 mA and the Device Status Bytes bit 3 is not set.
Bit 3
Loop Current Fixed:
This bit is set when Analog Output 1 does not follow the process. This bit is cleared
when Analog Output 1 follows the process.
Bit 4
More Status Available:
The “more status available” bit will be set when the device status condition occurs (i.e.
bit goes from 0 to 1) on at least one of the Additional Transmitter Status bits are set.
Bit 5
Cold Start:
This bit is set when a Master Reset is performed either by Command 42, or a power
cycle. This bit is cleared after the first response or burst. In the case of a burst, the
first burst always goes to the primary master.
Bit 6
Configuration Changed:
This bit is set when the last bytes of an EEPROM writing sequence is completed. Since
EEPROM writes are delayed until after the response, the immediate acknowledgement will not have the bit set, however, after the EEPROM write completes, the bit
will be set. This bit applies to all EEPROM writes whether or not they apply directly to
the configuration of the instrument or not. This bit is cleared when Command 38 is
executed.
Bit 7
Field Device Malfunction:
This bit is set when any of the following conditions is true, and cleared when all of the
following conditions are false:
a) There is at least one main board fault.
b) There is at least one Sensor 1 fault.
c) There is at least one Sensor 2 fault.
90
HART Communications
56 Analyzer Instruction Manual
Section 11: HART Communications
LIQ_MAN_51-56
11.4.2
May 2014
Status Information – Extended Device Status Bits (HART 7 only)
Bit 0
Maintenance Required:
Any calibration error will set this bit. Any calibration for either Sensor 1 or Sensor 2 that
fails will set this bit. This bit gets cleared if all device variables are calibrated successfully.
This bit is set to indicate that, while the device has not malfunctioned, the Field Device requires maintenance.
Bit 1
Device Variable Alert:
This bit will get set when at least one of device status byte of all the valid Device Variables does not equal to "good and not limited" (i.e. the higher 4 bits not equal to
0x1100).
This bit is set if any Device Variable is in an Alarm or Warning State. The host should
identify the Device Variable(s) causing this to be set using the Device Variable Status indicators.
Bit 3
Critical Power Failures:
This bit will always be cleared on the 56.
11.4.3
Additional Transmitter Status (Command 48)
Additional Transmitter Status provides diagnostic status bits specific to the condition of sensors, sensor boards, and the main board of the 56. Calibration errors and notification of events,
such as calibration in progress and relay activation are also indicated by status bits. Appendix 2
shows these bits organized according to the main board or applicable sensor board.
11.5
HART Hosts
A HART host can access live measurements, diagnostic messages, and provide a tool for configuring the measurement and calibrating the 56. Two examples are shown below:
11.5.1
AMS Intelligent Device Manager
The AMS Device Intelligent Device Manager is member of the AMS Suite of asset management
applications, which provides tools for configuration, calibration, diagnosing, and documenting
transmitters and valves. The following AMS windows are examples of these functions:
FIGURE 11-1 Main Measurement and Overall Status
HART Communications
91
Section 11: HART Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
FIGURE 11-2 Device Variables and Dynamic Variables
FIGURE 11-3 Diagnostic Messages (Additional Transmitter Status)
92
HART Communications
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 11: HART Communications
May 2014
FIGURE 11-4 Configuration
FIGURE 11-5 Calibration
HART Communications
93
Section 11: HART Communications
56 Analyzer Instruction Manual
May 2014
11.5.2
LIQ_MAN_51-56
475 Field Communicator
HART (and Fieldbus) devices can be access in the field using the 475, which provides the
same basic functionality as the AMS Intelligent Device Manager. Asset management
information can be uploaded into the AMS database from the 475 for a common database
for asset management data. The 475 uses a color menu driven display.
11.6
Wireless Communication using the 56
The 56 can communicate by Wireless HART using the Smart Wireless THUM Adaptor and
the 1420 Smart Wireless Gateway. All the information available with the wired device can
be accessed wirelessly, making it possible to have the measurements and benefits of HART
communication in locations where running cable would be difficult or prohibitively
expensive.
Although HART 5 or HART 7 can burst the Dynamic Variables (PV, SV, TV, & QV), HART 7
should be used with the THUM because up to 8 Device Variables can be continually burst
using Command 9.
11.7
Field Device Specification (FDS)
For more details on the implementation of HART in the 56 and its command structure, the
Field Device Specification for the relevant Device Revision should be consulted. They can be
downloaded from our website.
94
HART Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Section 12.0 Profibus Communications
12.1
General
The 56 can communicate using Profibus DP-V1 by purchasing a 56 with a Profibus
communication board installed (model code 56--0X--2X--3X--DP), or by adding a Profibus
communications board to the standard 56 with HART communications.
The standard HART 56 analyzer can be converted to Profibus DP by adding a Profibus board.
There are two different Profibus boards for the 56, which must be chosen to be compatible with
the main board software version of the 56.
• If the 56 software version begins with a 2 (2.XX), Profibus board PN 24285-02 should be chosen.
• If the 56 software version begins with a 1 (1.XX), Profibus board PN 24285-01 should be chosen.
A 56 with Profibus communications retains all the analog and discrete functionality of the
standard HART 56 transmitter. The four 4 to 20 mA analog outputs can be configured as analog
outputs or PID control outputs.
The four discrete relay outputs can be configured for concentration control (high/low alarms
with or without a delay timer), event based activation, timer based activation, or TPC (Duty
Cycle/ Pulse Width Modulation) control.
12.2
Profibus Features
As outlined in the Device Description Data Field (GSD), the 56 has the following Profibus DPV1 identification and capabilities:
Identification:
GSD Revision: 5
Vendor Name: Rosemount Analytical, Inc.
Model Name: MODEL 56 PROFIBUS DP
Revision: V1.00
Identification Number: 0x10AA
Hardware Release: V1.00
Software Release: V1.00
Implementation Type: VPC3+C
Device: DP-V1 Slave
Baud Rates Supported:
All Profibus DP baud rates are supported, from 9.6 kbps to 12 Mbps.
Additional Capabilities:
•
Profibus Communications
The 56 Profibus board supports the automatic transmission rate recognition. It
automatically sets its baud rate to the transmission rate of the Profibus network,
specified by the master.
95
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
•
The DP device supports the function Set_Slave_Add for setting the slave address via the
PROFIBUS. The slave address for the 56 can be set automatically or manually by the
Profibus master.
The DP device supports DP-V0 cyclic services and Identification and Maintenance (I&M)
function: - I&M 0
•
The Minimum Slave Interval is 100 μsec minimum interval between two poll cycles.
56 Modules
The 56 is a multiparameter device, which supports a number of combinations of 2 analytical
measurements, and four relays which can be configured for a number of discrete functions.
As a result, the 56 is defined as a modular station, to allow it to cover these measurements,
relay status, and associated information. The 56 has 36 modules available and 33 can be inserted into the device. They include:
•
The two main measurements, their units, and the measurement
•
The temperature measurements from each board and their units
•
Dual conductivity measurements: conductivity ration, percent passage and percent rejection
•
Dual flow board measurements: flow recovery, flow difference, and flow unit
•
Alarm relays (4 possible): their type, state, logic assignment and simulate status
Details of the module are in Section 12.3 Profibus Communications.
56 Device Diagnostics
A number of diagnostic bits are provided via Profibus, which cover the fault and warning messages
for the 56 main board, signal boards, and measurements. They are organized as follows:
Table 12-1. Device Diagnostic Bits
Diagnostic Bits*
16
Bit Description
Change in Diagnostic Data
24-27
Main Board Faults
33-38
Main Board Warnings
40-57
Signal Board 1 and Measurement Faults
72-89
Signal Board 2 and Measurement Faults
104-121
Signal Board 1 and Measurement Warnings
136-153
Signal Board 2 and Measurement Warnings
*Bit position after the header byte.
Details of the diagnostic bits appear in Tables 12.3.10 and 12.3.11.
12.3
Profibus Communications
12.3.1
Configuring the 56 for Profibus Communications
Baud Rates
The 56 supports all the baud rates of Profibus DP as shown in Table 12.2.1 below. Since the 56
supports transmission rate recognition, it will recognize the baud rate used on the network,
and set its own baud rate to match the network’s rate.
96
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Table 12-2. Baud Rates Supported
Baud
Rate
Maximum Response Time
for Responders
(unit is bit time)
9.6 kbps
60
19.2 kbps
60
45.45 kbps
60
93.75 kbps
60
187.5 kbps
60
500 kbps
100
1.5 Mbps
150
3 Mbps
250
6 Mbps
450
12 Mbps
800
Slave Address
The 56 supports the set slave address function, which makes it possible for the Profibus master to set the slave address of the 56. Some Profibus masters cannot set slave addresses, and
the slave addresses have to be set in each slave device. If this is case, the slave address can be
set in the 56 by pressing the Enter/Menu key and selecting DP: Profibus, which will launch the
Profibus configuration window:
•
The Profibus address is writeable up to a value of 126.
•
At the same time the slave address is being set, Profibus cable diagnostics should be
enabled. These diagnostics provide a useful tool for troubleshooting cable connection
problems.
Profibus Communications
97
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
12.3.1
LIQ_MAN_51-56
Profibus Faults and Warnings
Faults
The 56 displays the following hardware faults associated with problems related to the
Profibus board and its connection to the main board:
•
Profibus board mismatch - The revision of the installed Profibus board is not compatible
with the main board of the transmitter. Note the revisions of the Profibus board and the
main board of the transmitter and contact the factory.
•
Self-test failed - At power up, the Profibus board conducts several internal self tests. At
least one of them failed.
1. Try to reset the analyzer by cycling the power.
2. Replace the Profibus board
•
Internal communication failed - The main board detects the Profibus board, but the
Profibus board is not communicating with the main board.
1. Try to reset the analyzer by cycling the power.
2. Check the interconnecting cable between the Profibus and main board.
3. Replace the Profibus board.
Warnings
Warnings related to cable problems and software compatibility are also displayed, if Profibus
cable diagnostics are enabled:
•
Cable shorted or open - No Profibus cable is connected, or the A and B leads of the cable
are shorted.
•
Check Profibus Cable - The A and B leads in the Profibus cable may be reversed. If standard Profibus cable is used, the green lead should be wired to position A and the red lead
to position B. The bus may not have a terminator, or it may not be turned on. Check to
see that a terminator is present and turned on.
•
Profibus function limited - The Profibus board has new features that the main board
software does not support. If this is the case, the minor revision number of the Profibus
board will be greater than the minor revision number of the main board. Replace the
main board.
Details on these faults and warnings are accessed like any other fault or warning in the 56:
Press the INFO key and then select “Faults – fix now” or “Warnings – fix soon.”
LED Diagnostics
The Profibus board itself displays diagnostic information via an LED, which is located on the
upper right side of an installed board:
98
•
Slow blink – No cable is present or the connection is shorted
•
Slow, even blinking – reversed A and B cable connection
•
Quick flashing – Profibus message detected
•
1 second blinking – self-test failure
•
Off – no power or a defective or missing ribbon cable
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Diagnostic Information
The 56 provides diagnostic information on Profibus communications, which is accessed by
pressing the INFO key and then selecting “Profibus information” to launch the window below:
•
Address – displays the address being used by the 56 on the network
•
Baud rate – the baud rate of the 56 and the network
•
Master detected – If the master is not detected, the host and analyzer are not communicating. Check wiring connections, make sure the addresses match, and try changing the
baud rate.
•
POST, Lpf output, Rx character count – POST is the power-on self test. POST and the
remaining parameters on this window are used as factory diagnostics only.
Pressing the NEXT button brings up the final diagnostic parameter, the software version used
by the Profibus board:
Profibus Communications
99
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
12.4
LIQ_MAN_51-56
Data Transmission
General
The 56 uses RS-485 communication technology for transmission. When messages are
transmitted on Profibus networks, for word transfer (more than one byte), the high byte is
transferred first, followed by the low byte (Big-Endian/Motorola format).
56 Modules
The details of measurement modules and relay modules are shown below in the following tables:
12.4.1
Measurement Modules
Measurement modules provide the PV value for each signal board, the PV unit (Table 12-3),
and the PV type (Table 12-4)
Table 12-3. Measurement Modules
Module
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
100
Module Name
Signal Input Board 1
PV1 Value
PV1 Unit
PV1 Type
Temperature1
Temperature1 Unit
Signal Input Board 2
PV2 Value
PV2 Unit
PV2 Type
Temperature 2
Temperature 2 Unit
Dual Conductivity Input Boards
Conductivity Ratio
% Passage
% Rejection
Dual Flow Input Boards
Flow Recovery
Flow Difference
Flow Unit
Data
Type
Size
Bytes
Identifiers
float
uint
uchar
float
uint
4
2
1
4
2
0x41,0x83,0x01
0x41,0x81,0x02
0x41,0x80,0x03
0x41,0x83,0x04
0x41,0x81,0x05
float
uint
uchar
float
uint
4
2
1
4
2
0x41,0x83,0x06
0x41,0x81,0x07
0x41,0x80,0x08
0x41,0x83,0x09
0x41,0x81,0x0A
float
float
float
4
4
4
0x41,0x83,0x0B
0x41,0x83,0x0C
0x41,0x83,0x0D
float
float
uint
4
4
2
0x41,0x83,0x0E
0x41,0x83,0x0F
0x41,0x81,0x10
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Table 12-4. Measurement Unit Codes and Units
Unit
Code Unit
Unit
Code Unit
Unit
Code Unit
1001
1002
1034
1038
1043
1048
1059
1060
1061
1067
1105
1133
1137
1138
1140
1155
1157
1190
1211
1212
1213
1243
1274
1281
1282
1283
1284
1289
1290
1295
1302
1342
1349
1352
1353
1357
1358
1363
1364
1423
1424
1425
1521
1522
1523
1524
1525
1526
1527
1528
1551
1552
1555
1556
1558
1559
1560
°C
°F
m3
L
ft3
Gallon
h
d
m/sec
ft/sec
g/L
kPa
bar
mbar
atm
in Hg
mm Hg
kwatt
mA
µA
nA
mV
pH
Ohm
Gohm
Mohm
Kohm
mS
µS
Ohm-cm
mS/cm
%
m3/hour
L/min
L/hour
cu ft/min
cu ft/hour
GPM
GPH
ppm
ppb
ppth
ppb-ppm
ug/L-mg/L
None
% Oxygen in Gas
ppm Oxygen in Gas
NTU
FTU
FNU
S/cm
µS/cm
M Ohm-cm
k Ohm-cm
mg/L
µg/L
% Saturation
Table 12-5. Measurement (PV 1 and PV 2) Types
Types
0
1
2
3
4
5
6
7
8
9
10
11
12
13
Profibus Communications
Measurement
Unknown
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
Conductivity
Resistivity
NaOH
HCl
Low Range H2SO4
High Range H2SO4
NaCl
Types
14
15
16
17
18
19
20
21
22
23
24
25
26
Measurement
Custom Concentration
Total Dissolve Solids
Salinity
Oxygen
Ozone
Free Chlorine
pH Independent Free Chlorine
Total Chlorine
Chloramine
Flow
mA Input
Turbidity
Total Suspended Solids
101
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
12.4.2
LIQ_MAN_51-56
Relay Modules
Relay Modules provide the alarm type, state, logic, assignment, and simulation status as
shown in Tables 12-5 to 12-10:
Table 12-6. Measurement Modules
Module
Module Name
Relay 1
Alarm1 Type
Alarm1 State
Alarm1 Logic
Alarm1 Assignment
Alarm1 Simulate Status
Relay 2
Alarm2 Type
Alarm2 State
Alarm2 Logic
Alarm2 Assignment
Alarm2 Simulate Status
Relay 3
Alarm3 Type
Alarm3 State
Alarm3 Logic
Alarm3 Assignment
Alarm3 Simulate Status
Relay 4
Alarm4 Type
Alarm4 State
Alarm4 Logic
Alarm4 Assignment
Alarm4 Simulate Status
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
Data
Type
Size
Bytes
Identifiers
uchar
uchar
uchar
uchar
uchar
1
1
1
1
1
0x41,0x80,0x11
0x41,0x80,0x12
0x41,0x80,0x13
0x41,0x80,0x14
0x41,0x80,0x15
uchar
uchar
uchar
uchar
uchar
1
1
1
1
1
0x41,0x80,0x16
0x41,0x80,0x17
0x41,0x80,0x18
0x41,0x80,0x19
0x41,0x80,0x1A
uchar
uchar
uchar
uchar
uchar
1
1
1
1
1
0x41,0x80,0x1B
0x41,0x80,0x1C
0x41,0x80,0x1D
0x41,0x80,0x1E
0x41,0x80,0x1F
uchar
uchar
uchar
uchar
uchar
1
1
1
1
1
0x41,0x80,0x20
0x41,0x80,0x21
0x41,0x80,0x22
0x41,0x80,0x23
0x41,0x80,0x24
Table 12-7. Alarm Types
Alarm Type
Code
0
1
2
3
4
5
6
7
8
9
102
Alarm
Type
Set Point Alarm
Interval Timer
TPC (Duty Cycle)
Bleed Alarm
Feed Alarm
Setpoint Alarm + Delay Timer
Date and Time Timer
Totalizer Timer
Fault Alarm
No Alarm Assigned to Relay
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Table 12-8. Alarm States
0
1
Off
On
Table 12-10. Alarm Simulation
Table 12-9. Alarm Logic
Code
Code
Logic
State
0
High Alarm
0
Off
1
Low Alarm
1
De-Energize
2
USP Alarm (Contacting
Conductivity Only)
2
Energize
Table 12-11. Alarm Assignment Codes
Sensor 1
1
2
4
7
9
11
14
15
16
18
19
33
Sensor 2
41
42
44
47
49
51
54
55
56
58
59
73
Variable Assigned
Process Variable
Temperature
Amperometric Sensor Current
Millivolt Input
Glass electrode Impedance
Reference Electrode Impedance
Raw Conductivity
Raw Resistivity
Turbidity
Flow Velocity
Totalized Flow
Free Ions
Dual Sensor and Main Board Assignments:
0
201
202
203
204
205
206
207
208
209
Profibus Communications
Main Board Temperature
Conductivity Ratio
% Rejection (Conductivity)
% Passage (Conductivity)
Calculated pH (Conductivity)
Flow Ratio
Percent Flow Ratio
Flow Recovery
Flow Difference
Totalized Flow Difference
103
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
12.4.3
LIQ_MAN_51-56
Transmitter Diagnostic Bits
The 56 transmits diagnostic bits associated with main board and signal board faults and
warnings, as well as faults and warning specific to the sensors used to make the various measurements. These bits are shown organized by main board, signal board and measurement,
with and explanation of the fault or warning in Tables 12-11 to 12.13.
Note: The bit position shown is the bit position after the header byte.
Table 12-12. Main Board Diagnostic Bits – Main Board Faults
Main Board Faults
104
Bit
GSD Descriptor
Help
16
Change In
Diagnostic Data
A diagnostic bit has changed state
24
Main Board
CPU Error
The main board software is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Call the factory. The main board must be replaced (at the factory).
25
Main Board
User Data Error
The main board user EEPROM data is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Reset the analyzer to factory default settings. Go to main menu
and select Reset. Re-enter user settings and repeat calibration.
26
Main Board
Factory Data
Error
The main board factory EEPROM data is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Call the factory. The main board must be replaced (at the factory).
27
Main Board
Wrong Power
Supply
The existing power supply is incorrect for the turbidity sensor.
Replace power supply.
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Table 12-13. Main Board Diagnostic Bits – Main Board Warnings
Main Board Warnings
Bit
GSD Descriptor
Help
33
Unknown
Power Supply
The main board software cannot identify the power supply.
1. Check the ribbon cable connections between the power
supply board and main board.
2. Check the ribbon cable for damage.
3. Consult the factory.
34
Line Frequency Unknown
The analyzer has detected an AC power source, but the
frequency is not 50 Hz or 60 Hz.
1. Check the ribbon cable connections between the power
supply board and main board.
2. Check the mains frequency.
35
Main Board
Factory Cal
The analyzer was not calibrated at the factory.
1. Try to reset the analyzer by turning the power off then on
again.
2. Call the factory.
36
Keypad Error
The analyzer has detected shorted contacts in at least one key. Call
the factory.
37
Overheating
LCD Warning
High ambient temperature and high temperature inside the
analyzer is causing the LCD to overheat, reducing the operating
life of the display. The analyzer automatically begins reducing
the brightness of the display at 55°C (131°F). At 78°C (172°F)
analyzer turns off the display and stops storing data in the flash
drive (if one is installed). At 85°C (185°F) the analyzer turns off
the analog current outputs. At 90°C (194°F) the analyzer turns
off the relays. As the temperature cools, the functions are
restored. To prevent shutdown of the display and loss of other
analyzer functions, move the analyzer to a cooler place.
38
Excess Output
Current
One or more analog outputs are drawing too much current, and
all outputs have been shut down. The analyzer will check the current every 30 seconds and will return the outputs to normal operation once the current drops below the shutdown limit.
Check analog output wiring. A short from the positive terminal
to ground will cause excessive current to flow. Even if only one
output is shorted, all four outputs will be shut down.
If no wiring problems are found, the main board has been damaged and must be replaced.
Profibus Communications
105
Section 12: Profibus Communications
May 2014
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Table 12-14. Signal Board Diagnostic Bits
Diagnostic Bits
GSD Descriptor
Help
40 / 72
SN 1/2 Board
Unknown (Fault)
The analyzer automatically recognizes the sensor board on power up. Sensor
board unknown implies the sensor board is new and not supported or there is a
bad connection between the sensor board and the main board.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect ribbon cable for damage and replace if necessary.
3. Replace sensor board.
41 / 73
SN 1/2 HW SW
Mismatch error
(Fault)
The main board software does not match the sensor board software.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect ribbon cable for damage and replace if necessary.
3. Replace sensor board.
42 / 74
SN 1/2
Incompatible
Error (Fault)
The sensor board software is not supported by the main board. Either the sensor
board software is too old or the main board software is too old. Replace the main
board with one compatible with the sensor board. Call the factory for assistance.
The main board must be replaced at the factory.
43 / 75
SN 1/2 Not
Detected
The sensor board is no longer being detected.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect the ribbon cable for damage and replace if necessary.
3. Replace the sensor board.
44 / 76
SN 1/2 Not
Communicating
(Fault)
The analyzer automatically recognizes the sensor board on power up. Sensor not
communicating implies there is a bad connection between the sensor board and
the main board or there is a bad component on the sensor board.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect the ribbon cable for damage and replace if necessary.
3. Replace the sensor board.
45 / 77
SN 1/2 CPU error
(Fault)
The sensor board software is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. If cycling the power does not help, replace the sensor board.
46 / 78
SN 1/2 Factory
Data Error (Fault)
The factory-entered EEPROM data on the sensor board is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
47 / 79
SN 1/2 ADC Error
(Fault)
There is a bad hardware component on the sensor board. Replace the sensor
board.
50 / 82
SN 1/2 Hardware
Error (Fault)
Hardware error means there is a missing or bad hardware component on the sensor board. Replace the sensor board.
54 / 86
SN 1/2 User Data
Error (Fault)
The user-entered EEPROM data on the sensor board is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
55 / 87
SN 1/2 EEPROM
Write Error (Fault)
There is a bad CPU on the sensor board.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
SN 1/2 Need
Factory Cal
(Warning)
The sensor board was not calibrated at the factory. Call the factory.
Signal
Board 1 / 2
104 / 136
106
Profibus Communications
56 Analyzer Instruction Manual
LIQ_MAN_51-56
Section 12: Profibus Communications
May 2014
Table 12-14. Signal Board Diagnostic Bits – continued
Temperature Faults and Warnings
Signal
Board 1 / 2
GSD
Descriptor
Help
48 / 80
SN 1/2 RTD
Open (Fault)
The temperature measuring circuit is open. The RTD is improperly wired or the RTD IN or
RTD RTN are broken.
1. Check RTD wiring connections.
2. Check RTD for open or short circuits.
49 / 81
SN 1/2 RTD
Out Of Range
(Fault)
The resistance of the temperature sensing element (RTD) is too large or too small. The
sensor is wired improperly, or the temperature sensor is not recognized by the sensor board.
1. Check sensor wiring.
2. Check that the temperature sensing element is compatible with the analyzer. It must be
a Pt 100 or Pt 1000 RTD or a 22k NTC thermistor.
105 /
137
SN 1/2
Temperature
High
(Warning)
pH and Amperometric Measurements
The measured temperature is greater than 155 C.
1. Check wiring.
2. Disconnect the sensor and check the resistance between the RTD in and RTD return
wires. It should be about 110 ohm at room temperature.
3. Replace the sensor if the RTD has failed.
Contacting Conductivity Measurements
The measured temperature is greater than 300 C.
1. Check wiring.
2. Disconnect sensor and check resistance between RTD in and RTD return wires. It should
be about 1100 ohm at room temperature.
3. Replace sensor if RTD has failed.
Toroidal Conductivity Measurements
The measured temperature is greater than 200 C.
1. Check wiring.
2. Disconnect sensor and check resistance between RTD in and RTD return wires. It should
be about 110 ohm at room temperature.
3. Replace sensor if RTD has failed.
106 /
138
SN 1/2
Temperature
Low
(Warning)
pH and Amperometric Measurements
The measured temperature is less than -20 C.
1. Check wiring.
2. Disconnect the sensor and check the resistance between the RTD in and RTD return
wires. It should be about 110 ohm at room temperature.
3. Replace the sensor if the RTD has failed.
Contacting and Toroidal Conductivity Measurements
The measured temperature is less than -40 C.
1. Check wiring.
2. Disconnect the sensor and check the resistance between the RTD in and RTD return
wires. It should be about 110 ohm at room temperature.
3. Replace the sensor if the RTD has failed.
107 /
139
SN 1/2 RTD
Sense Line
Open
(Warning)
Profibus Communications
The analyzer uses the sense line to correct for the resistance of the RTD leads and for
changes in the lead wire resistance caused by variations in ambient temperature.
1. Check wiring.
2. Disconnect the sensor and check for continuity between the RTD return and RTD sense
leads at the sensor.
3. If sensor has failed, use a wire jumper to connect the RTD sense and return terminals.
4. Replace the sensor.
107
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Table 12-14. Signal Board Diagnostic Bits – continued
pH / ISE Faults and Warnings
Signal
Board 1 / 2
GSD Descriptor
Help
51 / 83
SN 1/2 Ref Z Too
High (Fault)
The impedance of the reference junction is too high.
1. Confirm that the sensor is fully immersed in the process liquid.
2. High reference impedance implies the reference junction is fouled or depleted.
Replace the sensor.
3. Some sensors are designed to have high reference impedance. If this is the case,
change the reference impedance setting (in the measurement menu) to high.
52 / 84
SN 1/2 Glass Z Too
High (Fault)
The sensing element in a pH sensor is a thin glass membrane. Normally, the impedance of the membrane is 80 – 100 megohm. As the glass ages, impedance increases. A large increase in impedance implies the sensor is near the end of its
useful life. Certain ion selective electrodes (fluoride, ammonia, and ammonium)
also have high impedance sensing elements.
1. Confirm that the sensor is fully immersed in the process liquid.
2. If sensor performance in buffers and the process is acceptable, increase the impedance limit.
3. Replace the sensor.
53 / 85
SN 1/2 Broken
Glass Fault (Fault)
The sensing element in a pH sensor is a thin glass membrane. In a good sensor,
the glass impedance is very high. If the glass bulb breaks or cracks, the impedance
drops to a very low value, and the broken glass fault appears. Certain ion selective
electrodes (fluoride, ammonia, and ammonium) also have high impedance sensing elements.
1. Inspect the sensor for obvious damage.
2. Check the sensor in buffers. If the sensor does not respond to pH changes, the
glass is broken.
3. Replace the sensor.
108 / 140
SN 1/2 Out Of
Range pH Brd
(Warning)
The pH input is outside the range -750 mV to +750 mV or the ORP or ISE input is
outside the range -1500 mV to +1500 mV.
Check sensor wiring.
Replace sensor.
111 / 143
SN 1/2 No
Solution Gnd
(Warning)
The sensor is not properly wired. Check wiring.
115 / 147
SN 1/2 Broken
Glass Disabled
(Warning)
The impedance of the sensing glass in a pH sensor is a strong function of temperature. For the impedance to be a useful indicator that the sensing glass is cracked
or broken, it must be corrected for temperature. If the process temperature exceeds the correction cut-off temperature, the broken glass fault is automatically
disabled. The analyzer recalculates the cut-off temperature every time the sensor
is calibrated. The cutoff temperature is about 115°C.
108
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
Table 12-14. Signal Board Diagnostic Bits – continued
Conductivity Faults and Warnings
Signal
Board 1 / 2
GSD Descriptor
Help
109 / 141
SN 1/2 Out Of
Range Cont/Tor
Brd
Contacting Conductivity Measurements
The raw conductance is greater than 1 S.
1. Check sensor wiring.
2. Check sensor cleanliness.
3. Check the sensor electrodes for shorts.
Toroidal Conductivity Measurements
The raw conductance is greater than 2 S.
1. Check sensor wiring.
2. Check sensor electrodes for shorts.
112 / 144
SN 1/2 -ve Rdng
Cont/Tor Brd
(Warning)
(Negative
conductivity
reading)
Contacting Conductivity Measurements
Negative Reading:
During zero calibration, the analyzer makes a correction for the conductance equivalent of the cable capacitance. If the difference between the measured conductance and the zero correction is less than -2 uS, the negative reading warning
appears.
1. Repeat zero calibration. Be sure the sensor is completely dry.
2. Confirm that the sensor is completely submerged in the process liquid.
Toroidal Conductivity Measurements
Negative Reading:
During zero calibration, the analyzer makes a correction for small offsets in the signal that exist even when the conductance is zero. If the difference between the
measured conductance and the zero correction is less than -50 uS, the negative
reading warning appears.
1. Repeat zero calibration. Be sure the sensor is completely dry.
2. Confirm that the sensor is completely submerged in the process liquid.
113 / 145
SN 1/2 % Out Of
Range
(Warning)
The calculated concentration or temperature is out of range.
1. NaOH: 0 – 12.00%, 0 - 100C.
2. HCl: 0 – 15.00%, 0 - 65C.
3. H2SO4 (low): 0 – 25.00%, -1 - 99C.
4. H2SO4 (high): 96.0 – 99.7%, 4 - 110C.
5. NaCl: 0 – 20.00%, 0 - 100C.
6. Salinity: 2 – 42 part per thousand, -2 - 35C.
Amperometric Faults and Warnings
Signal
Board 1 / 2
GSD Descriptor
Help
110 / 142
SN 1/2 -ve Rdng
Amp/Oxy Brd
(Warning)
(Negative
amperometric
reading)
The analyzer subtracts the zero current (the current generated by the sensor when
no analyte is present) from the measured current before converting the result into
a concentration value. A negative reading implies the zero step was done before the
sensor reached a stable zero current. Repeat the zero calibration. It may require
several hours or as long as overnight for the zero current to stabilize.
114 / 145
SN 1/2 Pressure
Sensor Broken
(Warning)
The pressure sensor on the oxygen sensor board has failed, or it is not installed, or it
is installed backwards. Replace the sensor board.
Profibus Communications
109
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Table 12-14. Signal Board Diagnostic Bits – continued
Flow / 4 to 20 mA Input Faults and Warnings
Signal
Board 1 / 2
GSD Descriptor
Help
117 / 149
SN 1/2 No Flow
Detected
(Warning)
The analyzer has sensed no flow for at least ten seconds.
1. Check that the flow exceeds the minimum flow recommended for the sensor.
2. Check wiring.
3. Check the sensor
118 / 150
SN 1/2 -ve Current
(Warning)
(Negative 4-20
mA input current)
The input current from the device is less than -0.1 mA.
1. The polarity is wrong. Reverse wiring.
2. Check the current source.
119 / 151
SN 1/2 Execessive
Current
(Warning)
The input current from the device is greater than 22.00 mA.
1. Check wiring.
2. Check the current source.
120 / 152
SN 1/2 Low
Current
(Warning)
The input current range has been set to 4-20 mA, and the input current is less
than 3.6 mA.
1. Check the input current range configuration.
2. Check configuration of current source.
Turbidity Faults and Warnings
Signal
Board 1 / 2
GSD Descriptor
Help
27
Main Board
Wrong Power
Supply (Fault)
The existing power supply is incorrect a the turbidity sensor. Replace power supply.
56 / 88
SN 1/2 Lamp Led
Fail (Fault)
The lamp or LED has failed.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor.
57 / 89
SN 1/2
Disconnected
(Fault)
The turbidity sensor cable is disconnected at either the sensor or the sensor board.
1. Confirm that the cable is tightly connected to the sensor receptacle. If necessary,
completely disconnect the cable and reconnect it. Check connectors for cleanliness.
2. Confirm that the cable is tightly connected to the board receptacle. If necessary,
completely disconnect the cable and reconnect it. Check connectors for cleanliness.
116 / 148
SN 1/2 Weak
Lamp (Warning)
The lamp intensity has decreased below the lower limit. The lamp is aging and
should be replaced as soon as possible.
121 / 153
SN 1/2 Needs
calibration
(Warning)
The analyzer continuously measures the intensity of the lamp and uses the measured intensity to correct for the effect of lamp drift on measured turbidity. As the
lamp ages and its intensity drops, the target intensity is adjusted. Readjusting the
target intensity introduces a significant error. To remove the error, the sensor must
be recalibrated.
1. Recalibrate the sensor in 20 NTU standard.
2. Order replacement sensor.
110
Profibus Communications
56 Analyzer Instruction Manual
Section 12: Profibus Communications
LIQ_MAN_51-56
May 2014
12.5
Installation and Wiring
12.5.1
General
This section shows the recommended method of wiring the Profibus cable to the Profibus
card of the 56.
Recommended Wiring Steps
1. Feed the Profibus cable through a cable gland fitting. Install the cable gland fitting into
the enclosure opening on the leftmost side of the enclosure nearest the front of the door
hinge. See photo. Note that the power cable should be installed in the leftmost side of
the enclosure farthest from the front of the door hinge, as shown in the photo below:
FIGURE 12-1. Proper enclosure opening locations for cable installation.
2. Secure the Profibus cable with a cable gland fittings and nut from the outside of the enclosure to ensure a proper seal. Note that the cable gland fittings (PN 23554-00 - Cable
Gland Kit, Quantity 5) do not require the securing nut inside the enclosure to properly
install the fitting. Simply thread the fitting into the grounding plate inside the enclosure
and tighten.
Once the gland fitting is tightly screwed into the enclosure and internal grounding plate, only
the external nut needs to be tightened to properly seal the Profibus cable.
FIGURE 12-2, Cable gland fittings secured on internal grounding plate.
Profibus Communications
111
Section 12: Profibus Communications
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
3. Remove the 4-lead terminal block which is installed on the Profibus slide-in board. This
photo shows the 4-lead terminal block installed onto the board:
FIGURE 12-3. Profibus board showing
removable 4-lead terminal block
FIGURE 12-4. 4-Lead Terminal Block
4. Wire the two leads of the Profibus cable (red and green) to the A and B positions of the
4-lead removable terminal block. A #0 Philips head screw driver is required to open and
close the terminal posts on the terminal block. The lead positions are labeled on the
Profibus card (A and B) to assist in wiring.
Note: For successful communication using standard Profibus cable:
The Green color lead must be wired to lead position A (RxD/TxD-N).
The Red color lead must be wired to lead position B (RxD/TxD-P).
The VP and DGND (digital ground) positions are used to a power bus terminator if required.
CAUTION
Make sure the Profibus cable is properly prepared. Unshielded wire leads can lead to
poor signal integrity from the Profibus device to the Master
5. After wiring the Profibus cable to the terminal block, slide the wired 4-lead terminal
block onto the 4 pins protruding from Profibus
FIGURE 12-5. Complete Profibus
board on the left side.
board installation with ribbon
Photo shows the 4-lead terminal block properly
wired to the Profibus board and installed onto the cable attached.
board.
Note: the Profibus cable must be directed downwards. The Profibus board can be partially or fully removed to allow easy insertion of the 4-lead terminal
block onto the Profibus board.
6. Ensure that the 10-lead ribbon cable is properly
connected from the Profibus board to the leftmost connector of the 3 10-pin shrouded
connectors on the right side of the main circuit
board.
Once power is wired to the unit (as shown in
photo), the Profibus-configured 56 is ready for
power up and communication on a Profibus
network.
112
Profibus Communications
56 Analyzer Instruction Manual
Section 13: Maintenance
LIQ_MAN_51-56
May 2014
Section 13.0 Maintenance
13.1
Overview
This section gives general procedures for routine maintenance of the 56 advanced analyzer. The
analyzer needs almost no routine maintenance. Sensors require periodic inspection and
cleaning. The calibration of the transmitter-sensor combination should be checked regularly,
and the loop recalibrated if necessary.
CAUTION
Always depower the analyzer or disconnect the analyzer from the main power supply
before opening the enclosure in a hazardous area.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this
symbol appears, consult the manual for appropriate action.
13.2
Analyzer Maintenance
Periodically clean the analyzer window and housing as needed with a cloth dampened with
water. Do not use abrasives or cleaning solutions.
13.3
USB Port
The USB communications port is protected by a NEMA-rated seal and cover. Do not remove
the cover during cleaning. Never remove the USB port cover when the instrument is operated in a hazardous rated area.
Maintenance
113
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
This page left blank intentionally
114
56 Analyzer Instruction Manual
Section 14: Return of Material
LIQ_MAN_51-56
May 2014
Section 14.0 Return of Material
14.1
General
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 1-949-7578500 for a Return Materials Authorization (RMA) number.
14.2
Warranty Repair
The following is the procedure for returning instruments still under warranty:
1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order
number. In the case of individual parts or sub-assemblies, the serial number
on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal”
(see Warranty). If possible, pack the materials in the same manner as they
were received.
4. Send the package prepaid to:
Rosemount Analytical
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ___________
IMPORTANT
Please see second section of “Return of Materials Request” form. Compliance with the OSHA
requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
14.3
Non-Warranty Repair.
The following is the procedure for returning for repair instruments that are no longer
under warranty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone
number of the individual to be contacted should additional information be needed.
3. Do Steps 3 and 4, above.
NOTE
Consult the factory for additional information regarding service or repair.
Return of Materials
115
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
This page left blank intentionally
116
56 Analyzer Instruction Manual
Appendix 1: HART and Device Variables
LIQ_MAN_51-56
May 2014
Appendix 1: HART and Device Variables
pH Device Variables
Assignable to Dynamic
Variables
Variable Range
pH (1)
PV, SV, TV or QV
0 to 14 pH
ORP (2), Redox (3)
PV, SV, TV or QV
-1500 to 1500 mV
Ammonia (4), Fluoride (5), Custom ISE (6)
PV, SV, TV or QV
0 to 1000 ppm
Device Variable Name
Sensor 1 / 2 Measurement Type
0 to 1000 ppb
0 to 1000 mg/L
0 to 1000 µg/L
0 to 100 %
Measurement Parameters
Sensor 1 / 2 Temperature
PV, SV, TV or QV
-15 to 200 °C
Sensor 1 / 2 Temperature resistance
TV or QV
0 to 100000 ohm
Sensor 1 / 2 mV input
PV, SV, TV or QV
-750 to 750 mV
Sensor 1 / 2 Glass impedance
PV, SV, TV or QV
0 to 2000 M
Sensor 1 / 2 Sensor impedance
PV, SV, TV or QV
0 to 2000 M
Sensor 1 / 2 Reference impedance
PV, SV, TV or QV
0 to 10000 k
5 to 360 °F
Appendix 1
117
Appendix 1: HART and Device Variables
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Contacting and Toroidal Conductivity Device Variables
Device Variable Name
Assignable to Dynamic Variables
Variable Range
Conductivity (7)
PV, SV, TV or QV
0 to 2000000 µS/cm
Resistivity (8)
PV, SV, TV or QV
0 to 50000000 ohm-cm
NaOH (9)
PV, SV, TV or QV
0 to 12 %
HCl (10)
PV, SV, TV or QV
0 to 15 %
Low H2SO4 (11)
PV, SV, TV or QV
0 to 25 %
High H2SO4 (12)
PV, SV, TV or QV
96 to 99.7 %
NaCl (13)
PV, SV, TV or QV
0 to 25 %
Custom Concentration (14)
PV, SV, TV or QV
0 to 1000 ppm
Sensor 1 / 2 Measurement Type
% Concentration:
0 to 1000 mg/L
0 to 100 g/L
0 to 100 %
0 to 1000 None
TDS (15)
PV, SV, TV or QV
0 to 10000 ppm
PV, SV, TV or QV
-15 to 200 °C
Measurement Parameters
Sensor 1 / 2 Temperature
5 to 360 °F
Sensor 1 / 2 Temperature resistance
TV or QV
0 to 100000 ohm
Sensor 1 / 2 Conductance
TV or QV
0 to 2000000 µS
Sensor 1 / 2 Input resistance
TV or QV
0 to 500 k
Sensor 1 / 2 Raw Conductivity
PV, SV, TV or QV
0 to 2000000 µS/cm
Sensor 1 / 2 Raw Resistivity
PV, SV, TV or QV
0 to 50000000 ohm-cm
Conductivity Ratio
PV, SV, TV or QV
0 to 10000
% Rejection
PV, SV, TV or QV
0 to 100 %
% Passage
PV, SV, TV or QV
0 to 100 %
Calculated pH (Contacting only)
PV, SV, TV or QV
0 to 14 pH
Dual Conductivity Measurments:
Turbidity Device Variables
Device Variable Name
Assignable to Dynamic Variables
Variable Range
PV, SV, TV or QV
0 to 200 FNU
Sensor 1 / 2 Measurement Type
Turbidity (25)
0 to 200 FTU
0 to 200 NTU
TSS (26)
PV, SV, TV or QV
0 to 1000 ppm
0 to 1000 mg/L
0 to 1000 None
Measurement Parameters
Sensor 1 / 2 Lamp Voltage
118
TV or QV
0 to 2000 mV
Appendix 1
56 Analyzer Instruction Manual
Appendix 1: HART and Device Variables
LIQ_MAN_51-56
May 2014
Amperometric Device Variables
Device Variable Name
Assignable to Dynamic Variables
Variable Range
Salinity (16)
PV, SV, TV or QV
0 to 36 ppth
Oxygen (17)
PV, SV, TV or QV
0 to 100 ppm
Sensor 1 / 2 Measurement Type
0 to 1000 ppb
0 to 100 mg/L
0 to 1000 µg/L
0 to 300 % Saturation
0 to 760 mmHg
0 to 30 inHg
0 to 1 bar
0 to 1000 mbar
0 to 100 kPa
0 to 1 atm
Ozone (18), Free Chlorine (19), pH Independent Free Chlorine (20), Total Chlorine (21),
PV, SV, TV or QV
Chloramine (22)
0 to 20 ppm
0 to 1000 ppb
0 to 20 mg/L
0 to 1000 µg/L
Measurement Parameters
Sensor 1 / 2 Temperature
PV, SV, TV or QV
-15 to 200 °C
Sensor 1 / 2 Temperature resistance
TV or QV
0 to 100000
Sensor 1 / 2 Sensor input current
PV, SV, TV or QV
0 to 100000 nA
Sensor 1 / 2 Polarizing voltage
TV or QV
-1000 to 1000 mV
Sensor 1 / 2 Pressure
TV or QV
0 to 5000 mmHg
5 to 360 °F
0 to 200 inHg
0 to 7 bar
0 to 7000 mbar
0 to 700 kPa
0 to 7 atm
Flow / mA Input Device Variables
Device Variable Name
Assignable to Dynamic Variables
Variable Range
Sensor 1 / 2 Measurement Type
Flow Rate (23)
PV, SV, TV or QV
0 to 100000 gal/min
0 to 100000 gal/hr
0 to 100000 cu.ft/min
0 to 100000 cu.ft/hr
0 to 100000 liter/min
0 to 100000 liter/hr
0 to 100000 m3/hr
Appendix 1
119
Appendix 1: HART and Device Variables
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Flow / mA Input Device Variables
Device Variable Name
(continued)
Assignable to Dynamic Variables
Variable Range
Sensor 1 / 2 Measurement Type
Scaled mA Input (24)
PV, SV, TV or QV
As per Configuration
Input Type:
Units:
Temperature
°C, °F
Pressure
mmHg, inHg, bar, mbar, kPa, atm
Flow
gal/min, gal/hr, cu.ft/hr, cu.ft/min,
liter/hr, m3/hr
Other
ft/sec, m/sec, None
pH/ORP
pH, mV
Conductivity
µS/cm, mS/cm
Resistivity
M -cm, K -cm
Concentration
%, parts per thousand, ppm, ppb
Weight per Volume
mg/L, µg/L, g/L
Turbidity
FNU, FTU, NTU
Sensor 1 / 2 Frequency
TV or QV
0 to 1000 Hz
Sensor 1 / 2 Velocity
PV, SV, TV or QV
0 to 10 ft/sec
Measurement Parameters
0 to 10 m/sec
Sensor 1 / 2 Totalizer
PV, SV, TV or QV
0 to 1000000 gal
0 to 1000000 liter
0 to 1000000 m3
0 to 1000000 cu.ft
Sensor 1 / 2 mA input
TV or QV
3.6 to 22 mA
Flow ratio
PV, SV, TV or QV
0 to 1000 NA
% Flow ratio
PV, SV, TV or QV
0 to 100 %
% Recovery
PV, SV, TV or QV
0 to 100 % Recovery
Flow difference
PV, SV, TV or QV
-10000 to 10000 gal/min
Dual Flow Measurements
-10000 to 10000 gal/hr
-10000 to 10000 cu.ft/min
-10000 to 10000 cu.ft/hr
-10000 to 10000 liter/min
-10000 to 10000 liter/hr
-10000 to 10000 m3/hr
Flow total difference
PV, SV, TV or QV
-100000 to 100000 gal
-100000 to 100000 liter
-100000 to 100000 m3
-100000 to 100000 cu.ft
120
Appendix 1
56 Analyzer Instruction Manual
Appendix 2: HART Status Bits
LIQ_MAN_51-56
May 2014
Appendix 2: HART Status Bits
Common Transmitter Status Bits
Byte
Bit
2
2
2
Severity
Device Status Bits Set
Sensor 1 Hardware Error
Error
Device Status: bit 7, 3
6
Sensor 1 User Data Error
Error
Device Status: bit 7, 3
2
7
Sensor 1 EEPROM Write Error
Error
Device Status: bit 7, 3
3
0
Sensor 1 Sensor Board Unknown
Error
Device Status: bit 7, 3
3
1
Sensor 1 Hardware/Software Mismatch
Error
Device Status: bit 7, 3
3
2
Sensor 1 Sensor Incompatible
Error
Device Status: bit 7, 3
3
3
Sensor 1 Sensor Not Detected
Error
Device Status: bit 7, 3
3
4
Sensor 1 Sensor not communicating
Error
Device Status: bit 7, 3
3
5
Sensor 1 CPU Error
Error
Device Status: bit 7, 3
3
6
Sensor 1 Factory Data Error
Error
Device Status: bit 7, 3
3
7
Sensor 1 ADC Error
Error
Device Status: bit 7, 3
5
2
Sensor 2 Hardware Error
Error
Device Status: bit 7, 3
5
6
Sensor 2 User Data Error
Error
Device Status: bit 7, 3
5
7
Sensor 2 EEPROM Write Error
Error
Device Status: bit 7, 3
6
0
Sensor 2 Sensor Board Unknown
Error
Device Status: bit 7, 3
6
1
Sensor 2 Hardware/Software Mismatch
Error
Device Status: bit 7, 3
6
2
Sensor 2 Sensor Incompatible
Error
Device Status: bit 7, 3
6
3
Sensor 2 Sensor Not Detected
Error
Device Status: bit 7, 3
6
4
Sensor 2 Sensor not communicating
Error
Device Status: bit 7, 3
6
5
Sensor 2 CPU Error
Error
Device Status: bit 7, 3
6
6
Sensor 2 Factory Data Error
Error
Device Status: bit 7, 3
6
7
Sensor 2 ADC Error
Error
Device Status: bit 7, 3
7
0
Maintenance required
Warning
Extended Device bit 0
7
1
Device variable alert
Warning
Extended Device bit 1
9
0
Simulation active
Mode
9
1
Non-volatile memory defect
Error
9
2
Volatile memory defect
Error
9
3
Watchdog reset executed
--
9
4
Voltage condition out of range
--
9
5
Environmental condition out of range
--
9
6
Electric defect
11
0
Analog channel-2 saturated
Warning
11
1
Analog channel-3 saturated
Warning
11
2
Analog channel-4 saturated
Warning
Appendix 2
Message
Error
121
Appendix 2: HART Status Bits
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Common Transmitter Status Bits
(continued)
Byte
Bit
Message
Severity
Device Status Bits Set
14
0
Analog channel-2 fixed
Mode
14
1
Analog channel-3 fixed
Mode
14
2
Analog channel-4 fixed
Mode
21
0
Main board CPU Error
Error
Device Status: bit 7, 3
21
1
User Data Error
Error
Device Status: bit 7, 3
21
2
Factory Data Error
Error
Device Status: bit 7, 3
22
1
Unknown power supply
Warning
Device Status: bit 4
22
2
Unknown line frequency
Warning
Device Status: bit 4
22
3
Factory calibration error
Warning
Device Status: bit 4
22
4
Keypad error
Warning
Device Status: bit 4
22
5
Overheating LCD
Warning
Device Status: bit 4
22
6
Excess output current
Warning
Device Status: bit 4
23
0
Relay-1 Energized
Mode
23
1
Relay-2 Energized
Mode
23
2
Relay-3 Energized
Mode
23
3
Relay-4 Energized
Mode
24
0
Sensor 1 Zeroing in Progress
Mode
Device Status: bit 4
24
1
Sensor 1 Calibration in Progress
Mode
Device Status: bit 4
24
2
Sensor 1 Standardization in Progress
Mode
Device Status: bit 4
24
3
Sensor 1 Stabilization in Progress
Mode
Device Status: bit 4
25
0
Sensor 2 Zeroing in Progress
Mode
Device Status: bit 4
25
1
Sensor 2 Calibration in Progress
Mode
Device Status: bit 4
25
2
Sensor 2 Standardization in Progress
Mode
Device Status: bit 4
25
3
Sensor 2 Stabilization in Progress
Mode
Device Status: bit 4
Temperature Status Bits
Byte
Bit
2
0
2
122
Message
Severity
Device Status Bits Set
Sensor 1 RTD Open
Error
Device Status: bit 7, 3
1
Sensor 1 RTD Out Of Range
Error
Device Status: bit 7, 3
5
0
Sensor 2 RTD Open
Error
Device Status: bit 7, 3
5
1
Sensor 2 RTD Out Of Range
Error
Device Status: bit 7, 3
17
1
Sensor 1 Temperature High
Warning
Device Status: bit 4
17
2
Sensor 1 Temperature Low
Warning
Device Status: bit 4
17
3
Sensor 1 RTD Sense Line Open
Warning
Device Status: bit 4
20
1
Sensor 2 Temperature High
Warning
Device Status: bit 4
20
2
Sensor 2 Temperature Low
Warning
Device Status: bit 4
20
3
Sensor 2 RTD Sense Line Open
Warning
Device Status: bit 4
Appendix 2
56 Analyzer Instruction Manual
Appendix 2: HART Status Bits
LIQ_MAN_51-56
May 2014
pH Board Status Bits
Byte
Bit
2
3
2
Message
Severity
Device Status Bits Set
Sensor 1 Reference Impedance Too High
Error
Device Status: bit 7, 3
4
Sensor 1 Glass Impedance Too High
Error
Device Status: bit 7, 3
2
5
Sensor 1 Broken Glass
Error
Device Status: bit 7, 3
5
3
Sensor 2 Reference Impedance Too High
Error
Device Status: bit 7, 3
5
4
Sensor 2 Glass Impedance Too High
Error
Device Status: bit 7, 3
5
5
Sensor 2 Broken Glass
Error
Device Status: bit 7, 3
16
3
Sensor 1 Broken Glass Disabled
Warning
Device Status: bit 4
17
7
Sensor 1 No Solution Ground
Warning
Device Status: bit 4
19
3
Sensor 2 Broken Glass Disabled
Warning
Device Status: bit 4
20
7
Sensor 2 No Solution Ground
Warning
Device Status: bit 4
24
6
Sensor 1 Offset Error
Mode
Device Status: bit 4
24
24
Extended Device Status: bit 0
7
Sensor 1 Slope Error
Mode
24
Device Status: bit 4
Extended Device Status: bit 0
25
6
Sensor 2 Offset Error
Mode
Device Status: bit 4
Extended Device Status: bit 0
25
7
Sensor 2 Slope Error
Mode
Device Status: bit 4
Extended Device Status: bit 0
Contacting and Toroidal Conductivity Status Bits
Byte
Bit
Severity
Device Status Bits Set
16
0
Sensor 1 Conductivity Negative Reading
Warning
Device Status: bit 4
1
Sensor 1 Percent Out Of range
Warning
Device Status: bit 4
17
5
Sensor 1 Conductivity Board Out Of Range
Warning
Device Status: bit 4
19
0
Sensor 2 Conductivity Negative Reading
Warning
Device Status: bit 4
1
Sensor 2 Percent Out Of range
Warning
Device Status: bit 4
20
5
Sensor 2 Conductivity Board Out Of Range
Warning
Device Status: bit 4
24
5
Sensor 1 Zero Error
Mode
Device Status: bit 4
Extended Device Status: bit 0
25
5
Sensor 2 Zero Error
Mode
Device Status: bit 4
Extended Device Status: bit 0
Appendix 2
Message
123
Appendix 2: HART Status Bits
56 Analyzer Instruction Manual
May 2014
LIQ_MAN_51-56
Amperometric Status Bits
Byte
Bit
16
2
17
Message
Severity
Device Status Bits Set
Sensor 1 Pressure Broken
Warning
Device Status: bit 4
6
Sensor 1 Amperometric Board Negative Reading
Warning
Device Status: bit 4
19
2
Sensor 2 Pressure Broken
Warning
Device Status: bit 4
20
6
Sensor 2 Amperometric Board Negative Reading
Warning
Device Status: bit 4
Flow / mA Input Status Bits
Byte
Bit
Severity
Device Status Bits Set
15
0
Sensor 1 Low Current
Message
Warning
Device Status: bit 4
16
5
Sensor 1 No Flow Detected
Warning
Device Status: bit 4
16
6
Sensor 1 Negative Current
Warning
Device Status: bit 4
16
7
Sensor 1 Excessive Current
Warning
Device Status: bit 4
18
0
Sensor 2 Low Current
Warning
Device Status: bit 4
19
5
Sensor 2 No Flow Detected
Warning
Device Status: bit 4
19
6
Sensor 2 Negative Current
Warning
Device Status: bit 4
19
7
Sensor 2 Excessive Current
Warning
Device Status: bit 4
Turbidity Input Status Bits
124
Byte
1
1
Bit
0
1
Message
Sensor 1 Lamp/LED Failed
Sensor 1 Disconnected
Severity
Error
Error
Device Status Bits Set
Device Status: bit 7, 3
Device Status: bit 7, 3
4
4
0
1
Sensor 2 Lamp/LED Failed
Sensor 2 Disconnected
Error
Error
Device Status: bit 7, 3
Device Status: bit 7, 3
15
1
Sensor 1 Need Calibration
Warning
Device Status: bit 4
18
1
Sensor 2 Need Calibration
Warning
Device Status: bit 4
21
3
Wrong Power Supply
Error
Device Status: bit 7, 3
Appendix 2
56 Analyzer Instruction Manual
LIQ_MAN_51-56
EC Declaration
EC Delaration
May 2014
125
LIQ_MAN_56
Rev. D
May 2014
facebook.com/EmersonRosemountAnalytical
8
AnalyticExpert.com
Credit Cards for U.S. Purchases Only.
twitter.com/RAIhome
youtube.com/user/RosemountAnalytical
Emerson Process Management
©2014 Rosemount Analytical, Inc. All rights reserved.
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark
of one of the Emerson Process Management family of companies. All other marks are the property
of their respective owners.
rosemountanalytical.com
© Rosemount Analytical Inc. 2014
The contents of this publication are presented for information purposes only, and while effort has
been made to ensure their accuracy, they are not to be construed as warranties or guarantees,
express or implied, regarding the products or services described herein or their use or applicability.
All sales are governed by our terms and conditions, which are available on request. We reserve the
right to modify or improve the designs or specifications of our products at any time without notice.