Yokogawa User's Manual Model DC402G Converter for Dual Cell Conductivity and Resistivity
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s
Manual
Model DC402G
Converter for Dual Cell
Conductivity and Resistivity
[Style: S2]
IM 12D08E02-01E
IM 12D08E02-01E
6th Edition
i
u Introduction
Thank you for purchasing the DC402G Converter for Dual Cell Conductivity and Resistivity.
Please read the following respective documents before installing and using the DC402G.
The related documents are as follows.
General Specifications:
User’s Manual:
GS 12D08E02-01E
IM 12D08E02-01E (this manual)
* the “E” in the document number is the language code.
NOTE
Electric discharge
The DC402G contains devices that can be damaged by electrostatic discharge. When servicing
this equipment, please observe proper procedures to prevent such damage. Replacement
components should be shipped in conductive packaging. Repair work should be done at
grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic
discharge.
Installation and wiring
The DC402G should only be used with equipment that meets the relevant IEC, American or
Canadian standards. Yokogawa accepts no responsibility for the misuse of this unit.
The Instrument is packed carefully with shock absorbing materials, nevertheless, the instrument
may be damaged or broken if subjected to strong shock, such as if the instrument is dropped.
Handle with care.
Although the instrument has a weatherproof construction, the transmitter can be harmed if it
becomes submerged in water or becomes excessively wet.
Do not use an abrasive or solvent in cleaning the instrument.
Yokogawa is not responsible for damage to the instrument, poor performance of the instrument
or losses resulting from such, if the problems are caused by:
• Improper operation by the user.
• Use of the instrument in improper applications
• Use of the instrument in an improper environment or improper utility program
• Repair or modification of the related instrument by an engineer not authorized by Yokogawa.
n How to dispose the batteries
This is an explanation about the new EU Battery Directive(DIRECTIVE 2006/66/EC).This
directive is only valid in the EU.
Batteries are included in this product.Batteries incorporated into this product cannot be removed
by yourself.Dispose them together with this product.
When you dispose this product in the EU,contact your local Yokogawa Europe B.V.office.Do not
dispose them as domestic household waste.
Battery type: silver oxide battery
Notice:
The symbol means they shall be sorted out and collected as ordained in
ANNEX II in DIRECTIVE 2006/66/EC.
n Product Disposal
The instrument should be disposed of in accordance with local and national legislation/
regulations.
Media No. IM 12D08E02-01E
6th Edition : Feb. 2016 (YK)
All Rights Reserved Copyright © 2007, Yokogawa Electric Corporation
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
ii
u
Safety Precautions
n Safety, Protection, and Modification of the Product
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specified in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modification of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order
to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes
what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Protective Ground Terminal.
This symbol indicates Function Ground Terminal. Do not use this terminal as the protective
ground terminal.
n Notes on Handling User’s Manuals
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written
consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you find mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
iii
n Warning and Disclaimer
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor
responsibility to any person or entity with respect to any direct or indirect loss or damage arising
from using the product or any defect of the product that YOKOGAWA can not predict in advance.
n Label
The next warning label is put on the product.
n Drawing Conventions
Some drawings may be partially emphasized, simplified, or omitted, for the convenience of
description.
Some screen images depicted in the user’s manual may have different display positions or
character types (e.g., the upper / lower case). Also note that some of the images contained in this
user’s manual are display examples.
n Warranty and service
Yokogawa products and parts are guaranteed free from defects in workmanship and material
under normal use and service for a period of (typically) 12 months from the date of shipment from
the manufacturer.
Individual sales organisations can deviate from the typical warranty period, and the conditions of
sale relating to the original purchase order should be consulted. Damage caused by wear and
tear, inadequate maintenance, corrosion, or by the effects of chemical processes are excluded
from this warranty coverage.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
iv
In the event of warranty claim, the defective goods should be sent (freight paid) to the service
department of the relevant sales organisation for repair or replacement (at Yokogawa discretion).
The following information must be included in the letter accompanying the returned goods:
•
•
•
•
•
•
•
Part number, model code and serial number
Original purchase order and date
Length of time in service and a description of the process
Description of the fault, and the circumstances of failure
Process/environmental conditions that may be related to the failure of the device.
A statement whether warranty or nonwarranty service is requested
Complete shipping and billing instructions for return of material, plus the name and phone
number of a contact person who can be reached for further information.
Returned goods that have been in contact with process fluids must be decontaminated/
disinfected before shipment. Goods should carry a certificate to this effect, for the health and
safety of our employees.
Material safety data sheets should also be included for all components of the processes to which
the equipment has been exposed.
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Toc-1
Model DC402G
Converter for Dual Cell
Conductivity and Resistivity [Style: S2]
IM 12D08E02-01E 6th Edition
TABLE OF CONTENTS
u Introduction.....................................................................................................i
u
Safety Precautions........................................................................................ii
1.
Introduction and General Description.................................................... 1-1
1.1
Instrument Check............................................................................................... 1-1
1.2 Application.......................................................................................................... 1-2
2.
3.
DC402G Specifications............................................................................. 2-1
2.1
General specifications...................................................................................... 2-1
2.2
Operating specifications................................................................................... 2-3
2.3
Model and suffix codes..................................................................................... 2-5
Installation and Wiring.............................................................................. 3-1
3.1
Installation and dimensions............................................................................. 3-1
3.1.1
Installation site.................................................................................... 3-1
3.1.2
Mounting methods.............................................................................. 3-1
3.2 Preparation......................................................................................................... 3-5
3.3
3.4
3.5
4.
Wiring the power supply................................................................................... 3-6
3.3.1
General precautions........................................................................... 3-6
3.3.2
Access to terminal and cable entry..................................................... 3-6
3.3.3
AC power............................................................................................ 3-8
3.3.4
Grounding the housing....................................................................... 3-8
3.3.5
Switching on the instrument................................................................ 3-9
Wiring the contact signals................................................................................ 3-9
3.4.1
General precautions........................................................................... 3-9
3.4.2
Contact outputs................................................................................... 3-9
3.4.3
Contact input..................................................................................... 3-10
Wiring the analog output signals................................................................... 3-10
3.5.1
General precautions......................................................................... 3-10
3.5.2
Analog output signals........................................................................ 3-11
3.6
Sensor wiring................................................................................................... 3-11
3.7
Sensor connection using junction box and extension cable..................... 3-12
3.8
Other sensor systems..................................................................................... 3-12
Operation; Display Functions And Setting............................................. 4-1
4.1
Operator interface.............................................................................................. 4-1
4.2
Explanation of operating keys.......................................................................... 4-2
IM 12D08E02-01E
6th Edition: Feb. 16, 2016-00
Toc-2
5.
4.3
Setting passcodes............................................................................................. 4-3
4.4
Display example................................................................................................. 4-3
4.5
Display functions............................................................................................... 4-4
Parameter setting...................................................................................... 5-1
5.1
5.2
Maintenance mode............................................................................................ 5-1
5.1.1
Manual activation of Hold.................................................................... 5-2
5.1.2
Setpoint adjustment............................................................................ 5-3
Commissioning mode....................................................................................... 5-4
5.2.1 Setpoints............................................................................................. 5-5
5.2.2 Range.................................................................................................. 5-7
5.2.3 Hold..................................................................................................... 5-9
5.2.4
Temperature compensation.............................................................. 5-11
5.2.5 Service.............................................................................................. 5-13
5.3
Notes for guidance in the use of service coded settings............................ 5-14
5.3.1
Parameter specific functions............................................................ 5-14
5.3.2
Temperature functions...................................................................... 5-15
5.3.3
Temperature compensation functions.............................................. 5-16
5.3.4
mA output functions.......................................................................... 5-17
5.3.5
Contact outputs................................................................................. 5-20
5.3.6
User interface.................................................................................... 5-24
5.3.7
Communication setup....................................................................... 5-26
5.3.8 General............................................................................................. 5-26
5.3.9
Test and setup mode......................................................................... 5-26
6. Calibration.................................................................................................. 6-1
6.1
When is calibration necessary?....................................................................... 6-1
6.2
Calibration procedure....................................................................................... 6-2
6.3
Calibration with HOLD active........................................................................... 6-3
7. Maintenance............................................................................................... 7-1
7.1
Periodic maintenance for the DC402G converter.......................................... 7-1
7.2
Periodic maintenance of the sensor................................................................ 7-1
7.3
Fuse Replacement............................................................................................. 7-2
8. Troubleshooting........................................................................................ 8-1
9.
Spare Parts................................................................................................. 9-1
10.
Appendix.................................................................................................. 10-1
10.1
User setting for non-linear output table (code 31, 35 and 36)..................... 10-1
10.2
User entered matrix data (code 23 to 28)...................................................... 10-1
10.3
Matrix data table (user selectable in code 22).............................................. 10-2
10.4
Sensor Selection.............................................................................................. 10-3
10.4.1 General............................................................................................. 10-3
10.4.2
Sensor selection............................................................................... 10-3
10.4.3
Selecting a temperature sensor....................................................... 10-3
IM 12D08E02-01E
6th Edition: Feb. 16, 2016-00
Toc-3
10.5
Setup for other functions................................................................................ 10-3
10.6
User setting table............................................................................................. 10-4
10.7
Configuration checklist for DC402G.............................................................. 10-6
10.8
USP <645> Water Purity Monitoring.............................................................. 10-7
10.9
WHAT IS DUAL CONDUCTIVITY?.................................................................. 10-8
Customer Maintenance Parts List.......................................CMPL 12D08E02-02E
Revision Information................................................................................................i
IM 12D08E02-01E
6th Edition: Feb. 16, 2016-00
Blank Page
1.
<1. Introduction and General Description>
1-1
Introduction and General Description
The Yokogawa DC402G is a 4-wire converter designed for industrial process monitoring,
measurement and control applications. This instruction manual contains the information needed
to install, set up, operate and maintain the unit correctly. This manual also includes a basic
troubleshooting guide to answer typical user questions.
Yokogawa can not be responsible for the performance of the DC402G if these instructions are
not followed.
1.1
Instrument Check
Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged
during shipment. If damage is found, retain the original packing materials (including the outer
box) and then immediately notify the carrier and the relevant Yokogawa sales office.
Make sure the model number on the nameplate affixed to the left side of the instrument agrees
with your order.
CAUTION
The nameplate will also contain the serial number and power supply selection.
Be sure to apply correct power to the unit.
Check that all the parts are present, including mounting bracket, as specified in the option codes
at the end of the model number. For a description of the model codes, refer to Chapter 2 of this
manual under General Specifications.
Basic Parts List:
Converter DC402G
Instruction Manual English
Optional mounting hardware when specified (See model code)
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
1-2
<1. Introduction and General Description>
1.2 Application
The DC402G converter is intended to be used for continuous on-line measurement in industrial
installations. The unit combines simple operation and microprocessor-based performance with
advanced self-diagnostics and enhanced communications capability to meet the most advanced
requirements. The measurement can be used as part of an automated process control system.
It can also be used to indicate dangerous limits of a process, to monitor product quality, or to
function as a simple controller for a dosing/neutralization system.
Yokogawa designed the DC402G to withstand harsh environments. The converter may be
installed either indoors or outside because the IP65 (NEMA 4X) housing and cabling glands
ensure the unit is adequately protected. The flexible polycarbonate window on the front door
of the DC402G allows pushbutton access to the keypad, thus preserving the water and dust
protection of the unit even during routine maintenance operations.
A variety of DC402G hardware is optionally available to allow wall, pipe, or panel mounting.
Selecting a proper installation site will permit ease of operation. Sensors should normally be
mounted closely to the converter in order to ensure easy calibration and peak performance. If
the unit must be mounted remotely from the sensors, WF10 extension cable can be used up to a
maximum of 50 metres (150 feet) with a BA10 junction box.
The DC402G is delivered with a general purpose default setting for programmable items.
(Default settings are listed in Chapter 5 and again in Chapter 10). While this initial configuration
allows easy start-up, the configuration should be adjusted to suit each particular application.
An example of an adjustable item is the type of temperature sensor used. The DC402G can be
adjusted for any one of five different types of temperature sensors.
To record such configuration adjustments, write changes in the space provided in Chapter 10
of this manual. Because the DC402G is suitable for use as a monitor, a controller or an alarm
instrument, program configuration possibilities are numerous.
Details provided in this instruction manual are sufficient to operate the DC402G with all
Yokogawa sensor systems and a wide range of third-party commercially available probes. For
best results, read this manual in conjunction with the corresponding sensor instruction manual.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
2-1
<2. DC402G Specifications>
2.
DC402G Specifications
2.1
General specifications
l Input specifications
Two inputs , each 2-electrode measurement with square wave excitation, using cell constants(C)
from 0.008 to 50.0 cm-1, with up to 60 metres (200ft) connection cable.
l Detection method
Frequency, read-pulse position and reference voltage are dynamically optimized.
l Input ranges
Minimum :
Maximum :
- Resistivity :
1μS x C at process temperature (underrange 0.000 μS/cm).
25 mS x C at process temperature (overrange 30 mS x C).
0.00 kΩ - 999 MΩ/C at 25 °C (77 °F) reference temperature.
Minimum :
Maximum :
40 Ω/C at process temperature (underrange 0.001 kΩ x cm).
1 MΩ/C at process temperature (overrange 999 MΩ x cm).
- Temperature
Pt1000 :
-20 to +250 °C (0 to 500 °F)
Pt100 and Ni100 :
-20 to +200 °C (0 to 400 °F)
8K55 NTC :
-10 to +120 °C (10 to 250 °F)
PB36 NTC : -20 to +120 °C (0 to 250 °F)
l Span
Conductivity/Resistivity: Min. span : 0.010 μS/cm; 0.001 kΩ x cm
up to 90% maximum zero suppression.
Max. span : 1500 mS/cm; 999 MΩ x cm
Ratio (cell1/cell2):
Min. span : 00.0
Max. span : 19.99
Difference (cell1- cell2): Min. span : 0.010 μS/cm
Max. span : 400 mS/cm
% Passage (100x[cell2/cell1] ): Min. span : 00.0
Max. span : 199.9
% Rejection (100x[( cell1-cell2)/cell1] ): Min. span : 0.1
Max. span : 400
% Deviation (100x[(cell2-cell1)/cell1] ): Min. span : 0.1
Max. span: 400
VGB-directive 450 L:
Min. span : 1.0 pH
Max. span : 14.0 pH
Temperature: Min. span : 25 °C (50 °F)
Max. span : 250°C (500 °F)
Difference Temperature:
Min. span : 25 °C (50 °F)
Max. span : 250 °C (500 °F)
l Transmission Signals
Two isolated outputs of 0/4-20 mA DC with common negative.
Max. load : 600 Ω.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
2-2
<2. DC402G Specifications>
Auxiliary output can be chosen from conductivity, linearized conductivity, resistivity, temperature,
differential temperature calculated value or PI control of conductivity/resistivity.
Burn up (22 mA) or Burn down (0/3.5 mA) to signal failure.
l Temperature compensation
Automatic, for temperature ranges mentioned under C (input ranges).
- Reference temperature :
programmable from 0 to 100 °C or 30 to 210 °F (default 25 °C).
l Compensation algorithm
According IEC 60746-3 NaCl tables (default). Two independent user programmable
temperature coefficients, from 0% to 3.5% per °C (°F) by adjustment or calibration.
- Matrix compensation : with conductivity function of concen-tration and temperature. Choice of 5
preprogrammed matrixes and a 25-points userprogrammable matrix.
l Display
Custom liquid crystal display, with a main display of 3 1/2 digits 12.5 mm high. Message display
of 6 alphanumeric characters, 7 mm high. Warning flags and units (mS/cm, kΩ·cm, μS/cm and
MΩ·cm) as appropriate.
l Contact Outputs
- General : Four (4) SPDT relay contacts with LED indicators. For S1, S2, and S3, the LED is
on when relay power is removed.
NOTE:
For S4 (FAIL) LED lights when relay is deenergised (Fail safe).
Contact outputs configurable for hysteresis and delay time.
- Switch capacity :
- Status :
Maximum values 100 VA, 250 VAC, 5 Amps.
Maximum values 50 Watts, 250 VDC, 5 Amps.
High/low process alarms, selected from conductivity, resistivity and temperature.
Contact output is also available to signal “Hold active”
- Control function :
On/Off
PI pulsed :
Proportional duty cycle control with integral term.
PI frequency : Proportional frequency control with integral term. (PI control on
Conductivity/Resistivity only) In addition FAIL alarm for system and
diagnostic errors on S4.
l Power Supply
Supply voltage rating: 115, 230 VAC
Applicable range:
97.8 to 132.2, 195.5 to 264.5 VAC
Supply frequency rating:
50 / 60 Hz
Applicable range:
50 Hz ± 5% / 60 Hz ± 5%
Power consumption:
Maximum 10 VA for steady operation
l Shipping Details
Package size : W x H x D 290 x 300 x 290 mm. 11.5 x 11.8 x 11.5 in.
Packed weight :
Approx. 2.5 kg (5lb).
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
2.2
<2. DC402G Specifications>
2-3
Operating specifications
l Performance : Conductivity
Linearity:
± 0.5 % FS
Repeatability: ± 0.5 % FS
Accuracy:
± 0.5 % FS
l Performance : Resistivity
Linearity:
± 0.5 % FS
Repeatability: ± 0.5 % FS
Accuracy:
± 0.5 % FS
l Performance : Temperature with Pt1000Ω, Ni100Ω and PB36 NTC
Linearity:
± 0.3 ºC
Repeatability: ± 0.3 ºC
Accuracy:
± 0.3 ºC
l Performance : Temperature with PT100Ω and 8k55Ω
Linearity:
± 0.4 ºC
Repeatability: ± 0.4 ºC
Accuracy:
± 0.4 ºC
Note: The following tolerance are added to above performance.
mA output tolerance :
± 0.02 mA of “0/4 - 20 mA”
Digital display tolerance:
+1 digit
l Performance : Temperature compensation
NaCl table:
± 1%
Matrix:
± 3%
Step response: 90 % (< 2 decades) in ≤ 6 seconds
l Ambient operating temperature
-10 to +55 °C (14 to 131 ºF)
l Storage temperature
-30 to +70 °C (-22 to 158 ºF)
l Humidity
10 to 90% RH non-condensing
l Housing
Cast aluminium case with chemically resistant coating, cover with flexible polycarbonate window.
Case color is off-white and cover is moss green. Cable entry is via six PG13.5 nylon glands.
Cable terminals are provided for up to 2.5 mm finished wires. Weather resistant to IP65, NEMA
4X, CSA Type 3S. Pipe wall or panel mounting, using optional hardware.
l Data protection
EEPROM for configuration and lithium battery for clock.
l Watchdog timer
Checks microprocessor
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<2. DC402G Specifications>
2-4
l Automatic safeguard
Return to measuring mode when no keystroke is made for 10 min.
l Power interruption
Less than 50 milliseconds no effect.
l Operation protection
3-digit programmable password.
l Safety and EMC conforming standards
Safety: conforms to EN 61010-1
EN 61010-2-030
EN 61010-2-201
CAN/CSA No.61010-1
UL Std. No. 61010-1
CSA C22.2 No. 94.2
UL 50E
EMC:
EN 61326-1* Class A, Table 2 (Note 1)
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
RCM: EN61326-1 Class A
Korea Electromagnetic Conformity Standard Class A
한국 전자파적합성 기준
*: Influence of immunity environment (Criteria A): Output shift is specified within ±25% of F.S.
Installation altitude:
2000 m or less
Category based on IEC 61010: II (Note 2)
Pollution degree based on IEC 61010:2 (Note 2)
Note 1: This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the
industrial environment only.
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
Note 2: Installation category, called overvoltage category, specifies impulse withstand voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor environment.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<2. DC402G Specifications>
2.3
2-5
Model and suffix codes
[Style: S2]
Model
Suffix code Option code
Description
DC402G
•••••••••••••••••• ••••••••••••••••• Dual Conductivity Converter
Type
-1
••••••••••••••••• General
Power Supply
-1
••••••••••••••••• 115 V +/-15% AC, 50/60 Hz (*2)
Voltage
-2
••••••••••••••••• 230 V +/-15% AC, 50/60 Hz
Language
-E ••••••••••••••••• English
-J ••••••••••••••••• Japanese
Pipe, wall mounting bracket (Stainless steel)
Options
/U
Panel mounting bracket (Stainless steel)
/PM
Hood for sun protection (Carbon steel)
/H3
Hood for sun protection (Stainless steel)
/H4
Stainless steel tag plate
/SCT
G 1/2
/AFTG
1/2 NPT
/ANSI
Epoxy baked finish (*1)
/X1
*1:
*2:
The housing is coated with epoxy resin.
When CSA safe standard conformity product is needed, select 115V “-1” of Power Supply Voltage.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page
3-1
<3. Installation and Wiring>
3.
Installation and Wiring
3.1
Installation and dimensions
CAUTION
Installation must be performed by a qualified personnel. If this rules is not followed and a damage
occurs, Yokogawa will not be held responsible.
3.1.1
Installation site
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
The DC402G converter is weatherproof and can be installed inside or outside. It should,
however, be installed as close as possible to the sensor to avoid long cable runs between sensor
and converter. In any case, the total cable length should not exceed 60 meters (200 feet). Select
an installation site where:
• Mechanical vibrations and shocks are negligible
• No relay/power switches are in the direct environment
• Access is possible to the cable glands (see figure 3.1)
• The converter is not mounted in direct sunlight or severe weather conditions
• Maintenance procedures are possible (avoiding corrosive environments)
The ambient temperature and humidity of the installation environment must be within the limits of
the instrument specifications. (See chapter 2).
3.1.2
Mounting methods
Refer to figures 3.2 and 3.3. Note that the DC402G converter has universal mounting
capabilities:
•
•
•
•
Panel mounting using optional brackets
Surface mounting on a plate (using bolts from the back)
Wall mounting on a bracket (for example, on a solid wall)
Pipe mounting using a bracket on a horizontal or vertical pipe (maximum pipe diameter 50 A)
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-2
<3. Installation and Wiring>
□ Hood (optional)
Option code : /H□
184
Four M6 screws, 8 (0.31) deep
220
(7.24)
80
(8.66)
(3.15)
72
(2.83)
144
20 (0.79)
(5.67)
80
(3.15)
112
23
144
(5.67)
Adaptor for conduit work
(option code : /AFTG, /ANSI)
(4.41)
(0.91)
A
D
B
E
36
C
F
A : For sensor cable
B : For sensor cable
36 (1.42) C : For output signal
D : For contact output (S3 and S4)
E : For contact output (S1 and S2)
38 (1.50)
F : For power supply
49
(1.93)
36
(1.42) (1.42)
Ground terminal
(M4 screw *)
Adaptor
Approx. 55 (2.17)
Cable inlet port (21 (0.83) dia. holes)
equivalent to DIN PG13.5 cable gland
Weight: Approx. 2 kg
*: Tightening torque: 1.2 N·m.
G 1/2 female ( / AFTG)
1/2 NPT female ( / ANSI)
F17.ai
Figure 3.1 Housing dimensions and layout of glands
23 (0.91)
12 max.(panel thickness)
M6, 4 screws *
Panel cutout dimensions
(0.47)
M5, 2 screws
137 +20
100
(3.94)
(5.43)
137 +20
178
(7.01)
(5.43)
*: Tightening torque: 2 N·m.
F14-2.ai
Figure 3.2 Panel mounting diagram (Option Code: /PM)
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-3
<3. Installation and Wiring>
Example of bracket used for pipe mounting
188
M6, 4 screws *
(7.40)
174
(6.85)
200
(7.87)
50
(1.97)
Nominal 50A (O.D 60.5mm)
(2 inch)
mounting pipe
100
(3.94)
Example of bracket used for wall mounting
135
(5.31)
13
M6, 4 screws *
(0.51)
224
200
(8.82)
(7.87)
35
(1.38)
15
(0.59)
70 (2.76)
*: Tightening torque: 2 N·m.
10mm dia., 3 holes
(0.39)
100 (3.94)
F14-1.ai
Figure 3.3 Wall and pipe mounting diagram (Option Code: /U)
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-4
<3. Installation and Wiring>
Figure 3.4 Internal view of DC402G wiring compartment
CAUTION
Wiring must be performed by a qualified personnel. If this rules is not followed and a damage
occurs, Yokogawa will not be held responsible.
WARNING
Never apply power to the DC402G converter and other instruments connected to the DC402G
converter until all wiring is completed.
WARNING
This product complies with the CE marking.
Where compliance with the CE marking and relevant standard is necessary, the following wiring
is required.
1.
2.
3.
4.
5.
Install an external switch or circuit breaker to the power supply of the DC402G converter.
Use an external switch or circuit breaker rated 5A and conforming to IEC 60947-1 or IEC
60947-3.
It is recommended that the external switch or circuit breaker be installed in the same room
as the DC402G converter.
The external switch or circuit breaker should be installed within reach of the operator and
identified with marking as a power supply switch to the DC402G converter.
Power lines such as power cables and contact outputs should be fixed securely onto a wall
or construction using cable racks, conduit tubing, nylon bands or other appropriate ways.
Accidental removal from terminals by pulling may result in electric shock.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-5
<3. Installation and Wiring>
3.2 Preparation
Refer to figure 3.4. The relay contact terminals and power supply connections are under the
screening (shielding) plate. These should be connected first. Connect the sensor and outputs.
To open the DC402G for wiring:
1.
Loosen the four frontplate screws and remove the cover.
2.
Use the rubber knob in the lower right hand corner and swing open the display board to the
left.
3.
The upper terminal strip is now visible.
4.
Remove the screen (shield) plate covering the lower terminal strip.
5.
Connect the power supply and contact outputs. Use the three glands at the back for these
cables.
6.
Replace the screen (shield) plate over the lower terminals.
WARNING
Always replace the screen plate over the power and contact outputs for safety and to avoid
interference.
7.
Connect the analog output(s) and the sensor input.
8.
Use the front three glands for analog output, sensor input, contact input and communication
cabling (see figure 3.5).
9.
Close the display board and switch on the power. Commission the instrument as required or
use the default settings.
10. Replace the cover and secure frontplate with the four screws.
Tighten four frontplate screws to 1.5 N·m torque.
CAUTION
For wiring, use cables with a heat resistance of at least 70°C.
CAUTION
Tighten the cable glands to 1.5 N•m torque. Torque less than 1.5 N•m cannot maintain the IP65
protection level.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-6
<3. Installation and Wiring>
High Voltage Section
CONTACT
(S3, S4/FAIL)
OUTPUT
CABLE(S)
SENSOR
CABLE(S)
CONTACT
(S1, S2)
OUTPUT
CABLE(S)
SENSOR
CABLE(S)
POWER
CABLE
ANALOG
OUTPUTS
CABLE
Suitable for cables with an outside diameter between 6 - 12 mm (0.24 - 0.47 in.)
Figure 3.5 Glands to be used for cabling
3.3
Wiring the power supply
3.3.1
General precautions
Make sure the power supply is switched off. Also, make sure that the power supply is correct for
the specifications of the DC402G and that the voltage specified on the nameplate on the left side
of the DC402G is satisfied.
Local health and safety regulations may require an external circuit breaker to be installed. The
instrument is protected internally by a fuse. The fuse rating is dependent on the supply to the
instrument. The 250 VAC fuses should be of the “time-lag” type, conforming to IEC60127.
The internal fuse is located next to the power terminals (in the lower right hand corner).
WARNING
Use only a fuse of the specified current, voltage and type ratings to prevent fire. For fuse
replacement, refer to Section 7.3, “Fuse Replacement.”
3.3.2
Access to terminal and cable entry
Terminals 1 and 2 on the bottom terminal strip are used for the power supply. Guide the power
cables through the gland closest to the power supply terminals. Use 2.5 mm2 (14 AWG) wires to
connect to the terminals.
We recommend that the cables be terminated with pin terminals (see figure 3.6).
Connect the wires as indicated in the wiring diagram (refer to figure 3.7).
CAUTION
Use cables whose outer diameter is 6 to 12 mm.
For power cables, use cables that comply with UL2556VW-1 or equivalent. Tighten the cables to
0.5 N•m torque.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-7
<3. Installation and Wiring>
2.2mm or less
Figure 3.6 Pin terminal
Table 3.1 Connection method
Power supply,
contact output
Others
Power supply
Pin form crimp terminal with the insulation sleeve
connection method
Conformity shape 2.5 mm2 (14 AWG)
0.13 to 2.5 mm2 (26 to 14 AWG)
Clamping torque
0.5 N•m
Crimp terminal
Product made in
Product made in wide Muller: (include the left model)
example
wide Muller: H2.5/7,
H0.25/5, H0.5/6, H0.5/10, H0.75/6, H0.75/10, H1/6, H1/10, H1.5/7, H1.5/10,
(Type)
H2.5/10, H2.5/12,
H1.5/12, H1.5/18, H0.14/10, H0.14/12, H0.25/10, H0.25/12, H0.34/10,
H2.5/18, H2.5/14D,
H0.34/12, H0.5/10, H0.5/12, H0.5/14, H0.5/16, H0.75/12, H0.75/14,
H2.5/18D, H2.5/24D H0.75/16, H0.75/18H1/12, H1/14, H1/16, H1/18, H1.5/14, H1.5/16, H1.5/24
WARNING
When connecting an external device, install and wire it so that requirements of the standard to
which the device complies are met.
FRONT GLANDS
REAR GLANDS
Sensors
Power
S1
0/4-20 mA Output
signals
Contact
output
0/4-20 mA
S2
S3
S4/FAIL
Contact
input
Figure 3.7 System configuration
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-8
<3. Installation and Wiring>
Sensor inputs
22 21 11
12 14
15
mA Outputs
11
12 14
15
Screen
23
Temp
Contact
Sensor 1
Temp
Sensor 2
63 66 65 62 61 95 94 93 92 91
mA1
mA2
Screen
mA Outputs
REFER TO INSTRUCTION MANUAL FOR CONNECTIONS
Relay Contacts
Power Supply
71 72 73 51 52 53 41 42 43 31 32 33
250VAC
5A
100VA
3
2
1
G
N
L
FUSE
250VDC
5A
50W
C NC NO C NC NO C NC NO C NC NO
S3
S1
S4
S2
250VAC; T 200mA
250VAC; T 100mA
115 VAC
230 VAC
Figure 3.8 Input and output connections
3.3.3
AC power
Connect terminal 1 to the phase line of the AC power and terminal 2 to the zero line. The size of
conductors should be at least 1.25 mm2. The overall cable diameter should be between 6 & 12
mm (0.24 & 0.47 in).
CAUTION
Should be used the cable beyond 300 V AC.
3.3.4
Grounding the housing
WARNING
Protective grounding must be made to prevent electric shock.
To protect the instrument against interference, the housing should be connected to ground by
a large area conductor. This cable can be fixed to the rear of the housing using a braided wire
cable. See figure 3.9. This is separated from input ground by a galvanic isolation.
The internal grounding terminal is not the protective grounding terminal.
CAUTION
Please be sure to connect protective grounding of DC402G with cable of 1.25 mm2 or larger
cross section in order to avoid the electrical shock to the operators and maintenance engineers
and prevent the influence of external noise. In addition, use cables whose colors are green and
yellow stripes.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-9
<3. Installation and Wiring>
And further connect the grounding wire to the
mark (100 Ω or less).
Location of grounding terminal
(M4 Screw)
F0308.ai
Figure 3.9 Grounding the housing
3.3.5
Switching on the instrument
After all connections are made and checked, the power can be switched on from the power
supply.
Make sure the LCD display comes on. All segments will illuminate, then the instrument will
momentarily display its unique serial number. After a brief interval, the display will change to
the measured value. If errors are displayed or a valid measured value is not shown, consult the
troubleshooting section (Chapter 8) before calling Yokogawa.
3.4
Wiring the contact signals
WARNING
There are high-voltage sections on the DC402G. Be sure to turn off the power before wiring.
3.4.1
General precautions
The contact output signals consist of voltage-free relay contacts for switching electrical
appliances (SPDT). They can also be used as digital outputs to signal processing equipment
(such as a controller or PLC). It is possible to use multi-core cables for the contact in and output
signals and shielded multi-core cable for the analog signals.
We recommend that the cables be terminated with pin terminals (see figure 3.6).
3.4.2
Contact outputs
The DC402G unit’s four contact outputs can be wired to suit your own custom requirements
(Figure 3.7).
In the Non-Alarm or Power Off states, contacts S1, S2 and S3 are OFF, Common (C) and
Normally Closed (NC) are in contact.
In the “Fail” or Power Off states, contact S4 is ON, Common (C) and Normally Closed (NC) are in
contact.
You can either use them to switch AC power, or switch a DC Voltage for digital interfacing.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3-10
<3. Installation and Wiring>
Default settings
•
•
•
•
The contact S1 is pre-programmed for high alarm function.
The contact S2 is pre-programmed for a low alarm function.
The contact S3 is pre-programmed for wash function.
The contact S4 is pre-programmed for FAIL.
The three control contacts (S1 to S3) can be used for simple process control by programming
their function (Chapter 5). The FAIL contact is programmed to signal a fault in the measuring
loop. Always connect the FAIL contact to an alarm device such as a warning light, sound
annunciator, or alarm panel to make full use of the fault detection possibilities (self diagnostics) of
the DC402G converter.
CAUTION
Should be used the cable beyond 300 V AC.
When using ON/OFF signals whose voltage is 33 VAC or 70 VDC or higher, use 2.5 mm2
(AWG14) wires. If using voltages lower these voltages, select wires in the range of 0.13 to 2.5
mm2 (AWG26 to AWG14) according to the current capacity of the connected load.
Use cables that comply with UL2556VW-1 or equivalent and whose outer diameter is 6 to 12
mm. Tighten the cables to 0.5 N•m torque.
3.4.3
Contact input
Voltage-free contact (do not apply voltage)
Wash start or input remote range change (either choice)
It is necessary to use screening/shielding on the output signal cables. Screw (M3) 23 is used to
connect the shielding.
Select wires in the range of 0.13 to 2.5 mm2 (AWG26 to AWG14) according to the current
capacity of the connected load.
Use cables that comply with UL2556VW-1 or equivalent and whose outer diameter is 6 to 12
mm. Tighten the cables to 0.5 N•m torque.
3.5
Wiring the analog output signals
3.5.1
General precautions
The analog output signals of the DC402G transmit low power standard industry signals to
peripherals like control systems or strip-chart recorders (Figure 3.7).
Do not apply voltage.
CAUTION
Should be used the cable beyond 30 V AC.
Select wires in the range of 0.13 to 2.5 mm2 (AWG26 to AWG14) according to the current
capacity of the connected load.
Use cables that comply with UL2556VW-1 or equivalent and whose outer diameter is 6 to 12
mm. Tighten the cables to 0.5 N•m torque.
We recommend that the cables be terminated with pin terminals (see figure 3.6).
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3.5.2
3-11
<3. Installation and Wiring>
Analog output signals
The output signals consist of active current signals of either 0-20 mA or 4-20 mA. The maximum
load can be 600 ohms on each.
It should be necessary to use screening/shielding on the output signal cables. Terminal 63 is
used to connect the shielding.
3.6
Sensor wiring
Refer to figure 3.9, which includes drawings that outline sensor wiring.
The DC402G can be used with a wide range of commercially available sensor types if provided
with shielded cables, both from Yokogawa and other manufacturers. The sensor systems from
Yokogawa fall into two categories, the ones that use fixed cables and the ones with separate
cables.
To connect sensors with fixed cables, simply match the terminal numbers in the instrument with
the identification numbers on the cable ends.
Note that the DC402G uses the 2 electrode measuring principle. Yokogawa sensors and cables
are prepared for compatibility with 4-electrode measuring systems. To avoid problems either cut
off and insulate the wires tagged 13 &16 or connect the wires in tandem 13 &14 into terminal 14
or 15 &16 into terminal 15.
CONDUCTIVITY RESISTIVITY CONVERTER
11 TEMPERATURE
12TEMPERATURE
14 CELL
15 CELL
SEPARATE SENSORS WITH WU40-LH.. CABEL
11 TEMPERATURE
12 TEMPERATURE
1
2
1
BROWN
BROWN
YELLOW/GREEN
2
RED
11 TEMPERATURE
12 TEMPERATURE
14 OUTER ELECTRODE
15 INNER ELECTRODE
SX42-SX . . - . F SENSORS
Note: Use shielded cable
14 OUTER ELECTRODE
15 INNER ELECTRODE
SC4A... SENSORS WITH INTEGRATED CABEL
Figure 3.10 Sensor wiring diagrams
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
3.7
<3. Installation and Wiring>
3-12
Sensor connection using junction box and
extension cable
Where a convenient installation is not possible using the standard cables between sensors and
converter, a junction box and extension cable may be used. The Yokogawa BA10 junction box
and the WF10 extension cable should be used. These items are manufactured to a very high
standard and are necessary to ensure that the specifications of the system can be met. The total
cable length should not exceed 60 metres (e.g. 10 m fixed cable and 50 m extension cable).
NOTE
Numbers 17 of both WF10 and BA10 do not need to be used.
3.8
Other sensor systems
To connect other sensor systems, follow the general pattern of the terminal connections as listed
below:
11 and 12 Always used for temperature compensation resistor input (Pt1000, Ni100, Pt100,
PB36 and 8k55)
14
Normally used for the outer electrode
15
Used for inner electrode
In case a 4-electrode measuring system will be used, 14 and 16 should be used for the current
electrodes. Please ensure that shielded cabling will be used.
In figure 3.10 this is shown in a schematic way.
11 12 14 15
TEMPERATURE
SENSOR
t
CELL
ELECTRODE
2-electrode configuration
Figure 3.11 Connection diagram for other sensors
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<3. Installation and Wiring>
Sensor Inputs
3-13
mA Outputs
22 21 11 12 14 15 11 12 14 15
63 66 65 62 61 95 94 93 92
SCREEN
91
mA1
mA2
SCREEN
23
CONT
SENSOR 1
SENSOR 2
mA OUTPUT
DC402 REFER TO INSTRUCTION MANUAL FOR CONNECTIONS
Relay Contacts
Power Supply
71 72 73 51 52 53 41 42 43 31 32 33
3
G
250VAC
250VDC
5A
100VA
5A
50W
1
L
FUSE
250V
T200mA
C NC NO C NC NO C NC NO C NC NO
S4
2
N
S3
S2
115 VAC
S1
Figure 3.12 Terminal identification labels example
Figure 3.13 Sensor cable connections
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page
4-1
<4. Operation; Display Functions And Setting>
4.
Operation; Display Functions And
Setting
4.1
Operator interface
This section provides an overview of the operation of the DC402G operator interface. The basic
procedures for obtaining access to the three levels of operation are described briefly. For a stepby-step guide to data entry, refer to the relevant section of this instruction manual. Figure 4.1
shows the DC402G operator interface.
l LEVEL 1: Maintenance
These functions are accessible by pushbutton through a flexible front cover window. The
functions make up the normal day-to-day operations that an operator may be required to
complete. Adjustment of the display and routine calibration are among the features accessible in
this way. (See table 4.1).
l LEVEL 2: Commissioning
A second menu is exposed when the DC402G front cover is removed and the display board is
revealed. Users gain access to this menu by pressing the button marked * in the lower right of
the display board. This menu is used to set such values as the output ranges and hold and wash
features. It also gives access to the service menu. (See table 4.1).
l LEVEL 3: Service
For more advanced configuration selections, press the button marked * , then press “NO”
repeatedly until you reach SERVICE. Now push the “YES” button. Selecting and entering
“Service Code” numbers in the commissioning menu provide access to the more advanced
functions. An explanation of the Service Codes is listed in chapter 5 and an overview table is
shown in chapter 10.
Table 4.1 Operations overview
Maintenance
Commissioning
Service
(Access to coded
entries from the
commissioning level)
Routine
SETPOINTS
CALIB 1(2)
DISPLAY 1(2)
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP
SERVICE
Function
Adjust alarm setpoints (when activated)
Calibration with a standard solution or a sample
Read auxiliary data or set message display
Switch hold on/off (when activated)
Adjust alarm setpoints
Adjust the output range
Activate the hold function
Select method of temperature compensation
Fine tune the specialized functions of the
converter
Chapter
5
6
4
5
5
5
5
5
5
NOTE: All three levels may be separately protected by a password. See Service Code 52 in
chapter 5 Service Code table for details on setting passwords.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
4-2
<4. Operation; Display Functions And Setting>
Output hold flag
Main display
Fail flag
HOLD
FAIL
Menu pointer flags
Units
MODE
cm
k
mS/cm
M
cm
S/cm
Message display
YES
NO
ENT
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
Key prompt flags
YES
NO
MODE
CONTACTS
*
S1
Selection keys
Yes: Accept setting
No: Change setting
Adjustment keys
: Choose digit to adjust
: Adjust digit
ENT: Confirm change
Commissioning
function menu
Commissioning
mode access key
S2
ENT
Relay contact
status indicators
S3
FAIL
YOKOGAWA
Measure/Maintenance
mode key
MARKINGS
WITHIN
INCLOSURE
Broken line indicates area
that can be seen through
front cover
Figure 4.1 DC402G operator interface
4.2
Explanation of operating keys
MODE key This key toggles between the Measuring and Maintenance modes. Press once to
obtain access to the maintenance function menu.
SETPOINTS
CAL.1/CAL.2
DISP.1/DISP.2
HOLD
Press again to return to the Measuring mode (press twice when hold is activated).
YES/NO keys These are used to select choices from the menu.
YES is used to accept a menu selection.
NO
is used to reject a selection, or to move ahead to the next option.
DATA ENTRY keys (˃ ˄ ENT)
˃ is used as a “cursor” key. Each press on this key moves the cursor or
flashing digit one place to the right. This is used to select the digit to be
changed when entering numerical data.
˄ is used to change the value of a selected digit. Each press on this key
increases the value by one unit. The value can not be decreased, so in
order to obtain a lower value, increase past nine to zero, then increase to
the required number.
ENT When the required value has been set using the > and ^ keys, press ENT
to confirm the data entry. Please note that the DC402G does not register
any change of data until the ENT key is pressed.
* key This is the Commissioning mode key. It is used to obtain access to the
Commissioning menu. This can only be done with the cover removed or opened.
Once this button has been used to initiate the Commissioning menu, follow the
prompts and use the other keys as described above.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
4.3
4-3
<4. Operation; Display Functions And Setting>
Setting passcodes
In Service Code 52, DC402G users can set passcode protection for each one of the three
operating levels, or for any one or two of the three levels. This procedure should be completed
after the initial commissioning (setup) of the instrument. The passcodes should then be recorded
safely for future reference.
When passcodes have been set, the following additional steps are introduced to the configuration
and programming operations:
l Maintenance
Press MODE key. The display shows 000 and *PASS*
Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Maintenance Mode
l Commissioning
Press * key. The display shows 000 and *PASS*
Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Commissioning Mode.
l Service
From the commissioning menu, select *Service by pressing YES key. The display shows 000 and
*PASS*
Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Service Mode.
NOTE
See Service Code 52 for the setting of passcodes.
4.4
Display example
The next page shows the sequence of button presses and screens displayed when working in
default configuration.
More or less options will be made available by the configuration of some service codes, or by
choices made in the Commissioning menu.
The following deviations are possible:
Items marked are omitted when switched off in commissioning mode and/or service
code 51.
Temperature compensation will be displayed dependent on chosen compensation
method: NaCl, TC 2.1 or matrix.
DISP.2 only appears if mA2 is configured for a 2nd (different) temperature
compensation or if % by weight.2 is enabled in code 55.
W/W % only appears if switched on in service code 55.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
4.5
4-4
<4. Operation; Display Functions And Setting>
Display functions
Computed value
µS / c m
Cell temperature
µS / c m
Displ. 1 = Sensor 1
Calculation type
Displ. 2 = Sensor 2
MODE
NO
NO
Uncompensated
SC (if USP is enabled
in service code 57)
µS / c m
YES
YES
NO
µS / c m
YES
NO
YES (See Setpoint
menu Chapter 5-1)
NO
NO
µS / c m
µS / c m
YES
NO
YES (See Calibration
menu Chapter 6)
NO
DISP.1
or
DISP.2
YES
NO
NO
NO
µS / c m
µS / c m
YES
Reference
temperature
Software release
number
NO
YES
NO
YES
NO
NO
YES
µS / c m
YES
µS / c m
NO
NO
NO
NO
µS / c m
YES
YES
µS / c m
NO
YES
NO
µS / c m
YES
NO
w/w %
2nd compensated
value
NO
NO
µS / c m
YES
µS / c m
Only if enabled
NO
Process
temperature
NO
HOLD
FAIL
NO
ENT
YES
NO
MODE
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
Actual Cell
Constant
SETPOINTS
RANGE
SET HOL D
TEMP.
SERVICE
CONTACTS
S1
S2
ENT
YOKOGAWA
YES
NO
NO
µS / c m
MODE
k W .c m
mS/cm
M W .c m
mS / c m
YES
NO
*
YES
NO
NO
Press YES to fix
the selected second
line of display
µS / c m
Current
output
YES
NO
S3
NO
FAIL/S4
MARKINGS
WITHIN
INCLOSURE
Note:The variety of display screens depends
on the configuration of the service
settings (see section 5)
Temperature
compensation
for conductivity 1
µS / c m
YES
IM 12D08E02-01E
NO
6th Edition: Feb. 19, 2016-00
5-1
<5. Parameter setting>
5.
Parameter setting
5.1
Maintenance mode
Standard operation of the DC402G instrument involves use of the Maintenance (or operating)
mode to set up some of the parameters.
Access to the Maintenance mode is available via the six keys that can be pressed through the
flexible window in the instrument front cover. Press the “MODE” key once to enter this dialog
mode.
(Note that at this stage the user will be prompted for a pass code where this has been previously
set up in service code 52, section 5.)
Setpoint
Select and adjust setpoint (when enabled in service menu section 5.3,
service code 51).
See adjustment procedure 5.2.2.
Calibrate
See “calibration” section 6.
Display setting
See “operation” section 4.
Hold
Manually switch on/off “hold” (when enabled in commissioning menu). See
adjustment procedure 5.2.4.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5.1.1
<5. Parameter setting>
5-2
Manual activation of Hold
MODE
S/cm
MEASURE
YES
NO
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
CONTACTS
MODE
*
S1
S2
S3
ENT
FAIL/S4
MARKINGS
WITHIN
INCLOSURE
YOKOGAWA
MODE
NO
S/cm
YES
CAL 1
NO
NO
NO
NO
NO
S/cm
YES
YES
S/cm
YES
NO
NO
HOLD
NO
YES
HOLD
S/cm
MEASURE
YES
NO
DISPLAY 1
Note: The HOLD feature must first be activated in the commissioning mode section 5.2.4
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-3
<5. Parameter setting>
5.1.2
Setpoint adjustment
Note: To enable adjustments of setpoints in
maintenance mode, Service Code 51
must be set to "ON".
Setpoints available will depend on their
configuration in the Service Code.
MODE
S/cm
MEASURE
YES
NO
CONTACTS
MODE
S1
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
*
S2
S3
ENT
FAIL/S4
MARKINGS
WITHIN
INCLOSURE
YOKOGAWA
MODE
S/cm
YES
CAL 1
NO
NO
NO
NO
NO
For adjustments,
follow procedures
as in section 5.2.2
S/cm
YES
NO
YES
YES
YES
NO
S/cm
YES
NO
S/cm
YES
YES
NO
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5.2
5-4
<5. Parameter setting>
Commissioning mode
In order to obtain peak performance from the DC402G, you must set it up for each custom
application.
Setpoints
Alarms are set by default S1 - high process alarm
S2 - low process alarm
S3 - not activated
S4 - Fail
The setpoints are at arbitrary default value. Therefore, you must set these
to meaningful values, or set them to off. (See service codes 40 to 49 and
user interface codes 50 to 59.)
Output ranges
mA output 1 is set as default to 0-100 μS/cm or 0-19.99 MΩ•cm.
For enhanced resolution in more stable measuring processes, it may be
desirable to select 5-10 μS/cm range, for example, and maybe 0-25 °C
temperature range.
Service codes 30 to 39 can be used to choose other output parameters
on mA output 2.
Choose from Table, temperature or PI control.
Hold
The DC402G converter has the ability to “HOLD” the output during
maintenance periods. This parameter should be set up to hold the last
measured value, or a fixed value to suit the process.
Service
This selection provides access to the service menu.
What follows are pictorial descriptions of typical frontplate pushbutton sequences for each
parameter setting function. By following the simple YES/NO prompts and arrow keys, users can
navigate through the process of setting range, setpoints, hold and service functions.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-5
<5. Parameter setting>
5.2.1 Setpoints
MODE
M
cm
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
*
YES
YES
NO
YES
NO
YES
NO
NO
NO
YES
YES
NO
NO
S/cm
YES
NO
ENT
ENT
repeated
keystrokes
NO
S/cm
YES
S/cm
NO
S/cm
ENT
ENT
S/cm
S/cm
ENT
ENT
S/cm
S/cm
ENT
ENT
NO
YES
NO
NO
YES
NO
NO
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-6
<5. Parameter setting>
Process Alarms on
S.3 and S.4 are
only available when
enabled in Service
Codes 40-49
Analog control setpoint
is only available when
enabled in Service Code 31
NO
YES
NO
YES
NO
NO
NO
YES
NO
YES
NO
YES
mS / c m
YES
NO
ENT
>
Adjust setpoint value
using > ENT keys
as shown for setpoint 1.
Setpoint confirmed.
Return to mode
ENT
commissioning.
mS / c m
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
mS / c m
ENT
Negative signs only appear for temp. settings.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-7
<5. Parameter setting>
5.2.2 Range
MODE
S/cm
S/cm
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
ENT
*
YES
S/cm
See facing
page
NO
ENT
NO
NO
YES
S/cm
YES
YES
NO
NO
ENT
NO
YES
S/cm
YES
YES
NO
NO
ENT
YES
NO
YES
S/cm
S/cm
ENT
ENT
NO
NO
ENT
mS/cm
S/cm
YES
NO
ENT
NO
YES
ENT
ENT
NO
NO
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-8
<5. Parameter setting>
Choose Range to adjust, then set begin scale (0%) and
end scale (100%) of the mA output signal, using the >
>,
and ENT keys. Selection of mA output(0-20 / 4-20 mA)
is in Service Code 30.
The decimal point and unit setting can be changedas
described before in Setpoint Settings.
Range Selection
Options are determined by
Service Code 31
Note: Range 2 does
not appear when PI
control set on mA2
YES
YES
YES
NO
NO
YES
YES
YES
NO
NO
S/cm
YES
NO
ENT
ENT
ENT
ENT
S/cm
ENT
ENT
ENT
ENT
Range values set, return
to commission mode.
YES
YES
NO
IM 12D08E02-01E
NO
6th Edition: Feb. 19, 2016-00
5-9
<5. Parameter setting>
5.2.3 Hold
MODE
S/cm
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
*
YES
NO
NO
YES
NO
NO
YES
YES
YES
YES
NO
NO
NO
YES
YES
NO
NO
NO
NO
HOLD deactivated, return
to commissioning menu.
YES
NO
YES
HOLD
HOLD
YES
NO
YES
NO
NO
YES
YES
NO
YES
NO
HOLD active
last measured
value.
NO
NO
YES
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-10
<5. Parameter setting>
HOLD values set,
return to commissioning
menu.
HOLD
HOLD
ENT
ENT
HOLD
HOLD
ENT
ENT
ENT
Set HOLD "fixed value"
for mA2.
HOLD
ENT
HOLD
YES
Set HOLD "fixed value"
for mA1.
ENT
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-11
<5. Parameter setting>
5.2.4
Temperature compensation
1. Why temperature compensation?
The conductivity of a solution is very dependent on temperature. Typically for every 1 °C change
in temperature the solution conductivity will change by approximately 2 %.
The effect of temperature varies from one solution to another and is determined by several
factors like solution composition, concentration and temperature range.
A coefficient (a) is introduced to express the amount of temperature influence in % change in
conductivity/°C.
In almost all applications this temperature influence must be compensated before the
conductivity reading can be interpreted as an accurate measure of concentration or purity.
Table 5.1 NaCl-compensation according to IEC 60746-3 with Tref = 25 °C
T
0
10
20
25
30
40
50
Kt
0.54
0.72
0.90
1.0
1.10
1.31
1.53
a
1.8
1.9
2.0
--2.0
2.0
2.1
T
60
70
80
90
100
110
120
Kt
1.76
1.99
2.22
2.45
2.68
2.90
3.12
a
2.2
2.2
2.2
2.2
2.2
2.2
2.2
T
130
140
150
160
170
180
190
200
Kt
3.34
3.56
3.79
4.03
4.23
4.42
4.61
4.78
a
2.2
2.2
2.2
2.2
2.2
2.2
2.2
2.2
2. Standard temperature compensation
From the factory the DC402G is calibrated with a general temperature compensation function
based on a sodium chloride salt solution. This is suitable for many applications and is compatible
with the compensation functions of typical laboratory or portable instruments.
A temperature compensation factor is derived from the following equation:
a=
Kt - Kref
T - Tref
x
100
K ref
In which:
a = Temperature compensation factor (in %/ °C)
T = Measured temperature (°C)
Kt = Conductivity at T
Tref = Reference temperature (°C)
Kref = Conductivity at Tref
3. Manual temperature compensation
If the standard compensation function is found to be inaccurate for the sample to be measured,
the converter can be set manually for a linear factor on site to match the application.
The procedure is as follows:
1. Take a representative sample of the process liquid to be measured.
2. Heat or cool this sample to the reference temperature of the converter (usually 25 °C).
3. Measure the conductivity of the sample with the DC402G and note the value.
4. Bring the sample to the typical process temperature (to be measured with the DC402G).
5. Adjust the display indication to the noted value at the reference temperature
6. Check that the temperature compensation factor has been changed
7. Insert the conductivity cell into the process again.
4. Other possibilities (section 5.3.3)
1. Enter calculated coefficient.
2. Enter matrix temperature compensation.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-12
<5. Parameter setting>
MODE
µS / c m
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
*
After briefly displaying*WAIT* it
will be possible to adjust the
display reading to the correct
value using>
>, ,ENT .
NO
µS / c m
YES
NO
ENT
NO
NO
YES
ENT
YES
YES
NO
NO
*TEMP.1
NO
NO
Briefly*WAIT*
YES
or
TEMP.2
*
YES
YES
YES
NO
NO
NO
After enabling TC it possible to
directly enter the coefficient in
service code 21
YES
YES
NO
NO
YES
NO
NO
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-13
<5. Parameter setting>
5.2.5 Service
MODE
M
cm
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
SETPOINTS
RANGE
SET HOLD
TEMP.
SERVICE
*
YES
After changing the parameter,
the instrument first goes into
reset to load the parameter
specific default values.
NO
NO
M
YES
cm
NO
NO
Example: Service Code 01
Select main parameter
ENT
for SC
YES
for RES
NO
ENT
>
With the >, ,ENT keys
NO
ENT
YES
NO
ENT
ENT
NO
YES
NO
ENT
NO
YES
YES
NO
ENT
NO
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5.3
5-14
<5. Parameter setting>
Notes for guidance in the use of service
coded settings
Don’t set or input service code numbers other than the code numbers defined in this manual.
Setting an undefined service code may make the converter malfunction.
When an undefined service code is input by some accident, push the MODE key and escape
from the service level.
5.3.1
Parameter specific functions
Code 1
*SC.RES
Choose the required parameter, either conductivity or resistivity. If the
parameter is changed the instrument will go into reset to load parameter
specific default values, followed by starting measurement. For all other
service codes the instrument will return to commissioning mode after the
service code setting is finished.
Note: For resistivity a fixed display format is used.
Code 3
*CC1
*CC2
Enter the factory calibrated cell constant mentioned on the nameplate or
on the fixed cable. This avoids the need for calibration. Any value
between 0.005 and 50.0 /cm may be entered. First choose to set the
constant for Cell 1or Cell 2 (*CC1 or *CC2). The cell constant is set as a
combination of a number in the main display, and a factor in the second
line. This gives the necessary resolution and decimal point placement.
Example: To set 0.00987 cm-1 first set factor 0.01xC1 in second line then
set number 0.987 in main display.
*NOTE: If the actual cell constant is changed after a calibration or if
the entered cell constant differs from previous value, then the
message “RESET?” will appear on the second line display. After
pressing “YES” the entered value becomes the new nominal
and calibrated cell constant. After pressing “NO” the update
procedure of the cell constant entry is cancelled.
Code 4
*AIR 1
*AIR 2
*AIR 1 and *AIR 2 select by the “NO” key
To eliminate cable influences on the measurement, a “zero” calibration
with a dry sensor may be done. If a connection box (BA10) and extension
cable (WF10) will be used, “zero” calibration should be done including the
connection equipment.
Code 5
*POL.CK
The DC402G has a polarisation check capable of monitoring the signal
from the cell for distortion of capacitive or polarisation errors. If there is a
problem with the installation or the cell becomes fouled, this will trigger
E1.
For some application this error detecting can cause unwanted signals
during operation. Therefore this code offers the possibility to disable/
enable this check.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-15
<5. Parameter setting>
Code Display
Function
Parameter Specific functions
01
*SC.RES Select main
parameter
02
03
04
05
*CC1/
*CC2
Set cell constant
*AIR 1/
Zero calibration
*AIR 2
*START
*”WAIT”
*END
*POL.CK Polarization check
0609
5.3.2
Function detail
X Y Z
Conductivity
Resistivity
Not used
Press NO to step through choice of
multiplying factors on the second display.
0.10xC
1.00xC
10.0xC
100.xC
0.01xC
Press YES to select a factor
Use >, ^, ENT keys to adjust MAIN digits
Zero calibration with dry cell connected
Press YES to confirm selection
Press YES to start, after briefly displaying
“WAIT”, *END will be displayed
Press YES to complete
Polarization check off
Polarization check on
Not used
0
1
Default values
0
Cond.
0.100
cm-1
0.10xC
1.000
0
1
1
On
Temperature functions
Code 10 *T.SENS
Selection of the temperature compensation sensor. The default selection
is the Pt1000 Ohm sensor, which gives excellent precision with the two
wire connections used. The other options give the flexibility to use a very
wide range of other conductivity/resistivity sensors.
Note: The temperature sensor for both conductivity cells must be the
same.
Code 11 *T.UNIT
Celsius or Fahrenheit temperature scales can be selected to suit user
preference.
Code 12 *T.ADJ 1
First select sensor 1 or sensor 2 for temperature adjustment (*T.ADJ 1 or
*T. ADJ 2).
With the process temperature sensor at a stable known temperature, the
temperature reading is adjusted in the main display to correspond.
The calibration is a zero adjustment to allow for the cable resistance,
which will obviously vary with length.
The normal method is to immerse the sensor in a vessel with water in it,
measure the temperature with an accurate thermometer, and adjust the
reading for agreement.
*T.ADJ 2
Code Display
Function
Temperature Measuring Functions
10
*T.SENS Temperature sensor
11
*T.UNIT
Display in °C or °F
12
*T.ADJ 1
*T.ADJ 2
Calibrate temperature
1319
Function detail
X Y Z Default values
Pt1000
Ni100
PB36
Pt100
8k55
°C
°F
Adjust reading to allow for cable
resistance.
Use >, ^ , ENT keys to adjust value
Not used
0
1
2
3
4
0
1
IM 12D08E02-01E
0
Pt1000
0
°C
None
6th Edition: Feb. 19, 2016-00
5-16
<5. Parameter setting>
5.3.3
Temperature compensation functions
Code 20 *T.R.°C
Choose a temperature to which the measured conductivity (or resistivity)
value must be compensated to. Normally 25°C is used, therefore this
temperature is chosen as default value. Limitations for this setting are: 0
to 100 °C.
If *T.UNIT in code 11 is set to °F, default value is 77°F and the limitations
are 32 - 212°F.
Code 21 *T.C.1/T.C.2 In addition to the procedure described in section 5.2.5 it is possible to
adjust the compensation factor directly. If the compensation factor of
the sample liquid is known from laboratory experiments or has been
previously determined, it can be introduced here, for sensor 1 and/or
sensor 2.
Adjust the value between 0.00 to 3.50 % per °C. In combination with
reference temperature setting in code 20 a linear compensation function
is obtained, suitable when enabled as in section 5.2.6 the matrix may be
selected for each sensor input for all kinds of chemical solutions.
Code 22 *MATRX
The DC402G is equipped with a matrix type algorithm for both inputs,
accurate temperature compensation in various applications. Select
the range as close as possible to the actual temperature/concentration
range. The DC402G will compensate by interpolation and extrapolation.
Consequently, there is no need for a 100% coverage.
If 9 is selected the temperature compensation range for the adjustable
matrix must be configured in code 23. Next the specific conductivity
values at the different temperatures must be entered in codes 24 to 28.
See section 5.2.6 for how to enable MATRIX compensation.
Code 23 *T1, *T2, *T3, Set the matrix compensation range. It is not necessary to enter equal
*T4 & *T5 °C temperature steps, but the values should increase from T1 to T5,
otherwise the entry will be rejected. Example: 0, 10, 30, 60 and 100 °C
are valid values for the T1....T5. The minimum span for the range (T5 T1) is 25 °C.
Code 24-28 *L1xT1 -
*L5xT5
In these access codes the specific conductivity values can be entered for
5 different concentrations of the process liquid; each one in one specific
access code (24 to 28). The table below shows a matrix entering example
for 1 - 15% NaOH solution for a temperature range from 0 - 100 °C.
Notes:
1. In chapter 10 a table is included to record your programmed values. It will make
programming easy for duplicate systems or in case of data loss.
2. Each matrix column has to increase in conductivity value.
3. Error code E4 occurs when two standard solutions have identical conductivity values at the
same temperature within the temperature range.
Table 5.2 Example of user adjustable matrix
Matrix
Code 23
Code 24
Code 25
Code 26
Code 27
Code 28
Temperature
Solution 1 (1%)
Solution 2 (3%)
Solution 3 (6%)
Solution 4 (10%)
Solution 5 (15%)
T1...T5
L1
L2
L3
L4
L5
Example
0 °C
31 mS/cm
86 mS/cm
146 mS/cm
195 mS/cm
215 mS/cm
Example
25 °C
53 mS/cm
145 mS/cm
256 mS/cm
359 mS/cm
412 mS/cm
Example
50 °C
76 mS/cm
207 mS/cm
368 mS/cm
528 mS/cm
647 mS/cm
Example
75 °C
98 mS/cm
264 mS/cm
473 mS/cm
692 mS/cm
897 mS/cm
IM 12D08E02-01E
Example
100 °C
119 mS/cm
318 mS/cm
575 mS/cm
847 mS/cm
1134 mS/cm
6th Edition: Feb. 19, 2016-00
Code Display
Function
Temperature compensation functions
20
*T.R.°C
Set reference temp.
21
*T.C.1
Set temp. coef. 1
22
23
24
25
26
27
28
29
5.3.4
5-17
<5. Parameter setting>
*T.C.2
Set temp. coef. 2
*MATRX
Select matrix
X= Sensor 1
Y= Sensor 2
*T1 °C (°F)
*T2..
*T3..
*T4..
*T5..
*L1xT1
*L1xT2
....
*L1xT5
*L2xT1
*L3xT1
*L4xT1
*L5xT1
Set temp. range
Enter conductivity
values for lowest
concentration
Concentration 2
Concentration 3
Concentration 4
Concentration 5
Function detail
X Y Z Default values
Use >, ^, ENT keys to set value
Adjust compensation factor for mA1
output, if set to TC in section 5.2.5.
Set value with >, ^, ENT keys
Adjust compensation factor for mA2
output, if set to TC in section 5.2.5.
Set value with >, ^, ENT keys
Choose matrix if set to matrix comp. in
section 5.2.5, using >, ^, ENT keys
Matrix selected in section 5.2.6
HCl (cation) pure water (0-80 °C)
Ammonia pure water (0-80 °C)
Morpholine pure water (0-80 °C)
HCl (0-5 %, 0-60 °C)
NaOH (0-5 %, 0-100 °C)
User programmable matrix
Enter 1st (lowest) matrix temp. value
Enter 2nd matrix temp. value
Enter 3rd matrix temp. value
Enter 4th matrix temp. value
Enter 5th (highest) matrix temp. value
Value for T1
Value for T2
25 °C
2.1 %
per °C
2.1 %
per °C
1
2
3
4
5
9
1
2
3
4
5
9
Value for T5
Similar to code 24
Similar to code 24
Similar to code 24
Similar to code 24
Not used
mA output functions
Code 30 *mA
Select 4-20mA or 0-20mA according to associated equipment (recorders,
controllers etc.)
Code 31 *OUTP.F Note: For resistivity measurement, read resistivity in stead of conductivity.
Output mA1
Conductivity linear
(terminals 61&62) Conductivity with 21 point output table. (The table can be
configured to give an output linear to concentration, see example at
the end of this page).
Output mA2
Conductivity linear
(terminals 65&66) Conductivity with 21 point output table.
Temperature linear
PI control on conductivity (analog output control signal with
proportional and integral functions).
Direct or reverse action of the mA control output. Direct gives rising
output with rising measurement. Reverse gives falling output with rising
measurement.
Code 32 *BURN
Diagnostic error messages can signal a problem by sending the output
signals upscale or downscale (22mA or 0/3.5mA). This is called upscale
or downscale burnout, from the analogy with thermocouple failure
signalling of a burned-out or open circuit sensor. In the case of the
DC402G the diagnostics are extensive and cover the whole range of
possible sensor faults.
Code 33 *RG.mA1(2) This function sets the proportional range for the mA output control signal.
The range setting is expressed in % of setpoint.
Code 34 *tI.mA1(2) This function sets the integral time for the mA output control signal
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<5. Parameter setting>
5-18
Code 35-36 *TABL1(2) The table function allows the configuration of an output curve by 21 steps
(intervals of 5%)
The following example shows how the table may be configured to
linearize the output with a W/W% curve. On the next page some other
possibilities are shown.
Code 37 *DAMP
Transmission signal damping (not mA control output).
Conductivity (mS/cm)
Output in %
1.000
100
800
80
600
60
400
40 % Output
200
20
0
0
2 4 6 8 10 12 14 16 18 20 22 24
Conductivity
0
Concentration (% by weight)
Fig. 5.1
Linearization of output Example: 0-25% Sulfuric acid
Table 5.3
Code Output
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
mA 0-20
mA 4-20
4
4.8
5.6
6.4
7.2
8
8.8
9.6
10.4
11.2
12
12.8
13.6
14.4
15.2
16
16.8
17.6
18.4
19.2
20.0
% H2SO4
0
1.25
2.5
3.75
5
6.25
7.5
8.75
10
11.25
12.5
13.75
15
16.25
17.5
18.75
20
21.25
22.5
23.75
25
mS/cm
0
60
113
180
218
290
335
383
424
466
515
555
590
625
655
685
718
735
755
775
791
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-19
<5. Parameter setting>
Code Display
Function
mA output functions
30
*mA
mA output range
31
*OUTP.F
Function detail
mA output functions
Direct/reverse action
(For PI control only)
32
33
34
35
36
37
3839
*BURN
Burn function
*RG.
PI range
mA1(2)
*tI.mA1(2) Integral time
(for PI control)
Output table for mA1
*TABL1
*0%
*5%
*10%
...
...
*90%
*100%
*TABL2
Output table for mA2
*DAMP
Damping time
X Y Z Default values
mA1 = 0-20 mA
mA1 = 4-20 mA
mA2 = 0-20 mA
mA2 = 4-20 mA
Calculated value (Service Code 58)
Meas. conductivity linear
Meas. conductivity table
Meas. temperature
T1-T2
For rising signal, set for increasing mA out.
For rising signal, set for decreasing mA out.
mA 1 No burnout
mA 1 Burnout downscale
mA 1 Burnout upscale
mA 2 No burnout
mA 2 Burnout downscale
mA 2 Burnout upscale
Proportional range for mA control signal
(use >, ^, ENT keys to adjust value)
0
1
0
1
2
3
4
5
6
0
1
2
1.1
0
1
0
1
2
3
4
5
6
0
1
2
Linearization table for mA1 in 5% steps.
The measured value is set in the main
display using the >, ^, ENT keys, for
each of the 5% interval steps.
Where a value is not known, that
value may be skipped, and a linear
interpolation will take place.
Similar to code 35
Damping on mA output 0-120 sec.
Not used
4-20
4-20
1.1
Cond.1/2
(direct)
(reverse)
0.0 No Burn.
No Burn.
0. sec.
% of output range
100
B
C
A
50
D
0
0
Fig. 5.2
10
50
% of conductivity range
100 110
Percentage of mA-output range versus percentage of conductivity range
EXAMPLES:
A = bi-lineair
B = hyperbolic (2 decades)
C = logarithmic (2 decades)
D = linear
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-20
<5. Parameter setting>
Table 5.4 Example of output tables
Output
0%
5%
10%
15%
20%
25%
30%
35%
40%
45%
50%
55%
60%
65%
70%
75%
80%
85%
90%
95%
100%
bi-lin
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
110.0
log 2
1.0
1.3
1.6
2.0
2.5
3.2
4.0
5.0
6.3
7.9
10.0
12.6
15.8
20.0
25.1
31.6
39.8
50.1
63.1
79.4
100.0
log 3
0.10
0.14
0.20
0.28
0.40
0.56
0.79
1.12
1.58
2.24
3.16
4.47
6.31
8.91
12.6
17.8
25.1
35.5
50.1
70.8
100.0
hyp 2
1.00
1.20
1.82
1.90
2.00
3.75
4.80
5.92
7.00
8.31
10.00
11.85
14.00
16.65
19.50
23.80
29.55
36.70
48.50
68.60
100.0
hyp 3
0.10
0.27
0.43
0.61
0.83
1.10
1.36
1.68
2.05
2.49
3.00
3.66
4.33
5.22
6.80
8.25
11.0
14.8
21.8
36.5
100.0
bi-lin = bi-lineair over 2 decades
log 2 = logarithmic over 2 decades
log 3 = logarithmic over 3 decades
hyp 2 = hyperbolic over 2 decades
hyp 3 = hyperbolic over 3 decades
NOTE:
Multiply the values from the table with appropriate factors to get the end-scale value you want.
5.3.5
Contact outputs
Code 40, *S1, *S2
41, 42
*S3, *S4
and 43
Process relays can be set for a variety of alarm and control function.
Digit “Y” sets the control action:
Process alarm is a simple On/Off trip controlled by the high/low setpoint.
Proportional duty cycle control has a pulse width modulation for
proportional dosing with solenoid valves.
Proportional frequency control is used for controlling electrically
positioned valves.
Temperature alarm is an On/Off trip on the measured temperature.
Digit “Z” sets the control parameter:
Alarm on main process
Control on main process
(Main process means conductivity/resistivity depending on the setting of
service code 01).
Digit “X” sets the type of trigger:
Off means that the relay is not active
Low setpoint means that the relay is triggered by a decreasing
measurement.
High setpoint means that the relay is triggered by an increasing
measurement “HOLD” active means that there is maintenance activity in
progress so the measurement is not live.
For *S4 There is the extra possibility to set up for “FAIL” indication.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-21
<5. Parameter setting>
Code Display
Function
Contact Settings
40
*S1
Relay 1 settings
Note: Main process
means cond. or resist.
whichever is set in
code #1
41
*S2
Relay 2 settings
Note: “HOLD” active
relay contact is used
to indicate when the
measuring mode is
interrupted
42
43
*S3
*S4
Relay 3 settings
Relay 4 settings
Note: “FAIL” relay
contact is used to
indicate when the
diagnostics detect a
problem
Function detail
Off
Low setpoint
High setpoint
“HOLD” active
Process alarm
Proportional duty cycle control **
Proportional frequency control **
PI duty cycle **
PI pulse freq. **
USP contact
Calculated value
Meas cond/res. value cell 1
Meas cond/res. value cell 2
Meas temp. value cell 1
Meas temp. value cell 2
Meas temp. T1-T2
Off
Low setpoint
High setpoint
“HOLD” active
Process alarm
Proportional duty cycle control
Proportional frequency control
PI duty cycle **
PI pulse freq. **
USP contact
Calculated value
Meas cond/res. value cell 1
Meas cond/res. value cell 2
Meas temp. value cell 1
Meas temp. value cell 2
Meas temp. T1-T2
Off
Low setpoint
High setpoint
“HOLD” active
Process alarm
Proportional duty cycle control
Proportional frequency control
PI duty cycle **
PI pulse freq. **
USP contact
Calculated value
Meas cond/res. value cell 1
Meas cond/res. value cell 2
Meas temp. value cell 1
Meas temp. value cell 2
Meas temp. T1-T2
Off
Low setpoint
High setpoint
“HOLD” active
Fail alarm
Process alarm
Proportional duty cycle control
Proportional frequency control
PI duty cycle **
PI pulse freq. **
USP contact
Calculated value
Meas cond/res. value cell 1
Meas cond/res. value cell 2
Meas temp. value cell 1
Meas temp. value cell 2
Meas temp. T1-T2
X Y Z Default values
0
1
2
3
0
1
2
3
0
1
2
3
0
1
2
3
4
IM 12D08E02-01E
2.0.1
High
0
1
2
3
4
5
Alarm
0
1
2
3
4
5
Cell 1
1.0.1
0
1
2
3
4
5
Low
Alarm
0
1
2
3
4
5
Cell 1
0.0.0 Off
0
1
2
3
4
5
0
1
2
3
4
5
4.0.0
0
1
2
3
4
5
FAIL
Alarm
0
1
2
3
4
5
6th Edition: Feb. 19, 2016-00
5-22
<5. Parameter setting>
Code 44 *D.TIME
The delay time sets the minimum relay switching time. This function can
be adjusted to give a good alarm function in a noisy process, preventing
the relay from “chattering” or repeatedly switching when the signal is
close to the setpoint.
The hysteresis is the value beyond the setpoint that the measured
value must exceed before the control function will start working. For
conductivity this setting is expressed in % of programmed setpoint value.
*SC.HYS
*T.HYST
*C.HYST
Code 45 *RANGE
Proportional range is the value above (or below) the setpoint that
generates full output in proportional control. This is expressed in % of the
programmed output span.
*PER.
The time period of the overall pulse control cycle (one ON and one OFF
period). See fig 5.4.
*FREQ.
The maximum frequency for the pulse frequency control. See fig 5.5.
Code 46 *tI.CNT
The integral time for the PI control settings.
Code 47 *EXPIR
When a system is set up to control on the relay outputs, the expiry time
can be enabled to warn of an ineffective control. In other words, when
the setpoint is exceeded for more than 15 minutes an error message is
generated. This can mean, for example, that the reagent tank is empty.
*tE.min
Code 48 *SC1
For proportional or PI control on the conductivity level, a working range
*SC1 must be set. When *SC1 is set as an output on mA1 and *SC2 is
set as an output on mA2, code 48 has no relevance. When using mA1 as
mA2 for calculated values or temperature etc. the range for control setting
of the relays should be done in code 48. After activating code 48 *SC1 is
displayed. Press YES to set range *SC1 or NO to go to *SC2.
Code Display
Function
Contacts Settings
44
*D.TIME Delay time
*SC.HYS Process hysteresis
*T.HYST
Hysteresis temp.
*C.HYST
*RANGE
Calculation hysteresis
Proportional range
*PER.
*FREQ.
Duty cycle period
Maximum frequency
46
*tI.CNT
Integral time
47
*EXPIR
Expiry time
48
*tE.mn
*SC1
Set expiry time
Set control range
0%
*100%
Set begin scale
Set end scale
45
49
Function detail
X Y Z Default values
Minimum relay switching time
Minimum change of process value for
relay reset after switching
Minimum temperature change for relay
reset after switching (fig. 5.3)
When proportional control selected in
code 40, 41, 42 or 43
Pulse control On time + Off time (fig. 5.4)
100% value for frequency control (fig.
5.5)
Integral time for relay controls when PI
is set
Warning of ineffective control action On
Warning of ineffective control action Off
Set expiring time using >, ^, ENT keys
Set range for *SC1 (*SC2) for
proportional contact control when *SC1
(*SC2) are not used on mA1 (mA2)
Use >,^, ENT keys to set value
Use >,^, ENT keys to set value
Not used
0.2 sec.
2.0 %
1 °C
0
10.0 %
10 sec.
70 p/m
1
0
IM 12D08E02-01E
0
100
sec.
Off
15 min
6th Edition: Feb. 19, 2016-00
<5. Parameter setting>
5-23
Cond./Resist.
Setpoint
Hys.
LED off
LED on
Delay time
LED off
Delay time
t (sec)
Fig. 5.3
% duty cycle control
10%
90%
100
50
50%
50%
ton
toff
10%
90%
0
Proportional
range
Setpoint
Fig. 5.4
SC
Pulse period
Time
% of output range
Duty cycle control
% controller output
100
Maximum pulse frequency
0.3s
50
0
50 % pulse frequency
Proportional
range
Setpoint
Fig. 5.5
SC
No pulses
Time
% of output range
Pulse frequency control
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5.3.6
<5. Parameter setting>
5-24
User interface
Code 50 *RET.
When Auto return is enabled, the converter reverts to the measuring
mode from anywhere in the configuration menus, when no button is
pressed during the set time interval of 10 minutes.
Code 51 *MODE
The adjustment of the contact setpoints can be setup for operation in the
maintenance mode. (Through the closed front cover).
Code 52 *PASS
Passcodes can be set on any or all of the access levels, to restrict access
to the instrument configuration.
Code 53 *Err.01
Error message configuration. Two different types of failure mode can be
set.
Hard fail gives a steady FAIL flag in the display, and a continuous contact
closure. All the other contacts (controls) are inhibited (except HOLD
contacts), and a Fail signal is transmitted on the outputs when enabled in
code 32.
Soft fail gives a flashing FAIL flag in the display, and the relay contacts
are pulsed. The other contacts (controls) are still functional, and the
controller continues to work normally. The call for maintenance is a good
example of where a SOFT fail is useful.
If set to 1, soft fail gives a flashing FAIL flag in the display and no pulsing
contact.
*SOFT
Code 54 *E5.LIM
*E6.LIM
Limits can be set for shorted and open measurement. Dependent on
the main parameter chosen in code 01, the DC402G will ask for the
absolute conductivity/resistivity (without influence of absolute cell
constant or Temp. comp).
Code 55 *%
For some applications the measured parameter values may be (more
or less) linear to concentration. For such applications it is not needed to
enter an output table, but 0 and 100% concentration values directly can
be set.
Code 56 *DISP
The display resolution is default set to autoranging for conductivity
reading. If a fixed display reading is needed, a choice can be made out of
7 possibilities. For resistivity the default reading is fixed to xx.xx MΩ·cm.
Code 57 *USP
Input 1 and/or Input 2 can be set to give automatic checking for
compliance with the water purity standard set in USP <645> (United
States Pharmacopea directive 23). For more detailed description see
Appendix 10.8.
Code 58 *CALC
A calculation type can be chosen to suit a variety of applications. See
appendix 10.9 for more details. For conductivity a choice can be made
from 7 options. For Resistivity only “differential” or “no calculation” is
possible.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5-25
<5. Parameter setting>
Code Display
User Interface
50
*RET
Auto return
51
*MODE
Mode setup
52
*PASS
53
Function
Passcode
Note # = 0 - 9, where 0
= no passcode, 1=111,
2=333, 3=777, 4=888,
5=123, 6=957, 7=331,
8=546, 9=847
*Err. 1. 1(2) Error setting
*Err. 5. 1(2)
*Err. 6. 1(2)
*Err. 7. 1(2)
*Err. 8. 1(2)
*Err. 13. 1(2)
*Err. 22. 1
*SOFT
54
SOFT FAIL CONTACT
*E5.LIM 1(2) E5 limit setting
*E6.LIM 1(2) E6 limit setting
55
*%
Display mA in w/w%
Set w/w% for range 1
56
*%1
*0%
*100%
*%2
*0%
*100%
*DISP
57
*USP
58
*CALC
USP setting
X=Cell 1 Y= Cell 2
Calculation setting
Set w/w% for range 2
Display resolution
Function detail
X Y Z Default values
Auto return to measuring mode Off
Auto return to measuring mode On
Setpoints in maintenance mode Off
Setpoints in maintenance mode On
Maintenance passcode Off
Maintenance passcode On
Commissioning passcode Off
Commissioning passcode On
Service passcode Off
Service passcode On
Polarisation too high
Soft fail
Polarisation too high
Hard fail
Shorted measurement
Soft fail
Shorted measurement
Hard fail
Open measurement
Soft fail
Open measurement
Hard fail
Temperature sensor open Soft fail
Temperature sensor open Hard fail
Temp. sensor shorted
Soft fail
Temp. sensor shorted
Hard fail
USP <645> limit exceeded Soft fail
USP <645> limit exceeded Hard fail
Control time-out
Soft fail
Control time-out
Hard fail
Pulsing contact
No Pulsing contact
Maximum conductivity value
(Minimum resistivity value)
Maximum conductivity value
(Minimum resistivity value)
mA1-range displayed in w/w% off
mA1-range displayed in w/w% on
mA2-range displayed in w/w% off
mA2-range displayed in w/w% on
Press YES to access 0% value adjustm.
Set 0% output value in w/w%
Set 100% output value in w/w%
Press YES to access 0% value adjustm.
Set 0% output value in w/w%
Set 100% output value in w/w%
Auto ranging display
Display fixed to X.XXX μS/cm or MΩ•cm
Display fixed to XX.XX μS/cm or MΩ•cm
Display fixed to XXX.X μS/cm or MΩ•cm
Display fixed to X.XXX mS/cm or kΩ•cm
Display fixed to XX.XX mS/cm or kΩ•cm
Display fixed to XXX.X mS/cm or kΩ•cm
Display fixed to XXXX mS/cm or kΩ•cm
Disable the E13 (USP limit passed)
Enable the E13 (USP limit passed)
Choose the main parameter for display
Ratio (a/b)
Differential (a-b)
% passage b/a
% rejection (100 (a-b)/a)
Deviation (100 X (b-a)/a)
No calculation SC1 in display
VGB directive 450L, pH calculation
0
1
0
1
0
#
1
0
0.0.0 Off
0
#
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
2
3
4
5
6
7
0
1
0
1
2
3
4
5
6
On
Off
Off
Off
0
#
1
Hard
1
Hard
1
Hard
1
Hard
1
0
Hard
Soft
0
Soft
0
Pulsing
25
0.04
1
1
0.0
mS
kΩ
μS
MΩ
Off
0
1
0
1
Off
0
Auto
0.0
Off/Off
5
No calc.
When set for resistivity measuring, only
selection 1 and 5 are available
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
5.3.7
<5. Parameter setting>
5-26
Communication setup
Please do not use the Code 61 due to the factory use.
Code 61 *HOUR
*MINUT
*SECND
*YEAR
*MONTH
*DAY
5.3.8 General
Code 70 *LOAD
5.3.9
The load defaults code allows the instrument to be returned to the default
set up with a single operation. This can be useful when wanting to change
from one application to another.
Test and setup mode
Code 80 *TEST
Not used
Note: Attempting to change data in this service code, or others in the series 80 and above
without the proper instructions and equipment, can result in corruption of the instrument
setup, and will impair the performance of the unit.
Code Display
Function
Communication
Clock setup
61
*HOUR
*MINUT
*SECND
*YEAR
*MONTH
*DAY
6269
Code Display
General
70
*LOAD
7179
Function detail
Adjust to current date and time using >, ^
and ENT keys
Not used
Function
Load defaults
Code Display
Function
Test and setup mode
80
*TEST
Test and setup
X Y Z Default values
Function detail
X Y Z Default values
Reset configuration to default values
Not used
Function detail
X Y Z Default values
Not used
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
6-1
<6. Calibration>
6. Calibration
6.1
When is calibration necessary?
Calibration of conductivity/resistivity instruments is normally not required, since Yokogawa
delivers a wide range of sensors, which are factory calibrated traceable to NIST standards. The
cell constant values are normally indicated on the top of the sensor or on the integral cable.
These values can be entered directly in service code 03 (section 5.3.1).
If the cell has been subjected to abrasion (erosion or coating) calibration may be necessary.
In the next section two examples are given. Alternatively calibration may be carried out with a
simulator to check the electronics only.
NOTE
During calibration the temperature compensation is still active. This means that the readings are
referred to the reference temperature as chosen in service code 20 (section 5.3.3, default 25 °C).
Calibration is normally carried out by measuring a solution with a known conductivity value at a
known temperature. The measured value is adjusted in the calibration mode. On the next pages
the handling sequence for this action is visualized.
Calibration solutions can be made up in a laboratory. An amount of salt is dissolved in water
to give a precise concentration with the temperature stabilized to the adjusted reference
temperature of the instrument (default 25 °C). The conductivity of the solution is taken from
literature tables or the table on this page.
Alternatively the instrument may be calibrated in an unspecified solution against a standard
instrument.
Care should be taken to make a measurement at the reference temperature since differences in
the type of temperature compensation of the instrument may cause an error.
NOTE
The standard instrument used as a reference must be accurate and based on an identical
temperature compensation algorithm. Therefore the Model SC72 Personal Conductivity Meter of
Yokogawa is recommended.
Typical calibration solutions.
The table shows some typical conductivity values for sodium-chloride (NaCl) solutions which can
be made up in a laboratory.
Table 6.1 NaCl values at 25 °C
Weight %
0.001
0.003
0.005
0.01
0.03
0.05
0.1
0.3
0.5
1
3
5
10
mg/kg
10
30
50
100
300
500
1000
3000
5000
10000
30000
50000
100000
Conductivity
21.4 µS/cm
64.0 µS/cm
106 µS/cm
210 µS/cm
617 µS/cm
1.03 mS/cm
1.99 mS/cm
5.69 mS/cm
9.48 mS/cm
17.6 mS/cm
48.6 mS/cm
81.0 mS/cm
140 mS/cm
NOTE
For resistivity measurement the standard
resistivity units of the calibration solution can be
calculated as follows:
R = 1000/G (kΩ•cm if G = μS/cm)
Example:
0.001% weight
R = 1000/21.4 = 46.7 kΩ•cm
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
6.2
6-2
<6. Calibration>
Calibration procedure
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
MΩ.cm
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
YES
NO
MODE
CONTACTS
S1
S2
ENT
S3
FAIL/S4
YOKOGAWA
MODE
MΩ.cm
YES
NO
Note: See section 4-5 for full
display access details
NO
MΩ.cm
YES
MΩ.cm
YES
NO
MΩ.cm
Put the sensor in standard
solution. Press YES.
ENT
Set the value
using the >, , ENT key.
MΩ.cm
>
ENT
MΩ.cm
Select the flashing digit with the > key.
Increase its value by pressing the key
>
ENT
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
MΩ.cm
YES
The calibration is now complete. Put the
sensor back in the process and press
YES to return to the measuring mode.
IM 12D08E02-01E
NO
6th Edition: Feb. 19, 2016-00
6.3
6-3
<6. Calibration>
Calibration with HOLD active
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
mS / c m
MEASURE
CAL 1
CAL 2
DISPLAY 1
DISPLAY 2
HOLD
YES
NO
MODE
CONTACTS
S1
S2
ENT
S3
FAIL/S4
YOKOGAWA
MODE
HOLD
mS / c m
YES
NO
Note: See section 4-5 for full
display access function
NO
HOLD
mS / c m
YES
NO
YES
HOLD
mS / c m
HOLD
YES
NO
mS / c m
HOLD
Put the sensor in standard
solution. Press YES.
ENT
mS / c m
HOLD
Set the value
using the >, , ENT key.
mS / c m
>
ENT
>
Select the flashing digit with the > key.
Increase its value by pressing the key
When the correct value is displayed,
press ENT to enter the change.
HOLD
mS / c m
After briefing displaying WAIT,
the CAL.END message appears.
YES
NO
The calibration is now complete. Put the
sensor back in the process and press YES.
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
HOLD
mS / c m
YES
IM 12D08E02-01E
NO
6th Edition: Feb. 19, 2016-00
Blank Page
7-1
<7. Maintenance>
7. Maintenance
7.1
Periodic maintenance for the DC402G
converter
The DC402G converter requires very little periodic maintenance. The housing is sealed to IP65
(NEMA 4X) standards, and remains closed in normal operation. Users are required only to make
sure the front window is kept clean in order to permit a clear view of the display and allow proper
operation of the pushbuttons. If the window becomes soiled, clean it using a soft damp cloth or
soft tissue. To deal with more stubborn stains, a neutral detergent may be used.
NOTE
Never used harsh chemicals or solvents. In the event that the window becomes heavily stained
or scratched, refer to the parts list (Chapter 9) for replacement part numbers.
When you must open the front cover and/or glands, make sure that the seals are clean and
correctly fitted when the unit is reassembled in order to maintain the housing’s weatherproof
integrity against water and water vapor. The measurement otherwise may be prone to problems
caused by exposure of the circuitry to condensation.
The DC402G instrument contains a lithium cell to support the clock function when the power is
switched off. This cell needs to be replaced at 5 yearly intervals (or when discharged). Contact
your nearest Yokogawa service centre for spare parts and instructions.
7.2
Periodic maintenance of the sensor
NOTE
Maintenance advice listed here is intentionally general in nature. Sensor maintenance is highly
application specific.
In general conductivity/resistivity measurements do not need much periodic maintenance. If the
DC402G indicates an error in the measurement or in the calibration, some action may be needed
(ref. chapter 8 trouble shooting). In case the sensor has become fouled an insulating layer
may be formed on the surface of the electrodes and consequently, an apparent increase in cell
constant may occur, giving a measuring error. This error is:
2x
Rv
Rcel
x 100 %
where:
Rv = the resistance of the fouling layer
Rcel = the cell resistance
NOTE
Resistance due to fouling or to polarization does not effect the accuracy and operation of a
4-electrode conductivity measuring system.
If an apparent increase in cell constant occurs cleaning the cell will restore accurate
measurement.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
7-2
<7. Maintenance>
Cleaning methods
1. For normal applications hot water with domestic washing-up liquid added will be effective.
2. For lime, hydroxides, etc., a 5 ...10% solution of hydrochloric acid is recommended.
3. Organic foulings (oils, fats, etc.) can be easily removed with acetone.
4. For algae, bacteria or moulds, use a chlorous solution (bleaching liquid*).
* Never use hydrochloric acid and bleaching liquid simultaneously. The very poisonous gas chlorine will result.
7.3
Fuse Replacement
This section describes how to replace the power supply fuse.
WARNING
• Use only a fuse of the specified current, voltage and type ratings to prevent fire.
• Power to the instrument must be turned off at the external breaker before replacing the fuse.
After replacing the fuse, install a high-voltage shield plate before turning the power on.
• Do not short circuit the fuse holder.
n Specified Ratings of the Power Supply Fuse
The power supply fuse (recommended fuse) used in the DC402G is as follows.
l 115 V AC version
Rated breaking current: 35 A or 10 times the rated current, whichever is greater (low breaking)
Maximum rated voltage: 250 V
Maximum rated current: 200 mA
Compliance:
UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law
Part number:
A1105EF
l 230 V AC version
Rated breaking current: 35 A or 10 times the rated current, whichever is greater (low breaking)
Maximum rated voltage: 250 V
Maximum rated current: 100 mA
Compliance:
UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law
Part number:
A1103EF
l How to replace the fuse
1.
Before replacing the fuse, turn off power to the instrument at the external breaker.
2.
Remove the instrument cover and the high-voltage shield plate.
3.
Remove the cover from the fuse holder by pulling out by hand.
4.
Remove the fuse and install a new, recommended or equivalent fuse on the holder.
5.
Put the fuse cover back on the holder securely.
6.
Install the high-voltage shield plate.
It is recommended that the fuse be replaced every two years even if it has not blown.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<7. Maintenance>
7-3
Fuse Cover
Fuse Holder
Shield Plate
Figure 7.1 How to replace the fuse
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page
<8. Troubleshooting>
8-1
8. Troubleshooting
The DC402G is a microprocessor-based analyzer that performs continuous self-diagnostics
to verify that it is working correctly. Error messages resulting from faults in the microprocessor
systems itself are few. Incorrect programming by the user can be corrected according to the limits
set in the following text.
In addition, the DC402G also checks the sensor to establish whether it is still functioning within
specified limits.
The DC402G makes a distinction among diagnostic findings. All errors are signaled by the FAIL
area in the display. Only faults in the measuring circuits activate the FAIL contact switching.
What follows is a brief outline of some of the DC402G troubleshooting procedures, followed by a
detailed table of error codes with possible causes and remedies.
n Diagnostics
l Off-line checks
The DC402G converter incorporates a diagnostic check of the adjusted cell constant value at
calibration. If the adjusted value stays within 80 - 120 % of the nominal value set in service code
03, it is accepted. Otherwise, the unit generates an error (E3).
The DC402G also checks the temperature compensation factor while performing manual
temperature compensation as described in section 5.2.5. If the the TC factor stays within 0.00%
to 3.50% per °C, it is accepted. Otherwise, E2 will be displayed.
l On-line checks
The DC402G performs several on-line checks to optimize the measurement and to indicate a
fault due to the fouling or polarization of the connected sensor. The fault will be indicated by the
activation of the FAIL-contact, the lighting of the LED and the flag in the display.
During measurement the DC402G adjusts the measuring frequency to give the best conditions
for the actual value being measured. At low conductivity there is a risk of error due to the
capacitive effects of the cable and the cell. These are reduced by using a low measuring
frequency. At high conductivity the capacitive effects become negligible and errors are more likely
to be caused by polarisation or fouling of the cell. These errors are decreased by increasing the
measuring frequency.
At all values the DC402G checks the signal from the cell to search for distortion which is typical of
capacitive or polarisation errors. If the difference between pulse front and pulse rear is > 20% an
error E1 will be displayed and the FAIL alarm will be activated. In service code 05 it is possible to
turn this check on and off.
The following error message table gives a list of possible problems that can be indicated by the
DC402G.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<8. Troubleshooting>
8-2
Table 8.1 Error Codes
Code
Error description
E1
Polarization detected on cell
E2
E3
E4
E5
E6
E7
E8
E9
E10
E13
E15
E17
E18
E19
E20
E21
E22
E24
Possible cause
Sensor surface fouled
Conductivity too high
Incorrect field calibration of TC
Suggested remedy
Clean sensor and calibrate
Replace sensor
Temperature coefficient out of
Re-adjust
limits (exceeds 0-3.5%/°C range)
Set calculated TC
Calibration out of limits
Calibrated value differs more
Check for correct sensor
than +/- 20 % of nominal value
Check for correct unit (μS/cm,
programmed in code 03.
mS/cm, kΩ•cm or MΩ•cm)
Repeat calibration
Matrix compensation error
Wrong data entered in 5x5 matrix Re-program
Conductivity too high or resistivity Incorrect wiring
Check wiring (Sec. 3.6)
too low (Limits set in service code Internal leakage of sensor
Replace sensor
54)
Defective cable
Replace cable
Conductivity too low or resistivity
Dry sensor
Immerse sensor
too high (Limits set in service code Incorrect wiring
Check wiring (Sec. 3.6)
54)
Defective cable
Replace cable
Temperature sensor open
Process temperature too high or Check process
(Pt1000 : T > 250°C or 500°F)
too low
Check model code sensor
(Pt100/Ni100 : T > 200°C or 400°F) Wrong sensor programmed
Check connections and cable
(8k55 : T < -10°C or 10°F)
Incorrect wiring
(PB36 : T < -20°C or 0°F)
Temperature sensor shorted
Process temperature too high or Check process
(Pt1000/Pt100/Ni100 :
too low
Check model code sensor
T < -20°C or 0°F)
Wrong sensor programmed
Check connections and cable
(8k55/PB36 : T > 120°C or 250°F) Incorrect wiring
Air set impossible
Too high zero due to cable
Replace cable
capacitance
EEPROM write failure
Fault in electronics
Try again, if unsuccessful contact
Yokogawa
USP Limit exceeded
Poor water quality
Check-ion exchangers
Cable resistance influence to
Cable resistance too high
Check cable
temperature exceeds +/- 15°C
Corroded contacts
Clean and reterminate
Wrong sensor programmed
Reprogram
Output span too small
Incorrect configuration by user
Reprogram
Table values make no sense
Wrong data programmed
Reprogram
Programmed values outside
Incorrect configuration by user
Reprogram
acceptable limits
All programmed data lost
Fault in electronics
Contact Yokogawa
Very severe interference
Checksum error
Software problem
Contact Yokogawa
Alarm activation time exceeded
Process control not effective
Check control equipment
within set time
Adjust value in code 47
Calculation out of limits
Incorrect configuration
Check settings
Extreme process condition
Check process
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
9.
<9. Spare Parts>
9-1
Spare Parts
See Customer Maintenance Parts List.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page
10-1
<10. Appendix>
10. Appendix
10.1
User setting for non-linear output table
(code 31, 35 and 36)
Output signal value
10.2
%
Output
mA
0-20
mA
4-20
000
005
010
015
020
025
030
035
040
045
050
055
060
065
070
075
080
085
090
095
100
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
00.4
04.8
05.6
06.4
07.2
00.8
08.8
09.6
10.4
11.2
0.12
12.8
13.6
14.4
15.2
0.16
16.8
17.6
18.4
19.2
20.0
User entered matrix data (code 23 to 28)
Medium:
Code 23 Temperature T1...T5
Code 24
Solution 1
L1
Code 25
Solution 2
L2
Code 26
Solution 3
L3
Code 27
Solution 4
L4
Code 28
Solution 5
L5
Medium:
Code 23 Temperature T1...T5
Code 24
Solution 1
L1
Code 25
Solution 2
L2
Code 26
Solution 3
L3
Code 27
Solution 4
L4
Code 28
Solution 5
L5
T1 data
T2 data
T3 data
T4 data
T5 data
T1 data
T2 data
T3 data
T4 data
T5 data
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
10-2
<10. Appendix>
10.3
Matrix data table (user selectable in code
22)
Matrix, Solution
HCL-p (cation)
selection 1
Ammonia-p
selection 2
Morpholine-p
selection 3
Hydrochloric Acid
selection 4
Temp (°C)
0
10
20
30
40
50
60
70
80
Data 1
0 ppb
0.0116 µS
0.0230 µS
0.0419 µS
0.0710 µS
0.1135 µS
0.173 µS
0.251 µS
0.350 µS
0.471 µS
Data 2
4 ppb
0.0228 µS
0.0352 µS
0.0550 µS
0.085 µS
0.129 µS
0.190 µS
0.271 µS
0.375 µS
0.502 µS
Data 3
10 ppb
0.0472 µS
0.0631 µS
0.0844 µS
0.115 µS
0.159 µS
0.220 µS
0.302 µS
0.406 µS
0.533 µS
Data 4
20 ppb
0.0911 µS
0.116 µS
0.145 µS
0.179 µS
0.225 µS
0.286 µS
0.366 µS
0.469 µS
0.595 µS
Data 5
100 ppb
0.450 µS
0.565 µS
0.677 µS
0.787 µS
0.897 µS
1.008 µS
1.123 µS
1.244 µS
1.373 µS
0
10
20
30
40
50
60
70
80
0 ppb
0.0116 µS
0.0230 µS
0.0419 µS
0.0710 µS
0.113 µS
0.173 µS
0.251 µS
0.350 µS
0.471 µS
2 ppb
0.0229 µS
0.0337 µS
0.0512 µS
0.0788 µS
0.120 µS
0.178 µS
0.256 µS
0.356 µS
0.479 µS
5 ppb
0.0502 µS
0.0651 µS
0.0842 µS
0.111 µS
0.149 µS
0.203 µS
0.278 µS
0.377 µS
0.501 µS
10 ppb
0.0966 µS
0.122 µS
0.150 µS
0.181 µS
0.221 µS
0.273 µS
0.344 µS
0.439 µS
0.563 µS
50 ppb
0.423 µS
0.535 µS
0.648 µS
0.758 µS
0.866 µS
0.974 µS
1.090 µS
1.225 µS
1.393 µS
0
10
20
30
40
50
60
70
80
0 ppb
0.0116 µS
0.0230 µS
0.0419 µS
0.0710 µS
0.113 µS
0.173 µS
0.251 µS
0.350 µS
0.471 µS
20 ppb
0.0272 µS
0.0402 µS
0.0584 µS
0.0851 µS
0.124 µS
0.181 µS
0.257 µS
0.357 µS
0.481 µS
50 ppb
0.0565 µS
0.0807 µS
0.108 µS
0.140 µS
0.181 µS
0.234 µS
0.306 µS
0.403 µS
0.528 µS
100 ppb
0.0963 µS
0.139 µS
0.185 µS
0.235 µS
0.289 µS
0.351 µS
0.427 µS
0.526 µS
0.654 µS
500 ppb
0.288 µS
0.431 µS
0.592 µS
0.763 µS
0.938 µS
1.12 µS
1.31 µS
1.52 µS
1.77 µS
0
15
30
45
60
1%
65 mS
91 mS
114 mS
135 mS
159 mS
2%
125 mS
173 mS
217 mS
260 mS
301 mS
3%
179 mS
248 mS
313 mS
370 mS
430 mS
4%
229 mS
317 mS
401 mS
474 mS
549 mS
5%
273 mS
379 mS
477 mS
565 mS
666 mS
1%
31 mS
53 mS
76 mS
97.5 mS
119 mS
2%
61 mS
101 mS
141 mS
182 mS
223 mS
3%
86 mS
145 mS
207 mS
264 mS
318 mS
4%
105 mS
185 mS
268 mS
339 mS
410 mS
5%
127 mS
223 mS
319 mS
408 mS
495 mS
Sodium Hydroxide
selection 5
0
25
50
75
100
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
10.4
<10. Appendix>
10-3
Sensor Selection
10.4.1 General
The inputs of the DC402G transmitter are freely programmable for ease of installation. Standard
2-electrode type sensors with a cell constant of 0.100/cm and a Pt1000 temperature sensor,
need no special programming. The DC402G indicates a fault with a signal in the display field if
there is a mismatch of sensors in the connection.
10.4.2
Sensor selection
The DC402G is pre/programmed to accept standard 2-electrode sensors with a Pt1000
temperature sensor. The DC402G is universally compatible with all 2- and 4-electrode type of
sensors with a cell constant within the range of 0.008/cm to 50.0/cm.
10.4.3
Selecting a temperature sensor
The DC402G reaches its highest accuracy when used with a PT1000 temperature sensor. This
may influence the choice of the conductivity/resistivity sensor, as in most cases the temperature
sensor is integrated in the conductivity/resistivity sensor.
10.5
Setup for other functions
● Contact Outputs
Alarms, trips and proportional control are all possible with the relay outputs, and the
configuration is by Service codes 40 - 49. In addition, FAIL alarm is available.
● Current Outputs
Transmission signals for the measured parameters and control signals can be set up in
service codes 30 - 39.
● Diagnostic checks
Polarization check and checks on the calibrated cell constant and the adjusted Temperature
Coefficient, are included in the DC402G.
Note: On the next page a reference list for the configuration of the DC402G is shown.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<10. Appendix>
10.6
10-4
User setting table
FUNCTION
SETTING DEFAULTS
Parameter specific functions
01 *SC.RES 0
SC
03 *0.10xC
0.10xC
Factor
C.C.1
1.000
/cm
C.C.2
1.000
/cm
04 *AIR
Perform zero calibration
05 *POL.CK 1
On
Temperature measuring functions
10 *T.SENS 0
Pt1000
11 *T.UNIT
0
°C
12 *T.ADJ 1
None
*T.ADJ 2
None
Temperature compensation functions
20 *T.R.°C
25
°C
21 *T.C.1
2.1
%/°C
*T.C.2
2.1
%/°C
22 *MATRX
None, see 5.2.5
23 *T1°C
T. range
See sep. table, §10.2
24 *L1xT1
Cond. C1 See sep. table, §10.2
25 *L2xT1
Cond. C2 See sep. table, §10.2
26 *L3xT1
Cond. C3 See sep. table, §10.2
27 *L4xT1
Cond. C4 See sep. table, §10.2
28 *L5xT1
Cond. C5 See sep. table, §10.2
mA outputs
30 *mA
1.1
both 4-20mA
31 *OUTP.F 1.1
SC1 & SC2.
*D/R
0
Reverse (control)
32 *BURN
0.0
both off
33 *RG.mA2 prop band only for PI control
34 *tI.mA2
integ. time only for PI control
35 *TABL1
21 pt table see code 31, §10.1
36 *TABL2
21 pt table see code 31, §10.1
37 *DAMP
0 sec
Output damping
Contacts
40 *S1
2.0.1
high SC1 Al.
41 *S2
1.0.1
low SC2 Al.
42 *S3
0.0.0
43 *S4
4.0.0
FAIL
44 *D.TIME 0.2
sec
*SC.HYS 2.0
% setpoint value
*T. HYST 1
°C
*C.HYST 0
% setpoint value
45 *RANGE 1
% output span
*PER
10
sec
*FREQ
70
p/min
46 *tI.CNT
100
sec
47 *EXPIR
0
off
*tE.min
15
min
48 *SC1
0
100µS/cm control range
0
20MΩ•cm
*SC2
0
100µS/cm control range
0
20MΩ•cm
User Interface
50 *RET
1
on
USER SETTINGS
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<10. Appendix>
FUNCTION
51 *MODE
52 *PASS
53 *Err.1.1/1.2
*Err.5.1/5.2
*Err.6.1/6.2
*Err.7.1/7.2
*Err.8.1/8.2
*Err.13.1/13.2
*Err.22
54 *E5.LIM1
*E6.LIM1
*E5.LIM2
*E6.LIM2
55
56
*%
*%1
*0%
*100%
*%2
*0%
*100%
*DISP
0
0.0.0
1
1
1
1
1
0
0
25
0.04
1
1
25
0.04
1
1
0.0
0
(2)
0.0
5
57 *USP
58 *CALC
Communication
61 *HOUR
General
70 *LOAD
reset
Test and setup mode
80 *TEST
SETTING DEFAULTS
off
all off
hard fail
hard fail
hard fail
hard fail
hard fail
soft fail
soft fail
mS/cm
kΩ•cm
µS/cm
MΩ•cm
mS/cm
kΩ•cm
µS/cm
MΩ•cm
Both off
10-5
USER SETTINGS
Auto ranging (SC)
(xx.xx MΩ•cm) (RES)
Both disabled
SC1
defaults
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
10-6
<10. Appendix>
10.7
Configuration checklist for DC402G
Standard Configuration
Options
Measured Variable(s)
primary inputs
Conductivity (SC) and Temp
conductivity range
0.000 µS/cm - 100 µS/cm
Resistivity i.s.o Conductivity
any span within 0.000µS/cm 1999mS/cm
Choice out of fixed µS/cm or
mS/cm
any span within 0.000kΩ•cm 999MΩ•cm
Auto ranging or other fixed
values
any span in -20 ... +250
Fahrenheit
code 01
“range”
0- 20 mA or 4- 20 mA
0- 20 mA or 4- 20 mA
SC, Resistivity, Temp, Table,
PI control
(4) freely programmable
code 30
code 30
code 31
µS/cm, mS/cm, kΩ•cm,
MΩ•cm,
temp, PI control, HOLD, FAIL
time: 0- 200 s; hyst 0.1- 100%
time out alarm
PI on contacts or mA output
–
code 40- 43
–
TC, %w/w, °C, mA1, mA2, CC,
TR, REL
burn low (3.5)/ high (22) on
mA1/ mA2
for maint/ comm./ serv level
enable or disable
setpoint adj
code 60
“display”
enabled
active
active
enable or disable
code 05
“calibrate”
“temp.”
SC40, SX42
Pt1000
0.100/cm
SC8SG, L&N compatible
Ni100, PB36, Pt100, 8k55
from 0.008/cm up to 50.0/cm
code 10
code 10
code 03
conductivity units
Auto ranging µS/cm - mS/cm
resistivity range
0 - 19.99 MΩ•cm
resistivity units
MΩ•cm
temperature range
temperature unit
Outputs
analog output
second output
output allocation
0 - 100 °C
Celsius
contact outputs
S1= high at 100 µS/cm
S2= low at 100 µS/cm
S4= FAIL
mS/cm and FAIL
contact allocation
4- 20 mA for SC
4- 20 mA for Temp
SC and Temp
contact variables
add. contact functions
control functions
digital outputs
Communication
digital interface
variables on display
dead time= 0.2 s; hyst= 0.1%
none
none
none (Don’t use.)
burn out
disabled
password protection
autoreturn
add. function in MAINT
Diagnostics
check on polarization
check on cell constant
check on TC coefficient
Compatibility
SC sensor
temperature sensor
cell constant
Special Features
temperature calibration
zero calibration
temp. comp.
disabled
return to measure in 10 min.
disabled
HOLD during
maintenance
contact during HOLD
soft fail alarm
Reference for change
disabled (Don’t use.)
µS/cm (mS/cm) and temp
code 56
“range”
code 56
“range”
code 11
“setpoint”
code 40. 41, 42, 43
code 44
code 47
code 45, 46, 34, 33
code 60
code 32
code 52
code 50
code 51
none
adjustment +/- 15 °C
none
adjustment -1 µS/cm
acc to NaCl tables IEC 60746-3 NaCl, manual TC, matrix code
20 - 28
disabled
hold last or hold fix
code 12
code 04
“temp.”,
disabled
disabled
code 40- 42
code 53
possible on S1,S2 or S3
possible for E1, E5...E13, E22
IM 12D08E02-01E
“Hold”
6th Edition: Feb. 19, 2016-00
10.8
10-7
<10. Appendix>
USP <645> Water Purity Monitoring
l What is USP?
USP stands for United States Pharmacopeia and it is responsible for issuing guidelines for the
pharmaceutical industry. Implementing these guidelines is highly recommended for companies
wishing to market drugs in the US. This means that USP is important for pharmaceutical
companies worldwide. This new USP, aims at the replacement of 5 antiquated laboratory tests by
simple conductivity analysis.
l How have we accomplished this in DC402G?
1.
In Software Rev. 1.1, (and later versions) we have defined an Error Code: E13. This is
independent of what range the customer is measuring or what temperature compensation
method he is using for water quality monitoring. When the display shows E13, then the
water quality exceeds the USP limits, and the FAIL contact closes to signal that the system
needs urgent attention.
2.
We have introduced uncompensated conductivity in the DISPLAY menu. In the LCD display
the user can read the temperature and the raw conductivity to compare his water quality
with the USP table.
3.
We have added a USP function to the contact allocation. All 4 contacts can be selected as
USP alarms. The contact closes when the USP limit is approached. It is possible to adjust
a safety margin for the water quality. For example, if the temperature is 64 °C. and the
safety margin is adjusted for 20%, then the contact closes at 0.8 x 2.2 μS/cm. = 1.76 μS/cm.
(2.2μS/cm is the USP limit at 64°C).
4.
We have kept all the DC402G functionality: It is even possible to have the mA Output and
Display readings in resistivity units. Most users will have very good water quality and in the
resistivity mode they will have better resolution on the recorder or DCS. The readings are
simply the reciprocal values of the conductivity values. In the example mentioned above the
contact will close at an uncompensated resistivity of 1/1.76 μS/cm. = 0.568 MΩ•cm.
l Setting up DC402G for USP
First enable USP in service code 57. Change the setting from 0 (default) to 1 (enabled).
This activates uncompensated conductivity in the display menu. The E13 alarm feature is
also enabled. For E13 the FAIL alarm (S4) is triggered when the uncompensated conductivity
exceeds the relevant value in the graph.
USP warning alarms can be selected for any of the 4 relay outputs.
Service codes 40 to 43 are for these alarms.
Set to 2.5.1 for input one, or 2.5.2 for input two.
The setpoint is the safety margin in %, and is set as described in sections 5.1.3 & 5.2.2.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<10. Appendix>
10-8
Conductivity limit as a
function of Temperature
3.5
3
μS/cm
2.5
2
1.5
1
0.5
0
0
25
50
75
Temperature in ºC
100
Fig. 10.1
10.9
WHAT IS DUAL CONDUCTIVITY?
Dual cell conductivity is a precise, comparative measurement.
The DC402G receives inputs from two conductivity cells located at different points in the process
and compares them according to one of six programmed user-selectable formula
- Ratio (a/b)
- Differential or linear difference (a-b)
- Percent passage (b/ax 100)
- Percent rejection {(a-b)/a x100}
- Deviation {(b/a) a x100}
- pH calculation acc. VGB directive 450L
The output signal corresponds directly to the formulas. The DC402G also displays the absolute
value of each cell on a second display line, as desired by the user.
The unit displays all values in conductivity units (μS/cm or mS/cm), percentage (%), resistivity
(MΩ•cm) or pH.
The pH value of the (boiler) water/condensate is calculated from the conductivity value before
and after the cation filter according the VGB directive 450L.
pH as function of differential Conductivity
12
11
pH
10
9
8
7
0.1
1.0
10
SC(before) - SC(cation) / 3
100
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
10-9
<10. Appendix>
This pH calculation requires two conductivity values which are compensated to 25ºC.
Normally matrix compensation for morpholine/ammonia and cation are used (service code 22
should be set to 2.1 or 3.1). Consequently the pH value displayed corresponds to 25ºC.
RETURN LINE
DUAL CELL
CONDUCTIVITY
COOLING
TOWER
MAKE-UP WATER
RATIO
A/B
B
A
BLOW DOWN LINE
COOLING WATER SYSTEM
Fig. 10.2
Ratio output controls blow-down of cooling tower based on concentration factor.
DIFFERENTIAL A-B
A
B
WASHING PROCESS
Fig. 10.3
Differential output water flow to optimize washing efficiency.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<10. Appendix>
10-10
DEVIATION
B-A
X 100
A
B
SEA WATER
DIFFERENTIAL
TEMPERATURE (TA - TB)
SEA WATER
A
CONDENSATE
Fig. 10.4
HEAT EXCHANGER
Deviation output alarms directly after leakage in the heat exchanger.
SAMPLE
COOLER
A
pH = f(A, B)
B
CONDENSATE
OR STEAM
CATION EXCHANGER
Fig. 10.5
pH output as a function of two conductivity values before and after the cation exchanger.
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
<10. Appendix>
10-11
l Error codes
Code
Error description
E1
Polarization detected on cell
E2
E3
E4
E5
E6
E7
E8
E9
E10
E13
E15
E17
E18
E19
E20
E21
E22
E24
Possible cause
Sensor surface fouled
Conductivity too high
Incorrect field calibration of TC
Suggested remedy
Clean sensor
Replace sensor
Temperature coefficient out of limits
Re-adjust
(exceeds 0 to 3.5%/ºC range)
Set calculated TC
Calibration out of limits
Calibrated value differs more
Check for correct sensor
than +/- 20 % of nominal value
Check for correct unit (µS/cm,
programmed in code 03.
mS/cm, kΩ•cm or MΩ•cm)
Repeat calibration
Matrix compensation error
Wrong data entered in 5x5 matrix Re-program
Conductivity too high or resistivity too low Incorrect wiring
Check wiring (Sec. 3.6)
(Limits set in service code 54)
Internal leakage of sensor
Replace sensor
Defective cable
Replace cable
Conductivity too low or resistivity too high Dry sensor
Immerse sensor
(Limits set in service code 54)
Incorrect wiring
Check wiring (Sec. 3.6)
Defective cable
Replace cable
Temperature sensor open
Process temperature too high or Check process
(Pt1000 : T > 250°C or 500°F)
too low
(Pt100/Ni100 : T > 200°C or 400°F)
Wrong sensor programmed
Check model code sensor
(8k55 : T < -10°C or 10°F)
Incorrect wiring
Check connections and cable
(PB36 : T < -20°C or 0°F)
Temperature sensor shorted
Process temperature too high or Check process
(Pt1000/Pt100/Ni100 : T < -20°C or 0°F) too low
(8k55/PB36 : T > 120°C or 250°F)
Wrong sensor programmed
Check model code sensor
Incorrect wiring
Check connections and cable
Air set impossible
Too high zero due to cable
Replace cable
capacitance
EEPROM write failure
Fault in electronics
Try again, if unsuccessful
contact Yokogawa
USP Limit exceeded
Poor water quality
Check-ion exchangers
Check cable
Cable resistance influence to
Cable resistance too high
Clean and reterminate
temperature exceeds +/- 15°C
Corroded contacts
Reprogram
Wrong sensor programmed
Output span too small
Incorrect configuration by user
Reprogram
Table values make no sense
Wrong data programmed
Reprogram
Programmed values outside acceptable Incorrect configuration by user
Reprogram
limits
All programmed data lost
Fault in electronics
Contact Yokogawa
Very severe interference
Checksum error
Software problem
Contact Yokogawa
Alarm activation time exceeded
Process control not effective
Check control equipment
within set time
Adjust value in code 47
Calculation out of limits
Incorrect configuration
Check settings
Extreme process condition
Check process
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page
Customer
Maintenance
Parts List
Model DC402G [Style: S2]
Dual Cell Conductivity and Resistivity
Converter
5
13
15
4
3
7
2
8
10
9
14
11
16
Item
1
12
Part No.
Qty
Description
K9664CL
K9215DB
K9663MC
K9672AV
1
1
1
1
K9672HB
K9672HC
1
1
Cover Assembly
Hinge pin
Flat Cable
Digital Assembly
Analog Assembly
For 115 V AC
For 230 V AC
K9432AH
K9215DJ
K9663MM
1
1
1
1
A1105EF
A1103EF
1
1
Cable Gland Assembly 6 units
Fuse (Time Lag Fuse)
For 115 V AC Power, 0.200 A,T
For 230 V AC Power, 0.100 A,T
10
11
12
A1565EF
K9672BJ
K9664DH
1
1
1
Fuse Holder
Bracket Assembly
Stainless steel screw assembly to fix cover
13
14
K9664DK
1
K9672BG
K9672BH
K9664AR
1
1
1
K9171SU
K9316AF
1
1
Screw Assembly
Bracket Assembly for Analog Board
For 115 V AC Power
For 230 V AC Power
Bracket Assembly for Digital Board
Adapter Assembly
For G1/2 screw when /AFTG specified (5 units).
For 1/2NPT screw when /ANSI specified (5 units).
1
2
3
*4
*5
7
8
9
15
*16
Housing
Polyurethane Coating
Epoxy-polyester Coating
* Do not exchange these parts. Call serviceman.
All Rights Reserved, Copyright © 2007, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 12D08E02-02E
1st Edition : Mar. 2007 (YK)
2nd Edition : Sep. 2012 (YK)
Pipe/Wall Mounting Hardware
(Option Code : /U)
Panel Mounting Hardware
(Option Code : /PM)
7
2
8
3
3
9
4
5
6
1
3
Sun Protection Cover
(Option Code: /H3, /H4)
10
Item
1
2
3
4
5
Part No.
K9171SS
Y9608KU
D0117XL-A
K9171SY
K9171SX
Qty
1
4
1
1
1
Description
Mounting Set (/U)
Screw
U-Bolt Assembly
Plate
Bracket
6
7
8
9
K9171ST
Y9520LU
K9171SW
Y9608KU
1
2
2
4
Mounting Set (/PM)
Screw
Bracket
Screw
K9664CA
K9664CC
1
1
Sun Protection Cover
Cabon steel (/H3)
Stainless steel (/H4)
10
CMPL 12D08E02-02E
2nd Edition : Sep. 2012 (YK)
i
Revision Information
Title
: Model DC402G Converter for Dual Cell Conductivity and Resistivity [Style: S2]
Manual No. : IM 12D08E02-01E
Feb. 2016/6th Edition
EN 61010-2-201 addition, etc.
Aug. 2015/5th Edition
Review all for a document software change.
Note addition for CSA safety standards, etc. (Pages iii, 1-1, 2-3, 2-4, 2-5, Chapter 3, 7-2)
Delete chapter 11.
Feb. 2014/4th Edition
Page 2-3, 3-3 Safety standard is revised.
Appendix 2 Quality Inspection added.
PREFACE, Addition of “How to dispose the batteries”.
Page Appendix 2 11-1 to 11-4 QIS 12D08E02-01E, QIC-12D08E02-01 revised (some change of
withstanding voltage test).
CMPL 12D08E02-02E revised to 2nd edition (some of P/N changed).
Oct. 2007/3rd Edition
Appendix 2 Quality Inspection added.
Jul. 2007/2nd Edition
Page 2-3: Some of E. Housing and K. Safety and EMC conforming standards description revised
(addition of CSA certified); Page 3-3: Terminal label of Figure 3-4 revised, CSA added to WARNING;
Page 3-4: The torque to tighten frontplate screws added; Page 3-9: Terminal label of Figure 3-11
revised; Page 5-14: Some caution added to 5-3. Notes for guidance in the use of service code
settings; Page 5-15: Code 12 description page moved; Page 5-17: Code 22 *MATRX description
corrected; Page 5-25: Default value of code 58 added
Mar. 2007/1st Edition
Newly published
IM 12D08E02-01E
6th Edition: Feb. 19, 2016-00
Blank Page