Yokogawa Model IR250 Infrared Gas Analyzer User's Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s Manual Model IR250 Infrared Gas Analyzer IM 11G02Q01-01EN IM 11G02Q01-01EN 3rd Edition i u PREFACE We are grateful for your purchase of Yokogawa’s Infrared Gas Analyzer, Model: IR250. Read this instruction manual carefully. The related documents are as follows. General Specifications: GS 11G02Q01-01EN User’s Manual: IM 11G02Q01-01EN IM 11G02Q01-51EN IM 11G00E01-01EN (this manual) Communication Functions (MODBUS) ZX8D Zirconia Oxygen Analyzer The “EN” in the document number is the language code, meaning English WARNING This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent explosion, fire or other serious accidents. CAUTION • For installation, observe the rule on it given in the instruction manual and select a place where the weight of gas analyzer can be endured. Installation at an unsuited place may cause turnover or fall and there is a risk of injury. • During installation work, care should be taken to keep the unit free from cable chips or otherforeign objects. Otherwise, it may cause fire, trouble or malfunction of the unit. WARNING In piping, the following precautions should be observed. Wrong piping may cause gas leakage. If the leaking gas contains a toxic component, there is a risk of serious accident being induced. Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring. • Connect pipes correctly referring to the instruction manual. • Exhaust should be led outdoors so that it will not remain in the locker and installation room.Exhaust from the analyzer should be relieved in the atmospheric air in order that an unnecessary pressure will not be applied to the analyzer. Otherwise, any pipe in the analyzer may be disconnected to cause gas leakage. • For piping, use a pipe and a pressure reducing valve to which oil and grease are not adhering. If such a material is adhering, a fire or the like accident may be caused. CAUTION • Wiring work must be performed with the main power set to OFF to prevent electric shocks. • Enforce construction of class-D grounding wire by all means. If the specified grounding construction is neglected, a shock hazard or fault may be caused. • Wires should be the proper one meeting the ratings of this instrument. If using a wire which cannot endure the ratings, a fire may occur. • Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating may cause fire. Media No.IM 11G02Q01-01EN 3rdEdition :May 31, 2017 (YK) All Rights Reserved Copyright © 2016, Yokogawa Electric Corporation IM 11G02Q01-01EN 3rd Edition : May. 31, 2017-00 ii WARNING • Do not smoke nor use a flame near the gas analyzer. Other wise, a fire may be caused. • Do not allow water to go into the gas analyzer. Otherwise, hazard shock or fire in the instrument may be caused. WARNING • Purge not only inside of IR250 but all measuring gas lines with zero gas sufficiently, when you provide maintenance or inspection on IR250 with its cover or door open. Otherwise it may cause hazardous accidents such as gas leakage, fire and explosion. CAUTION • Do not operate the analyzer for a long time with its cover or door left open. Otherwise, dust, foreign matter, etc. may stick on internal walls, thereby causing faults. CAUTION • Be sure to observe the following for safe operation avoiding the shock hazard and injury. • Replacement parts such as a maintenance part should be disposed of as incombustibles. For details, follow the local ordinance. l Inspection of product On receipt of the product, inspect the package and note it has no damage caused by the delivery. Confirm the specification of the product comforms with your order and accessories are all delivered. Check the product code labeld on a model plate is correct. See 2.2 Model and Suffix Code and 2.3 Standard Accessories IM 11G02Q01-01EN 3rd Edition : May. 31, 2017-00 iii u Safety Precautions n Safety, Protection, and Modification of the Product • This manual is intended for the following personnel : • Engineers responsible for installation, wiring, and maintenance of the equipment • Personnel responsible for normal daily operation of the equipment. • In order to protect the system controlled by the product and the product itself and ensure safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product. • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this instrument may be impaired. • If any protection or safety circuit is required for the system controlled by the product or for the product itself, prepare it separately. • Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables. • Modification of the product is strictly prohibited. • The following safety symbols are used on the product as well as in this manual. WARNING This symbol indicates that an operator must follow the instructions laid out in this manual in order to avoid the risks for the human body and health including risk of injury, electric shock, or fatalities. or the damages to instruments. The manual describes what special care the operator must take to avoid such risks. CAUTION This symbol indicates that the operator must refer to the instructions in this manual in order to prevent the instrument (hardware) or software from being damaged, or a system failure from occurring. The following are signal words to be found only in our instruction manuals. CAUTION This symbol gives information essential for understanding the operations and functions. NOTE This symbol indicates information that complements the present topic. n Notes on Handling User’s Manuals • Please hand over the user’s manuals to your end users so that they can keep the user’s manuals on hand for convenient reference. • Please read the information thoroughly before using the product. • The purpose of these user’s manuals is not to warrant that the product is well suited to any particular purpose but rather to describe the functional details of the product. • No part of the user’s manuals may be transferred or reproduced without prior written consent from YOKOGAWA. • YOKOGAWA reserves the right to make improvements in the user’s manuals and product at any time, without notice or obligation. • If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your local distributor. • This manual is an essential part of the product ; keep it in a safe place for future reference. IM 11G02Q01-01EN 3rd Edition : May. 31, 2017-00 iv n Drawing Conventions Some drawings may be partially emphasized, simplified, or omitted, for the convenience of description. Some screen images depicted in the user’s manual may have different display positions or character types (e.g., the upper / lower case). Also note that some of the images contained in this user’s manual are display examples. n How to dispose the batteries This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This directive is only valid in the EU. Batteries are included in this product. Batteries incorporated into this product cannot be removed by yourself. Dispose them together with this product. When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office. Do not dispose them as domestic household waste. Battery type: Poly-carbon monofluoride Lithium battery NOTICE: The symbol above means they shall be sorted out and collected as ordained in ANNEX II in DIRECTIVE 2006/66/EC. n Authorized Representative in EEA The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2, 3825 HD Amersfoort, The Netherlands). n Trademark policy • All names of company, brand of product used in this manual are registered trademarks. • TM or R to signify brand or trademarks are not used in this manual. IM 11G02Q01-01EN 3rd Edition : May. 31, 2017-00 v  After-Sales Warranty  Do not modify the product.  Yokogawa warrants the product for the period stated in the pre-purchase quotation. Yokogawa shall conduct defined warranty service based on its standard. When the customer site is located outside of the service area, a fee for dispatching the maintenance engineer will be changed to the customer.  During the warranty period, for repair under warranty carry or send the product to the local sales representative or service office. Yokogawa will replace or repair any damaged parts and return the product to you.  Before returning a product for repair under warranty, provide us with the model name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.  If we replace the product with a new one, we won’t provide you with a repair report.  In the following cases, customer will be charged repair fee regardless of warranty period.  Failure of components which are out of scope of warranty stated in instruction manual.  Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa did not supply.  Failure due to improper or insufficient maintenance by user.  Failure due to misoperation, misuse or modification which Yokogawa does not authorize.  Failure due to power supply (voltage, frequency) being outside specifications or abnormal.  Failure caused by any usage out of scope of recommended usage  Any damage from fire, earthquake, a storm and flood, lightning, disturbance, riot, warfare, radiation and other natural changes.  Yokogawa does not warrant conformance with the specific application at the user site. Yokogawa will not bear direct/indirect responsibility for damage due to a specific application.  Yokogawa will not bear responsibility when the user configures the product into systems or resells the product.  Maintenance service and supplying repair parts will be covered for five years after the production ends. For repair this product, please contact the nearest sales office described in this instruction manual. IM 11G02Q01-01EN 3rd Edition : May. 31, 2017-00 Blank Page i Model IR250 Infrared Gas Analyzer IM 11G02Q01-01EN 3rd Edition CONTENTS u PREFACE........................................................................................................i u Safety Precautions.......................................................................................iii  After-Sales Warranty.....................................................................................v 1. OVERVIEW................................................................................................. 1-1 2. 3. 4. 5. 1.1 Name and description....................................................................................... 1-2 1.2 Sampling system configuration....................................................................... 1-3 1.3 Measurement Principle..................................................................................... 1-5 SPECIFICATIONS...................................................................................... 2-1 2.1 Standard Specification...................................................................................... 2-1 2.2 Model and Suffix Code...................................................................................... 2-7 2.3 Standard Accessories..................................................................................... 2-13 2.4 External dimensions........................................................................................ 2-14 2.5 NO2 / NO Converter........................................................................................... 2-19 INSTALLATION.......................................................................................... 3-1 3.1 Where to install equipment............................................................................... 3-1 3.2 How to install...................................................................................................... 3-2 3.3 Piping.................................................................................................................. 3-3 3.4 Sampling............................................................................................................. 3-5 3.5 Wiring ................................................................................................................. 3-7 3.6 Solenoid valve drive signal for calibration................................................... 3-16 3.6.1 Manual calibration............................................................................. 3-16 3.6.2 Auto calibration................................................................................. 3-17 OPERATION............................................................................................... 4-1 4.1 Warm-up operation and regular operation...................................................... 4-1 4.2 Keys and function on Panel . ........................................................................... 4-2 4.3 Mode and screens.............................................................................................. 4-3 4.4 General operation.............................................................................................. 4-6 SETUP (User Mode)................................................................................... 5-1 5.1 Switch Ranges................................................................................................... 5-2 5.2 Calibration Parameters..................................................................................... 5-3 5.2.1 Calibration Value................................................................................. 5-4 5.2.2 About ZERO Calibration..................................................................... 5-5 5.2.3 About Calibration Range..................................................................... 5-6 IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 ii 5.2.4 6. 7. 8. Auto Calibration Component / Range................................................. 5-7 5.3 Alarm Setting...................................................................................................... 5-8 5.4 Setting Auto calibration................................................................................... 5-11 5.5 Setting of Auto Zero Calibration..................................................................... 5-16 5.6 Peak Alarm Setting.......................................................................................... 5-20 5.7 Parameter Setting............................................................................................ 5-22 5.7.1 Output Hold....................................................................................... 5-23 5.7.2 Response time, Average Period, Backlight, Contrast...................... 5-26 5.7.3 Maintenance Mode........................................................................... 5-28 CALIBRATION............................................................................................ 6-1 6.1 Zero Calibration.................................................................................................. 6-1 6.2 Span Calibration................................................................................................ 6-2 MAINTENANCE.......................................................................................... 7-1 7.1 Daily check and maintenance procedures..................................................... 7-1 7.2 Replacement of power fuse.............................................................................. 7-2 7.3 Cleaning of sampling cell.................................................................................. 7-3 7.3.1 Disassembly and assembly of sampling cell (Pipe cell, Block cell)................................ 7-3 7.3.2 How to clean sampling cell................................................................. 7-6 ERROR MESSAGE ................................................................................... 8-1 Revision Information................................................................................................i IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 1. 1-1 <1. OVERVIEW> OVERVIEW This product measures the concentration of NO, SO2, CO2, CO and CH4 contained in sampling gas on the principle that different atomic molecules have an absorption spectrum in the wave band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law. Since this instrument incorporates a compact O2 sensor, it allows measuring up to 4 components simultaneously by using the built-in O2 sensor (up to 3 components if O2 measurement is excluded). Furthermore, use of a microprocessor and large sized liquid crystal display realizes improvement of operability, accuracy and multi- functions. This product is being purposely designed to be a part of large-scale fixed installations (LSFI.) Intended uses of the product being part of LSFI are • Combustion gas measuring / control of industrial furnaces / incinerators. • Steel gas analysis (blast furnace, converter, thermal treatment furnace, sintering ((pellet equipment)), coke furnace) • Chemical analysis (petroleum refinery plant, petrochemical plant, gas generation plant) • Environmental management and others IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 1.1 1-2 <1. OVERVIEW> Name and description (1)Power switch (2) Panel (9)Flow checker (1) Power switch (only for IR250-C) (7)Sampling gas outlet (12)Com.Connector (13) Analog input (5) Purge gas inlet (RS-485) connector(A/O) (6)Sampling gas inlet (12) External input connector (10)Terminal block 1 to 10 from the left (4)Power (5)Purge gas supply inlet (3)Fuse (2) panel (11)External input connector(A/I) (14)Digital i/o connector (from the bottom DIO1,DIO2, DIO3) (6)sampling gas inlet (7)Sampling gas outlet (in IR250-C, shared for reference gas) (8)Reference gas inlet (only for IR250-C) (3)Fuse (4)Power supply (4)Power terminal M4 Terminal block type (Power cable code: -NN) (2)Panel Mounting plate for wall mounting (Code”-C”) (16)COMP2(second component)I/O terminal (15)COMP1(1st Component) I/O terminal (1)Power switch (3)Fuse (4)Power supply terminal (7)Sampling gas outlet (6)Sampling gas inlet Wiring port (Rubber grommets) (5)Purge gas inlet Name Description Name Description Checks flow rate of sampling gas (1) Power Switch ON/OFF the analyzer (9) Flow checker (2) Display/Operation panel LCD, set functions (10)Terminal block (3) Fuse (11) External input connector connecting to the output of external O2 analyzer (4) Power supply (switch) connect to power supply (12) communication connector RS-485 connector (5) Purge gas inlet connect to purge gs (13) Analog output connector (D-sub 25 pin) connector for analog output (6) Sampling gas inlet connect to measuring gas (14)Digital I/O tube Connector for digital I/O (7) sampling gas outlet connect to exhaust line (15) COMP1 I/O I/O for 1st component. (8) Reference gas inlet fuse inside (16) COMP2 I/O I/O for 2nd component IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 1-3 <1. OVERVIEW> 1.2 Sampling system configuration The figure below shows a general configuration of sampling system. The system configuration may vary depending on sample gas. Consult with Yokogawa l Measurement of sample gas with high moisture content (2°C saturation level or NO, SO2, CO (0-200l ppm) measurement (Excluding IR250-C) (9)Flowmeter (10) Secondary filter (1)Primary filter (3)pump (12)3-way solenoid valve sample gas inlet (11)NO2/NO Converter for NOx (7)Solenoid valve (3)pump atmosphere (7) * (4)Electric dehumidifier (2)Safety drain trap (7) (7) Gas Analyzer (IR250) (5)Drain pot O2 Analyzer (External) (For zioconia) (6) Ball valve (8) Standard gas for zero calibration Drain Exhaust(atmospheric pressure) Standard gas for span calibration * : A dehumidifier must be used for NO, SO2, and CO (0-200 ppm) measurements.(approx. equal to 2°C saturation) F03.ai l In IR250-C the measurement of sample gas with high moisture content (2°C saturation level or above) (10)Secondary filter *3 (13)Reference gas purifier Air filter (3)pump Reference gas (Humidified N2 or atmosphere) (1)Primary filter (3)Pump (9)Flowmeter (11)NO2/NO Gas Analyzer Converter (IR250-C) for NOx measurement (7)Solenoid valve Sample gas *1 (2)Safety drain trap (4)Electric dehumidifier Exhaust (atomspheric pressure) *2 O2 Analyzer (7) (7) (7)Solenoid valve Exhaust (atomspheric pressure) (5) Drain pot (6)Ball valve Drain (8)Standard gas Standard gas for zero calibration for span calibration *1: Make the electric dehumidifier removes the moisture so that the temperature of sample gas goes below 5°C. The water concentration should be equalized in reference gas and sample gas. *2: Use No2/NO converter in case of measuring NOx. *3: The concentration of measuring component in a reference gas determines whether a reference gas purifier is required or not. If the concentration level is lower than 0.1 % FS, a reference gas purifier is not required. If atmospheric air is used as reference gas and the measuring range is lower level than the description below, the use of a reference gas purifier is recommended. However, for measuring CO2, atmospheric air should not be used as reference gas. NO analyzer : 0-100 ppm or less, SO2 analyzer : 0-50 ppm or less, CO analyzer : 0-1000 ppm or less. Consult us on reference gas purifier suited to each component. F03-1.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 1-4 <1. OVERVIEW> Item Description Item Description (1) Primary filter Removes dust and mist (8) Standard gas for calibration Used for zero/span calibration (2) Safety drain trap Separates and dischages drain (9) Flowmeter Adjust and monitors sample gas flow rate. (3) Pump Sucks in sample gas. (10) Secondary filter (membrane filter) Removes fine dust. (4) Electric dehumidifier Dehumidifies sample gas (11) NO2/NO converter Converter NO2 gas into NO gas (5) Drain pot Collects discharged water from dehumidifier (12) 3-way solenoid valve Used for introducing humidified gas (6) Ball valve Used for discharging drain (13)Reference gas purifier Eliminates particles out of the reference gas (7) Solenoid Valve Used for switching sampling and calibration lines IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 1.3 1-5 <1. OVERVIEW> Measurement Principle l Infrared gas analyzer (CO2, CO, CH4, SO2, NO) Front expansion chamber Gas inlet Infrared-ray light source Gas outlet Rear expansion chamber Detector Motor Mass flow sensor Measuring cell Chopper Preamplifier Signal processing and calculation block Indication Output F01.ai l Consecutive zero correction with IR250-C Sample and reference gas are measured at each respective timing. Measured gas concentration is output by converting the difference value between the sample and reference gas into a concentration value. Continuous monitoring on the difference value between sample and reference gas leads to an elimination of influence from interference component and zero-drift Gas Analyzer Sample gas Reference gas SV1 com NC Solenoid valve com NO Infrared light source sample cell Exhaust gas Detector motor NC SV2 chopper NO NC: Normal close NO: Normal open Gas flow l Paramagnetic oxygen analyzer (O2) Permanent magnets Mirror Measuring cell Gas inlet Magnetic field Gas outlet Preamplifier LED Permanent magnets Photodiode Indication Output Signal processing and calculation block F02.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Blank Page 2-1 <2.SPECIFICATIONS> 2. SPECIFICATIONS 2.1 Standard Specification Measurement principle: CO2, CO, CH4 SO2 NO; Non-dispersive infrared method Single light source-single beam O2; Paramagnetic type (built-in), or zirconia type (external) Measurable gas components and measuring ranges: Component NO SO2 CO2 CO CH4 O2 (Built-in paramagnetic) O2 (Dedicated Zirconia) • • • • IR250-A Min.range Max.range 0-500 ppm 0-5000 ppm 0-500 ppm 0-5000 ppm 0-500 ppm 0-100 vol% 0-200 ppm 0-100 vol% 0-1000 ppm 0-100 vol% 0-5 vol% 0-100 vol% IR250-B Min.range Max.range ― ― 0-500 ppm 0-5000 ppm 0-500 ppm 0-100 vol% 0-500 ppm 0-100 vol% 0-1000 ppm 0-100 vol% ― ― IR250-C Min.range Max.range 0-50 ppm 0-5000 ppm 0-50 ppm 0-5000 ppm 0-50 ppm 0-5000 ppm 0-50 ppm 0-5000 ppm ― ― 0-5 vol% 0-100 vol% IR250-D Min.range Max.range 0-200 ppm 0-5000 ppm 0-200 ppm 0-10 vol% 0-100 ppm 0-100 vol% 0-200 ppm 0-100 vol% 0-500 ppm 0-100 vol% 0-5 vol% 0-100 vol% 0-5 vol% ― 0-5 vol% 0-5 vol% 0-25 vol% ― 0-25 vol% 0-25 vol% IR250-A, IR250-C, IR250-D are capable of measuring up to 4 components including O2. 1 or 2 measuring ranges per component. Measuring range ratio: IR250-A, ≤1:10 (except built-in paramagnetic O2 analyzer) ≤1:20 (built-in paramagnetic O2 analyzer) IR250-B, ≤1:10 IR250-C, IR250-D ≤1:10 (except built-in paramagnetic O2 analyzer) ≤1:20 (built-in paramagnetic O2 analyzer) For measurable components and possible combinations of measuring ranges, see pages 2-18 Display: Digital, 4-digit LCD with LED backlight • Instantaneous value of each component • Instantaneous value after O2 correction (only in CO, SO2, NO with O2 measurement) • Average value after O2 correction (only in CO, SO2, NO with O2 measurement) • Average O2 value (Only with O2 measurement) Analog output signal: 4 to 20 mA DC or 0 to 1 V DC, (isolated from ground wire or internal circuit, nonisolated between output cables) . Analog output corresponds one-to-one with measured value indication. Permissible load resistance; 550 Ω max. for 4 to 20 mA DC 100 kΩ min. for 0 to 1 V DC Max points of IR250-A; 7 points IR250-B; 4 points IR250-C, IR250-D; 11 points See page 3-15 “Measurable components and corresponding channle numbers”. Analog input signal (IR250-A, IR250-C. IR250-D): For signal input from external O2 sensor* Input section is not isolated. Signal requirement: Signal from Yokogawa’s zirconia type O2 sensor (Model ZX8D) 0 to 1 V DC of full scale * External O2 sensor should be purchased separately. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-2 <2SPECIFICATIONS> Contact output: Model IR250-A IR250-C IR250-D Contacts 1a relay contact 1c relay contact 1c relay contact Contact Capacity 250 V AC/2 A resistive load 24 V DC/1 A resistive load Max. number of output points 1a relay 13 points 1c relay 5 points 15 points Isolation of the relay contacts and the internal circuit All relay contacts are isolated mutually and from the internal circuit. Output signal Instrument error, Calibration error, Range identification, Auto calibration status, Solenoid valve drive signal for automatic calibration, Pump ON/OFF*1, High / Low limit alarm*1, Peak alarm*2 * 1 : * 2 : High / Low limit alarm on 4 component (including O2) analyzer of IR250-C, IR250-D have 3 points of output signal. applies only to IR250-A. Contact input: Model IR250-A IR250-D Non-voltage input*3 Voltage input contact Applied voltage ― 12 to 24 V DC Required max.voltage ― 15 mA Max of input points 7 points 9 points Isolation of the relay contacts Mutual contacts ; non isolation and the internal circuit Internal circuit: photocoupler isolation Input *3: *4: IR250-C Contacts Isolation with photocoupler Remote range changeover, auto calibration remote start, remote (output) hold, average value reset*4 When it is ON: 0 V DC, OFF: 5 V DC, ON: 5 mA. applies only to IR250-C, IR250-D Note: See 3-10 I/O Wiring Diagram for further information. Consult YOKOGAWA if you need contact input / output for IR250-B. Power supply: Voltage rating; 100 to 240 V AC Allowable range; 85 to 264 V AC Frequency; 50/60 Hz Maximum voltage rating; 100 VA Inlet; Conform to EN60320 Protection Class l (When IR250-A, IR250-C, IR250-D use other than “-NN” as power cable) Operating conditions: Ambient temperature; Ambient humidity; Storage conditions: Ambient temperature; Ambient humidity; -5 to 45°C (when Measurable component code -C, -G, -H, -P is specified and you run the analyzer with electric power of larger than 200 V, the temperature is 40 °C max) 95 % RH max., non-condensing -20 to 60°C 90%RH max., non-condensing Dimensions (H × W × D): IR250-A; 177 × 483 × 480 mm (with slide rail for 19-inch rack mounting type) IR250-B; 484 × 294 × 170 mm IR250-C,IR250-D; 133× 483 × 382 mm IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Weight: 2-3 <2.SPECIFICATIONS> Approx. 11 kg (IR250-A, IR250-C, IR250-D, ), Approx. 7 kg (IR250-B) Finish color: Front panel; Casing; Enclosure: Off-white (Munsell 10Y7.5/ 0.5 equivalent) steel-blue Steel casing, for indoor use Material of gas-contacting parts: Gas inlet/outlet; SUS304 Sample cell; SUS304 / polychloroprene Infrared-ray transmitting window; CaF2 Internal tubing; polyvinyl chloride, Fluoropolymer (PTFE), polypropylene Gas inlet/outlet: Rc1/4 or 1/4 NPT internal thread Purge gas flow rate: approx. 1 L/min (when necessary) Safety and EMC conforming standards: Only IR250-A, IR250-C or IR250-D as CE marking compliance product Select the suffix code of Power cable : -NN or -EU. Installation altitude; 2000 m or less Pollution degree; 2 (Note) Installation category; II (Note) ● Installation category, so called overvoltage category, specifies impulse withstanding voltage. Category II is for electrical equipment. ●  ollution degree indicates the degree of existence of solid, liquid, gas or other P inclusions which may reduce dielectric strength. Degree 2 indicates the normal indoor environment. ; EN61010-1 ; EN61326-1 Class A, Table 2 (for Industrial application) EN61326-2-3, EN610003-2, EN61000-3-3 EMC (RCM) of Australia, New Zealand, EN61326-1 Class A Safety * EMC * Note: This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only. n Standard Functions Output signal hold: Output signals is held during manual and auto calibrations by activation of holding (turning on its setting). Indication of instantaneous values are not held. The values to be held are the last ones indicated just before calibration mode starts or selected setting value. Indicated values are not held. Remote output hold: (not standard function in IR250-B) Output signal is held at any preset value or at the last value just before contact input signal is received while voltage is input to the terminals. Holding is maintained while the voltage is input to the terminals. Indication of instantaneous values are not held. Range changeover: Set a changeover among Auto, Manual, Remote. Only selected mode is effective during the operation. Manual ; Key operation switches ranges IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-4 <2SPECIFICATIONS> Auto ; When a measured value level is higher than 90 % of full scale concentration for the first range, the range switches from the first range to the second one. When a measured value level is lower than 80 % of full scale concentration for the first range, the range switches from the second range to the first one. Remote ; (not standard function in IR250-B) When the contact input is active, a remote range switch is available. Remote range switch is active only when the first range applies, and when it is not, the second range applies Range identification signal: (not standard function in IR250-B) The current measuring range is identified by a contact position. IR250-A,  hen the first range is selected, the contact output terminals are W closed, and when the second range is selected, they are open for each component. IR250-C, IR250-D,  hen the first range is selected, the NO side is closed, and W when the second range is selected, NC side is closed for each component. Auto calibration: (not standard function in IR250-B) Auto calibration is carried out periodically at the preset cycle. When a standard gas cylinder for calibration and a solenoid valve for opening/ closing the gas flow line are prepared externally by the customer, calibration will be carried out with the solenoid valve drive contacts for zero calibration and each span calibration turned on/ off sequentially at the set auto calibration timing. Auto calibration cycle setting; Auto calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). Gas flow time setting; The time for flowing each calibration gas in auto calibration is set. Settable within 60 to 599 seconds (in increments of 1 second) Auto calibration remote start: (not standard function in IR250-B) Auto calibration is carried out only once according to an external input signal. Calibration sequence is settable in the same way as the cyclic auto calibration. Calibration starts by opening the auto calibration remote start input terminals after shortcircuiting for 1.5 seconds or longer. Auto calibration is started when the contacts open. Auto zero calibration: (not standard function in IR250-B) Auto zero calibration is carried out periodically at the preset cycle. This cycle is independent of “Auto calibration” cycle. When zero calibration gas and solenoid valve for opening/closing the calibration gas flow line are prepared externally by the customer, zero calibration will be carried out with the solenoid valve drive contact for zero calibration turned on/off at the set auto zero calibration timing. Auto zero calibration cycle setting; Auto zero calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-5 <2.SPECIFICATIONS> Gas flow time setting: The time for flowing zero gas in auto zero calibration is set. Settable 60 to 900 seconds (in increments of 1 second) High/low limit alarm: (not standard function in IR250-B) Alarm contact output turns on when measurement value reaches the preset high or low limit alarm value Contacts close when the instantaneous value of each channel exceeds the high alarm limit value or falls below the low alarm limit value. Instrument error contact output: Contact signals are output at occurrence of analyzer error. Contacts close at occurrence of analyzer error No. 1, 3 or 10. Calibration error contact output: (not standard function in IR250-B) Contacts close at occurrence of manual or auto calibration error (any of errors No.4 to 9). Auto calibration status contact output: (not standard function in IR250-B) Contacts close during auto calibration. Pump ON/OFF contact output: (available only in IR250-A) During measurement, contacts close. While calibration gas is flowing, contacts open. Contacts are connected in power supply of pump, and stop the sample gas while calibration gas is flowing. n Optional Functions (Specify with suffix code.) O2 correction (when /K is selected): Conversion of measured CO, SO2 and NO gas concentrations into values at reference O2 concentration. Correction formula: C = 21 – On x Cs 21 – Os Where: C: ample gas concentration after O2 correction Cs: Measured concentration of sample gas Os: Measured O2 concentration (limit set 1 to 20 % O2, initial value : 20 % ) On: R  eference O2 concentration (value changeable by setting 0 to 19 % O2, initial value : 12 % ) The upper limit value of the fractional part in this calculation is 4. The result of calculation is indicated and output in an analog output signal. Average value after O2 correction (when /K is selected): The result of O2 correction or instantaneous O2 value can be output as an average value in the determined period of time. Moving average method is used. Sampling is carried out at intervals of 30 seconds. (Output is updated every 30 seconds. It is updated to the averaged value of the most recent elapsed averaging time period Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4 hours (in increments of 1 hour). Average value reset (when IR250-C specifies /K or IR250-D specifies /K): The above-mentioned output of average value is started from the initial state by opening the average value resetting input terminals after short-circuiting for 1.5 seconds or longer. Output is reset by short-circuiting and restarted by opening. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-6 <2SPECIFICATIONS> CO concentration peak alarm (when /A is selected): (available only for CO - O2 measurement with IR250-A) Alarm output turns on according to the preset concentration and count. Whenever the instantaneous value of CO exceeds the preset concentration value, count increments. If the count exceeds the preset value in one hour, the alarm contacts close. Communication function (when /C is selected): RS-485C (9 pins D-sub) Half-duplex bit serial Start-stop synchronization MODBUS protocol Contents; Read/write parameters, measurement concentration and instrument status n Performance Model Repeatability Linearity Zero drift Spam drift Response time (90% FS response) IR250-A IR250-B ±0.5% FS (less than ±0.5% FS 50 ppm: 1% FS) ±1% FS ±2% FS / week ±2% FS / week IR250-C ±0.5% FS IR250-D ±0.5% FS ±0.5% FS/ week ±2% FS / week(O2 measurement) ±2% FS / week (When measurement NO, SO2 of less than 500 ppm uses auto zero.calibration ) ±2% FS / week within10 to 30 sec.(including sample replacement) Influence from other gases: If the following gases are contained in sample gas, a measurement result may be interfered. For detailed solution, consult our representatives. IR250-A, IR250-C, IR250-D; Interference component CO 1000 ppm CO2 15% NO analyzer ≤1.0%FS ≤1.0%FS (the range of 500ppm: ≤2%FS) SO2 analyzer CO2 analyzer ≤1.0%FS ≤1.0%FS ≤1.0%FS ― H2O 20°C saturation ― ― ≤1.0%FS H2O 2°C saturation CH4 1000 ppm ≤2%FS ≤20ppm ― ≤1.0%FS ≤2%FS ― CO analyzer ― ≤1.0%FS (the range of 200ppm: ≤2%FS) ≤1.0%FS (the range of 500ppm: ≤2%FS) ≤2%FS ≤1.0%FS CH4 analyzer ≤1.0%FS ≤1.0%FS ≤1.0%FS ― ― Note: Range of 0-200 ppm is excluded. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-7 <2.SPECIFICATIONS> IR250-B; Interference component NO analyzer SO2 analyzer CO2 analyzer CO analyzer CH4 analyzer CO 1000ppm CO2 15% ― ― ― ≤1.0%FS ≤1.0%FS ― ≤1.0%FS ― ≤1.0%FS ― ≤1.0%FS ≤1.0%FS ≤1.0%FS ≤1.0%FS ≤1.0%FS H2O 2°C saturation CH4 1000ppm ― ≤2%FS ― ≤2%FS ― ― ≤20ppm ≤1.0%FS ≤1.0%FS ― H2O 20°C saturation 2.2 Model and Suffix Code [Style : S1] Model IR202 Suffix code --------------------------------------- Specification - - - - - - - - - - - - - - Infrared Gas Analyzer ----------------------------------------------------- 19 inch rack mounting type*1 Wall mounting type*2 Low-concentration 19inch rack mounting type*3 19 inch rack mounting D-Sub terminal type ------------------------------------------------------------------------------------------------------------------------------------------------------------- NO SO2 NO+SO2 CO CO2 CO2+CO NO+CO NO+SO2+CO CH4 CH4+CO CO2+CH4 CO2+CO+CH4 ----------------------------------------------------- Without O2 Analyzer External Zirconia type O2 Sensor (ZX8D) *5 External O2 Analyzer*6 Built-in paramagnetic type O2 Sensor Measuring range -NN for NO *7,*8 -01 -02 -03 -04 -07 ------------------------------------------------------------------------------- Not available Min. 0-50 ppm / Max. 0-500 ppm Min. 0-100 ppm / Max. 0-1000 ppm Min. 0-200 ppm / Max. 0-2000 ppm Min. 0-250 ppm / Max. 0-2500 ppm Min. 0-500 ppm / Max. 0-5000 ppm Measuring range for -NN -01 SO2*7,*8 -02 -03 -04 -06 -07 -08 -09 -10 -11 -12 ------------------------------------------------------------------------------------------------------------------------------------------------------------- Not available Min. 0-50 ppm /Max. 0-500 ppm Min. 0-100 ppm /Max. 0-1000 ppm Min. 0-200 ppm /Max. 0-2000 ppm Min. 0-250 ppm /Max. 0-2500 ppm Min. 0-500 ppm /Max. 0-2500 ppm Min. 0-500 ppm /Max. 0-5000 ppm Min. 0-1000 ppm /Max. 0-5000 ppm Min. 0-1000 ppm/ Max. 0-1 % Min. 0-2000 ppm/ Max. 0-2 % Min. 0-5000 ppm/ Max. 0-5 % Min. 0-1 % / Max. 0-10 % Measuring range for CO2 -NN *7,*8 -01 -02 -03 -04 -07 -09 -10 -11 -12 -13 -14 -15 -16 --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- Not available Min. 0-50 ppm / Max. 0--500 ppm Min. 0-100 ppm / Max. 0--1000 ppm Min. 0-200 ppm / Max. 0--2000 ppm Min. 0-250 ppm / Max. 0-2500 ppm Min.0-500 ppm / Max. 0-5000 ppm Min.0-1000 ppm/ Max. 0-1 % Min.0-2000 ppm/ Max. 0-2 % Min.0-5000 ppm/ Max. 0-5 % Min.0-1%/ Max. 0-10% Min.0-2%/ Max. 0-20% Min.0-3%/ Max. 0-25% Min.0-5%/ Max. 0-50% Min.0-10%/ Max. 0-100% Type -A -B -C -D Optional suffix code Measurable -A Component -B *4 -C -D -E -F -G -H -K -L -M -P O2 Analyzer N 1 2 3 IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-8 <2SPECIFICATIONS> [Style : S1] Optional suffix code Model Suffix code IR202 --------------------------------------- Measuring range for CO *7,*8 Measuring range for CH4 *7,*8 Measuring range for O2 *7, Analog Output *9 -NN -01 -02 -03 -04 -07 -09 -10 -11 -12 -13 -14 -15 -16 -NN -07 -09 -10 -11 -12 -13 -14 -15 -16 N 1 2 3 4 5 6 7 8 -1 -4 Power Cable *10 Mounting Type *11 -NN -JA -US -EU -CN -A -B -C -D ----------------------------------------------------------------------------------------------------------------------------------- Not available Min. 0-500 ppm / Max. 0-5000 ppm Min. 0-1000 ppm / Max.0-1% Min. 0-2000 ppm / Max.0-2% Min. 0-5000 ppm / Max.0-5% Min. 0-1% / Max. 0-10% Min. 0-2% / Max. 0-20% Min. 0-3% / Max. 0-25% Min. 0-5% / Max. 0-50% Min. 0-10% / Max. 0-100% --------------------------------------------------------------------------------------------------------------------- Not available 0-5/10 vol% 0-5/25 vol% 0-10/25 vol% 0-5 vol% 0-10 vol% 0-25 vol% 0-50 vol% 0-100 vol% ----------------------------------------------------- None (Terminal block for power supply) Rated voltage 125V AC(PSE) Rated voltage 125V AC(UL) Rated voltage 250V AC(CEE) Rated voltage 250V AC(CCC) ----------------------------------------------------- Rack mounting with Slide rail Panel mounting Wall mounting Rack mounting - - - - - - - - - - - - - Always "-N” N ― Not available Min. 0-50 ppm/ Max. 0-500 ppm Min. 0-100 ppm/ Max. 0-1000 ppm Min. 0-200 ppm/ Max. 0-2000 ppm Min. 0-250 ppm/ Max. 0-2500 ppm Min. 0-500 ppm/ Max. 0-5000 ppm Min. 0-1000 ppm/ Max.0-1% Min. 0-2000 ppm/ Max. 0-2% Min. 0-5000 ppm/ Max. 0-5% Min. 0-1% / Max. 0-10% Min. 0-2%/ Max. 0-20% Min. 0-3%/ Max. 0-25% Min. 0-5%/ Max. 0-50% Min. 0-10%/ Max. 0-100% - - - - - - - - - - - - - - Japanese - - - - - - - - - - - - - - English - - - - - - - - - - - - - - Chinese -N ― --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- - - - - - - - - - - - - - - Rc1/4 - - - - - - - - - - - - - - 1/4NPT J E C Language - - - - - - - - - - - - - - Infrared Gas Analyzer - - - - - - - - - - - - - - 0-1V DC, non-isolation - - - - - - - - - - - - - - 4-20mA DC, non-isolation R T Piping - - - - - - - - - - - - - Always "N" /K /A /C /U Option Specification O2 correction output*12 Peak count alarm output*13 RS-485 communication*14 Conversion of unit (mg/m3, g/m3)*15 Note: *1: Same size as IR200 (former model) in size of an installation dimension. *2: For IR202-B, measuring component including NO cannot be selected. Measuring of 3 component cannot be selected. O2 measurement is not available. *3: For IR202-C, components including CH4 cannot be selected. *4 : For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately. *5 : Purchase separately ZX8D, Zirconia O2 analyzer. Prepare ZX8D StyleD as CE compliance product. *6: Purchase a O2 analyzer separately. A signal from the external O2 analyzer has to be set as 0-1 V DC linear corresponding *7: *8 *9 to full scale. . You can choose any range within the measurable range. Refer to Measurable component list For IR202-B, both of these codes are available to choose to make output. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <2.SPECIFICATIONS> 2-9 *10 Select appropriate power cables based on the required voltage and specification. The abbreviation represents each country or region that regulate the standard and certify each product. PSE-Japan, UL-US, CEE-Europe, CCC-China. When you select IR202-B, the power cable is always -NN. “-A “ cannot be specified with “-NN” . *11: Mounting of IR202-A is always -A , IR202-B: -B or C. IR202-C and IR 250-D: always -D. *12: Not selectable for IR202-B. Not selectable for the unit without O2 analyzer. O2 correction output applies only to NO, CO, and SO2. As an option, O2 correction average value is available. NO is indicated as NOx. *13: Selectable only for IR202-A. Peak count alarm can be provided only for CO measurement. *14: Should be specified when using MODBUS™ communication. *15 : Not selectable for IR202-B. Only the range for NO, SO2 or CO are convertable.Select a range by ppm. Preset value unit of ranges before shipping is converted one ; mg/m3 range (for 10000mg/m3 or above, the unit is g/m3) See the following conversion table as reference. ppm range range mg/m3 NO SO2 CO 0-50 ppm 0-65.0 mg/m3 0-140 mg/m3 0-60.0 mg/m3 0-100 ppm 0-130 mg/m3 0-280 mg/m3 0-125 mg/m3 0-200 ppm 0-260 mg/m3 0-570 mg/m3 0-250 mg/m3 0-250 ppm 3 0-325 mg/m 3 0-700 mg/m 0-300 mg/m3 0-300 ppm 0-400 mg/m3 0-850 mg/m3 0-375 mg/m3 0-500 ppm 0-650 mg/m3 0-1400 mg/m3 0-600 mg/m3 0-1000 ppm 0-1300 mg/m3 0-2800 mg/m3 0-1250 mg/m3 0-2000 ppm 0-2600 mg/m3 0-5600 mg/m3 0-2500 mg/m3 0-2500 ppm 0-3300 mg/m3 0-7100 mg/m3 0-3000 mg/m3 0-5000 ppm 0-6600 mg/m3 0-14.00 g/m3 0-6250 mg/m3 IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <2SPECIFICATIONS> 2-10 n Measurable Components and Ranges A : IR250-A, B : IR250-B, C: IR250-C, D : IR250-D (1) Single-component analyzer (NO, SO2, CO2, CO, CH4) Range Code Component Minimum/Maximum NO *1 SO2 CO2 CO CH4 *2 -01 0-50ppm/0-500ppm C C C C ― -02 0-100ppm/0-1000ppm C C CD C ― -03 0-200ppm/0-2000ppm CD CD CD ACD ― -04 0-250ppm/0-2500ppm CD CD CD CD ― -05 0-300ppm/0-2500ppm ― ― ― ― ― -06 0-500ppm/0-2500ppm ― ― ― ― ― -07 0-500ppm/0-5000ppm ACD ABCD ABCD ABCD D -08 0-1000ppm/0-5000ppm ― ― ― ― ― -09 0-1000ppm/0-1% ― D ABD ABD ABD -10 0-2000ppm/0-2% ― D ABD ABD ABD -11 0-5000ppm/0-5% ― D ABD ABD ABD -12 0-1%/0-10% ― D ABD ABD ABD -13 0-2%/0-20% ― ― ABD ABD ABD -14 0-3%/0-25% ― ― ABD ABD ABD -15 0-5%/0-50% ― ― ABD ABD ABD -16 0-10%/0-100% ― ― ABD ABD ABD *1 : *2 : IR250-B cannot measure NO. IR250-C cannot measure CH4. (2) Two-component analyzer Refer to the table (1) regarding measuring range (Minimum / Maximum.) a. NO+SO2 (1st component is NO, 2nd component is SO2) IR250-B is not applicable. 2nd comp. range 1st comp.range NO -01 -02 -03 -04 -07 SO2 -01 -02 -03 -04 -07 C C C C C C C C C C C C CD CD CD C C CD CD CD C C CD CD ACD IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-11 <2.SPECIFICATIONS> b. CO2+CO (1st component is CO2, 2nd is CO) Blank represents “not available.” CO 2nd comp. range 1st comp.range CO2 -01 -02 -03 -04 -07 -09 -10 -11 -12 -13 -14 -15 -16 -01 -02 -03 -04 C C C C C CD D D CD D D -07 -09 C C CD CD BCD BD BD -10 -11 BD BD BD BD BD ABD ABD ABD ABD ABD ABD ABD ABD ABD ABD ABD ABD BD ABD ABD ABD ABD -12 BD BD ABD ABD ABD ABD ABD ABD -13 -14 ABD ABD ABD ABD ABD ABD BD ABD ABD ABD ABD ABD ABD -15 BD BD ABD ABD ABD ABD ABD -16 ABD ABD ABD ABD ABD c. NO+CO(1st component is NO, 2nd is CO) IR250-B is not applicable. 2nd comp range 1st comp. range NO -01 -02 -03 -04 -07 -01 C C C C C -02 C C C C C CO -03 C C CD CD ACD -04 C C CD CD CD -07 C C CD CD ACD d. CH4+CO(1st component is CH4, 2nd is CO) IR250-C is not applicable. Blank represents "not available." 2nd.comp. range 1st comp. range CH4 -07 -09 -10 -11 -12 -13 -14 -15 -16 CO -03 -04 D D -07 D D AD -09 D AD AD AD AD AD -10 -11 D AD AD AD AD AD -12 AD AD AD AD -14 D AD AD -15 AD AD -16 AD AD AD IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <2SPECIFICATIONS> 2-12 e. CH4+CO(1st component is CH4, 2nd is CO) IR250-C is not applicable. Blank represents “not available.” 2nd comp.. range 1st comp.range CO2 -02 -03 -04 -07 -09 -10 -11 -12 -13 -14 -15 -16 CH4 -07 -09 D D D D D D D D D D D AD AD AD AD -10 D D D D AD AD AD AD AD -11 -12 AD AD AD AD AD AD D -13 -14 -15 D AD AD AD AD AD AD AD AD AD AD AD -16 AD AD AD (3) Three-component analyzer (NO and CO) IR250-B is not designed for three component analyzer. a. NO+SO2+CO (1st component is NO, 2nd is SO2, 3rd is CO) IR250-B is not applicable. Blank represents "not available." Code Min / Max range 1st component NO -01 0-50ppm/0-500ppm C -02 0-100ppm/0-1000ppm C -03 0-200ppm/0-2000ppm CD -04 0-250ppm/0-2500ppm CD -05 0-300ppm/0-2500ppm ― -06 0-500ppm/0-2500ppm ― -07 0-500ppm/0-5000ppm ACD -08 0-1000ppm/0-5000ppm ― -09 0-1000ppm/0-1% ― -10 0-2000ppm/0-2% ― -11 0-5000ppm/0-5% ― -12 0-1%/0-10% ― -13 0-2%/0-20% ― -14 0-3%/0-25% ― -15 0-5%/0-50% ― -16 0-10%/0-100% ― 3rd comp. range 2nd comp. range SO2 -01 -02 -03 -04 -06 -07 -08 CO -01 -02 -03 -04 C C C C C C C C C C CD CD AD C CD CD AD C -07 AD C AD IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-13 <2.SPECIFICATIONS> b. CO2+CO+CH4 (1st component is NO, 2nd is SO2, 3rd is CH4) IR250-B, IR250-C are not applicable. Blank represents "not available." 2nd comp. range 1st comp. range -11 -12 -13 -14 -15 -16 CO2 CO -07 -09 -10 -11 -12 -13 -14 -15 D AD AD AD D AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD AD Code Min./Max range 3rd component CH4 -01 0-50ppm/0-500ppm ― -02 0-100ppm/0-1000ppm ― -03 0-200ppm/0-2000ppm ― -04 0-250ppm/0-2500ppm ― -05 0-300ppm/0-2500ppm ― -06 0-500ppm/0-2500ppm ― -07 0-500ppm/0-5000ppm ― -08 0-1000ppm/0-5000ppm ― -09 0-1000ppm/0-1% ― -10 0-2000ppm/0-2%% ― -11 0-5000ppm/0-5 AD *1 AD -12 0-1%/0-10% -13 0-2%/0-20% AD -14 0-3%/0-25% AD *2 -15 0-5%/0-50% ― -16 0-10%/0-100% ― *1: *2: AD AD AD AD -16 AD AD AD AD AD CO range -16 is not applicable. CO range -07 is not applicable. 2.3 Standard Accessories Name Qty Description Code Specification Fuse 2 250 V/2A delay type (K9642CJ) Input/output terminal module 1 Connector for external O2 analyzer O2 analyzer "1", "2" Type, “-C”, “-D” Power cable 1 (varies with type of power cables) come with a cable stopper power cable "-JA", "-US", "-EU", "-CN" Ferrite Cores for Power Cable 1 power cable "-NN" Ferrite Cores for digital I/O 3 Type; ”A","-B" Analog output, digital I/O connector 4 D-sub connector 25 pins with 8 male screws Type, "-C", "-D" RS-485 Connector 1 D-sub connector 9 pins with 2 male screws Option "-C" Mounting plate 4 Mounting type "-B" Slide rail 2 Mounting type, Factory-installed delivery. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2.4 2-14 <2SPECIFICATIONS> External dimensions l IR250-A Power switch Unit: mm M4 (for slide rail) 7 101.6 177 463 483 169 24 480 or less 429 Terminal Terminal Terminal block 3 block 5 Commun. block 1 Connector Terminal Terminal block 2 block 4 Fuse Sampling gas outlet (Rc1/4 or 1/4NPT) Sampling gas inlet (Rc1/4 or 1/4NPT) Purge gas inlet (Rc1/4 or 1/4NPT) Power supply 4-M6 101.6 450 or more 465 Rack mounting dimensions IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-15 <2.SPECIFICATIONS> l IR250-B Unit: mm 70 Mounting hardware for panel mounting *1 Mounting plate for Wall mounting (code: -C) 4-M6 or Φ8 440 454 484 454 70 175 70 294 Mounting dimensions for Wall mounting (code: -C) 281 +2 0 Wiring port (Rubber grommets) Purge gas inlet (Rc1/4 or 1/4NPT) Sampling gas inlet (Rc1/4 or 1/4NPT) Sampling gas outlet((Rc1/4 or 1/4NPT) 430 +2 0 150 350 *1: 4 mounting brackets come in package for Panel cut-out dimensions and spacing Panel mounting (code: -B) for Panel mounting (code: -B) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-16 <2SPECIFICATIONS> l IR250-C or IR250-D Unit: mm Power switch (M4 Terminal Block Type) (Power Cable code: -NN) Flow checker (Only for IR250-C) 25 7 57.2 463 483 Analog output connector (A/O) Communication connector (RS-485) 429 External input connector (A/I) 124.5 132.5 24 16 380 Purge gas inlet (Rc1/4 or 1/4NPT) Sampling gas inlet (Rc1/4 or 1/4NPT) Sampling (Reference) gas outlet (Rc1/4 or 1/4NPT) Reference gas inlet (only for IR250-C) (Rc1/4 or 1/4NPT) Fuse 4-M6 Power supply 57.2 Digital input/output connector (M4 Terminal Block Type) (Power Cable code: -NN) (DIO1 to 3 from the left) 450 or more 465 Rack mounting dimensions IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-17 <2.SPECIFICATIONS> Accessories l Slide rail (Model : 305A-20/Acuride International Inc.) Only when Mounting type -A is selected, the slide rail is available. The rail is pre-installed at factory before delivery. Unit: mm Closed 530±3 Cabinet member 1.52t Closed position flush or under (22.7) Intermediate member (9.5) Drawer member 1.27t 15.9±0.5 Open 508.0±0.8 476.2±0.3 463.5±0.3 352.8±0.3 111.1±0.3 (24.4) 35.3±0.3 25.4±0.3 12.7±0.3 12.7±0.3 4.5 × 5.3 123.8±0.3 4.5 × 5.3 238.1±0.3 15.9±0.5 19.1±0.3 365.1±0.3 505.0±0.8 F05.EPS Reinforcement plate IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2-18 <2SPECIFICATIONS> The instrument weight should be supported at the bottom of the unit (or the side of the unit when mounted with the slide rails). For easy maintenance, it is recommended to have an installation that allows the instrument to be withdrawn along the slide rail. Slide rail mounted type Guide rail mounted type * Rack size Unit: mm 101.6 or 100 101.6 or 100 Rack size 450 or more 465 Mounting diagram Slide rails 450 or more 465 EIA standard EIA standard Mounting diagram Guide rails * For the guide rail mounted type, a maintenance space (200 mm or more) should be provided on top of the main unit.   IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 2.5 2-19 <2.SPECIFICATIONS> NO2 / NO Converter Use for NOx measurement. Part number: K9350LE (Non-CE conformity) K9350LF (CE conformity) Mounting: Indoor surface mounting Target Gases: General boiler exhaust gas, atmosphere Catalyst: Amount; Replacement cycle; 2 cm3 Approx. 8 months (at flow rate of 0.5 L/min with 5% O2,10 ppm NO2) Temperature set-point; 220 ±10°C (Sensing tip; K thermocouple) Wetted materials:  ceramic, Fluoropolymer (PTFE), glass filter, SUS316 Conversion efficiency: 90% or higher, conforms to JIS Gas Flow Rate: 0.5 L/min Ambient Temperature: -5 to 45°C Power Supply:  K9350LE; 100 V AC, 50/60 Hz K9350LF; 100 to 240 V AC, 50/60 Hz Power Consumption: Approx. 85 VA Weight: Approx. 1.1 kg Sample gas requirements: Dust/drain removed, gas temperature at 150°C or lower Safety conforming standards: Installation altitude; 2000 m or less Pollution degree; 2 (Note) Installation category; II (Note) ・ Installation category, called over voltage category, specifies impulse with standing voltage. Category II is for electrical equipment. ・ Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which reduce dielectric strength. Degree 2 is the normal indoor environment. Note Safety; EN61010-1 EMC; EN61326-1 Class A, Table 2 (For use in industrial locations), EN61326-2-3, EN61000-3-2, EN61000-3-3 EMC Regulatory Arrangement in Australia and New Zealand (RCM) Note: The product mounted in a steel enclosure conforms to the requirements of EMC directive. CAUTION: The instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only. Spare Parts for one-year-use Item Part No. Qty 2 Catalyst for NO2/NO converter K9350LP Glass wool for NO2/NO converter K9350LQ 2 Fitting for NO2/NO converter K9350LV 2 IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 K9350LE <2SPECIFICATIONS> 2-20 Unit: mm (150) 135 Gas Outlet ∅9.5/∅5.5 Wiring Diagram (120) 105 (210) 7 8 4-∅5 Gas Inlet ∅9.5/∅5.5 Approx. 130 110 K9350LF 100 V AC (Shared with internal wiring) Unit: mm Unit:mm 148 135 124 130 (NP) Gas Outlet ø9.5/5.5 27 105 160 212 (NP) LNE 4-ø5 Power Inlet Front View 12 Gas Inlet ø9.5/5.5 Side View IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3. 3-1 <3. INSTALLATION> INSTALLATION WARNNING IR250 is Non-Explosion-proof type. Do not use it in a place with explosive gases to prevent explosion, fire or other serious accidents. CAUTION • Inappropriate installation may cause hazardous accident such as tipover, electric shock, fire, and injury. • IR250 is heavy product. It should be handled carefully to install. Otherwise it may cause tipover, fall, injury and hazardous accident. • During installation work, care should be taken to keep the unit free from cable chips or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit. 3.1 Where to install equipment Installation Requirements • The product is for indoor use. Avoid exposure to direct sunlight, weather, and radiant heat from hot substances. If exposure to such conditions are unavoidable, a protective hood or cover should be prepared. • Vibration-free environment • A clean atmosphere Ambient temperature: -5 to 45°C (40°C at max if suffix code -C, -G, -H, -P is selected and the electric power being used is larger than 200 V power supply) Ambient humidity: 95%RH or less, no condensation Clear the space as below if the analyzer is installed by itself. analyzer (when installed) Approx. 500mm Approx. 400mm analyzer (drawer out) Approx. 500mm Approx. 300mm IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-2 <3. INSTALLATION> To install the analyzer for optimum performance, select a location that meets the following conditions; • IR250 is system built-in type, embedded in a steel rack, or enclosure of steel sheet. Keep a minimum clearance of 10 cm above the analyzer for heat dissipation. The same size of clearance is required for each analyzer when you install several units on a multistage rack. • Circuit breaker rating should meet the analyzer rating, that meets IEC60947-1 and IEC60947-3. • Circuit breaker should be installed near the analyzer where an operator can access it. • A label that clearly identifies the breaker should be placed on it. 3.2 How to install The next table shows the type of mounting and its code of equipment. Rack-mounting type of equipment can be installed with sliding rails. Mounting type (Code) IR250-A Rack-mounting with sliding rails (-A) Panel-mounting (-B) Wall-mounting (-C) Rack-mounting (-D)  See “2.4 IR250-B IR250-C IR250-D     External dimensions” on page 2-14. Follow the next instruction and install equipment properly. • Front side of equipment must stand vertically. • The analyzer weight must be supported at the bottom of the casing.(except IR250-B) • Protect the analyzer from vibrating. For example, install rubber material around the case to minimize the vibration. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3.3 3-3 <3. INSTALLATION> Piping WARNING In piping, the following precautions should be observed. Wrong piping may cause gas leakage. If the leaking gas contains a toxic component, there is a risk of serious accident being induced. Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring. • Connect pipes correctly referring to the instruction manual. • Exhaust should be led outdoors so that it will not remain in the locker and installation room. • Exhaust from the analyzer should be relieved in the atmospheric air in order that an unnecessary pressure will not be applied to the analyzer. Otherwise, any pipe in the analyzer may be disconnected to cause gas leakage. • For piping, use a pipe and a pressure reducing valve to which oil and grease are not adhering. If such a material is adhering, a fire or the like accident may be caused. CAUTION Piping connection must be secured. Gas inlets have several types: reference gas inlet, outlet or purge gas inlet. When the connection is insecure or wrong, combustible, toxic, explosive gas may be accumulated inside the analyzer or system. Observe the following when connecting the gas pipes. • The pipes should be connected to the gas inlet and outlet on the analyzer, respectively. • Connect the sampling system to the instrument by using corrosion-resistant tube. The material should be such as Fluoropolymer (PTFE), stainless steel, or polyethylene. In case where there is no danger of corrosion, don’t use rubber or soft vinyl tube. Analyzer indication may become inaccurate due to the adsorption of gases. • Piping connections are Rc1/4 (1/4 NPT) female-threaded. Cut the pipe as short as possible for obtaining quick response. Pipe of approx. ∅4 mm (inside diameter) is recommended. • Entry of dust in the instrument may cause operation fault. Use clean pipes and couplings. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <3. INSTALLATION> 3-4 Sampling gas outlet (Rc1/4 or 1/4NPT) Sampling gas outlet (Rc1/4 or 1/4NPT) Purge gas inlet (Rc1/4 or 1/4NPT) Purge gas inlet (Rc1/4 or 1/4NPT) Sampling gas inlet (Rc1/4 or 1/4NPT) Sampling gas outlet(shared with reference gas outlet in IR250-C) (Rc1/4 or 1/4NPT) Reference gas inlet (only for IR250-C) (Rc1/4 or 1/4NPT) Purge gas inlet (Rc1/4 or 1/4NPT) Sampling gas inlet (Rc1/4 or 1/4NPT) Sampling gas outlet(Rc1/4 or 1/4NPT) Sampling gas inlet: Connect the pipe so that zero/span calibration standard gas or measured gas pretreated with dehumidification is supplied properly. The gas flow rate should be kept constant within the range of 1.0 ± 0.5 L/min. for IR250-A, 0.5 ± 0.2 L/min. for IR250-B,IR250-D, and 1.0 ± 0.2 L/min. for IR250-C. Sampling gas (reference gas ) outlet: Measured gas is exhausted after measurement. Connect the pipe so that the gas may escape through the gas outlet into the atmosphere. Purge gas inlet: It is used for purging the inside of the total gas analyzer. When the analyzer must be purged, refer to Item 3.4, Purging inside Analyzer. Use dry gas N2 or instrumentation air for purge gas. (flow rate of 1 L/min or more should be used and no dust or mist is contained). Reference gas inlet: For IR250-C, introduce reference gas for the sample switching. Connect the pipe so that atmosphere pretreated with dehumidification, dry air, dry N2 is supplied properly. The gas flow rate should be kept constant within the range of 1.0 ± 0.2 L/min. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3.4 3-5 <3. INSTALLATION> Sampling See “1.2 Sampling system configuration” on page 1-3 l Standard requirements of sampling gas Flow rate: IR250-A ; 1.0 ± 0.5 L/min IR250-B, IR250-D ;0.5 ± 0.2 L/min IR250-C; 1.0 ± 0.2 L/min Temperature: 0 to 50°C Pressure: less than 10 kPa (atmosphere open at gas outlet) Dust : Particle size should be smaller than 0.3µm and the mass concentration should be less than 100µg / Nm3. Mist : Not allowed Moisture lower than 2°C saturation Corrosive element : less than 1 ppm Use sampling gas with care as below; • The dust contained in sample gas should be eliminated completely with filters. The filter at the final stage should be capable of eliminating dust of 0.3 microns. • The dew point of sample gas must be lower than the ambient temperature for preventing formation of drain in the analyzer. If water vapor is contained in sample gas, its dew point should be reduced down to about 0°C through a dehumidifier. • If SO3 mist is contained in sample gas, the mist should be eliminated with a mist filter, cooler, etc. Eliminate other mist in the same way. • If a large amount of highly corrosive gas such as Cl2, F2 or HCl is contained in sample gas, the service life of analyzer will be shortened. So, avoid such gases. • Sample gas temperature is allowed within a range from 0 to 50°C. Pay attention not to flow hot gas directly into the analyzer. • Flow rate must not be changed during a measurement. • Install a flowmeter as shown in a configuration (1.2), and check the flow rate. l Standard gas for calibration Prepare the standard gas for zero / span calibration. Periodical calibration are necessary. (weekly is recommended) Measuring components excluding O2 Zero gas: N2 gas Span gas: Gas with concentration of 90% level or higher in the range of each corresponding component (recommended) (the rest of it is N2) Measuring O2 (Built-in paramagnetic O2 analyzer) N2 gas Zero gas: Span gas: Gas with concentration of 90% level or higher in the range of each corresponding component (the rest of it is N2 gas or atmosphere (21% O2 )) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-6 <3. INSTALLATION> Measuring O2 (External Zirconia O2 analyzer) Zero gas Dry air, or atmospheric air, but atmospheric air should not be contained in zero gas if CO2 meter is provided. Span gas 1 to 2% gas (the rest of it is N2) It is understood to influence the calibration model of the infrared gas analyzer when a lot of H2, He, and Ar are included in the measurement gas (pressure broadening). The span gas must use a gas near the composition of the measurement gas when you measure the gas like the above-mentioned. l Purging of instrument inside The inside of instrument need not be purged generally except for the following cases. • A combustible gas component is contained in sample gas. • Corrosive gas is contained in the atmospheric air at the installation site. • The same gas as the sample gas component is contained in the atmospheric air at the installation site. In such cases as above, the inside of analyzer should be purged with the air for instrumentation or N2. Purging flow rate should be about 1 L/min. If dust or mist is contained in purging gas, it should be eliminated completely in advance. l Pressure at sampling gas outlet Pressure at the sampling gas outlet should be adjusted to atmospheric pressure. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3.5 3-7 <3. INSTALLATION> Wiring CAUTION • Wiring work must be performed with the main power set to OFF to prevent electric shocks. • Enforce construction of class-D grounding wire by all means, 100 Ω or less. If the specified grounding construction is neglected, a shock hazard or fault may be caused. • Wires should be the proper one meeting the ratings of this instrument. If using a wire which cannot endure the ratings, a fire may occur. • Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating may cause fire. See the following pages on wiring instruction. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <3. INSTALLATION> 3-8 Commun.Connector Terminal Terminal Terminal block 5 block 1 Terminal block 3Terminal block 2 Power supply Fuse block 4 Communication External input connector connector (RS-485) (A/I) Analog output connector (A/O) Fuse Power Digital input/output connector (DIO1 to DIO3 from bottom to the top) supply Power terminal (M4 Terminal Block Type) (Power Cable code: -NN) COMP2(2nd comp.)I/O terminal COMP1(1st comp.)I/O terminal Power supply Wiring port (Rubber grommets) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-9 <3. INSTALLATION> n Power Supply Connect the given power supply to the power terminal, and connect the ground wire to the grounding terminal (P). Be sure to perform protective earth connection. Use solderless terminals (for M4) for connection to the terminals (power and earth). Application line diameter ø9.5 to ø10.5 mm. M4 thread terminal P N Power inlet L attached ferrite core AC 100 to 240V Grounding 2-pole plug Earth cable CAUTION After wiring , be sure to place the protective cover for the M4 thread terminal blocks to assure safety. l When noise source is in the vicinity • Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder) that generates much electrical noise. If using the instrument near such a noise generating unit is unavoidable, use a different power line to avoid noise • Mount a noise suppressor such as varistor or spark quencher as shown below to the noise generating unit when noise is generated from relays or solenoid valves. Mount the suppressor near the noise generating source, otherwise it will have no effect. Main unit power supply varistor or spark quencher install near the source Noise generating source n I/O Wiring Diagram NOTE • The analog output signals of the instrument are not isolated individually. It is recommended to isolate the signals individually to eliminate the interference from the unnecessary signals or the effect of external interference, especially it the cable exceeds 30 meters or leads to outdoors. • To avoid external interference, wiring of analog output signal, O2 sensor input and contact input should be run separately from that of power supply and contact output. • To avoid the effect of noise generated from external units, be sure to ground the analyzer main unit. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-10 <3. INSTALLATION> n I/O Wiring Diagram l IR250-A Applicable diameter of conducting wire is within Ø 1.9(AWG20) to Ø 2.1 mm(AWG18) Terminal block 1 CH5 output – (CH5_OUT) + 1 11 2 12 CH4 output – (CH4_OUT) + 3 13 4 14 CH3 output – (CH3_OUT) + 5 15 6 16 CH2 output – (CH2_OUT) + 7 17 8 18 CH1 output – (CH1_OUT) + 9 19 10 20 Unassigned * O2 sensor input – (O2_IN) + Unassigned – Unassigned + Unassigned – CH7 output + (CH7_OUT) Unassigned – CH6 output + (CH6_OUT) Unassigned 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 (M3.5 screw) Terminal block 3 Terminal block 4 11 2 12 3 13 4 14 Remote hold input (R_HOLD) 5 15 6 16 Average value reset input (RESET) 7 17 8 18 Auto calibration remote start input (R_CAL) 9 19 10 20 Unassigned Unassigned (M3.5 screw) 1 Unassigned Terminal block 2 Instrument error (FAULT) Contact output for CH4 span calibration (SPAN_CH4) 1 11 2 12 CH4 range identification signal output (RNG_IDCH4) Contact output for CH3 span calibration (SPAN_CH3) 3 13 4 14 CH3 range identification signal output (RNG_IDCH3) Contact output for CH2 span calibration (SPAN_CH2) 5 15 6 16 CH2 range identification signal output (RNG_IDCH2) Contact output for CH1 span calibration (SPAN_CH1) 7 17 8 18 CH1 range identification signal output (RNG_IDCH1) Contact output for zero calibration (ZERO) 9 19 10 20 (M3.5 screw) Unassigned CH4 remote range changeover input (R_RNG_CH4) CH3 remote range changeover input (R_RNG_CH3) CH2 remote range changeover input (R_RNG_CH2) CH1 remote range changeover input (R_RNG_CH1) Unassigned Auto calibration status contact output (ACAL/MNT) Calibration error contact output (CAL_ALM) Pump ON/OFF contact output (PUMP) Unassigned (M3.5 screw) Terminal block 5 Unassigned CH3 alarm output (ALM_CH3) CH2 alarm output (ALM_CH2) CH1 alarm output (ALM_CH1) 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 *O2 sensor input is used when an external O2 analyzer is selected. Unassigned Note: Unassigned terminals are used for internal connection. So they should not be used as repeating terminals either. Peak alarm output (PEAK_ALM) CH4 alarm output (ALM_CH4) (M3.5 screw) NOTE After connecting contact I/O signal cables, install ferrite core on each terminal stand. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-11 <3. INSTALLATION> l IR250-B Applicable diameter of conducting wire is within Ø 1.9(AWG20) to Ø 2.1 mm(AWG18) AUTO CAL Unassigned Unassigned Unassigned 1 7 2 8 3 9 4 10 5 11 6 12 Unassigned Unassigned Unassigned COMP2 (2nd component) 0-1 V DC 7 Unassigned 1 + (instantaneous − 8 2 value ) Unassigned 9 4-20mA DC + 3 (instantaneous − 4 10 Unassigned value) 5 11 Unassigned Unassigned 6 12 (M3.5 screw) COMP1 (1st component) 0-1 V DC 7 Unassigned 1 + (instantaneous value) 4-20mA DC (instantaneous value) FAULT (M3.5 screw) − + − 2 8 3 9 4 10 5 11 6 12 Unassigned Unassigned Unassigned (M3.5 screw) (Note)Unassigned terminals are used for internal connection. So they should not be used as repeating terminals, either. When a Contact Output error occurs on IR250, the connection between 5 and 6 is conductive. 1a contact 250 VAC, 2A (Resistive load) NOTE After connecting contact I/O signal cables, install ferrite core on each terminal stand. l IR250-C, IR250-D Connector for external input(A/I) Connector for analog output(A/O) (for O2 input) AI- 13 25 1 5 1 14 9 6 D-sub 25pin Male AI+ D-sub 9pin Female (GND) AO1+ 6 AO12 RTxD+ AO2+ 7 AO23 RTxDAO3+ 8 AO34 AO4+ 9 AO45 AO5+ AO5AO6+ AO6AO7+ AO7AO8+ AO8AO9+ AO9AO10+ AO10AO11+ AO11AO12+ (Note) Display Ch number is same as the AO number under standard specifications. AO12Do not use connectors in blank. 1 1 14 2 15 3 16 4 17 5 18 6 19 7 20 8 21 9 22 10 23 11 24 12 25 13 RS-485 Connector IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 <3. INSTALLATION> 3-12 Connector for analog output (A/O) Output signal : 4 20 m A DC or 0 to 1 V DC (select on order) Minus lines for the insulation and signal are common from the ground and internal circuit. Allowable load: 4 to 20 mA DC, 100 kΩ or more Connector for external input (A/I) Input signal: External zirconia O2 Sensor (ZX8D) : Zirconia O2 sensor signal External O2 analyzer : 0 to 1 V DC (DC input resistor of 1MΩ or more) • It is used when the cord 1 (external zirconia O2 Sensor(ZX8D)) or cord 2 (an external O2 analyzer) is specified as ordered. DO NOT connect when cord 3 (Built in paramagnetic O2 analyzer) is installed. • Connect the dedicated connector (accessory) to the output of the external Zirconia analyzer or the external O2 analyzer (received separately). • In case of an external O2 analyzer, input a signal of 0 to 1 V DC with respect to O2 full scale of the analyzer. The O2 concentration display, output, and O2 correction can be performed. NOTE In IR250-C and IR250-D, O2 sensor external input is not isolated. It is recommended to isolate when an external O2 analyzer is installed apart from IR250. Zirconia O2 sensor (ZX8D) should be installed at a location that is as close to this instrument as possible How to connect the O2 signal to the dedicated connector (accessory). Turn + To O2 sensor - Solder the signal cable to the terminal. RS-485 Connector For MODBUS protocol connection Only when (-C) is selected. Contact input / output (DIO) : Connector for digital input / output (DIO1,DIO2, DIO3) Contact input signal : Voltage is applied from the external 12 to 24 V DC, max 15mA Photocoupler isolation (from each DI and ground) Contact capacity : C contact relay output 24V/1A AC/DC resistive load IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Connector for digital input/output (DIO1, DIO2, DIO3) 13 1 1 14 25 3-13 <3. INSTALLATION> 14 D-sub 25pin female 2 15 3 16 4 17 5 18 (Note) DIO1, DIO2, DIO3 have the same internal circuit of the connector. Do not use pins in blank. 6 19 7 20 8 21 9 22 10 23 11 24 NC com NO NC com NO NC com NO NC com NO NC com NO DIO1 Connector DI1+ DI1DI2+ DI2DI3+ DI3- DIO2 Connector DI4+ DI4DI5+ DI5DI6+ DI6- DIO3 Connector DI7+ DI7Digital input DI8+ OFF : 0 V DI8ON : 12 to 24 V DC DI9+ DI9- DO1 DO6 DO11 DO2 DO7 DO12 DO3 DO8 DO13 DO4 DO9 DO14 DO5 DO10 DO15 Digital output max.contact load rating 24 V DC/1A 12 25 13 IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-14 <3. INSTALLATION> ● digital input signal DI1 DI2 DI3 DI4 Remote hold Average value reset Auto calibration start Auto zero calibration start Remote range 1 Remote range 2 Remote range 3 Remote range 4 Remote range 5 DI5 DI6 DI7 DI8 DI9 ● digital output signal DO4 DO5 DO6 1-compo.analyzer instrument error calibration error (auto calibration status) (zero) (For span gas Ch1 ) ― 2-compo.analyzer 3-compo.analyzer instrument error instrument error calibration error calibration error (auto calibration status) (auto calibration status) (zero) (zero) (For span gas Ch1 ) (For span gas Ch1 ) (For span gas Ch2 ) (For span gas Ch2 ) 4-compo.analyzer instrument error calibration error (auto calibration status) (zero) (For span gas Ch1 ) (For span gas Ch2 ) DO7 DO8 ― ― ― ― (For span gas Ch3 ) (For span gas Ch4 ) DO9 ― DO10 Range identification Ch1 DO11 (Alarm 1) DO12 (Alarm 2) DO13 (Alarm 3) DO14 (Alarm 4) Range identification Ch1 Range identification Ch2 (Alarm 1) (Alarm 2) (Alarm 3) (Alarm 4) (For span gas Ch3 ) Range identification Ch1 Range identification Ch2 Range identification Ch3 (Alarm 1) (Alarm 2) (Alarm 3) (Alarm 4) DO15 (Alarm 5) (Alarm 5) (Alarm 5) DO1 DO2 DO3 Range identification Ch1 Range identification Ch2 (Alarm 1) (Alarm 2) (Alarm 3) Range identification Ch3 Range identification Ch4 Note The normal open side (NO) of digital output is close when the function is active without range ID. In case of range ID, normal open (NO) side is close with L-range. The normal close (NC) side is close with H-range. NOTE • Isolated output (from each DO and ground) To avoid external interference, wiring of analog output signal, O2 sensor input and contact input should be run separately from that of power supply and contact output. • To avoid the effect of noise generated from external units, be sure to ground the analyzer main unit IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3-15 <3. INSTALLATION> Suffix/option code Option comp O2 Analyzer code* -A, -C, -D -A N not specified -A, -B, -C, -D -B N not specified -A, -C, -D -C N not specified -A, -B, -C, -D -D N not specified -A, -B, -C, -D -E N not specified -A, -B, -C, -D -F N not specified -A, -C, -D -G N not specified -A, -C, -D -H N not specified -A, -B, -D -K N not specified -A, -D -L N not specified -A, -D -M N not specified -A, -D -P N not specified -A, -C, -D -A 1, 2, 3 not specified -A, -C, -D -B 1, 2, 3 not specified -A, -C, -D -C 1, 2, 3 not specified -A, -C, -D -D 1, 2, 3 not specified -A, -C, -D -E 1, 2, 3 not specified -A, -C, -D -F 1, 2, 3 not specified -A, -C, -D -G 1, 2, 3 not specified -A, -C, -D -H 1, 2, 3 not specified -A, -D -K 1, 2, 3 not specified -A, -D -L 1, 2, 3 not specified -A, -D -M 1, 2, 3 not specified -A, -D -P 1, 2, 3 not specified NO SO2 NO CO CO2 CO2 NO NO CH4 CH4 CO2 CO2 NOx SO2 NOx CO CO2 CO2 NOx NOx CH4 CH4 CO2 CO2 -A, -C, -D -A 1, 2, 3 /K NOx O2 -A, -C, -D -B 1, 2, 3 /K SO2 O2 -A, -C, -D -C 1, 2, 3 /K -A, -C, -D -D 1, 2, 3 /K -A, -C, -D -F 1, 2, 3 /K -A, -C, -D -G 1, 2, 3 /K -A -H 1, 2, 3 /K -C, -D -H 1, 2, 3 /K NOx SO2 CO -A, -D -L 1, 2, 3 /K CH4 CO O2 -A, -D -P 1, 2, 3 /K CO2 CO CH4 Type *: ● ● ● Output and corresponding channels Ch1 Ch2 Ch3 Ch4 Ch5 Ch6 Ch7 Ch8 Ch9 Ch10 SO2 CO CO SO2 CO CO CH4 CO O2 O2 SO2 O2 O2 CO CO SO2 O2 CO CH4 CO CH4 O2 O2 O2 CO O2 O2 O2 CH4 O2 Correct Correct NOx NOx Ave. Correct Correct SO2 SO2 Ave. Correct Correct Correct NOx NOx SO2 Ave Correct Correct CO CO O2 CO Ave. Correct Correct CO2 CO O2 CO CO Ave. Correct Correct Correct NOx CO O2 NOx NOx CO Ave Correct Correct NOx SO2 CO O2 NOx SO2 NOx SO2 O2 Correct SO2 Ave Correct CO Ave Correct CO Correct Correct Correct Correct Correct Correct NOx SO2 CO NOx SO2 CO Ave Ave Ave Correct Correct CO CO Ave. Correct Correct O2 CO CO Ave O2 In the column of Option code, “Not specified” refers any option code except /K (with O2 correction value output). When CH1 has SO2 in its raw, it means the Ch1 corresponds with the SO2 component. When option /K is specified, NO is indicated as NOx. “Correct” represents O2 correction. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3.6 3-16 <3. INSTALLATION> Solenoid valve drive signal for calibration 3.6.1 Manual calibration (See “6. CALIBRATION”). ・Zero calibration Calibration end on Cal.gas flow Zero Calibration output off Ch1to 5 Span calibration output off Output hold function Output signal hold (With hold ON setting) Hold extension time ・Span calibration Zero Calibration output Calibration end when selecting CH2 with up/down keys Ch1Span calibration output Ch2 Span calibration output Output hold function (With hold ON setting) Cal.gas flow Output signal hold Hold extension time * The hold extension time depends on the gas flow time of the automatic calibration settings. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 3.6.2 3-17 <3. INSTALLATION> Auto calibration (See “5.3 Alarm Setting”Automatic calibration) Ch1 span Ch3 span calibration calibration Zero Ch4 span Ch2 span calibration calibration calibration Zero gas Auto cal. start Zero calibration output Ch5 span calibration 350S Ch1 span calibration output Ch2 span calibration output Ch3 span calibration output Ch4 span calibration output Ch5 span calibration output Ch1span gas 350S Ch2span gas 350S Ch3span gas 350S Ch4span gas 350S Ch5span gas 350S Auto. Cal. Contact Output hold function Hold extension time (with hold ON setting) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Blank Page 4. 4-1 <4. OPERATION> OPERATION Check that the pipes are correctly connected to the gas sampling port and drain port. Check that the analyzer is correctly wired as specified. 4.1 Warm-up operation and regular operation Procedure 1. Turn ON the power. In one or two seconds, the measurement screen will appear at the front panel. 2. IR250-A, IR250-C, IR250-D need about 4 hours, IR250-B need about 2 hours for warm-up. During the warm-up, concentration reading may indicate “ ------”, which is not a sign or error operation. 3. Set the required values according to Chapter 5, “Setup (User mode)”. 4. Perform zero / span calibration after the warm-up. See “6. CALIBRATION” on page 6-1. 5. Start supplying measuring gas to IR250. Start measuring. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 4.2 4-2 <4. OPERATION> Keys and function on Panel 1 2 3 4 5 6 Displays Controls 7 8 Control keys Keys Description Keys Description (1) MODE switch the mode. (5) DOWN move cursor downward to select an item or decrement numeral value. (2) ZERO go to zero calibration. (6) UP move cursor upward to select an item or increment numeral value. (3) SPAN go to span calibration. (7) ESCAPE return to a previous screen or cancel the setting midway (4) SIDE change the selected item (by moving the cursor) and numeral digit. (8) ENTER confirm selected items or values, or execute calibration. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 4.3 4-3 <4. OPERATION> Mode and screens IR250 has two modes, Measuring mode and User mode. User mode includes Maintenance mode. During the measuring mode, calibration is performed. The picture below shows how to navigate through modes . Detail varies with specification. Measurement mode *1 ZERO Calibration (see 6.1) Measurement mode Span Calibration (See 6.2) Measurement mode User mode Calibration value(5.2.1) Zero calibration (5.2.2) Calibration range (5.2.3) Auto calibration components / range (5.2.4) ●Changeover of Range(5.1) ●Setting about Calibration(5.2) ●Alarm Setting (5.3) *3 ●Setting Auto calibration(5.4) *3 ●Setting Auto Zero Calibration(5.5) *3 ●Setting of Peak Alarm(5.6) *2 ●Parameter Setting(5.7) *1 Scroll screen on measure mode. *2 only when optio code /A is selected *3 not displayed on IR202-B Start time Cycle Flow time ON / OFF Auto calibration Run / Stop Start time Cycle Flow time ON / OFF Auto zero calibration Run / Stop ON / OFF Peak alarm value Peak alarm count Hysteresis Current time Key lock Output Hold (5.7.1) Response time (5.7.2) Average period (5.7,2) Backlight timer (5.7.2) Contrast(5.7.2) Maintenance (Enter password) (5.7.3) IM 11G02Q01-01EN F0802.ai 3rd Edition : May 31, 2017-00 4-4 <4. OPERATION> n Measuring mode When power turns on, always a measuring mode (see a sample below) appears. To display more than 6 channels on a page, scroll the page with keys. (2) (1) (2) (1) Ch 0.0 0.0 0.0 0 0.0 0.0 0 NOX 0-200 Ch SO2 0-200 Ch CO2 0-20 Ch CO 0-200 Ch O2 0-25 (3) Ch 8 ppm Ch 9 ppm Scroll vol% Ch 10 Ch ppm 11 vol% 12 Ch (4) AV CV 0.0 0.0 0.0 0.0 0.0 NOX 0-200 AV CV NOX AV CV SO2 AV CV CO 0-200 0-200 0-200 AV O2 0-25  1h ppm 1h ppm 1h ppm 1h ppm 1h vol% (3) (5) F0803.ai Number Display (1) Component (2) Concentration (3) Range (4) Unit (5) Average time Description Displays which component of measured gas showing instantaneous value, converted instantaneous value, converted average value, etc. Displays measured value of concentration. Displays range values. Displays unit , ppm or vol % Displays average time. l Instantaneous concentration value: As a sample screen picture shows above, gas component such as NOx on Ch1(shown above), SO2 on Ch2 (shown above) indicate what gas component is currently being measured and each corresponding instantaneous concentration value. l O2 correction concentration values: Channels (Ch) with “cv” indicate O2 correction concentration value of each designated gas component. The O2 correction concentration value is calculated from the following equation, based on which sampling component is measured, what is the O2 instantaneous concentration value, and O2 correction reference value. See 5. Other Parameters in 5.7.3 Maintenance Mode , 21 - On On: The value of the O2 correction reference value x Cs 21 - Os (Value set by application) Os: Oxygen concentration (%) Cs: Concentration of relevant measured component  Os is not more than O2 concentration limit in value. The correction components are NO, CO and SO2 only. Correction output= l O2 correction concentration average value: AV AV In the Ch (component) and O2 average value where “ CV ” is displayed as “ CV CO” in the component display, a value obtained by averaging O2 corrected concentration value or O2 average value in a fixed time is output every 30 seconds. Averaging time can be changed between 1 to 59 minutes or 1 to 4 hours according to the average time settings . IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 4-5 <4. OPERATION> (The averaging set time is displayed as “1h” , for instance, in the range display.) * The measurement ranges of O2 correction concentration value and O2 correction concentration average value are the same as that of the measuring components. Also, the measurement range of O2 average value is the same as that of O2. n User mode, Calibration mode The setting/selection screen is configured as shown below: • In the status display area, the current status is displayed. • In the message display area, messages associated with operation are displayed. • When the display area shows items or numerical values to change or set, move cursor with keys. Message display area Status display area Setting item/selection item display area Cursor Fig5-4.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 4-6 <4. OPERATION> 4.4 General operation l Measuring mode In the measurement mode, up to five channels are displayed on a single page. To view further channels beyond the first five ones, press keys to scroll. Measuring mode NOX Ch 0-100 SO2 Ch 0-100 Ch CO Ch O2 0-100 0-25 Ch CV NOX 0-200 Ch SO2 Ch CO 2 0-100 3 0-100 O2 Ch 4 0-25 Ch CVNOX 5 0-200 Ch CVSO2 6 0-200 Ch CV SO2 6 0-200 Ch CV CO 7 0-200 Ch CV NOX 8 0-100 Ch 9 Ch AV CV SO2 AV CV CO 0-100 10 0-100 0.0 0.0 0.0 0.0 0 0.0 0 0.0 0.0 0.0 0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ppm ppm ppm vol% vol% ppm ppm vol% See 6.1 See 6.2 ppm ppm ppm ppm 1h ppm 1h ppm 1h ppm or l User Mode Switch Ranges Calibration Parameters Alarm setting* Setting of Auto Calibration* Setting of Auto Zero Calibration* Peak Alarm Setting Parameter Setting. User Mode Select an item with UP/DOWN and ENT Back with ESC Switch Ranges Calibration Parameters Alarm Setting Setting Auto Calibration Setting Auto Zero Calibration Setting Peak Alarm Parameter Setting For the further information, refer to “Chapter 5. Set up (User mode)”. F0805.ai * : Availability depends on specification. l How to close the mode Press to cancel the operation midway or complete setting. A previous screen will return. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5. 5-1 <5.SETUP (User Mode)> SETUP (User Mode) In User Mode, select an item and go to each setting. (1) On the measuring mode, press MODE display User Mode. and (2) Move cursor and select an item to set. Press . User Mode Select an item with UP/DOWN and ENT Back with ESC Switch Ranges Calibration Parameters Alarm Setting Setting of Auto Calibration Setting of Auto Zero Calibration Peak Alarm Setting Parameter Setting ( ) to each screen IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.1 5-2 <5. SETUP (User Mode> Switch Ranges See “5.7.3 Maintenance Mode” on page 5-28 when you change a measuring range. This section describes how to change over preset range. MR: manual, select range manually on the current screen. RR: remote, select range according to remote range switch contact input. AR: auto, when the measured concentration value exceeds 90 % of Range1, switch automatically from Range1 to Range 2. When the measured concentration value is smaller than 80 % of Range1, automatically switch from Range2 to Range 1. User mode Switch Ranges Switch range (1) Move cursor with keys. Select Ch(components) and press . Ch1 NOx Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select range with UP/DOWN and ENT Back with ESC 0–50.00 Range1 MR 0–500.0 Range2 0–50.00 Range1 AR 0–500.0 Range2 0–10.00 Range1 RR 0–20.00 Range2 Range1 0–50.00 MR Range2 0–500.0 Range1 0–10.00 MR Range2 0–25.00 ( (2) Move cursor with keys. Select one out of MR, AR, or RR and press inverted.. . The selected one is (3) When you select MR, Range1 or Range 2 are inverted. Move cursor with keys. Select Range 1 or Range2 and press .Measuring Switch range Ch1 NOx Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 MR AR RR AR AR (4) After selection, press the key. CAUTION Regardless of how you switch ranges , the range of O2 correction values or O2 correction average values is automatically switched by changing the range of corresponding component. ) Select method of Switch ranges with UP/DOWN and ENT. Back with ESC Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–10.00 vol% Range2 0–20.00 vol% Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–10.00 vol% Range2 0–25.00 vol% ( ) End of Swich Ranges starts with the selected range. (3) The “Channel Selection” screen appears. Press the or key until the cursor goes to a desired Ch (component). ppm ppm ppm ppm vol% vol% ppm ppm vol% vol% Switch range Ch1 NOx Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 MR AR RR MR MR Select range with UP/DOWN and ENT Back with ESC Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–10.00 vol% Range2 0–20.00 vol% Range1 0–50.00 ppm Range2 0–500.0 ppm Range1 0–10.00 vol% Range2 0–25.00 vol% ( ) End of Swich Ranges IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-3 <5.SETUP (User Mode)> l Range identification contact operation Regardless of how you switch ranges (MR, AR, RR), the range identification contact output corresponding to each Ch (component) is conductive when Range 1 is active, and open when Range 2 is active. If the measurement value is held by remote hold contact input or during calibration routine and range switch conditions are met, the contact will change position only after the hold condition is removed. (See “3.5 Wiring” on page 3-7) 5.2 Calibration Parameters Set Calibration concentration and other value. User mode Calibration Parameters Move cursor with keys. Select an item among Calibration Value, About ZERO Calibration, About Calibration Range, About Calibration Components / Range. Select an item with UP/ Cal. Parameters Down and ENT Back with ESC Calibration Value About ZERO Calibration About Calibration Range Auto Calibration Components / Range ( ) screen of each mode IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-4 <5. SETUP (User Mode> 5.2.1 Calibration Value It allows you to set concentrations of the standard gas (zero and span) of each channel used for calibration. (1) Move cursor with keys. Select User mode Cal.Parameters Cal.Value Ch(components) and press . Cal. Settings Cal. Value Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select an item with UP/ DOWN and ENT Back with ESC RANGE 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10vol% 0–20vol% 0–50.00ppm 0–500.0ppm 0–10vol% 0–25vol% ZERO +0000. 0 +00000 +0000. 0 +00000 +00. 000 +000. 00 +0000. 0 +00000 21. 00 21. 00 ( keys and select a (2) Move cursor with concentration to set. Press . then the value is inverted. (3) Enter concentration value of ZERO or SPAN. Enter keys. numerical value in each digit with Press to shift to another digit. After entering, press .to settle the value. Next calibration will use those values. • Applicable value range ( NO EXCEPTIONAL ranging value other than these) NOx, SO2, CO2, CO, CH4, and built-in O2 sensor Span gas:  to 105% FS (full scale of each 1 range) Zirconia O2 measurement Zero gas: Span gas: 5 to 25 vol% (21.00 vol% for atmospheric air) 0.01 to 5 vol% Cal. Settings Cal. Value Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 SPAN 0050. 0 0500. 0 0050. 0 0500. 0 10. 000 020. 00 0050. 0 0500. 0 01. 00 01. 00 ) Select an item with UP/ DOWN and ENT Back with ESC RANGE 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10vol% 0–20vol% 0–50.00ppm 0–500.0ppm 0–10vol% 0–25vol% ZERO +0000. 0 +00000 +0000. 0 +00000 +00. 000 +000. 00 +0000. 0 +00000 21. 00 21. 00 SPAN 0050. 0 0500. 0 0050. 0 0500. 0 10. 000 020. 00 0050. 0 0500. 0 01. 00 01. 00 Cursor for setting value Cal. Settings Cal. Value Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Set calibration value RANGE 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10vol% 0–20vol% 0–50.00ppm 0–500.0ppm 0–10vol% 0–25vol% ZERO +0000. 0 +00000 +0000. 0 +00000 +00. 000 +000. 00 +0000. 0 +00000 21. 00 21. 00 SPAN 0050. 0 0500. 0 0050. 0 0500. 0 10. 000 020. 00 0050. 0 0500. 0 01. 00 01. 00 END of (setting) Calibration Value CAUTION For Zirconia O2 measurement, set the concentration value of ZERO to 21.00 ( when atmospheric air is used). Set to the concentration value as indicated on the cylinder when cylinder air is used. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.2.2 5-5 <5.SETUP (User Mode)> About ZERO Calibration Manual calibration has two mode, “at once” and “each.” “At once” enables you to calibrate some desired component all at once. “Each” enables you to select each component and calibrate one by one. User mode Cal.Parameters Zero Calibration keys. Select (1) Move cursor with Ch(components) and press . Cal. Settings ZERO Cali. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select Ch No. for setting calibration value Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( (2) Move the cursor to select At Once or Each. •Setting “each” Select the Ch (component) on the manual zero calibration screen and then perform the zero calibration. •Setting “at once” At a manual zero calibration, Ch (components) for which “at once” was selected can simultaneously be zero-calibrated NOTE Cal. Settings ZERO Cali. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 at once at once at once each ) Select each or both Ch at ZERO Calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( Set “At Once” when you use cylinder or ambient air as zero gas. at once at once at once at once at once each ) End of Zero Calibration Setting .to activate the designated calibration. (3) After setting completion, press l Manual calibration (See “6. CALIBRATION” on page 6-1) When selecting “each” on every Ch. When selecting “at once” on each Ch only one cursor is inverted. all cursors to show selected ranges are inverted. ZERO Cali Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 ENT ; Go on calibration of selected CH ESC : Not calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0 0 0. 00 0 21. 00 ZERO Cali Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 ENT ; Go on calibration of selected CH ESC : Not calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0 0 0. 00 0 21. 00 F0907.ai  IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-6 <5. SETUP (User Mode> 5.2.3 About Calibration Range This mode is used to set if the range of each CH (component) at the zero and span calibration (manual calibration or auto calibration) should be calibrated with a single range or 2 ranges. User Mode Cali.Value Call range (1) Move cursor with keys. Select Ch(components) and press . Cali. settings Cali range (2) Select Both or Current with keys. • When selecting “both”, Range 1 and Range 2 of the set Ch are calibrated together as zero or span calibration. • When selecting “current”, the range alone displayed at the set Ch is calibrated. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select Ch No. Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( Cali. settings Cali range both current current both current ) Set calibration range current or both range Ch1 NOx Range 1: 0 to 50 ppm Range 2: 0 to 500 ppm both Ch2 SO2 Range 1: 0 to 50 ppm Range 2: 0 to 500 ppm current Ch1: Range 1 and Range 2 are calibrated together, with zero or span calibration. Ch2: Only currently displayed range is calibrated, with zero or span calibration. (3) After setting completion, press calibration range. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( .to save the both current current both current ) End of Calibration Range setting l Manual Calibration (see “6. CALIBRATION” on page 6-1) When you set “both” for Nox or CO, two highlighted cursors appear in both ranges. ZERO Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 ENT: Go on Calibration of selected Ch ESC : Not calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0 0 0. 00 0 21. 00 F0910.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.2.4 5-7 <5.SETUP (User Mode)> Auto Calibration Component / Range Select the Ch (component) and the range for which auto calibration is to be performed. The Ch for which “AR” has been selected as range switch mode is calibrated in the range set here. Auto calibration and the manual calibration of the component for which “AR” has been selected as range switch mode are performed in the range selected here. In this case, once the calibration is started, the range is automatically switched, and on completion of the calibration, the original range is resumed. The range identification contact is interlocked with the range after the switch. However, if the hold setting is set to “ON,” the contact status before calibration is maintained. (1) Move cursor with keys. Select Ch(components) and press . (2) Select range to calibrate and press . Calibration proceeds within the selected range. (3) During (2)(above), press “disable”. Press . (4) Move cursor with “disable”, and press . to invert “enable” or keys to select “enable” or When “enable” is selected, Auto calibration is made in the following rules according to the setting. • Zero calibration is carried out on Chs at once which were set to enable . • The span of Chs (components) which were set to enable is calibrated in the ascending order of Ch number. NOTE User Mode Cali.Value Auto Cal.Comp./ Range Cali.setting Auto Cali. /range Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select Ch No. Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( Cali.setting Auto Cali. /range Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 enable enable enable enable enable ) Select range for auto calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% enable enable enable enable enable ( The components which were set to enable are calibrated at once at the time of auto calibration or auto zero calibration, regardless of setting in 5.2.2 Setting zero calibration . ) End of Auto Cal.range setting Cali.setting Auto Cali. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Set enable or disable for auto calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( enable enable enable enable enable ) End of Auto Cali. Component setting IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.3 5-8 <5. SETUP (User Mode> Alarm Setting This setting does not apply to IR250-B. This mode is used to set the upper and lower limit value to provide an alarm output against measured concentration during measurement. Before changing the alarm setting, set the ON/ OFF to OFF. User mode Alarm setting Alarm setting (1) Move cursor with No. and press . Select Alarm or Hysteresis setting keys. Select Alarm Alarm-1 Alarm-2 Alarm-3 Alarm-4 Alarm-5 Hysteresis 00 %FS 00 %FS ( See the next page (2) Select item to set and press setting is completed. . Now alarm l Setting items Alarm number corresponds to the same number of alarm contact. Channel: Channel setting targeted for issuance of alarm. One Ch No. can be selected for multiple alarms. H-Limit value: S  ets the high limit value (concentration) of alarm. Settable range is 0 to 100 % FS. Settable on each range. L-Limit value: S  ets the low limit value (concentration) of alarm. Settable range is 0 to 100 % FS.Settable on each range. Kind of Alarm: Select one among High limit alarm, Low limit alarm, and High limit or Low limit alarm, HH limit alarm, and LL limit alarm. Alarm Setting Alarm-1 ) Select an item with UP/DOWN and ENT Back with ESC Channel Ch1 H-Limit Range1 100.0 ppm Range2 2000 ppm L-Limit Range1 000.0 ppm Range2 0000 ppm Kind of Alarm  High ON/OFF  OFF ( Cursor for setting value Alarm Setting Alarm-1 ) Select an item with UP/DOWN and ENT Back with ESC Ch1 Channel H-Limit Range1 100.0 ppm Range2 2000 ppm L-Limit Range1 000.0 ppm Range2 0000 ppm Kind of Alarm  High ON/OFF  OFF End of Alarm setting High, HH …… Alarm contact closes when above H-limit alarm. Low, LL …… Alarm contact closes when below L-limit alarm. High or Low… Alarm contact closes when above H-limit value or below lower limit value. ON/OFF: Enables the alarm function if set at ON, or disables it if set at OFF. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-9 <5.SETUP (User Mode)> NOTE Set the values so that upper limit value > lower limit value. Also set the value (upper limit valuelower limit value) > hysteresis error. l When an alarm occurs (sample screen image) When an upper limit alarm occurs, the “H-alarm” message lights. (Lower alarm message appears L-alarm. High alarm” message lights at CH. ( HH-alarm, LL-alarm are available as well for HH or LL limit.) Ch ---NOX H-alarm 0-100 Ch SO2 NOTE Ch CO2 For 10 minutes after turning on power, the alarm judgment is inactive. Ch CO Ch O2 0-100 0-10 0-100 0-25 ppm 0.0 0.0 0 3 0.0 2 1. 0 0 ppm vol% ppm vol% F0915.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-10 <5. SETUP (User Mode> n Hysteresis setting To prevent chattering of an alarm output near the alarm setting values, set hysteresis. 0 to 20% of full scale [% full scale (% FS)] represents the percentage with the width of the component measurement range regarded as 100%. NOTE Hysteresis is common to all alarms. (1) Move cursor with and press . keys. Select hysteresis. Alarm Setting Select Alarm or Hysteresis setting Alarm-1 Alarm-2 Alarm-3 Alarm-4 Alarm-5 Hysteresis (2) Enter hysteresis value with keys in each digit. Press to switch to another digit. Press .to validate the value. 00 %FS ( Alarm Setting ) ENT Select Alarm or Hysteresis setting Alarm-1 Alarm-2 Alarm-3 Alarm-4 Alarm-5 Hysteresis 00 %FS ENT End of Hysteresis setting F0916.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-11 <5.SETUP (User Mode)> l Hysteresis (In case of upper limit alarm) If hysteresis values exceed the upper limit as shown in the graph below, an alarm output is provided. Once the alarm output is turned “ON”, it remains “ON” until the value falls below the set lower limit of the hysteresis indication. Upper limit value Hysteresis width ON (conductive) Alarm OFF (open) F6_3_2_1.ai 5.4 Setting Auto calibration This calibration does not apply to IR250-B. CAUTION • When an auto calibration starts, the measurement screen appears automatically. • During Auto calibration, none of keys works other than Key Lock ON/OFF (see 5.7) or key to stop forcedly auto calibration. “Stop Auto Calibration” cannot be performed with the key lock to ON. To cancel auto calibration forcedly, set the key lock to OFF and then execute “Stop Auto Calibration”. • When power failure or power off occurs, after resuming, the auto calibration starts again at previously-set start time and repeats on the set cycle. Remote start Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the remote start input short-circuited for at least 1.5 seconds. With input (Hold at least 1.5 sec) Remote start input Without input F0921.ai It is carried out automatically at the timing set for ZERO / SPAN calibration. To change the setting of auto calibration, switch auto calibration setting ON/OFF to “OFF” and change the parameter. Auto calibration status contact output is conductive during auto calibration (NO side), and is open in other cases.(See 3.5 Wiring) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-12 <5. SETUP (User Mode> User Mode Auto Cali.Setting (1) Move cursor with press . Set Auto Cal. keys. Select item and First Set ON/OFF to “OFF”, start changing value. Select setting item Start Time Cycle Flow Time ON/OFF SUN 12 : 00 07 day OFF Time:MON 12 : 34 (2) Enter numerical values with shift to another digit. . Press to ( See next page ) Set Auto Cal. Select Start Time Setting items • Start Time : S  etting at the first calibration (day of the week, hour, minute) • Cycle : A period between the start time of one calibration and another (unit : hour/day) • Flow Time : The time required for the calibration gas to be replaced in the cell • ON/OFF : Auto Calibration Run Auto zero calibration ON or OFF (3) After setting, press . The set value will be activated next time you enter the auto calibration. Set back ON/OFF to “ON”. Start Time Cycle Flow Time ON/OFF SUN 12 : 00 07 day OFF Time:MON 12 : 34 Auto Calibration Run End of Auto Calibration setting IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-13 <5.SETUP (User Mode)> n Setting Flow rate (1) Select “Flow time” and press .Go to set Flow time. User Mode Auto Cali.Setting Flowtime Set Auto Cal. ZERO Ch1Span Ch2Span Ch3Span Ch4Span Ch5Span Ex. time Select setting item 350 sec. 350 sec. 350 sec. 350 sec. 300 sec. 300 sec. 300 sec. ( (2) Move cursor with press . keys. Select gas. and Set Auto Cal. ZERO Ch1Span Ch2Span Ch3Span Ch4Span Ch5Span Ex. time (3) You can change the parameter, which is inverted. (60 to 900 sec) ) Set flow item of calibration gas 60 to 900 sec 350 sec. 350 sec. 350 sec. 350 sec. 300 sec. 300 sec. 300 sec. End of Gas flow time setting NOTE Set the flow time value of Zero / Span so that the value is five times larger than that of Response (5.7.2) (4) After changing values, press . Press to go back to Auto Calibration Setting. CAUTION • The Chs only currently being used are displayed on the screen. • The Ex. time is the output signal hold extension time after the completion of calibration. It is valid only when the hold setting is set to “ON.” . IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-14 <5. SETUP (User Mode> l Example Start Time Cycle Flow Time SUN 1 12:00 day 350 sec. 350 sec. 350 sec. 350 sec. 300 sec. 300 sec. 300 sec. ZERO Ch1Span Ch2Span Ch3Span Ch4Span Ch5Span Ex. time ON/OFF ON If you perform auto calibration as shown above, Sunday 12:00 Monday 12:00 Cycle Tuesday 12:00 :Auto Calibration Ch1 Zero Cal Span Ca 350sec. 350sec Ch2 Span Ca 350sec Ch3 Span Ca 350sec Ch4 Span Ca 300sec Ch5 Replace Span Cal ment time 300sec 300sec Flow time F0920.ai The table (above) shows an example when Ch1 through Ch5 are set “enable” in an auto calibration setting of item 5.2.4. n Execute auto calibration and cancellation This mode is used to perform auto calibration one time or cancel the auto calibration forcibly. l One-time auto calibration (1) In User mode, go to Set Auto Cal, and Auto Calibration. Press . Set Auto Cal. Select setting item SUN 12 : 00 07 day Start Time Cycle Flow Time ON/OFF OFF Time:MON 12 : 34 Auto Calibration Run ( Set Auto Cal. (2) “Run” is highlighted and a message appears. Press .to execute Auto Calibration or to cancel. ) Auto Cal Run ENT:Run / Stop ESC:Cancel Start Time Cycle Flow Time ON/OFF SUN 12 : 00 07 day OFF Time:MON 12 : 34 Auto Calibratio Run IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-15 <5.SETUP (User Mode)> l Forced cancellation of auto calibration (While Auto Calibration being conducted) (1) In the User mode, go to Set Auto Cal, and Auto Calibration. Press . If you do while auto calibration is being conducted, Auto Calibration Stop is displayed. Set Auto Cal. Select setting item SUN 12 : 00 07 day Start Time Cycle Flow Time ON/OFF OFF Time:MON 12 : 34 Auto Calibration Run ( Set Auto Cal. (2) “Stop” is inverted. A message appears, prompting you to verify that you want to cancel or continue auto calibration. To cancel the auto calibration forcibly, press the . ey. If you press the key, auto calibration is cancelled (not stopped). NOXCal ZERO 0-100 SO2Cal ZERO 0-100 Ch CO2 Ch CO Ch O2 0-10 0-100 0-25 Ch NOXCal SPAN 0-100 Ch SO2 Ch CO2 Ch CO 0-100 0-10 • Ch1 span calibration A message, “Span cal.” blinks at Ch1. 0-100 Ch O2 0-25 Ch NOX Ch SO2Cal SPAN 0-100 0-100 • Ch2 span calibration A message, “Span cal.” blinks at Ch2. OFF Auto Calibratio Stop Ch A message, “Zero cal.” blinks at Ch1 and Ch2. SUN 12 : 00 07 day Time:MON 12 : 34 Ch • Zero calibration Auto Cal Run ENT:Run / Stop ESC:Cancel Start Time Cycle Flow Time ON/OFF n Auto Calibration screen” When the auto calibration components (see Section 5.2.4) to “Ch1 is “enable” and “Ch2 is “enable” , ) Ch CO2 Ch CO 0-10 0-100 Ch O2 0-25 0.5 0.3 0.0 0 0 0.0 2 1. 0 2 9 0.8 0.0 0.0 0 0.0 0.0 0 0.0 9 5.0 0.0 0 0.0 0.0 0 ppm ppm vol% ppm vol% ppm ppm vol% ppm vol% ppm ppm vol% ppm vol% F0924.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.5 5-16 <5. SETUP (User Mode> Setting of Auto Zero Calibration This function is not available in IR250-B. CAUTION • When an auto zero calibration starts, the measurement screen automatically appears. • Any operation other than “Auto Zero Calibration Stop” (see Section 6.5.2) is not permitted during auto zero calibration. “Auto Zero Calibration Stop” cannot be performed with the key lock set to ON. To cancel auto zero calibration forcibly, set the key lock to OFF and then execute “Auto Zero Calibration Stop”. • When power is turned off (including power failure) and resumes, auto zero calibration is carried out at the preset start time and repeats in the programmed cycle. • If the auto calibration period and auto zero calibration period have overlapped, the auto calibration is retained, ignoring the auto zero calibration of that period. • When the hold setting is set to ON, the hold time of auto calibration contact and measurement value output signal is extended after calibration for gas replacement time. Remote start Whether the auto zero calibration is set at ON or OFF, an auto zero calibration is available by remote start input. With input (Hold at least 1.5 sec) Remote start input Without input F0921.ai Auto zero calibration is automatically carried out when zero calibration are set. Components for which a calibration is to be made are determined by setting of auto calibration component in 6.2.5. Before changing the setting of auto zero calibration, set the ON/OFF to “OFF”. In Auto calibration status, contact output is conductive during auto zero calibration (NO side), and is open in other cases.(See 3.5 Wiring) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-17 <5.SETUP (User Mode)> (1) Move cursor with and press . First Set ON/OFF to “OFF”, start changing value. (2) Enter numerical values with . Press to shift cursor to right. Setting items • Start Time : S  etting at the first calibration (day of the week, hour, minute) • Cycle : A period between the start time of one calibration and another (unit : hour/day) Setting range ; 1 to 99 hours or 1 to 40 days(initial : 7 days) • Flow Time : The time required for the calibration gas to be replaced in the cell Setting range ; 60 to 900 sec for each (initial ; 300 sec) • ON/OFF : User Mode Set Auto Zero Cal keys. Select item Auto zero calibration ON or OFF (3) After setting completion, press . The set value will be activated next time you perform auto zero calibration. Set back ON/OFF to “ON”. Set Auto Zero Cali Start Time Cycle Flow Time ON/OFF Select setting item SUN 12 : 00 07 day OFF Time:MON 12 : 34 Auto Zero Calibration Run See next page Set Auto Zero Cali Set start time Start Time Cycle Flow Time ON/OFF SUN 12 : 00 07 day 300 s OFF Time:MON 12 : 34 Auto Zero Calibration Run End of ZeroCalibration setting l Example Start time SUN 12:00 Cycle 12 hour Flow time 300 sec ON / OFF ON when you perform Auto Zero Calibration under the condition above SUNDAY 12:00 Cycle MONDAY 00:00 MONDAY 12:00 :Auto Zero Calibration Zero Cal Replacement time 300sec 300sec Flow time Gas replacement time after calibration is same as the value F0926.ai of the flow time. (An example of “Ch1: through Ch5: enable,” as given in item 5.2.4 “Setting of auto calibration components / range”) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-18 <5. SETUP (User Mode> n Forced run / stop of auto zero calibration This mode is used to cancel the auto zero calibration forcibly or perform auto zero calibration one time only. l Execution auto zero calibration (one time only) (1) In the User mode, go to Set Auto Cal, and Auto Zero Calibration.Press . If you do while auto zero calibration is being conducted, “Auto Zero Calibration Stop” is displayed. Set Auto Zero Cal. Select setting item SUN 12 : 00 07 day Start Time Cycle Flow Time ON/OFF OFF Time : MON 12 : 34 Auto Zero Calibration Run ( Set Auto Cal. (2) “Run” is inverted. A message appears, prompting you to verify that you want to continue auto zero calibration. Press the to execute the auto zero calibration, and to cancel. ) Auto Cal Run ENT : Run / Stop ESC : Cancel SUN 12 : 00 07 day Start Time Cycle Flow Time ON/OFF OFF Time : MON 12 : 34 Auto Zero Cal. Run l Forced cancellation of auto calibration (While Auto Calibration being conducted) (1) In the User mode, go to Set Auto Cal. Move cursor down to Auto Calibration.Press . If you do while auto calibration is being conducted, “Auto Calibration Stop” is displayed. Set Auto Cal. Select setting item SUN 12 : 00 07 day Start Time Cycle Flow Time ON/OFF OFF Time:MON 12 : 34 Auto Calibration Run ( Set Auto Cal. (2) “Stop” is inverted. A message appears, prompting you to verify that you want to cancel or continue auto calibration. To cancel the auto calibration forcibly, press the . If you press the , auto calibration is cancelled (not stopped). ) Auto Cal Run ENT:Run / Stop ESC:Cancel Start Time Cycle Flow Time ON/OFF SUN 12 : 00 07 day OFF Time:MON 12 : 34 Auto Calibratio Stop IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-19 <5.SETUP (User Mode)> n Auto zero calibration (sample image) In case where setting the auto calibration components (see Section 5.2.4) to “Ch1: enable” and “Ch2: enable • Zero calibration A message, “Zero cal.” blinks at Ch1 and Ch2 Ch NOXcal ZERO 0-100 Ch SO2cal ZERO 0-100 Ch CO2 Ch CO 0-10 0-100 Ch O2 0-25 0.5 0.3 0. 0 0 0.0 2 1. 0 2 ppm ppm vol% ppm vol% F0929.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.6 5-20 <5. SETUP (User Mode> Peak Alarm Setting Peak alarm appears only when option code (/A) is specified. During a measurement when the peak count exceeds the setting value to indicate the CO concentration is over the upper limit, the peak alarm is output. This section describes how to perform various settings of peak alarm. NOTE The peak count is not implemented during the first 10 minutes after the power is turned on. Peak Alarm : Setting of peak alarm is performed with ON/OFF Alarm Value : If a peak concentration exceeds the set alarm value one time, the peak counter increments by 1. settable value 0 to 1000 ppm  5 ppm step (initial value: 500 ppm) Alarm Count : When a peak in excess of the setting time occurs, a peak count alarm is output. settable 1 to 99 counts (initial value: 5 counts) Hysteresis : To prevent possible chattering when the peak value may exceed the set peak concentration by only 1 time, the peak count has an allowance in the hysteresis width. 0 to 20% of full scale (initial value: 0% of full scale) [% full scale] represents the percentage with the CO range regarded as 100%. User mode Peak Alarm setting Peak Alarm (1) Point the cursor to an item to set. Press Select setting item . Peak Alarm Alarm Value Alarm Count Hysteresis OFF 0500 ppm 05 times 00 %FS ( Set Peak Alarm ON or OFF Peak Alarm (2) Execute setting or enter each numerical value. to write the value in each digit and Move press to shift between digits. Press to store values. ) Peak Alarm Alarm Value Alarm Count Hysteresis OFF 0500 ppm 05 times 00 % FS ( ) End of Peak Alarm Setting IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-21 <5.SETUP (User Mode)> l Sample Action of peak alarm (1) A peak count alarm is conductive at 5 times/hour. (ON) (2) A peak count alarm is open at 4 times/hour. (OFF) Peak concentration Measured values 3 hours ago 2 hours ago 1 hour ago Current hour Peak count alarm F6_6graph.ai If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set times per hour, a peak alarm contact output becomes closed (ON). If it is less than the set section from times per hour, it is open (OFF). Since 5 times of peaks /hour is marked at (1) the above graph, the peak count alarm is turned ON. Since peaks of more than 5 times per 1 hour occur at the interval between (1) and (2) , the peak count alarm remains ON. Since at (2), peaks are reduced to 4 times per hour, it is turned OFF. Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when measured gas is fluctuated near the alarm value. l Releasing peak count alarm To release the peak count alarm, set the peak alarm to OFF. Turning on the peak alarm initiates counting from 0. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-22 <5. SETUP (User Mode> 5.7 Parameter Setting It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be set are as follows: Current Time : Sets the current day of the week, hour and time. Key Lock : Output Hold : determines ON/OFF of key lock operation except the key lock OFF determines whether Calibration Output is held or not. (See 5.7.1) Response time : Sets the response time of electrical system. Range ; NDIR measurement component: 1 to 9 cycles (initial: 1 cycle) O2 sensor : 1 to 60s (initial : 15s) * Cycle value can be set by switching from Maintenance mode to Sample Switch setting. One cycle value equals a sum of (s):Flow time of reference gas plus (s): Flow time of Sample gas. Average Period : Sets the moving average time. Range ; 1 to 59 minutes or 1 to 4 hours (initial ; 1 hour). Value by unit , 1 to 59 min, 1 to 4 hour. Back Light : Set display automatic light off and the time until backlight out. Contrast : Adjust contrast of LCD Maintenance mode : Requires passwords to go to the Maintenance mode.(See 5.7.3) Range ; 0000 to 9999 (initial : 0000) (1) Move cursor with . to an item to set. Press User mode Parameter Parameter Select setting item Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF OFF Current ON 05min 0000 ( Parameter (2) Enter numerical value and execute setting. Change parameter or enter value by keys. To move cursor to the right, press . Press to establish value. ) Set day of week Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF OFF Current ON 05min 0000 End of Parameter Setting IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.7.1 5-23 <5.SETUP (User Mode)> Output Hold By setting an output hold to ON, an output signal of each channel are held during the calibration (manual calibration and auto calibration) and for the gas flow time (refer to Item 5.4, Setting of Auto Calibration). Regardless of Hold ON/OFF setting, an output signal can be held via an external input. You can select the value for hold from the value immediately before entering output hold, “current”, and arbitrary value, “setting.”. If you cancel calibration after the calibration gas starts being supplied regardless automatically or manually, the holding extension time will be performed. Instantaneous measurement value cannot be held. Output only can be held. If set value is selected for hold, instantaneous O2 correction value is calculated and held based on the set value. Range identification contact output cannot be switched even if the range is switched during the hold. l Screen display during Holding The “Hold ON” message blinks on the measuring screen. Since the screen displays the process of calibration during the manual calibration, “Hold ON” is not displayed even if the output signal is held, but the screen is displayed with the hold extending time. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-24 <5. SETUP (User Mode> l How Output Hold works • Manual calibration or to flow gas to calibrate then press Calibration Hold extending time Output hold Hold for a length of the time set at gas flow times (see 5.4 Auto Calibration Setting) F0934.ai • Auto Calibration Auto calibration start Auto calibration end Calibration Hold extending time Output hold Hold for a length of the time set at gas flow times F0935.ai • Remote hold with input Remote hold input Output hold without input F0936.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-25 <5.SETUP (User Mode)> (1) Point the cursor to Output Hold and press User mode Parameter . Select Hold ONOFF Parameter (2) “ON” or “OFF” is highlighted. Select one with keys. (3) Press to go select “Current” or “Setting”. Current : Hold the value immediately before the hold Setting : Holds the value arbitrarily set. Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF ON Pevious value ON 05分 0000 Select Hold setting Parameter Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF ON Setting ON 05分 0000 ( Parameter Hold (4) Press to go to the screen to set value. (5) Move cursor with (component). Press Select Ch No. NOX SO2 CO2 CO O2 010 020 015 012 022 %FS %FS %FS %FS %FS ( ) keys and select a Ch (6) A value is highlighted and changeable. Enter parameter in each digit with keys. Press to shift to another digit. After changing parameter, press . NOTE The setting is expressed as 1/1 full scale range for both respective ranges. When 0 to 1000 ppm is selected as the range, and 10% FS is selected as hold setting, the output equivalent to 100 ppm is held irrespective of the measurement value at that time. (7) Press screen. Ch1 Ch2 Ch3 Ch4 Ch5 ) to return to the parameter setting Parameter Hold Ch1 Ch2 Ch3 Ch4 Ch5 Set Hold value 0 to 100 %FS NOX SO2 CO2 CO O2 010 020 015 012 022 %FS %FS %FS %FS %FS End of Hold Setting Parameter Setting Screen IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.7.2 5-26 <5. SETUP (User Mode> Response time, Average Period, Backlight, Contrast n Response time Parameter Select Ch No. Response Time The response time of the electrical system can be changed. Setting is available by components. Ch1 Ch2 Ch3 Ch4 Ch5 CAUTION Setting value as second (s) here are rough and not exact ones. The setting value can be modified as requested by customer. NOX SO2 CO2 CO O2 10 20 15 12 15 s 0020 s 0020 s 0020 s 0020 s F0940.ai n Average Period It allows you to set an average period of the average value of O2 correction. It enables you to set an average time of 1 to 59 minutes (1-minute step) or 1 to 4 hours (1-hour step). Parameter Select Ch No Average Period Ch8 Ch9 Ch10 Changing the setting resets the average value of O2 correction and O2 average value. (Pressing the ENT validates the resetting only for components whose setting was changed.) AV CU AV CU AV CU NOX SO2 CO 01 hour 01 hour 50 min l Example of average action Suppose the average period is 1 hour. Reset Average value Time Time F6-7graph2.ai • Sampling occurs every 30 seconds. • Every 30 seconds, the average for last 1 hour (time setting) is output. • At the instant of resetting, zero is assumed for all past values. It means that the average value will not be correct for 1 hour after resetting. l Average value reset This mode is used to clear all average values O2 correction average and O2 average, and restarts averaging. With an option code /K, you can reset from external input. The indication value and output value is 0 ppm, vol% or so at the time of the reset input (based on average period settings). With input (hold at least 1.5 sec) Reset input without input With input, resetting lasts. Shift from “with input” to “without input” to “without input” starts averaging. F0939.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-27 <5.SETUP (User Mode)> n Backlight Timer Backlight of the LCD automatically turns off, whose timer you can setup. When the specified time elapses during the measurement screen display, the backlight is automatically turned off. Press any key to reset backlight OFF. Only when ON is selected, the time to turn off the to set time with backlight is displayed. Press keys. Press to confirm. If you select “OFF”, the backlight does not go off. Select ON or OFF Parameter 12/01/11 WED 13 : 50 OFF ON Previous value Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode ON 05min 0000 F0943.ai n Contrast Adjust the contrast of LCD. Press to settle. adjust. Press Parameter keys to Select ON or OFF Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF ON Previous value ON 05 min 0000 F0944-2.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5.7.3 5-28 <5. SETUP (User Mode> Maintenance Mode In this mode, you can check of sensor input value, display of error log file, set password and others. Set password first. The password is required next time you enter the mode. User mode Parameter Parameter (1) Select the Maintenance Mode . Press Select an item Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode Parameter keys. Press (2) Change parameter with go right. After entering value, press . 12/01/11 WED 13 : 50 OFF ON Previous Value ON 05 min 0000 Select an item to Default password preset at factory is “0000.” You can go to Maintenance Mode before you change the password. Current Time Key Lock Output Hold Response Time Average Period Backlight Timer Contrast to Maintenance Mode 12/01/11 WED 13 : 50 OFF ON Previous Value ON 05 min 0000 ( Maintenance mode (3) Enter the password to go to Maintenance Mode. Select an item to set. Move cursor to the item to set and press . 1. 2. 3. 4. 5. 6. ) Select operating item Sensor Input Value Error log Cal.Log Output Adj. Other Pramameter To Factory Mode ( ) To each Maintenance screen The following shows sample images of each Maintenance Mode. Press to return. CAUTION “6.Factory Mode” is not for customer use, only for our service engineers IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-29 <5.SETUP (User Mode)> 1. Sensor Input Value • Input 1 to 4 : NDIR sensor digital value • Input 5 : O2 sensor digital value Maintenance Sensor input Input 1 Input 2 Input 3 Input 4 Input 5 CAUTION Customers must not change the setting. 100821 96118 102241 82856 11050 Gas  Sample 2. Error Log Error log traces error history of the latest 14 occurrence. Maintenance ENT :Clear Error log Error Log ESC :Bacj Refer to “8 Error message” to know the contents of error. ErrorNo Y M D H M Ch To erase all error in a log, press “Clear Error Log.” after selecting No. 4 04 No. 1 04 No. 6 03 No. 9 03 No. 5 03 No. 9 03 Next Page 2 1 12 12 12 12 11 10 1 1 1 1 18 12 10 10 0 0 10 2 10 10 0 0 5 1 2 2 2 2 Page1 Clear Error Log 3. Calibration Log Calibration Log traces past calibration history. The history data include sensor input value, concentration value, and the date when zero / span calibration was performed. Each Ch(component) records latest 10 calibration data. R : Z  1 : Zero calibration (Z) of Range 1 S1 : Span calibration (S) of Range 1 Z2 : Zero calibration (Z) of Range 2 S2 : Span calibration (S) of Range 2 Cnt :  alue of measuring detector at the time of V calibration Con : Maintenance Cal.Log Ch1 Ch2 Ch3 Ch4 Ch5 Select Ch No NOX SO2 CO2 CO O2 Clear Error Log  oncentration value displayed before C calibration Move the cursor to select Clear Calibration Log and to erase all of the calibration history. press Maintenance Cal.Log Ch1 NOx R Z1 S1 Cnt 485231 441764 Con -0.2 189.5 YDHM 12111810 12111815 F0947.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-30 <5. SETUP (User Mode> 4. Output adjustment Connect the digital multi- meter to the output terminal corresponding to the number of OUT to be adjusted, and adjust the value so that 4mA or 0V is output at zero and 20mA or 1V is output at span. Move cursor with keys to select OUTNo. or . The value you select is ZERO / SPAN, and press highlighted. Adjust the numerical value with keys to shift to another according to output value. Press digit. After the adjustment press . 5. Other parameter Password Set : Set password to move from the parameter setting screen to the maintenance mode. Arbitrary 4-digit number can be selected. O2 ref.value : Set the oxygen concentration reference value at the time of O2 correction calculation. Settable in the range from 00 to 19%.(Initial value : 12%) Maintenance Output Adj. OUT 1 2 3 4 5 6 Zero 01245 01245 01245 01245 01245 01245 Maintenance Output Adj. OUT 1 2 3 4 5 6 Zero 01245 01245 01245 01245 01245 01245 Adjust OUTPUT ZERO and SPAN Span 11845 11845 11845 11845 11845 11845 OUT 7 8 9 10 11 12 Zero 01900 01900 01900 01900 01900 01900 Span 12500 12500 12500 12500 12500 12500 Adjust OUTPUT ZERO and SPAN Span 11845 11845 11845 11845 11845 11845 OUT 7 8 9 10 11 12 Zero 01900 01900 01900 01900 01900 01900 Span 12500 12500 12500 12500 12500 12500 Limit : S  et the O2 concentration limit at the time of O2 correction calculation. Settable in the range from 01 to 20%. (initial value : 20%) * See 4.3 Mode and screens for O2 correction concentration value to calculate O2 correction. This function is not available on IR250-B. . Station No. : Set Station No. in MODBUS. Settable parameter goes 00 to 32. Range setting : Moves to the screen on which measuring range is changed. Set Sample Switching : Moves to the screen on which parameters about Sample switching is changed. (Only IR250-C displays) Maintenance mode setting Select an item Password Set  2465 O2 ref.value 12% O2 Limit 20% O2 Station No.01 Range setting Set Sample Switching F0952.ai keys to select an item to Move cursor with modify. Password, O2 ref. Value, Limit, Station No. are inverted. Move keys to change parameter. Change the parameter and press CAUTION Keep the password in safe and do not forget. Otherwise, you cannot enter the maintenance mode. In case of lost of the password, contact us immediately. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-31 <5.SETUP (User Mode)> l How to set / change range The measuring range can be arbitrarily selected in the minimum and the maximum range specified at the time of purchase. The range to be used can be selected 1 or 2. (1) Move cursor with keys. Go to Range setting and press . Select Ch No. Maintenance Mode Range set keys to select Ch (2) Move cursor with (component) to modify. Press . Ch1 Ch2 Ch3 Ch4 Ch5 NOX SO2 CO2 CO O2 ( (3) Select the value to change and press . • Settable range The value for range 1 and range 2 must fall within the range from the MIN and the MAX range (including the MIN and the MAX range), and at the same time range 1 must be smaller than range 2. The number of ranges is 1 or 2. keys. Press (4) Change the value with to shift to another digit. Unit cannot be changed. When decimal point is highlighted, the decimal point can move. (5) After completion of the setting, press . CAUTION You must conduct Zero / Span calibration each time you change the range setting. Otherwise, the measurement result may not be correct in value. Maintenance range set Ch1 NOx MIN range Range 1 Range 2 MAX range Range num. ) Select range or range num. 100.0 500.0 1000. 2000. 2 ppm ppm ppm ppm ( ) Maintenance range set Ch1 NOx MIN range Range 1 Range 2 MAX range Range num. 100.0 500.0 1000. 2000. 2 ppm ppm ppm ppm End of range setting/change F0954.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-32 <5. SETUP (User Mode> l Sample switch setting Only available for IR250-C. The following pages describe how to set gas flow time, interference compensation coefficient. These values are required for sample switch. Flow time of reference gas and sample gas are switched by sample switch method at the following timing. Measuring time of reference gas Flow time of reference gas Measuring time of reference gas Flow time of sample gas Measuring time of reference gas Flow time of reference gas Measuring time of reference gas Flow time of sample gas close DO26contact open close DO27contact open measuring position 1 cycle F09S4.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-33 <5.SETUP (User Mode)> (1) How to change the setting of Gas Flow time CAUTION Gas flow time is essential value for sample switch. DO NOT CHANGE unnecessarily. Here describes how to set up Reference Gas Measuring Time, Sample Gas Switching Time, Sample Gas Flow Time, Sample Gas Measuring Time. User mode Maint.Mode Sample Switching (1) Move cursor to Gas Flow Time and Press . Maint. Mode Sample Switch setting select an item Gas Flow Time Interference Compensation Coefficient (2) Move cursor with keys to select Reference Gas Switching Time or Reference Gas Measuring time. Reference Gas Switching Time (initial value : 5 sec.) :1 to 30s. Reference Gas Measuring Time (initial value : 5 sec.) :1to 60s. Maint.Mode Sample Switch Gas Flow time Select an item Ref.gas Switching t 05 s Ref.gas Measuring t 05 s Ref.gas Flow time 10 s Sample Switching 05 s Sample Measuring 05 s Sample Flow time 10 s ( (3) Press to settle each value. Reference Gas Flow Time, Sample Gas Switching Time, Sample Gas Measuring Time, Sample Gas Flow Time are calculated according to the following relational expression. Switching time of sample gas (t1) = Switching time of reference gas (t3) ) Set gas replacement time. Maint.Mode Sample Switch Setting range is from 1 to Gas Flow time 30 seconds Ref.gas Switching t Ref.gas Measuring Ref.gas Flow time Sample Switching Sample Measuring Sample Flow time 05 s 05 s 10 s 05 s 05 s 10 s Flow time of sample gas (Ts) = Flow time of reference gas (Tr) Flow time of reference gas (Tr) = Switching time of reference gas (t3) + measuring time of reference gas (t4) Flow time of sample gas (Ts) = Switching time of sample gas (t1) + measuring time of sample gas (t2) 1 cycle = flow time of reference gas (Tr) + flow time of sample gas (Ts) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 5-34 <5. SETUP (User Mode> (2) Interference compensation coefficient (Measuring / Calibration) (1) Move cursor to Interference Compensation Coefficient and press . (2) Move cursor with keys to select Ch (component) to modify. Press to change the value. (3) After changing the value,press value. to settle the User mode Maint.Mode Sample Switching Maint. Mode Sample Switch setting select an item Gas Flow Time Interference Compensation Coefficient ( ) Maint. Mode Select the component to Sample Switch be adjusted. Interference Adj. Ch1 NO Ch2 SO2 Ch3 CO Ch4 CO2 44445 0 44445 44445 IM 11G02Q01-01EN 1.000000 1.000000 1.000000 1.000000 3rd Edition : May 31, 2017-00 6. 6-1 <6. CALIBRATION> CALIBRATION This chapter describes manual calibration. 6.1 Zero Calibration Zero Calibration is for zero point adjustment. Proper zero gas, suitable for the application, should be used. Refer to Section 3.4 “Sampling” (1) On the Measurement screen, press to show Manual Zero Calibration screen. Select Ch No. with UP/DOWN and ENT Back with ESC ZERO Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( keys to select Ch (2) Move cursor with (component) to calibrate. Press then the zero gas will be supplied. CAUTION For the Ch (components) in which “at once” is set in the zero calibration range setting (5.2.2), zero calibration is carried out at once. If “both” is set in the calibration range (5.2.3), both ranges are calibrated simultaneously. (3) Wait until the indication is stabilized with the zero gas supplied. After the indication has been key. Zero calibration in stabilized, press the range selected by the cursor is carried out l How to close Zero Calibration To close the Zero Calibration or cancel the operation midway, press . A previous screen will return. CAUTION Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 ZERO Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 0.0 0. 00 0.0 20. 09 ) Select Ch No. with UP/DOWN and ENT Back with ESC ZERO Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 0.0 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0.0 0.0 0. 00 0.0 20. 09 ENT : Go on calibration of selected CH. ESC : Not calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0.0 0.9 0. 34 1.1 20. 09 to Measuring after Manual Zero Cali. F0955.ai For the Ch (component) for which “AR” is selected in 5.1, the cursor automatically moves to the range selected in “5.2.4 Setting of auto calibration component / range” , and calibration is carried out within that range. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 6.2 6-2 <6. CALIBRATION> Span Calibration Span Calibration is used to perform a span point adjustment. Supply calibration gas with concentration set to the span value to perform the span calibration. For the span calibration gas for the NOx, SO2, CO2, CO, CH4measurement, use the standard gas with a concentration of 90 to 100% of its measuring range value. For the span calibration gas for the O2 measurement, use the standard gas with a concentration of 90 to 100% of its measuring range value when measuring with the built-in O2 sensor, and use the standard gas of 1 to 2 vol% when measuring with an external zirconia O2 sensor. SPAN Cal. (1) On the Measurement screen press the Manual Span Calibration screen. to display (2) Move cursor with keys to select Ch (component) to calibrate. Press then the span gas will be supplied. CAUTION For the Ch (components) in which “both” is set in 5.2.3 Setting of calibration range, span calibration is carried out on both ranges. (3) Wait until the indication is stabilized with the span gas supplied. After the indication has been key. Span calibration in stabilized, press the range selected by the cursor is carried out l How to close Span Calibration TO close the Span Calibration or cancel the operation . A previous screen will return. midway, press Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 Select Ch No. with UP/DOWN and ENT Back with ESC Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% ( Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 SPAN Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 0.0 0. 00 0.0 00. 00 ) Select Ch No. with UP/DOWN and ENT Back with ESC SPAN Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 0.0 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 0.0 0.0 0. 00 0.0 00. 00 ENT : Go on calibration of selected CH. ESC : Not calibration Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 Range1 Range2 0–50.00ppm 0–500.0ppm 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–20.00vol% 0–50.00ppm 0–500.0ppm 0–10.00vol% 0–25.00vol% 98.0 0.0 0. 00 0.0 00. 00 CAUTION For the Ch (component) for which “AR” is selected in “Section 5.1 Changeover of range” the cursor automatically moves to the range selected in “5.2.4 Setting of auto calibration component / range”, and calibration is carried out within that range. to Measuring after Manual SPAN Cali. F0602.ai IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7. 7-1 <7. MAINTENANCE> MAINTENANCE WARNING Purge not only inside of IR250 but all measuring gas lines with zero gas sufficiently, when you provide maintenance or inspection on IR250 with its cover or door open. Otherwise it may cause hazardous accidents such as gas leakage, fire and explosion. CAUTION DO NOT operate IR250 with its cover left open for a long period of time. It may have dust and dirt, which damages the equipment. 7.1 Daily check and maintenance procedures Maintenance and check list Parts to be checked Indication Value Phenomena Indication values are lower. Dust contamination Indication values are higher. in sampling cell Daily Weekly Yearly Causes Remedy Consult us.Clean sampling cell and inspect sampling device, especially gas filter. Air is absorbed midway in the sampling tube. Check for any leak of the sampling line and repair if any. Sample gas flow rate and purge gas flow rate (if purging takes place in the instrument). Standard flow rate is not within the range of the specified flow rate. — Adjust the flowmeter with flow rater needle valve. Zero point of gas analyzer Deviation from zero point — Zero adjustment Span point of gas analyzer Deviation from span point — Span adjustment Gas analyzer — — Overhaul or service in accordance with proper service ● Flow rate check of sampling gas or purge gas. Flow rate of sampling gas must be within 1.0 ± 0.5 L/min. Flow rate of sampling gas must be within 0.5 ± 0.2 L/min. Flow rate of sampling gas or reference gas must be within 1.0 ± 0.2 L/min. Flow rate of purge gas must be approximately 1 L/min. Maintenance and inspection should be carried out every day, if required. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7-2 <7. MAINTENANCE> ● Zero calibration and span calibration For performing a zero calibration, refer to 6.1, Zero calibration. After the zero calibration, perform a span calibration by referring to 6.2 Span calibration. Zero calibration and span calibration should be performed once a week, if required. n Long term maintenance Create a long-term maintenance component procurement plan in advance based on the gas analyzer annual inspection plan described below. Each period described below is standard value, thus each one may vary depending on measuring gas conditions or operating environment. The recommended replacement period is not the warranty period. Consult us for the maintenance service in detail. Recommended replacement period Components Q’ty Recommended replacement period (year) IR light source unit 1 or 2 5 Varies with conditions of measuring gases or their range. O-ring for sampling cell 2 to 5 2 Varies with conditions of measuring gases or their range. Detector 1 or 2 8 Varies with conditions of measuring gases or their range. LCD 1 3 Solvent valve 2 3 only for IR250-C AC/DC main power 1 5 IR250-B need 3 pieces DC/DC power PCB 2 5 Annual inspection — 1 7.2 Notes Replacement of power fuse CAUTION Prior to the following work, be sure to repair blown down fuse (short, etc.), if any. The part number of fuse is K9642CJ. (1) Turn “OFF” the main power switch of the analyzer. Refer to “1.1 Names and description ” to know where each switch or fuse is. (2) Turn the fuse holder cap (shown in the figure above) counterclockwise and pull it out. The cap will be removed. Remove a fuse out of the holder. Replace it with a new one. (250VAC/2A, Time-lag type). (3) Reinstall the fuse holder cap, turn ON the power supply switch. Confirm the analyzer starts up normally. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7.3 7-3 <7. MAINTENANCE> Cleaning of sampling cell Entry of dust or water drops in the sampling cell contaminates the interior of the cell, thus resulting in a drift. Clean the inside if dirty. Then, check the sampling device, especially the filter, to prevent the cell from being contaminated by dust or mist. CAUTION Service personnel from YOKOGAWA will take care of the maintenance. Contact YOKOGAWA representatives for maintenance service. 7.3.1 Disassembly and assembly of sampling cell (Pipe cell, Block cell) There are two kinds of sampling cells, block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and pipe cells (Cell length: 64 mm, 125 mm, 200 mm and 250 mm). 2-component analyzer may incorporate both sampling cells in one unit. In such a case, detach the pipe cell and then block cell CAUTION Special care should be taken when assembling or disassembling the measuring cell to avoid the application of force to the detector pipe or light source unit pipe. If the pipe is deformed or damaged by excessive force, there is a danger of gas leak, thus resulting in faulty operation. l How to remove pipe cell (1) Stop measured gas. If it is harmful, purge the pipe cell thoroughly with zero gas. (2) Turn OFF the power switch. (3) Remove the screws then remove the cover.(Open the front door on IR250-B). (4) Remove the tubes connected to the pipe cell. (5) See Figure. 7.1. Loosen and remove screws (Screw A) from the cell retainer fastening the pipe cell (both ends). (6) Remove the cell from the measuring unit and unscrew the infrared transmission window at both ends in the right direction.Don’t miss the O-ring. (7) For assembly, reverse the disassembly procedure.Make sure to put the space in 0.5mm between light source unit and the pipe cell, the pipecell and the detector. Screw the pipe part of gas inlet with window on the both side to the measuring cell. Screw it so that tubing (both end of pipe cell) and the pipe cell align facing up. (Marking of pipe cell faces up.) Screw A Pipe cell O-ring Cell retainer (Infrared transmission) Window Detector Light source unit Figure 7.1 Sample of Measuring cell( Pipe cell) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7-4 <7. MAINTENANCE> l How to remove block cell Block cell is attached with the detector. Remove them together as a whole unit out of the base plate. (1) Stop measured gas. If it is harmful, purge the pipe cell thoroughly with zero gas. (2) Turn OFF the power switch. (3) Remove the screws then remove the cover.(Open the front door on IR250-B). (4) See Figure 7.2. Remove the tube connected to the block cell. (5) Unscrew 6 screws holding the power board (Screw B) and remove the measuring cell. DO NOT pull it off forcibly, because some cables are still connected. (6) Unscrew the screws (Screw C) on the back of the base plate, holding the detector. The number of detector varies with the specification. One detector is fixed with two screws. (7) Unscrew the two screws (Screw D) holding the detector to the light source unit. (8) Unscrew the two screws (Screw E) holding the detector to the light souce unit and the block cell. Handle the unit horizontally. Otherwise all of the parts may slip off. Remove the detector and the block cell together. Then take away the block cell. Be careful not to miss the small parts. The (infrared transmission) window is easy to fall (not fixed) between the detector and the block cell. Don’t miss the O-ring, either. (9) For assembly, reverse the disassembly procedures. Place the O-ring between the window and the block cell. Make sure the position is correct. If the analyzer has two detectors, install them so that no space is left in-between. Connect cables to the detector correctly. Screw E Block cell O-ring (Infrared transmission) Window Detector Light source unit Screw D Screw B Base plate Screw C Light source power board Figure 7.2 Sample of Measuring cell (Block cell) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7-5 <7. MAINTENANCE> l How to remove pipe cell and block cell First remove pipe cell then remove block cell. Block cell is attached to the detector, so you need to take the whole cell out of the base plate. (1) Follow step on “How to remove pipe cell,” from (1) to (6) to remove the pipe cell. See Figure 7.3. (7) For assembly, reverse the disassembly procedures. (8) See Fig. 7.3.Unscrew the two screws (Screw B) holding the base plate, and take away the measuring cell. DO NOT pull it off forcibly, because some cables are still connected. (9) Unscrew the screws (Screw C) on the back of the base plate, holding the detector. (10) Unscrew the two screws (Screw D) holding the detector to the light source unit. (11) Unscrew the two screws (Screw E) holding the detector to the light souce unit and the block cell. Handle the unit horizontally. Otherwise all of the parts may slip off. Remove the detector and the block cell together. Then take away the block cell. Be careful not to miss the small parts. The (infrared transmission) window is easy to fall (not fixed) between the detector and the block cell. Don’t miss the O-ring, either. (12) For assembly, reverse the disassembly procedures. Place the O-ring between the window and the block cell. Make sure the position is correct. Make sure to put the space in 0.5mm between light source unit and the pipe cell, the pipe cell and the detector. Screw the pipe part of gas inlet with window on the both side to the measuring cell. Screw it so that tubing (both end of pipe cell) and the pipe cell align facing up. (Marking of pipe cell faces up.) Screw E Block Cell (Infrared transmission) Window Detector Light source unit O-ring Pipe cell Screw A Cell retainer Screw D O-ring (Infrared transmission) Window Detector Screw B Base plate Screw C Figure 7.3 Light source power board Sample of Measuring cell (Pipe cell and Block cell) IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 7.3.2 7-6 <7. MAINTENANCE> How to clean sampling cell CAUTION Handle the fragile window with care. Use care not to rub off the dirt from the window roughly. • To clean the sampling cell inside or infrared ray transmission window, first clear large dirt of it with a soft brush and then wipe lightly with soft cloth. Do not use abrasive or paper cloth. • If the window or the sampling cell interior is very dirty, use a soft lint-free cloth moistened with absolute alcohol. • If the window is corroded, rub off the scale from the window lightly with a soft cloth to which chrome oxide powder is applied. If it is excessively corroded, it should be replaced with new one. • When the sampling cell or window cleaning is completed, assemble according to the sampling cell disassembly and assembly procedures. Assemble the pipe carefully. If it becomes bent or damaged, replace it with a new part. • Do not wash the sample cell components with water. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 8. 8-1 <8.ERRROR MESSAGE> ERROR MESSAGE Error display Error contents Probable causes Error No.1 Light source or motor rotation is faulty. • Infrared ray light source is faulty. • Sector motor rotation is faulty or stopped. • Amplifier circuit is faulty. Error No.2 Detector failure. • Detector voltage circuit is faulty. • Detector element is broken or faulty. • Amplifier circuit is faulty Error No.3 A/D error • A/D conversion circuit is faulty. Error No.4 Zero calibration is not within the allowable range. Error No.5 A amount of zero calibration (indication value) is over 50% of full scale. • Zero gas is not supplied. • Zero point is deflected much due to dirty cell. • Detector is faulty. Error No.6 Span calibration is not within the allowable range. Error No.7 A amount of span calibration (difference between indication value and calibrated concentration) is over 50% of full scale. Error No.8 Measured values fluctuate too much during zero and span calibration. • Calibration gas is not supplied. • Time for supplying calibration gas is short. Error No.9 Calibration is abnormal during auto calibration. • Error corresponding to No. 4 to No. 8 occurred during auto calibration. Error No.10 Output cable connection is improper. • DIO circuit is failure. • Internal wiring to the DIO circuit is broken. • Gas is not supplied long enough. • Span gas is not supplied. • Calibrated concentration setting does not match cylinder concentration. • Zero calibration is not performed properly. • Span is deflected much due to dirty cell. • Detector sensitivity is deteriorated. When one of those errors among No.1 to No.3 or No.10 occurs, instrument error (FAULT) contact output is closed. When one of those errors among No.4 to No.9 occurs, calibration error contact output is closed. n Trouble shooting When Error No. 1 occurs, remove the top cover of the analyzer and check the LED on the light source power PCB. If the LCD light is turned off, light source is disconnected. When Error NO.1 to No.3 or No.10 occurs, the analyzer is faulty. Contact us immediately. When Error No.4 to No.8 occurs, the calibration procedure may be wrong. Check the following again. If the error is still indicated, contact us immediately. • Is the calibration gas supplied in the analyzer? • Does the calibration operation match the supplied gas? (For example, zero calibration is performed while flowing the span gas by mistake.) • Does the supplied gas concentration match the gas concentration set at the calibration concentration setting?[ Also, when Error No.5 or NO.7 occurs, you can perform calibration forcibly by following the procedure described in the next section. Use it as fault recovery when calibration fails and calibration contents are missed. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 8-2 <8.ERRROR MESSAGE> n How to operate when error appears l Error No.1 to No.4, No.6, No.8 to No.10 Measurement screen NONo.9 2 Error 0-25 Ch Ch SO2 Ch CO2 Ch CO Ch O2 0-200 0-10 0-200 0-25 0 0.8 1 3.6 0.0 0 0.0 2 1.0 0 Error Contents display Auto Cal. error ESC: Back to MEAS Error No.9 ppm ppm Zero NOx Calibration Error Cause vol% ppm ● Calibration gas is not flowing. ● Gas flowing time is short. vol% ● delete error display. ● Even after pressing if trouble shooting is not completed, error is displayed again.     ● Setting concent. is different from gas concent. ● Lower sensitivity due to sample cell contamination. ● When more than two errors occur at a time, press to move to another error display. l Error No. 5 or No. 7 ZERO Cal. Ch1 NOX Ch2 SO2 Ch3 CO2 Ch4 CO Ch5 O2 ● Press  ENT : Go on calibration of selected Ch. ESC : Not calibration レンジ1 0–200 ppm Error No.5 レンジ2 0–2000ppm Range1 0–200 ppm Range2 0–2000ppm Range1 0–10 vol% Range2 0–20 vol% Range1 0–200 ppm Range2 0–1000ppm Range1 0–10 vol% Range2 0–25 vol% Zero cali. more than 50%FS ENT : Force Cali. ESC : Stop cal. and back to MEAS. NOx Calibration error Cause ● Zero gas is not flowing. ● Deviation of zero point due to contamination. ● Low sensitivity of detector Error No.5 3083 -13.6 -0.006 0.2 -0.09 to delete the error display. Calibrated forcedly Calibration is continued. Unless another calibration error occurs, calibration is carried out to the end, and the Measurement screen returns. Ch NOx Ch SO2 Ch CO2 Ch CO Ch O2 0-25 0-200 0-10 0-200 0-25 IM 11G02Q01-01EN 9 0.8 1 3.6 0.0 0 0.0 0.0 9 ppm ppm vol% ppm vol% 3rd Edition : May 31, 2017-00 8-3 <8.ERRROR MESSAGE> n Error log file If an error occurs, the error history is saved in an error log file. You can find the error log file in the maintenance mode. l Error log screen Maintenance Mode Error Log Errors that occurred ErrorNo. Y No. 10 16 No. 9 16 No. 5 15 No. 9 15 No. 7 15 No. 7 15 Next Page Date and time when an error occurred ENT :Clear Error Log ESC:Back M 2 1 12 12 12 11 D 11 10 1 1 1 2 T 18 12 10 10 0 2 M 10 2 10 10 0 43 Component for which the error occurred. Ch 5 5 1 1 2 Page1 New Old Clear Error Log * Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every time a new error occurs. * If the power supply is turned OFF, the contents in the error log file will not be lost or damaged. They will be saved. l Deletion of error history Press the ENTER key on the Error Log screen (looks like the above one), and the “Error Log Clear” will be inverted. Further pressing the key will clear the error history. IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Blank Page i Revision Information Model IR250 Infrared Gas Analyzer  Title:  Manual No.: IM 11G02Q01-01EN Edition Date Remark (s) 3rd 2nd 1st Revised p.1-2, 1-5, 2-2, 2-8, 2-9, 2-13, 2-16, 3-4, 3-8 Modified in p.2-13, pp.3-9 to 3-12 Newly published May. 2017 Feb. 2017 Feb. 2017 Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN http://www.yokogawa.com/an/ IM 11G02Q01-01EN 3rd Edition : May 31, 2017-00 Blank Page