User’s
Manual
Model IR250
Infrared Gas Analyzer
IM 11G02Q01-01EN
IM 11G02Q01-01EN
3rd Edition
i
u PREFACE
We are grateful for your purchase of Yokogawa’s Infrared Gas Analyzer, Model: IR250. Read this
instruction manual carefully.
The related documents are as follows.
General Specifications: GS 11G02Q01-01EN
User’s Manual:
IM 11G02Q01-01EN
IM 11G02Q01-51EN
IM 11G00E01-01EN
(this manual)
Communication Functions (MODBUS)
ZX8D Zirconia Oxygen Analyzer
The “EN” in the document number is the language code, meaning English
WARNING
This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fire or other serious accidents.
CAUTION
• For installation, observe the rule on it given in the instruction manual and select a place where
the weight of gas analyzer can be endured. Installation at an unsuited place may cause turnover
or fall and there is a risk of injury.
• During installation work, care should be taken to keep the unit free from cable chips or
otherforeign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
WARNING
In piping, the following precautions should be observed. Wrong piping may cause gas leakage.
If the leaking gas contains a toxic component, there is a risk of serious accident being induced.
Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring.
• Connect pipes correctly referring to the instruction manual.
• Exhaust should be led outdoors so that it will not remain in the locker and installation
room.Exhaust from the analyzer should be relieved in the atmospheric air in order that an
unnecessary pressure will not be applied to the analyzer. Otherwise, any pipe in the analyzer
may be disconnected to cause gas leakage.
• For piping, use a pipe and a pressure reducing valve to which oil and grease are not adhering. If
such a material is adhering, a fire or the like accident may be caused.
CAUTION
• Wiring work must be performed with the main power set to OFF to prevent electric shocks.
• Enforce construction of class-D grounding wire by all means. If the specified grounding
construction is neglected, a shock hazard or fault may be caused.
• Wires should be the proper one meeting the ratings of this instrument. If using a wire which
cannot endure the ratings, a fire may occur.
• Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating
may cause fire.
Media No.IM 11G02Q01-01EN
3rdEdition :May 31, 2017 (YK)
All Rights Reserved Copyright © 2016, Yokogawa Electric Corporation
IM 11G02Q01-01EN
3rd Edition : May. 31, 2017-00
ii
WARNING
• Do not smoke nor use a flame near the gas analyzer. Other wise, a fire may be caused.
• Do not allow water to go into the gas analyzer. Otherwise, hazard shock or fire in the instrument
may be caused.
WARNING
• Purge not only inside of IR250 but all measuring gas lines with zero gas sufficiently, when you
provide maintenance or inspection on IR250 with its cover or door open. Otherwise it may
cause hazardous accidents such as gas leakage, fire and explosion.
CAUTION
• Do not operate the analyzer for a long time with its cover or door left open. Otherwise, dust,
foreign matter, etc. may stick on internal walls, thereby causing faults.
CAUTION
• Be sure to observe the following for safe operation avoiding the shock hazard and injury.
• Replacement parts such as a maintenance part should be disposed of as incombustibles. For
details, follow the local ordinance.
l Inspection of product
On receipt of the product, inspect the package and note it has no damage caused by the delivery.
Confirm the specification of the product comforms with your order and accessories are all
delivered. Check the product code labeld on a model plate is correct. See 2.2 Model and Suffix
Code and 2.3 Standard Accessories
IM 11G02Q01-01EN
3rd Edition : May. 31, 2017-00
iii
u
Safety Precautions
n Safety, Protection, and Modification of the Product
• This manual is intended for the following personnel :
• Engineers responsible for installation, wiring, and maintenance of the equipment
• Personnel responsible for normal daily operation of the equipment.
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specified in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modification of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual
in order to avoid the risks for the human body and health including risk of injury, electric
shock, or fatalities. or the damages to instruments. The manual describes what special
care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in
order to prevent the instrument (hardware) or software from being damaged, or a system
failure from occurring.
The following are signal words to be found only in our instruction manuals.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
n Notes on Handling User’s Manuals
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written
consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you find mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
• This manual is an essential part of the product ; keep it in a safe place for future reference.
IM 11G02Q01-01EN
3rd Edition : May. 31, 2017-00
iv
n Drawing Conventions
Some drawings may be partially emphasized, simplified, or omitted, for the convenience of
description.
Some screen images depicted in the user’s manual may have different display positions or
character types (e.g., the upper / lower case). Also note that some of the images contained in this
user’s manual are display examples.
n How to dispose the batteries
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office. Do not
dispose them as domestic household waste.
Battery type:
Poly-carbon monofluoride Lithium battery
NOTICE: The symbol above means they shall be sorted out and collected as ordained in ANNEX II in
DIRECTIVE 2006/66/EC.
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
n Trademark policy
• All names of company, brand of product used in this manual are registered trademarks.
• TM or R to signify brand or trademarks are not used in this manual.
IM 11G02Q01-01EN
3rd Edition : May. 31, 2017-00
v
After-Sales Warranty
Do not modify the product.
Yokogawa warrants the product for the period stated in the pre-purchase
quotation. Yokogawa shall conduct defined warranty service based on its
standard. When the customer site is located outside of the service area, a fee
for dispatching the maintenance engineer will be changed to the customer.
During the warranty period, for repair under warranty carry or send the product
to the local sales representative or service office. Yokogawa will replace or
repair any damaged parts and return the product to you.
Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
If we replace the product with a new one, we won’t provide you with a repair report.
In the following cases, customer will be charged repair fee regardless of
warranty period.
Failure of components which are out of scope of warranty stated in instruction manual.
Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa did
not supply.
Failure due to improper or insufficient maintenance by user.
Failure due to misoperation, misuse or modification which Yokogawa does not authorize.
Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
Failure caused by any usage out of scope of recommended usage
Any damage from fire, earthquake, a storm and flood, lightning, disturbance, riot, warfare,
radiation and other natural changes.
Yokogawa does not warrant conformance with the specific application at the
user site. Yokogawa will not bear direct/indirect responsibility for damage due
to a specific application.
Yokogawa will not bear responsibility when the user configures the product
into systems or resells the product.
Maintenance service and supplying repair parts will be covered for five years
after the production ends. For repair this product, please contact the nearest
sales office described in this instruction manual.
IM 11G02Q01-01EN
3rd Edition : May. 31, 2017-00
Blank Page
i
Model IR250
Infrared Gas Analyzer
IM 11G02Q01-01EN 3rd Edition
CONTENTS
u PREFACE........................................................................................................i
u
Safety Precautions.......................................................................................iii
After-Sales Warranty.....................................................................................v
1.
OVERVIEW................................................................................................. 1-1
2.
3.
4.
5.
1.1
Name and description....................................................................................... 1-2
1.2
Sampling system configuration....................................................................... 1-3
1.3
Measurement Principle..................................................................................... 1-5
SPECIFICATIONS...................................................................................... 2-1
2.1
Standard Specification...................................................................................... 2-1
2.2
Model and Suffix Code...................................................................................... 2-7
2.3
Standard Accessories..................................................................................... 2-13
2.4
External dimensions........................................................................................ 2-14
2.5
NO2 / NO Converter........................................................................................... 2-19
INSTALLATION.......................................................................................... 3-1
3.1
Where to install equipment............................................................................... 3-1
3.2
How to install...................................................................................................... 3-2
3.3
Piping.................................................................................................................. 3-3
3.4
Sampling............................................................................................................. 3-5
3.5
Wiring ................................................................................................................. 3-7
3.6
Solenoid valve drive signal for calibration................................................... 3-16
3.6.1
Manual calibration............................................................................. 3-16
3.6.2
Auto calibration................................................................................. 3-17
OPERATION............................................................................................... 4-1
4.1
Warm-up operation and regular operation...................................................... 4-1
4.2
Keys and function on Panel . ........................................................................... 4-2
4.3
Mode and screens.............................................................................................. 4-3
4.4
General operation.............................................................................................. 4-6
SETUP (User Mode)................................................................................... 5-1
5.1
Switch Ranges................................................................................................... 5-2
5.2
Calibration Parameters..................................................................................... 5-3
5.2.1
Calibration Value................................................................................. 5-4
5.2.2
About ZERO Calibration..................................................................... 5-5
5.2.3
About Calibration Range..................................................................... 5-6
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
ii
5.2.4
6.
7.
8.
Auto Calibration Component / Range................................................. 5-7
5.3
Alarm Setting...................................................................................................... 5-8
5.4
Setting Auto calibration................................................................................... 5-11
5.5
Setting of Auto Zero Calibration..................................................................... 5-16
5.6
Peak Alarm Setting.......................................................................................... 5-20
5.7
Parameter Setting............................................................................................ 5-22
5.7.1
Output Hold....................................................................................... 5-23
5.7.2
Response time, Average Period, Backlight, Contrast...................... 5-26
5.7.3
Maintenance Mode........................................................................... 5-28
CALIBRATION............................................................................................ 6-1
6.1
Zero Calibration.................................................................................................. 6-1
6.2
Span Calibration................................................................................................ 6-2
MAINTENANCE.......................................................................................... 7-1
7.1
Daily check and maintenance procedures..................................................... 7-1
7.2
Replacement of power fuse.............................................................................. 7-2
7.3
Cleaning of sampling cell.................................................................................. 7-3
7.3.1
Disassembly and assembly of sampling cell (Pipe cell, Block cell)................................ 7-3
7.3.2
How to clean sampling cell................................................................. 7-6
ERROR MESSAGE ................................................................................... 8-1
Revision Information................................................................................................i
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
1.
1-1
<1. OVERVIEW>
OVERVIEW
This product measures the concentration of NO, SO2, CO2, CO and CH4 contained in sampling
gas on the principle that different atomic molecules have an absorption spectrum in the wave
band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law.
Since this instrument incorporates a compact O2 sensor, it allows measuring up to 4 components
simultaneously by using the built-in O2 sensor (up to 3 components if O2 measurement is
excluded).
Furthermore, use of a microprocessor and large sized liquid crystal display realizes improvement
of operability, accuracy and multi- functions.
This product is being purposely designed to be a part of large-scale fixed installations (LSFI.)
Intended uses of the product being part of LSFI are
• Combustion gas measuring / control of industrial furnaces / incinerators.
• Steel gas analysis (blast furnace, converter, thermal treatment furnace, sintering ((pellet
equipment)), coke furnace)
• Chemical analysis (petroleum refinery plant, petrochemical plant, gas generation plant)
• Environmental management and others
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
1.1
1-2
<1. OVERVIEW>
Name and description
(1)Power switch (2) Panel
(9)Flow checker
(1) Power switch (only for IR250-C)
(7)Sampling gas outlet
(12)Com.Connector (13) Analog input (5) Purge gas inlet
(RS-485)
connector(A/O)
(6)Sampling
gas inlet
(12) External input
connector
(10)Terminal block
1 to 10 from the left
(4)Power (5)Purge gas
supply
inlet
(3)Fuse
(2) panel
(11)External input
connector(A/I)
(14)Digital i/o connector
(from the bottom
DIO1,DIO2, DIO3)
(6)sampling gas inlet
(7)Sampling gas outlet
(in IR250-C, shared for
reference gas)
(8)Reference gas inlet
(only for IR250-C)
(3)Fuse
(4)Power supply
(4)Power terminal
M4 Terminal block type
(Power cable code: -NN)
(2)Panel
Mounting plate for
wall mounting
(Code”-C”)
(16)COMP2(second component)I/O terminal
(15)COMP1(1st Component) I/O terminal
(1)Power switch
(3)Fuse
(4)Power supply terminal
(7)Sampling gas outlet
(6)Sampling gas inlet
Wiring port
(Rubber grommets)
(5)Purge gas inlet
Name
Description
Name
Description
Checks flow rate of
sampling gas
(1) Power Switch
ON/OFF the analyzer
(9) Flow checker
(2) Display/Operation
panel
LCD, set functions
(10)Terminal block
(3) Fuse
(11) External input
connector
connecting to the output
of external O2 analyzer
(4) Power supply (switch)
connect to power supply
(12) communication
connector
RS-485 connector
(5) Purge gas inlet
connect to purge gs
(13) Analog output
connector (D-sub 25 pin)
connector for analog
output
(6) Sampling gas inlet
connect to measuring gas (14)Digital I/O
tube
Connector for digital I/O
(7) sampling gas outlet
connect to exhaust line
(15) COMP1 I/O
I/O for 1st component.
(8) Reference gas inlet
fuse inside
(16) COMP2 I/O
I/O for 2nd component
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
1-3
<1. OVERVIEW>
1.2
Sampling system configuration
The figure below shows a general configuration of sampling system. The system configuration
may vary depending on sample gas. Consult with Yokogawa
l Measurement of sample gas with high moisture content (2°C saturation level or NO,
SO2, CO (0-200l ppm) measurement (Excluding IR250-C)
(9)Flowmeter
(10) Secondary filter
(1)Primary filter (3)pump (12)3-way solenoid valve
sample
gas inlet
(11)NO2/NO
Converter
for NOx
(7)Solenoid
valve
(3)pump
atmosphere
(7)
* (4)Electric
dehumidifier
(2)Safety
drain trap
(7)
(7)
Gas Analyzer
(IR250)
(5)Drain
pot
O2 Analyzer (External)
(For zioconia)
(6) Ball valve
(8) Standard gas
for zero calibration
Drain
Exhaust(atmospheric
pressure)
Standard gas
for span calibration
* : A dehumidifier must be used for NO, SO2, and CO (0-200 ppm) measurements.(approx. equal to 2°C saturation)
F03.ai
l In IR250-C the measurement of sample gas with high moisture content (2°C
saturation level or above)
(10)Secondary
filter
*3
(13)Reference
gas purifier
Air filter (3)pump
Reference gas
(Humidified N2
or atmosphere)
(1)Primary filter (3)Pump
(9)Flowmeter
(11)NO2/NO
Gas Analyzer
Converter
(IR250-C)
for NOx measurement
(7)Solenoid
valve
Sample
gas
*1
(2)Safety drain
trap
(4)Electric
dehumidifier
Exhaust
(atomspheric
pressure)
*2
O2 Analyzer
(7)
(7)
(7)Solenoid valve
Exhaust
(atomspheric
pressure)
(5) Drain
pot
(6)Ball valve
Drain
(8)Standard gas
Standard gas
for zero calibration for span calibration
*1: Make the electric dehumidifier removes the moisture so that the temperature of sample gas goes below 5°C.
The water concentration should be equalized in reference gas and sample gas.
*2: Use No2/NO converter in case of measuring NOx.
*3: The concentration of measuring component in a reference gas determines whether a reference gas purifier
is required or not.
If the concentration level is lower than 0.1 % FS, a reference gas purifier is not required. If atmospheric air
is used as reference gas and the measuring range is lower level than the description below, the use of a
reference gas purifier is recommended. However, for measuring CO2, atmospheric air should not be used
as reference gas. NO analyzer : 0-100 ppm or less, SO2 analyzer : 0-50 ppm or less, CO analyzer : 0-1000 ppm
or less. Consult us on reference gas purifier suited to each component.
F03-1.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
1-4
<1. OVERVIEW>
Item
Description
Item
Description
(1) Primary filter
Removes dust and mist
(8) Standard gas for
calibration
Used for zero/span
calibration
(2) Safety drain trap
Separates and
dischages drain
(9) Flowmeter
Adjust and monitors
sample gas flow rate.
(3) Pump
Sucks in sample gas.
(10) Secondary filter
(membrane filter)
Removes fine dust.
(4) Electric dehumidifier Dehumidifies sample
gas
(11) NO2/NO converter
Converter NO2 gas into
NO gas
(5) Drain pot
Collects discharged
water from dehumidifier
(12) 3-way solenoid
valve
Used for introducing
humidified gas
(6) Ball valve
Used for discharging
drain
(13)Reference gas
purifier
Eliminates particles out
of the reference gas
(7) Solenoid Valve
Used for switching
sampling and
calibration lines
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
1.3
1-5
<1. OVERVIEW>
Measurement Principle
l Infrared gas analyzer (CO2, CO, CH4, SO2, NO)
Front expansion chamber
Gas inlet
Infrared-ray
light source
Gas outlet
Rear expansion chamber
Detector
Motor
Mass flow sensor
Measuring cell
Chopper
Preamplifier
Signal
processing
and calculation
block
Indication
Output
F01.ai
l Consecutive zero correction with IR250-C
Sample and reference gas are measured at each respective timing. Measured gas concentration
is output by converting the difference value between the sample and reference gas into a
concentration value. Continuous monitoring on the difference value between sample and
reference gas leads to an elimination of influence from interference component and zero-drift
Gas Analyzer
Sample gas
Reference gas
SV1
com
NC
Solenoid
valve
com
NO
Infrared
light source
sample
cell
Exhaust gas
Detector
motor
NC
SV2
chopper
NO
NC: Normal close NO: Normal open
Gas flow
l Paramagnetic oxygen analyzer (O2)
Permanent
magnets
Mirror
Measuring cell
Gas inlet
Magnetic field
Gas outlet
Preamplifier
LED
Permanent
magnets
Photodiode
Indication
Output
Signal processing
and calculation block
F02.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Blank Page
2-1
<2.SPECIFICATIONS>
2.
SPECIFICATIONS
2.1
Standard Specification
Measurement principle:
CO2, CO, CH4 SO2 NO;
Non-dispersive infrared method
Single light source-single beam
O2;
Paramagnetic type (built-in), or zirconia type (external)
Measurable gas components and measuring ranges:
Component
NO
SO2
CO2
CO
CH4
O2 (Built-in
paramagnetic)
O2 (Dedicated Zirconia)
•
•
•
•
IR250-A
Min.range Max.range
0-500 ppm 0-5000 ppm
0-500 ppm 0-5000 ppm
0-500 ppm 0-100 vol%
0-200 ppm 0-100 vol%
0-1000 ppm 0-100 vol%
0-5 vol%
0-100 vol%
IR250-B
Min.range Max.range
―
―
0-500 ppm 0-5000 ppm
0-500 ppm 0-100 vol%
0-500 ppm 0-100 vol%
0-1000 ppm 0-100 vol%
―
―
IR250-C
Min.range Max.range
0-50 ppm
0-5000 ppm
0-50 ppm
0-5000 ppm
0-50 ppm
0-5000 ppm
0-50 ppm
0-5000 ppm
―
―
0-5 vol%
0-100 vol%
IR250-D
Min.range Max.range
0-200 ppm 0-5000 ppm
0-200 ppm 0-10 vol%
0-100 ppm 0-100 vol%
0-200 ppm 0-100 vol%
0-500 ppm 0-100 vol%
0-5 vol%
0-100 vol%
0-5 vol%
―
0-5 vol%
0-5 vol%
0-25 vol%
―
0-25 vol%
0-25 vol%
IR250-A, IR250-C, IR250-D are capable of measuring up to 4 components including O2.
1 or 2 measuring ranges per component.
Measuring range ratio:
IR250-A,
≤1:10 (except built-in paramagnetic O2 analyzer)
≤1:20 (built-in paramagnetic O2 analyzer)
IR250-B,
≤1:10
IR250-C, IR250-D
≤1:10 (except built-in paramagnetic O2 analyzer)
≤1:20 (built-in paramagnetic O2 analyzer)
For measurable components and possible combinations of measuring ranges, see pages 2-18
Display:
Digital, 4-digit LCD with LED backlight
• Instantaneous value of each component
• Instantaneous value after O2 correction (only in CO, SO2, NO with O2 measurement)
• Average value after O2 correction (only in CO, SO2, NO with O2 measurement)
• Average O2 value (Only with O2 measurement)
Analog output signal:
4 to 20 mA DC or 0 to 1 V DC, (isolated from ground wire or internal circuit, nonisolated between output cables) .
Analog output corresponds one-to-one with measured value indication.
Permissible load resistance;
550 Ω max. for 4 to 20 mA DC
100 kΩ min. for 0 to 1 V DC
Max points of
IR250-A; 7 points
IR250-B; 4 points
IR250-C, IR250-D;
11 points
See page 3-15 “Measurable components and corresponding channle numbers”.
Analog input signal (IR250-A, IR250-C. IR250-D):
For signal input from external O2 sensor*
Input section is not isolated.
Signal requirement:
Signal from Yokogawa’s zirconia type O2 sensor (Model ZX8D)
0 to 1 V DC of full scale
* External O2 sensor should be purchased separately.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-2
<2SPECIFICATIONS>
Contact output:
Model
IR250-A
IR250-C
IR250-D
Contacts
1a relay contact
1c relay contact
1c relay contact
Contact Capacity
250 V AC/2 A
resistive load
24 V DC/1 A resistive load
Max. number of output
points
1a relay 13 points
1c relay 5 points
15 points
Isolation of the relay
contacts and the internal
circuit
All relay contacts are isolated mutually and from the internal
circuit.
Output signal
Instrument error, Calibration error, Range identification, Auto
calibration status, Solenoid valve drive signal for automatic
calibration, Pump ON/OFF*1, High / Low limit alarm*1, Peak
alarm*2
* 1 :
* 2 :
High / Low limit alarm on 4 component (including O2) analyzer of IR250-C, IR250-D have 3 points of output signal.
applies only to IR250-A.
Contact input:
Model
IR250-A
IR250-D
Non-voltage input*3
Voltage input contact
Applied voltage
―
12 to 24 V DC
Required max.voltage
―
15 mA
Max of input points
7 points
9 points
Isolation of the relay contacts Mutual contacts ; non isolation
and the internal circuit
Internal circuit: photocoupler isolation
Input
*3:
*4:
IR250-C
Contacts
Isolation with photocoupler
Remote range changeover, auto calibration remote start, remote (output)
hold, average value reset*4
When it is ON: 0 V DC, OFF: 5 V DC, ON: 5 mA.
applies only to IR250-C, IR250-D
Note: See 3-10 I/O Wiring Diagram for further information.
Consult YOKOGAWA if you need contact input / output for IR250-B.
Power supply:
Voltage rating; 100 to 240 V AC
Allowable range;
85 to 264 V AC
Frequency; 50/60 Hz
Maximum voltage rating;
100 VA
Inlet; Conform to EN60320 Protection Class l (When IR250-A,
IR250-C, IR250-D use other than “-NN” as power cable)
Operating conditions:
Ambient temperature;
Ambient humidity;
Storage conditions:
Ambient temperature;
Ambient humidity;
-5 to 45°C (when Measurable component code -C, -G,
-H, -P is specified and you run the analyzer
with electric power of larger than 200 V, the
temperature is 40 °C max)
95 % RH max., non-condensing
-20 to 60°C
90%RH max., non-condensing
Dimensions (H × W × D):
IR250-A;
177 × 483 × 480 mm (with slide rail for 19-inch rack mounting
type)
IR250-B; 484 × 294 × 170 mm
IR250-C,IR250-D;
133× 483 × 382 mm
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Weight:
2-3
<2.SPECIFICATIONS>
Approx. 11 kg (IR250-A, IR250-C, IR250-D, ), Approx. 7 kg (IR250-B)
Finish color:
Front panel;
Casing;
Enclosure:
Off-white (Munsell 10Y7.5/ 0.5 equivalent)
steel-blue
Steel casing, for indoor use
Material of gas-contacting parts:
Gas inlet/outlet;
SUS304
Sample cell;
SUS304 / polychloroprene
Infrared-ray transmitting window;
CaF2
Internal tubing;
polyvinyl chloride, Fluoropolymer (PTFE), polypropylene
Gas inlet/outlet:
Rc1/4 or 1/4 NPT internal thread
Purge gas flow rate:
approx. 1 L/min (when necessary)
Safety and EMC conforming standards:
Only IR250-A, IR250-C or IR250-D as CE marking compliance product
Select the suffix code of Power cable : -NN or -EU.
Installation altitude; 2000 m or less
Pollution degree;
2 (Note)
Installation category; II (Note)
●
Installation category, so called overvoltage category, specifies impulse
withstanding voltage. Category II is for electrical equipment.
●
ollution degree indicates the degree of existence of solid, liquid, gas or other
P
inclusions which may reduce dielectric strength. Degree 2 indicates the normal
indoor environment.
; EN61010-1
; EN61326-1 Class A, Table 2 (for Industrial application) EN61326-2-3, EN610003-2, EN61000-3-3
EMC (RCM) of Australia, New Zealand, EN61326-1 Class A
Safety *
EMC *
Note: This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the
industrial environment only.
n Standard Functions
Output signal hold:
Output signals is held during manual and auto calibrations by activation of holding
(turning on its setting). Indication of instantaneous values are not held. The values to be
held are the last ones indicated just before calibration mode starts or selected setting
value. Indicated values are not held.
Remote output hold: (not standard function in IR250-B)
Output signal is held at any preset value or at the last value just before contact input
signal is received while voltage is input to the terminals.
Holding is maintained while the voltage is input to the terminals. Indication of
instantaneous values are not held.
Range changeover:
Set a changeover among Auto, Manual, Remote. Only selected mode is effective during
the operation.
Manual ;
Key operation switches ranges
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-4
<2SPECIFICATIONS>
Auto ; When a measured value level is higher than 90 % of full scale
concentration for the first range, the range switches from the first
range to the second one.
When a measured value level is lower than 80 % of full scale
concentration for the first range, the range switches from the
second range to the first one.
Remote ; (not standard function in IR250-B)
When the contact input is active, a remote range switch is
available. Remote range switch is active only when the first
range applies, and when it is not, the second range applies
Range identification signal:
(not standard function in IR250-B)
The current measuring range is identified by a contact position.
IR250-A,
hen the first range is selected, the contact output terminals are
W
closed, and when the second range is selected, they are open
for each component.
IR250-C, IR250-D,
hen the first range is selected, the NO side is closed, and
W
when the second range is selected, NC side is closed for each
component.
Auto calibration: (not standard function in IR250-B)
Auto calibration is carried out periodically at the preset cycle.
When a standard gas cylinder for calibration and a solenoid valve for opening/ closing the
gas flow line are prepared externally by the customer, calibration will be carried out with
the solenoid valve drive contacts for zero calibration and each span calibration turned on/
off sequentially at the set auto calibration timing.
Auto calibration cycle setting;
Auto calibration cycle is set.
Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in
increments of 1 day).
Gas flow time setting;
The time for flowing each calibration gas in auto calibration is set.
Settable within 60 to 599 seconds (in increments of 1 second)
Auto calibration remote start: (not standard function in IR250-B)
Auto calibration is carried out only once according to an external input signal. Calibration
sequence is settable in the same way as the cyclic auto calibration.
Calibration starts by opening the auto calibration remote start input terminals after shortcircuiting for 1.5 seconds or longer. Auto calibration is started when the contacts open.
Auto zero calibration: (not standard function in IR250-B)
Auto zero calibration is carried out periodically at the preset cycle.
This cycle is independent of “Auto calibration” cycle.
When zero calibration gas and solenoid valve for opening/closing the calibration gas
flow line are prepared externally by the customer, zero calibration will be carried out with
the solenoid valve drive contact for zero calibration turned on/off at the set auto zero
calibration timing.
Auto zero calibration cycle setting;
Auto zero calibration cycle is set.
Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in
increments of 1 day).
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-5
<2.SPECIFICATIONS>
Gas flow time setting:
The time for flowing zero gas in auto zero calibration is set. Settable 60 to 900 seconds (in
increments of 1 second)
High/low limit alarm: (not standard function in IR250-B)
Alarm contact output turns on when measurement value reaches the preset high or low
limit alarm value
Contacts close when the instantaneous value of each channel exceeds the high alarm
limit value or falls below the low alarm limit value.
Instrument error contact output:
Contact signals are output at occurrence of analyzer error.
Contacts close at occurrence of analyzer error No. 1, 3 or 10.
Calibration error contact output: (not standard function in IR250-B)
Contacts close at occurrence of manual or auto calibration error (any of errors No.4 to 9).
Auto calibration status contact output: (not standard function in IR250-B)
Contacts close during auto calibration.
Pump ON/OFF contact output: (available only in IR250-A)
During measurement, contacts close.
While calibration gas is flowing, contacts open. Contacts are connected in power supply
of pump, and stop the sample gas while calibration gas is flowing.
n Optional Functions (Specify with suffix code.)
O2 correction (when /K is selected):
Conversion of measured CO, SO2 and NO gas concentrations into values at reference
O2 concentration.
Correction formula: C =
21 – On
x Cs
21 – Os
Where:
C: ample gas concentration after O2 correction
Cs: Measured concentration of sample gas
Os: Measured O2 concentration (limit set 1 to 20 % O2, initial value : 20 % )
On: R
eference O2 concentration (value changeable by setting 0 to 19 %
O2, initial value : 12 % )
The upper limit value of the fractional part in this calculation is 4.
The result of calculation is indicated and output in an analog output signal.
Average value after O2 correction (when /K is selected):
The result of O2 correction or instantaneous O2 value can be output as an average value
in the determined period of time.
Moving average method is used. Sampling is carried out at intervals of 30 seconds.
(Output is updated every 30 seconds. It is updated to the averaged value of the most
recent elapsed averaging time period
Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4
hours (in increments of 1 hour).
Average value reset (when IR250-C specifies /K or IR250-D specifies /K):
The above-mentioned output of average value is started from the initial state by opening
the average value resetting input terminals after short-circuiting for 1.5 seconds or longer.
Output is reset by short-circuiting and restarted by opening.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-6
<2SPECIFICATIONS>
CO concentration peak alarm (when /A is selected):
(available only for CO - O2 measurement with IR250-A)
Alarm output turns on according to the preset concentration and count.
Whenever the instantaneous value of CO exceeds the preset concentration value, count
increments. If the count exceeds the preset value in one hour, the alarm contacts close.
Communication function (when /C is selected):
RS-485C (9 pins D-sub)
Half-duplex bit serial
Start-stop synchronization MODBUS protocol
Contents; Read/write parameters, measurement concentration and instrument
status
n Performance
Model
Repeatability
Linearity
Zero drift
Spam drift
Response time
(90% FS response)
IR250-A
IR250-B
±0.5% FS (less than ±0.5% FS
50 ppm: 1% FS)
±1% FS
±2% FS / week
±2% FS / week
IR250-C
±0.5% FS
IR250-D
±0.5% FS
±0.5% FS/ week
±2% FS / week(O2
measurement)
±2% FS / week
(When
measurement NO,
SO2 of less than
500 ppm uses auto
zero.calibration )
±2% FS / week
within10 to 30 sec.(including sample replacement)
Influence from other gases:
If the following gases are contained in sample gas, a
measurement result may be interfered. For detailed solution,
consult our representatives.
IR250-A, IR250-C, IR250-D;
Interference
component
CO 1000 ppm
CO2 15%
NO analyzer
≤1.0%FS
≤1.0%FS
(the range of
500ppm: ≤2%FS)
SO2 analyzer CO2 analyzer
≤1.0%FS
≤1.0%FS
≤1.0%FS
―
H2O 20°C saturation ―
―
≤1.0%FS
H2O 2°C saturation
CH4 1000 ppm
≤2%FS
≤20ppm
―
≤1.0%FS
≤2%FS
―
CO analyzer
―
≤1.0%FS
(the range
of 200ppm:
≤2%FS)
≤1.0%FS
(the range
of 500ppm:
≤2%FS)
≤2%FS
≤1.0%FS
CH4 analyzer
≤1.0%FS
≤1.0%FS
≤1.0%FS
―
―
Note: Range of 0-200 ppm is excluded.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-7
<2.SPECIFICATIONS>
IR250-B;
Interference
component
NO analyzer
SO2 analyzer
CO2 analyzer
CO analyzer
CH4 analyzer
CO 1000ppm
CO2 15%
―
―
―
≤1.0%FS
≤1.0%FS
―
≤1.0%FS
―
≤1.0%FS
―
≤1.0%FS
≤1.0%FS
≤1.0%FS
≤1.0%FS
≤1.0%FS
H2O 2°C saturation
CH4 1000ppm
―
≤2%FS
―
≤2%FS
―
―
≤20ppm
≤1.0%FS
≤1.0%FS
―
H2O 20°C saturation
2.2
Model and Suffix Code
[Style : S1]
Model
IR202
Suffix code
---------------------------------------
Specification
- - - - - - - - - - - - - - Infrared Gas Analyzer
-----------------------------------------------------
19 inch rack mounting type*1
Wall mounting type*2
Low-concentration 19inch rack mounting type*3
19 inch rack mounting D-Sub terminal type
-------------------------------------------------------------------------------------------------------------------------------------------------------------
NO
SO2
NO+SO2
CO
CO2
CO2+CO
NO+CO
NO+SO2+CO
CH4
CH4+CO
CO2+CH4
CO2+CO+CH4
-----------------------------------------------------
Without O2 Analyzer
External Zirconia type O2 Sensor (ZX8D) *5
External O2 Analyzer*6
Built-in paramagnetic type O2 Sensor
Measuring range -NN
for NO *7,*8
-01
-02
-03
-04
-07
-------------------------------------------------------------------------------
Not available
Min. 0-50 ppm / Max. 0-500 ppm
Min. 0-100 ppm / Max. 0-1000 ppm
Min. 0-200 ppm / Max. 0-2000 ppm
Min. 0-250 ppm / Max. 0-2500 ppm
Min. 0-500 ppm / Max. 0-5000 ppm
Measuring range for -NN
-01
SO2*7,*8
-02
-03
-04
-06
-07
-08
-09
-10
-11
-12
-------------------------------------------------------------------------------------------------------------------------------------------------------------
Not available
Min. 0-50 ppm /Max. 0-500 ppm
Min. 0-100 ppm /Max. 0-1000 ppm
Min. 0-200 ppm /Max. 0-2000 ppm
Min. 0-250 ppm /Max. 0-2500 ppm
Min. 0-500 ppm /Max. 0-2500 ppm
Min. 0-500 ppm /Max. 0-5000 ppm
Min. 0-1000 ppm /Max. 0-5000 ppm
Min. 0-1000 ppm/ Max. 0-1 %
Min. 0-2000 ppm/ Max. 0-2 %
Min. 0-5000 ppm/ Max. 0-5 %
Min. 0-1 % / Max. 0-10 %
Measuring range for CO2 -NN
*7,*8
-01
-02
-03
-04
-07
-09
-10
-11
-12
-13
-14
-15
-16
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Not available
Min. 0-50 ppm / Max. 0--500 ppm
Min. 0-100 ppm / Max. 0--1000 ppm
Min. 0-200 ppm / Max. 0--2000 ppm
Min. 0-250 ppm / Max. 0-2500 ppm
Min.0-500 ppm / Max. 0-5000 ppm
Min.0-1000 ppm/ Max. 0-1 %
Min.0-2000 ppm/ Max. 0-2 %
Min.0-5000 ppm/ Max. 0-5 %
Min.0-1%/ Max. 0-10%
Min.0-2%/ Max. 0-20%
Min.0-3%/ Max. 0-25%
Min.0-5%/ Max. 0-50%
Min.0-10%/ Max. 0-100%
Type
-A
-B
-C
-D
Optional suffix
code
Measurable -A
Component -B
*4
-C
-D
-E
-F
-G
-H
-K
-L
-M
-P
O2 Analyzer
N
1
2
3
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-8
<2SPECIFICATIONS>
[Style : S1]
Optional suffix
code
Model
Suffix code
IR202
---------------------------------------
Measuring range for CO
*7,*8
Measuring range for CH4
*7,*8
Measuring range for O2 *7,
Analog Output *9
-NN
-01
-02
-03
-04
-07
-09
-10
-11
-12
-13
-14
-15
-16
-NN
-07
-09
-10
-11
-12
-13
-14
-15
-16
N
1
2
3
4
5
6
7
8
-1
-4
Power Cable *10
Mounting Type *11
-NN
-JA
-US
-EU
-CN
-A
-B
-C
-D
-----------------------------------------------------------------------------------------------------------------------------------
Not available
Min. 0-500 ppm / Max. 0-5000 ppm
Min. 0-1000 ppm / Max.0-1%
Min. 0-2000 ppm / Max.0-2%
Min. 0-5000 ppm / Max.0-5%
Min. 0-1% / Max. 0-10%
Min. 0-2% / Max. 0-20%
Min. 0-3% / Max. 0-25%
Min. 0-5% / Max. 0-50%
Min. 0-10% / Max. 0-100%
---------------------------------------------------------------------------------------------------------------------
Not available
0-5/10 vol%
0-5/25 vol%
0-10/25 vol%
0-5 vol%
0-10 vol%
0-25 vol%
0-50 vol%
0-100 vol%
-----------------------------------------------------
None (Terminal block for power supply)
Rated voltage 125V AC(PSE)
Rated voltage 125V AC(UL)
Rated voltage 250V AC(CEE)
Rated voltage 250V AC(CCC)
-----------------------------------------------------
Rack mounting with Slide rail
Panel mounting
Wall mounting
Rack mounting
- - - - - - - - - - - - - Always "-N”
N
―
Not available
Min. 0-50 ppm/ Max. 0-500 ppm
Min. 0-100 ppm/ Max. 0-1000 ppm
Min. 0-200 ppm/ Max. 0-2000 ppm
Min. 0-250 ppm/ Max. 0-2500 ppm
Min. 0-500 ppm/ Max. 0-5000 ppm
Min. 0-1000 ppm/ Max.0-1%
Min. 0-2000 ppm/ Max. 0-2%
Min. 0-5000 ppm/ Max. 0-5%
Min. 0-1% / Max. 0-10%
Min. 0-2%/ Max. 0-20%
Min. 0-3%/ Max. 0-25%
Min. 0-5%/ Max. 0-50%
Min. 0-10%/ Max. 0-100%
- - - - - - - - - - - - - - Japanese
- - - - - - - - - - - - - - English
- - - - - - - - - - - - - - Chinese
-N
―
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- - - - - - - - - - - - - - Rc1/4
- - - - - - - - - - - - - - 1/4NPT
J
E
C
Language
- - - - - - - - - - - - - - Infrared Gas Analyzer
- - - - - - - - - - - - - - 0-1V DC, non-isolation
- - - - - - - - - - - - - - 4-20mA DC, non-isolation
R
T
Piping
- - - - - - - - - - - - - Always "N"
/K
/A
/C
/U
Option
Specification
O2 correction output*12
Peak count alarm output*13
RS-485 communication*14
Conversion of unit (mg/m3, g/m3)*15
Note:
*1: Same size as IR200 (former model) in size of an installation dimension.
*2: For IR202-B, measuring component including NO cannot be selected. Measuring of 3 component cannot be selected. O2
measurement is not available.
*3: For IR202-C, components including CH4 cannot be selected.
*4 : For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately.
*5 : Purchase separately ZX8D, Zirconia O2 analyzer. Prepare ZX8D StyleD as CE compliance product.
*6: Purchase a O2 analyzer separately. A signal from the external O2 analyzer has to be set as 0-1 V DC linear corresponding
*7:
*8
*9
to full scale. .
You can choose any range within the measurable range.
Refer to Measurable component list
For IR202-B, both of these codes are available to choose to make output.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<2.SPECIFICATIONS>
2-9
*10
Select appropriate power cables based on the required voltage and specification. The abbreviation represents each country
or region that regulate the standard and certify each product. PSE-Japan, UL-US, CEE-Europe, CCC-China. When you select
IR202-B, the power cable is always -NN. “-A “ cannot be specified with “-NN” .
*11: Mounting of IR202-A is always -A , IR202-B: -B or C. IR202-C and IR 250-D: always -D.
*12: Not selectable for IR202-B. Not selectable for the unit without O2 analyzer. O2 correction output applies only to NO, CO, and
SO2. As an option, O2 correction average value is available. NO is indicated as NOx.
*13: Selectable only for IR202-A. Peak count alarm can be provided only for CO measurement.
*14: Should be specified when using MODBUS™ communication.
*15 : Not selectable for IR202-B. Only the range for NO, SO2 or CO are convertable.Select a range by ppm. Preset value unit of
ranges before shipping is converted one ; mg/m3 range (for 10000mg/m3 or above, the unit is g/m3) See the following conversion
table as reference.
ppm range
range mg/m3
NO
SO2
CO
0-50 ppm
0-65.0 mg/m3
0-140 mg/m3
0-60.0 mg/m3
0-100 ppm
0-130 mg/m3
0-280 mg/m3
0-125 mg/m3
0-200 ppm
0-260 mg/m3
0-570 mg/m3
0-250 mg/m3
0-250 ppm
3
0-325 mg/m
3
0-700 mg/m
0-300 mg/m3
0-300 ppm
0-400 mg/m3
0-850 mg/m3
0-375 mg/m3
0-500 ppm
0-650 mg/m3
0-1400 mg/m3
0-600 mg/m3
0-1000 ppm
0-1300 mg/m3
0-2800 mg/m3
0-1250 mg/m3
0-2000 ppm
0-2600 mg/m3
0-5600 mg/m3
0-2500 mg/m3
0-2500 ppm
0-3300 mg/m3
0-7100 mg/m3
0-3000 mg/m3
0-5000 ppm
0-6600 mg/m3
0-14.00 g/m3
0-6250 mg/m3
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<2SPECIFICATIONS>
2-10
n Measurable Components and Ranges
A : IR250-A, B : IR250-B, C: IR250-C, D : IR250-D
(1) Single-component analyzer (NO, SO2, CO2, CO, CH4)
Range
Code
Component
Minimum/Maximum
NO *1
SO2
CO2
CO
CH4 *2
-01
0-50ppm/0-500ppm
C
C
C
C
―
-02
0-100ppm/0-1000ppm
C
C
CD
C
―
-03
0-200ppm/0-2000ppm
CD
CD
CD
ACD
―
-04
0-250ppm/0-2500ppm
CD
CD
CD
CD
―
-05
0-300ppm/0-2500ppm
―
―
―
―
―
-06
0-500ppm/0-2500ppm
―
―
―
―
―
-07
0-500ppm/0-5000ppm
ACD
ABCD
ABCD
ABCD
D
-08
0-1000ppm/0-5000ppm
―
―
―
―
―
-09
0-1000ppm/0-1%
―
D
ABD
ABD
ABD
-10
0-2000ppm/0-2%
―
D
ABD
ABD
ABD
-11
0-5000ppm/0-5%
―
D
ABD
ABD
ABD
-12
0-1%/0-10%
―
D
ABD
ABD
ABD
-13
0-2%/0-20%
―
―
ABD
ABD
ABD
-14
0-3%/0-25%
―
―
ABD
ABD
ABD
-15
0-5%/0-50%
―
―
ABD
ABD
ABD
-16
0-10%/0-100%
―
―
ABD
ABD
ABD
*1 :
*2 :
IR250-B cannot measure NO.
IR250-C cannot measure CH4.
(2) Two-component analyzer
Refer to the table (1) regarding measuring range (Minimum / Maximum.)
a. NO+SO2 (1st component is NO, 2nd component is SO2)
IR250-B is not applicable.
2nd comp.
range
1st comp.range
NO
-01
-02
-03
-04
-07
SO2
-01
-02
-03
-04
-07
C
C
C
C
C
C
C
C
C
C
C
C
CD
CD
CD
C
C
CD
CD
CD
C
C
CD
CD
ACD
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-11
<2.SPECIFICATIONS>
b. CO2+CO (1st component is CO2, 2nd is CO)
Blank represents “not available.”
CO
2nd comp.
range
1st comp.range
CO2
-01
-02
-03
-04
-07
-09
-10
-11
-12
-13
-14
-15
-16
-01
-02
-03
-04
C
C
C
C
C
CD
D
D
CD
D
D
-07
-09
C
C
CD
CD
BCD
BD
BD
-10
-11
BD BD
BD BD
BD ABD ABD ABD
ABD ABD
ABD
ABD ABD
ABD
ABD ABD
ABD
BD ABD
ABD
ABD
ABD
-12
BD
BD
ABD
ABD
ABD
ABD
ABD
ABD
-13
-14
ABD
ABD
ABD
ABD
ABD
ABD
BD
ABD
ABD
ABD
ABD
ABD
ABD
-15
BD
BD
ABD
ABD
ABD
ABD
ABD
-16
ABD
ABD
ABD
ABD
ABD
c. NO+CO(1st component is NO, 2nd is CO)
IR250-B is not applicable.
2nd comp
range
1st comp. range
NO
-01
-02
-03
-04
-07
-01
C
C
C
C
C
-02
C
C
C
C
C
CO
-03
C
C
CD
CD
ACD
-04
C
C
CD
CD
CD
-07
C
C
CD
CD
ACD
d. CH4+CO(1st component is CH4, 2nd is CO)
IR250-C is not applicable. Blank represents "not available."
2nd.comp.
range
1st comp. range
CH4
-07
-09
-10
-11
-12
-13
-14
-15
-16
CO
-03
-04
D
D
-07
D
D
AD
-09
D
AD
AD
AD
AD
AD
-10
-11
D
AD
AD
AD
AD
AD
-12
AD
AD
AD
AD
-14
D
AD
AD
-15
AD
AD
-16
AD
AD
AD
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<2SPECIFICATIONS>
2-12
e. CH4+CO(1st component is CH4, 2nd is CO)
IR250-C is not applicable. Blank represents “not available.”
2nd comp..
range
1st comp.range
CO2
-02
-03
-04
-07
-09
-10
-11
-12
-13
-14
-15
-16
CH4
-07
-09
D
D
D
D
D
D
D
D
D
D
D
AD
AD
AD
AD
-10
D
D
D
D
AD
AD
AD
AD
AD
-11
-12
AD
AD
AD
AD
AD
AD
D
-13
-14
-15
D
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
-16
AD
AD
AD
(3) Three-component analyzer (NO and CO)
IR250-B is not designed for three component analyzer.
a. NO+SO2+CO (1st component is NO, 2nd is SO2, 3rd is CO)
IR250-B is not applicable. Blank represents "not available."
Code
Min / Max range
1st component NO
-01
0-50ppm/0-500ppm
C
-02
0-100ppm/0-1000ppm
C
-03
0-200ppm/0-2000ppm
CD
-04
0-250ppm/0-2500ppm
CD
-05
0-300ppm/0-2500ppm
―
-06
0-500ppm/0-2500ppm
―
-07
0-500ppm/0-5000ppm
ACD
-08
0-1000ppm/0-5000ppm
―
-09
0-1000ppm/0-1%
―
-10
0-2000ppm/0-2%
―
-11
0-5000ppm/0-5%
―
-12
0-1%/0-10%
―
-13
0-2%/0-20%
―
-14
0-3%/0-25%
―
-15
0-5%/0-50%
―
-16
0-10%/0-100%
―
3rd comp.
range
2nd comp. range
SO2
-01
-02
-03
-04
-06
-07
-08
CO
-01
-02
-03
-04
C
C
C
C
C
C
C
C
C
C
CD
CD
AD
C
CD
CD
AD
C
-07
AD
C
AD
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-13
<2.SPECIFICATIONS>
b. CO2+CO+CH4 (1st component is NO, 2nd is SO2, 3rd is CH4)
IR250-B, IR250-C are not applicable. Blank represents "not available."
2nd comp.
range
1st comp. range
-11
-12
-13
-14
-15
-16
CO2
CO
-07
-09
-10
-11
-12
-13
-14
-15
D
AD
AD
AD
D
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
AD
Code
Min./Max range
3rd component CH4
-01
0-50ppm/0-500ppm
―
-02
0-100ppm/0-1000ppm
―
-03
0-200ppm/0-2000ppm
―
-04
0-250ppm/0-2500ppm
―
-05
0-300ppm/0-2500ppm
―
-06
0-500ppm/0-2500ppm
―
-07
0-500ppm/0-5000ppm
―
-08
0-1000ppm/0-5000ppm
―
-09
0-1000ppm/0-1%
―
-10
0-2000ppm/0-2%%
―
-11
0-5000ppm/0-5
AD *1
AD
-12
0-1%/0-10%
-13
0-2%/0-20%
AD
-14
0-3%/0-25%
AD *2
-15
0-5%/0-50%
―
-16
0-10%/0-100%
―
*1:
*2:
AD
AD
AD
AD
-16
AD
AD
AD
AD
AD
CO range -16 is not applicable.
CO range -07 is not applicable.
2.3
Standard Accessories
Name
Qty
Description
Code Specification
Fuse
2
250 V/2A delay type (K9642CJ)
Input/output terminal
module
1
Connector for external O2 analyzer
O2 analyzer "1", "2"
Type, “-C”, “-D”
Power cable
1
(varies with type of power cables)
come with a cable stopper
power cable "-JA", "-US", "-EU",
"-CN"
Ferrite Cores for Power
Cable
1
power cable "-NN"
Ferrite Cores for digital
I/O
3
Type; ”A","-B"
Analog output, digital I/O
connector
4
D-sub connector 25 pins with 8 male
screws
Type, "-C", "-D"
RS-485 Connector
1
D-sub connector 9 pins with 2 male
screws
Option "-C"
Mounting plate
4
Mounting type "-B"
Slide rail
2
Mounting type, Factory-installed
delivery.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2.4
2-14
<2SPECIFICATIONS>
External dimensions
l IR250-A
Power switch
Unit: mm
M4 (for slide rail)
7
101.6
177
463
483
169
24
480 or less
429
Terminal
Terminal Terminal
block 3
block 5
Commun. block 1
Connector
Terminal
Terminal
block 2
block 4
Fuse
Sampling gas outlet
(Rc1/4 or 1/4NPT)
Sampling gas inlet
(Rc1/4 or 1/4NPT)
Purge gas inlet
(Rc1/4 or 1/4NPT)
Power supply
4-M6
101.6
450 or more
465
Rack mounting dimensions
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-15
<2.SPECIFICATIONS>
l IR250-B
Unit: mm
70
Mounting hardware
for panel mounting *1
Mounting plate for
Wall mounting (code: -C)
4-M6 or Φ8
440 454 484
454
70
175
70
294
Mounting dimensions
for Wall mounting (code: -C)
281 +2
0
Wiring port
(Rubber grommets)
Purge gas inlet (Rc1/4 or 1/4NPT)
Sampling gas inlet (Rc1/4 or 1/4NPT)
Sampling gas outlet((Rc1/4 or 1/4NPT)
430 +2
0
150
350
*1: 4 mounting brackets come in package for
Panel cut-out dimensions and spacing
Panel mounting (code: -B)
for Panel mounting (code: -B)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-16
<2SPECIFICATIONS>
l IR250-C or IR250-D
Unit: mm
Power switch
(M4 Terminal Block Type)
(Power Cable code: -NN)
Flow checker
(Only for IR250-C)
25
7
57.2
463
483
Analog output connector (A/O)
Communication
connector (RS-485)
429
External input connector
(A/I)
124.5
132.5
24
16
380
Purge gas inlet (Rc1/4 or 1/4NPT)
Sampling gas inlet (Rc1/4 or 1/4NPT)
Sampling (Reference) gas outlet
(Rc1/4 or 1/4NPT)
Reference gas inlet (only for IR250-C)
(Rc1/4 or 1/4NPT)
Fuse
4-M6
Power
supply
57.2
Digital input/output connector (M4 Terminal Block Type)
(Power Cable code: -NN)
(DIO1 to 3 from the left)
450 or more
465
Rack mounting dimensions
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-17
<2.SPECIFICATIONS>
Accessories
l Slide rail (Model : 305A-20/Acuride International Inc.)
Only when Mounting type -A is selected, the slide rail is available. The rail is pre-installed at
factory before delivery.
Unit: mm
Closed
530±3
Cabinet member
1.52t
Closed position flush or under
(22.7)
Intermediate
member
(9.5)
Drawer member
1.27t
15.9±0.5
Open
508.0±0.8
476.2±0.3
463.5±0.3
352.8±0.3
111.1±0.3
(24.4)
35.3±0.3
25.4±0.3
12.7±0.3
12.7±0.3
4.5 × 5.3
123.8±0.3
4.5 × 5.3
238.1±0.3
15.9±0.5
19.1±0.3
365.1±0.3
505.0±0.8
F05.EPS
Reinforcement plate
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2-18
<2SPECIFICATIONS>
The instrument weight should be supported at the bottom of the unit (or the side of the
unit when mounted with the slide rails). For easy maintenance, it is recommended to
have an installation that allows the instrument to be withdrawn along the slide rail.
Slide rail mounted type
Guide rail mounted type *
Rack size
Unit: mm
101.6
or
100
101.6
or
100
Rack size
450 or more
465
Mounting diagram
Slide rails
450 or more
465
EIA standard
EIA standard
Mounting diagram
Guide rails
* For the guide rail mounted type, a maintenance space (200 mm or more)
should be provided on top of the main unit.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
2.5
2-19
<2.SPECIFICATIONS>
NO2 / NO Converter
Use for NOx measurement.
Part number: K9350LE (Non-CE conformity)
K9350LF (CE conformity)
Mounting:
Indoor surface mounting
Target Gases: General boiler exhaust gas, atmosphere
Catalyst: Amount;
Replacement cycle;
2 cm3
Approx. 8 months (at flow rate of 0.5 L/min with 5% O2,10
ppm NO2)
Temperature set-point; 220 ±10°C (Sensing tip; K thermocouple)
Wetted materials:
ceramic, Fluoropolymer (PTFE), glass filter, SUS316
Conversion efficiency: 90% or higher, conforms to JIS
Gas Flow Rate:
0.5 L/min
Ambient Temperature: -5 to 45°C
Power Supply:
K9350LE;
100 V AC, 50/60 Hz
K9350LF;
100 to 240 V AC, 50/60 Hz
Power Consumption: Approx. 85 VA
Weight:
Approx. 1.1 kg
Sample gas requirements:
Dust/drain removed, gas temperature at 150°C or lower
Safety conforming standards:
Installation altitude;
2000 m or less
Pollution degree;
2 (Note)
Installation category; II (Note)
・ Installation category, called over voltage category, specifies impulse with standing voltage. Category II
is for electrical equipment.
・ Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which reduce dielectric strength.
Degree 2 is the normal indoor environment.
Note
Safety;
EN61010-1
EMC; EN61326-1 Class A, Table 2 (For use in industrial locations),
EN61326-2-3, EN61000-3-2, EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
Note: The product mounted in a steel enclosure conforms to the requirements of EMC directive.
CAUTION: The instrument is a Class A product, and it is designed for use in the industrial environment. Please use this
instrument in the industrial environment only.
Spare Parts for one-year-use
Item
Part No.
Qty
2
Catalyst for NO2/NO converter
K9350LP
Glass wool for NO2/NO
converter
K9350LQ 2
Fitting for NO2/NO converter
K9350LV
2
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
K9350LE
<2SPECIFICATIONS>
2-20
Unit: mm
(150)
135
Gas Outlet ∅9.5/∅5.5
Wiring Diagram
(120)
105
(210)
7 8
4-∅5
Gas Inlet ∅9.5/∅5.5
Approx.
130
110
K9350LF
100 V AC
(Shared with internal wiring)
Unit: mm
Unit:mm
148
135
124
130
(NP)
Gas Outlet
ø9.5/5.5
27
105 160 212
(NP)
LNE
4-ø5
Power Inlet
Front View
12
Gas Inlet
ø9.5/5.5
Side View
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.
3-1
<3. INSTALLATION>
INSTALLATION
WARNNING
IR250 is Non-Explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fire or other serious accidents.
CAUTION
• Inappropriate installation may cause hazardous accident such as tipover, electric shock, fire,
and injury.
• IR250 is heavy product. It should be handled carefully to install. Otherwise it may cause
tipover, fall, injury and hazardous accident.
• During installation work, care should be taken to keep the unit free from cable chips or other
foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
3.1
Where to install equipment
Installation Requirements
• The product is for indoor use. Avoid exposure to direct sunlight, weather, and radiant heat
from hot substances. If exposure to such conditions are unavoidable, a protective hood or
cover should be prepared.
• Vibration-free environment
• A clean atmosphere
Ambient temperature:
-5 to 45°C
(40°C at max if suffix code -C, -G, -H, -P is selected and the electric power being used is
larger than 200 V power supply)
Ambient humidity:
95%RH or less, no condensation
Clear the space as below if the analyzer is installed by itself.
analyzer
(when installed)
Approx.
500mm
Approx.
400mm
analyzer
(drawer out)
Approx. 500mm
Approx.
300mm
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-2
<3. INSTALLATION>
To install the analyzer for optimum performance, select a location that meets the following
conditions;
• IR250 is system built-in type, embedded in a steel rack, or enclosure of steel sheet.
Keep a minimum clearance of 10 cm above the analyzer for heat dissipation. The same
size of clearance is required for each analyzer when you install several units on a multistage
rack.
• Circuit breaker rating should meet the analyzer rating, that meets IEC60947-1 and
IEC60947-3.
• Circuit breaker should be installed near the analyzer where an operator can access it.
• A label that clearly identifies the breaker should be placed on it.
3.2
How to install
The next table shows the type of mounting and its code of equipment. Rack-mounting type of
equipment can be installed with sliding rails.
Mounting type (Code)
IR250-A
Rack-mounting with
sliding rails (-A)
Panel-mounting (-B)
Wall-mounting (-C)
Rack-mounting (-D)
See “2.4
IR250-B
IR250-C
IR250-D
External dimensions” on page 2-14.
Follow the next instruction and install equipment properly.
• Front side of equipment must stand vertically.
• The analyzer weight must be supported at the bottom of the casing.(except IR250-B)
• Protect the analyzer from vibrating. For example, install rubber material around the case to
minimize the vibration.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.3
3-3
<3. INSTALLATION>
Piping
WARNING
In piping, the following precautions should be observed.
Wrong piping may cause gas leakage.
If the leaking gas contains a toxic component, there is a risk of serious accident being induced.
Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring.
• Connect pipes correctly referring to the instruction manual.
• Exhaust should be led outdoors so that it will not remain in the locker and installation room.
• Exhaust from the analyzer should be relieved in the atmospheric air in order that an unnecessary
pressure will not be applied to the analyzer. Otherwise, any pipe in the analyzer may be
disconnected to cause gas leakage.
• For piping, use a pipe and a pressure reducing valve to which oil and grease are not adhering. If
such a material is adhering, a fire or the like accident may be caused.
CAUTION
Piping connection must be secured. Gas inlets have several types: reference gas inlet, outlet or
purge gas inlet. When the connection is insecure or wrong, combustible, toxic, explosive gas may
be accumulated inside the analyzer or system.
Observe the following when connecting the gas pipes.
• The pipes should be connected to the gas inlet and outlet on the analyzer, respectively.
• Connect the sampling system to the instrument by using corrosion-resistant tube. The
material should be such as Fluoropolymer (PTFE), stainless steel, or polyethylene. In case
where there is no danger of corrosion, don’t use rubber or soft vinyl tube. Analyzer indication
may become inaccurate due to the adsorption of gases.
• Piping connections are Rc1/4 (1/4 NPT) female-threaded. Cut the pipe as short as possible
for obtaining quick response. Pipe of approx. ∅4 mm (inside diameter) is recommended.
• Entry of dust in the instrument may cause operation fault. Use clean pipes and couplings.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<3. INSTALLATION>
3-4
Sampling gas outlet
(Rc1/4 or 1/4NPT)
Sampling gas outlet
(Rc1/4 or 1/4NPT)
Purge gas inlet
(Rc1/4 or 1/4NPT)
Purge gas inlet (Rc1/4 or 1/4NPT)
Sampling gas inlet (Rc1/4 or 1/4NPT)
Sampling gas outlet(shared with reference gas outlet in IR250-C)
(Rc1/4 or 1/4NPT)
Reference gas inlet (only for IR250-C)
(Rc1/4 or 1/4NPT)
Purge gas inlet (Rc1/4 or 1/4NPT)
Sampling gas inlet (Rc1/4 or 1/4NPT)
Sampling gas outlet(Rc1/4 or 1/4NPT)
Sampling gas inlet: Connect the pipe so that zero/span calibration standard gas or measured
gas pretreated with dehumidification is supplied properly.
The gas flow rate should be kept constant within the range of
1.0 ± 0.5 L/min. for IR250-A, 0.5 ± 0.2 L/min. for IR250-B,IR250-D, and
1.0 ± 0.2 L/min. for IR250-C.
Sampling gas (reference gas ) outlet: Measured gas is exhausted after measurement.
Connect the pipe so that the gas may escape through
the gas outlet into the atmosphere.
Purge gas inlet: It is used for purging the inside of the total gas analyzer. When the
analyzer must be purged, refer to Item 3.4, Purging inside Analyzer.
Use dry gas N2 or instrumentation air for purge gas. (flow rate of 1 L/min
or more should be used and no dust or mist is contained).
Reference gas inlet: For IR250-C, introduce reference gas for the sample switching. Connect
the pipe so that atmosphere pretreated with dehumidification, dry air, dry
N2 is supplied properly. The gas flow rate should be kept constant within
the range of 1.0 ± 0.2 L/min.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.4
3-5
<3. INSTALLATION>
Sampling
See “1.2
Sampling system configuration” on page 1-3
l Standard requirements of sampling gas
Flow rate:
IR250-A ;
1.0 ± 0.5 L/min
IR250-B, IR250-D ;0.5 ± 0.2 L/min
IR250-C;
1.0 ± 0.2 L/min
Temperature:
0 to 50°C
Pressure:
less than 10 kPa (atmosphere open at gas outlet)
Dust : Particle size should be smaller than 0.3µm and the mass concentration
should be less than 100µg / Nm3.
Mist :
Not allowed
Moisture
lower than 2°C saturation
Corrosive element :
less than 1 ppm
Use sampling gas with care as below;
• The dust contained in sample gas should be eliminated completely with filters. The filter at
the final stage should be capable of eliminating dust of 0.3 microns.
• The dew point of sample gas must be lower than the ambient temperature for preventing
formation of drain in the analyzer. If water vapor is contained in sample gas, its dew point
should be reduced down to about 0°C through a dehumidifier.
• If SO3 mist is contained in sample gas, the mist should be eliminated with a mist filter,
cooler, etc. Eliminate other mist in the same way.
• If a large amount of highly corrosive gas such as Cl2, F2 or HCl is contained in sample gas,
the service life of analyzer will be shortened. So, avoid such gases.
• Sample gas temperature is allowed within a range from 0 to 50°C. Pay attention not to flow
hot gas directly into the analyzer.
• Flow rate must not be changed during a measurement.
• Install a flowmeter as shown in a configuration (1.2), and check the flow rate.
l Standard gas for calibration
Prepare the standard gas for zero / span calibration. Periodical calibration are necessary. (weekly
is recommended)
Measuring components excluding O2
Zero gas:
N2 gas
Span gas: Gas with concentration of 90% level or higher in the range of each
corresponding component (recommended) (the rest of it is N2)
Measuring O2 (Built-in paramagnetic O2 analyzer)
N2 gas
Zero gas:
Span gas: Gas with concentration of 90% level or higher in the range of each
corresponding component
(the rest of it is N2 gas or atmosphere (21% O2 ))
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-6
<3. INSTALLATION>
Measuring O2 (External Zirconia O2 analyzer)
Zero gas Dry air, or atmospheric air, but atmospheric air should not be
contained in zero gas if CO2 meter is provided.
Span gas
1 to 2% gas (the rest of it is N2)
It is understood to influence the calibration model of the infrared gas analyzer when a lot of H2,
He, and Ar are included in the measurement gas (pressure broadening).
The span gas must use a gas near the composition of the measurement gas when you measure
the gas like the above-mentioned.
l Purging of instrument inside
The inside of instrument need not be purged generally except for the following cases.
• A combustible gas component is contained in sample gas.
• Corrosive gas is contained in the atmospheric air at the installation site.
• The same gas as the sample gas component is contained in the atmospheric air at the
installation site.
In such cases as above, the inside of analyzer should be purged with the air for instrumentation
or N2. Purging flow rate should be about 1 L/min.
If dust or mist is contained in purging gas, it should be eliminated completely in advance.
l Pressure at sampling gas outlet
Pressure at the sampling gas outlet should be adjusted to atmospheric pressure.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.5
3-7
<3. INSTALLATION>
Wiring
CAUTION
• Wiring work must be performed with the main power set to OFF to prevent electric shocks.
• Enforce construction of class-D grounding wire by all means, 100 Ω or less.
If the specified grounding construction is neglected, a shock hazard or fault may be caused.
• Wires should be the proper one meeting the ratings of this instrument. If using a wire which
cannot endure the ratings, a fire may occur.
• Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating
may cause fire.
See the following pages on wiring instruction.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<3. INSTALLATION>
3-8
Commun.Connector
Terminal
Terminal
Terminal
block 5
block 1 Terminal block 3Terminal
block 2
Power supply
Fuse
block 4
Communication
External input connector connector (RS-485)
(A/I)
Analog output connector (A/O)
Fuse
Power
Digital input/output connector
(DIO1 to DIO3 from bottom to the top) supply
Power terminal
(M4 Terminal Block Type)
(Power Cable code: -NN)
COMP2(2nd comp.)I/O terminal
COMP1(1st comp.)I/O terminal
Power supply
Wiring port
(Rubber grommets)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-9
<3. INSTALLATION>
n Power Supply
Connect the given power supply to the power terminal, and connect the ground wire to the
grounding terminal (P). Be sure to perform protective earth connection. Use solderless terminals
(for M4) for connection to the terminals (power and earth).
Application line diameter ø9.5 to ø10.5 mm.
M4 thread terminal
P
N
Power inlet
L
attached ferrite core
AC 100 to 240V
Grounding 2-pole plug
Earth cable
CAUTION
After wiring , be sure to place the protective cover for the M4 thread terminal blocks to assure
safety.
l When noise source is in the vicinity
• Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder)
that generates much electrical noise. If using the instrument near such a noise generating unit is
unavoidable, use a different power line to avoid noise
• Mount a noise suppressor such as varistor or spark quencher as shown below to the noise
generating unit when noise is generated from relays or solenoid valves. Mount the suppressor
near the noise generating source, otherwise it will have no effect.
Main unit
power supply
varistor or
spark quencher
install near the source
Noise
generating
source
n I/O Wiring Diagram
NOTE
• The analog output signals of the instrument are not isolated individually. It is recommended
to isolate the signals individually to eliminate the interference from the unnecessary signals
or the effect of external interference, especially it the cable exceeds 30 meters or leads to
outdoors.
• To avoid external interference, wiring of analog output signal, O2 sensor input and contact
input should be run separately from that of power supply and contact output.
• To avoid the effect of noise generated from external units, be sure to ground the analyzer
main unit.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-10
<3. INSTALLATION>
n I/O Wiring Diagram
l IR250-A
Applicable diameter of conducting wire is within Ø 1.9(AWG20) to Ø 2.1 mm(AWG18)
Terminal block 1
CH5 output –
(CH5_OUT) +
1
11
2
12
CH4 output –
(CH4_OUT) +
3
13
4
14
CH3 output –
(CH3_OUT) +
5
15
6
16
CH2 output –
(CH2_OUT) +
7
17
8
18
CH1 output –
(CH1_OUT) +
9
19
10
20
Unassigned
* O2 sensor input –
(O2_IN) +
Unassigned
– Unassigned
+
Unassigned
– CH7 output
+ (CH7_OUT)
Unassigned
– CH6 output
+ (CH6_OUT)
Unassigned
1
11
2
12
3
13
4
14
5
15
6
16
7
17
8
18
9
19
10
20
(M3.5 screw)
Terminal block 3
Terminal block 4
11
2
12
3
13
4
14
Remote hold input
(R_HOLD)
5
15
6
16
Average value reset
input (RESET)
7
17
8
18
Auto calibration
remote start
input (R_CAL)
9
19
10
20
Unassigned
Unassigned
(M3.5 screw)
1
Unassigned
Terminal block 2
Instrument error
(FAULT)
Contact output for CH4
span calibration
(SPAN_CH4)
1
11
2
12
CH4 range identification
signal output (RNG_IDCH4)
Contact output for CH3
span calibration
(SPAN_CH3)
3
13
4
14
CH3 range identification
signal output (RNG_IDCH3)
Contact output for CH2
span calibration
(SPAN_CH2)
5
15
6
16
CH2 range identification
signal output (RNG_IDCH2)
Contact output for CH1
span calibration
(SPAN_CH1)
7
17
8
18
CH1 range identification
signal output (RNG_IDCH1)
Contact output for
zero calibration
(ZERO)
9
19
10
20
(M3.5 screw)
Unassigned
CH4 remote range
changeover input
(R_RNG_CH4)
CH3 remote range
changeover input
(R_RNG_CH3)
CH2 remote range
changeover input
(R_RNG_CH2)
CH1 remote range
changeover input
(R_RNG_CH1)
Unassigned
Auto calibration status contact
output (ACAL/MNT)
Calibration error
contact output
(CAL_ALM)
Pump ON/OFF contact
output (PUMP)
Unassigned
(M3.5 screw)
Terminal block 5
Unassigned
CH3 alarm output
(ALM_CH3)
CH2 alarm output
(ALM_CH2)
CH1 alarm output
(ALM_CH1)
1
11
2
12
3
13
4
14
5
15
6
16
7
17
8
18
9
19
10
20
*O2 sensor input is used when an external O2 analyzer is selected.
Unassigned
Note: Unassigned terminals are used for internal connection.
So they should not be used as repeating terminals either.
Peak alarm
output (PEAK_ALM)
CH4 alarm output
(ALM_CH4)
(M3.5 screw)
NOTE
After connecting contact I/O signal cables, install ferrite core on each terminal stand.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-11
<3. INSTALLATION>
l IR250-B
Applicable diameter of conducting wire is within Ø 1.9(AWG20) to Ø 2.1 mm(AWG18)
AUTO CAL
Unassigned
Unassigned
Unassigned
1
7
2
8
3
9
4
10
5
11
6
12
Unassigned
Unassigned
Unassigned
COMP2
(2nd component)
0-1 V DC
7
Unassigned
1
+
(instantaneous −
8
2
value )
Unassigned
9
4-20mA DC + 3
(instantaneous − 4 10
Unassigned
value)
5 11
Unassigned
Unassigned
6 12
(M3.5 screw)
COMP1
(1st component)
0-1 V DC
7
Unassigned
1
+
(instantaneous
value)
4-20mA DC
(instantaneous
value)
FAULT
(M3.5 screw)
−
+
−
2
8
3
9
4
10
5
11
6
12
Unassigned
Unassigned
Unassigned
(M3.5 screw)
(Note)Unassigned terminals are used for internal connection. So they should not be
used as repeating terminals, either.
When a Contact Output error occurs on IR250, the connection between 5 and 6 is
conductive.
1a contact 250 VAC, 2A (Resistive load)
NOTE
After connecting contact I/O signal cables, install ferrite core on each terminal stand.
l IR250-C, IR250-D
Connector for external input(A/I) Connector for analog output(A/O)
(for O2 input)
AI-
13
25
1
5
1
14
9
6
D-sub 25pin Male
AI+
D-sub 9pin Female
(GND)
AO1+
6
AO12
RTxD+
AO2+
7
AO23
RTxDAO3+
8
AO34
AO4+
9
AO45
AO5+
AO5AO6+
AO6AO7+
AO7AO8+
AO8AO9+
AO9AO10+
AO10AO11+
AO11AO12+ (Note) Display Ch number is same as the AO number under standard
specifications.
AO12Do not use connectors in blank.
1
1
14
2
15
3
16
4
17
5
18
6
19
7
20
8
21
9
22
10
23
11
24
12
25
13
RS-485 Connector
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
<3. INSTALLATION>
3-12
Connector for analog output (A/O)
Output signal : 4 20 m A DC or 0 to 1 V DC (select on order)
Minus lines for the insulation and signal are common from the
ground and internal circuit.
Allowable load:
4 to 20 mA DC, 100 kΩ or more
Connector for external input (A/I)
Input signal:
External zirconia O2 Sensor (ZX8D) : Zirconia O2 sensor signal
External O2 analyzer : 0 to 1 V DC (DC input resistor of 1MΩ or more)
• It is used when the cord 1 (external zirconia O2 Sensor(ZX8D)) or cord 2 (an external O2
analyzer) is specified as ordered. DO NOT connect when cord 3 (Built in paramagnetic O2
analyzer) is installed.
• Connect the dedicated connector (accessory) to the output of the external Zirconia analyzer or
the external O2 analyzer (received separately).
• In case of an external O2 analyzer, input a signal of 0 to 1 V DC with respect to O2 full scale of
the analyzer. The O2 concentration display, output, and O2 correction can be performed.
NOTE
In IR250-C and IR250-D, O2 sensor external input is not isolated. It is recommended to isolate
when an external O2 analyzer is installed apart from IR250. Zirconia O2 sensor (ZX8D) should be
installed at a location that is as close to this instrument as possible
How to connect the O2 signal to the dedicated connector (accessory).
Turn
+
To O2 sensor
-
Solder the signal cable to the terminal.
RS-485 Connector
For MODBUS protocol connection
Only when (-C) is selected.
Contact input / output (DIO) : Connector for digital input / output (DIO1,DIO2, DIO3)
Contact input signal : Voltage is applied from the external 12 to 24 V DC, max 15mA Photocoupler isolation (from each DI and ground)
Contact capacity : C contact relay output 24V/1A AC/DC resistive load
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Connector for digital input/output
(DIO1, DIO2, DIO3)
13
1
1
14
25
3-13
<3. INSTALLATION>
14
D-sub 25pin female
2
15
3
16
4
17
5
18
(Note) DIO1, DIO2, DIO3 have the same
internal circuit of the connector.
Do not use pins in blank.
6
19
7
20
8
21
9
22
10
23
11
24
NC
com
NO
NC
com
NO
NC
com
NO
NC
com
NO
NC
com
NO
DIO1
Connector
DI1+
DI1DI2+
DI2DI3+
DI3-
DIO2
Connector
DI4+
DI4DI5+
DI5DI6+
DI6-
DIO3
Connector
DI7+
DI7Digital input
DI8+
OFF : 0 V
DI8ON : 12 to 24 V DC
DI9+
DI9-
DO1
DO6
DO11
DO2
DO7
DO12
DO3
DO8
DO13
DO4
DO9
DO14
DO5
DO10
DO15
Digital output max.contact
load rating
24 V DC/1A
12
25
13
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-14
<3. INSTALLATION>
● digital input signal
DI1
DI2
DI3
DI4
Remote hold
Average value reset
Auto calibration start
Auto zero calibration
start
Remote range 1
Remote range 2
Remote range 3
Remote range 4
Remote range 5
DI5
DI6
DI7
DI8
DI9
● digital output signal
DO4
DO5
DO6
1-compo.analyzer
instrument error
calibration error
(auto calibration
status)
(zero)
(For span gas Ch1 )
―
2-compo.analyzer
3-compo.analyzer
instrument error
instrument error
calibration error
calibration error
(auto calibration status) (auto calibration
status)
(zero)
(zero)
(For span gas Ch1 )
(For span gas Ch1 )
(For span gas Ch2 )
(For span gas Ch2 )
4-compo.analyzer
instrument error
calibration error
(auto calibration
status)
(zero)
(For span gas Ch1 )
(For span gas Ch2 )
DO7
DO8
―
―
―
―
(For span gas Ch3 )
(For span gas Ch4 )
DO9
―
DO10 Range identification
Ch1
DO11 (Alarm 1)
DO12 (Alarm 2)
DO13 (Alarm 3)
DO14 (Alarm 4)
Range identification
Ch1
Range identification
Ch2
(Alarm 1)
(Alarm 2)
(Alarm 3)
(Alarm 4)
(For span gas Ch3 )
Range identification
Ch1
Range identification
Ch2
Range identification
Ch3
(Alarm 1)
(Alarm 2)
(Alarm 3)
(Alarm 4)
DO15 (Alarm 5)
(Alarm 5)
(Alarm 5)
DO1
DO2
DO3
Range identification
Ch1
Range identification
Ch2
(Alarm 1)
(Alarm 2)
(Alarm 3)
Range identification
Ch3
Range identification
Ch4
Note The normal open side (NO) of digital output is close when the function is active without range ID.
In case of range ID, normal open (NO) side is close with L-range.
The normal close (NC) side is close with H-range.
NOTE
• Isolated output (from each DO and ground)
To avoid external interference, wiring of analog output signal, O2 sensor input and contact input
should be run separately from that of power supply and contact output.
• To avoid the effect of noise generated from external units, be sure to ground the analyzer main
unit
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3-15
<3. INSTALLATION>
Suffix/option code
Option
comp
O2
Analyzer
code*
-A, -C, -D
-A
N
not specified
-A, -B, -C, -D -B
N
not specified
-A, -C, -D
-C
N
not specified
-A, -B, -C, -D -D
N
not specified
-A, -B, -C, -D -E
N
not specified
-A, -B, -C, -D -F
N
not specified
-A, -C, -D
-G
N
not specified
-A, -C, -D
-H
N
not specified
-A, -B, -D
-K
N
not specified
-A, -D
-L
N
not specified
-A, -D
-M
N
not specified
-A, -D
-P
N
not specified
-A, -C, -D
-A
1, 2, 3
not specified
-A, -C, -D
-B
1, 2, 3
not specified
-A, -C, -D
-C
1, 2, 3
not specified
-A, -C, -D
-D
1, 2, 3
not specified
-A, -C, -D
-E
1, 2, 3
not specified
-A, -C, -D
-F
1, 2, 3
not specified
-A, -C, -D
-G
1, 2, 3
not specified
-A, -C, -D
-H
1, 2, 3
not specified
-A, -D
-K
1, 2, 3
not specified
-A, -D
-L
1, 2, 3
not specified
-A, -D
-M
1, 2, 3
not specified
-A, -D
-P
1, 2, 3
not specified
NO
SO2
NO
CO
CO2
CO2
NO
NO
CH4
CH4
CO2
CO2
NOx
SO2
NOx
CO
CO2
CO2
NOx
NOx
CH4
CH4
CO2
CO2
-A, -C, -D
-A
1, 2, 3
/K
NOx O2
-A, -C, -D
-B
1, 2, 3
/K
SO2 O2
-A, -C, -D
-C
1, 2, 3
/K
-A, -C, -D
-D
1, 2, 3
/K
-A, -C, -D
-F
1, 2, 3
/K
-A, -C, -D
-G
1, 2, 3
/K
-A
-H
1, 2, 3
/K
-C, -D
-H
1, 2, 3
/K
NOx SO2 CO
-A, -D
-L
1, 2, 3
/K
CH4 CO
O2
-A, -D
-P
1, 2, 3
/K
CO2 CO
CH4
Type
*:
●
●
●
Output and corresponding channels
Ch1 Ch2
Ch3
Ch4
Ch5
Ch6
Ch7
Ch8
Ch9
Ch10
SO2
CO
CO
SO2 CO
CO
CH4
CO
O2
O2
SO2
O2
O2
CO
CO
SO2
O2
CO
CH4
CO
CH4
O2
O2
O2
CO
O2
O2
O2
CH4
O2
Correct
Correct
NOx
NOx
Ave.
Correct
Correct
SO2
SO2
Ave.
Correct
Correct Correct
NOx
NOx SO2
Ave
Correct
Correct
CO
CO O2
CO
Ave.
Correct
Correct
CO2 CO O2
CO
CO
Ave.
Correct
Correct Correct
NOx CO O2
NOx
NOx CO
Ave
Correct Correct
NOx SO2 CO
O2
NOx SO2
NOx SO2 O2
Correct
SO2
Ave
Correct
CO
Ave
Correct
CO
Correct Correct Correct
Correct Correct Correct
NOx SO2
CO
NOx SO2
CO
Ave
Ave
Ave
Correct
Correct
CO
CO
Ave.
Correct Correct
O2
CO
CO Ave
O2
In the column of Option code, “Not specified” refers any option code except /K (with O2 correction value output).
When CH1 has SO2 in its raw, it means the Ch1 corresponds with the SO2 component.
When option /K is specified, NO is indicated as NOx.
“Correct” represents O2 correction.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.6
3-16
<3. INSTALLATION>
Solenoid valve drive signal for calibration
3.6.1
Manual calibration
(See “6.
CALIBRATION”).
・Zero calibration
Calibration end
on
Cal.gas flow
Zero Calibration output
off
Ch1to 5 Span
calibration output
off
Output hold function
Output signal hold
(With hold ON setting)
Hold extension time
・Span calibration
Zero Calibration output
Calibration end
when selecting
CH2 with up/down keys
Ch1Span calibration
output
Ch2 Span calibration
output
Output hold function
(With hold ON setting)
Cal.gas flow
Output signal hold
Hold extension time
* The hold extension time depends on the gas flow time
of the automatic calibration settings.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
3.6.2
3-17
<3. INSTALLATION>
Auto calibration
(See “5.3
Alarm Setting”Automatic calibration)
Ch1 span
Ch3 span
calibration
calibration
Zero
Ch4 span
Ch2 span
calibration
calibration
calibration
Zero gas
Auto cal. start
Zero calibration output
Ch5 span
calibration
350S
Ch1 span calibration output
Ch2 span calibration output
Ch3 span calibration output
Ch4 span calibration output
Ch5 span calibration output
Ch1span gas
350S
Ch2span gas
350S
Ch3span gas
350S
Ch4span gas
350S
Ch5span gas
350S
Auto. Cal. Contact
Output hold function
Hold
extension time
(with hold ON setting)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Blank Page
4.
4-1
<4. OPERATION>
OPERATION
Check that the pipes are correctly connected to the gas sampling port and drain port. Check that
the analyzer is correctly wired as specified.
4.1
Warm-up operation and regular operation
Procedure
1. Turn ON the power.
In one or two seconds, the measurement screen will appear at the front panel.
2. IR250-A, IR250-C, IR250-D need about 4 hours, IR250-B need about 2 hours for warm-up.
During the warm-up, concentration reading may indicate “ ------”, which is not a sign or error
operation.
3. Set the required values according to Chapter 5, “Setup (User mode)”.
4. Perform zero / span calibration after the warm-up. See “6. CALIBRATION” on page 6-1.
5. Start supplying measuring gas to IR250. Start measuring.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
4.2
4-2
<4. OPERATION>
Keys and function on Panel
1
2
3
4
5
6
Displays
Controls
7
8
Control keys
Keys
Description
Keys
Description
(1) MODE
switch the mode.
(5) DOWN
move cursor downward to select an
item or decrement numeral value.
(2) ZERO
go to zero calibration.
(6) UP
move cursor upward to select an
item or increment numeral value.
(3) SPAN
go to span calibration.
(7) ESCAPE
return to a previous screen or
cancel the setting midway
(4) SIDE
change the selected item (by
moving the cursor) and numeral
digit.
(8) ENTER
confirm selected items or values, or
execute calibration.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
4.3
4-3
<4. OPERATION>
Mode and screens
IR250 has two modes, Measuring mode and User mode. User mode includes Maintenance
mode. During the measuring mode, calibration is performed.
The picture below shows how to navigate through modes . Detail varies with specification.
Measurement
mode
*1
ZERO Calibration (see 6.1)
Measurement
mode
Span Calibration (See 6.2)
Measurement
mode
User mode
Calibration value(5.2.1)
Zero calibration (5.2.2)
Calibration range (5.2.3)
Auto calibration components / range (5.2.4)
●Changeover of
Range(5.1)
●Setting about
Calibration(5.2)
●Alarm Setting
(5.3)
*3
●Setting Auto
calibration(5.4)
*3
●Setting Auto Zero
Calibration(5.5)
*3
●Setting of
Peak Alarm(5.6)
*2
●Parameter
Setting(5.7)
*1 Scroll screen on measure mode.
*2 only when optio code /A is selected
*3 not displayed on IR202-B
Start time
Cycle
Flow time
ON / OFF
Auto calibration Run / Stop
Start time
Cycle
Flow time
ON / OFF
Auto zero calibration Run / Stop
ON / OFF
Peak alarm value
Peak alarm count
Hysteresis
Current time
Key lock
Output Hold (5.7.1)
Response time (5.7.2)
Average period (5.7,2)
Backlight timer (5.7.2)
Contrast(5.7.2)
Maintenance (Enter password) (5.7.3)
IM 11G02Q01-01EN
F0802.ai
3rd Edition : May 31, 2017-00
4-4
<4. OPERATION>
n Measuring mode
When power turns on, always a measuring mode (see a sample below) appears.
To display more than 6 channels on a page, scroll the page with
keys.
(2)
(1)
(2)
(1)
Ch
0.0
0.0
0.0 0
0.0
0.0 0
NOX
0-200
Ch
SO2
0-200
Ch
CO2
0-20
Ch
CO
0-200
Ch
O2
0-25
(3)
Ch
8
ppm
Ch
9
ppm
Scroll
vol%
Ch
10
Ch
ppm
11
vol%
12
Ch
(4)
AV
CV
0.0
0.0
0.0
0.0
0.0
NOX
0-200
AV
CV
NOX
AV
CV
SO2
AV
CV
CO
0-200
0-200
0-200
AV
O2
0-25
1h
ppm
1h
ppm
1h
ppm
1h
ppm
1h
vol%
(3)
(5)
F0803.ai
Number
Display
(1)
Component
(2)
Concentration
(3)
Range
(4)
Unit
(5)
Average time
Description
Displays which component of measured gas showing instantaneous
value, converted instantaneous value, converted average value, etc.
Displays measured value of concentration.
Displays range values.
Displays unit , ppm or vol %
Displays average time.
l Instantaneous concentration value:
As a sample screen picture shows above, gas component such as NOx on Ch1(shown above),
SO2 on Ch2 (shown above) indicate what gas component is currently being measured and each
corresponding instantaneous concentration value.
l O2 correction concentration values:
Channels (Ch) with “cv” indicate O2 correction concentration value of each designated gas
component. The O2 correction concentration value is calculated from the following equation,
based on which sampling component is measured, what is the O2 instantaneous concentration
value, and O2 correction reference value. See 5. Other Parameters in 5.7.3 Maintenance Mode ,
21 - On
On:
The value of the O2 correction reference value
x Cs
21 - Os
(Value set by application)
Os:
Oxygen concentration (%)
Cs: Concentration of relevant measured component
Os is not more than O2 concentration limit in value.
The correction components are NO, CO and SO2 only.
Correction output=
l O2 correction concentration average value:
AV
AV
In the Ch (component) and O2 average value where “ CV
” is displayed as “ CV CO” in the
component display, a value obtained by averaging O2 corrected concentration value or O2
average value in a fixed time is output every 30 seconds.
Averaging time can be changed between 1 to 59 minutes or 1 to 4 hours according to the
average time settings .
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
4-5
<4. OPERATION>
(The averaging set time is displayed as “1h” , for instance, in the range display.)
* The measurement ranges of O2 correction concentration value and O2 correction concentration average value are the same as that of
the measuring components. Also, the measurement range of O2 average value is the same as that of O2.
n User mode, Calibration mode
The setting/selection screen is configured as shown below:
• In the status display area, the current status is displayed.
• In the message display area, messages associated with operation are displayed.
• When the display area shows items or numerical values to change or set, move cursor with
keys.
Message display area
Status display area
Setting item/selection item
display area
Cursor
Fig5-4.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
4-6
<4. OPERATION>
4.4
General operation
l Measuring mode
In the measurement mode, up to five
channels are displayed on a single page.
To view further channels beyond the first five
ones, press
keys to scroll.
Measuring mode
NOX
Ch
0-100
SO2
Ch
0-100
Ch
CO
Ch
O2
0-100
0-25
Ch
CV NOX
0-200
Ch
SO2
Ch
CO
2
0-100
3
0-100
O2
Ch
4
0-25
Ch
CVNOX
5
0-200
Ch
CVSO2
6
0-200
Ch
CV SO2
6
0-200
Ch
CV CO
7
0-200
Ch
CV NOX
8
0-100
Ch
9
Ch
AV
CV
SO2
AV
CV
CO
0-100
10
0-100
0.0
0.0
0.0
0.0 0
0.0 0
0.0
0.0
0.0 0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
See 6.1
See 6.2
ppm
ppm
ppm
ppm
1h
ppm
1h
ppm
1h
ppm
or
l User Mode
Switch Ranges
Calibration Parameters
Alarm setting*
Setting of Auto Calibration*
Setting of Auto Zero Calibration*
Peak Alarm Setting
Parameter Setting.
User Mode
Select an item with
UP/DOWN and ENT
Back with ESC
Switch Ranges
Calibration Parameters
Alarm Setting
Setting Auto Calibration
Setting Auto Zero Calibration
Setting Peak Alarm
Parameter Setting
For the further information, refer to “Chapter 5.
Set up (User mode)”.
F0805.ai
* : Availability depends on specification.
l How to close the mode
Press
to cancel the operation midway or complete setting. A previous screen will return.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.
5-1
<5.SETUP (User Mode)>
SETUP (User Mode)
In User Mode, select an item and go to each setting.
(1) On the measuring mode, press MODE
display User Mode.
and
(2) Move cursor and select an item to set. Press .
User Mode
Select an item with
UP/DOWN and ENT
Back with ESC
Switch Ranges
Calibration Parameters
Alarm Setting
Setting of Auto Calibration
Setting of Auto Zero Calibration
Peak Alarm Setting
Parameter Setting
(
)
to each screen
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.1
5-2
<5. SETUP (User Mode>
Switch Ranges
See “5.7.3
Maintenance Mode” on page 5-28 when you change a measuring range. This
section describes how to change over preset range.
MR: manual, select range manually on the current screen.
RR: remote, select range according to remote range switch contact input.
AR: auto, when the measured concentration value exceeds 90 % of Range1, switch
automatically from Range1 to Range 2. When the measured concentration value is smaller
than 80 % of Range1, automatically switch from Range2 to Range 1.
User mode Switch Ranges
Switch range
(1) Move cursor with
keys. Select
Ch(components) and press
.
Ch1
NOx
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select range with
UP/DOWN and ENT
Back with ESC
0–50.00
Range1
MR
0–500.0
Range2
0–50.00
Range1
AR
0–500.0
Range2
0–10.00
Range1
RR
0–20.00
Range2
Range1
0–50.00
MR
Range2
0–500.0
Range1
0–10.00
MR
Range2
0–25.00
(
(2) Move cursor with
keys. Select one out of
MR, AR, or RR and press
inverted..
. The selected one is
(3) When you select MR, Range1 or Range 2 are
inverted. Move cursor with
keys. Select
Range 1 or Range2 and press
.Measuring
Switch range
Ch1
NOx
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
MR
AR
RR
AR
AR
(4) After selection, press the
key.
CAUTION
Regardless of how you switch ranges , the range of
O2 correction values or O2 correction average values
is automatically switched by changing the range of
corresponding component.
)
Select method of Switch
ranges with UP/DOWN
and ENT. Back with ESC
Range1
0–50.00 ppm
Range2
0–500.0 ppm
Range1
0–50.00 ppm
Range2
0–500.0 ppm
Range1
0–10.00 vol%
Range2
0–20.00 vol%
Range1
0–50.00 ppm
Range2
0–500.0 ppm
Range1
0–10.00 vol%
Range2
0–25.00 vol%
(
)
End of Swich Ranges
starts with the selected range.
(3) The “Channel Selection” screen appears.
Press the
or
key until the
cursor goes
to a desired Ch (component).
ppm
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
vol%
Switch range
Ch1
NOx
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
MR
AR
RR
MR
MR
Select range with
UP/DOWN and ENT
Back with ESC
Range1 0–50.00 ppm
Range2 0–500.0 ppm
Range1 0–50.00 ppm
Range2 0–500.0 ppm
Range1 0–10.00 vol%
Range2 0–20.00 vol%
Range1 0–50.00 ppm
Range2 0–500.0 ppm
Range1 0–10.00 vol%
Range2 0–25.00 vol%
(
)
End of Swich Ranges
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-3
<5.SETUP (User Mode)>
l Range identification contact operation
Regardless of how you switch ranges (MR, AR, RR), the range identification contact output
corresponding to each Ch (component) is conductive when Range 1 is active, and open when
Range 2 is active.
If the measurement value is held by remote hold contact input or during calibration routine and
range switch conditions are met, the contact will change position only after the hold condition is
removed. (See “3.5
Wiring” on page 3-7)
5.2
Calibration Parameters
Set Calibration concentration and other value.
User mode Calibration Parameters
Move cursor with
keys. Select an item
among Calibration Value,
About ZERO Calibration,
About Calibration Range,
About Calibration Components / Range.
Select an item with UP/
Cal. Parameters Down and ENT
Back with ESC
Calibration Value
About ZERO Calibration
About Calibration Range
Auto Calibration Components / Range
(
)
screen of each mode
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-4
<5. SETUP (User Mode>
5.2.1
Calibration Value
It allows you to set concentrations of the standard gas (zero and span) of each channel used for
calibration.
(1) Move cursor with
keys. Select
User mode Cal.Parameters Cal.Value
Ch(components) and press
.
Cal. Settings
Cal. Value
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select an item with UP/
DOWN and ENT
Back with ESC
RANGE
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10vol%
0–20vol%
0–50.00ppm
0–500.0ppm
0–10vol%
0–25vol%
ZERO
+0000. 0
+00000
+0000. 0
+00000
+00. 000
+000. 00
+0000. 0
+00000
21. 00
21. 00
(
keys and select a
(2) Move cursor with
concentration to set. Press
. then the value is
inverted.
(3) Enter concentration value of ZERO or SPAN. Enter
keys.
numerical value in each digit with
Press
to shift to another digit.
After entering, press
.to settle the value. Next
calibration will use those values.
• Applicable value range ( NO EXCEPTIONAL
ranging value other than these)
NOx, SO2, CO2, CO, CH4, and built-in O2 sensor
Span gas:
to 105% FS (full scale of each
1
range)
Zirconia O2 measurement
Zero gas:
Span gas:
5 to 25 vol% (21.00 vol% for
atmospheric air)
0.01 to 5 vol%
Cal. Settings
Cal. Value
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
SPAN
0050. 0
0500. 0
0050. 0
0500. 0
10. 000
020. 00
0050. 0
0500. 0
01. 00
01. 00
)
Select an item with UP/
DOWN and ENT
Back with ESC
RANGE
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10vol%
0–20vol%
0–50.00ppm
0–500.0ppm
0–10vol%
0–25vol%
ZERO
+0000. 0
+00000
+0000. 0
+00000
+00. 000
+000. 00
+0000. 0
+00000
21. 00
21. 00
SPAN
0050. 0
0500. 0
0050. 0
0500. 0
10. 000
020. 00
0050. 0
0500. 0
01. 00
01. 00
Cursor for setting value
Cal. Settings
Cal. Value
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Set calibration value
RANGE
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10vol%
0–20vol%
0–50.00ppm
0–500.0ppm
0–10vol%
0–25vol%
ZERO
+0000. 0
+00000
+0000. 0
+00000
+00. 000
+000. 00
+0000. 0
+00000
21. 00
21. 00
SPAN
0050. 0
0500. 0
0050. 0
0500. 0
10. 000
020. 00
0050. 0
0500. 0
01. 00
01. 00
END of (setting)
Calibration Value
CAUTION
For Zirconia O2 measurement, set the concentration value of ZERO to 21.00 ( when atmospheric
air is used). Set to the concentration value as indicated on the cylinder when cylinder air is used.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.2.2
5-5
<5.SETUP (User Mode)>
About ZERO Calibration
Manual calibration has two mode, “at once” and “each.” “At once” enables you to calibrate some
desired component all at once. “Each” enables you to select each component and calibrate one
by one.
User mode Cal.Parameters Zero Calibration
keys. Select
(1) Move cursor with
Ch(components) and press
.
Cal. Settings
ZERO Cali.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select Ch No.
for setting calibration value
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
(2) Move the cursor to select At Once or Each.
•Setting “each”
Select the Ch (component) on the manual zero
calibration screen and then perform the zero
calibration.
•Setting “at once”
At a manual zero calibration, Ch (components)
for which “at once” was selected can
simultaneously be zero-calibrated
NOTE
Cal. Settings
ZERO Cali.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
at once
at once
at once
each
)
Select each or both Ch
at ZERO Calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
Set “At Once” when you use cylinder or ambient air as
zero gas.
at once
at once
at once
at once
at once
each
)
End of Zero Calibration Setting
.to activate the designated calibration.
(3) After setting completion, press
l Manual calibration (See “6.
CALIBRATION” on page 6-1)
When selecting “each” on every Ch.
When selecting “at once” on each Ch
only one cursor is inverted.
all cursors to show selected ranges are inverted.
ZERO Cali
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
ENT ; Go on calibration of
selected CH
ESC : Not calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0
0
0. 00
0
21. 00
ZERO Cali
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
ENT ; Go on calibration of
selected CH
ESC : Not calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0
0
0. 00
0
21. 00
F0907.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-6
<5. SETUP (User Mode>
5.2.3
About Calibration Range
This mode is used to set if the range of each CH (component) at the zero and span calibration
(manual calibration or auto calibration) should be calibrated with a single range or 2 ranges.
User Mode Cali.Value Call range
(1) Move cursor with
keys. Select
Ch(components) and press
.
Cali. settings
Cali range
(2) Select Both or Current with
keys.
• When selecting “both”, Range 1 and Range 2 of
the set Ch are calibrated together as zero or span
calibration.
• When selecting “current”, the range alone displayed
at the set Ch is calibrated.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select Ch No.
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
Cali. settings
Cali range
both
current
current
both
current
)
Set calibration range
current or both range
Ch1
NOx
Range 1: 0 to 50 ppm
Range 2: 0 to 500 ppm
both
Ch2
SO2
Range 1: 0 to 50 ppm
Range 2: 0 to 500 ppm
current
Ch1: Range 1 and Range 2 are calibrated together, with zero or span
calibration.
Ch2: Only currently displayed range is calibrated, with zero or span
calibration.
(3) After setting completion, press
calibration range.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
.to save the
both
current
current
both
current
)
End of Calibration Range setting
l
Manual Calibration (see “6.
CALIBRATION” on page 6-1)
When you set “both” for Nox or CO, two highlighted cursors appear in both ranges.
ZERO Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
ENT: Go on Calibration of
selected Ch
ESC : Not calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0
0
0. 00
0
21. 00
F0910.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.2.4
5-7
<5.SETUP (User Mode)>
Auto Calibration Component / Range
Select the Ch (component) and the range for which auto calibration is to be performed. The Ch
for which “AR” has been selected as range switch mode is calibrated in the range set here. Auto
calibration and the manual calibration of the component for which “AR” has been selected as
range switch mode are performed in the range selected here. In this case, once the calibration is
started, the range is automatically switched, and on completion of the calibration, the original
range is resumed. The range identification contact is interlocked with the range after the switch.
However, if the hold setting is set to “ON,” the contact status before calibration is maintained.
(1) Move cursor with
keys. Select
Ch(components) and press
.
(2) Select range to calibrate and press
.
Calibration proceeds within the selected range.
(3) During (2)(above), press
“disable”. Press
.
(4) Move cursor with
“disable”, and press
.
to invert “enable” or
keys to select “enable” or
When “enable” is selected, Auto calibration is made
in the following rules according to the setting.
• Zero calibration is carried out on Chs at once
which were set to enable .
• The span of Chs (components) which were set to
enable is calibrated in the ascending order of Ch
number.
NOTE
User Mode Cali.Value Auto Cal.Comp./ Range
Cali.setting
Auto Cali. /range
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select Ch No.
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
Cali.setting
Auto Cali. /range
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
enable
enable
enable
enable
enable
)
Select range for
auto calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
enable
enable
enable
enable
enable
(
The components which were set to enable are
calibrated at once at the time of auto calibration or auto
zero calibration, regardless of setting in 5.2.2 Setting
zero calibration .
)
End of Auto Cal.range setting
Cali.setting
Auto Cali.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Set enable or disable
for auto calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
enable
enable
enable
enable
enable
)
End of Auto Cali. Component setting
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.3
5-8
<5. SETUP (User Mode>
Alarm Setting
This setting does not apply to IR250-B.
This mode is used to set the upper and lower limit value to provide an alarm output against
measured concentration during measurement. Before changing the alarm setting, set the ON/
OFF to OFF.
User mode Alarm setting
Alarm setting
(1) Move cursor with
No. and press
.
Select Alarm or
Hysteresis setting
keys. Select Alarm
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Hysteresis 00 %FS 00 %FS
(
See the next page
(2) Select item to set and press
setting is completed.
. Now alarm
l Setting items
Alarm number corresponds to the same number
of alarm contact.
Channel: Channel setting targeted for
issuance of alarm. One Ch No.
can be selected for multiple
alarms.
H-Limit value: S
ets the high limit value
(concentration) of alarm.
Settable range is 0 to 100 % FS.
Settable on each range.
L-Limit value: S
ets the low limit value
(concentration) of alarm.
Settable range is 0 to 100 %
FS.Settable on each range.
Kind of Alarm: Select one among High limit
alarm, Low limit alarm, and High
limit or Low limit alarm, HH limit
alarm, and LL limit alarm.
Alarm Setting
Alarm-1
)
Select an item
with UP/DOWN and ENT
Back with ESC
Channel
Ch1
H-Limit Range1 100.0 ppm
Range2 2000 ppm
L-Limit Range1 000.0 ppm
Range2 0000 ppm
Kind of Alarm
High
ON/OFF OFF
(
Cursor for setting value
Alarm Setting
Alarm-1
)
Select an item
with UP/DOWN and ENT
Back with ESC
Ch1
Channel
H-Limit Range1 100.0 ppm
Range2 2000 ppm
L-Limit Range1 000.0 ppm
Range2 0000 ppm
Kind of Alarm
High
ON/OFF OFF
End of Alarm setting
High, HH …… Alarm contact
closes when
above H-limit alarm.
Low, LL …… Alarm contact closes when below L-limit alarm.
High or Low… Alarm contact closes when above H-limit value or below lower
limit value.
ON/OFF:
Enables the alarm function if set at ON, or disables it if set at OFF.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-9
<5.SETUP (User Mode)>
NOTE
Set the values so that upper limit value > lower limit value. Also set the value (upper limit valuelower limit value) > hysteresis error.
l When an alarm occurs (sample screen image)
When an upper limit alarm occurs, the “H-alarm” message lights. (Lower alarm message appears
L-alarm. High alarm” message lights at CH. ( HH-alarm,
LL-alarm are available as well for HH or LL limit.)
Ch
---NOX
H-alarm
0-100
Ch
SO2
NOTE
Ch
CO2
For 10 minutes after turning on power, the alarm judgment is
inactive.
Ch
CO
Ch
O2
0-100
0-10
0-100
0-25
ppm
0.0
0.0 0 3
0.0
2 1. 0 0
ppm
vol%
ppm
vol%
F0915.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-10
<5. SETUP (User Mode>
n Hysteresis setting
To prevent chattering of an alarm output near the alarm setting values, set hysteresis.
0 to 20% of full scale [% full scale (% FS)] represents the percentage with the width of the
component measurement range regarded as 100%.
NOTE
Hysteresis is common to all alarms.
(1) Move cursor with
and press
.
keys. Select hysteresis.
Alarm Setting
Select Alarm or
Hysteresis setting
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Hysteresis
(2) Enter hysteresis value with
keys in each
digit. Press
to switch to another digit. Press
.to validate the value.
00 %FS
(
Alarm Setting
) ENT
Select Alarm or
Hysteresis setting
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Hysteresis
00 %FS
ENT
End of Hysteresis setting
F0916.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-11
<5.SETUP (User Mode)>
l Hysteresis (In case of upper limit alarm)
If hysteresis values exceed the upper limit as shown in the graph below, an alarm output is
provided.
Once the alarm output is turned “ON”, it remains “ON” until the value falls below the set lower limit
of the hysteresis indication.
Upper limit value
Hysteresis
width
ON (conductive)
Alarm
OFF (open)
F6_3_2_1.ai
5.4
Setting Auto calibration
This calibration does not apply to IR250-B.
CAUTION
•
When an auto calibration starts, the measurement screen appears automatically.
•
During Auto calibration, none of keys works other than Key Lock ON/OFF (see 5.7) or key to
stop forcedly auto calibration.
“Stop Auto Calibration” cannot be performed with the key lock to ON. To cancel auto
calibration forcedly, set the key lock to OFF and then execute “Stop Auto Calibration”.
•
When power failure or power off occurs, after resuming, the auto calibration starts again at
previously-set start time and repeats on the set cycle.
Remote start
Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the
remote start input short-circuited for at least 1.5 seconds.
With input (Hold at least 1.5 sec)
Remote start input
Without input
F0921.ai
It is carried out automatically at the timing set for ZERO / SPAN calibration.
To change the setting of auto calibration, switch auto calibration setting ON/OFF to “OFF” and
change the parameter.
Auto calibration status contact output is conductive during auto calibration (NO side), and is open
in other cases.(See 3.5 Wiring)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-12
<5. SETUP (User Mode>
User Mode Auto Cali.Setting
(1) Move cursor with
press
.
Set Auto Cal.
keys. Select item and
First Set ON/OFF to “OFF”, start changing value.
Select setting item
Start Time
Cycle
Flow Time
ON/OFF
SUN 12 : 00
07
day
OFF
Time:MON 12 : 34
(2) Enter numerical values with
shift to another digit.
. Press
to
(
See next page
)
Set Auto Cal. Select Start Time
Setting items
• Start Time : S
etting at the first calibration (day of the
week, hour, minute)
• Cycle : A period between the start time of one
calibration and another (unit : hour/day)
• Flow Time : The time required for the calibration gas
to be replaced in the cell
• ON/OFF :
Auto Calibration Run
Auto zero calibration ON or OFF
(3) After setting, press
. The set value will be
activated next time you enter the auto calibration.
Set back ON/OFF to “ON”.
Start Time
Cycle
Flow Time
ON/OFF
SUN 12 : 00
07
day
OFF
Time:MON 12 : 34
Auto Calibration Run
End of Auto Calibration setting
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-13
<5.SETUP (User Mode)>
n Setting Flow rate
(1) Select “Flow time” and press
.Go to set Flow time.
User Mode Auto Cali.Setting Flowtime
Set Auto Cal.
ZERO
Ch1Span
Ch2Span
Ch3Span
Ch4Span
Ch5Span
Ex. time
Select setting item
350 sec.
350 sec.
350 sec.
350 sec.
300 sec.
300 sec.
300 sec.
(
(2) Move cursor with
press
.
keys. Select gas. and
Set Auto Cal.
ZERO
Ch1Span
Ch2Span
Ch3Span
Ch4Span
Ch5Span
Ex. time
(3) You can change the parameter, which is inverted.
(60 to 900 sec)
)
Set flow item of
calibration gas 60 to 900
sec
350 sec.
350 sec.
350 sec.
350 sec.
300 sec.
300 sec.
300 sec.
End of Gas flow time setting
NOTE
Set the flow time value of Zero / Span so that the value is five times larger than that of Response
(5.7.2)
(4) After changing values, press
. Press
to go back to Auto Calibration Setting.
CAUTION
• The Chs only currently being used are displayed on the screen.
• The Ex. time is the output signal hold extension time after the completion of calibration. It is
valid only when the hold setting is set to “ON.” .
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-14
<5. SETUP (User Mode>
l Example
Start Time
Cycle
Flow Time
SUN
1
12:00
day
350 sec.
350 sec.
350 sec.
350 sec.
300 sec.
300 sec.
300 sec.
ZERO
Ch1Span
Ch2Span
Ch3Span
Ch4Span
Ch5Span
Ex. time
ON/OFF
ON
If you perform auto calibration as shown above,
Sunday
12:00
Monday
12:00
Cycle
Tuesday
12:00
:Auto Calibration
Ch1
Zero Cal Span Ca
350sec. 350sec
Ch2
Span Ca
350sec
Ch3
Span Ca
350sec
Ch4
Span Ca
300sec
Ch5
Replace
Span Cal ment time
300sec
300sec
Flow time
F0920.ai
The table (above) shows an example when Ch1 through Ch5 are set “enable” in an auto
calibration setting of item 5.2.4.
n Execute auto calibration and cancellation
This mode is used to perform auto calibration one time or cancel the auto calibration forcibly.
l One-time auto calibration
(1) In User mode, go to Set Auto Cal, and Auto
Calibration. Press
.
Set Auto Cal.
Select setting item
SUN 12 : 00
07
day
Start Time
Cycle
Flow Time
ON/OFF
OFF
Time:MON 12 : 34
Auto Calibration
Run
(
Set Auto Cal.
(2) “Run” is highlighted and a message appears. Press
.to execute Auto Calibration or
to cancel.
)
Auto Cal Run
ENT:Run / Stop
ESC:Cancel
Start Time
Cycle
Flow Time
ON/OFF
SUN 12 : 00
07
day
OFF
Time:MON 12 : 34
Auto Calibratio Run
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-15
<5.SETUP (User Mode)>
l Forced cancellation of auto calibration (While Auto Calibration being
conducted)
(1) In the User mode, go to Set Auto Cal, and Auto
Calibration. Press
. If you do while auto
calibration is being conducted, Auto Calibration
Stop is displayed.
Set Auto Cal.
Select setting item
SUN 12 : 00
07
day
Start Time
Cycle
Flow Time
ON/OFF
OFF
Time:MON 12 : 34
Auto Calibration
Run
(
Set Auto Cal.
(2) “Stop” is inverted. A message appears, prompting
you to verify that you want to cancel or continue
auto calibration. To cancel the auto calibration
forcibly, press the
. ey. If you press the
key,
auto calibration is cancelled (not stopped).
NOXCal
ZERO
0-100
SO2Cal
ZERO
0-100
Ch
CO2
Ch
CO
Ch
O2
0-10
0-100
0-25
Ch
NOXCal
SPAN
0-100
Ch
SO2
Ch
CO2
Ch
CO
0-100
0-10
• Ch1 span calibration
A message, “Span cal.” blinks at Ch1.
0-100
Ch
O2
0-25
Ch
NOX
Ch
SO2Cal
SPAN
0-100
0-100
• Ch2 span calibration
A message, “Span cal.” blinks at Ch2.
OFF
Auto Calibratio Stop
Ch
A message, “Zero cal.” blinks at Ch1 and Ch2.
SUN 12 : 00
07
day
Time:MON 12 : 34
Ch
• Zero calibration
Auto Cal Run
ENT:Run / Stop
ESC:Cancel
Start Time
Cycle
Flow Time
ON/OFF
n Auto Calibration screen”
When the auto calibration components (see Section 5.2.4) to
“Ch1 is “enable” and “Ch2 is “enable” ,
)
Ch
CO2
Ch
CO
0-10
0-100
Ch
O2
0-25
0.5
0.3
0.0 0 0
0.0
2 1. 0 2
9 0.8
0.0
0.0 0
0.0
0.0 0
0.0
9 5.0
0.0 0
0.0
0.0 0
ppm
ppm
vol%
ppm
vol%
ppm
ppm
vol%
ppm
vol%
ppm
ppm
vol%
ppm
vol%
F0924.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.5
5-16
<5. SETUP (User Mode>
Setting of Auto Zero Calibration
This function is not available in IR250-B.
CAUTION
• When an auto zero calibration starts, the measurement screen automatically appears.
• Any operation other than “Auto Zero Calibration Stop” (see Section 6.5.2) is not permitted
during auto zero calibration. “Auto Zero Calibration Stop” cannot be performed with the key
lock set to ON. To cancel auto zero calibration forcibly, set the key lock to OFF and then
execute “Auto Zero Calibration Stop”.
• When power is turned off (including power failure) and resumes, auto zero calibration is
carried out at the preset start time and repeats in the programmed cycle.
• If the auto calibration period and auto zero calibration period have overlapped, the auto
calibration is retained, ignoring the auto zero calibration of that period.
• When the hold setting is set to ON, the hold time of auto calibration contact and
measurement value output signal is extended after calibration for gas replacement time.
Remote start
Whether the auto zero calibration is set at ON or OFF, an auto zero calibration is available by
remote start input.
With input (Hold at least 1.5 sec)
Remote start input
Without input
F0921.ai
Auto zero calibration is automatically carried out when zero calibration are set.
Components for which a calibration is to be made are determined by setting of auto calibration
component in 6.2.5.
Before changing the setting of auto zero calibration, set the ON/OFF to “OFF”.
In Auto calibration status, contact output is conductive during auto zero calibration (NO side),
and is open in other cases.(See 3.5 Wiring)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-17
<5.SETUP (User Mode)>
(1)
Move cursor with
and press
.
First Set ON/OFF to “OFF”, start changing
value.
(2)
Enter numerical values with
.
Press
to shift cursor to right.
Setting items
• Start Time : S
etting at the first calibration (day
of the week, hour, minute)
• Cycle : A period between the start time of
one calibration and another
(unit : hour/day)
Setting range ; 1 to 99 hours or 1
to 40 days(initial : 7 days)
• Flow Time : The time required for the
calibration gas to be replaced in
the cell
Setting range ; 60 to 900 sec for
each (initial ; 300 sec)
• ON/OFF :
User Mode Set Auto Zero Cal
keys. Select item
Auto zero calibration ON or OFF
(3) After setting completion, press
. The set
value will be activated next time you perform
auto zero calibration. Set back ON/OFF to
“ON”.
Set Auto
Zero Cali
Start Time
Cycle
Flow Time
ON/OFF
Select setting item
SUN 12 : 00
07
day
OFF
Time:MON 12 : 34
Auto Zero Calibration
Run
See next page
Set Auto
Zero Cali
Set start time
Start Time
Cycle
Flow Time
ON/OFF
SUN 12 : 00
07
day
300 s
OFF
Time:MON 12 : 34
Auto Zero Calibration
Run
End of ZeroCalibration setting
l Example
Start time
SUN
12:00
Cycle
12
hour
Flow time
300
sec
ON / OFF
ON
when you perform Auto Zero Calibration under the condition above
SUNDAY
12:00
Cycle
MONDAY
00:00
MONDAY
12:00
:Auto Zero Calibration
Zero Cal
Replacement time
300sec
300sec
Flow
time
Gas replacement time after calibration is same as the value
F0926.ai
of the flow time.
(An example of “Ch1: through Ch5: enable,” as given in item 5.2.4 “Setting of auto calibration
components / range”)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-18
<5. SETUP (User Mode>
n Forced run / stop of auto zero calibration
This mode is used to cancel the auto zero calibration forcibly or perform auto zero calibration one
time only.
l Execution auto zero calibration (one time only)
(1) In the User mode, go to Set Auto Cal, and Auto
Zero Calibration.Press
. If you do while auto
zero calibration is being conducted, “Auto Zero
Calibration Stop” is displayed.
Set Auto Zero
Cal.
Select setting item
SUN 12 : 00
07
day
Start Time
Cycle
Flow Time
ON/OFF
OFF
Time : MON 12 : 34
Auto Zero Calibration
Run
(
Set Auto Cal.
(2) “Run” is inverted. A message appears, prompting
you to verify that you want to continue auto zero
calibration. Press the
to execute the auto zero
calibration, and
to cancel.
)
Auto Cal Run
ENT : Run / Stop
ESC : Cancel
SUN 12 : 00
07
day
Start Time
Cycle
Flow Time
ON/OFF
OFF
Time : MON 12 : 34
Auto Zero Cal. Run
l Forced cancellation of auto calibration (While Auto Calibration being
conducted)
(1) In the User mode, go to Set Auto Cal. Move cursor
down to Auto Calibration.Press
. If you do
while auto calibration is being conducted, “Auto
Calibration Stop” is displayed.
Set Auto Cal.
Select setting item
SUN 12 : 00
07
day
Start Time
Cycle
Flow Time
ON/OFF
OFF
Time:MON 12 : 34
Auto Calibration
Run
(
Set Auto Cal.
(2) “Stop” is inverted. A message appears, prompting
you to verify that you want to cancel or continue
auto calibration. To cancel the auto calibration
forcibly, press the
. If you press the
, auto
calibration is cancelled (not stopped).
)
Auto Cal Run
ENT:Run / Stop
ESC:Cancel
Start Time
Cycle
Flow Time
ON/OFF
SUN 12 : 00
07
day
OFF
Time:MON 12 : 34
Auto Calibratio Stop
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-19
<5.SETUP (User Mode)>
n Auto zero calibration (sample image)
In case where setting the auto calibration components (see Section 5.2.4) to “Ch1: enable” and
“Ch2: enable
• Zero calibration
A message, “Zero cal.” blinks at Ch1 and Ch2
Ch
NOXcal
ZERO
0-100
Ch
SO2cal
ZERO
0-100
Ch
CO2
Ch
CO
0-10
0-100
Ch
O2
0-25
0.5
0.3
0. 0 0
0.0
2 1. 0 2
ppm
ppm
vol%
ppm
vol%
F0929.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.6
5-20
<5. SETUP (User Mode>
Peak Alarm Setting
Peak alarm appears only when option code (/A) is specified.
During a measurement when the peak count exceeds the setting value to indicate the CO
concentration is over the upper limit, the peak alarm is output. This section describes how to
perform various settings of peak alarm.
NOTE
The peak count is not implemented during the first 10 minutes after the power is turned on.
Peak Alarm : Setting of peak alarm is performed with ON/OFF
Alarm Value : If a peak concentration exceeds the set alarm value one time, the peak counter
increments by 1.
settable value 0 to 1000 ppm
5 ppm step
(initial value: 500
ppm)
Alarm Count : When a peak in excess of the setting time occurs, a peak count alarm is output.
settable 1 to 99 counts
(initial value: 5 counts)
Hysteresis : To prevent possible chattering when the peak value may exceed the set peak
concentration by only 1 time, the peak count has an allowance in the hysteresis
width. 0 to 20% of full scale
(initial value: 0% of full scale)
[% full scale] represents the percentage with the CO range regarded as 100%.
User mode Peak Alarm setting
Peak Alarm
(1) Point the cursor to an item to set. Press
Select setting item
.
Peak Alarm
Alarm Value
Alarm Count
Hysteresis
OFF
0500 ppm
05 times
00 %FS
(
Set Peak Alarm
ON or OFF
Peak Alarm
(2) Execute setting or enter each numerical value.
to write the value in each digit and
Move
press
to shift between digits. Press
to store
values.
)
Peak Alarm
Alarm Value
Alarm Count
Hysteresis
OFF
0500 ppm
05 times
00 % FS
(
)
End of Peak Alarm Setting
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-21
<5.SETUP (User Mode)>
l Sample Action of peak alarm
(1) A peak count alarm is conductive at 5 times/hour. (ON)
(2) A peak count alarm is open at 4 times/hour. (OFF)
Peak concentration
Measured values
3 hours ago
2 hours ago
1 hour ago
Current hour
Peak count alarm
F6_6graph.ai
If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over
the set times per hour, a peak alarm contact output becomes closed (ON). If it is less than the set
section from
times per hour, it is open (OFF). Since 5 times of peaks /hour is marked at (1)
the above graph, the peak count alarm is turned ON. Since peaks of more than 5 times per 1
hour occur at the interval between (1) and (2)
, the peak count alarm remains ON. Since at
(2), peaks are reduced to 4 times per hour, it is turned OFF.
Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when
measured gas is fluctuated near the alarm value.
l Releasing peak count alarm
To release the peak count alarm, set the peak alarm to OFF.
Turning on the peak alarm initiates counting from 0.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-22
<5. SETUP (User Mode>
5.7
Parameter Setting
It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to
be set are as follows:
Current Time :
Sets the current day of the week, hour and time.
Key Lock :
Output Hold :
determines ON/OFF of key lock operation except the key lock OFF
determines whether Calibration Output is held or not. (See 5.7.1)
Response time : Sets the response time of electrical system.
Range ; NDIR measurement component: 1 to 9 cycles (initial: 1 cycle)
O2 sensor : 1 to 60s (initial : 15s)
*
Cycle value can be set by switching from Maintenance mode to Sample Switch setting. One cycle value equals a sum of (s):Flow
time of reference gas plus (s): Flow time of Sample gas.
Average Period : Sets the moving average time.
Range ; 1 to 59 minutes or 1 to 4 hours (initial ; 1 hour). Value by unit , 1
to 59 min, 1 to 4 hour.
Back Light :
Set display automatic light off and the time until backlight out.
Contrast :
Adjust contrast of LCD
Maintenance mode : Requires passwords to go to the Maintenance mode.(See 5.7.3)
Range ; 0000 to 9999 (initial : 0000)
(1) Move cursor with
.
to an item to set. Press
User mode Parameter
Parameter
Select setting item
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
OFF Current
ON
05min
0000
(
Parameter
(2) Enter numerical value and execute setting. Change
parameter or enter value by
keys. To move
cursor to the right, press
. Press
to establish
value.
)
Set day of week
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
OFF Current
ON
05min
0000
End of Parameter Setting
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.7.1
5-23
<5.SETUP (User Mode)>
Output Hold
By setting an output hold to ON, an output signal of each channel are held during the calibration
(manual calibration and auto calibration) and for the gas flow time (refer to Item 5.4, Setting
of Auto Calibration). Regardless of Hold ON/OFF setting, an output signal can be held via an
external input.
You can select the value for hold from the value immediately before entering output hold,
“current”, and arbitrary value, “setting.”.
If you cancel calibration after the calibration gas starts being supplied regardless automatically or
manually, the holding extension time will be performed.
Instantaneous measurement value cannot be held. Output only can be held.
If set value is selected for hold, instantaneous O2 correction value is calculated and held based
on the set value.
Range identification contact output cannot be switched even if the range is switched during the
hold.
l Screen display during Holding
The “Hold ON” message blinks on the measuring screen. Since the screen displays the process
of calibration during the manual calibration, “Hold ON” is not displayed even if the output signal is
held, but the screen is displayed with the hold extending time.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-24
<5. SETUP (User Mode>
l How Output Hold works
• Manual calibration
or
to flow gas
to calibrate
then press
Calibration
Hold extending time
Output hold
Hold for a length of the time set at gas flow times
(see 5.4 Auto Calibration Setting)
F0934.ai
• Auto Calibration
Auto calibration start
Auto calibration end
Calibration
Hold extending time
Output hold
Hold for a length of the time set at gas flow times
F0935.ai
• Remote hold
with input
Remote hold input
Output hold
without input
F0936.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-25
<5.SETUP (User Mode)>
(1) Point the cursor to Output Hold and press
User mode Parameter
.
Select Hold ONOFF
Parameter
(2) “ON” or “OFF” is highlighted. Select one with
keys.
(3) Press
to go select “Current” or “Setting”.
Current : Hold the value immediately before the
hold
Setting : Holds the value arbitrarily set.
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
ON Pevious value
ON
05分
0000
Select Hold setting
Parameter
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
ON Setting
ON
05分
0000
(
Parameter
Hold
(4) Press
to go to the screen to set value.
(5) Move cursor with
(component). Press
Select Ch No.
NOX
SO2
CO2
CO
O2
010
020
015
012
022
%FS
%FS
%FS
%FS
%FS
(
)
keys and select a Ch
(6) A value is highlighted and changeable. Enter
parameter in each digit with
keys. Press
to shift to another digit. After changing parameter,
press
.
NOTE
The setting is expressed as 1/1 full scale range for both
respective ranges. When 0 to 1000 ppm is selected as
the range, and 10% FS is selected as hold setting, the
output equivalent to 100 ppm is held irrespective of the
measurement value at that time.
(7) Press
screen.
Ch1
Ch2
Ch3
Ch4
Ch5
)
to return to the parameter setting
Parameter
Hold
Ch1
Ch2
Ch3
Ch4
Ch5
Set Hold value
0 to 100 %FS
NOX
SO2
CO2
CO
O2
010
020
015
012
022
%FS
%FS
%FS
%FS
%FS
End of Hold Setting
Parameter Setting Screen
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.7.2
5-26
<5. SETUP (User Mode>
Response time, Average Period, Backlight, Contrast
n Response time
Parameter
Select Ch No.
Response Time
The response time of the electrical system can be
changed. Setting is available by components.
Ch1
Ch2
Ch3
Ch4
Ch5
CAUTION
Setting value as second (s) here are rough and not exact
ones. The setting value can be modified as requested by
customer.
NOX
SO2
CO2
CO
O2
10
20
15
12
15 s
0020 s
0020 s
0020 s
0020 s
F0940.ai
n Average Period
It allows you to set an average period of the average
value of O2 correction. It enables you to set an average
time of 1 to 59 minutes (1-minute step) or 1 to 4 hours
(1-hour step).
Parameter
Select Ch No
Average Period
Ch8
Ch9
Ch10
Changing the setting resets the average value of O2
correction and O2 average value. (Pressing the ENT
validates the resetting only for components whose setting
was changed.)
AV
CU
AV
CU
AV
CU
NOX
SO2
CO
01 hour
01 hour
50 min
l Example of average action
Suppose the average period is 1 hour.
Reset
Average value
Time
Time
F6-7graph2.ai
• Sampling occurs every 30 seconds.
• Every 30 seconds, the average for last 1 hour (time setting) is output.
• At the instant of resetting, zero is assumed for all past values. It means that the average
value will not be correct for 1 hour after resetting.
l Average value reset
This mode is used to clear all average values O2 correction average and O2 average, and
restarts averaging. With an option code /K, you can reset from external input. The indication
value and output value is 0 ppm, vol% or so at the time of the reset input (based on average
period settings).
With input (hold at least 1.5 sec)
Reset input
without input
With input, resetting lasts.
Shift from “with input” to “without input” to “without input” starts averaging.
F0939.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-27
<5.SETUP (User Mode)>
n Backlight Timer
Backlight of the LCD automatically turns off, whose timer you can setup.
When the specified time elapses during the measurement screen display, the backlight is
automatically turned off. Press any key to reset backlight OFF.
Only when ON is selected, the time to turn off the
to set time with
backlight is displayed. Press
keys. Press
to confirm.
If you select “OFF”, the backlight does not go off.
Select ON or OFF
Parameter
12/01/11 WED 13 : 50
OFF
ON Previous value
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
ON
05min
0000
F0943.ai
n Contrast
Adjust the contrast of LCD. Press
to settle.
adjust. Press
Parameter
keys to
Select ON or OFF
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
ON Previous value
ON
05 min
0000
F0944-2.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5.7.3
5-28
<5. SETUP (User Mode>
Maintenance Mode
In this mode, you can check of sensor input value, display of error log file, set password and
others. Set password first. The password is required next time you enter the mode.
User mode Parameter
Parameter
(1) Select the Maintenance Mode . Press
Select an item
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
Parameter
keys. Press
(2) Change parameter with
go right. After entering value, press
.
12/01/11 WED 13 : 50
OFF
ON Previous Value
ON
05 min
0000
Select an item
to
Default password preset at factory is “0000.” You
can go to Maintenance Mode before you change
the password.
Current Time
Key Lock
Output Hold
Response Time
Average Period
Backlight Timer
Contrast
to Maintenance
Mode
12/01/11 WED 13 : 50
OFF
ON Previous Value
ON
05 min
0000
(
Maintenance
mode
(3) Enter the password to go to Maintenance Mode.
Select an item to set. Move cursor to the item to
set and press
.
1.
2.
3.
4.
5.
6.
)
Select operating item
Sensor Input Value
Error log
Cal.Log
Output Adj.
Other Pramameter
To Factory Mode
(
)
To each Maintenance screen
The following shows sample images of each Maintenance Mode. Press
to return.
CAUTION
“6.Factory Mode” is not for customer use, only for our service engineers
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-29
<5.SETUP (User Mode)>
1. Sensor Input Value
• Input 1 to 4 : NDIR sensor digital value
• Input 5 : O2 sensor digital value
Maintenance
Sensor input
Input 1
Input 2
Input 3
Input 4
Input 5
CAUTION
Customers must not change the setting.
100821
96118
102241
82856
11050
Gas Sample
2. Error Log
Error log traces error history of the latest 14 occurrence. Maintenance ENT :Clear Error log
Error Log
ESC :Bacj
Refer to “8 Error message” to know the contents of
error.
ErrorNo Y
M
D
H
M Ch
To erase all error in a log, press
“Clear Error Log.”
after selecting
No. 4 04
No. 1 04
No. 6 03
No. 9 03
No. 5 03
No. 9 03
Next Page
2
1
12
12
12
12
11
10
1
1
1
1
18
12
10
10
0
0
10
2
10
10
0
0
5
1
2
2
2
2
Page1
Clear Error Log
3. Calibration Log
Calibration Log traces past calibration history. The history data include sensor input value,
concentration value, and the date when zero / span calibration was performed.
Each Ch(component) records latest 10 calibration
data.
R :
Z
1 : Zero calibration (Z) of Range 1
S1 : Span calibration (S) of Range 1
Z2 : Zero calibration (Z) of Range 2
S2 : Span calibration (S) of Range 2
Cnt :
alue of measuring detector at the time of
V
calibration
Con :
Maintenance
Cal.Log
Ch1
Ch2
Ch3
Ch4
Ch5
Select Ch No
NOX
SO2
CO2
CO
O2
Clear Error Log
oncentration value displayed before
C
calibration
Move the cursor to select Clear Calibration Log and
to erase all of the calibration history.
press
Maintenance
Cal.Log
Ch1 NOx
R
Z1
S1
Cnt
485231
441764
Con
-0.2
189.5
YDHM
12111810
12111815
F0947.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-30
<5. SETUP (User Mode>
4. Output adjustment
Connect the digital multi- meter to the output terminal
corresponding to the number of OUT to be adjusted,
and adjust the value so that 4mA or 0V is output at zero
and 20mA or 1V is output at span.
Move cursor with
keys to select OUTNo. or
. The value you select is
ZERO / SPAN, and press
highlighted. Adjust the numerical value with
keys
to shift to another
according to output value. Press
digit. After the adjustment press
.
5. Other parameter
Password Set : Set password to move from the
parameter setting screen to the
maintenance mode. Arbitrary 4-digit
number can be selected.
O2 ref.value : Set the oxygen concentration reference
value at the time of O2 correction
calculation. Settable in the range from
00 to 19%.(Initial value : 12%)
Maintenance
Output Adj.
OUT
1
2
3
4
5
6
Zero
01245
01245
01245
01245
01245
01245
Maintenance
Output Adj.
OUT
1
2
3
4
5
6
Zero
01245
01245
01245
01245
01245
01245
Adjust OUTPUT
ZERO and SPAN
Span
11845
11845
11845
11845
11845
11845
OUT
7
8
9
10
11
12
Zero
01900
01900
01900
01900
01900
01900
Span
12500
12500
12500
12500
12500
12500
Adjust OUTPUT
ZERO and SPAN
Span
11845
11845
11845
11845
11845
11845
OUT
7
8
9
10
11
12
Zero
01900
01900
01900
01900
01900
01900
Span
12500
12500
12500
12500
12500
12500
Limit : S
et the O2 concentration limit at the time of O2 correction calculation. Settable in the
range from 01 to 20%. (initial value : 20%)
* See 4.3 Mode and screens for O2 correction concentration value to calculate O2 correction. This function is not available on
IR250-B. .
Station No. : Set Station No. in MODBUS.
Settable parameter goes 00 to
32.
Range setting : Moves to the screen on which
measuring range is changed.
Set Sample Switching : Moves to the screen on which
parameters about Sample
switching is changed. (Only
IR250-C displays)
Maintenance
mode
setting
Select an item
Password Set 2465
O2 ref.value 12% O2 Limit 20% O2
Station No.01
Range setting
Set Sample Switching
F0952.ai
keys to select an item to
Move cursor with
modify. Password, O2 ref. Value, Limit, Station No. are inverted.
Move
keys to change parameter. Change the parameter and press
CAUTION
Keep the password in safe and do not forget. Otherwise, you cannot enter the maintenance
mode. In case of lost of the password, contact us immediately.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-31
<5.SETUP (User Mode)>
l How to set / change range
The measuring range can be arbitrarily selected in the minimum and the maximum range
specified at the time of purchase. The range to be used can be selected 1 or 2.
(1) Move cursor with
keys. Go to Range setting and press
.
Select Ch No.
Maintenance
Mode
Range set
keys to select Ch
(2) Move cursor with
(component) to modify. Press
.
Ch1
Ch2
Ch3
Ch4
Ch5
NOX
SO2
CO2
CO
O2
(
(3) Select the value to change and press
.
• Settable range
The value for range 1 and range 2 must fall within the
range from the MIN and the MAX range (including
the MIN and the MAX range), and at the same
time range 1 must be smaller than range 2. The
number of ranges is 1 or 2.
keys. Press
(4) Change the value with
to shift to another digit. Unit cannot be changed.
When decimal point is highlighted, the decimal
point can move.
(5) After completion of the setting, press
.
CAUTION
You must conduct Zero / Span calibration each
time you change the range setting. Otherwise, the
measurement result may not be correct in value.
Maintenance
range set
Ch1 NOx
MIN range
Range 1
Range 2
MAX range
Range num.
)
Select range or
range num.
100.0
500.0
1000.
2000.
2
ppm
ppm
ppm
ppm
(
)
Maintenance
range set
Ch1 NOx
MIN range
Range 1
Range 2
MAX range
Range num.
100.0
500.0
1000.
2000.
2
ppm
ppm
ppm
ppm
End of range setting/change
F0954.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-32
<5. SETUP (User Mode>
l Sample switch setting
Only available for IR250-C. The following pages describe how to set gas flow time, interference
compensation coefficient. These values are required for sample switch.
Flow time of reference gas and sample gas are switched by sample switch method at the
following timing.
Measuring time of
reference gas
Flow time of
reference gas
Measuring time of
reference gas
Flow time of
sample gas
Measuring time of
reference gas
Flow time of
reference gas
Measuring time of
reference gas
Flow time of
sample gas
close
DO26contact
open
close
DO27contact
open
measuring
position
1 cycle
F09S4.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-33
<5.SETUP (User Mode)>
(1) How to change the setting of Gas Flow time
CAUTION
Gas flow time is essential value for sample switch. DO NOT CHANGE unnecessarily.
Here describes how to set up Reference Gas Measuring Time, Sample Gas Switching Time,
Sample Gas Flow Time, Sample Gas Measuring Time.
User mode Maint.Mode Sample Switching
(1) Move cursor to Gas Flow Time and Press
.
Maint. Mode
Sample Switch
setting
select an item
Gas Flow Time
Interference Compensation
Coefficient
(2) Move cursor with
keys to select
Reference Gas Switching Time or Reference
Gas Measuring time.
Reference Gas Switching Time
(initial value : 5 sec.)
:1 to 30s.
Reference Gas Measuring Time
(initial value : 5 sec.)
:1to 60s.
Maint.Mode
Sample Switch
Gas Flow time
Select an item
Ref.gas Switching t 05 s
Ref.gas Measuring t 05 s
Ref.gas Flow time 10 s
Sample Switching 05 s
Sample Measuring 05 s
Sample Flow time 10 s
(
(3) Press
to settle each value. Reference
Gas Flow Time, Sample Gas Switching Time,
Sample Gas Measuring Time, Sample Gas Flow
Time are calculated according to the following
relational expression.
Switching time of sample gas (t1) =
Switching time of reference gas (t3)
)
Set gas replacement time.
Maint.Mode
Sample Switch Setting range is from 1 to
Gas Flow time 30 seconds
Ref.gas Switching t
Ref.gas Measuring
Ref.gas Flow time
Sample Switching
Sample Measuring
Sample Flow time
05 s
05 s
10 s
05 s
05 s
10 s
Flow time of sample gas (Ts) = Flow time of reference gas (Tr)
Flow time of reference gas (Tr) =
Switching time of reference gas (t3) + measuring time of reference gas (t4)
Flow time of sample gas (Ts) =
Switching time of sample gas (t1) + measuring time of sample gas (t2)
1 cycle = flow time of reference gas (Tr) + flow time of sample gas (Ts)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
5-34
<5. SETUP (User Mode>
(2) Interference compensation coefficient (Measuring / Calibration)
(1) Move cursor to Interference Compensation
Coefficient and press
.
(2) Move cursor with
keys to select Ch
(component) to modify. Press
to change the
value.
(3) After changing the value,press
value.
to settle the
User mode Maint.Mode Sample Switching
Maint. Mode
Sample Switch
setting
select an item
Gas Flow Time
Interference Compensation
Coefficient
(
)
Maint. Mode
Select the component to
Sample Switch be adjusted.
Interference Adj.
Ch1 NO
Ch2 SO2
Ch3 CO
Ch4 CO2
44445
0
44445
44445
IM 11G02Q01-01EN
1.000000
1.000000
1.000000
1.000000
3rd Edition : May 31, 2017-00
6.
6-1
<6. CALIBRATION>
CALIBRATION
This chapter describes manual calibration.
6.1
Zero Calibration
Zero Calibration is for zero point adjustment. Proper zero gas, suitable for the application, should
be used. Refer to Section 3.4 “Sampling”
(1) On the Measurement screen, press
to show
Manual Zero Calibration screen.
Select Ch No. with
UP/DOWN and ENT
Back with ESC
ZERO Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
keys to select Ch
(2) Move cursor with
(component) to calibrate. Press
then the zero
gas will be supplied.
CAUTION
For the Ch (components) in which “at once” is set in the
zero calibration range setting (5.2.2), zero calibration is
carried out at once. If “both” is set in the calibration range
(5.2.3), both ranges are calibrated simultaneously.
(3) Wait until the indication is stabilized with the
zero gas supplied. After the indication has been
key. Zero calibration in
stabilized, press the
range selected by the cursor is carried out
l How to close Zero Calibration
To close the Zero Calibration or cancel the operation
midway, press
. A previous screen will return.
CAUTION
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
ZERO Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
0.0
0. 00
0.0
20. 09
)
Select Ch No. with
UP/DOWN and ENT
Back with ESC
ZERO Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
0.0
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0.0
0.0
0. 00
0.0
20. 09
ENT : Go on calibration of
selected CH.
ESC : Not calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0.0
0.9
0. 34
1.1
20. 09
to Measuring after Manual Zero Cali.
F0955.ai
For the Ch (component) for which “AR” is selected in
5.1, the cursor automatically moves to the range selected in “5.2.4 Setting of auto calibration
component / range” , and calibration is carried out within that range.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
6.2
6-2
<6. CALIBRATION>
Span Calibration
Span Calibration is used to perform a span point adjustment. Supply calibration gas with
concentration set to the span value to perform the span calibration. For the span calibration gas
for the NOx, SO2, CO2, CO, CH4measurement, use the standard gas with a concentration of 90
to 100% of its measuring range value. For the span calibration gas for the O2 measurement, use
the standard gas with a concentration of 90 to 100% of its measuring range value when
measuring with the built-in O2 sensor, and use the standard gas of 1 to 2 vol% when measuring
with an external zirconia O2 sensor.
SPAN Cal.
(1) On the Measurement screen press
the Manual Span Calibration screen.
to display
(2) Move cursor with
keys to select Ch
(component) to calibrate. Press
then the span
gas will be supplied.
CAUTION
For the Ch (components) in which “both” is set in 5.2.3
Setting of calibration range, span calibration is carried
out on both ranges.
(3) Wait until the indication is stabilized with the
span gas supplied. After the indication has been
key. Span calibration in
stabilized, press the
range selected by the cursor is carried out
l How to close Span Calibration
TO close the Span Calibration or cancel the operation
. A previous screen will return.
midway, press
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
Select Ch No. with
UP/DOWN and ENT
Back with ESC
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
(
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
SPAN Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
0.0
0. 00
0.0
00. 00
)
Select Ch No. with
UP/DOWN and ENT
Back with ESC
SPAN Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
0.0
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
0.0
0.0
0. 00
0.0
00. 00
ENT : Go on calibration of
selected CH.
ESC : Not calibration
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
Range1
Range2
0–50.00ppm
0–500.0ppm
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–20.00vol%
0–50.00ppm
0–500.0ppm
0–10.00vol%
0–25.00vol%
98.0
0.0
0. 00
0.0
00. 00
CAUTION
For the Ch (component) for which “AR” is selected
in “Section 5.1 Changeover of range” the cursor
automatically moves to the range selected in “5.2.4
Setting of auto calibration component / range”, and
calibration is carried out within that range.
to Measuring after Manual SPAN Cali.
F0602.ai
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7.
7-1
<7. MAINTENANCE>
MAINTENANCE
WARNING
Purge not only inside of IR250 but all measuring gas lines with zero gas sufficiently, when you
provide maintenance or inspection on IR250 with its cover or door open. Otherwise it may cause
hazardous accidents such as gas leakage, fire and explosion.
CAUTION
DO NOT operate IR250 with its cover left open for a long period of time. It may have dust and
dirt, which damages the equipment.
7.1
Daily check and maintenance procedures
Maintenance and check list
Parts to be checked
Indication Value
Phenomena
Indication values are lower. Dust contamination
Indication values are higher. in sampling cell
Daily
Weekly
Yearly
Causes
Remedy
Consult us.Clean
sampling cell and
inspect sampling device,
especially gas filter.
Air is absorbed
midway in the
sampling tube.
Check for any leak of the
sampling line and repair
if any.
Sample gas flow
rate and purge gas
flow rate (if purging
takes place in the
instrument).
Standard flow rate is not
within the range of the
specified flow rate.
—
Adjust the flowmeter with
flow rater needle valve.
Zero point of gas
analyzer
Deviation from zero point
—
Zero adjustment
Span point of gas
analyzer
Deviation from span point
—
Span adjustment
Gas analyzer
—
—
Overhaul or service in
accordance with proper
service
● Flow rate check of sampling gas or purge gas.
Flow rate of sampling gas must be within 1.0 ± 0.5 L/min.
Flow rate of sampling gas must be within 0.5 ± 0.2 L/min.
Flow rate of sampling gas or reference gas must be within 1.0 ± 0.2 L/min.
Flow rate of purge gas must be approximately 1 L/min.
Maintenance and inspection should be carried out every day, if required.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7-2
<7. MAINTENANCE>
● Zero calibration and span calibration
For performing a zero calibration, refer to 6.1, Zero calibration. After the zero calibration, perform
a span calibration by referring to 6.2 Span calibration.
Zero calibration and span calibration should be performed once a week, if required.
n Long term maintenance
Create a long-term maintenance component procurement plan in advance based on the gas
analyzer annual inspection plan described below.
Each period described below is standard value, thus each one may vary depending on
measuring gas conditions or operating environment. The recommended replacement period is
not the warranty period. Consult us for the maintenance service in detail.
Recommended replacement period
Components
Q’ty
Recommended
replacement period (year)
IR light source unit
1 or 2
5
Varies with conditions of measuring gases or their
range.
O-ring for sampling cell
2 to 5
2
Varies with conditions of measuring gases or their
range.
Detector
1 or 2
8
Varies with conditions of measuring gases or their
range.
LCD
1
3
Solvent valve
2
3
only for IR250-C
AC/DC main power
1
5
IR250-B need 3 pieces
DC/DC power PCB
2
5
Annual inspection
—
1
7.2
Notes
Replacement of power fuse
CAUTION
Prior to the following work, be sure to repair blown down fuse (short, etc.), if any.
The part number of fuse is K9642CJ.
(1) Turn “OFF” the main power switch of the analyzer. Refer to “1.1 Names and description ” to
know where each switch or fuse is.
(2) Turn the fuse holder cap (shown in the figure above) counterclockwise and pull it out.
The cap will be removed. Remove a fuse out of the holder. Replace it with a new one.
(250VAC/2A, Time-lag type).
(3) Reinstall the fuse holder cap, turn ON the power supply switch. Confirm the analyzer starts
up normally.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7.3
7-3
<7. MAINTENANCE>
Cleaning of sampling cell
Entry of dust or water drops in the sampling cell contaminates the interior of the cell, thus
resulting in a drift. Clean the inside if dirty. Then, check the sampling device, especially the filter,
to prevent the cell from being contaminated by dust or mist.
CAUTION
Service personnel from YOKOGAWA will take care of the maintenance. Contact YOKOGAWA
representatives for maintenance service.
7.3.1
Disassembly and assembly of sampling cell (Pipe cell, Block cell)
There are two kinds of sampling cells, block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and
pipe cells (Cell length: 64 mm, 125 mm, 200 mm and 250 mm).
2-component analyzer may incorporate both sampling cells in one unit. In such a case, detach
the pipe cell and then block cell
CAUTION
Special care should be taken when assembling or disassembling the measuring cell to avoid
the application of force to the detector pipe or light source unit pipe. If the pipe is deformed or
damaged by excessive force, there is a danger of gas leak, thus resulting in faulty operation.
l How to remove pipe cell
(1) Stop measured gas. If it is harmful, purge the pipe cell thoroughly with zero gas.
(2) Turn OFF the power switch.
(3) Remove the screws then remove the cover.(Open the front door on IR250-B).
(4) Remove the tubes connected to the pipe cell.
(5) See Figure. 7.1. Loosen and remove screws (Screw A) from the cell retainer fastening the
pipe cell (both ends).
(6) Remove the cell from the measuring unit and unscrew the infrared transmission window at
both ends in the right direction.Don’t miss the O-ring.
(7) For assembly, reverse the disassembly procedure.Make sure to put the space in 0.5mm
between light source unit and the pipe cell, the pipecell and the detector. Screw the pipe part
of gas inlet with window on the both side to the measuring cell. Screw it so that tubing (both
end of pipe cell) and the pipe cell align facing up. (Marking of pipe cell faces up.)
Screw A
Pipe cell
O-ring
Cell retainer
(Infrared transmission)
Window
Detector
Light source unit
Figure 7.1 Sample of Measuring cell( Pipe cell)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7-4
<7. MAINTENANCE>
l How to remove block cell
Block cell is attached with the detector. Remove them together as a whole unit out of the base plate.
(1) Stop measured gas. If it is harmful, purge the pipe cell thoroughly with zero gas.
(2) Turn OFF the power switch.
(3) Remove the screws then remove the cover.(Open the front door on IR250-B).
(4) See Figure 7.2. Remove the tube connected to the block cell.
(5) Unscrew 6 screws holding the power board (Screw B) and remove the measuring cell. DO
NOT pull it off forcibly, because some cables are still connected.
(6) Unscrew the screws (Screw C) on the back of the base plate, holding the detector. The
number of detector varies with the specification. One detector is fixed with two screws.
(7) Unscrew the two screws (Screw D) holding the detector to the light source unit.
(8) Unscrew the two screws (Screw E) holding the detector to the light souce unit and the block
cell. Handle the unit horizontally. Otherwise all of the parts may slip off.
Remove the detector and the block cell together. Then take away the block cell.
Be careful not to miss the small parts. The (infrared transmission) window is easy to fall (not
fixed) between the detector and the block cell. Don’t miss the O-ring, either.
(9) For assembly, reverse the disassembly procedures.
Place the O-ring between the window and the block cell. Make sure the position is correct.
If the analyzer has two detectors, install them so that no space is left in-between. Connect
cables to the detector correctly.
Screw E
Block cell
O-ring
(Infrared transmission)
Window
Detector
Light source
unit
Screw D
Screw B
Base plate
Screw C
Light source
power board
Figure 7.2 Sample of Measuring cell (Block cell)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7-5
<7. MAINTENANCE>
l How to remove pipe cell and block cell
First remove pipe cell then remove block cell. Block cell is attached to the detector, so you need
to take the whole cell out of the base plate.
(1) Follow step on “How to remove pipe cell,” from (1) to (6) to remove the pipe cell. See Figure
7.3.
(7) For assembly, reverse the disassembly procedures.
(8) See Fig. 7.3.Unscrew the two screws (Screw B) holding the base plate, and take away the
measuring cell. DO NOT pull it off forcibly, because some cables are still connected.
(9) Unscrew the screws (Screw C) on the back of the base plate, holding the detector.
(10) Unscrew the two screws (Screw D) holding the detector to the light source unit.
(11) Unscrew the two screws (Screw E) holding the detector to the light souce unit and the block
cell. Handle the unit horizontally. Otherwise all of the parts may slip off.
Remove the detector and the block cell together. Then take away the block cell.
Be careful not to miss the small parts. The (infrared transmission) window is easy to fall (not
fixed) between the detector and the block cell. Don’t miss the O-ring, either.
(12) For assembly, reverse the disassembly procedures.
Place the O-ring between the window and the block cell. Make sure the position is correct.
Make sure to put the space in 0.5mm between light source unit and the pipe cell, the pipe
cell and the detector. Screw the pipe part of gas inlet with window on the both side to the
measuring cell. Screw it so that tubing (both end of pipe cell) and the pipe cell align facing
up. (Marking of pipe cell faces up.)
Screw E
Block Cell
(Infrared transmission)
Window
Detector
Light source
unit
O-ring
Pipe cell
Screw A
Cell retainer
Screw D
O-ring
(Infrared transmission)
Window
Detector
Screw B
Base plate
Screw C
Figure 7.3
Light source
power board
Sample of Measuring cell (Pipe cell and Block cell)
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
7.3.2
7-6
<7. MAINTENANCE>
How to clean sampling cell
CAUTION
Handle the fragile window with care. Use care not to rub off the dirt from the window roughly.
• To clean the sampling cell inside or infrared ray transmission window, first clear large dirt of
it with a soft brush and then wipe lightly with soft cloth. Do not use abrasive or paper cloth.
• If the window or the sampling cell interior is very dirty, use a soft lint-free cloth moistened
with absolute alcohol.
• If the window is corroded, rub off the scale from the window lightly with a soft cloth to which
chrome oxide powder is applied. If it is excessively corroded, it should be replaced with new
one.
• When the sampling cell or window cleaning is completed, assemble according to the
sampling cell disassembly and assembly procedures. Assemble the pipe carefully. If it
becomes bent or damaged, replace it with a new part.
• Do not wash the sample cell components with water.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
8.
8-1
<8.ERRROR MESSAGE>
ERROR MESSAGE
Error display
Error contents
Probable causes
Error No.1
Light source or motor rotation is
faulty.
• Infrared ray light source is faulty.
• Sector motor rotation is faulty or stopped.
• Amplifier circuit is faulty.
Error No.2
Detector failure.
• Detector voltage circuit is faulty.
• Detector element is broken or faulty.
• Amplifier circuit is faulty
Error No.3
A/D error
• A/D conversion circuit is faulty.
Error No.4
Zero calibration is not within the
allowable range.
Error No.5
A amount of zero calibration
(indication value) is over 50% of full
scale.
• Zero gas is not supplied.
• Zero point is deflected much due to dirty cell.
• Detector is faulty.
Error No.6
Span calibration is not within the
allowable range.
Error No.7
A amount of span calibration
(difference between indication
value and calibrated concentration)
is over 50% of full scale.
Error No.8
Measured values fluctuate too
much during zero and span
calibration.
• Calibration gas is not supplied.
• Time for supplying calibration gas is short.
Error No.9
Calibration is abnormal during auto
calibration.
• Error corresponding to No. 4 to No. 8 occurred during
auto calibration.
Error No.10
Output cable connection is
improper.
• DIO circuit is failure.
• Internal wiring to the DIO circuit is broken.
• Gas is not supplied long enough.
• Span gas is not supplied.
• Calibrated concentration setting does not match
cylinder concentration.
• Zero calibration is not performed properly.
• Span is deflected much due to dirty cell.
• Detector sensitivity is deteriorated.
When one of those errors among No.1 to No.3 or No.10 occurs, instrument error (FAULT) contact output is closed.
When one of those errors among No.4 to No.9 occurs, calibration error contact output is closed.
n Trouble shooting
When Error No. 1 occurs,
remove the top cover of the analyzer and check the LED on the light source power PCB. If
the LCD light is turned off, light source is disconnected.
When Error NO.1 to No.3 or No.10 occurs,
the analyzer is faulty. Contact us immediately.
When Error No.4 to No.8 occurs,
the calibration procedure may be wrong. Check the following again. If the error is still
indicated, contact us immediately.
• Is the calibration gas supplied in the analyzer?
• Does the calibration operation match the supplied gas? (For example, zero calibration is
performed while flowing the span gas by mistake.)
• Does the supplied gas concentration match the gas concentration set at the calibration
concentration setting?[
Also, when Error No.5 or NO.7 occurs,
you can perform calibration forcibly by following the procedure described in the next section.
Use it as fault recovery when calibration fails and calibration contents are missed.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
8-2
<8.ERRROR MESSAGE>
n How to operate when error appears
l Error No.1 to No.4, No.6, No.8 to No.10
Measurement screen
NONo.9
2
Error
0-25
Ch
Ch
SO2
Ch
CO2
Ch
CO
Ch
O2
0-200
0-10
0-200
0-25
0 0.8
1 3.6
0.0 0
0.0
2 1.0 0
Error Contents display
Auto Cal. error
ESC: Back to MEAS
Error No.9
ppm
ppm
Zero NOx Calibration Error
Cause
vol%
ppm
● Calibration gas is not flowing.
● Gas flowing time is short.
vol%
●
delete error display.
● Even after pressing
if trouble shooting
is not completed, error is displayed again.
● Setting concent. is different from gas concent.
● Lower sensitivity due to sample cell contamination.
● When more than two errors occur at a time,
press
to move to another error display.
l Error No. 5 or No. 7
ZERO Cal.
Ch1
NOX
Ch2
SO2
Ch3
CO2
Ch4
CO
Ch5
O2
● Press
ENT : Go on calibration of
selected Ch.
ESC : Not calibration
レンジ1
0–200 ppm
Error No.5
レンジ2
0–2000ppm
Range1 0–200 ppm
Range2 0–2000ppm
Range1 0–10 vol%
Range2 0–20 vol%
Range1 0–200 ppm
Range2 0–1000ppm
Range1 0–10 vol%
Range2 0–25 vol%
Zero cali. more than 50%FS
ENT : Force Cali.
ESC : Stop cal. and
back to MEAS.
NOx Calibration error
Cause
● Zero gas is not flowing.
● Deviation of zero point due to contamination.
● Low sensitivity of detector
Error No.5
3083
-13.6
-0.006
0.2
-0.09
to delete the error display.
Calibrated
forcedly
Calibration is continued. Unless another
calibration error occurs, calibration is carried out to the
end, and the Measurement screen returns.
Ch
NOx
Ch
SO2
Ch
CO2
Ch
CO
Ch
O2
0-25
0-200
0-10
0-200
0-25
IM 11G02Q01-01EN
9 0.8
1 3.6
0.0 0
0.0
0.0 9
ppm
ppm
vol%
ppm
vol%
3rd Edition : May 31, 2017-00
8-3
<8.ERRROR MESSAGE>
n Error log file
If an error occurs, the error history is saved in an error log file. You can find the error log file in the
maintenance mode.
l Error log screen
Maintenance
Mode
Error Log
Errors that
occurred
ErrorNo. Y
No. 10 16
No. 9 16
No. 5 15
No. 9 15
No. 7 15
No. 7 15
Next Page
Date and time when an
error occurred
ENT :Clear Error Log
ESC:Back
M
2
1
12
12
12
11
D
11
10
1
1
1
2
T
18
12
10
10
0
2
M
10
2
10
10
0
43
Component for which
the error occurred.
Ch
5
5
1
1
2
Page1
New
Old
Clear Error Log
* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every time a new error occurs.
* If the power supply is turned OFF, the contents in the error log file will not be lost or damaged. They will be saved.
l Deletion of error history
Press the ENTER key on the Error Log screen (looks like the above one), and the “Error Log
Clear” will be inverted. Further pressing the key will clear the error history.
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Blank Page
i
Revision Information
Model IR250 Infrared Gas Analyzer
Title:
Manual No.: IM 11G02Q01-01EN
Edition Date
Remark (s)
3rd
2nd
1st
Revised p.1-2, 1-5, 2-2, 2-8, 2-9, 2-13, 2-16, 3-4, 3-8
Modified in p.2-13, pp.3-9 to 3-12
Newly published
May. 2017
Feb. 2017
Feb. 2017
Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
http://www.yokogawa.com/an/
IM 11G02Q01-01EN
3rd Edition : May 31, 2017-00
Blank Page